Dr. M i c h a e l B a e r t s c h iPhD (Biomedicine), MSc.Optom. et M.med.Education, FAAO, FEAOO
eyeness ag, Bern / Switzerland
Contact LensesSwiss Eye Week 2020
![Page 2: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/2.jpg)
eyeness AG
![Page 3: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/3.jpg)
eyeness Team
![Page 4: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/4.jpg)
www.eyeness.ch/downloads
D o w n l o a d
![Page 5: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/5.jpg)
DisclosuresStudien für: Consulting Mandate:
Abbott Falco Kontaktlinsen
Alcon Vision Care Haag Streit Diagnostics
Bausch&Lomb RaayonNova
Contamac Vistakon (Johnson&Johnson)
Cooper Vision
Falco Kontaktlinsen
Haag Streit Diagnostics
Sensimed AG (Triggerfish)
Tissot Medical Research
Vistakon (Johnson&Johnson)
![Page 6: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/6.jpg)
Goals
• Understand the basics of contact lenses.
• Know their differentiations and frequency of CL wearers,
understand the specific optics and physiology, distinguish
wearing modalities, know the fields of applications and the
fitting basics, discuss the importance of hygiene and lens
care, differentiate complications and how to avoid them
and take a look into novelties.
• Have Fun with Contacts !
![Page 7: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/7.jpg)
History and Demography
Leonardo da Vinci, 1508 and Rene Descartes 1637
![Page 8: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/8.jpg)
(Glass) Scleral Lenses / Shells
Over All Diameter
OAD 20-24 mm
![Page 9: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/9.jpg)
(Soft) Hydrogel Lenses
Otto Wichterle, Prague 1961
Patent sold to Bausch&Lomb 1968
Over All Diameter OAD 13-15 mm, Water content 25-85%
![Page 10: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/10.jpg)
(Hard) gas permeable Lenses
Over All Diameters between 8 mm to 18 mm, no water
Invented after WW II (1945), High-tech today
10,2 mm 9,3 mm 8,3 mm
11,2 mm 13,0 mm 16,0 mm
![Page 11: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/11.jpg)
Movement, centration and stabilization
![Page 12: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/12.jpg)
Demography
Contact lenses
1/3 vs 2/3
![Page 13: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/13.jpg)
CL Market (Alcon Data, 2015)
![Page 14: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/14.jpg)
Optics and Physiology
Whatever contact lens you fit,
always respect the physiology
and the safety of the eye !
![Page 15: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/15.jpg)
Spheric Calculations
• As a rule of thumb: 0.1mm radius difference results in a power change of 0.5dpto If the lens is chosen 0.1mm steeper than the cornea you have to
increase the amount of minus with -0.5dpt
o If the lens is chosen 0.1mm flatter than the cornea you have to decrease the amount of minus with +0.5dpt
![Page 16: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/16.jpg)
Aniseikonia between CL and glasses
Myopic Px profit the most from CL regarding less retinal image size
reduction, respectively image size magnification.
Hyperopic Px profit the most from CL regarding aesthetics.
![Page 17: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/17.jpg)
Visual acuity comparison
Myopic Px profit the most of CL regarding VA
![Page 18: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/18.jpg)
Visual field comparison
Both Px groups profit from CL regarding VF
![Page 19: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/19.jpg)
Accommodation
• Hyperopic Px with contact lenses needs to accommodate less
(more accommodation success A0) for a given distance.
• Myopic Px with contact lenses needs to accommodate more.
![Page 20: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/20.jpg)
Convergence
• Myopic Px with contact lenses needs to
converge more (less comfortable) compared
to a correction with spectacles for a given
distance.
• Hyperopic Px needs to converge less with
contact lenses.
![Page 21: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/21.jpg)
Example Distance Myopia
![Page 22: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/22.jpg)
Example Reading Myopia
comfortReading Target
![Page 23: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/23.jpg)
Regarding Presbyopia and CL
Due to accommodation and convergence disadvantages, myopic
Px are more difficult to correct with CL than with glasses.
But they always profit from better optics (less aberrations).
![Page 24: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/24.jpg)
Higher Order Aberrations
K.Gumus et al The Impact of the Boston Ocular Surface
Prosthesis on wavefront Higher-Order Aberrations, Am J
Ophthalmol 2011;151:682–690
The corrective rates of higher-
order aberrations are about
70-90% with (hard) scleral
lenses in place !
![Page 25: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/25.jpg)
CL Materials: Physics and Physiology
Differentiate important physical properties :
• Oxygen Permeability
• Hardness (Shore), Resistance
• Stiffness (Flex Modulus)
• Wettability (Wetting angle)
• Silicone, Fluor, Water content
• Refractive Index
• UV Filter
![Page 26: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/26.jpg)
Oxygen Permeability Dk
26 42
56
86
110 118 121 125
138 147
156 160 163 175
0
20
40
60
80
100
120
140
160
180
200
etafilcon A omafilcon Filcon II 3 somofilcon A balafilcon A Narafilcon A senofilcon A enfilcon A lotrafilcon B senofilcon A delefilcon A comfilcon A samfilcon A lotrafilcon A
Sauerstoffdurchlässigkeit Dk/t
DW
EW
CW
![Page 27: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/27.jpg)
Hardness, Resistance
Scratches and Edgedefects
![Page 28: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/28.jpg)
Wettability, Silicone, Water, Fluor
Deposits (Lipids and Proteins)
![Page 29: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/29.jpg)
Contact Lens Manufacturing
Understand manufacturing steps:
- Calculation
- Preparation
- Back Geometry
- Over All Diameter
- Front Geometry and Power
- Finish
- Quality Check
Disposable lenses (daily, weekly, monthly)
![Page 30: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/30.jpg)
CL Laboratories
Calculation and Preparation
Infrastructure
Cutting and lathingFinish and Quality Check
FALCO, Switzerland
![Page 31: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/31.jpg)
Calculation and Preparation
Finish and Quality CheckCalculation and Preparation
GELFLEX, Perth / Australia
CL Laboratories
![Page 32: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/32.jpg)
Contact Lens Design
Distinguish different Backcurve Designs
• spherical
• Multi-curve
• aspheric
• toric
![Page 33: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/33.jpg)
Sphere versus Asphere
„ “
![Page 34: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/34.jpg)
KK: Steiler Apex mit hoher Abflachung
zur Peripherie hin
Numeric Eccentricity nE
r1
r2
r2 > r1
![Page 35: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/35.jpg)
Numeric Eccentricity nE – Rule of Thumb
Numerical Eccentricity / peripheral flattening
0.1 = 0.02 mm -> Elliptic
0.2 = 0.05 mm
0.3 = 0.10 mm -> 0.1 mm
0.4 = 0.16 mm -> 0.2 mm
0.5 = 0.23 mm
0.6 = 0.35 mm -> 0.3 mm
0.8 = 0.55 mm
1.0 = 0.80 mm -> Parabolic (Keratoconus)
1.2 = 1.20 mm -> Hyperbolic (Keratoconus)
1.5 = 2.00 mm
![Page 36: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/36.jpg)
Spherical Design
0.0
0.0 0.0
0.0
BC = r2/0
![Page 37: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/37.jpg)
Multi curve Design
0.0
0.0 0.0
0.0
BC = r2/0
![Page 38: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/38.jpg)
Aspherical Design
BC = r2/0
0.0
0.0 0.0
0.0
0.4
0.4
0.4
0.4
![Page 39: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/39.jpg)
Toric Design
0.0
0.0 0.0
0.0
BC = r2/0
BC = r2/1
0.6 0.6
![Page 40: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/40.jpg)
Lens Design
![Page 41: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/41.jpg)
Fitting Technics,Ocular Topography and
Fluorescein
![Page 42: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/42.jpg)
Technique Oculus
• 22’000 measuring pointso Nose and forehead make shadows and reduce the
amount of measure points
o Only about the central 8mm diameter are really measured
o Accuracy +/-0.1D
![Page 43: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/43.jpg)
Diagnostic Software (Mapping)
![Page 44: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/44.jpg)
Diagnostic Software (Fourier)
• Fourier Analysiso Mathematical separation of the topographic map into
individual components
o Optical aberration (spherical component, decentration or decentration with spherical component, regular astigmatism, irregularities or irregularities with spherical component)
![Page 45: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/45.jpg)
Diagnostic Software (Fourier)
![Page 46: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/46.jpg)
3D ImagesRegular Astigmatism
![Page 47: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/47.jpg)
3D Images
Irregular Astigmatism
post Trauma
![Page 48: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/48.jpg)
3D ImagesPseudo Keratoconus (high riding contact lens)
![Page 49: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/49.jpg)
3D Images
Keratoconus
![Page 50: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/50.jpg)
3D Images
Keratoplasty
![Page 51: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/51.jpg)
3D Images
LASIK / Orthokerathology
![Page 52: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/52.jpg)
3D Images
Pellucid marginal Degeneration
![Page 53: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/53.jpg)
Scheimpflug Technology
• Slit illumination creating a cross-sectional image of the anterior chambero True elevation data rather than curvature values
o Measurement of anterior and posterior surface, center of cornea really measured
o Independent of tear layer quality
o Detection of opacities in cornea and cristalline lens
![Page 54: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/54.jpg)
Scheimpflug Technology
![Page 55: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/55.jpg)
Diagnostic Software
![Page 56: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/56.jpg)
Diagnostic Software (Mapping)
• Topography and elevation data of anterior and
posterior corneal surface
– Keratoconus detection (very early due to the posterior
surface evaluation) and classification
– Advanced topographical calculation due to refractive
surgery
– Comparative displays for progression control and
follow-up
![Page 57: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/57.jpg)
Diagnostic Software (Pachy & Topo)
![Page 58: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/58.jpg)
Diagnostic Software (Holladay Report)
![Page 59: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/59.jpg)
Diagnostic Software (Belin/Ambrosio)
• Sagittal and tangential mapping shows similar
results and no evidence of abnormality
![Page 60: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/60.jpg)
OCT Technology
• Optical Coherence Tomography (OCT)o Mapping and Keratoconus detection software already
available
o Resolution between 5μm - 60 μm
o Maximum of four scans, to date
![Page 61: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/61.jpg)
OCT Technology
![Page 62: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/62.jpg)
• OCT and contact lenses
o Measurement coverage up to 16.00mm and include
Sclera/Conjunctiva data
o Simplify scleral lens fitting process
o Allows in-situ validation of any contact lens, even
Hydrogel’s
OCT Technology
![Page 63: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/63.jpg)
OCT Technology
![Page 64: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/64.jpg)
OCT Technology
Toric Soft Lens Fit Analysis
![Page 65: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/65.jpg)
Fitting procedure
1. Anamnese and History taking
2. Perform Slitlamp observation and Topography measurement
3. Check central and peripheral corneal astigmatism
4. Check ametropia and total eye astigmatism
5. Define wearing modality
6. Choose your ideal fitting strategy (Physiology and Optic)
7. Define the ideal material, lens geometry and over all diameter OAD
8. Check centration, rotation (stabilization axes) and vertical movement
9. Define Lens Care regime
10. Organize regular CL and Eye Checks
11. Send final report to Ophthalmologist
![Page 66: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/66.jpg)
CL Fitting Software
![Page 67: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/67.jpg)
CL Fitting Software
![Page 68: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/68.jpg)
Fitting Software Oculus
![Page 69: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/69.jpg)
Fitting Software
• “Real Life” peripheral Astigmatism
– The lens rocks around flat meridian
– High or Low Riding lens as a consequence
![Page 70: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/70.jpg)
Final Fitting Result
• Perfect centration
• Outstanding comfort
• Fluorescein pattern:
typical pooling in steep
mid-periphery and alignment fit in periphery
• Markings of flat meridian
![Page 71: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/71.jpg)
Rigid Lens Fluorescein PatternsProducts (Small molecular / Large molecular NaFl)
Small
Large
RGP, screening Soft lenses
![Page 72: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/72.jpg)
Rigid Lens Fluorescein PatternsTechnical approach (Kobalt Filter, Yellow (Wratten) Filter)
Kobalt Filter
![Page 73: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/73.jpg)
Rigid Lens Fluorescein Patterns
Wratten
Yellow Filter 12
Mobile Alternatives
LED hand held
![Page 74: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/74.jpg)
Rigid Lens Fluorescein PatternsApplication / Installation of NaFl
Install only a small (!) drop of NaFl, diluted with NaCl
![Page 75: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/75.jpg)
Characteristics: Diameter
Tear exchange
Oxygen
Movement
Price (standard lenses)
Comfort
Stabilization
Safety
Less dust (foreign body)
Handling
Small lenses :
Large lenses (intra-limbal):
![Page 76: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/76.jpg)
Characteristics: Diameter
Very large lenses :
(“Miniscleral” or Scleral)
Comfort
Stabilization
Less dust (foreign body)
Protection
Visual acuity !
![Page 77: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/77.jpg)
Contact Lenses for Astigmatism
Available for Soft and Hard GP Lenses up to +/- 20dpt cyl
![Page 78: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/78.jpg)
Contact Lenses for Irregular CorneasDue to :
Contact lens use (Deformation, unaceptable fit, Pseudo Keratoconus)
Trauma (Physical, chemical)
Infections (Microorganism)
Degeneration (Keratoconus, Marginal Deg., Salzmann, Band Deg,...)
Dystrophy (EBM Dystrophy, Mapdot-fingerprint, Meesmann,
Fuch’s, bullous Epitheldys., Granular, Lattice, ….)
Systemic (e.g. Lyell Syndrom)
Post surgery (PKP, lamellar Keratoplasty, Trabecelectomy / Filtering bleb)
Refractive surgery (RK, PRK, LASIK/LASEK, SMILE…)
![Page 79: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/79.jpg)
Typical Topographies
Keratoconus Pellucid marginal
or Terrien Degeneration
![Page 80: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/80.jpg)
Optical Axes
Center of Pupil
Apex caudal
decentered
Keratoconus
Scars and
folds(Vogt Striae
or Haab
Lines)
![Page 81: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/81.jpg)
Quadrant-specific Design
1.6 1.6
1.4
0.0
cross sectional view
Spherical optic zone with quadrant specific
periphery for best physiological fit and
maximized visual acuity.
![Page 82: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/82.jpg)
Quadrant-specific Design
![Page 83: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/83.jpg)
Penetrating Keratoplasty
![Page 84: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/84.jpg)
Cross section view Comparison
![Page 85: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/85.jpg)
Typical Topographies
Keratoplasty
![Page 86: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/86.jpg)
Case Keratoplasty
![Page 87: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/87.jpg)
Case Keratoplasty
![Page 88: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/88.jpg)
Case Keratoplasty
Multicurve, reverse geometry with 2 fenestration holes
![Page 89: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/89.jpg)
Quadrant specific Keratoplasty CL
• New Geometry (Falco Switzerland) for tilted grafts– Quadrant specific geometry with tilted reverse geometry
![Page 90: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/90.jpg)
Advantages
• Perfect centrationo Vaulting entire Cornea and Limbus
o The lens rests only at the Scleral Zone
• Simplyfied Fittingo Sclerals are fitted by Sagittal Depth rather than BC
or K readings
• Outstanding Optic Resultso Flatter BC and bigger Optical Zone Diameter
Scleral Lenses
![Page 91: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/91.jpg)
Scleral Fitting TechnicCentral Area - Fluorescein Pattern
![Page 92: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/92.jpg)
Scleral Fitting TechnicThe entire fit, centration and stabilization of the whole CL lays on the conjunctiva/sclera
- OAD 15 - 18mm
- Extreme high Dk
![Page 93: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/93.jpg)
Contact Lenses for Presbyopia
![Page 94: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/94.jpg)
Translating bifocal Systems
© by Bach Optik
![Page 95: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/95.jpg)
Multifocal, Simultaneous DC
© by Bach Optik
![Page 96: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/96.jpg)
Soft and disposable Contact Lenses• Distance or Near central optic zone, with
progressive or degressive power changes
![Page 97: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/97.jpg)
Translational Soft (BYO Royal™)
![Page 98: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/98.jpg)
The Future is Now!
Segmented-Simultaneous Design (Falco Kontaktlinsen Switzerland)
![Page 99: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/99.jpg)
Contact Lenses for Children
![Page 100: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/100.jpg)
Reasons for CL for Children
Myopia Aphakia
Anisometropia
Hyperopia
Astigmatism
![Page 101: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/101.jpg)
Children and eye glassesChildren loves eye glasses ? (Hyperopia +8.0 dpt)
At least their parents believe and unfortunately some eye “experts” as well !
![Page 102: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/102.jpg)
Children and contact lensesChildren loves contact lenses ! (Hyperopia +9.0 dpt)
This is what the children believes and their contact lens specialists !
![Page 103: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/103.jpg)
Children and contact lenses
The two most common reasons today are :
- Myopia Management - Aphakia
- Special soft and „hard“ GP contact lenses are suitable for children.
- Well tolerated and at least as safe as for adults.
- Hygiene and regular CL/eye checks are evident.
![Page 104: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/104.jpg)
Contact Lenses for Therapy• Pain reduction
• Protection against environmental factors
• dirt, mechanical friction, evaporation of tear film
• Improving the epithelial healing process
• Stabilization and modulatio of corneal surface
• Improvement of VA
• Optical Rehabilitation
• Cosmetic Rehabilitation
• (carrier for pharmacologic drugs)
• Improvement of overall comfort, vision and safety !
![Page 105: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/105.jpg)
Method
Fitting of therapeutic and refractive effective soft or hard bandage
contact lenses as ocular surface protection layer.
Without contact lens
With protecting contact lens
OCT image of anterior ocular segment
![Page 106: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/106.jpg)
Materials
• Hydrogel (“old style”)
• Traditional (BG 80%, Benz-G)
• Disposable (Daily, Monthly)
• Silicone Hydrogel (new style)
• Disposable (Daily, Bi-weekly, Monthly)
• Gas Permeable (Hard) Specialty CL and Scleral Lenses
• Hybrid Lenses (Hard GP center, Soft skirt)
![Page 107: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/107.jpg)
Main Material for Clinic
•Past…:
• Hydrogel with H2O content
•…Present:
• Silicone Hydrogel !
![Page 108: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/108.jpg)
O2-Permeability of Silicone Hydrogel CL
Conv. Hydro
![Page 109: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/109.jpg)
Excellent Resultse.g. Px with Epidermolysis Bullosa
- Eliminate pain
- Stabilize Epithelium
- Reduce inflammation
- Increase or restore vision
![Page 110: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/110.jpg)
Epithelial basement membrane Dystrophies EBMD :(Cogan or Map-Dot-Fingerprint)
- Eliminate Discomfort
- Increase physiological
and optical/visual
Stabilization
Excellent Results
![Page 111: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/111.jpg)
Clinical Case: EBMD
![Page 112: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/112.jpg)
Clinical Case: EBMD
B.H., m, 59 y/o
Since 6/2010:
Biofinity sph / Biofinity Toric
Vacc: OD 0.9 , OS 0.8
Constant wear 1 week, cleaning, changes every 4 weeks.
Rewetting drops as needed.
![Page 113: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/113.jpg)
Conclusion
Short term prognosis :
• Therapeutic bandage contact lenses are useful to protect the
compromised cornea from negative environmental influences.
• They are helpful to protect the patient from discomfort and
pain.
• They are usable as a drug deposit with care.
![Page 114: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/114.jpg)
Intermediate and long term prognosis :
• Therapeutic bandage contact lenses are useful to protect the long term dystrophic cornea from negatives environmental influences.
• They help to remodel the irregular cornea for better vision.
• They help to protect the patient from discomfort and pain.
They make patients happy!
Conclusion
![Page 115: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/115.jpg)
Contact Lenses for OrthokeratologyHistory: Dr. J. Ball 1850 developed eye cup it spring mounted mallet to pound the cornea flat through the closed eye
![Page 116: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/116.jpg)
History
With CL initially reported 1960 by Jessen (Old)*o Spherical lens, daily wear, PMMA (Dk = 0)
1997 Mountford introduced new Design (accelerated Ortho-K)
o 4-5 Zones Reverse Geometry, Overnight wear, extrem high Dk >100
*Please: Don‘t mix up old and new Ortho-K
![Page 117: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/117.jpg)
How it works• Change in epithelial thickness
o 4 weeks in Cats results in 8.4microns reduction, results in -4.0D of Ortho-K effect !
• Munnerlyn Formula (tissue displacement)o Optical Zone Diameter2 x refractive Error / 3
o e.g. -5.0D correction with Treatment Zone of 3.0mm (tangential mapping) results in 15microns Epithelium displacement
![Page 118: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/118.jpg)
• Central epithelial thinning o Focal cell compression (Wing and Basal cells) rather than cell migration
or loss of cell layers
o Microvilli and Desmosomes unaffected
• Midperipheral epithelial/stromal thickeningo Increased surface (squamous) cell and nucleus sizeo After 24h: Stromal +2.5% / Epithelial +16.89%o No classical stromal Edema, more likely increased cell retention!
Histology in Ortho-K
• Endothelium o No changes at all
• Anterior Radii vs posterior Radiio Ortho-K alters the anterior corneal shape rather than the posterior
radius of the corneal curvature and the anterior chamber depth
o There is no overall “corneal bending”
Tsukiyama, Eye & Contact Lens 34(1): 17–20, 2008
![Page 119: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/119.jpg)
• Pediatric and Geriatric Patients
• Therapeutic Indications
• Extreme High Altitude Climbers
Contact lenses for Continuous wear
![Page 120: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/120.jpg)
Indication for continuous wear
• Bandage contact lens after surgery or injury
• Therapeutic use in cornea disorders• EBMD (Map-dot-fingerprint dystrophy) and recurrent corneal erosion
• Epidermolysis Bullosa
• Elagophthalmus
• Incomplete lid closure during sleep
Post LASIK: Pre-Therapy CL With Therapy CL
![Page 121: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/121.jpg)
Products OverviewPurevision
(Toric, multi)
Night&Day(Children)
AirOptix(Toric, multi)
Oasys(Toric, multi)
Biofinity(Toric, multi)
DK/t 110 175 138 154 160
H2O 36% 24% 33% 40% 48%
Radius 8.60 8.40 / 8.60 8.60 8.40 8.60 (8.7)
Ø 14.00 13.80 14.2 14.0014.00
(14.50)
Diopter +6.0 / -12.0 +6.0 / -10.0 +6.0 / -10.0+6.0 /
-12.0
+10.0 /
-10.0
Cylinder
-0.75
-1.25
-1.75
-2.25
*******
-0.75
-1.25
-1.75
-2.25
-0.75
-1.25
-1.75
-2.25
-0.75
to
-5.75
CW 30 days 30 days 7 days 7 days 30 days
Replace 1 month 1 month 1 month 1 week 1 month
Modulus 1.50 1.52 1.2 0.72 0.75
![Page 122: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/122.jpg)
Risk factors for infiltrative event
Szczotka-Flynn L. et al.: Predictive Factors for Corneal Infiltrates with Continuous Wear of Silicone Hydrogel
Contact Lenses; Arch Ophthalmol. 2007;125:488-492
Extended or continuous wear is not for
everybody and deserves special care !
![Page 123: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/123.jpg)
Complications
• Evaporation / Dehydration
• Lack of Oxygen
• Microbial Inflammations or
Infections
• Chemical contamination
• Allergies
![Page 124: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/124.jpg)
1
2
Complications
Superior Epithelial Arcurat Lesion SEAL Gigantopapillary Conjunctivitis GPC
vs
Follicle (FoCoSi)
Contact Lens induced acute red Eye CLARE
![Page 125: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/125.jpg)
Complications CLPC
Differential diagnosis Papillae versus Follicle
Naumann
Pathologie
des Auges
1980;
12:252
![Page 126: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/126.jpg)
Complications
![Page 127: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/127.jpg)
Contact Lens related Peripher Ulcus (CLPU)
“CLPU is an inflammatory reaction of the cornea that is
characterized in its active stage by focal excavation of the
epithelium, infiltration, and necrosis of the anterior stroma.
However, Bowman’s layer is intact”Holden BA, Reddy MK, Sankaridurg PR, et al. The histopathology of contact lens induced peripheral corneal ulcer. Invest Ophthalmol Vis Sci 1997;38:s201.
![Page 128: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/128.jpg)
Microbial (infectious) Keratitis
“MK is an infection of the cornea by microbes that is
characterized by excavation of the corneal epithelium,
Bowman’s layer, and stroma with infiltration and necrosis
of the tissue.”Cokington CD, Hyndiuk RA. Bacterial keratitis. In: Tabbara KF, Hyndiuk RA (eds): Infection of the Eye. 2nd ed. Little, Brown & Co, 1996.
![Page 129: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/129.jpg)
Differential Dx: CLPU vs MK
Active phase: CLPU Fluo „negativ“ MK Fluo strong positive
![Page 130: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/130.jpg)
1
3
Complications
Chronic Red Eye (Ideology ?)
Bacterial Blepharitis due to chronic MGD
=>
Intense Lidcare, warm compresses, Omega-3 substitution
Final result
after 6 months
![Page 131: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/131.jpg)
Predisposing factors for adverse events
Blepharitis and Meibomian Gland Dysfunction (MGD)
Picture: www.healtarena.org
![Page 132: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/132.jpg)
Hygiene and
Safety:
Important
Informations for
CL wearers
![Page 133: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/133.jpg)
Contact Lens Care
Soft Peroxid Hard GPAll-in-One
Rinsing (NaCl 0.9%)
Cleaning
Desinfection
& Storage
Specials (De-Protein)
Always Rub & Rinse with All-in-One products !
![Page 134: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/134.jpg)
Contact Lens Care
![Page 136: Contact Lenses - eyeness€¦ · Optics and Physiology Whatever contact lens you fit, always respect the physiology and the safety of the eye ! Spheric Calculations • As a rule](https://reader033.vdocuments.us/reader033/viewer/2022050216/5f61fc6842b86d274f39204e/html5/thumbnails/136.jpg)
Conclusion and Discussion
Contact lenses are offering a high amount of
technical and physiological possibilities to correct
nearly every kind of ametropia.
They offer an acceptable safety level for all ages, but
fitting and communication has to be age related.
Contact lenses are much more than just “another
kind of glasses”.