curso seguridad paciente m7 diapo 21 release of foreign bodies by clini. use intravenous infusion...
TRANSCRIPT
8/2/2019 Curso Seguridad Paciente M7 Diapo 21 Release of Foreign Bodies by Clini. Use Intravenous Infusion Sets
http://slidepdf.com/reader/full/curso-seguridad-paciente-m7-diapo-21-release-of-foreign-bodies-by-clini-use 1/5
"
l
~I$,,
ELSEVIER
Biomateriols 17 (1996) 663-666
i[l199fj Elsevier Science Lim ited
Pnnl"d in Crnat Brita in. All r ights reserved
0142-9612/961£15.00
Release of foreign bodies (particles) by
clinical use of intravenous infusionsets
Henrik Madsen and Ole Windingtnst ttute of Hygiene, UnlvEifs/('! 01 Copenhagen, Oenmark
in clinical practice, stripping the plastic tubes of intravenous (iv.) intusron sets 'Nlth a scissor blade IS
a commonly used method for re-establishing tlow in malfunctioning i.v. sets. The present Investigation
concludes that this procedure results In release of plastic particles from the luminal wall of the tube.
Particles are subsequently flushed into the patient. The average amount of particles released
exclusively from the i.v . infusion sets under these circumstances may exceed the standards for
acceptable particle content per millilitre in large volume i.v. injectable fluids. according to the British
Pharmacopoeia (1960) and the United Slates Pharmacopoeia (1990).
Keywords: Inrravenous .ntu ston sets. Siripping, oeructes , contsminetioo
Received 21 ;-.Jovember1994: accepted 25 ,'.~ay1985
-•1,
It has been known for decades that intravenous (i.v.l
infusion therapy is a source of i.v. particle
contamination, Particles are partly contained in the
infusion fluid itself, and partly derived from the inside
of the utensiis used, /\11 parts of the infusion system
(bottle". 'I, set, cannula) ha v e been shown to produce
particles ~-4. Considerable efforts have been made to
minimize particle contamination of infusion fluids,
Thus, empirical limits for maximum particle content
In parenteral solutions are given in the British
Pharmacopoeias and American Pharmacopoeia"
(Table 1), No limits are set for the contribution of
particles from devices and handling connected with
the administration of parenteral fluids.
As regards i,v. infusion sets. some particulate matter
is released, even from new sterile infusion sets 2---l , The
influence of clinical use and handling is included in
this investigation.
In clinical practice, flow malfunction in i.v. infusion
sets is often seen, even when correctlv installed and
changed within recommended intervals. I conducted a
questionnaire (unpublished) of 144 nurses in four
Danish and one Swedish hospitals, which showed that
more than 70% of the personnel stripped the infusion
sets-with a pair of scissors as first choice procedure for
re-establishing a normal flow under such circumstances.
B y stripping an i.v. infusion tube, high pressure is
applied to the luminal fluid, thus flushing the tube,
This is done b y ' initially bending the tuoe proximally
near to the drop chamber to prevent fluid escaping
backwards, The tube is then compressed between a
finger and a scissor blade and stripped downwards
i1
CDrrespondence to Dr H. Madsen, Lanojergparken J, OK 4000
Roskilde. Denmark.
towards the cannula, so that the fluid cleans the distal
part of the set by the pressure obtained,
The high prevalence of this rather violent method
prompted the present in vestigation. the purpose of
which was to examine the possible additional particle
release by stripping i.v, infusion sets.
MATERIALS A_. ' \ rD ~1ETHODS
r\ total of nine polyvinylchloride (PVC) i.v. infusion
sets of three different commercially available brands
were stud led. Each set was flushed three times with
100 ml membrane-filtered (0.2 pm pore size) distilled
water. The water was filtered separately for each
tlushing through a 13mrn diameter, 5.um pore size
Nucleopore filter, mounted on a stainless steel funnel
filter holder (total volume 40 rnll, thus collecting what
was shed from the inside of the set. Between the
second and the third flushings. the infusion tube was
stripped three times at a length of approximately
50 cm. by compressing the tube between the thumb
and a scissor blade. as shown in Figure 1. Each of the
total of 27 samples was labelled first, second or third
and also with the set number.
The entire filtering procedure was performed in a
Table 1 Hygienic standard of parenteral solutions (no. 01
particles allowed In 1ml\
British Pharmacopoeia (1980) 10eD 100
United Slates PharmacDpoela
XXII (1990)
50
66 3 Biomatenals \9%. Vol. 17 ~o. 7
5
8/2/2019 Curso Seguridad Paciente M7 Diapo 21 Release of Foreign Bodies by Clini. Use Intravenous Infusion Sets
http://slidepdf.com/reader/full/curso-seguridad-paciente-m7-diapo-21-release-of-foreign-bodies-by-clini-use 2/5
A n:
VG S tuttgart-H r. W olfsch rn ittVan:
KM /C -H r. D ittm arZeichen:
KM/DIT /FR
ZUT Kenntni~ (Verbleib):
Telefan
46 51Datum
08.05.96
B!BRAUNInterne Mi tteilunq
S d 1!~ g ee in ter H err W o lfsch rn itt,
anliegerul erhalten S ie, w ie versprochen , eine aktuelle V er6ffen tlichung zurn Problem des
Partikelabriebs bei m echan ischer M altratierung des In fusionsschlauches, beispielsw else
m it H ilf'e einer S chere, w ie in dieser Un tersuchung dargestellt.
Ich denke, dan sich diese Un tersuchung in etwa auch Falle ubertragen la13t, wo m it H ilfe
der geschlossen en R o llen klern me versucht w ird, ein halbw egs leergelaufenes ln fusiorisge-
rat w ieder luftfrei zu bekornmen , u rn so eine Neuverwendurig des B esteckes zu errnogli-
chen .
L eider ist die A rbeit in E nglisch, daher nachfolgend eine kurze Uhersetzung d er Z usarn -
rnenfassung un ter der Uberschrift auf S eite 663:
"In del' k lin ischen R outine ist das S trippen des Plastikschlauches eines in travenosen In fu-
sionssets mit einer S chere eine haufig verw endet M ethode, um den Fluf bei einern
schlecht funktio n ierenden In fusi 0 nsgera t w ieder herzustellen . D ie vo rliegende Un tersu-chung korn rn t zu dern S chluf}, daB dieses V erfahren zu einer Frellassung von Kunststoff-
partikeln aus der in neren W and des S chlauches fuhrt. Die Partikel werden in der Folge in
d en P atie nte n eingespult. Die durchschnittliche Partikelm enge, die un ter diesen Urnstan-den aus den lnfusiorissets freigesetzt wird, kann die von der B ritish Pharm acopoeia (1980)
und der U nited S tates P harm acopoeia (1990) festgesetzten S tandards fur akzeptable Parti-kelm engen pro M illiliter in grof3volum igen I nfu sio nslo su ng en u bersteig en ".
lch hoffe, d a1 3 I hn en d iese V e r6 ffen tlic hu ng ein wenig weiterhilft und verbleibe
m it freundlichen G rii13en
Anlage
8/2/2019 Curso Seguridad Paciente M7 Diapo 21 Release of Foreign Bodies by Clini. Use Intravenous Infusion Sets
http://slidepdf.com/reader/full/curso-seguridad-paciente-m7-diapo-21-release-of-foreign-bodies-by-clini-use 3/5
66 4 Particle production from i.v. sets: H. Madsen and O. Winding
Figure 1 The stripping procedure. Left hand bending the
Iruusron tube. right hand stripping downwards using
scissor blade
horizontal laminar air flow bench in a clean room. as
descnbed by Winding7, in order to avoid airborne
contamination of the sampling procedure from the
env ironment.
Subsequently. particles were counted in a scanning
electron microscope at x 500 magnification. tilt 45'.
using the equipment and the method described by
Windin~:( From each sample. particles were counted
in 100' randoml v chosen counting fields. Energy
dispersive X-ray equipment was used for further
investigation.
All utensils that came into contact with the infusionsets were cleaned in an ultrasound bath and flushed
with membrane-filtered distilled water to avoid
external contamination.
RESULTS
Handling by stripping the i.v. infusion sets produced
particles of various size and shape. Table 2 presents
the numbers and sizes 0 f particles counted after the
three flushings. (See Figures 2 and 3.)
As expected and reported previously':", some
particles were flushed from new sterile sets (i.e, first
flushing). Second flushing reveals considerably fewerparticles, probably because most of the loose particles
from the lumen have already been flushed out. Third
flushing snows a significant increase in particle
number. compatible 'with the effect 0[ the stripping
procedure, Median particle counts and range are
presented in Table 2.
Numerous plaques were found covering the filter
surface as a thin film. Energy dispersive analysis of X-
rays. which enables the detection of inorganic
Bandular fibre
100 urn r---I 100)(
20!-1m r---I 500 "
Figure 2 Example at bandular fibre. length approximately
BOO.um. from the third flushing. set no. 4. Prooably plastic
t ibr e.
Biornater iais 1996. Vol. 17 No.7
8/2/2019 Curso Seguridad Paciente M7 Diapo 21 Release of Foreign Bodies by Clini. Use Intravenous Infusion Sets
http://slidepdf.com/reader/full/curso-seguridad-paciente-m7-diapo-21-release-of-foreign-bodies-by-clini-use 4/5
Particle orooucuon tram i.v. sets: H. . '.1adsen and O. Winding 66 5
Table 2 . Particle counts accordinq to size for each flushing
separately (particles/100 rnl) Median values and range are
Dresented
First flushing Second flushing Third tlusning
5-IO.um
10--25 pm
25-S0!1m
50-75 pm
75--I00 pm
> 100ur n
2~60 (880--1 I 100)
1793 (610-6810)
685 (130-3070)
226 (20-78'0)
61 (0-120)52 (0-120)
2698 (650-<\060)
1856 (570-5400)
855 (120-2710)
293 (30- 1150)
112 (0-330)111 (10-220)
898 (300-2860)
616 (200-1300)
226 (30--480)
58 (10-140)
12 (0-40)28 ((hq0)
280 um particle
50 urn t-----1 200 x
20 ~ r------t 500:<
Figure:] Example of particle from tile firs: flushmg, set no.
2. ,\"ay be glass or metal.
chemica] elements (between nine and 92). showed a
silicon content in these plaques compatible with
silicon.e (see Fizure 4). There was no oarticular. ~ .lncrease in production of these plaques following
strip pi.ng , and we did not find their source. They may
derive from the i.v. sets or from the utensils (syringe.
cannula) used to feed the sets. Most likely, the silicon
detected could come from the syringe and cannula. asthese are normally siliconized. while the i..I. sets are
ce inside the venous system. particles can reach the
act as emboli and form granulomas or'8.9. Particles may even be shunted into the
system via non-obliterated canals or shunt
vessels, probably causing thromboembolic
llcations in other organs like brain and eyes.
Plaques showing
Silicium content
2000 x
500 x lSl250 urn
Figure 4 Plaques showing silicon con ten! compatible 'Nilh
silicone.
Lung granulomas created around intravascular
particles will probably have a larger diameter than the
particle itself. In critically ill patients this may have an
effect, especially considering the vast amount of
relatively large particles created by the stripping
procedure.
Even the parncte contribution without stripping
should be considered. Previously, calls have beenmade for in-line filtration of i.v.-infusion sets l. 10.11
,,\voiding the stripping procedure rnay have an equal
effective role in reducing particle production in i,v .
infusion therapy.
From the counts found in this investigation, particle
release exclusively from i.v. infusion sets may exceed
the limits (particles per millilitre) given for large
volume injectable l1uids bv the British
Pharmacopoeias and the United Stotes
Pnatmocoooeio".
It is not~d that the reason for a now malfunction in an
Biomateriais )ggl). Vo!. 17 :-' :0.7
8/2/2019 Curso Seguridad Paciente M7 Diapo 21 Release of Foreign Bodies by Clini. Use Intravenous Infusion Sets
http://slidepdf.com/reader/full/curso-seguridad-paciente-m7-diapo-21-release-of-foreign-bodies-by-clini-use 5/5
66 6 Particle production from i.v, sets: H. Madsen and O. 'Nind/fJg
I.V. infusion set may well be luminal thrombus
forma.tion at the tip of the catbetert2. Re-establishing
now by stripping the i.v. infusion tube will
subsequently deliver thrombi as well as particles to thepatient's venous system
REFERENCES
.
Ahnfeld FW, Klaus E. Quantitative Analysen uoer den
Partikelhalt von Infusicnslosungen. Zubehor und
Medikamenten. Anaestesist 1977: 26: 476-4801.
IlIum L. Character-ization of particulate contamination
released by application of parenteral solutions. r .
Particulate matter from administration sets, "rch
Pbarm. Chern Sc i Edn 1978: 6: 93-108.
Winding 0 Particle release from angiographic utensils .
Eut: I Radio! 198 l: 11-1-116.
Cooper OF. Baret: CW o Particulate matter from giv:ng
sets. Pharm r 1970: 205: 186-187.
fi
7
The British Pharmacopoeia 1980. London: The
Pharmaceutical Press, 1980.
The United States Pharmacopoeia 1990.
Winding O. A method for determination and
element analysis of particulate contamination in
injectable solutions. Am !Hosp Pharm 1976: 33:
1154-1159.
Franke RP. Ouanrifizierung Psrtikularer Bestandt.eile in
der Mlkrozirkulation der Lunge. Beitr Elektronetutu-
kroskop Direktabb Obert] 1986: 19: 429--434.
Carvan ) 1 v 1 . Gunnar BW. The harmful effects of
particles in intravenous solutions. Meii f ;-\.ust 1964:
2: 1-6.
Rvan PB. In-line final fihration=-A method of
minimizing contamination in intravenous therapy. Bull
Parent Drug Assoc 1973: 27: 1-14.
Faichuk KH et al. Microparticulate induced phlebitis.
New Eng!! Med 1985: 312: 78-82.
Bait T N . Petersen RV. Surface characteristics of plastic
intravenous catheters. Am I Hasp Piiartti 1979: 36:
1707-1711.
11
Biornatenals lY96. Vol. 17 No. 7