![Page 1: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/1.jpg)
ID/AAPM/10
Indra J. Das, PhD, FACRDepartment of Radiation OncologyIndiana University of School of Medicine & Midwest Proton Radiation Therapy Institute (MPRI), Indiana
Commissioning and Calibrating a Linear Accelerator – State-of-the-Art in 2010
![Page 2: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/2.jpg)
ID/AAPM/10
Preface
Med. Phys. 35(9), 4186-4214, 2008
![Page 3: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/3.jpg)
ID/AAPM/10
![Page 4: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/4.jpg)
ID/AAPM/10
2008
![Page 5: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/5.jpg)
ID/AAPM/10
Planning for Commissioning Data
![Page 6: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/6.jpg)
ID/AAPM/10
Time= [(PDD + 5 profiles)/beam energy]x ( open + 4 wedges) x 60 points/scanx [(1 s/pts + (1s/movement and delay)]x (15 fields x 2 energies)
∼105 s∼ 30 h
Planning for Commissioning Time
![Page 7: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/7.jpg)
ID/AAPM/10
Rational For Commissioning Beam DataFirst, it is not evident that manufacturing procedures for all linear accelerators have produced a level of reproducibility acceptable for clinical use. For example, variations in beam parameters have been noted between beams with the same nominal energies.
Second, on-site changes made during installation and acceptance of the user’s accelerator e.g., changes in beam energy and/or profiles from beam steering will not be modeled in the golden data.
Third, the beam characteristics of the soft wedges are made by moving jaws that depend on the speed parameters of the jaws and a deviation at site could affect the beam profile of the soft wedge.
Fourth, although acceptable agreement with the golden data set may be found in individual checks, it may be that some clinical setups will have multiple errors, which combine to produce unacceptable results.
TG-106
![Page 8: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/8.jpg)
ID/AAPM/10
Hrbacek, et al Med. Phys. 34, 2917–2927, 2007.
Rational For Not Using Golden Data
![Page 9: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/9.jpg)
ID/AAPM/10
Hrbacek, et al “Quantitative evaluation of a beam-matching procedure using one dimensional gamma analysis,” Med. Phys. 34, 2917–2927, 2007.
Rational For Not Using Golden Data
|xi − x¯ | <∆, ∀ xi
∆ =? (0.5, 1.0 0r 2%)
![Page 10: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/10.jpg)
ID/AAPM/10
Time required for commissioning adual energy linear accelerator with photon and electron beam is
Answer: 4Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008
Question
1. 1 day2. 3 days3. 1 week4. 4-6 weeks5. 2 months
![Page 11: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/11.jpg)
ID/AAPM/10
Standard chamber 10−1 cm3—The active volume for a standard Farmer-type ionization chamber is on average 0.6 cm3.
Minichamber 10−2 cm3—The active volume for a mini-ionization chamber is on average 0.05 cm3.
Microchamber 10−3 cm3—The active volume for a microionization chamber is on average 0.007 cm3
and ideally suited for small field dosimetry such as radiosurgery,gamma knife, CyberKnife, and IMRT
Definition of Detectors
![Page 12: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/12.jpg)
ID/AAPM/10
Water
Air
1 2 3 4 5
Setting Water Tank & Detector
![Page 13: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/13.jpg)
ID/AAPM/10
Know Your Connectors
![Page 14: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/14.jpg)
ID/AAPM/10
Understand Detector, Connector & Cable
See Poster:Srivastava et al, SU-GG-T-270, 2010
![Page 15: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/15.jpg)
ID/AAPM/10
Quality of Cables
![Page 16: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/16.jpg)
ID/AAPM/10
When setting ion chamber in water tank, the correct position of the chamber as viewed in water tank (as seen in figure) is:
Water
Air
1 2 3 4 5
Answer: 3Reference : Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008
Question
1. Position # 12. Position # 23. Position # 34. Position # 45. Position # 5
![Page 17: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/17.jpg)
ID/AAPM/10
Setup and Possible Errors
![Page 18: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/18.jpg)
ID/AAPM/10
Electrometer Null Setting Cable subtraction Proper bias >300 V for ion chamber 100 V for diamond 0 v for all diodes
Proper gain Proper mode
![Page 19: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/19.jpg)
ID/AAPM/10
Choose Consistent & Correct Polarity
![Page 20: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/20.jpg)
ID/AAPM/10
Chambers & Gain
![Page 21: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/21.jpg)
ID/AAPM/10
Selection of detector for beam data
![Page 22: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/22.jpg)
ID/AAPM/10
XYZ
XY
Z
Scan Direction
Chamber Orientations
Radiation beam
Choice of Detector Orientation
![Page 23: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/23.jpg)
ID/AAPM/10
Detector Orientation
![Page 24: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/24.jpg)
ID/AAPM/10
0102030405060708090
100110
0 0.5 1 1.5 2
Rel
ativ
e Dos
e
Distance Off Axis (cm)
6MV, 2x2 cm2 field, Illustration of chamber volume effects
Diode, dmax
PTW Pinpoint, dmax
RK chamber, dmax
![Page 25: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/25.jpg)
ID/AAPM/10
![Page 26: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/26.jpg)
ID/AAPM/10
The possible setup error that causes the photon beam dose profile in figure is due to:
0
10
20
30
40
50
60
70
80
90
100
-4 -3 -2 -1 0 1 2 3 4Distance (cm)
Rel
ativ
e D
ose 1.5 cm
10 cm20 cm
Answer: 3Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008
Question
1. Gantry tilt
2. Collimator rotation
3. Tank arm tilt
4. Gantry and arm tilt5. Gantry tilt, collimator rotation and
tank arm tilt
![Page 27: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/27.jpg)
ID/AAPM/10
0 2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0
Depth (mm)
10
20
30
40
50
60
70
80
90
100
0
Dos
e (%
)
Surface & Buildup Dose, 6 MV
0.6 cc0.3 cc0.125 ccMarkusA16IC-4PinPointDiamondPFDSFD
Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008
Answer: 2
Question
The accurate measurement of the buildup and surface dose could be achieved by a:
1. Diode detector2. Parallel plate chamber3. Cylindrical chamber4. Diamond detector5. PinPoint chamber
![Page 28: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/28.jpg)
ID/AAPM/10
0
50
100
150
200
250
-20 -15 -10 -5 0 5 10 15 20
Rel
ativ
e D
ose
Distance Off Axis(cm)
6 MV 60 Deg Wedge, 10 cm depth: water vs Profiler
Diode Array Profile
Water Profile
![Page 29: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/29.jpg)
ID/AAPM/10
0
20
40
60
80
100
120
140
160
180
200
-20 -15 -10 -5 0 5 10 15 20
Rel
ativ
e D
ose
Distance Off Axis(cm)
18 MV 60 Degree Wedge, 10 cm depth, 100 cm SAD: Water vs Profiler
Diode Array profile
Water Profile
![Page 30: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/30.jpg)
ID/AAPM/10
Scanning Speed
![Page 31: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/31.jpg)
ID/AAPM/10
Scanning Speed
![Page 32: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/32.jpg)
ID/AAPM/10
0
20
40
60
80
100
0 5 10 15 20 25 30 35
Perc
ent D
epth
Dos
e
Depth(cm)
Arithmetic Mean (AM) smoothing: 60 Degree Wedge PDD
UnsmoothedAM x 1AM x 2AM x 3AM x 5
![Page 33: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/33.jpg)
ID/AAPM/10
![Page 34: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/34.jpg)
ID/AAPM/10
Electron beam depth doses shown in figure represents the problem of
0
0.2
0.4
0.6
0.8
1.0
1.2
0 2 4 6 8 10 12 14 16 18 20
Depth, z (cm)
Dep
th D
ose
(%)
6 MeV9 MeV
12 MeV
15 MeV
18 MeV
21 MeV
Answer: 4Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008
Question
1. Noise in the cable2. Electrometer gain3. Bias on the electrometer4. Speed of scanning5. Tuning of accelerator
![Page 35: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/35.jpg)
ID/AAPM/10
Future of Beam Data Commissioning
Standardization of linear acceleratorsMonte Carlo based commissioning Newer Radiation Detectors & Cables Newer Scanning Systems Smart algorithms
![Page 36: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/36.jpg)
ID/AAPM/10
Aubin et al., Med Phys, 37(5), 2279-2288, 2010
![Page 37: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/37.jpg)
ID/AAPM/10
Monte Carlo Codes
Aubin et al., Med Phys, 37(5), 2279-2288, 2010
![Page 38: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/38.jpg)
ID/AAPM/10
Simulation of intensity at target
Aubin et al., Med Phys, 37(5), 2279-2288, 2010
![Page 39: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/39.jpg)
ID/AAPM/10
Simulated Profiles
Aubin et al., Med Phys, 37(5), 2279-2288, 2010
![Page 40: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/40.jpg)
ID/AAPM/10
Profiles for different fields
![Page 41: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/41.jpg)
ID/AAPM/10
Depth Dose Simulation
Aubin et al., Med Phys, 37(5), 2279-2288, 2010
![Page 42: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/42.jpg)
ID/AAPM/10
Detector Manufacturer Type volumeSFD Scanditronix Photon diode 1.7x10-5cm3
PFD Scanditronix Photon diode 1.9x10-4cm3
Exradin A-16 Standard Imaging Ion chamber 0.007cm3
Wellhofer-IC4 Scanditronix Ion chamber 0.40 cm3
Pinpoint PTW Ion chamber 0.015cm3
0.125cc PTW Ion chamber 0.125cm3
0.3cc PTW Ion chamber 0.3 cm3
0.6cc PTW Ion chamber 0.6 cm3
Diamond PTW Diamond 0.003cm3
Markus PTW Parallel plate 0.055cm3
Edge Detector Sun Nuclear Diode 10-5cm3
Other
Detectors
![Page 43: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/43.jpg)
ID/AAPM/10
Sensitivity vs Volume of Detectors
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
1.0E-04 1.0E-03 1.0E-02 1.0E-01 1.0E+00
Volume (cm3)
Rela
tive s
ensit
ivity
PFD
SFD
A-16 PinPointMarkus
IC4
0.125cc
0.3cc
0.6 cc
![Page 44: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/44.jpg)
ID/AAPM/10
Beam Profile
0
10
20
30
40
50
60
70
80
90
100
110
0 10 20 30 40 50 60 70 80Distance (mm)
Dos
e (%
)
0.6 cc0.3 cc0.125 ccMarkusA16IC-4PinPointDiamondPFDSFD
Answer: 5Reference: Das et al, TG-106, Med. Phys. 35(9), 4186-4214, 2008
Question
Photon beam dose profiles taken with various detectors as shown in figure is possibly due to:
1. Speed of scanning2. Beam asymmetry3. Pb piece in the beam4. Hysteresis of scanning system5. Orientation of scanning detector
![Page 45: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/45.jpg)
ID/AAPM/10
Comparison of Large Tank and Small SRS Cylinder Tank for SRS, TMR & Profiles
ARM Inc., Port Saint Lucie, FL 34983
Moving Tank System for TMR
![Page 46: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/46.jpg)
ID/AAPM/10
No SSD to SAD calculations required, No cubic spline fit of a limited number of fixed data points needed
0
0.2
0.4
0.6
0.8
1
1.2
0
0.2
0.4
0.6
0.8
1
1.2
0 50 100 150 200 250 300 350
Calculated TPR
Measured TPR
Calculated TPR ~2% less at depth Cubic spline fit of 12 data points
Nikesch et al, CyberKnife Center , Palm Beach, FL
Direct TMR Data Acquisition
![Page 47: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/47.jpg)
ID/AAPM/10
Scanner Orientation Advantage
![Page 48: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/48.jpg)
ID/AAPM/10
Sun Nuclear 3D Scanner
2
1
1
32
1. Ring drive maintains consistent scanning direction
2. Diameter drive has maximum scanning range of 640mm
3. Vertical drive has maximum travel of 400mm
![Page 49: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/49.jpg)
ID/AAPM/10
Conclusions Golden Data should be taken as a
reference only Understand time and amount of data
to be taken View each parameters properly,
double check by another individual Use proper detector for each type of
data collection Set optimum speed for scanning, do
not rush
![Page 50: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/50.jpg)
ID/AAPM/10
-Conclusions
Understand the limits and measuring condition
Question every unusual data set Do not smooth data too much Write report for future reference Future technology & resources
could help commissioning simpler
![Page 51: Commissioning and Calibrating a Linear Accelerator – State ... · Fourth, although acceptable agreement with the golden data set may be found ... 2. Parallel plate chamber 3. Cylindrical](https://reader030.vdocuments.us/reader030/viewer/2022020303/5b3c157b7f8b9a5e1f8d2b69/html5/thumbnails/51.jpg)
ID/AAPM/10
Thanks