Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF ACETABULAR COMPONENTS OF THRs IN DIGITAL STANDARD RADIOGRAPHS PETER MAYRHOFER, revisited April 2012

Upload: herbert-hopkins

Post on 16-Dec-2015

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

EBRA-CUP:

A METHODS FOR MIGRATION MEASUREMENT OF ACETABULAR COMPONENTS OF THRs IN DIGITAL STANDARD RADIOGRAPHS

PETER MAYRHOFER, revisited April 2012

Page 2: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

EBRA INSTALLATIONSOUTSIDE EUROPE• San Francisco, US

• Los Angeles, US

• New Orleans, US

• Memphis Arlington, US

• New York, US

• Cleveland Ohio, US

• Ottawa, Canada

• Montreal, Canada

• Tokyo (2), Japan

• Osaka, Japan

• Adelaide, Australia

• Melbourne, Australia

• Shanghai, China

EUROPE• Belgium (1)

• Austria (8)

• Belgium (1)

• Denmark (5)

• France (5)

• Germany (23)

• Greece (3)

• Italy (2)

• Netherlands (1)

• Norway (1)

• Sweden (2)

• Switzerland (5)

• United Kingdom (13)

April 2012:

84 Installations

EBRA has become an International Standard.

Page 3: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

EARLY MIGRATION > 1mm IS PREDICTIVE FOR PROSTHETIC LOOSENING!

J Bone Joint Surg Br. 1996 May;78(3):422-26. Early migration predicts late aseptic failure of hip sockets.Krismer M, Stockl B, Fischer M, Bauer R, Mayrhofer P, Ogon M.

J Bone Joint Surg Br. 1999 Mar;81(2):273-80. The prediction of failure of the stem in THR by measurement of early migration using EBRA-FCA. Einzel-Bild-Roentgen-Analyse-femoral component analysis.Krismer M, Biedermann R, Stockl B, Fischer M, Bauer R, Haid C.

EBRA ENABLES QUALITY CONTROL FOR

IMPLANT DESIGN & SURGICAL TECHNIQUE

Page 4: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

Migration is only measurable in serial radiographs of the same patient whose position (exterior orientation) is never exactly the same.

time

mm

EBRA requires at least 4 x-rays per series!

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 5: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EBRA-CUP REQUIRES FULL SIZED AP PELVIS RADIOGRAPHS OF SIZE 430 x 350 mm

Resolution min. 75dpi

digitized with film scanner

or

retrieved from PACS system including a scale for calibration!

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 6: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EBRA-FCA ALSO ACCEPTS HIP RADIOGRAPHS NOT SHOWING THE WHOLE PELVIS.

Resolution min. 75dpi

As the image is calibrated by the true head size, no scale is needed.

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Tip of stem must be visible!

Page 7: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EBRA RSA

• 1 standard x-ray per observation• no ball markers• retrospective studies

• 2 special x-rays per observation• markers required• prospective studies

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 8: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EBRA-CUP PARAMETERS

x

zy

cup inclination& anteversion

head & cup

x-migrationy-migration

z-migration not possible

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 9: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EXACT INTRINSICORIENTATIONIS UNKNOWN INSTANDARD X-RAY

X-ray focus

principal point

film planeFocus-film-distance and position of central beam have to be assessed approximatively.

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 10: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EBRA - LANDMARKS

film center

approximative principal point

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 11: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

cup

principal point film

focus

spatial parameter of migration

3D-METHOD EBRA-CUP

reference plane

reference line

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 12: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

ERRORS CAUSED BY CHANGES OF INTRINSIC ORIENTATION

Focus variation within cuboid 60 x 110 x 400 mm affects parameters of migration in a magnitude of only 0.01 mm (experimental study based on 78 radiographs)

900mm

1300mm

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 13: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

COMPENSATION OF ERRORS

• by use of an approximative intrinsic orientation bound to pelvic contours(based on study of 78 random x-rays)

• by use of variable reference planes passing through the x-ray focus

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 14: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EXTRINSIC ORIENTATION VARIES REMARKABLY

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 15: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

ERRORS CAUSED BY CHANGES OF EXTERIOR ORIENTATIONRotations of the pelvic bone cause remarkable errors up to 7 mm of simulated migration!

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 16: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

COMPENSATION OF ERRORS

• by use of a comparability strategy which compares radiographs with respect to similar conditions at exposure

• migration is measured only in pairs of comparable radiographs

• overlapping trends are weighted and averaged

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 17: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EBRA COMPARABILITY ALGORITHM

time

mig

ratio

n

non comparable x-ray

raw data

smoothed data

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 18: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

ACCURACY TESTS

• zero migration experiment (27 x-rays)• 33 pairs of x-rays taken within 1 month

• comparison with RSA (T. Ilchman et al.)

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

J Arthroplasty. 1992 Jun;7(2):121-27. Measurement accuracy in acetabular cup migration. A comparison of fourradiologic methods versus roentgen stereophotogrammetric analysis.Ilchmann T, Franzen H, Mjoberg B, Wingstrand H.

Page 19: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EBRA - ACCURACY

MIGRATION MEASUREMENT

• medio-lateral mig: 0.8 mm (95% CI)• cranial migration: 1.0 mm (95% CI)

ANGULAR MEASUREMENT

• cup anteversion: 3.3°• cup inclination: 2.5°

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria

Page 20: Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria EBRA-CUP: A METHODS FOR MIGRATION MEASUREMENT OF

EBRA GRAPHICS PROGRAM

PRINTER

EXCEL

Department of Engineering Mathematics, Geometry and Computer Science, University of Innsbruck, Austria