Critical Care Critical Care Monitoring Monitoring
Nuts and BoltsNuts and Bolts
Mike McEvoy, PhD, REMT-P, RN, CCRNAlbany Medical Center, Albany, New York, USA
Cardiothoracic Surgical ICU
![Page 2: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/2.jpg)
![Page 3: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/3.jpg)
Mike McEvoy, PhD, RN, CCRN, REMT-P
www.mikemcevoy.com
![Page 4: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/4.jpg)
DisclosuresDisclosures• I serve on the speakers bureaus for
Masimo Corp. and Medtronic Corp.• I have no other financial
relationships to disclose.• I am the EMS editor for Fire
Engineering magazine.• I do not intend to discuss any
unlabeled or unapproved uses of drugs or products.
![Page 5: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/5.jpg)
Jones & Bartlett, 2010Jones & Bartlett, 2010
![Page 6: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/6.jpg)
Goals for this talk:Goals for this talk:• Objectives of hemodynamic
monitoring• Blood pressure measurement• Art lines in practice• Preventing complications• Troubleshooting• Treatment parameters/goals• Non-invasive monitoring• New technologies
![Page 7: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/7.jpg)
Goal of hemodynamic Goal of hemodynamic monitoringmonitoring
• Assess tissue perfusion– Oxygenation and distribution (flow)
• Others?– Respirations
– Hydration
– Labs:» Chemistries
» Hematology
» Toxicology
» Microbiology
![Page 8: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/8.jpg)
3 types of shock:3 types of shock:
1. Distributive (septic)
2. Volume (hypovolemic)
3. Pump (cardiogenic)
![Page 9: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/9.jpg)
Are physical findings Are physical findings enough?enough?
• HR
• LOC
• BP
• UO
![Page 10: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/10.jpg)
Apparently not…Apparently not…
• 50% of physical assessments wrong • Therapeutic interventions altered with
invasive assessment 34 - 56% of the time:– 1980 Del Guercio - 1984 Connors – 1984 Eisenberg - 1990 Bailey – 1991 Steinberg - 1993 Coles – 1994 Minoz - 1998 Staudinger – 2002 Jacka
![Page 11: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/11.jpg)
Lung Sounds in HFLung Sounds in HF
• If rales were present, all had a wedge pressure >18, very specific
• Only 9 of 37 with a wedge pressure >18 had rales, very insensitive
• So…clear lung fields tell you very little about fluid status in heart failure
Butman et al. J Amer Coll Cardiol. 10/93
![Page 12: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/12.jpg)
So we dove right in…So we dove right in…
Swan Ganz Catheterization
![Page 13: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/13.jpg)
Connors et al 1996 Connors et al 1996 JAMAJAMA
5734 adult ICU patients 1989-1994, 5 ICUs at 15 tertiary med centers
PA cath = 30 day mortality, ICU LOS, costs of care
![Page 14: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/14.jpg)
Harvey et al: PAC-Man Harvey et al: PAC-Man 2005 Lancet 2005 Lancet - - Game Game Over?Over?
1014 patients at 65 UK institutions:NO DIFFERENCE between PA cath versus no PA cath
![Page 15: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/15.jpg)
![Page 16: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/16.jpg)
Cochrane R & R: 2006 Cochrane R & R: 2006 (Review and Reappraisal)(Review and Reappraisal)
“The PAC is a monitoring tool; if it is used to direct therapy
and there is noimprovement in outcome, then the therapy does not
help.”
![Page 17: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/17.jpg)
Two Problems:Two Problems:
1. Define “normal”
2. Who’s behind the wheel?
![Page 18: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/18.jpg)
Blood Pressure
BP = CO x SVR
Indirect Pressure Measurement Direct Pressure Measurement
** A diastolic pressure of 60 is necessary to maintain coronary artery perfusion.
![Page 19: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/19.jpg)
Why do we measure Why do we measure BP?BP?• Because we can.
![Page 20: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/20.jpg)
Purpose of blood Purpose of blood pressurepressure
![Page 21: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/21.jpg)
Biventricular CV Biventricular CV SystemSystemL (systemic) R (pulmonic)
LV 110/10 RV 25/ 0-5
AO 120/80 PA 25/10
capillary 30-50 capillary 12-17
RA 0-5 LA 8-12
![Page 22: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/22.jpg)
![Page 23: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/23.jpg)
![Page 24: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/24.jpg)
Arterial Pressure Arterial Pressure MonitoringMonitoring
Direct
Pressure
Indirect
Flow
![Page 25: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/25.jpg)
Flow Measurements Flow Measurements Not Accurate:Not Accurate:
• Low blood flow states
• High SVR states– Avg 33.1 mmHg difference
cuff vs. a-line
– Cuff consistently underestimates pressure
- Cohn, JM (JAMA 199:972, 1967)
![Page 26: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/26.jpg)
Flow measurements Flow measurements • Pulses• Cuff• NIBP• Doppler
All sense pulsatile flow
![Page 27: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/27.jpg)
27
What We What We KnowKnow about about FlowFlow
(Indirect Measurement)(Indirect Measurement)• Pulses: – Carotid = SBP > 60– Femoral = SBP > 70– Radial = SBP > 80
• Cuff– Errors in measurement r/t size and heart level
• NIBP– Calculates systolic and diastolic based on MAP and HR
• Doppler– PEA
• Flow based measurements are NOT accurate in low flow states or with high SVR, e.g. shock
![Page 28: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/28.jpg)
Avoid assumptions !Avoid assumptions !
1. BP ≠ blood flow
levophed…
2. Blood flow ≠ perfusion
O2 or nutrient deficiency…
![Page 29: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/29.jpg)
Arterial Pressure Monitoring Indications
Patient in shock not rapidly responsive to therapy
Insertion SitesRadialBrachialAxillaryFemoralDorsalis Pedis
![Page 30: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/30.jpg)
Arterial Pressure Monitoring
Radial artery has the benefit of collateral circulation from the ulnar artery
Allen Test used to evaluate the collateral flow prior to radial artery cannulation
![Page 31: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/31.jpg)
A-line Monitoring Set-up
![Page 32: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/32.jpg)
Invasive Monitoring Equipment
Flush solution -- usually heparinized Continuous flush system (usually a
pressure bag or pump) Pressure transducer and pressure tubing Invasive catheter Monitor
![Page 33: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/33.jpg)
Transducers
Convert one form of energy to another
Sense pressure Convert it to an
electrical signal Electrical signal
causes monitor reading
+
-
Dom e
![Page 34: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/34.jpg)
Leveling, Referencing, Balancing…
Placing the air-fluid interface of the catheter system at the phlebostatic axis
This negates the weight effect of the fluid in the catheter tubing (hydrostatic pressure)
“Setting the correct reference point is the single most important step in setting up a pressure monitoring system.” … Gardner, 1993
![Page 35: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/35.jpg)
Leveling: the # 1 Cause of Error in Pressure Monitoring
![Page 36: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/36.jpg)
Phlebostatic Axis Located at the
intersection of the 4th ICS and midway between the anterior and posterior surfaces of the chest
Midaxillary line is NOT interchangeable with mid anteroposterior level in all persons … Bartz, et al, 1988
![Page 37: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/37.jpg)
Phlebostatic Axis
As the patient moves from flat to upright, the phlebostatic level rotates on the axis and remains horizontal. This position confirmed by CT by Paolella, et al, 1988.
![Page 38: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/38.jpg)
Phlebostatic Axis
The phlebostatic axis moves to midchest at the 4th ICS when patient is in the lateral position.
![Page 39: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/39.jpg)
Leveling
Air fluid interface is the point in the system that is opened to air during zeroing
Inaccuracies are produced if the air-fluid interface is above or below the phlebostatic axis – 1.86 mmHg/inch
Phlebostatic axis determined by Windsor and Burch (1945) as correct reference for measurement of venous pressures
![Page 40: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/40.jpg)
Give 500 ml of LR for CVP < 5 Transducer leveled 2 inches too
high 1.86 mmHg/inch x 2 =
underestimation of actual CVP by 3.72 mmHg (Ooops!)
Recorded CVP = 3 500 cc bolus of LR given Actual CVP = 7 (before LR
bolus)
![Page 41: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/41.jpg)
Zeroing
Opening the system to air to establish atmospheric pressure as zero (0)
This negates all pressure contributions from the atmosphere
Allows only pressure values that exist within the heart or vessel to be measured
![Page 42: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/42.jpg)
When to Zero Before insertion After disconnecting transducer from
pressure cable When values are in question
Ahrens, T. et al. Frequency requirements for zeroing transducers in hemodynamic monitoring, Am J Crit Care, 1995;4:466-471
![Page 43: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/43.jpg)
Arterial Pressure Waveform
1. Systole
2. Dicrotic notch
3. Diastole
![Page 44: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/44.jpg)
![Page 45: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/45.jpg)
Placement of Arterial Placement of Arterial LineLine
150
90
60
![Page 46: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/46.jpg)
MAPMAP
![Page 47: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/47.jpg)
Art Line Placement:Art Line Placement:• The farther out, the higher the SBP• Cuff has no correlation
– Pressure vs. Flow
• Mean Pressure always consistent
![Page 48: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/48.jpg)
If BP increases, does If BP increases, does flow increase?flow increase?
• Think of levophed…
NOPE
![Page 49: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/49.jpg)
Preventing Complications with Arterial Lines
Complication Preventive Measure
Confusion with venous line Clearly label tubing as “Arterial”
Hemorrhage ALARMS, routinely check connectors
Infection Label dressing, careful sampling
Loss of waveform, catheter movement Immobilize extremity, secure catheter to skin
Blood back-up into tubing, loss of waveform
Pressurize system > patient blood pressure
Loss of circulation distal to site ?spasm vs thrombus? - may need to d/c
Clotted Use continuous flush system
![Page 50: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/50.jpg)
Troubleshooting Common Arterial Line Problems
Damping of waveformCauses:
Flush bag empty or pressure < patient pressure Catheter tip against vessel wall Clot at catheter tip Air bubbles in system Kinked catheter or tubing
![Page 51: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/51.jpg)
Troubleshooting Common Arterial Line Problems
Damping of waveform Interventions:
Keep flush bag at 300 mmHg Reposition extremity, use splint if necessary Gently aspirate clot, then flush line Clear system of air bubbles (limit to 1
stopcock) Remove kinks in tubing, check site, consider
suturing catheter to skin
![Page 52: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/52.jpg)
Troubleshooting Common Arterial Line Problems
Abnormally high/low readingsCauses:
Transducer not level Hypertension/hypotension System error
Interventions: Re-level system Assess pressure with alternate means Determine and optimize system dynamic
response
![Page 53: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/53.jpg)
Two Problems:Two Problems:
1. Define “normal”
2. Who’s behind the wheel?
![Page 54: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/54.jpg)
What is “normal”?What is “normal”?
Blood Pressure:
• Bland, ShoemakerJ Surg Obst 1978 -– 74 % of survivors
achieved normal values– 76% of NON-
SURVIVORS achieved normal vital signs
![Page 55: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/55.jpg)
Is it the car or the Is it the car or the driver?driver?• If you don’t know how to drive, a car can be a dangerous weapon.
• If you don’t know how to interpret the data, a monitoring catheter can be a dangerous weapon.
![Page 56: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/56.jpg)
What we really treat:What we really treat:NL Treat
HR 60 – 100 < 40 or > 120
MAP 70 – 105 < 60 – 70
CVP 2 – 6 < 5, Panic > 20
Sa02 95 – 100% < 90
ScvO2 > 70% (sepsis) <70% if hct >30
CI 2.5 – 4.0 < 2.2 – 2.5
SI 25 - 45 < 25 concerns you
Lactate 1 – 2 > 4 mmol/L and pH < 7.30 suggests tissue hypoxia
pH 7.35 – 7.45 Hmmm… 6.888, BE + 5…
![Page 57: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/57.jpg)
Critical Information
• Oxygenation
• Perfusion
![Page 58: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/58.jpg)
Evaluating PerfusionEvaluating Perfusion• Tools we have
![Page 59: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/59.jpg)
Lactate (Lactic Acid)Lactate (Lactic Acid)
Hypoperfusion severity index
NL < 2, concerned when > 4
> 15 often fatal
More helpful as trend
![Page 60: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/60.jpg)
POC Lactate TestingPOC Lactate Testing
• Developed for athletes & climbers
• Not FDA approved
• Currently underinvestigation in EMSand Fire service
![Page 61: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/61.jpg)
SvOSvO22
• Reflects O2 reserve & extraction
• < 60% requires investigation:– Hct, CO, SaO2
– VO2
• The lower the level, the worse…
• < 40% typically fatal
![Page 62: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/62.jpg)
ScvOScvO22
• O2 reserve & extraction upper body
• Typically 5 – 13% > SvO2 (avg 7.5)
• NL > 70%
• Sampled from CVC (oximetric CVL available)
![Page 63: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/63.jpg)
Gastric Mucosal COGastric Mucosal CO22
• Recent data suggest PgCO2 may reflect perfusion
• CO2 clearance reflects perfusion
• A-g CO2 gap < 10 is normal
• Pa CO2 - PgCO2 gap > 10 is bad
![Page 64: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/64.jpg)
Sublingual COSublingual CO22 – PslCO – PslCO22 Very proximal gut
NL = 44 – 64 mmHg
levels correlate with perfusion
Studies halted August 2004
![Page 65: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/65.jpg)
Nellcor Capnoprobe™ Nellcor Capnoprobe™ US approval Jan 2003.
Research study: Children's Medical Center – Dallas TX.
11 kids infected Burkholderia cepacia, 2 died.
Traced to saline solution packaging of probes.
5,600 units @ 30 centers recalled
![Page 66: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/66.jpg)
CapnographyCapnography
• CO2 clearance reflects perfusion!
• Available for intubated and non-intubated patients
• Developmentscoming: IPI
![Page 67: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/67.jpg)
Decision to Call the Decision to Call the CodeCode• 120 prehospital patients in
nontraumatic cardiac arrest
• EtCO2 had 90% sensitivity in predicting ROSC
• Maximal level of <10mmHg during the first 20 minutes after intubation was never associated with ROSC
Source: Canitneau J. P. 1996. End-tidal carbon dioxide during cardiopulmonary resuscitation in humans presenting mostly with asystole, Critical Care Medicine 24: 791-796
![Page 68: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/68.jpg)
End-tidal COEnd-tidal CO22 (EtCO (EtCO22))
Normal a-A gradient– 2-5mmHg difference between the EtCO2
and PaCO2 in a patient with healthy lungs
– Wider differences found » In abnormal perfusion and ventilation » Incomplete alveolar emptying» Poor sampling
![Page 69: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/69.jpg)
Future DevelopmentsFuture Developments
• Perfusion assessment derived from exhaled CO coupled with bioimpedance data.
![Page 70: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/70.jpg)
Integrated Pulmonary Integrated Pulmonary IndexIndex™™
![Page 71: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/71.jpg)
IPI Values – fuzzy logicIPI Values – fuzzy logic
IPI Patient Status10 Normal
8-9 Within normal range
7 Close to normal range; requires attention
5-6 Requires attention
3-4 Requires attention or intervention
1-2 Requires intervention
![Page 72: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/72.jpg)
Acoustic Resp Acoustic Resp MonitoringMonitoring• Likely release 2010
• Electrical sensor based• Initially will report RR
• Future versions VT
• Telemetry based• May replace capnography?
![Page 73: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/73.jpg)
Esophageal doppler Esophageal doppler (TED)(TED) Transducer probe inserted into distal esophagus
Blood flow measured by doppler principle
![Page 74: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/74.jpg)
TTE (Trans Thoracic TTE (Trans Thoracic Echo)Echo)• Also nurse or medic driven
• Chest wall instead of esophageal
![Page 75: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/75.jpg)
TED/TTE:TED/TTE:
![Page 76: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/76.jpg)
Bioimpedance (TEB)Bioimpedance (TEB)
• Thoracic electrical bioimpedance
• 4 electrodes sent current through thorax, analyze flow resistance
• With age, gender, height, weight calculate SV, ejection time, CO, thoracic fluid content, acceleration index, velocity index, etc.
![Page 77: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/77.jpg)
TEB (CardioDynamics TEB (CardioDynamics BioZBioZ®®))
![Page 78: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/78.jpg)
RELIANT RELIANT Non Invasive Hemodynamic MonitorNon Invasive Hemodynamic Monitor
![Page 79: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/79.jpg)
SVR MAP-CVP / CO
CO Cardiac Output
CI Cardiac Index
SV Stoke Volume
SVV Stroke Volume Variance
SVI Stroke Volume Index
HR Heart Rate
TPR Total Peripheral Resistance
VET Ventricular Ejection Time
MAP Mean Arterial Pressure
NIBP Non Invasive Blood Pressure
TFC Thoracic Fluid Content
TFCd % Directional Change in TFC/Time
CP Cardiac Power
CPI Cardiac Power Index
CAPTURES (14 ) PARAMETERSCAPTURES (14 ) PARAMETERS“In Real Time”“In Real Time”
TPR : Dynes – (MAP / CO)*80 TPR : mmHg * min./liters – (MAP / CO)
TFCd : % Change in TFC over 15 mins. Vs. baseline TFC CP: MAP*CO/451
CPI: CP/BSA
![Page 80: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/80.jpg)
Current of a known amplitude & frequency is applied on outer electrodes
Voltage signal captured on inner electrodes
100% Noninvasive100% NoninvasiveMonitors Any Patient, AnywhereMonitors Any Patient, Anywhere
Change in phase of the frequency is recorded and the signal translated to flow (similar to Doppler in concept)
![Page 81: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/81.jpg)
0 I
II
II
I
Volts
Io
Vo
Io
Vo
0 I
II
II
I
Amp.
Bioimpedance
Bioreactance
![Page 82: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/82.jpg)
PASSIVE LEG RAISE TEST (PLRT)PASSIVE LEG RAISE TEST (PLRT)
12%12%
FLUIDFLUIDRESPONSIVERESPONSIVE
12%12%
NOT FLUIDNOT FLUIDRESPONSIVERESPONSIVE
3 MINS. BASE LINE CO/CI
11
VENOUS BLOOD SHIFT
BEFORE PLRBEFORE PLR DURING PLRDURING PLR3 MINS. CO/CI POST SHIFT
22
CO & CI IMPACTCO & CI IMPACT
![Page 83: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/83.jpg)
• Volume expansion 1st line of therapy.
• Only ½ of patients respond to fluids with increased CO.
• Need a reliable means to determine ability to respond to fluid.
Problem: Enough Volume? Problem: Enough Volume?
![Page 84: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/84.jpg)
Chest 2002;121;1245-1252
![Page 85: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/85.jpg)
PLR?? PLR?? Passive
Leg Raising45 °
Semi-Fowler’s Passive
Leg Raising45 °
Semi-Fowler’s
• 150 – 300 ml volume• Effects < 30 sec.. Not more than 4 minutes• Self-volume challenge• Reversible
![Page 86: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/86.jpg)
Ocular ScannerOcular Scanner
EyeMarker Systems™
![Page 87: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/87.jpg)
Retinal imagingRetinal imaging• Pattern
recognition:– Botulism,
neurotoxins– Nerve Agents– Carbon Monoxide– Cyanide
![Page 88: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/88.jpg)
Hydration StatusHydration Status
• Saliva osmolality
![Page 89: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/89.jpg)
Perfusion IndexPerfusion Index• Perfusion Index is an objective method for
measuring a patient’s peripheral perfusion• Perfusion Index is an early indicator of
deterioration
![Page 90: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/90.jpg)
Perfusion IndexPerfusion Index
Infrared
Saturation
Red
![Page 91: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/91.jpg)
• 108 healthy, 37 critically ill adults (finger sensors)
• PI range: 0.3% to 10%, median 1.4%
• ROC used to determine the “cutoff” value
• 1.4% PI best discriminated normal from abnormal
What is a “Normal” PI ?What is a “Normal” PI ?
Lima, et al. CCM 2002
![Page 92: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/92.jpg)
Clinical Uses for PI Normals have been suggested to be:
>1.4% adults, >1.27% neonates1.Site selection (varies between patients and sites)2.Chorioamnionitis (placental membrane/amniotic
fluid infection)3.Effectiveness of Servoflorane anesthesia4.Monitor onset/effectiveness of epidural anesthesia5.Predict illness severity scores (good correlation)6.Monitor/quantify peripheral perfusion7.Detect shock states
8.PI trend may best reflect changes in condition
![Page 93: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/93.jpg)
PhotoplethysmographyPhotoplethysmography
Abs
orpt
ion
TimeTime
R IR
Photodetector
Pleth Waveform
![Page 94: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/94.jpg)
Pleth WaveformPleth Waveform
![Page 95: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/95.jpg)
A-line versus Pulse Ox Pleth
![Page 96: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/96.jpg)
Definition of PVIDefinition of PVI• Pleth Variability Index (PVI) is a
measure of dynamic changes in PI that occur during the respiratory cycle
• PVI is a percentage from 1 to 100%: 1 = no variability and 100 = maximum variability
![Page 97: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/97.jpg)
Fluid Status/Volume Responsiveness
• High variability (high PVI) = volume depletion
• 15 – 50% of patients are fluid non-responders – low variability (low PI) suggests the patient is a non-responder– The ventricle more sensitive to respiratory
changes is more responsive to preload
![Page 98: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/98.jpg)
Pulse CO-Oximetry
Oxygenated Hb and reduced Hb absorb different amounts of Red (RD) and Infrared (IR) Light
![Page 99: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/99.jpg)
Pulse CO-Oximetry
Oxygenated Hb and reduced Hb absorb different amounts of Red (RD) and Infrared (IR) Light
1. Carboxyhemoglobin
2. Methemoglobin
3. Hemoglobin
4. ? Glucose
5. ? Cyanide
6. ?
![Page 100: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/100.jpg)
SummarySummary
• Perfusion is the goal
• Perfusion = oxygenation + flow
• You cannot do it alone
• Less invasive is better
• Technology should make you a better clinician (only as good as u)
![Page 101: Critical Care Monitoring Nuts and Bolts Mike McEvoy, PhD, REMT-P, RN, CCRN Albany Medical Center, Albany, New York, USA Cardiothoracic Surgical ICU](https://reader037.vdocuments.us/reader037/viewer/2022110322/56649d585503460f94a38376/html5/thumbnails/101.jpg)
Thanks for your Thanks for your attention!attention!