corrosion behaviour of ti–15mo alloy for dental implant applications

8
Corrosion behaviour of Ti–15Mo alloy for dental implant applications Satendra Kumar *, T.S.N. Sankara Narayanan * National Metallurgical Laboratory, Madras Centre, CSIR Complex, Taramani, Chennai 600 113, India 1. Introduction Titanium and titanium alloys are widely used for many biomedical applications due to their low density, excellent biocompatibility, corrosion resistance and mechanical proper- ties. 1–4 Among the various types of Ti alloys, Ti–6Al–4V has been the choice in many instances due to its ideal mechanical properties and corrosion resistance for implant applications. However, if the leaching of V and Al exceeds a threshold level, then it may cause peripheral neuropathy, osteomalacia and Alzheimer diseases. 5–7 The development of Ti–6Al–7Nb 8 and Ti–5Al–2.5Fe, 9 where Nb and Fe were substituted for V in Ti– 6Al–4V alloy, has not received much success as they still contains Al. 10,11 The titanium alloys developed in the early stage are mainly a + b type ones. Recently, mechanical biocompatibility of biomaterials is also regarded as an important criterion in the selection of biomaterial. Hence, the research and development on b-type titanium alloys, which are considered advantageous in terms of mechanical biocompatibility, are increasing. 12,13 Since the b-phase in Ti alloys exhibits a significantly lower modulus than the a-phase, the development of low modulus b-Ti alloys which retain a single b-phase microstructure on rapid cooling from high temperatures assumes significance. Several b-phase Ti alloys, having Nb, Ta, Zr and Mo as alloying elements (b-stabilizer elements) such as, Ti–12Mo–6Zr–2Fe and Ti–13Mo–7Zr–3Fe (‘TMZF’), 14,15 Ti–15Mo–5Zr–3Al, 16 Ti– 15Mo–3Nb–3O (‘TIMETAL 21Srx’), 16 Ti–14Nb–13Zr, 17 Ti–35Nb– journal of dentistry 36 (2008) 500–507 article info Article history: Received 7 November 2007 Received in revised form 7 March 2008 Accepted 25 March 2008 Keywords: Corrosion Titanium alloys Biocompatibility X-ray diffraction Electrochemical characterization Dental implant abstract The corrosion behaviour of Ti–15Mo alloy in 0.15 M NaCl solution containing varying concentrations of fluoride ions (190, 570, 1140 and 9500 ppm) is evaluated using potentio- dynamic polarization, electrochemical impedance spectroscopy (EIS) and chronoampero- metric/current–time transient (CTT) studies to ascertain its suitability for dental implant applications. The study reveals that there is a strong dependence of the corrosion resistance of Ti–15Mo alloy on the concentration of fluoride ions in the electrolyte medium. Increase in fluoride ion concentration from 0 to 9500 ppm shifts the corrosion potential (E corr ) from 275 to 457 mV vs. SCE, increases the corrosion current density (i corr ) from 0.31 to 2.30 mA/cm 2 , the passive current density (i pass ) from 0.07 to 7.32 mA/cm 2 and the double-layer capacitance (C dl ) from 9.63 10 5 to 1.79 10 4 F and reduces the charge transfer resistance (R ct ) from 6.58 10 4 to 6.64 10 3 V cm 2 . In spite of the active dissolution, the Ti–15Mo alloy exhibit passivity at anodic potentials at all concentrations of the fluoride ions studied. In dental implants since the exposure of the alloy will be limited only to its ‘neck’, the amount of Mo ions released from Ti–15Mo alloy is not likely to have an adverse and hence, in terms of biocompatibility this alloy seems to be acceptable for dental implant applications. The results of the study suggest that Ti–15Mo alloy can be a suitable alternative for dental implant applications. # 2008 Elsevier Ltd. All rights reserved. * Corresponding authors. Tel.: +91 44 2254 2077; fax: +91 44 2254 1027. E-mail addresses: [email protected] (S. Kumar), [email protected] (T.S.N. Sankara Narayanan). available at www.sciencedirect.com journal homepage: www.intl.elsevierhealth.com/journals/jden 0300-5712/$ – see front matter # 2008 Elsevier Ltd. All rights reserved. doi:10.1016/j.jdent.2008.03.007

Upload: satendra-kumar

Post on 25-Oct-2016

218 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Corrosion behaviour of Ti–15Mo alloy for dental implant applications

Corrosion behaviour of Ti–15Mo alloy for dental implantapplications

Satendra Kumar *, T.S.N. Sankara Narayanan *

National Metallurgical Laboratory, Madras Centre, CSIR Complex, Taramani, Chennai 600 113, India

j o u r n a l o f d e n t i s t r y 3 6 ( 2 0 0 8 ) 5 0 0 – 5 0 7

a r t i c l e i n f o

Article history:

Received 7 November 2007

Received in revised form

7 March 2008

Accepted 25 March 2008

Keywords:

Corrosion

Titanium alloys

Biocompatibility

X-ray diffraction

Electrochemical characterization

Dental implant

a b s t r a c t

The corrosion behaviour of Ti–15Mo alloy in 0.15 M NaCl solution containing varying

concentrations of fluoride ions (190, 570, 1140 and 9500 ppm) is evaluated using potentio-

dynamic polarization, electrochemical impedance spectroscopy (EIS) and chronoampero-

metric/current–time transient (CTT) studies to ascertain its suitability for dental implant

applications. The study reveals that there is a strong dependence of the corrosion resistance

of Ti–15Mo alloy on the concentration of fluoride ions in the electrolyte medium. Increase in

fluoride ion concentration from 0 to 9500 ppm shifts the corrosion potential (Ecorr) from�275

to �457 mV vs. SCE, increases the corrosion current density (icorr) from 0.31 to 2.30 mA/cm2,

the passive current density (ipass) from 0.07 to 7.32 mA/cm2 and the double-layer capacitance

(Cdl) from 9.63 � 10�5 to 1.79 � 10�4 F and reduces the charge transfer resistance (Rct) from

6.58 � 104 to 6.64 � 103 V cm2. In spite of the active dissolution, the Ti–15Mo alloy exhibit

passivity at anodic potentials at all concentrations of the fluoride ions studied. In dental

implants since the exposure of the alloy will be limited only to its ‘neck’, the amount of Mo

ions released from Ti–15Mo alloy is not likely to have an adverse and hence, in terms of

biocompatibility this alloy seems to be acceptable for dental implant applications. The

results of the study suggest that Ti–15Mo alloy can be a suitable alternative for dental

implant applications.

# 2008 Elsevier Ltd. All rights reserved.

avai lable at www.sc iencedi rec t .com

journal homepage: www. int l .e lsev ierhea l th .com/ journals / jden

1. Introduction

Titanium and titanium alloys are widely used for many

biomedical applications due to their low density, excellent

biocompatibility, corrosion resistance and mechanical proper-

ties.1–4 Among the various types of Ti alloys, Ti–6Al–4V has

been the choice in many instances due to its ideal mechanical

properties and corrosion resistance for implant applications.

However, if the leaching of V and Al exceeds a threshold level,

then it may cause peripheral neuropathy, osteomalacia and

Alzheimer diseases.5–7 The development of Ti–6Al–7Nb8 and

Ti–5Al–2.5Fe,9 where Nb and Fe were substituted for V in Ti–

6Al–4V alloy, has not received much success as they still

contains Al.10,11

* Corresponding authors. Tel.: +91 44 2254 2077; fax: +91 44 2254 1027E-mail addresses: [email protected] (S. Kumar), tsnsn@redi

0300-5712/$ – see front matter # 2008 Elsevier Ltd. All rights reservedoi:10.1016/j.jdent.2008.03.007

The titanium alloys developed in the early stage are mainly

a + b type ones. Recently, mechanical biocompatibility of

biomaterials is also regarded as an important criterion in the

selection of biomaterial. Hence, the research and development

on b-type titanium alloys, which are considered advantageous

in terms of mechanical biocompatibility, are increasing.12,13

Since the b-phase in Ti alloys exhibits a significantly lower

modulus than the a-phase, the development of low modulus

b-Ti alloys which retain a single b-phase microstructure on

rapid cooling from high temperatures assumes significance.

Several b-phase Ti alloys, having Nb, Ta, Zr and Mo as alloying

elements (b-stabilizer elements) such as, Ti–12Mo–6Zr–2Fe

and Ti–13Mo–7Zr–3Fe (‘TMZF’),14,15 Ti–15Mo–5Zr–3Al,16 Ti–

15Mo–3Nb–3O (‘TIMETAL 21Srx’),16 Ti–14Nb–13Zr,17 Ti–35Nb–

.ffmail.com (T.S.N. Sankara Narayanan).

d.

Page 2: Corrosion behaviour of Ti–15Mo alloy for dental implant applications

j o u r n a l o f d e n t i s t r y 3 6 ( 2 0 0 8 ) 5 0 0 – 5 0 7 501

7Zr–5Ta and Ti–34Nb–9Zr–8Ta (‘TNZT’),14,15 Ti–29Nb–4.6Zr–

13Ta,17 Ti–15Mo14, etc., were developed. Ho et al.18 and Oliveira

et al.19 have studied the structure and properties of a series of

binary Ti–Mo alloy with Mo content ranging up to 20 wt.%.

Based on the microstructural evolution and strengthening

mechanisms of Ti–15Mo alloy, Nag et al.15 have recommended

it as one of the promising biocompatible Ti alloy.

In the solution annealed condition, Ti–15Mo alloy has a

modulus of elasticity that is about two-third of the modulus of

Ti–6Al–4V alloy, along with considerably improved ductility

and fatigue properties.20,21 Ti–15Mo alloy has superior notch

sensitivity resistance when compared with conventional a + b

titanium alloys, Ti–6Al–4V ELI and Ti–6Al–7Nb.22 This is an

important material property to be considered for devices

subject to scratching and for components with holes, threads,

sharp edges, and in contact with other components in the

design of an implant. Considering the improved mechanical

properties and better corrosion resistance of the b-phase Ti–

15Mo alloy, it is worthwhile to evaluate its corrosion resistance

in a variety of simulated body fluids and is being explored by

our research group.

Titanium alloys have also been used in dental implants.

Since the oral environment could involve fluoride medium,

the degree of corrosion resistance offered by the Ti alloys in

fluoride containing medium becomes an important criterion

in the selection of a metallic biomaterial to be used in the oral

medium. Many researchers have confirmed the negative

influence of fluoride ions on the corrosion of titanium.23–31

During the last four decades, the use of gels and solutions

containing high concentrations of fluorides has indeed

become more frequent, reaching a noticeable impact on the

dental caries prevention.32 There has been a rapid increase in

the utilization of fluoridated prophylactic gels and rinses in

the odontological field. Many commercially available fluori-

nated gels contain very high concentrations of fluoride ions,

up to 10,000 ppm, with a pH, ranging between 7.2 and 3.2.28,33

Such a high concentration of fluoride ions would have a

deleterious influence on the corrosion resistance of Ti alloys.

Hence, it is essential to understand the electrochemical

behaviour of Ti alloys in fluoride medium.

The biocompatibility of Ti and its alloys in dental

application is decided based on the osseointegration response

and cell adhesion behaviour. Wang and Li34 have studied the

biocompatibility of Ti alloys for dental restoration. They have

found that Ti and its alloys were not mutagenic but no

significant difference in the cell attachment was observed.

Surface modification of Ti alloys, namely TiN coating, plasma

sprayed hydroxyapatite coatings, pulsed laser deposited of TiC

coating, etc., were explored to improve the biocompatibility of

Ti alloys for dental implant applications.35–37 It has been

established that the chemical properties of the oxide layer on

the Ti alloys play an important role in deciding its biocompat-

ibility with the surrounding tissues. If the medium is acidic

and contains appreciable amount of fluoride ions, then it

would lead to the formation of hydrofluoric acid (HF). When

the concentration of HF exceeds 30 ppm, the passive film on

the Ti alloy will get destructed and its mechanical properties

will be drastically affected.27 Alloying of certain elements

along with Ti is found to offer a better corrosion resistance in

fluoride containing medium.38,39 Alloying of 0.5 wt.% of Pt or

Pd with Ti promotes the formation of a passive film on the

titanium surface in presence of fluoride and offer better

corrosion resistance than that of CP–Ti, Ti–6Al–4V and Ti–6Al–

7Nb alloys.32 It is important to ascertain whether the b-phase

Ti–Mo alloys having 10–20 wt.% of Mo could offer a better

corrosion resistance in presence of fluoride. Alves et al.23 have

evaluated the corrosion resistance of Ti–10Mo alloy in 0.15 M

NaCl containing 570 ppm of fluoride ions and compared with

that of Ti–6Al–4V alloy. According to them, both Ti–10Mo and

Ti–6Al–4V alloys exhibit similar electrochemical character-

istics and the passive current density of Ti–10Mo alloy is

relatively lower.23 Much remains to be explored on the

corrosion behaviour of Ti–Mo alloys in fluoride containing

electrolytes. In this perspective, the present work aims to

study the corrosion behaviour of Ti–15Mo alloy in 0.15 M NaCl

containing varying concentrations of fluoride ions (190, 570,

1140 and 9500 ppm) and to predict its suitability for dental

implant applications.

2. Materials and methods

Ti–15Mo alloy (chemical composition, in wt.%—N: 0.01; C: 0.02;

H: 0.011; Fe: 0.012; O: 0.10; Mo: 15.04; Ti: balance) was used in

the present study. The microstructure of the Ti–15Mo alloy

was examined using a Leica DMLM optical microscope with

image analyzer software. For metallographic studies, the Ti–

15Mo alloy was mechanically polished using various grades of

SiC paper (60, 100, 220, 320, 400, 600, 1/0, 2/0, 3/0 and 4/0,

respectively) followed by polishing using a 0.3 mm diamond

paste and subsequently etched for 20–30 s approximately

using a mixture of 15 vol.% HNO3, 5 vol.% HF and balance

water. This etching solution has been used earlier by Ho et al.18

for evaluating the microstructure of Ti–Mo alloys containing

up to 20 wt% Mo. The structural characteristic of the Ti–15Mo

alloy was evaluated by X-ray diffraction (XRD) measurement,

using Cu Ka radiation. The microhardness was determined

using a Leica VMHTMOT microhardness tester at a load of

200 g applied for 15 s.

The corrosion behaviour of Ti–15Mo alloy was evaluated

using 0.15 M NaCl solution as the base electrolyte to which

varying concentrations of fluoride ions (190, 570, 1140 and

9500 ppm), was added (as NaF) to study the influence of

fluoride. Nakagawa et al.28 have suggested that many

commercially available fluorinated gels may contain up to

10,000 ppm of fluoride ions, with a pH ranging from 7.2 to 3.2.

Alves et al.23 have evaluated the corrosion resistance of Ti–

10Mo and Ti–6Al–4V alloys in 0.15 M NaCl containing 570 ppm

of fluoride ion. Based on these reports it was decided to use

0.15 M NaCl as the base electrolyte and to represent a wide

range, the concentration of fluoride ions were chosen as 190,

570, 1140 and 9500 ppm.

Potentiodynamic polarization, electrochemical impedance

spectroscopy (EIS) and chronoamperometric/current–time

transient (CTT) studies were carried out using a potentio-

stat/galvanostat/frequency response analyzer of ACM instru-

ments (model: Gill AC) to assess the corrosion performance of

the alloy. For corrosion studies, the Ti–15Mo alloy was

mechanically polished using various grades of SiC paper,

rinsed with deionized water, pickled using a mixture of

Page 3: Corrosion behaviour of Ti–15Mo alloy for dental implant applications

Fig. 1 – Microstructure of the Ti–15Mo alloy.

j o u r n a l o f d e n t i s t r y 3 6 ( 2 0 0 8 ) 5 0 0 – 5 0 7502

35 vol.%, HNO3, 5 vol.% HF and balance water at 40 8C,

thoroughly rinsed in deionized water and dried using a

stream of compressed air. The cleaned Ti–15Mo alloy forms

the working electrode while a saturated calomel electrode

(SCE) and a graphite rod were used as the reference and

auxiliary electrodes, respectively. The electrodes were placed

within a flat cell in such a way that only 1 cm2 area of the

working electrode was exposed to the electrolyte solution.

Potentiodynamic polarization measurements were carried out

in the potential range from �250 mV in the cathodic direction

to +250 mV in the anodic direction from open-circuit potential

(OCP) vs. SCE at a scan rate of 100 mV/min and the corrosion

potential (Ecorr) and corrosion current density (icorr) were

determined from the polarization curves using Tafel extra-

polation method. The potentiodynamic polarization experi-

ments were also carried out in the potential range of �250 to

+3000 mV with respect to OCP vs. SCE at a scan rate of 100 mV/

min, to study the passivation behaviour and the average

passive current density (ipass) was calculated from the

polarization curves. EIS study was conducted by applying a

sinusoidal signal with an amplitude of 32 mV and the

electrode response was analyzed in the frequency range

between 10,000 and 0.01 Hz in 0.15 M NaCl containing varying

concentrations of fluoride ions (190, 570, 1140 and 9500 ppm),

at their respective OCP. The charge transfer resistance (Rct)

and double-layer capacitance (Cdl) values were determined

from the Nyquist plot by fitting the data using Boukamp

software. The CTT studies of Ti–15Mo alloy in 0.15 M NaCl

containing varying concentrations of fluoride ions (190, 570,

1140 and 9500 ppm), were performed at three different

potentials, namely, +500, +1250 and +2000 mV vs. SCE for

1 h. The potentiodynamic polarization, EIS and CTT studies

were repeated at least three times so as to ensure reprodu-

cibility of the test results.

Fig. 2 – Potentiodynamic polarization curve of Ti–15Mo

alloy in 0.15 M NaCl with varying concentrations of

fluoride ions (190, 570, 1140 and 9500 ppm): scan rate—

100 mV/min (potential in mV vs. SCE).

3. Results

3.1. Microstructure, structure and microhardness of theTi–15Mo alloy

The microstructure of the Ti–15Mo alloy (Fig. 1) reveals

equiaxed b-grains as the dominant phase, which is homo-

geneous and evenly distributed. The X-ray diffraction pattern

of the alloy indicates that only the b-phase is retained in the

structure (figure not shown). The microhardness of Ti–15Mo

alloy is found to be 238 � 5 HV0.2.

3.2. Potentiodynamic polarization studies of the Ti–15Moalloy

Potentiodynamic polarization studies of the Ti–15Mo alloy

were conducted in the potential range of�250 to +250 mV with

respect to OCP vs. SCE at a scan rate of 100 mV/min to observe

the effect of fluoride ions on the corrosion behaviour (Fig. 2).

Though the shape of the curves is quite similar, the active

region of the curves is extended to higher current region in

presence of fluoride ions. The corrosion potential (Ecorr) and

corrosion current density (icorr), calculated using Tafel extra-

polation method, are compiled in Table 1. There is a cathodic

shift in Ecorr from �275 to �457 mV vs. SCE and a remarkable

increase in icorr from 0.31 to 2.30 mA/cm2 with increase in

fluoride ion concentration from 0 to 9500 ppm.

To observe the effect of fluoride ion on the passivity of

Ti–15Mo alloy, the potentiodynamic polarization studies

were performed in the potential range of �250 to +3000 mV

with respect to OCP vs. SCE at a scan rate of 100 mV/min.

The anodic branch of the polarization curve exhibits an

active–passive transition in all the cases (Fig. 2). The average

passive current density (ipass) of the Ti–15Mo alloy is

measured from the polarization curves and compiled in

Table 1.

Page 4: Corrosion behaviour of Ti–15Mo alloy for dental implant applications

Table 1 – Corrosion potential (Ecorr), corrosion current density (icorr) and average passive current density (ipass) of Ti–15Moalloy in 0.15 M NaCl containing varying concentrations of fluoride ions

Electrolyte medium Corrosion potential,Ecorr (mV vs. SCE)

Corrosion currentdensity, icorr (mA/cm2)

Average passive currentdensity, ipass (mA/cm2)

0.15 M NaCl �275 0.31 0.07

0.15 M NaCl + 190 ppm of F� �282 0.65 0.11

0.15 M NaCl + 570 ppm of F� �376 1.22 0.42

0.15 M NaCl + 1140 ppm of F� �409 1.77 0.93

0.15 M NaCl + 9500 ppm of F� �457 2.30 7.32

Fig. 3 – Bode impedance and phase angle plots of Ti–15Mo

alloy in 0.15 M NaCl with varying concentrations of

fluoride ions (190, 570, 1140 and 9500 ppm): (*) 0 ppm; ( )

190 ppm; (*) 570 ppm; (&) 1140 ppm; (4) 9500 ppm of FS.

j o u r n a l o f d e n t i s t r y 3 6 ( 2 0 0 8 ) 5 0 0 – 5 0 7 503

3.3. Electrochemical impedance spectroscopy studies of theTi–15Mo alloy

Fig. 3 shows the Bode impedance and Bode phase angle plots of

the Ti–15Mo alloy in 0.15 M NaCl containing varying concen-

tration of fluoride ions (190, 570, 1140 and 9500 ppm), at their

respective OCP vs. SCE. The Bode impedance plot of Ti–15Mo

alloy clearly reveals the dependence of the impedance values

on the fluoride ion concentration in the base solution (0.15 M

NaCl), which is also reflected in the change in diameter of the

semicircle in the Nyquist plots (figure not shown). The Rct and

Cdl, determined from the Nyquist plots after fitting the data

using Boukamp software, are presented in Table 2. The Rct is

decreased from 6.58 � 104 to 6.64 � 103 V cm2 whereas the Cdl

Table 2 – Charge transfer resistance (Rct) and double-layer capacitance (Cdl) of the Ti–15Mo alloy in 0.15 M NaClcontaining varying concentrations of fluoride ions

Electrolyte medium Rct (V cm2) Cdl (F)

0.15 M NaCl 6.58 � 104 9.63 � 10�5

0.15 M NaCl + 190 ppm of F� 4.76 � 104 8.40 � 10�5

0.15 M NaCl + 570 ppm of F� 2.16 � 104 1.31 � 10�4

0.15 M NaCl + 1140 ppm of F� 1.19 � 104 1.57 � 10�4

0.15 M NaCl + 9500 ppm of F� 6.64 � 103 1.79 � 10�4

is increased from 9.63 � 10�5 to 1.79 � 10�4 F with increase in

fluoride ion concentration from 0 to 9500 ppm.

The Bode phase angle plots (Fig. 3) indicate that the phase

angle shift is nearly�308 at low frequency in 0.15 M NaCl and it

reduces to �58 with increase in fluoride ion concentration

from 0 to 9500 ppm in the base solution. However, at

intermediate frequencies, the phase angle maximum is

around �658 in 0.15 M NaCl and remained constant over a

wide range of frequency. The phase angle shift increases from

approximately �658 to �708 with increase in fluoride ion

concentration and the frequency range in which the phase

angle remains constant is also reduced. When the fluoride ion

concentration in the base solution is higher (9500 ppm), the

phase angle shift starts to reduce immediately after reaching

the peak value of �708. The flat portion of the phase-angle

spectrum reduces with increase in fluoride ion concentration

in 0.15 M NaCl solution.

3.4. Chronoamperometric/current–time transient studiesof the Ti–15Mo alloy

CTT studies of Ti–15Mo alloy in 0.15 M NaCl containing

varying concentrations of fluoride ions (190, 570, 1140 and

9500 ppm) were performed at three different potentials,

namely, +500, +1250 and +2000 mV vs. SCE for 1 h. The CTT

curve of Ti–15Mo alloy in 0.15 M NaCl containing 0, 190, 570

and 1140 ppm of fluoride ions, at +1250 mV vs. SCE is

presented as a representative curve in Fig. 4. In the absence

Fig. 4 – Current–time transient curves of Ti–15Mo in 0.15 M

NaCl with varying concentrations of fluoride ions (190, 570

and 1140 ppm) at +1250 mV vs. SCE.

Page 5: Corrosion behaviour of Ti–15Mo alloy for dental implant applications

Table 3 – Average steady state current density of the Ti–15Mo alloy in 0.15 M NaCl containing varying concentrations offluoride ions at different impressed potentials

Electrolyte medium Average steady state current density at different impressed potentials(mA/cm2)

+500 mV +1250 mV +2000 mV

0.15 M NaCl 0.003 0.012 0.020

0.15 M NaCl + 190 ppm of F� 0.018 0.027 0.121

0.15 M NaCl + 570 ppm of F� 0.086 0.102 0.184

0.15 M NaCl + 1140 ppm of F� 0.168 0.530 0.732

0.15 M NaCl + 9500 ppm of F� 1.305 4.652 6.756

j o u r n a l o f d e n t i s t r y 3 6 ( 2 0 0 8 ) 5 0 0 – 5 0 7504

of fluoride ions as well as in presence of 190 ppm of fluoride

ions, the CTT curves exhibit a rapid decrease in current

density followed by a slow decay to attain a steady state. In

presence of 1140 ppm of fluoride ions, the CTT curves exhibit

a decrease in current density during the initial period,

reaching a minimum followed by an increase in current

density, reaching a peak current, which subsequently decays

to attain a steady state. In presence of 9500 ppm of fluoride

ions, the CTT curves exhibit a sharp increase in current

density during the initial periods which then decays to attain

steady state, at all applied potentials (figure not shown). The

average steady state current density calculated at all the three

potentials is complied in Table 3.

4. Discussion and conclusions

4.1. Microstructure, structure and microhardness of theTi–15Mo alloy

The retention of the b-phase in Ti alloys with higher Mo

content has been reported earlier by Davis et al.40 Ho et al.18

have reported that Ti–Mo alloys having 9 wt.% Mo have a

significant amount of equiaxed b-phase whereas in alloys

containing equal to or greater than 10 wt.% Mo, the b-phase

becomes the only dominant phase. Bania41 has also

reported that a minimum of 10 wt.% Mo is needed to fully

stabilize b-phase at room temperature. Ho et al.18 and

Oliveira et al.19 have suggested that the crystal structure of

the Ti–Mo alloys is sensitive to the Mo content. Based on the

XRD measurements, they have confirmed that a significant

retention of the b-phase for the Ti–Mo alloy containing

10 wt.% Mo, while in alloys having higher Mo concentrations

(15 and 20 wt.%) only the b-phase is retained. Ho et al.18

have reported that the hardness of Ti–Mo alloys (containing

6–20 wt.% Mo), in general, is higher than that of CP–Ti.

They have found that the hardness of Ti–7.5Mo alloy is

lower by 10.5% than that of Ti–6Al–4V alloy whereas the Ti–

20Mo alloy had a higher hardness than Ti–6Al–4V alloy.

The results of the present study reveal that the microhard-

ness of the Ti–15Mo alloy is lesser than the observation

made by Ho et al.18 As several factors could influence the

microhardness of Ti–Mo alloys, which include solid solution

strengthening, precipitation hardening, strain aging, grain

size and crystal structure/phase (a, a + b, b),18 the inter-

pretation of the measured microhardness values becomes

quite complex.

4.2. Potentiodynamic polarization studies of the Ti–15Moalloy

The shift in the active region of the polarization curves

towards higher current region (Fig. 2) suggesting the negative

influence of fluoride ions on the corrosion resistance of Ti–

15Mo alloy. The cathodic shift in Ecorr from �275 to �457 mV

vs. SCE and a remarkable increase in icorr from 0.31 to 2.30 mA/

cm2 with increase in fluoride ion concentration from 0 to

9500 ppm confirm this phenomenon. The negative influence

of fluoride ions on the corrosion resistance of Ti alloys has also

been confirmed by many researchers.23–31 Thomson-Neal

et al.42 have reported that 3 ppm NaF is sufficient to tarnish

the surface of Ti while autoclaving the Ti implants. Oshida

et al.33 have reported that commercially available fluoride

treatment agent could cause discoloration of Ti–6Al–4V

brackets. According to Schutz and Thomas,43 20 ppm of NaF

may destroy the protective oxide layer on Ti while Nakagawa

et al.27 have reported such an occurrence at 30 ppm of NaF.

Hence, the increase in the active region of Ti–15Mo alloy in

presence of fluoride ions is due to the formation of a porous or

defective oxide layer that reduces its corrosion protectiveness,

which is reflected in the Ecorr and icorr values.

The active–passive transition observed in the anodic

branch of the polarization curves, both in the absence and

in presence of 190, 570, 1140 and 9500 ppm of fluoride ions,

suggests that the presence of fluoride ions in 0.15 M NaCl did

not hinder the formation of passive oxide film on the surface

of the Ti–15Mo alloy. The occurrence of active–passivation

transition of Ti alloys in presence of fluoride ions has also been

reported earlier.30,44 Huang et al.30 have observed the active–

passive transition of Ti–6Al–4V alloy in acidic artificial saliva

(pH: 5) except when the NaF concentration is 0.5% whereas

Schmidt et al.44 have observed a similar phenomenon in

lactated Ringer’s serum (pH: 6.5) in presence of fluoride. The

occurrence of active–passive transition depends mainly on the

concentration of HF in the medium. According to Nakagawa

et al.,27 in acidic conditions, 30 ppm of HF could lead to the

destruction of the passive film on Ti surface. Since, the pH of

the electrolyte used in the present study is 6.0, the concentra-

tion of HF will be much lower than 30 ppm. Kwon et al.31 have

estimated the concentration of HF in solutions of 0.05, 0.1, and

0.2% NaF, i.e., 950, 1900 and 3800 ppm of fluoride ions, at pH 6

is of the order of only 12–13 ppm and confirmed that at these

HF concentrations the passive oxide layer on Ti alloys could be

sustained. Hence, the occurrence of active–passive transition

in 0.15 M NaCl as well as in presence of 190, 570, 1140 and

Page 6: Corrosion behaviour of Ti–15Mo alloy for dental implant applications

j o u r n a l o f d e n t i s t r y 3 6 ( 2 0 0 8 ) 5 0 0 – 5 0 7 505

9500 ppm of fluoride ions is due to the lower concentrations of

HF and confirms the ability of Ti–15Mo alloy to impart

passivity. The passive films are believed to be composed of

one or more protective oxide films.

The ipass is found to increase from 0.07 to 7.32 mA/cm2 with

increase in fluoride ion concentration from 0 to 9500 ppm. This

observation suggests that the oxide film formed on the surface

is having less insulating property, thus allowing the passage of

higher current for oxidation of species in the electrolyte.

Huang et al.30 have also reported about the increase in passive

current density of Ti–6Al–4V with increase in NaF concentra-

tion in the electrolyte. The increase in ipass of Ti–15Mo alloy

may be due to the dissolution of the alloying element, namely

Mo, induced by fluoride ions present in the medium.

4.3. Electrochemical impedance spectroscopy studies of theTi–15Mo alloy

The decrease in Rct from 6.58 � 104 to 6.64 � 103 V cm2 and

increase in Cdl from 9.63 � 10�5 to 1.79 � 10�4 F suggests the

negative influence of fluoride ions on the corrosion resistance

of Ti–15Mo alloy and support the observations made by

potentiodynamic polarization studies. The phase angle max-

imum and the phase shifts observed in the Bode phase angle

plots is typical for a passive surface indicating good corrosion

resistance with a near capacitive behaviour and indicate the

difficulty in charge transfer process. The two peaks observed

in phase angle plots indicate the involvement of two

relaxation time constants. The two distinct capacitive beha-

viours can be attributed due to the bi-layered oxide surface

consisting of a porous outer layer and a barrier inner layer.45,46

The change in the shape of the Bode plots as a function of

fluoride ion concentration is mainly caused by diffusion

phenomena due to dissolution reaction. Therefore, the ease of

charge transfer is more with the increase of fluoride ion

concentration in the base solution. Besides, at higher fluoride

ion concentration (9500 ppm) with 0.15 M NaCl, the phase

angle shift values are higher over the frequency range of 1000–

100 Hz. This is believed to be due to the formation of passive

oxide film consisting of oxides of titanium and molybdenum

immediately after the fast interaction of the alloy with fluoride

ion whereas the interaction is delayed in the absence or

presence of lower concentrations of fluoride ions. The passive

film formed in presence of 9500 ppm of fluoride ions may not

be stable for a longer time and starts dissolving due to the

negative effect of higher concentration of fluoride ion. This

could be one of the reasons for the decrease in phase angle just

after reaching the peak value. The changes in phase angle

confirm the negative influence of fluoride ion and support the

observation made in potentiodynamic polarization studies.

4.4. Chronoamperometric/current–time transient studiesof the Ti–15Mo alloy

The rapid decrease in current density followed by a slow decay

to attain a steady state observed in the CTT curves in the

absence as well as in presence of 190 ppm of fluoride ions can

be explained based on the decrease in active area due to the

growth of a passive film. The decrease in current density

during the initial period, reaching a minimum followed by an

increase in current density, reaching a peak current, which

subsequently decays to attain a steady state observed in

presence of 1140 ppm of fluoride ions can be due to the

instability of passive oxide film that leads to the exposure of

active surface area and the consequent repassivation or re-

precipitation of the oxides on the surface. The maximum of

the CTT curves may be associated with the active/passive

transition peak, which is commonly attributed to electro-

oxidation process or new phase formation. Alves et al.23 have

also observed a similar trend in the CTT curves of Ti–10Mo

alloy in 0.15 M NaCl containing 570 ppm of fluoride ions. The

electro-oxidation process may be mainly related to titanium or

titanium species along with the respective alloying element

species present in the initially formed film.

The extent of change in current as a function of time and

the average steady state current density has a strong

dependence on the concentration of fluoride ions in the

electrolyte medium and the applied potentials. The increase in

the average steady state current density with increase in

fluoride ion concentration and increase in applied potentials is

due to the dissolution of the protective oxide film as well as the

substrate (Table 3). It is important to note that in spite of the

increase in extent of dissolution with increase in applied

potential, the Ti–15Mo alloy exhibit passivity at all the three

potentials, namely, +500, +1250 and +2000 mV vs. SCE. Alves

et al.23 have reported that the passive current density of Ti–

10Mo alloy is lower than Ti–6Al–4V alloy in 0.15 M NaCl

containing 570 ppm of fluoride ions. Al-Mayouf et al.39 have

also reported that corrosion current density of Ti–6Al–4V alloy

is lower than CP–Ti and Ti–30Cu–10Ag alloy in presence of

9500 ppm of fluoride ions. The results of the current–time

transient of the present study further confirm the observa-

tions made earlier by other researchers.23,39

The CTT studies reveal greater extent of dissolution of the

Ti–15Mo alloy in presence of fluoride ions. Hence, the toxicity

of Mo should be considered as a criterion in the choice of Ti–

15Mo alloy to be used for dental implant application. Some

positive as well as negative influence of Mo has been reported

in the published literature.47–49 Mo is considered to be

instrumental in regulating the pH balance in the body. It

has been shown to act as a cofactor for a limited number of

enzymes in humans.47 However, excessive intake of Mo, in

rare cases, has been found to cause joint pain and swelling, leg

cramps and, a burning sensation upon urination. It has been

reported that 10–100 ppm of Mo ion could alter cell metabolic

functions.48 The recommended dietary allowance (RDA) for

adult men and women is 45 mg/day. The tolerable upper intake

level is 2 mg/day, a level based on impaired reproduction and

growth in animals.49 Fortunately, the exposure of dental

implants to fluoride ion containing gels, etc., would be limited

only to the ‘neck’ of the implant and for very short periods of

time. Hence, the amount of Mo ions released from Ti–15Mo

alloy is not likely to have an adverse effect and in terms of

biocompatibility Ti–15Mo alloy seem to be acceptable for

dental implant applications.

The corrosion behaviour of Ti–15Mo alloy in 0.15 M NaCl

solution containing varying concentrations of fluoride ions (190,

570, 1140 and 9500 ppm), was evaluated using potentiodynamic

polarization, electrochemical impedance spectroscopy and

chronoamperometric studies to ascertain the protective ability

Page 7: Corrosion behaviour of Ti–15Mo alloy for dental implant applications

j o u r n a l o f d e n t i s t r y 3 6 ( 2 0 0 8 ) 5 0 0 – 5 0 7506

of the alloy in presence of fluoride ion. The microstructure,

structural characteristics and hardness were also evaluated.

The study leads to the following conclusions:

� T

he Ti–15Mo alloy exhibits the presence of only the

equiaxed b-phase, which is homogeneous and evenly

distributed and its microhardness is 238 � 5 HV0.2.

� T

he active–passive transition is observed in presence of all

concentrations of fluoride ions. However, the active region is

extended to higher current region in presence of fluoride

ions. In spite of the active dissolution in presence of fluoride

ions, the Ti–15Mo alloy exhibit passivity at anodic poten-

tials.

� T

here is a strong dependence of the Ecorr, icorr, ipass, Rct, Cdl

and the average steady state current density values of the

Ti–15Mo alloy on the concentration of fluoride ions in the

electrolyte medium. Increase in fluoride ion concentration

increases the icorr, ipass, average steady state current density

and Cdl values, causes a cathodic shift in Ecorr and a decrease

in Rct values, suggesting the negative influence of fluoride

ion and a decrease in corrosion protective ability of Ti–15Mo

alloy. The average steady state current density also exhibits

a linear dependence on the applied potential, suggesting the

dissolution of the protective oxide film as well as the

substrate.

� A

s the exposure of dental implants to fluoride ion containing

gels, etc., would be limited only to the ‘neck’ of the implant

and for very short periods of time, the amount of Mo ions

released from Ti–15Mo alloy is not likely to have an adverse

effect. Hence, in terms of biocompatibility Ti–15Mo alloy

seem to be acceptable for dental implant applications.

� B

ased on the results of the study, Ti–15Mo alloy can be a

suitable alternative for dental implant applications.

Acknowledgement

The authors express their sincere thanks to Prof. S.P.

Mehrotra, Director, National Metallurgical Laboratory, Jam-

shedpur, for his keen interest and permission to publish this

paper.

r e f e r e n c e s

1. Kovacs P, Davidson JA. Chemical and electrochemicalaspects of the biocompatibility of titanium and its alloys. In:Brown SA, Lemons JE, editors. Medical applications of titaniumand its alloys: the materials and biological issues, ASTM STP1272. 1996:163–78.

2. Hunt JA, Stoichet M. Biomaterials: surface interactions.Current Opinion in Solid State and Materials Science2001;5:161–2.

3. Williams DF, Titanium and titanium alloys.Williams DF,editor. Biocompatibility of clinical implant materials, vol. 2. CRCPress; 1981. p. 9–44.

4. Long MJ, Rack HJ. Titanium alloys in total jointreplacement—a materials science perspective. Biomaterials1998;19:1621–39.

5. Rao S, Ushida T, Tateishi T, Okazaki Y, Asao S. Effect of Ti,Al, and V ions on the relative growth rate of fibroblasts

(L929) and osteoblasts (MC3T3-E1) cells. Bio-Medical MaterialsEngineering 1996;6:79–86.

6. Walker PR, Leblanc J, Sikorska M. Effects of aluminium andother cations on the structure of brain and liver chromatin.Biochemistry 1990;28:3911–5.

7. Yumoto S, Ohashi H, Nagai H, Kakimi S, Ogawa Y, Iwata Y,et al. Aluminum neurotoxicity in the rat brain. InternationalJournal of PIXE 1992;2:493–504.

8. Semlitsch MF, Weber H, Streicher RM, Schon R. Jointreplacement components made of hot-forged andsurface-treated Ti–6Al–7Nb alloy. Biomaterials 1992;13:781–8.

9. Zwicker J, Etzold U, Moser T. Abrasive properties of oxidelayers on Ti–Al15–Fe2.5 in contact with high densitypolyethylene. Titanium’84 science and technology, vol. 2.Munich: Deutsche Gesellschaft Fur Metallkunde EV; 1985. p.1343–50.

10. Perl DP, Brody AR. Alzheimer’s disease: X-ray spectrometricevidence of aluminum accumulation in neurofibrillarytangle-bearing neurons. Science 1980;208:297–9.

11. Crapper DR, McLachlan DR, Farnell B, Galin H, Karlik S,Eichhorn G, et al. Aluminum in human brain disease. In:Sarkar B, editor. Biological aspects of metals and metals-relateddiseases. New York: Raven Press; 1993. p. 209–18.

12. Kuroda D, Niinomi M, Morinaga M, Kato Y, Yashiro T.Design and mechanical properties of new b-type titaniumalloys for implant materials. Materials Science and EngineeringA 1998;243:244–9.

13. Ahmed T, Lomg M, Silvestri J, Ruiz C, Rack HJ. A newlow modulus, biocompatible titanium alloy. In:Blenkinsop PA, Evans WJ, Flower HM, editors. Titanium’95:science and technology. The Institute of Materials; 1996. p.1760–7.

14. Nag S, Banerjee R, Fraser HL. Comparison of microstructuralevolution in Ti–Mo–Zr–Fe and Ti–15Mo biocompatible alloys.Journal of Materials Science Materials in Medicine 2005;16:679–85.

15. Nag S, Banerjee R, Fraser HL. Microstructural evolution andstrengthening mechanisms in Ti–Nb–Zr–Ta, Ti–Mo–Zr–Feand Ti–15Mo biocompatible alloys. Materials Science andEngineering C 2005;25:357–62.

16. Steinemann SG, Ausli PAM, Szmuklermoncler S, SemlitschM, Pohler O, Hintermann HE, et al. Beta-titanium alloy forsurgical implants. Beta titanium in the 1990’s. Warrendale,PA: The Mineral, Metals and Materials Society; 1993. p. 2689–96.

17. Mishra AK, Davidson JA, Kovacs P, Poggie RA. Ti–13Nb–13Zr:a new low modulus, high strength, corrosion resistant nearbeta alloy for orthopaedic implants. Beta titanium in the1990’s. Warrendale, PA: The Mineral, Metals and MaterialsSociety; 1993. p. 61–72.

18. Ho WF, Ju CP, Chern Lin H. Structure and properties of castbinary Ti–Mo alloys. Biomaterials 1999;20:2115–22.

19. Oliveira NTC, Aleixo G, Caram R, Guastaldi AC.Development of Ti–Mo alloys for biomedical applications:microstructure and electrochemical characterization.Materials Science and Engineering A 2007;452–453:727–31.

20. IMI Titanium 205, Alloy data sheet, IMI Titanium Limited,Birmingham, England.

21. Steinemann S, et al. Beta-Titanium Alloy for SurgicalImplants. Seventh World Conference on Titanium. 1992.

22. Disegi J. Titanium alloys for fracture fixation implants.Injury International Journal of the Care of the Injured 2000;31:S-D2-6.

23. Alves APR, Santana FA, Rosa LAA, Cursino SA, Codaro EN. Astudy on corrosion resistance of the Ti–10Mo experimentalalloy after different processing methods. Materials Scienceand Engineering C 2004;24:693–6.

Page 8: Corrosion behaviour of Ti–15Mo alloy for dental implant applications

j o u r n a l o f d e n t i s t r y 3 6 ( 2 0 0 8 ) 5 0 0 – 5 0 7 507

24. Wilhelmsen W, Grande AP. The influence of hydrofluoricacid and fluoride ion on the corrosion and passive behaviorof titanium. Electrochemica Acta 1987;32:1469–72.

25. Toumelin-Chemla F, Rouelle F, Burdairon G. Corrosiveproperties of fluoride-containing odontologic gels againsttitanium. Journal of Dentistry 1996;24:109–15.

26. Reclaru L, Meyer JM. Effects of fluorides on titanium andother dental alloys in dentistry. Biomaterials 1998;19:85–92.

27. Nakagawa M, Matsuya S, Shiraishi T, Ohta M. Effect offluoride concentration and pH on corrosion behavior oftitanium for dental use. Journal of Dental Research1999;78:1568–72.

28. Nakagawa M, Matsuya S, Udoh K. Corrosion behavior ofpure titanium and titanium alloys in fluoride-containingsolutions. Dental Material Journal 2001;20:305–14.

29. Nakagawa M, Matsuya S, Udoh K. Effect of fluoride anddissolved oxygen concentrations on the corrosion behaviorof pure titanium and titanium alloys. Dental Material Journal2002;21:83–92.

30. Huang HH. Effects of fluoride and albumin concentration onthe corrosion behavior of Ti–6Al–4V alloy. Biomaterials2003;24:275–82.

31. Kwon YH, Seol HJ, Kim H, Hwang KJ, Lee SG, Kim KH. Effectof acidic fluoride solution on b-titanium alloy wire. Journal ofBiomedical Materials Research Part B Applied Biomaterials2005;73:285–90.

32. Stookey GK. Critical evaluation of the composition and useof topical fluorides. Journal of Dental Research 1990;69:805–12.

33. Oshida Y, Sellers CB, Mirza K, Farzin-Nia F. Corrosion ofdental metallic materials by dental treatment agents.Materials Science and Engineering C 2005;25:343–8.

34. Wang RR, Li Y. In-vitro evaluation of biocompatibility ofexperimental titanium alloys for dental restorations. TheJournal of Prosthetic Dentistry 1998;80:495–500.

35. Huang HH, Hsu CH, Pan SJ, He JL, Chen CC, Lee TL. Corrosionand cell adhesion behaviour of TiN-coated and ion-nitridedtitanium for dental applications. Applied Surface Science2005;244:252–6.

36. Lavos-Valereto IC, Wolynec S, Deboni MCZ, Knig Jr B. In-vitroand in-vivo biocompatibility testing of Ti–6Al–7Nb alloy withand without plasma-sprayed hydroxyapatite coatings. Journalof Biomedical Materials Research 2001;58:727–33.

37. Brama M, Rhodes N, Hunt J, Ricci A, Teghil R, Migliaccio S,et al. Effect of titanium carbide coating on theosseointegration response in vitro and in vivo. Biomaterials2007;28:595–608.

38. Nakagawa M, Matono Y, Matsuya S, Udoh K, Ishikawa K.The effect of Pt and Pd alloying additions on the corrosionbehavior of titanium in fluoride-containing environments.Biomaterials 2005;26:2239–46.

39. Al-Mayouf AM, Al-Swayih AA, Al-Mobarak NA, Al-Jabab AS.Corrosion behavior of a new titanium alloy for dentalimplant applications in fluoride media. Materials Chemistryand Physics 2004;86:320–9.

40. Davis R, Flower HM, West DRF. Martensitic transformationsin Ti–Mo alloys. Journal of Materials Science 1979;14:712–22.

41. Bania PJ. Beta titanium alloys and their role in the titaniumindustry. Beta titanium alloys in the 1990’s. Warrendale, PA:The Mineral, Metals and Materials Society; 1993. p. 3–14.

42. Thomson-Neal D, Evans GH, Meffert RM. Effects ofprophylactic treatments on titanium, sapphire andhydroxyapatite-coated implants. A SEM study. TheInternational Journal of Periodontics and Restorative Dentistry1989;9:300–11.

43. Schutz RW, Thomas DE. Corrosion of titanium and titaniumalloys. Metals handbook, vol. 13. OH: Metals Park, ASMInternational; 1987. p. 669–706.

44. Schmidt AM, Azambuja DS. Effect of fluoride ions on Ti–6Al–4V alloy passivation in lactated Ringer’s serum. MaterialsResearch 2003;6:239–46.

45. Souto MR, Laz MM, Reis RL. Biomaterials 2003;24:4213.46. Venugopalan R, Wiemer JJ, George MA, Lucas LC.

Biomaterials 2000;21:1669.47. Rajagopalan KV. Molybdenum: an essential trace element in

human nutrition. Annual Review of Nutrition 1988;8:401–27.48. Messer RLW, Lucas LC. Evaluations of metabolic activities as

biocompatibility tools: a study of individual ions’ effects onfibroblasts. Dental Materials 1999;15:1–6.

49. Trumbo P, Yates AA, Schlicker SA, Poos MI. Dietaryreference intakes: vitamin A, vitamin K, arsenic, boron,chromium, copper, iodine, iron, manganese, molybdenum,nickel, silicon, vanadium, and zinc. Journal of the AmericanDietetic Association 2001;101:294–301.