Design of a Rotational Stability Measurement Device
For Analysis of ACL Reconstruction
University of PittsburghSenior Design – BioE 1160/1161
Stephanie BechtoldKatie Dillon
Kara Wagner
Mentor: Thore Zantop, M.D.
April 18, 2005
![Page 2: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/2.jpg)
Goal
To develop and test a novel device to measure the rotational stability of the knee after an Anterior Cruciate Ligament (ACL) reconstruction
![Page 3: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/3.jpg)
ACL anatomy
• Two distinct fiber bundles
• Anteromedial (AM)
• Translational stability
• Posterolateral (PL)
• Rotational stability
www.aclsolutions.com/ images
![Page 4: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/4.jpg)
Current Reconstruction Methods
• Single bundle
• Restore only the AM bundle
• Susceptible to reinjuries by pivoting
• Double bundle
• Restores both AM and PL bundles
• More anatomically correct
![Page 5: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/5.jpg)
Devices to Evaluate ACL Reconstruction
• Current devices only measure translational stability of the knee
• Double bundle technique creates need for device to measure rotational stability
• Evaluate effectiveness of PL bundle reconstruction
http://www.medmetric.com/kt1.htm http://www.aircast.com
![Page 6: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/6.jpg)
Problem Statement
• A device is needed to measure the rotational stability of the knee
• Comprehensive analysis of reconstruction techniques
• An immobilization device is needed to comfortably restrict movement in the hip and ankle joints
• Ensure pure rotation of the knee is measured
![Page 7: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/7.jpg)
Market Considerations
• Market size
• 16,000 Orthopedic Surgeons (AAOS)
• Predicate Devices• KT1000: $3900
• Rolimeter: $850
![Page 8: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/8.jpg)
• Adjust boot to keep ankle immobile
• Fix knee at various flexion angles (0, 30, 45, 60, 90 degrees)
• Comfortably immobilize hip
• Consistently and safely apply known moment
• Collect repeatable data
Initial Design Considerations
![Page 9: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/9.jpg)
• Universal force-moment sensor (JR3, Woodland, California)
Prototype Development- Boot• Aircast® Pneumatic Walker Brace• Nest of Birds Ascension Technology Corporation, Burlington, VT/USA
Photos courtesy of Ferguson Lab
![Page 10: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/10.jpg)
Adjust boot to keep ankle immobile
• Fix knee at various flexion angles (0, 30, 45, 60, 90 degrees)
• Comfortably immobilize hip
Consistently apply known moment
• Collect repeatable data
Initial Design Considerations
![Page 11: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/11.jpg)
Prototype Development- Hip Brace
• Lateral Decubitus position
• Wheelchair leg rest
![Page 12: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/12.jpg)
Adjust boot to keep ankle immobile
Fix knee at various flexion angles (0, 30, 45, 60, 90 degrees)
• Comfortably immobilize hip
Consistently and safely apply known moment
• Collect repeatable data
Initial Design Considerations
![Page 13: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/13.jpg)
Prototype development
• Redesign of initial prototype • Patient placed in supine position
• Lower leg horizontal
• Adjustable leg rest
• Knee flexion angle
• Length of femur
• Minimize metal components
![Page 14: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/14.jpg)
Prototype Fabrication
![Page 15: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/15.jpg)
Materials Selection
• Constructed of acrylic
• Eliminates metal components
• Minimizes interference with magnetic sensors
• Durable
• Comfortable for patient
![Page 16: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/16.jpg)
Evaluation of Device
• 4 subjects - at 3 flexion angles
• 5 trials each
• Ensure repeatability
• Goal: Range of motion within ±1°
• Subject reports stability of ankle and hip joints
![Page 17: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/17.jpg)
Experimental Methods
• Subjective knee evaluation performed by a clinician to determine health of knee
• Subject fitted with boot and brace to comfort
• Nest of Birds (NOB) sensors placed on
• Proximal Tibia
• Distal Femur
• Front of Boot
![Page 18: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/18.jpg)
Experimental Methods
• Lower leg leveled at horizontal
• Creates neutral start position
• Moment applied by clinician (10Nm)
• Range of motion recorded by Nest of Birds
![Page 19: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/19.jpg)
Experimental Methods
![Page 20: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/20.jpg)
Sample Results
![Page 21: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/21.jpg)
Preliminary Results
• Range of Motion, in degrees
• 4 subjects, 5 trials each
FlexionAngle Subject 1 Subject 2 Subject 3 Subject 4
0 degrees 40.75 ± 1.7 62.20 ± 1.4 43.40 ± 1.5 64.40 ± 1.9
45 degrees 49.60 ± 1.8 52.00 ± 1.0 31.60 ± 0.9 57.00 ± 0.7
90 degrees 37.00 ± 1.6 37.00 ± 0.7 27.20 ± 1.5 54.20 ± 1.8
![Page 22: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/22.jpg)
Adjust boot to keep ankle immobile
Fix knee at various flexion angles (0, 30, 45, 60, 90 degrees)
Comfortably immobilize hip
Consistently and safely apply known moment
Collect repeatable data
Initial Design Considerations
![Page 23: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/23.jpg)
Discussion
• Range of motion was repeatable within ±2° for all subjects
• Subjects reported ease of use and comfort of the boot and brace
• Overall apparatus is heavy for operator
![Page 24: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/24.jpg)
Competitive Analysis
Our Device• Strengths
• More comprehensive analysis of ACL reconstruction
• Weaknesses• Complex design• Higher cost
Predicate Devices• Strengths
• Lightweight• Simple design
• Weaknesses• Limited analysis of
reconstruction techniques
• Race for “first to market”• Competition to develop similar device
![Page 25: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/25.jpg)
Constraints limiting Phase I testing
• IRB approval
• Create baseline data for normal subjects
• Overhead cost of measurement devices
![Page 26: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/26.jpg)
Quality System Considerations
• Class I device: non-invasive
• Human factors• Biocompatibility
• Ease of use for clinician
• Patient’s Comfort
• Ethical issues• Broad range of normal subjects
• Reduce cost
![Page 27: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/27.jpg)
Future• Streamline design
• Minimize cost• Lower weight of apparatus
• Testing on reconstruction patients• Evaluate effectiveness of technique
• Possible combination device based on KT1000 or Rolimeter
• To measure both translational and rotational
stability in the same device
![Page 28: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/28.jpg)
Acknowledgements
• Kevin Bell M.S.
• Volker Musahl M.D.
• Ryan Costic M.S.
• Larry Herman
• Department of Bioengineering
• Drs. Hal Wrigley and Linda Baker
![Page 29: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/29.jpg)
IKDC form
• Standard, subjective knee evaluation
• Patient• History of injury
• Symptoms
• Activity Level
• Clinician• Translational stability evaluated using
KT1000 or rolimeter
• Subjective evaluation of rotational stability
![Page 30: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/30.jpg)
Background
• 100,000 ACL reconstructions per year
• 6th most common orthopedic procedure in the US
• Debate over best method• Single vs. double bundle
• Complex anatomy• AM bundle - translation• PL bundle - rotation
Courtesy of Ferguson Lab
![Page 31: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/31.jpg)
Work Breakdown
• Kara Wagner• Background research, contact with
machine shop
• Katie Dillon• Testing
• Stephanie Bechtold• Solidworks prototype
• Everyone• Design History File, SBIR
![Page 32: Design of a Rotational Stability Measurement Device For Analysis of ACL Reconstruction University of Pittsburgh Senior Design – BioE 1160/1161 Stephanie](https://reader036.vdocuments.us/reader036/viewer/2022062410/56649ec45503460f94bce157/html5/thumbnails/32.jpg)
Milestones
• Initial meeting with mentor – October 2004
• Initial draft of design history file – December 8, 2004
• Initial draft of SBIR – December 17, 2004
• Ordered materials – First week of March 2005
• Completed Solidworks – Third week of March
• Prototype Completed – Last week of March
• Preliminary testing – First week of April