COMBINED DENSITY AND SIZE-BASED SORTING IN DETERMINISTIC LATERAL DISPLACEMENT DEVICES Stefan H. Holm 1* , Jason P. Beech 1 and Jonas O. Tegenfeldt 1 1 Solid State Physics and Nanometer Structure Consortium (nmC@LU), Lund University, Sweden ABSTRACT We present a deterministic-lateral-displacement (DLD) device that extends the capabilities of this traditionally size- based particle separation technique to also be sensitive to density. By the use of T-shaped posts instead of the normally cylindrical posts the particle trajectory through the device will be a function of its vertical position which in turn is de- termined by the buoyancy of the particles. The potential lies in fast sorting of complex biological samples together with diagnosis and treatment-monitoring of diseases affecting cell-density, eg. cancer, sickle-cell anemia and malaria. We demonstrate proof-of-principle of combined size-and-density-based sorting, specifically particles of identical size but different density. KEYWORDS: Deterministic Lateral Displacement, Density, Fractionation INTRODUCTION DLD is a powerful microfluidic technique first shown capable of separating particles based on size[1] and recently shown by our group to have the potential to discriminate cells based on shape[2] and deformability[3]. In this work we implement separation based on density in a DLD device by the use of T-shaped posts instead of the traditionally cylindrical posts, (Fig 1C). As the critical size above which particles are laterally displaced is a function of the distance between the posts (the gap, G) we achieve a device with two separate critical sizes. As the vertical position of a particle depends on the buoyancy force the trajectory through the device will be a function of both the particle size and density. It has been shown that cell density is a highly specific biomarker with an intrinsic cell-to-cell variation a 100-fold lower than that of volume or mass variation[4]. Further , the cell density is known to vary over the course of the cell- cycle[5], due to apoptosis[6] and diseases such as malaria, sickle cell anemia and cancer[7]. With these characteristics, cell density is a highly interesting biomarker which combined with the power of DLD have the potential to open up for new separation schemes. THEORY In DLD the sample flows through a symmetric array of micrometer-sized posts under laminar flow conditions. Hydrodynamic and steric interactions between posts and particles give rise to a separation where larger particles are deflected at a specific angle with respect to the flow. However, small particles are not affected by these interactions and travel through the device without being laterally displaced, Fig 1A, thus achieving a separation of the particles. The specific size above which particles are deflected is known as the critical size. This value, D c , is a function of the periodicity of the array, N, and more important for this work the distance between two posts, the gap (G), see the unit cell in Fig 1B. Traditionally the posts of a DLD device have been cylindrical with a single gap, and a single critical size as a result. However, by the use of T-shaped posts we obtain a device with two distinct gaps and consequently two separate critical sizes. The particles trajectory through the device will thus be a function of both its size and also its lateral position. Figure 1. (A) Schematic illustration showing a section of a DLD device, with posts colored blue. The hydro- dynamic center of the small yellow particle is able to fit within the black flow stream while the larger red parti- cle is laterally displaced to the adjacent flow stream for each row. (B) A unit cell in the DLD array together with definitions of the period and the critical size. (C) Illustration showing the normal cylindrical posts and the T-shaped posts used in this work. With T-shaped posts two different gaps, and consequently two separate criti- cal diameters are present, where the particles' vertical positions determine their behavior in the device. 978-0-9798064-6-9/μTAS 2013/$20©13CBMS-0001 1224 17th International Conference on Miniaturized Systems for Chemistry and Life Sciences 27-31 October 2013, Freiburg, Germany

Upload: others

Post on 19-May-2020

11 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: COMBINED DENSITY AND SIZE-BASED SORTING IN … · COMBINED DENSITY AND SIZE-BASED SORTING IN DETERMINISTIC LATERAL DISPLACEMENT DEVICES Stefan H. Holm1*, Jason P. Beech1 and Jonas

COMBINED DENSITY AND SIZE-BASED SORTING IN DETERMINISTIC LATERAL DISPLACEMENT DEVICES

Stefan H. Holm1*, Jason P. Beech1 and Jonas O. Tegenfeldt1 1Solid State Physics and Nanometer Structure Consortium (nmC@LU), Lund University, Sweden

ABSTRACT

We present a deterministic-lateral-displacement (DLD) device that extends the capabilities of this traditionally size-based particle separation technique to also be sensitive to density. By the use of T-shaped posts instead of the normally cylindrical posts the particle trajectory through the device will be a function of its vertical position which in turn is de-termined by the buoyancy of the particles. The potential lies in fast sorting of complex biological samples together with diagnosis and treatment-monitoring of diseases affecting cell-density, eg. cancer, sickle-cell anemia and malaria. We demonstrate proof-of-principle of combined size-and-density-based sorting, specifically particles of identical size but different density. KEYWORDS: Deterministic Lateral Displacement, Density, Fractionation

INTRODUCTION

DLD is a powerful microfluidic technique first shown capable of separating particles based on size[1] and recently shown by our group to have the potential to discriminate cells based on shape[2] and deformability[3].

In this work we implement separation based on density in a DLD device by the use of T-shaped posts instead of the traditionally cylindrical posts, (Fig 1C). As the critical size above which particles are laterally displaced is a function of the distance between the posts (the gap, G) we achieve a device with two separate critical sizes. As the vertical position of a particle depends on the buoyancy force the trajectory through the device will be a function of both the particle size and density.

It has been shown that cell density is a highly specific biomarker with an intrinsic cell-to-cell variation a 100-fold lower than that of volume or mass variation[4]. Further , the cell density is known to vary over the course of the cell-cycle[5], due to apoptosis[6] and diseases such as malaria, sickle cell anemia and cancer[7]. With these characteristics, cell density is a highly interesting biomarker which combined with the power of DLD have the potential to open up for new separation schemes. THEORY

In DLD the sample flows through a symmetric array of micrometer-sized posts under laminar flow conditions. Hydrodynamic and steric interactions between posts and particles give rise to a separation where larger particles are deflected at a specific angle with respect to the flow. However, small particles are not affected by these interactions and travel through the device without being laterally displaced, Fig 1A, thus achieving a separation of the particles. The specific size above which particles are deflected is known as the critical size. This value, Dc, is a function of the periodicity of the array, N, and more important for this work the distance between two posts, the gap (G), see the unit cell in Fig 1B.

Traditionally the posts of a DLD device have been cylindrical with a single gap, and a single critical size as a result. However, by the use of T-shaped posts we obtain a device with two distinct gaps and consequently two separate critical sizes. The particles trajectory through the device will thus be a function of both its size and also its lateral position.

Figure 1. (A) Schematic illustration showing a section of a DLD device, with posts colored blue. The hydro-dynamic center of the small yellow particle is able to fit within the black flow stream while the larger red parti-cle is laterally displaced to the adjacent flow stream for each row. (B) A unit cell in the DLD array together with definitions of the period and the critical size. (C) Illustration showing the normal cylindrical posts and the T-shaped posts used in this work. With T-shaped posts two different gaps, and consequently two separate criti-cal diameters are present, where the particles' vertical positions determine their behavior in the device.

978-0-9798064-6-9/µTAS 2013/$20©13CBMS-0001 1224 17th International Conference on MiniaturizedSystems for Chemistry and Life Sciences27-31 October 2013, Freiburg, Germany

Page 2: COMBINED DENSITY AND SIZE-BASED SORTING IN … · COMBINED DENSITY AND SIZE-BASED SORTING IN DETERMINISTIC LATERAL DISPLACEMENT DEVICES Stefan H. Holm1*, Jason P. Beech1 and Jonas

EXPERIMENTAL

Flow simulations of the device was carried out using COMSOL Multiphysics 4.3 (COMSOL AB, Sweden). The result for the flow through a unit cell is shown in Fig 2A together with horizontal slices of the unit cell showing the flow field for the small and the large gaps respectively. The graph displays the flow profiles along the shortest lines between the posts in the gaps.

The final device is ~10 cm long and ~1 cm in width, shown in Fig 2B-C. The particles enter through the sample inlet and if there is no interaction with the device they end up in outlet 1. However, if they are larger than the critical size at their lateral position they will be laterally displaced into outlet 2.

The unit cell width, λ, is 36 µm and the T-shaped post radii are 8 and 13 µm. Together with a periodicity, N, of 18 the resulting critical sizes in the device are expected to be 3.50 µm and 7.00 µm.

A master for replica molding was fabricated using two precisely aligned UV-lithography (MJB4, SUSS MicroTec, Germay) steps in 8µm thick SU8 (MicroChem, USA) on silicon wafers. Devices were then produced using standard soft lithography and bonded to a glass slide treated with oxygen plasma (Plasmatic Systems Inc, USA). Overpressure was used to induce flow in the device (MFCS-FLEX, Fluigent, France). Devices were flushed with 0.1% Tween for 5 minutes prior to the experiment to avoid non-specific sticking.

Microspheres of various sizes and materials were used: silicon (Si) microspheres, ρ=2.33 g/cm3, 3.13±0.42 µm and 4.63±0.46 µm (Bangs laboratories Inc, USA) and fluorescent polystyrene (PS) microspheres, ρ=1.05 g/cm3 4.87±0.28 µm and 3.10±0.15 µm (Polysciences Inc, USA. Concentrations were ~104/µL for samples with just one type of material and ~103/µL for the sample containing particles of different materials. In our device the small spheres are expected to be below the smaller critical size while the larger spheres are between the smaller critical size and the larger critical size. The buffer used was 15% sucrose (Sigma-Aldrich LLC, USA) in DI-water solution. The lateral positions of the particles at the end of the DLD array were determined using our own particle tracking software coded in MATLAB R2012b (MathWorks, MA, USA).

RESULT AND DISCUSSION

As expected the heavy silicon microspheres (negative buoyancy) sediment to the bottom and depending on the orienta-tion of the device they experience the low or the high critical size (Fig 3A). In this case they are only separated for the small gap with an expected critical size of 3.50 µm and not in the large gap with an expected critical size of 7.00 µm.

Figure 2. (A) FEM simulations performed in COMSOL Multiphysics 4.3a (Burlington , Ma, USA) to investigate how T-shaped posts affect the flow. First shown is a 3D model of a unit cell with the velocity field color coded (red - high speed and blue - low speed). The 2D-plots are horizontal slices of the unit cell showing the flow profiles at the center height of the small and large gap section respectively. The flow pro-file between two posts, which determines the critical diameter in the DLD device was then plotted (gap size normalized). From here the critical particle sizes, Dc, can readily be determined. (B) SEM-micrograph showing an overview of the device with the inlets and outlets and the separation array. (C) SEM-micrographs of the device inlet followed by higher magnified micrographs of the center section of the de-vice showing the T-shaped posts.

1225

Page 3: COMBINED DENSITY AND SIZE-BASED SORTING IN … · COMBINED DENSITY AND SIZE-BASED SORTING IN DETERMINISTIC LATERAL DISPLACEMENT DEVICES Stefan H. Holm1*, Jason P. Beech1 and Jonas

More interestingly, mixing particles of different density we demonstrate that we can separate particles that have similar size and that differ in density only (Fig. 3C). Here, the buoyance on the PS particles forces them to flow into the upper sec-tion of the device with a critical size above the particle sizes while, again the silicon particles sediment into lower level with a lower critical size which is between the size of the particles, and consequently only the large Si particles are displaced to-wards outlet 2.

CONCLUSION

We have successfully demonstrated a proof-of-principle device where separation is sensitive not only to size but also density of the particles. In contrast to other alternative single-cell techniques for density characterization[4], this ap-proach is entirely passive. ACKNOWLEDGEMENTS

The work presented herein was made possible by the support of the Nanometer Structure Consortium at Lund Uni-versity (nmC@LU), the Swedish Research Council and the Crafoord Foundation. REFERENCES [1] Huang, L.R., et al., Science, 2004. 304(5673): p. 987-990. [2] Holm, S.H., et al., Lab on a Chip, 2011. 11(7): p. 1326-1332. [3] Beech, J.P., et al., Lab on a Chip, 2012. 12(6): p. 1048-1051. [4] Grover, W.H., et al., Proceedings of the National Academy of Sciences, 2011. [5] Pertoft, H. and A.D. Wolff, J Cell Bio, 1972. 55(3): p. 579-585. [6] Morris, A.H. and R.G. Wyllie, Am J Pathol., 1982: p. 78-87. [7] Bosslet, R.R., P. Altevogt, and V. Schirrmacher, Br J Cancer., 1981. 44(3): p. 356-362. CONTACT *S.H. Holm, tel: +46 (0)70-4099380; [email protected]

Figure 3. (A) False-colored time-integrated micrograph of a 4.87 µm PS bead (green) and a 4.63 µm Si sphere (yellow). The different lateral position can clearly be seen with the green particle above the small gaps. (B) Graph showing the distribution of silicon (Si) microspheres, 3.13 µm (red) and 4.63 µm (yellow), in DI water at the outlet of the device. The illustrations in the middle show how the device was oriented. When the device is flipped over the Si spheres sediment into the level with smaller gaps and the larger spheres are later-ally displaced into outlet 2. (C) Outlet distribution of both large and small Si (yellow, red) and PS (green, black) microspheres in 15% sucrose solution. The buoyancy of the PS microspheres forces them into a vertical position with large gaps while the silicon microspheres sediment into a vertical position with small gaps. Combining of size and density-based separation allows for the sorting out a single specific particle type.