control of scatter radiation. objectives begin discussing factors that influence image detail or...

62
Control of Control of Scatter Scatter Radiation Radiation

Upload: nicholas-powell

Post on 26-Dec-2015

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Control of Control of Scatter RadiationScatter Radiation

Page 2: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

ObjectivesObjectives Begin discussing factors that influence Begin discussing factors that influence

image detail or visibility of detailimage detail or visibility of detail Spatial & Contrast resolutionSpatial & Contrast resolution Radiographic NoiseRadiographic Noise Scatter RadiationScatter Radiation Ways to reduce scatter & improve image Ways to reduce scatter & improve image

qualityquality Primary beam restriction & GridsPrimary beam restriction & Grids Technique adjustments when using Technique adjustments when using

gridsgrids

Page 3: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

What are some factors that What are some factors that increase scatter radiation?increase scatter radiation?

Page 4: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

3 factors contribute to an 3 factors contribute to an increase in scatter increase in scatter

Increased kVpIncreased kVp

Increased x-ray field sizeIncreased x-ray field size

Increased patient thicknessIncreased patient thickness

Page 5: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

2 principal characteristics of 2 principal characteristics of any image are Spatial & any image are Spatial &

Contrast ResolutionContrast Resolution Spatial resolution Spatial resolution

Resolution is the ability to image two separate Resolution is the ability to image two separate objects and visually distinguish one from the objects and visually distinguish one from the otherother

Spatial resolution is the ability to image small Spatial resolution is the ability to image small objects that have high subject contrast (eg. objects that have high subject contrast (eg. bone-soft tissue interface, calcified lung bone-soft tissue interface, calcified lung nodules)nodules)

Determined by focal-spot size and other factors Determined by focal-spot size and other factors that contribute to blurthat contribute to blur

Diagnostic x-ray has excellent spatial resolution. Diagnostic x-ray has excellent spatial resolution. It is measured in line pairs per mm. It is measured in line pairs per mm. (CT measured (CT measured in cm)in cm)

Page 6: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence
Page 7: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

SMPTE Test Pattern

Page 8: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence
Page 9: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Image-forming x-raysImage-forming x-rays

Two kinds of photons are responsible Two kinds of photons are responsible for the OD and contrast on an image: for the OD and contrast on an image: Photons that pass through without Photons that pass through without interacting and those that are interacting and those that are scattered though Compton.scattered though Compton.

X-rays that exit the patient are X-rays that exit the patient are remnant and those that interact with remnant and those that interact with the IR are image-forming. the IR are image-forming.

Page 10: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Ideally, only those x-rays that do Ideally, only those x-rays that do not interact with the patient not interact with the patient

should reach the IR….should reach the IR…. However, scatter radiation is a However, scatter radiation is a

factor that must be managedfactor that must be managed Proper collimation has the PRIMARY Proper collimation has the PRIMARY

effect of reducing patient dose by effect of reducing patient dose by _________ ?_________ ?

Proper collimation also improved Proper collimation also improved image contrast by reducing image contrast by reducing radiographic noise or fog caused by radiographic noise or fog caused by scatterscatter

Page 11: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Fog or Noise

Page 12: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Contrast changes with the use of a grid

Less scatter radiation & less radiographic noise – shorter scale = “better contrast”

With Grid No Grid

Page 13: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

kVpkVp

As x-ray energy increases Photoelectric As x-ray energy increases Photoelectric and Compton interactions decrease. and Compton interactions decrease. Explain?Explain?

At 50 kVp 79% photoelectric, 21% At 50 kVp 79% photoelectric, 21% Compton & less than 1% transmissionCompton & less than 1% transmission

At 80 kVp 46% photoelectric, 52% At 80 kVp 46% photoelectric, 52% Compton & 2% transmissionCompton & 2% transmission

Pg. 225Pg. 225

Page 14: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

How does increasing kVp How does increasing kVp affect patient dose? affect patient dose?

Page 15: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Patient ThicknessPatient Thickness

Imaging thick parts of the body Imaging thick parts of the body results in more scatter radiation results in more scatter radiation than thin partsthan thin parts

IMAGE TEST TOOL

Page 16: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Is patient thickness Is patient thickness something the radiographer something the radiographer

can control? can control?

Page 17: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Patient thicknessPatient thickness

Normally, NoNormally, No

Compression devices improves Compression devices improves spatial resolution by reducing spatial resolution by reducing patient thickness and bringing the patient thickness and bringing the object closer to the IR. Compression object closer to the IR. Compression also reduces patient dose and also reduces patient dose and contrast resolution contrast resolution

Page 18: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

CompressionCompression

Improves spatial Improves spatial resolutionresolution

Reduces OIDReduces OID Reduces patient Reduces patient

dose dose Improves contrast Improves contrast

resolution resolution (reducing fog or (reducing fog or noise)noise)

Page 19: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

CompressionCompression

Page 20: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Field SizeField Size

As field size increases, intensity of As field size increases, intensity of scatter radiation also increases scatter radiation also increases rapidly. Especially during rapidly. Especially during fluoroscopyfluoroscopy

Page 21: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Compare images: What do you think about radiographic contrast & image noise?

Page 22: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Control of Scatter Control of Scatter RadiationRadiation

Technologists routinely use two Technologists routinely use two types of devices to reduce the types of devices to reduce the amount of scatter radiation reaching amount of scatter radiation reaching the IRthe IR

Beam restrictorsBeam restrictors

GridsGrids

Page 23: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

3 Types of beam-restricting 3 Types of beam-restricting devicesdevices

Aperture Aperture DiaphragmDiaphragm

Cones or Cones or

Cylinders Cylinders

Variable aperture Variable aperture collimatorcollimator

Page 24: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Variable Aperture Variable Aperture CollimatorCollimator

The most common beam-restricting The most common beam-restricting device is the light-localizing variable device is the light-localizing variable aperture collimatoraperture collimator

The first part of the collimator The first part of the collimator serves to control off-focus radiation. serves to control off-focus radiation. What is off-focus radiation?What is off-focus radiation?

Page 25: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Off - focus RadiationOff - focus Radiation

X-ray tubes are designed so that the X-ray tubes are designed so that the projectile e- interacts with the projectile e- interacts with the target. However, some of the e- target. However, some of the e- bounce off the target and land on bounce off the target and land on other areasother areas

This caused x-rays to be produced This caused x-rays to be produced out side the focal spotout side the focal spot

Page 26: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Extrafocal RadiationExtrafocal Radiation

These rays can also be called off-These rays can also be called off-focus radiationfocus radiation

Extrafocal radiation is undesirable Extrafocal radiation is undesirable because it extends the size of the because it extends the size of the focal spot, increases patient skin focal spot, increases patient skin dose & reduces image contrastdose & reduces image contrast

Page 27: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Off-focus radiationOff-focus radiation

Page 28: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Fixed diaphragm in the Fixed diaphragm in the tube housingtube housing

Using a grid Using a grid

does not reducedoes not reduce

extrafocal extrafocal

radiationradiation

Page 29: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

First-stage entrance First-stage entrance shuttering deviceshuttering device

Has multiple Has multiple collimator collimator blades blades protruding from protruding from the top of the the top of the collimator into collimator into the tube housingthe tube housing

Page 30: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

The collimator lamp & mirrorThe collimator lamp & mirror

Must be adjusted Must be adjusted so that the so that the projected light projected light field coincides with field coincides with the x-ray beamthe x-ray beam

Misalignment of Misalignment of the light field and the light field and beam can result in beam can result in collimator cutoff of collimator cutoff of anatomic anatomic structuresstructures

Page 31: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Total FiltrationTotal Filtration Filtration review…Filtration review… Total Filtration = _______________ + Total Filtration = _______________ +

_____________?_____________?

The collimator assembly is usually The collimator assembly is usually equivalent to approximately _______ mm Al equivalent to approximately _______ mm Al filtration.filtration.

Minimum filtration for tubes that can Minimum filtration for tubes that can operate about 70 kVp is _______ mm Al or operate about 70 kVp is _______ mm Al or equiv.equiv.

Page 32: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

The GridThe Grid

Page 33: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Contrast & Contrast Contrast & Contrast ResolutionResolution

Two devices are used to reduce Two devices are used to reduce Compton effect beam-restricting Compton effect beam-restricting devices and radiographic gridsdevices and radiographic grids

Beam-restricting devices effects Beam-restricting devices effects what reaches the patient. Grids what reaches the patient. Grids effect the remnant beam effect the remnant beam

Page 34: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Contrast & Contrast Contrast & Contrast ResolutionResolution

Contrast = the comparison of areas Contrast = the comparison of areas of light, dark and shades of gray on of light, dark and shades of gray on the imagethe image

Contrast Resolution = the ability to Contrast Resolution = the ability to image adjacent similar tissuesimage adjacent similar tissues

Page 35: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid RatioGrid Ratio

High-ratio grids are more effective in High-ratio grids are more effective in cleaning up scatter radiation than low-ratio cleaning up scatter radiation than low-ratio gridsgrids

The angle of deviation is smaller for high-The angle of deviation is smaller for high-ratio grids. (the photon must be traveling in ratio grids. (the photon must be traveling in a straighter line to make it through the grid)a straighter line to make it through the grid)

However, the higher the ratio the more However, the higher the ratio the more radiation exposure necessary to get a radiation exposure necessary to get a sufficient number of x-rays through the grid sufficient number of x-rays through the grid to the IRto the IR

Page 36: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid RatioGrid Ratio

3 important dimensions on a grid: The 3 important dimensions on a grid: The thickness of the grid strips, the width of thickness of the grid strips, the width of the interspace material, and the height the interspace material, and the height of the gridof the grid

The grid ratio is the The grid ratio is the HEIGHTHEIGHT of the grid of the grid divided by the divided by the INTERSPACE WIDTHINTERSPACE WIDTH::

Grid ratio = Grid ratio = h h

DD

Page 37: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

h = height of the grid, T = h = height of the grid, T = thickness of the grid strip, D = thickness of the grid strip, D =

width of the interspace materialwidth of the interspace material

Page 38: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

The higher the ratio the The higher the ratio the straighter the photon must straighter the photon must

travel to reach the IRtravel to reach the IR Grid ratios Grid ratios

rangerange

from 5:1 to 16:1from 5:1 to 16:1 Most commonMost common

8:1 to 10:18:1 to 10:1 A 5:1 grid will A 5:1 grid will

clean up 85% clean up 85%

16:1 clean up 97%16:1 clean up 97%

Page 39: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid FrequencyGrid Frequency

The number of grid strips or grid The number of grid strips or grid lines per inch or centimeter lines per inch or centimeter

The higher the frequency the more The higher the frequency the more strips and less interspace material strips and less interspace material and the higher the grid ratioand the higher the grid ratio

As grid frequency increases, patient As grid frequency increases, patient does is increase because more does is increase because more scatter will be absorbedscatter will be absorbed

Page 40: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid FrequencyGrid Frequency

Some grids reduce the thickness of Some grids reduce the thickness of the strips to reduce the exposure to the strips to reduce the exposure to the patient, this over all reduces the the patient, this over all reduces the grid clean upgrid clean up

Grids have frequencies in the range Grids have frequencies in the range of 25 to 45 lines per centimeter (60 of 25 to 45 lines per centimeter (60 to 110 lines per inch)to 110 lines per inch)

Page 41: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Higher frequency with the same Higher frequency with the same interspace distance reduces the interspace distance reduces the

grid effectivenessgrid effectiveness

Page 42: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid PerformanceGrid Performance

The principal function of a grid is to The principal function of a grid is to improve image contrastimprove image contrast

Contrast Improvement FactorContrast Improvement Factor (k)(k) = the = the ratio of the contrast of a radiograph ratio of the contrast of a radiograph made with a grid to the contrast of the made with a grid to the contrast of the radiograph made without a grid. A radiograph made without a grid. A contrast improvement factor of 1 contrast improvement factor of 1 indicates no improvements indicates no improvements

The higher the grid ratio & frequency The higher the grid ratio & frequency the higher the kthe higher the k

Page 43: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Bucky FactorBucky Factor

Using grids require more patient Using grids require more patient dose. Why is this?dose. Why is this?

When a grid is used technique must When a grid is used technique must be increased to maintain ODbe increased to maintain OD

The amount of increase is given by The amount of increase is given by the Bucky factor (B) or grid factorthe Bucky factor (B) or grid factor

Page 44: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Bucky Factor or grid Bucky Factor or grid factorfactor

The higher the grid ratio or frequency The higher the grid ratio or frequency the higher the bucky factorthe higher the bucky factor

The Bucky factor increases with The Bucky factor increases with increasing kVpincreasing kVp

Pg 235: We will use the average Pg 235: We will use the average values for calculations. values for calculations.

Page 45: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Selectivity or ability to Selectivity or ability to “clean up”“clean up”

the heavier the grid the more Pb it the heavier the grid the more Pb it containscontains

Page 46: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid Types Grid Types

Parallel Grid – simplest type of gridParallel Grid – simplest type of grid All the lead strips are parallelAll the lead strips are parallel Only clean up scatter in one Only clean up scatter in one

direction (along the axis of the grid)direction (along the axis of the grid)

Easy to make, however can cause Easy to make, however can cause grid cutoff with short SID’s. grid cutoff with short SID’s.

Page 47: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid cutoffGrid cutoff Distance to cutoffDistance to cutoff SIDSID Grid ratioGrid ratio With decreasing With decreasing SID more potential SID more potential for grid cutofffor grid cutoff IR size will alsoIR size will alsoInfluence grid Influence grid

cutoffcutoff

Page 48: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid Cutoff – Parallel grid

THE HIGHER-RATIO THE MORE CUTOFF POTENTIAL

Page 49: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Crossed GridCrossed Grid

Have lead strips running along the Have lead strips running along the long and short axes of the gridlong and short axes of the grid

Made by placing two parallel grid on Made by placing two parallel grid on top of each othertop of each other

Page 50: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Crossed GridCrossed Grid

Have twice the grid ratio as linear Have twice the grid ratio as linear

gridsgrids

However, CR vs grid placement is However, CR vs grid placement is critical. The CR must align with the critical. The CR must align with the center of the grid and the grid and CR center of the grid and the grid and CR must be exactly parallel or grid cutoff must be exactly parallel or grid cutoff will occurwill occur

Page 51: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Focused GridFocused Grid

Designed to minimize grid cutoffDesigned to minimize grid cutoff Lead strips are aligned with the Lead strips are aligned with the

divergence of the x-ray beamdivergence of the x-ray beam

Each focused grid must be identified Each focused grid must be identified with the appropriate SIDwith the appropriate SID

Wrong SID = Grid cutoffWrong SID = Grid cutoff

Page 52: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Focused grid have a little SID Focused grid have a little SID latitude (eg. 100cm grid could be latitude (eg. 100cm grid could be

used at 90cm – 110cm)used at 90cm – 110cm)

Page 53: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Moving GridsMoving Grids

All stationary grids will give you grid All stationary grids will give you grid lines on your radiograph. Thinner Pb lines on your radiograph. Thinner Pb strips will give you less noticeable strips will give you less noticeable lines. However, thinner strips have lines. However, thinner strips have less Pb content not “cleaning up” as less Pb content not “cleaning up” as wellwell

Grid Lines are made when primary x-Grid Lines are made when primary x-rays are absorbed in the grid strips. rays are absorbed in the grid strips.

Page 54: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Focused grids are usually Focused grids are usually used as moving gridsused as moving grids

The grid is placed in a holding The grid is placed in a holding mechanism that begins moving just mechanism that begins moving just before the x-ray exposure and before the x-ray exposure and continues moving after the exposure continues moving after the exposure endsends

2 types of movement Reciprocating 2 types of movement Reciprocating & Oscillating& Oscillating

Page 55: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid MotionGrid Motion

Reciprocating = moves several times Reciprocating = moves several times about 2cm back and forth during the about 2cm back and forth during the exposureexposure

Oscillating = moves several times Oscillating = moves several times about 2 – 3 cm in a circular patternabout 2 – 3 cm in a circular pattern

Most grids are moving. Except for Most grids are moving. Except for portable imaging portable imaging

Page 56: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid ProblemsGrid Problems

Increased OID, especially with Increased OID, especially with moving gridsmoving grids

The biggest problem with grids is The biggest problem with grids is misalignmentmisalignment

GRID PROBLEMS RESULT IN:UNDEREXPOSED IMAGEOR UNDEREXPOSEDEDGES OF IMAGE

Page 57: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid Problems – Off LevelGrid Problems – Off Level

Page 58: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid Problems – Off Grid Problems – Off CenterCenter

A problem with focused & crossed gridsA problem with focused & crossed grids

Page 59: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid Problems – Grid Problems – Off FocusOff Focus (wrong SID)(wrong SID)

Page 60: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid Problems – Grid Problems – Upside-Upside-Down Down

A problem with focused & crossed gridsA problem with focused & crossed grids

Page 61: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Grid SelectionGrid Selection

Patient Dose Patient Dose Pg 241 – mAs changesPg 241 – mAs changes

ExamExam Detail requiredDetail required Part thicknessPart thickness Desired technique (kVp)Desired technique (kVp) Equipment availabilityEquipment availability

Page 62: Control of Scatter Radiation. Objectives Begin discussing factors that influence image detail or visibility of detail Begin discussing factors that influence

Questions….?