cr/dr image artifacts

47
George David Associate Professor of Radiol Medical College of Georgia

Upload: jared56

Post on 15-Jun-2015

1.506 views

Category:

Documents


9 download

TRANSCRIPT

Page 1: CR/DR Image Artifacts

George DavidAssociate Professor of Radiology

Medical College of Georgia

Page 2: CR/DR Image Artifacts

Quick Review of Technology

Page 3: CR/DR Image Artifacts

Computed Radiography (CR)Re-usable metal imaging plates

replace film & cassette

                                                                 

Page 4: CR/DR Image Artifacts

CR Exposure

Page 5: CR/DR Image Artifacts

Reading Imaging PlateReader scans

plate with laser light using rotating mirror

Plate pulled through scanner by rollers

Light given off by plate measured byPM tube &recorded by computer

Page 6: CR/DR Image Artifacts

CR Operation

after read-out, plate erased using a bright light

plate can be erased virtually without limit

Plate life defined not by erasure cycles but by physical wear

Page 7: CR/DR Image Artifacts

CR LatitudeVery largePlate responds to

decades of input exposure

Computer scale inputs exposure to viewable densities

               

Page 8: CR/DR Image Artifacts

CR Very Sensitive to Scatter

Page 9: CR/DR Image Artifacts

Digital Radiography (DR)Digital bucky incorporated into

equipmentDirect digital outputHigh Lattitude

Page 10: CR/DR Image Artifacts

Raw Data ImageUnprocessed image as read from receptor

Not a readable diagnostic image

Requires computer processing before presentation as finished radiograph

Page 11: CR/DR Image Artifacts

Enhancing Raw Image (Image Segmentation)

1. Identify collimated image border

2. Separate raw radiation from anatomy

3. Apply appropriate tone-scale to image

Create look-up table (LUTLUT) Maps pixel value to display gray

shade

This process is specific to a

particular body part and

projection

*

Page 12: CR/DR Image Artifacts

Image SegmentationEstablish location of

collimated border

• Define anatomic region

• Produce look-up table based on histogram from anatomical area• Body part & projection specific

Page 13: CR/DR Image Artifacts
Page 14: CR/DR Image Artifacts

Gross Overexposure

Properly exposed

• 8X increase in mAs

Page 15: CR/DR Image Artifacts

Gross Underexposure

Properly exposed

• mAs reduced by ~ 100

Page 16: CR/DR Image Artifacts

Artifacts in Digital Radiography

Page 17: CR/DR Image Artifacts

Artifact CategoriesNon-digital-related technical errorsLook-up table / image processing errorsExposure artifactsCR artifactsDR artifactsInterference artifactsDisplay artifacts

Page 18: CR/DR Image Artifacts

Radiography ArtifactsDigital or Not

Mis-positioningMotionIncorrect patient IDDouble exposuresGrid cut-off

Page 19: CR/DR Image Artifacts

No grid102 cm SID.33R ESE

GridGrid122 cm SID122 cm SID2.2 R ESE2.2 R ESE

http://http://radiographics.rsnajnls.org/radiographics.rsnajnls.org/cgi/reprint/11/2/307?cgi/reprint/11/2/307?maxtoshow=&HITS=10&hitmaxtoshow=&HITS=10&hits=10&RESULTFORMAT=s=10&RESULTFORMAT=&fulltext=grid+artifacts&and&fulltext=grid+artifacts&andorexactfulltext=and&searchorexactfulltext=and&searchid=1&FIRSTINDEX=0&sortid=1&FIRSTINDEX=0&sortspec=relevance&resourcetspec=relevance&resourcetype=HWCIType=HWCIT

Page 20: CR/DR Image Artifacts

References

“Digital Radiography Artifacts”from Veterinary Radiology & UltrasoundWm Tod Drost, David J. Reese, William J. Hornof

“Artifacts in Digital Radiography” from Veterinary Radiology & UltrasoundDavid A. Jimenez, Lauara J. Armbrust, Robert T.

O’Brien, David S. BillerRadiographicshttp://www.slideshare.net/ricksw78/digital-

radiography

Page 21: CR/DR Image Artifacts

Look-up Table / Image Processing Errors

Page 22: CR/DR Image Artifacts

•Diagonal collimation caused algorithm to improperly recognize collimated area•Gonadal shields can fool algorithm•2 exposures on one plate can fool algorithm

http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?maxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=amaxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resortifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resourcetype=HWCITurcetype=HWCIT

Page 23: CR/DR Image Artifacts

Chest incorrectly specified by technologist to CR reader as lumbar spine

http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?maxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifact&maxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resourcetype=HWsearchid=1&FIRSTINDEX=10&sortspec=relevance&resourcetype=HWCITCIT

Page 24: CR/DR Image Artifacts

Uberschwinger Artifact

Causes appearance of fine black line halo

Result of implementation of edge enhancement

Can be mistaken for evidence of loosening of orthpedic device

Can mimic pneumothorax

Page 25: CR/DR Image Artifacts

More Uberschwinger

Page 26: CR/DR Image Artifacts
Page 27: CR/DR Image Artifacts

ClippingOccurs because of applying LUT during pre-

processingKodak recommends technologists do not

window/level on qc workstation.Raw data: 12-14 bit imageLUT during pre-processing: Only 10-12 bits

savedInformation may be lost

Page 28: CR/DR Image Artifacts

Low Exposure Results in Image Noise

Increasing mAs

Page 29: CR/DR Image Artifacts

Exposure ArtifactVERY high exposures can result in saturationMaximum detector response reachedNo response to increased doseUniformly darkCannot be windowed/leveled

Page 30: CR/DR Image Artifacts

CR ArtifactsFadingLight leakPhysical damage to imaging plates

Cracks, scuffs, scratchesContaminationDust / dirt

Dirt in readerHighly sensitive to scatter radiationUpside-down insertion into bucky

Page 31: CR/DR Image Artifacts

CR Artifact: FadingCR latent image consists of excited electrons

stuck in high energy state orbitsOver time if image not read then electrons

fall back to ground energy stateSeveral days required before significant

fading occurs

Page 32: CR/DR Image Artifacts

CR Artifact: Light Leak in CassetteCR plates erased by exposure to bright lightLight leaks in cassette can cause partial

premature erasure of plate

Page 33: CR/DR Image Artifacts

•On imaging plate during reading•In reader

•Can get on plate•Can be blocked in light guide of reader

•White line in direction of plate movement

•CR readers require periodic preventive maintenance & cleaning http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?maxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=amaxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resortifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resourcetype=HWCITurcetype=HWCIT

Page 34: CR/DR Image Artifacts

Dirt on CR plate. Right image repeated on screen-film

http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?maxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifact&searmaxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resourcetype=HWCITchid=1&FIRSTINDEX=10&sortspec=relevance&resourcetype=HWCIT

Page 35: CR/DR Image Artifacts

What are white lines?

Page 36: CR/DR Image Artifacts

Cracks on image plate due to mechanical wear

http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?maxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifmaxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resourcetact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resourcetype=HWCIType=HWCIT

Page 37: CR/DR Image Artifacts

http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?http://radiographics.rsnajnls.org/cgi/reprint/11/5/795?maxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=amaxtoshow=&HITS=10&hits=10&RESULTFORMAT=&fulltext=artifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&resortifact&searchid=1&FIRSTINDEX=10&sortspec=relevance&reso

urcetype=HWCITurcetype=HWCIT

Page 38: CR/DR Image Artifacts

Transport problems in CR reader

Laser

RotatingMirror

Beam scannedacross film

PM Tube

Analog/Digital

Computer

Light

Plate Travel

Page 39: CR/DR Image Artifacts

CR Plates are Very Susceptible to Scatter

Do not keep plates in room during other exposures

Do not store where likely to receive scatter

Erase plate before use if it has been sitting a long time

Page 40: CR/DR Image Artifacts

DR ArtifactsDead detector elements

Software correction to a point…

R/F interferenceDetector shielded to

prevent thisPeriodic pattern

Page 41: CR/DR Image Artifacts

DR Artifacts: Spatial Variations in Background Signal & Gain

Addressed by acquiring flood and building calibration maskSimilar to uniformity correction in nuclear

medicineArtifact results if attenuator in beam during

calibration, then removedContrastTape

Page 42: CR/DR Image Artifacts

Tape artifact seen after tape removed

Page 43: CR/DR Image Artifacts

Interferance

Page 44: CR/DR Image Artifacts

CR Grid Interference103 lines / inch grids have same

frequency as CR laser scanner. This can cause “Moire” pattern artifact

Align grid lines perpendicular to scan orientation whenever possibleReduces chances of artifacts caused by laser

scanner.

Page 45: CR/DR Image Artifacts

Moire effect because of interference between scan frequency of matrix and spiral

Page 46: CR/DR Image Artifacts

DisplayPoor calibrationDead pixelsBacklight issues

Page 47: CR/DR Image Artifacts

The End

?