controls for human papilloma virus (hpv) dna testing · pdf filecontrols for human papilloma...
TRANSCRIPT
![Page 1: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/1.jpg)
Controls for Human Papilloma Virus (HPV) DNA testing
Mark Manak, Ph.D.Boston Biomedica, Inc.
SoGAT XVIIParis, FranceMay 27, 2004
![Page 2: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/2.jpg)
Human Papilloma Virus InfectionsOne of most common sexually transmitted viruses
– 5.5 million new infections in US per year– Most common cause of mortality
from gynecological malignancy world wide
80% of population will be infected with HPV in their lifetime.– 28-46% by age 25.– Most infections spontaneously cleared, – 5-10% of women over 30 are persistent carriers of HPV– 15,000 US women progress to cervical cancer annually – 5,000 die
Fully preventable if detected and treated earlyDetected by PAP smear FDA approved combination of cervical cytology and HPV DNA
testing for primary screening for women aged 30 yr and older
![Page 3: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/3.jpg)
Cervical Cytology/HPV DNA Testing
HPV Sample Collection and Testing Cervical SwabsUsed for both cytology and HPV DNA testing
Cell Morphology %HPV DNA +Normal 10%ASC-US 20-40%
Atypical squamous cells of undetermined significanceLSIL 60-85%
Low grade squamous intraepithelial lesionsHSIL 80-90%
High grade squamous intraepithelial lesionsSquamous Cell Carcinoma 99%
![Page 4: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/4.jpg)
Digene’s Hybrid Capture 2 HPV DNA Test
Hybrid Capture of DNA-RNA hybrids: MAb DetectionUses two RNA probe cocktails Differentiate between high and low risk HPV types
– High Risk Subtypes16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 68
– Low Risk Subtypes6, 11, 42, 43 44
From Digene Web site
![Page 5: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/5.jpg)
ACCURUN 370: HPV DNA ControlConvenient Liquid Cytology “Liquid Pap Smear” Sample
– Packaged in Centrifuge Tube– Processing identical to Clinical Sample– In 4 ml PreservCyt or 2 ml SurePath
Challenging, yet consistent low level HPV Control– Cultured human epithelial diluted
with excess non-infected human cells– Positive cells: 1-2 copies/cell of HPV Type 16 DNA– HPV-16 most common high risk subtype in the US
Thoroughly characterized and quantitated– Cell Count, hemocytometer and Guava – Cell staining specific for SiHa cells– Digene HC2 – HPV TaqMan
![Page 6: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/6.jpg)
>50
37.6
34.1
31.0
27.4
Avg Ct
0
3
29
245
3,100
Av Copies/assay
0.5 (green)
5 (dark blue)
50 (lavender)
500 (blue)
5,000 (red)
Number of cells
Real time Detection of Serial Dilutions of SiHa cells
![Page 7: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/7.jpg)
Real-time Quantitation: Cell Mixtures
Red: SiHa cells 8,480 copies/ml.Blue: SiHa & Uninfected cells 8,370 copies/ml.
![Page 8: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/8.jpg)
Titration of S iHa Cells
0.1
1.0
10.0
100.0
1 10 100Number of SiHA Cells (x 103)
HPV Stability Study Test (37o C)
0.1
1.0
10.0
100.0
0 5 10 15 20Days
RLU
/CO
HPV Freeze-Thaw
0.1
1.0
10.0
100.0
0 1 2 3 4
Number of Freeze-Thaws
RLU
/CO
HPV Stability Study (22-25o C)
0.1
1.0
10.0
100.0
0 20 40 60 80Days
RLU
/CO
Stability of HPV DNA Controls
![Page 9: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/9.jpg)
Validation Study Design
• Single lot of control material • 4 testing sites
– 3 Clinical Labs, 1 IVD manufacturer• Results reported as RLU/cutoff ratio• Standard Digene Protocol• 52 observations• 8 assay runs
– 4 assay runs manual equipment – 4 assays automated “Rapid Capture”
• 2 kit reagent lots
![Page 10: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/10.jpg)
Field Study Results
0
3
6
9
12
15
Reference Lab
RLU
/CO
RL
-1
RL
-2
RL
-3
0
2
4
6
8
10
12
14
3 4 5 6 7 8 9 10 11 12 13 14Mo
re
RLU/CO
Frequency
N = 52
Mean = 7.7
Median = 7.8
SD = 1.08
![Page 11: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/11.jpg)
HPV DNA Control: Conclusion• Convenient format compatible with Cytology specimen collection
– Well accepted by technologists – Well developed cell button formation – Minimal training for use and interpretation
• Mimics Low HPV Level liquid cytology sample – Consistent, well characterized control– Controls for entire testing process
• Stable, consistent product – Stable up to 65 days at 22-250 C– Reflects >2 years of refrigerated storage
HPV ACCURUN ProductsIn 4 ml PreservCyt In 2 ml SurePathA370 – HPV Positive Control A870 – HPV Positive ControlA371 – HPV Negative Control A871 – HPV Negative Control
![Page 12: Controls for Human Papilloma Virus (HPV) DNA testing · PDF fileControls for Human Papilloma Virus (HPV) DNA testing Mark Manak, Ph.D. Boston Biomedica, Inc. SoGAT XVII Paris, France](https://reader031.vdocuments.us/reader031/viewer/2022030502/5aaeb22a7f8b9a3a038c7641/html5/thumbnails/12.jpg)
Acknowledgements
BBI Diagnostics– Alan Doty– Barbara Weiblen– Harmesh Sharma
BBI Biotech– Janet Lathey– Yiping Zhang– Anekella Bharathi– Ron Meixell
Digene – Iwona Mielznska
Cleveland Clinic – Susan Schindler
Johns Hopkins Clinical Lab
Quest Diagnostics Lab
SiHa cells are used under a Material License Agreement with the NIH.Y. Ito Exp. Biol. Med. 135: 543-545, 1970.