Page 1: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP5 ColorPro V Color Analyzer Form 7722 Operation Manual

i

CP5 Tri-stimulus Contact

Colorimeter ColorPro V

Operation Manual

3200 Sencore Drive, Sioux Falls, SD 57107

Page 2: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP5 ColorPro V Color Analyzer Form 7722 Operation Manual

ii

WARNING

PLEASE OBSERVE THESE SAFETY PRECAUTIONS

There is always a danger present when using electronic test equipment. Unexpected voltages can be present at unusual locations in defective

equipment and distribution systems. Become familiar with the equipment with which you are working, and observe the following safety precautions.

Every precaution has been taken in the design of the Tri-stimulus Contact Colorimeter to insure that it is as safe as possible. However, safe operation depends on you, the operator. 1. Never exceed the limits of the meter hardware, as given in the specifications

section or other special warnings provided in this manual.

2. Always be sure that your equipment is in good working order. Broken or frayed test leads or cables can be extremely dangerous and may expose you to high voltages.

3. Remove test leads immediately following measurements to reduce the possibility of shock.

4. Do not work alone when working under hazardous conditions. Always have another person available in case of an accident.

5. Never assume that a cable shield is at earth ground potential. Both static and electrical voltages can be present on a cable’s sheath.

6. Always follow standard safety procedures.

Page 3: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

1

CP5 Tri-stimulus Contact Colorimeter

ColorPro V

Operations Manual

3200 Sencore Drive, Sioux Falls, SD 57107

Page 4: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

2

TABLE OF CONTENTS

SAFETY WARNING ............................................................................... inside front cover

DESCRIPTION Introduction ................................................................................................................................... 3 Features ........................................................................................................................................ 3 Specifications ................................................................................................................................ 4 CP6000 ColorPro V Sensor and Attachments .............................................................................. 5 Supplied/Optional Accessories ..................................................................................................... 6 OPERATION Connecting Equipment .................................................................................................................. 7 Running the ColorPro Software .................................................................................................... 7 Dark Calibration ............................................................................................................................ 7 Operational Hints .......................................................................................................................... 7 ColorPro by CalMAN software ...................................................................................................... 8 USB to Serial Converter Driver Installation ................................................................................. 10 APPENDIX A: Measuring Light as the Eye Sees Light .............................................................. 13 APPENDIX B: CIE-UCS Chromaticity Coordinates .................................................................... 16 WARRANTY AND SERVICE INFORMATION ....................................... inside back cover

Page 5: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

3

DESCRIPTION

Introduction The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to calibrate video displays quickly and accurately, with minimal manual adjustment, especially when used with a Sencore video generator. The CP5 provides excellent low light level sensitivity and accuracy and can be used on all direct view and rear projection display types.

CP5 Colorimeter Features The CP5 Colorimeter provides precise and convenient measurement and calibration of all direct-view and rear projection displays, including LCD, plasma, CRT, DLP, SXRD, and D-ILA displays. Note: Use the Sencore OTC1000 Optical ColorPro sensor for proper calibration of all front projection systems. The CP5 colorimeter measurements closely match human eye response, using highly accurate tri-stimulus filtered sensors. The CP5 is capable of accurately measuring down to 0.01 foot Lamberts, with fast measurement updates, giving more accurate background and picture colors, especially in the darkest pictures, where the human eye is most critical of color accuracy. Fast, constant-rate measurement updates means you never wait for a measurement while making display calibration adjustments. Auto Dark Calibration eliminates a manual calibration step to simplify your calibration and eliminate measurement errors.

Page 6: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

4

SPECIFICATIONS Performance for 1.0 cd/m2 < Y < 1000 cd/m2

Accuracy—Display:

White (x/y) ± 0.002 Color (RGB, x/y) ± 0.004 Luminance (Y) ± 3% Y

Repeatability (same sensor, no dismount):

White (x/y) ± 0.001 Color ± 0.001 Luminance ± 0.5% Y Measurement Time: Approximately 1 second

Performance for 0.05 cd/m2 < Y < 1.0 cd/m2

Accuracy—Display:

White (x/y) ± 0.006 Color (RGB, x/y) ± 0.006 Luminance (Y) ± 5% Y

Repeatability (same sensor, no dismount):

White (x/y) ± 0.002 Color ± 0.002 Luminance ± 0.5% Y Measurement Time: Approximately 1 second

Sensor linearity throughout dynamic range

Measured on integrating sphere with tungsten illumination: ± 2%

Luminance resolution:

Measuring LCD: ~ 0.01 cd/m2 Measuring CRT: ~ 0.25 cd/m2

Weight & Size (approximate): unpacked

Total weight: 0.45 lb. Diameter: 2.7” Height: 1.6” Cable length: 72”

Page 7: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

5

ColorPro V Colorimeter and Attachments

1. ColorPro V Meter: Connects to the PC via a USB Port and cable. The colorimeter is positioned to the center of a display using one of the two attachments (2, 3). The colorimeter is used to provide measurement values to the PC and the application software when analyzing the color and luminance parameters of a display.

2. Front Attachment – Suction Cups: Attaches to the sensor side of the ColorPro V colorimeter. This attachment is used to mount the ColorPro V to the center of a display for measurement using the multiple suction cups to physically attach and hold the sensor.

3. Front Attachment – Felt: Attaches to the sensor side of the ColorPro V colorimeter. Use to cushion the sensor close to the display screen and direct light to the sensor without physically attaching the sensor to the screen.

Attachment Instructions: Lineup and insert the two notches on the attachment with the front side holes of the ColorPro V colorimeter. Gently turn a few degrees clockwise. To remove, turn a few degrees counterclockwise to release the notches and pull apart.

4. ColorPro V Cable Counter-weight: Attaches to the cable of the ColorPro V colorimeter. Lies on top or above the display screen as a counter-weight to hold the ColorPro V sensor in position.

Important: Use the #3 attachment and counter-weight when you fear that attaching the ColorPro V sensor to the display screen via the suction cup attachment may pose potential damage or be difficult to clean the surface upon completion of the calibration. Example: LCD screens, displays without protection front surfaces etc. When in doubt, use the felt attachment and counter-weight.

ColorPro

V

SENCORE

1.

2.

3.

ColorPro V Sensor/Attachments Description

4.

Page 8: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

6

CP5 Colorimeter Accessories

Sencore Part #s

Description Photo

39G523

Six foot null-modem serial cable with a 9-pin female D-sub on each end

1 supplied

(40B497) Insert (10B897) Case

Video suite case – storage for: the ColorPro V probe,

software, and USB cable a product from the VP400

VideoPro Multimedia Video Generator family

leads, serial adapter, power adapters and laptop PC storage space in the cover

1 supplied (VideoPro is a separate purchase)

39G1016

USB extension cable 10 ft USB extension cable for the ColorPro V. 1 supplied

USB-304 (39G1077)

USB to Serial Adapter (39G1077) A USB to serial cable with a 9-pin male D-sub on one end and a USB connector on the other end. 1 supplied

Page 9: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

7

OPERATION

Startup Connecting Equipment A USB port is required on your PC to connect and power the ColorPro V meter. If you connect the CP5 to your computer before you run the application software, most software utilities designed for use with this colorimeter, will automatically detect and initialize communication with the CP5.

Running the Application Software When you have connected the ColorPro V to the PC, double-click the application software shortcut icon on your PC desktop to start the program. Confirm communications with the sensor by making measurements with the application software or navigating to the software section dedicated to the meter device and establishing communications. Dark Calibration Dark calibration is not required on the CP5 ColorPro sensor. Calibration is automatic upon initialization with the application software. Operational Hints Ambient Light The ColorPro V is sensitive to ambient light. When you calibrate a display be sure to minimize or eliminate any room light. If room light reaches the front glass, it will enter the cabinet, be internally reflected, and come back out of the display, affecting the ColorPro readings. Use the Ambi-Block feature, if provided by the application software, to electronically remove low levels of ambient light from the measurement. You may also use some type of shroud to block light from the race of the display. Sensor Placement The position of the ColorPro V colorimeter on the display face should remain constant during measurement. A position change can cause color coordinate and luminance readings to change as the uniformity of the display is not constant. Sensor Cleaning To be sure that the ColorPro sensor with the suction cups attachment stays attached to the face of the display, be sure that both the face of the display and the suction cup surfaces are free from dirt and grime. Use a paper towel, moistened with liquid glass cleaner, to clean the rubber suction cup surfaces and the face of the display. Alternately, use pre-packaged alcohol-moistened prep pads.

Note: Do not use cleaning solutions other than liquid glass cleaner or isopropyl alcohol. Also, do not apply the cleaner directly to the sensor, only to the paper towel.

Page 10: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

8

Application Software – ColorPro by CalMan The CP5 Colorimeter requires a software utility to power the sensor, perform dark level calibration, and convert/present the measurement data to a user. Sencore recommends using the ColorPro by CalMAN calibration software. ColorPro by CalMAN is world’s most powerful display calibration software, with interface control of popular luminance/color meters and test pattern sources, that when combined with its unique guided workflow wizards provides unprecedented display calibration ease and accuracy.

ColorPro by CalMAN features:

Works with all popular industry light/color meters from Sencore (CP5, OTC100-HW), X-Rite, Klein, Photo Research, Orb and others, to meet all testing requirements and applications.

Provides automated video test pattern selections from many video generators and video

processors for fast multi-point grayscale, gamma and gamut testing. Controls reference optical discs for source calibration.

Page 11: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

9

Dynamic changes within the software, based upon the color meter, signal source, user profile, display type, and display input selections, insures an accurate calibration with little chance of user error.

Sencore exclusive step-by-step ColorPro wizards guide you through each calibration step.

A button press provides the proper test pattern(s), meter reading(s) and prompts you through each adjustment.

Full support and integration of 3D CMS Color Gamut and 3D multipoint

Grayscale/Gamma measurement tools to accurately calibrate Color Management systems (CMS) and multipoint RGB white balance features within displays or video processors.

Comprehensive graphs show the results of each measurement and adjustment, including

Single Temperature, Correlated Color Temperature, Luminance, Point Gamma, CIE Gamut, Delta E, Bulls-eye Color, RGB Level Tracking, RGB Color Balance, and Bulls-eye Grayscale. Customize or create your own graphs.

Create and customize your calibration reports by choosing and arranging the report

graphs plus adding your logo, letterhead or other graphics. User configurable Design mode makes it easy to change the calibration workflow, data

charts, note boxes, and layout to match the way you work. Help information, adapted for each software section, accompanies and clarifies each

menu and calibration step. Also includes hot key information. Calibration Optimization and Reporting Environment (CORE) engine with 64 bit double

precision floating math performs advanced color matrix operations and statistical analysis for unsurpassed measurement accuracy.

All measured, calculated, and target values are available in an extensive raw data table

for easy integration into a custom spreadsheet or report.

Page 12: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

10

USB to Serial Adapter – Driver Installation A USB to Serial Adapter is provided with the CP5 Colorimeter to provide a connection from the application software, such as ColorPro by CalMAN, to a Sencore video generator. This connection from the PC to the generator provides PC software controlled test pattern selection and subsequently automated display calibration testing. The USB to Serial Adapter requires a driver installation on the PC to operate properly. Improper driver installation may initially present an inability to select or control test pattern selection within the ColorPro by CalMan software or earlier CP6000 ColorPro software. The Sencore CP6000 ColorPro software provided a “Cannot Initialize the Video Generator” message when attempting to connect with the generator. The following instructions explain how to load the needed driver file to correct this problem. The ColorPro by CalMAN and original CP6000 software provides commands from the laptop’s serial port to operate a Sencore VP40x series Video Generator. This automation lets the software automatically select video test patterns corresponding with the calibration steps. To enable the software to control the video generator, a serial connection needs to be made from the computer to the generator. A serial cable (39G523) is supplied for this connection.

(39G523) A six foot null-modem serial cable with a 9-pin female D-sub on each end used to connect the PC to the Sencore VP40x generator. In the event that your laptop or PC does not contain a serial port, you can produce a virtual serial port with the use of USB to Serial Port Adapter (Sencore part # USB304). This adapter is supplied with the CP5, enabling use of the ColorPro from common USB laptop/computer ports. The current Sencore USB to Serial Adapter (USB304) is a Chip Technology (Model UC232R-10).

The USB to Serial Adapter requires a driver file be installed so that it can operate. When the driver file is not properly installed, the ColorPro 6000 software presents the message “Cannot initialize the video generator.”

Please note that this message also may indicate an improper serial cable or improper cable connection. A null modem serial cable is required and supplied (39G523). Check the cable and connections before proceeding.

The following are the instructions for installing a driver for the current USB to Serial Adapter. This adapter is sourced from FTDI Chip Technology (Model UC232R-10). Please note that other USB to Serial Adapter models from other companies would require a different driver file than listed in the following steps.

Page 13: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

11

The driver for the USB to Serial Adapter (USB304) is loaded into the installed program file folder for the ColorPro by CALMAN and/or ColorPro 6000 (since version 1.0.2). The needed driver file is loaded into the computer with the initial software installation. To locate and load the driver, navigate to My Computer C:/Program files/Sencore/ColorPro by CalMAN/Drivers or C:/Program files/Sencore/ColorPro 6000/Drivers and find the file labeled CDM 2.02.04.exe. Run the program to execute the installation of the driver.

The driver file is located in C:/Program files/Sencore/ColorPro by CALMAN/Drivers. Run the CDM 2.02.04.exe to execute the installation of the driver. To verify that the driver for the USB to Serial Adapter is installed correctly, take a look in the Device Manager. To get to the Device Manager, right click on My Computer, and go to Properties. Select the Hardware Tab, then the Device Manager tab. Select the Ports entry. You should see a listing such as USB Serial Cable (com 4). Any com number is fine and indicates a successful installation. If you do not see a listing in the port section, an additional step is required. Again within the Device Manager locate the Universal Serial Bus Controllers listing. Locate the FTDI USB Serial Converter listing and double click on it. Navigate to the Advanced tab, enable the VCP listing, and click the OK box. Unplug the USB to Serial Adapter and plug it back in to the PC’s USB port. This should enable a USB Serial Port (Com _) listing within the Device Manager in the Ports section.

Page 14: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

12

A USB Serial Port (Com 4) listing should appear when properly installed. Any Com number is OK.

Page 15: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

13

APPENDIX A

Measuring Light as the Eye Sees Light The human eye sees light through rod and cone type light receptors. There are red, green, and blue cone type receptors, while the rod receptors give us black and white vision, especially in small detail and low light.

The red, green, and blue cone receptors each have a different response to different colors (frequencies) of light. The average response of the human eye receptors to light across the visible spectrum is shown by the Standard Observer Response graph, developed by the International Commission on Illumination (CIE).

We call this tristimulus vision, as there are three types of receptors that individually send information to our brain and allow us to perceive different colors for the different mixtures of light energy within the visible spectrum. Tristimulus Measuring Devices

Page 16: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

14

Tristimulus color measurement devices are called colorimeters. This type of device works much the same as the human eye. A filter for each light sensor allows only a precise amount of each color of light to reach the sensor from the source to be measured. The response of each of the three filters is designed to mimic the response of one of the three types of cones in the average human eye.

The measurement information from each of the three light sensors allows us to compute a different measurement result for the different mixtures of light energy within the visible spectrum, in a way that duplicates the response of the human eye/brain combination. To accurately predict the response of the human eye to a combination of light energy at different frequencies, a tristimulus color measurement device must “see” light exactly the same way the human eye sees light. Display Technology Changes Display types that are now on the market produce spectrums of light that are often very different from each other and from the average spectrum produced by CRTs. Some displays produce strong peaks of light at color frequencies where CRTs produce much less light. The mixing results of these color frequencies still give us the familiar red, green, and blue that CRTs produce, but start with different levels of some color frequencies. The important difference between these new display types and the older CRT technology is that new technology displays may produce strong peaks of light at just about any color frequency.

Page 17: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

15

Courtesy Joe Kane Productions

This means that, if a colorimeter is to produce accurate measurement results for all display types, its optical filters must accurately duplicate the CIE standard observer response at all color frequencies, not just at the particular frequencies at which CRTs produce high light output. Its three color sensors must see light over the entire visible spectrum with the exact same amplitude response that the three color sensors of our eye see light. The CP6000 'ColorPro V' uses optical filters which are accurate to the CIE standard observer response at all frequencies of light, not just high output CRT frequencies, and accurately measures all displays of the past, present, and future.

Page 18: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

16

APPENDIX B CIE-UCS Chromaticity Coordinates

Color Temp CIE 1931 CIE 1976 UCS Kelvin Chromaticity Coordinates Chromaticity Coordinates x y u v

Black Body Values

3000 0.4369 0.4041 0.2482 0.5214

3500 0.4053 0.3907 0.2347 0.5113

4000 0.3804 0.3768 0.2249 0.5016

4500 0.3608 0.3635 0.2175 0.4927

5000 0.3451 0.3516 0.2118 0.4847

5500 0.3324 0.3410 0.2074 0.4775

6000 0.3221 0.3318 0.2039 0.4712

6500 0.3135 0.3236 0.2011 0.4655

7000 0.3064 0.3165 0.1988 0.4605

7500 0.3003 0.3103 0.1969 0.4561

8000 0.2952 0.3048 0.1952 0.4521

8500 0.2908 0.2999 0.1939 0.4486

9000 0.2869 0.2956 0.1927 0.4454

9300 0.2849 0.2932 0.1921 0.4436

9500 0.2836 0.2917 0.1917 0.4425

10000 0.2806 0.2883 0.1908 0.4399

10500 0.2780 0.2852 0.1900 0.4375

11000 0.2757 0.2824 0.1894 0.4354

11500 0.2736 0.2799 0.1888 0.4334

12000 0.2718 0.2776 0.1882 0.4317

CIE Standard Illuminants

D50 0.3457 0.3586 0.2100 0.4881

E 0.3333 0.3333 0.2105 0.4737

D55 0.3320 0.3480 0.2049 0.4810

D65 0.3127 0.3290 0.1989 0.4683

C 0.3101 0.3162 0.2013 0.4609

D75 0.2990 0.3149 0.1945 0.4585

u' = 4x / -2x + 12y + 3 v' = 9y / -2x +12y + 3

Page 19: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

17

WARRANTY AND SERVICE INFORMATION WARRANTY Your ColorPro Color Analyzer has been built to the highest quality standards in the industry. Each unit has been tested, aged under power for at least 24 hours and the every function and range was retested to insure it met all published specifications. Your instrument is fully protected with a 1-year warranty. SERVICE Save the original shipping carton and packing material for reuse should you ever need to ship your generator, or return it to the Sencore factory for repair. 1. If you are returning the unit to Sencore for service, call the service department at the number

below. The Service department Administrator will issue you a return authorization number (RUA). Please enclose the following information with the instrument when returned for service: owner's address, billing information, purchase order (if applicable), name and phone number of contact person, description of problem and reason for return.

2. Completely enclose the unit inside a plastic bag to protect its finish and prevent foreign

material from getting inside. 3. Cushion the unit equally on all sides with a minimum of 3 inches of padding material. Pack

the padding tightly enough to prevent the unit from shifting during shipment. 4. Seal all seams on the container with packing or strapping tape. 5. Send the packed unit to the following address (we recommend shipping via United Parcel

Service): SENCORE FACTORY SERVICE 3200 Sencore Drive Sioux Falls, SD 57107 Most service repairs are completed within 72 hours. If you need to ask about your unit, the Sencore Service Department phone number is: 1-800-SENCORE (736-2673) or FAX (605) 339-7032 Fill in for your records: Purchase date: _________________________ Serial Number (on opening power-on display) _______________________ Software version (in Utilities About menu) _________________________

Page 20: CP5 Tri-stimulus Contact Colorimeter - AVPro Store Manual Form7722.pdf · The CP5 tri-stimulus contact colorimeter works with application software, such as ColorPro by CalMAN, to

CP6000 ColorPro V Color Analyzer Form 7722 Operation Manual

18

3200 Sencore Drive Sioux Falls, SD 57107 Call 1-800-SENCORE (736-2673) Innovatively designed www.sencore.com with your time in mind. Form7722 Printed in U.S.A.