close coupled field accredited research results...penicillium chrysogenum, bacillus subtilis var...
TRANSCRIPT
CLOSE COUPLED FIELD
ACCREDITED RESEARCH RESULTS
This document contains summaries of reports presented by laboratories carrying out
validation research into the efficacy of the technology
The laboratories concerned, and their respective reports, are:
Microsearch Laboratories Ltd
A. A Summary of the Anti-Microbial Performance of the AirManager
Technology applied to a range of Micro-organisms and Viral Particles
B. A Study into the Removal of Tobacco Smoke Analytes
University of Central Lancashire
A. Assessment of Ozone Leakage from the AirManager Unit
B. Passive Sampling of Ozone Leakage over a prolonged period
CCFRA Technology Ltd
A. A Summary of the Anti-Microbial performance of the AirManager
Technology applied to Pseudomonas aeruginosa, Staphylococcus aureus,
Penicillium chrysogenum, Bacillus subtilis var globigii and Neurospora
(Chrysonilia) sitophila.
Mountain Heath Services Ltd
A. Volatile Organic Compounds tested at 100 x maximum OEL
Microbial Innovations Ltd
A. Methylene Chloride Removal using Close Coupled Field technology.
Clean Air Solutions Ltd
A. Particle reduction and arrest capabilities of Close Coupled Field
technology.
A summary of the performance characteristics of the
Airmanager range of atmospheric treatment devices
Quest International (Airmanager) have developed a range of devices designed for
microbiological decontamination of atmospheres. This technology employs closed
coupled field technology for the contained generation of Ozone and ions, in tandem
with terminal electrostatic filtration. of the air stream.
Combining these technologies in a manner which affords a high flow rate permits
the effective treatment of large volumes of atmosphere in a manner which we
believe represents a significant advance in the area of air management.
This document summarises the anti-microbial performance data, obtained by an
independent assessment laboratory.
Ozone production characteristics
Ozone is a highly oxidative molecule and may be generated by UV radiation or the effects of electrical
discharge on molecular Oxygen in air.
The Airmanger device generates ozone by electrical discharge employing patent pending technology.
The European standard for atmospheric levels of ozone is currently 0.2 ppm while according to
various literature sources the required dosage of ozone required to inactivate microbial systems, on
contact , varies between 0.05 ppm and 0.4 ppm.
An important aspect of this validation effort has been to demonstrate compliance of the device with
European ozone emission standards, whilst additionally producing evidence of sufficient ozone
generation to accommodate effective competence with regard to the task of broad scope anti-
microbial activity.
A key advantage of the Air manager Device is the claim that ozone generation and reactions with
micro-organisms, occur contained solely within the device resulting in decontamination with no
measurable emission of ozone.
Addtionally it is claimed that the electrostatic filter stage serves only to capture oxidised debris and
remove particulates from the air.
Ozone measurements
These claims have been investigated employing a novel probe by which ozone production is
determined by measurement of the degree of oxidation obtained with a d-alpha tocopherol coating
during exposure Trials have been conducted to measure ozone production within the device and the
potential for environmental accumulation during use, with and without filter in situ.
The following tables summarise our findings.
Measurement of ozone production by d-alpha Tocopherol probe oxidation with filter in place
Run Time hours
03 ppm within
treatment chamber 0
3 ppm within 60 m
3
Room
0 24 <0.2
6 103 <0.2
12 94 <0.2
18 107 <0.2
24 102 <0.2
Measurement of ozone production by d-alpha Tocopherol probe oxidation without filter in place
Time hours
03 ppm within
treatment chamber 0
3 ppm within 60 m
3
Room
0 41 <0.2
6 96 <0.2
12 97 <0.2
18 104 <0.2
24 106 <0.2
Our data indicates no significant emission of ozone from the device could be measured over a 24
hour period in the operating environment.
Measurement indicates that significantly higher levels of Ozone are produced within the closed
coupled field device than predictably are required for contact inactivation all classes of micro-
organisms for which susceptibility has been published.
Microbiological aspects of Filter performance
Electrostatic air filtration is know to produce reduction in the levels air borne microbial contaminants.
A possible problem with stand alone filtration devices is therefore the accumulation of possibly
infective or otherwise unwanted viable contamination within the structure of the filter during life
span. We have conducted trials to monitor these possibilities and present below the data obtained
showing the recovery of differing classes of organism from the interior surfaces of the terminal filter
are different periods of operation in waste processing room.
Recovery of viable micro-organisms from electrostatic filter material after differing periods of usage
Operation interval
TVC cm3
Filter material
Moulds cm
3 Filter
material
Yeasts cm
3 Filter
material
Bacillus sps cm
3 Filter
material
Gram neg sps cm
3 Filter
material
Gram Pos sps cm
3 Filter
material
1 day <10 <10 <10 <10 <10 <10
1 week <10 <10 <10 <10 <10 <10
1 month <10 <10 <10 <10 <10 <10
4 months <10 30 <10 20 <10 80
Conclusions
Our data demonstrates that in an environment know to exhibit high levels of airborne microbial
contamination significant build up of viable organisms occurred in the filtration unit up to and
including three months of use. This effect may caused by impingement of residual zone on the active
surfaces, loss of viability due to dehydration in the high flow rate of air, nutrient scarcity or a
combination of these and other factors.
Such findings to some degree support the anti-microbial efficiency of the ozone generation system
presented below. More importantly these findings suggest that in respect of bacteria and fungi the
filtration stage is unlikely to represent a biological hazard during replacement.
Demonstration of Anti-microbial competence
The following experimental data reports on the performance of the M4/4 device in relation to the
reduction of single pass microbial challenges. Performance at each of four levels level of flow rate has
been determined for a range of organisms with and without electrostatic filtration in place.
M4/4 Single pass performance with electrostatic filtration
Organism Challenge level cfu/l-
1 Speed 1 Recovery
cfu/l-1
Speed 2 Recovery
cfu/l-1
Speed 3 Recovery
cfu/l-1
Speed 4 Recovery
cfu/l-1
A.niger 8.80E+06 <1 <1 <1 <1
S.typhimurium 7.40E+06 <1 <1 <1 <1
C.Albicans 6.00E+06 <1 <1 <1 <1
S.aureus 7.10E+06 <1 <1 <1 <1
B.cereus 2.20E+06 <1 <1 <1 1.30E+02
M4/4 Single performance with no electrostatic filtration
Organism Challenge level cfu/l-
1 Speed 1 Recovery
cfu/l-1
Speed 2 Recovery
cfu/l-1
Speed 3 Recovery
cfu/l-1
Speed 4 Recovery
cfu/l-1
A.niger 7.00E+06 <1 <1 <1 <1
S.typhimurium 8.40E+06 <1 <1 <1 <1
C.albicans 8.30E+06 <1 <1 <1 <1
S.aureus 9.20E+06 <1 <1 <1 <1
B.cereus 4.70E+06 <1 <1 3.10E+02 9.80E+02
Speed 1 = standard flow rate 190 m3
Conclusions :
All conditions of treatment produced significant levels of reduction in the levels of air borne
challenges.
Under these challenge conditions Increment 3 gave a 100 % performance with filter in place while
increment 2 gave a 100 % performance with no filter in place.
With filter in place all organisms were reduced to non detectable levels at increment 4 with the
exception of Bacillus cereus where only a 4 log reduction was achieved.
My recommendation is to limit flow to increment 3 (570 m3/hour) with the filter in situ, to guarantee a
consistent and rapid degree of air processing.
Continuos dosage lethality with a range of micro-organisms
In this series of trials a wide range of microbial types were continuously introduced at the intake
section of the Airmanager device for a period of 1 hour. During the exposure time periodic
measurements were taken at the output section and the levels of survivors were determined. The
following results were obtained.
Airmanager performance against continuos input of Bacteria and Fungi
Organism Class Mean cfu/m3/Hr
at input Treatment stream
Mean cfu/m
3/Hr
post Treatment exit stream
Mean decline Log/cfu/m
3/Hr
post Treatment exit stream
Apparent percentage reduction
Escherichia coli Gram –ve 2.1E+05 0.0E+00 >5 >99.999
S.tyhpi murium Gram –ve 4.6E+05 0.0E+00 >5 >99.999
E.agglormerans Gram –ve 3.9E+05 0.0E+00 >5 >99.999
E.gergoviae Gram –ve 4.2E+05 0.0E+00 >5 >99.999
A.aerogens Gram –ve 7.1E+05 0.0E+00 >5 >99.999
S.marcescens Gram –ve 8.2E+05 0.0E+00 >5 >99.999
E.sakazakii Gram –ve 3.4E+05 0.0E+00 >5 >99.999
Ecoli 0157 H:7 Gram –ve 3.5E+05 0.0E+00 >5 >99.999
P.aeruginosa Gram –ve 6.1E+05 0.0E+00 >5 >99.999
P.putida Gram –ve 8.2E+05 0.0E+00 >5 >99.999
S.aureus oxford Gram +ve 4.3E+05 0.0E+00 >5 >99.999
S.aureus MSRA Gram +ve 4.8E+05 0.0E+00 >5 >99.999
S.epidermidis Gram +ve 3.7E+05 0.0E+00 >5 >99.999
M.luteus Gram +ve 9.0E+05 0.0E+00 >5 >99.999
S.faecalis Gram +ve 7.3E+05 0.0E+00 >5 >99.999
S.pyogenes Gram +ve 3.6E+05 0.0E+00 >5 >99.999
B.cereus Gram +ve 7.1E+05 0.0E+00 >5 >99.999
B.globigii G+ve Spore 7.9E+05 1.0E+01 >5 99.999
B.subtilis G+ve Spore 2.1E+05 3.0E+01 >5 99.986
B. megaterium G+ve Spore 6.2E+05 9.0E+01 >5 99.985
S.cerevisiea Yeast 4.3E+05 0.0E+00 >5 >99.999
S.bailli Yeast 7.2E+05 0.0E+00 >5 >99.999
Pichia mixed sps Yeast 6.3E+05 0.0E+00 >5 >99.999
S.ludwigii Yeast 6.0E+05 0.0E+00 >5 >99.999
A.niger Mould mycelial 6.2E+05 0.0E+00 >5 >99.999
A.flavus Mould mycelial 7.8E+05 0.0E+00 >5 >99.999
F.poea Mould mycelial 7.2E+05 0.0E+00 >5 >99.999
P.digitatum Mould mycelial 6.9E+05 0.0E+00 >5 >99.999
F graminerium Mould mycelial 4.3E+05 0.0E+00 >5 >99.999
A.niger Mould Spore 8.2E+05 7.0E+01 >5 99.991
A.flavus Mould Spore 6.7E+05 5.0E+01 >5 99.993
F.poea Mould Spore 8.2E+05 0.0E+00 >5 >99.999
P.digitatum Mould Spore 6.7E+05 0.0E+00 >5 >99.999
F graminerium Mould Spore 2.9E+05 0.0E+00 >5 >99.999
Airmanager performance against continuos input of viral particles
Organism Class Mean cfu/m3/Hr
at input Treatment stream
Mean cfu/m
3/Hr
post Treatment exit stream
Mean decline Log/cfu/m
3/Hr
post Treatment exit stream
Apparent percentage reduction
CTX SS DNA 4.3E+12 8.1E+02 >12 >99.999
ScV-L-BC DS RNA 9.2E+12 4.6E+02 >12 >99.999
FcoV (attenuated) SS + RNA 7.1E+12 3.0E+02 >12 >99.999
T4 Phage DS DNA 5.3E+12 7.4E+02 >12 >99.999
Conclusions
The air manager device has demonstrated a high level of competence in the inactivation of a wide
range of micro-organisms including, Bacteria cells, Bacterial spores, virus particles, Mould, Mould
spores and Yeasts. Kill efficiencies in excess of Log 12 were obtained consistently for all classes of
viral particle examined, while for all other classes of organism no less than a Log 5 kill was obtained
on a continuous basis. In summary the device is highly effective at killing micro-organisms.
In Use Assessments
Detailed below are the results of environmental trials during which the capacity of the Airmanager
device was assessed in relation to the clean up of natural airborne population of micro-organisms.
The efficiency of the Airmanager ‘M4’ air treatment unit in the
sanitisation of a Laboratory incubator room
Introduction :
In spite of good compliance to GLP standards it is possible that laboratories may still develop
problems associated with airborne microbial contamination. Usually such problems are detected by
routine environmental surveillance or incidences of contamination on solid agar plates.
In this study a problem was investigated relating to a persistent environmental contamination in a
commercial grain testing laboratory. This facility had reported significant levels of in lab mould
contamination of both blank plates and plates intended for the isolation of yeasts and moulds from
samples. In house environmental analysis by settle plate had determined the presence of identical
isolates to those found on the plates in the atmosphere of the incubation room.
This isolate responsible for the contamination was confirmed as Fusarium poae. This organism is
common in temperate regions and is associated with commodities such as Wheat and Maize both of
which were commonly handled by the facility. It demonstrates growth over the range 2.5 'c to 33'c and
characteristically, on common mycological media produces profuse growth with Salmon or pale pink
colonies.
Trial outline :
The trial was conducted in two stages. During the first month of monitoring the M4 device was not in
operation and air sampling was conducted on an hourly basis between the hours of 9.00am and 6.00
p.m. over a six day working week Sampling was conducted employing a Cassela volumetric
sampler with impaction onto Oxytetracycline glucose yeast agar. During week one (device off) 0.1,
0.2, and 0.5 L-1
air volumes were taken at the specified interval.
The device was operative during month two. Sampling was conducted to the same schedule described
above with an identical sampling procedure.
Simultaneously during the trial records were kept of non compliant contaminated agar intended for
use in analytical procedures.
Results :
Table : Mean air quality in a laboratory incubation area and media quality over a two month period
with and without the M4 device in operation.
Device status
Week F. poae cfu/L-1
/Air Percentage in lab plate
contamination
OFF 1 17100 3
OFF 2 21300 2
OFF 3 16700 9
OFF 4 18900 3
Mean 18500 4.25
ON 5 20 <1
ON 6 40 <1
ON 7 2 <1
ON 8 3 <1
Mean 16.25 <1
Conclusions ;
In this laboratory an overt problem had been experienced in relation to media contamination which
was directly related to environmental cross contamination with Fusarium poae.
The operation of the M4 device in the area which was the source of this problem successfully reduced
the level of contamination on a consistent basis by between 2 and 3 log cycles L-1 air
. This magnitude
of effect was sufficient to reduce the level of media contamination to a non detectable level.
On this basis the M4 device has been shown to be an effective tool in the maintenance of a
microbiological laboratory air quality.
A laboratory investigation of the Microbiological and Sensory
efficiency of the Quest Ozonation device
Quest International (UK) Ltd have developed a innovative device which through a novel piece of
engineering is capable of treating atmospheres with very high doses of Ozone with a free fan
transfer volume of 190 m3/hour. Additionally the device is capable of operating with either 'cutting
edge' replaceable Electrostatic filters ( FR = 160 m3/hour) or alternatively with H.E.P.A. filters ( FR =
65 m3/hour) as a final air treatment.
It was theorised that usage of the device with Electrostatic filtration in combination with ozonation
would afford both reduction of airborne micro-organisms and good odour decontamination
characteristics while employment of H.E.P.A. filtration was anticipated to produce superior
microbiological performance.
Prior to this work no microbiological performance data had been obtained but it was known that
ozonation in combination with electrostatic filtration produced superior sensory results.
Quest International (UK) ltd invited Microsearch laboratories Ltd to conduct pilot testing to determine
the anti-microbial efficiency of the device.
We have achieved this goal by monitoring the reduction of air borne Gram negative bacteria
attributable to the Quest device over a seven day period in a microbiological laboratory waste
processing room. Measurement has included the performance characteristics of both filtration system
and has also included determination of sensory factors.
Trial 1 : Efficiency of the Quest Device in the sanitisation of the atmosphere in a Class II
microbiological laboratory waste room.
Outline :
The vast majority of contemporary microbiological laboratories are equipped with a designated area
designed to afford physical segregation of contaminated biological waste intended for sanitation by
autoclaving preceding safe disposal. Such waste consists of agar plates, cultures and implements
employed in microbiological manipulations.
In general such waste is extremely biologically active prior to treatment and may contain billions of
organisms per gram. While every effort in GLP is to prevent transfer of contaminants the nature of the
autoclaving process requires that storage bags are open to the atmosphere at the start of processing. As
a consequence the opportunity exists for introduction of large masses of organisms or spores into the
environment. Factually such areas exhibit high levels of airborne contamination.
These distribution factors coupled with the thermal currents created by autoclave operation engender a
demonstrably abundant and sustained level of airborne micro-organisms of many differing types. It is
true that such contamination is unlikely to present as a direct health risk through inhalation but such
an environment provides a useful model for efficiency studies of devices which purport to reduce
airborne levels of micro-organisms.
In this trial the regime involved the sampling of the atmosphere in the test environment by impaction
of air onto the surface of agar plates through the use of a Cassela air sampling device. The Cassela unit
is capable of accurately sampling a known volume atmosphere over a 30 second period and
continuously delivers the sampled air to an enclosed chamber. In this chamber an agar plate is exposed
to column of intake air whilst rotating thus distributing micro-organisms evenly over the surface of the
plate. Subsequent incubation of the plates allows enumeration of numbers of organisms present in the
original volume of atmosphere examined. Through the use of differing types of agar and a diagnostic
tests it is possible differentially count different types or classes of micro-organism.
Room Conditions : The room comprised of a 24.3 m3 cube. It contained an autoclave with treated
waste in one half and 25 Kg storage bags of untreated waste in the remaining floor area. At any one
time the area contained a minimum of 16 untreated waste bags of which between 8 and 10 bags would
be handled and processed in a working day between the hours of 9.00 am and 6.00 p.m. The sampling
device waste located centrally. Normally the room atmosphere is vented by forced extraction, this was
not the case during this trial. The autoclave hot exhausted was vented via an enclosed circuit which
we believe did not effect the atmospheric composition of the test environment.
Sampling Plan : Sampling occurred over a twenty four hour period at the intervals given in table one
below.
Such sampling extended over a seven day period with the Quest device running without any form of
filtration in place and without the ozone generator switched on. This was intended as a protocol to
demonstrate the background level of contamination. The data obtained is given in tables 1 and 3
below. The data gathered in this exercise was employed as the comparison set for all information
gathered during the subsequent period when the Quest device was operational as a sanitising unit.
Two further identically scheduled sampling periods were conducted sequentially separated by a four
day recovery gap. Firstly the device was operated with ozone on and an Electrostatic filter in place. In
the second session the device was operated with a H.E.P.A. filter in place again with an identical
sampling plan.
Microbiological analysis : The agar employed in all test was Violet Red Bile glucose agar ( VRBGA)
intended for the recovery of Gram positive organisms from the atmosphere through the use of the
Cassela device. Colonies were recovered on this agar after incubation at 35 ‘c for 24 hours. As the
trials were intended primarily to show overall comparisons , It was assumed that all isolates obtained
on VRBGA were Gram negative and all isolates were counted. Colonies were further differentiated on
the basis of Oxidase reaction. All sampling was conducted in duplicate.
Results : Table 1 below and associated graph provides the data obtained for Gram negative (Ox +ve
and Ox –Ve) isolates during the unsanitised control period and that for the data obtained during the
period of Ozone treatment associated with electrostatic filtration of the return air flow. Table 2
illustrates the average percentage kill through out the day attributable to the action of Ozonation and
Electrostatic filtration. Tables 3 and 4 summarise the same categories of data obtained for the period
when sanitisation was attempted employing Ozonation and H.E.P.A filtration.
Table 1:
Mean Microbial Levels over a 24 hour period in a microbiological waste room
( 23.4 M3) with and without continuous operation of the Quest Device
with electrostatic filtration
Time Condition Oxidase Pos Gram negative isolates L
-1/air
Oxidase Neg Gram negative isolates L-1/air
06:00 AM o3 off EF- 4.7E+03 8.6E+03
10:00 AM o3 off EF- 5.6E+03 9.2E+03
02:00 PM o3 off EF- 9.8E+03 2.8E+04
06:00 PM o3 off EF- 2.0E+04 3.7E+04
08:00 PM o3 off EF- 1.8E+04 3.3E+04
02:00 AM o3 off EF- 9.2E+03 1.9E+04
04:00 AM o3 off EF- 3.7E+03 8.7E+03
06:00 AM o3 on EF+ 6.0E+02 1.3E+03
10:00 AM o3 on EF+ 7.7E+02 1.3E+03
02:00 PM o3 on EF+ 1.6E+03 4.4E+03
06:00 PM o3 on EF+ 3.4E+03 6.2E+03
08:00 PM o3 on EF+ 3.4E+03 5.8E+03
02:00 AM o3 on EF+ 1.4E+03 3.0E+03
04:00 AM o3 on EF+ 4.8E+02 1.2E+03
06:00
AM10:00
AM02:00
PM06:00
PM08:00
PM02:00
AM04:00
AM
0.0E+00
5.0E+03
1.0E+04
1.5E+04
2.0E+04
2.5E+04
3.0E+04
3.5E+04
4.0E+04
time of day
Mean Gram negative levels per L-1
/air over a 24 hour
period in a 24.3 m3 waste room
OX Neg G Neg No t reat ment
OX Pos G Neg No t reat ment
OX Neg G Neg t reat ed
OX Pos G Neg t reated
Table 2 :
Mean Microbial % reduction levels over a 24 hour period in a microbiological waste
room ( 23.4 M3) with the QUEST DEVICE operating and with electrostatic filtration
Time Condition Oxidase Pos Gram negative isolates L
-1/air
Oxidase Neg Gram negative isolates L-1/air
06:00 AM o3 on EF+ 87.3 85.1
10:00 AM o3 on EF+ 86.2 85.4
02:00 PM o3 on EF+ 83.2 84.3
06:00 PM o3 on EF+ 82.8 83.2
08:00 PM o3 on EF+ 81.3 82.2
02:00 AM o3 on EF+ 85.3 84.6
04:00 AM o3 on EF+ 86.9 85.8
06:00 AM10:00 AM
02:00 PM06:00 PM
08:00 PM02:00 AM
04:00 AM
OX Neg G Neg
Ox Pos G Neg
78
79
80
81
82
83
84
85
86
87
88
Time of day
Mean Gram -ve % reduction per L-1/air over 24 hours
in a 24.3 m3 waste room
OX Neg G Neg
Ox Pos G Neg
Table 3:
Mean Microbial Levels over a 24 hour period in a microbiological waste room
( 23.4 M3) with and without continuous operation of the Quest Device
with H.E.P.A filtration
Time Condition Oxidase Pos Gram negative isolates L
-1/air
Oxidase Neg Gram negative isolates L-1/air
06:00 AM o3 off HEPA- 4.7E+03 8.6E+03
10:00 AM o3 off HEPA- 5.6E+03 9.2E+03
02:00 PM o3 off HEPA- 9.8E+03 2.8E+04
06:00 PM o3 off HEPA- 2.0E+04 3.7E+04
08:00 PM o3 off HEPA- 1.8E+04 3.3E+04
02:00 AM o3 off HEPA- 9.2E+03 1.9E+04
04:00 AM o3 off HEPA- 2.1E+03 8.7E+03
06:00 AM o3 On HEPA + 3.1E+03 5.7E+03
10:00 AM o3 On HEPA + 3.9E+03 6.2E+03
02:00 PM o3 On HEPA + 7.0E+03 2.0E+04
06:00 PM o3 On HEPA + 1.6E+04 2.8E+04
08:00 PM o3 On HEPA + 1.5E+04 2.7E+04
02:00 AM o3 On HEPA + 6.5E+03 1.4E+04
04:00 AM o3 On HEPA + 1.4E+03 5.7E+03
06:00
A M10:00
A M02:00
P M06:00
P M08:00
P M02:00
A M04:00
A M
0.0E+00
5.0E+03
1.0E+04
1.5E+04
2.0E+04
2.5E+04
3.0E+04
3.5E+04
4.0E+04
tim e of day
Mean Gram negative levels per L-1
/air over a 24 hour
period in a 24.3 m3 waste room
OX Neg G Neg No t r eat ment
OX Pos G Neg No t r eat ment
OX Neg G Neg t r eat ed
OX Pos G Neg t r eat ed
Table 4 :
Mean Microbial % reduction levels over a 24 hour period in a microbiological waste
room ( 23.4 M3) with the QUEST DEVICE operating and with H.E.P.A. filtration.
06:00 AM o3 On HEPA + 34.5 33.9
10:00 AM o3 On HEPA + 30.3 33.1
02:00 PM o3 On HEPA + 28.6 28.7
06:00 PM o3 On HEPA + 22.1 24.6
08:00 PM o3 On HEPA + 19.2 17.4
02:00 AM o3 On HEPA + 29.6 28.2
04:00 AM o3 On HEPA + 34.8 34.2
Trial 2 : This work was intended as a primer and involved the introduction of a number of single
calibrated doses of Aspergillus niger hyphae fragments and spore particles into a chamber having a 8
l-1
volume. The chamber was constructed in a manner as to permit access of the Quest device intake
grill to the interior of the chamber while the output section vented directly into a second chamber of
identical volume. Both chambers were vented by membrane filters for the purpose of pressure
equalisation. The purpose of the trial was to attempt to demonstrate a single pass efficiency in lethality
in a known airborne pathogen.
Dosing conditions : The biological material was delivered in the form of a fungal hyphae and spores
dispersed in Calcium silicate matrix. Both Chambers equipped with fans intended to assist dispersion.
Sampling was conducted via suction with collection in a 2 % sucrose/saline solution an involved 2 L-1
volume for each chamber. The device was not operational during dosing for 2 minutes after the
introduction of the biological material but had been previously stabilised for 30 minutes . After the
post dose period the device was operated for period of 1 minute and after which the atmosphere in the
delivery chamber was samples.
Analysis : Recovery solutions were examined by serial dilution and survivors were estimated on
Oxytetracycline glucose yeast agar ( 5 days at 25 @c). The results of these counts provided estimates
of the level of dosage and the level of survivors per l-1
of atmosphere before and after treatment.
Results ;
Tables 5 and 6 below provides the data obtained in this trial for instances of the device operating with
either the electrostatic or H.E.P.A. filter in place.
Table 5
Single pass efficiency for Aspergillus Niger (mixed Hyphae and spores)employing Electrostatic filtration and 0
3 dosing
Challenge level cfu/l
-1/air pre in take
Recovery level cfu/l
-1/air post
filter
Percentage Kill
8.3E+05 7.6E+04 90.807
4.2E+05 2.0E+03 99.524
6.1E+05 4.1E+03 99.328
7.3E+05 9.2E+03 98.740
7.2E+05 6.2E+03 99.139
7.4E+05 7.1E+03 99.041
8.2E+05 8.4E+03 98.976
6.3E+05 9.2E+03 98.540
Mean 98.012
Table 6
Single pass efficiency for Aspergillus Niger (mixed Hyphae and spores)employing H.E.P.A. filtration and 0
3 dosing
Challenge level cfu/l
-1/air pre in take
Recovery level cfu/l
-1/air post
filter
Percentage Kill
5.5E+05 8.0E+01 99.985
6.1E+05 9.0E+01 99.985
2.8E+05 3.0E+01 99.989
6.1E+05 9.0E+01 99.985
6.3E+05 8.0E+01 99.987
5.2E+05 8.0E+01 99.985
6.3E+05 1.1E+02 99.983
5.8E+05 5.0E+01 99.991
Mean 99.986
Trial 3 ; Sensory appreciation ;
Outline : A sensory evaluation was conducted each day during operation of the device in the
microbiological waste processing facility. This involved subjective scoring by four people according to
the Key given with Table 7, below. Evaluations were made for each type of filter and with the Ozone
generator operating.
Table 7
Sensory appreciation scores obtained during operation of
the Quest Device with either Electrostatic filtration or H.E.P.A.
Day Electrostatic H.E.P.A.
0 1 1
1 3 2
2 6 2
3 6 3
4 6 2
5 6 3
6 6 2
7 6 3
Key to scores
1 Unpleasant
2 Change perceived but unpleasant
3 Improvement
4 Acceptable but some odour detected
5 Acceptable environment
6 Markedly improved odour free
Conclusions ;
Firstly considering the performance of the device in the Microbiological waste room From the data
given in tables 1-4 it is possible to conclude that the test environment under conditions of no treatment
did exhibit elevated levels of airborne microbial contamination. In the same tables it is observed that
irrespective of the filter type employed with device, measurable reduction of airborne levels of Gram
negative bacteria was achieved.
On a continuos use basis with active replacement of micro-organisms into the environment, operation
with Electrostatic filtration gave an average of 84 % reduction of gram negative bacteria (Ox +ve and
-ve). This amounts to a continuos overall reduction of between 1 and 2 log cycles. By comparison the
unit gave only 28 % when operated with H.E.P.A. filtration.
In theory H.E.P.A. filtration should provide greater efficiency with respect to microbial removal but
under the trial conditions we calculated that with this form of filtration in place the device was capable
of only 2.7 room changes per hour. It is apparent this was an insufficient flow rate to achieve high
levels of reduction in an environment to which micro-organisms are constantly being added.
In the case of operation with Electrostatic filtration we obtained 7.1 room changes per hour , a factor
which produced a much higher degree of impingement on the levels of airborne gram negative
bacteria.
It is interesting to note that both forms of filtration gave very high kill efficiencies during the single
pass trials with Aspergillus niger. In this case H.E.P.A. in combination with Ozonation gave 99.986
reduction of challenge which is close to theoretical performance. On the other hand Electrostatic
filtration in combination with ozonation gave 98.012 reduction of challenge.
______________________________________________________________________
Over all our data favours the combination of Ozonation with electrostatic filtration. This
configuration affords high flow rate with very high levels of kill in an environment where
recontamination of sanitised air is a continuos phenomena. By comparison with other commercial
units the kill rate in the waste room environment may be considered very significant. Additionally our
sensory trials indicate this conformation improves the human perception of the space in which the
device is operating.
Laboratories details and points of contact ;
CEO MR R D O.Connor
Tech Manager Mrs W Ingham
Microsearch Laboratories Ltd
Units 3-7 Scotts Trading Complex
Mytholmroyd
Halifax
West Yorkshire
England
HX7 5LH
Phone + 44 (0) 1422 885087 Fax +44 (0) 1422 883721
Email [email protected]
Interim report of an ongoing study into the efficiency of the
Airmanager (P8) Atmospheric Processing device in the
removal of Tobacco Smoke Analytes in a Public House Pool
room
Introduction
A study has been commissioned to determine the effectiveness of the Airmanager P8
atmospheric processing device in the removal of tobacco smoke analytes in a public
house pool room. This report details our preliminary findings in a single test
environment.
The Airmanager P8 atmospheric processing device produces improvement in the
quality of air by the oxidation of organic compounds and harmful micro-organisms
in the presence of high contained levels of Ozone. In this case Ozone is produced by
closed coupled field technology resulting in a unique highly oxidative environment
which produces no release of 03 to the environment. Additionally the output from the
Airmanager device is further purified after passing through a state of the art
electrostet filter assembly.
In this report we have examined the efficiency of the device in reduction of eight
types of tobacco related toxic substances in the test environment. These substances
primarily occur in the atmosphere due to combustion of tobacco and the associated
exhalation of smoke from combusted tobacco. A list of the analytes determined is
given in table 2 below.
The test environment consisted of a Public house pool room with a volume of 84 m3
into which an Airmanager P8 unit was installed. During operation the Airmanager
device, per specification, was predicted to change and process the environment within
this room at a rate 9 times per hour.
Common practice prior to the trial was to evacuated the atmosphere by forced and
passive ventilation. These systems of air purification were considered unsatisfactory
by the Landlord especially during the winter, due to the requirement to compensate
for massive heat loss.
Trial outline :
After installation of the Airmanager P8 device, air sampling was conducted in the
pool room for seven days, between the hours of 8.00 pm and 9.00 pm at a rate of 5
m3
per hour, without the Airmanager device in operation. This provided background
control data for all analytes. A further set of control data was obtained at 5.0 pm
which represents a point after normal ventilation and when the room is not used for
pool or smoking. The data relating to this point may be considered base level for all
analytes.
During the subsequent seven days the sampling procedure was repeated with the
Airmanager device in operation, with the goal of determining the efficiency of
atmospheric clean up.
During the trial an estimate was made of the daily cigarette consumption during the
sampling interval.
Sampling was conducted by the use of a vacuum device with collection of sampled
atmosphere in either Phosphate Buffer or an Acetonitrile : Methanol phase.
Analytes were determined quantitatively employing the following analytical
techniques : Gas Liquid Chromatography, HPLC diode array and Differential pulse
polarography.
Results ;
Table 1 below describes the pattern of cigarette consumption recorded for the test
environment during the sampling periods.
Table 2 describes the mean levels of analytes recorded during the control period and
during the period of sampling when the Airmanager P8 device was activated. This
table also describes the contribution to air quality attributable to the Airmanager P8
device in terms of percentage reduction of airborne toxic substances. Comparative
levels of each analyte for the both sampling periods are provided in graphical form.
Table 1
Mean cigarette consumption in a Public House pool room between
Between 8.00 PM and 9.00 PM
Day Cigarette consumption per
hour
Monday 5
Tuesday 11
Wednesday 7
Thursday 9
Friday 19
Saturday 23
Sunday 16
Mean 13
Table 2
Mean level of Tobacco smoke analytes in the atmosphere in a Public house pool room
for a seven day period with and without the Airmanager unit in operation
ANALTYE UNIT Mean Environmental level 5 PM No
treatment
Mean Environmental level 9 PM No
treatment
Mean Environmental level 9 PM with air treatment
Mean Environmental
reduction of Analyte due to
treatment
Device off Device off Device on
Carbon monoxide Mg/m3
0.82 7.1 0.4 94.4
3-Ethenylpyridine Mg/m3
0.17 37.6 0.4 98.9
Formaldehyde Mg/m3
0.33 84.2 0.2 99.8
Acetaldehyde Mg/m3
0.01 196.3 0.4 99.8
Ammonia Mg/m3
0.01 103.5 0.8 99.2
Nicotine Mg/m3
0.96 61.4 1.06 98.3
Total Phenolics Mg/m3
0.11 12.7 0.2 98.4
Total cresols mg/m3
0.06 3.8 0.08 97.9
0
50
100
150
200
mg
/m3
Atm
os
ph
ere
Mean levels of atmospheric tobacco smoke residue with
and with out the Airmanger unit in operation
Analyte Levels No
treatment
7.1 37.6 84.2 196.3 103.5 61.4 12.7 3.8
Analyte Levels with
Treatment
0.4 0.4 0.2 0.4 0.8 1.06 0.2 0.08
1 2 3 4 5 6 7 8
Conclusions & Discussion ;
There should be no doubt that the test environment represented a significant
challenge in terms of the levels of airborne tobacco related contaminants. Peak
usage of cigarettes was recorded at 23 per hour ( mean 13 ) in the test environment.
In addition both the Landlord and several of his staff anecdotally considered the
environment to be excessively ‘smoky’.
Our test data demonstrates that activation of Airmanager device produced highly
significant reduction in all levels of tobacco smoke analytes with overall analyte
clearance rates over the range 97 .9 to 99.8 %. The level of reduction is such that the
residue levels during device operation are not significantly different from background
level during periods when the room was in disuse. Considering our findings and that
there was virtually constant replacement of the analytes to the atmosphere, it is our
opinion that Airmanager P8 device has performed in a highly efficient manner in the
removal of toxic tobacco smoke contaminants
R.D.O’Connor B.Sc. Ci.Biol., M.R.I.P.H.H.
Trevor Smith B.Sc.
FINAL REPORT
Assessment of ozone leakage from an
air filtration unit
FOR
David Hallam
Quest International (UK) Limited
Unit 8F, Kayley Industrial Estate
Ashton-under-Lyne
Lancashire
OL7 0AU
Written by
Dr. Alexis J. Holden BSc PhD CChem
MRSC
Senior Lecturer, Analytical Science
University of Central Lancashire
Faculty of Science
Preston
PR12HE
Tel No. 01772 893521
Fax No. 01772 892903
Email: [email protected]
2 June 2003
1. Instructions from customer
The company, Quest International Limited, supplied to UCLAN, 1 AirManager
Air Filtration Unit (AM4). The company required tests to be undertaken which
determined if ozone was leaking from the air filtration system when it was
operational.
2. Tests performed
The leakage of ozone was measured when the air filtration system was operated in
4 different modes: (i) filter in and korona on; (ii) filter out and korona on; (iii)
filters in and korona off, and (iv) filters out and korona off.
The ozone levels were measured at 0, 0.5 and 1.0 m from the emitting face of the
unit. The distance was measured using a metre rule and was checked at intervals
during the experiment by the operator. The experiment was performed by 2
operators, Dr. Alexis Holden and Mr. James Donnelly (Technician).
The experiment was performed on 19 June 2002 in a laboratory that was at a
temperature of 22 °C.
The ozone measurement was performed using Gastec detection tubes (No. 18L).
The 18L range provides a rapid, fully quantitative analysis of the concentration of
ozone in air with an accuracy of ± 25%. The manufacturer states that the
minimum detectable concentration as 0.01 ppm. The Gastec tubes were purchased
specifically for this work and were marked valid until May 2005. A Gastec multi-
stroke gas sampling pump was used in conjunction with the tubes.
The principle of the gas tube operation is described by equation 1 below.
2O3 + C16H10N2O2 → 2C8H5NO2 + 2O2 Eqn (1).
The ozone in air, once sucked up through the tube, bleaches the indigo
(C16H10N2O2, blue) to form isatin (C8H5NO2), which is white in colour.
For each position, i.e. 0, 0.5 and 1.0 m from the emitting surface, (at an
approximate angle of 90º) and each operational mode, a tube was placed in the
pump and held in position manually. The system was left to stabilize for 5 minutes
and then 10 pumps (equivalent to 1000 cm3
volume) were drawn on the hand
pump. Each pump lasted an average of 30 seconds. The measurement for each
combination of position and operational mode was repeated five times.
3. Results
The individual results for each tube are shown in Table 1.
Table 1. Individual raw results (ppm) for the Gastec tubes.
Running
mode
Orde
r1
Replicate results
(ppm)
Mean Actual
value2
Operator
1 2 3 4 5 (ppm) (ppm)
Filter in;
Korona on
0 m 4th
0.1 0.1 0.1 0.1 0.1 0.1 0.05 J. Donnelly
0.5 m 5th
0 0 0 0 0 0 0 J. Donnelly
1.0 m 6th
0 0 0 0 0 0 0 J. Donnelly
Filter out;
Korona on
0 m 3rd
<0.05 <0.05 <0.05 <0.05 <0.05 <0.05 <0.025 A. Holden
0.5 m 7th
0 0 0 0 0 0 0 J. Donnelly
1.0 m 8th
0 0 0 0 0 0 0 J. Donnelly
Filter in;
Korona off
0 m 1st 0 0 0 0 0 0 0 J. Donnelly
0.5 m 11th 0 0 0 0 0 0 0 J. Donnelly
1.0 m 12th 0 0 0 0 0 0 0 J. Donnelly
Filter out;
Korona off
0 m 2nd
0 0 0 0 0 0 0 J. Donnelly
0.5 m 9th
0 0 0 0 0 0 0 J. Donnelly
1.0 m 10th 0 0 0 0 0 0 0 J. Donnelly
1this shows the order in which the replicates where run.
2As 10 pumps were used, the values read from the tubes were halved as per the
manufacturers instructions.
4. Discussion and conclusions
The readings were very small such that the highest readings only coloured the first
graduation on the Gastec tube. The highest reading was recorded when the tube
was placed at the emitting surface and the filter was in and the korona was on. The
next highest reading was recorded with the korona on, but the filter out. All other
positions and operational combinations produced no change of colour on the
Gastec tube indicating the levels of ozone, if present, were less than 0.01 ppm.
The average gap in the Gastec tube through which the air is drawn was 1 mm. The
analysis system used is known as active sampling. Five replicate tubes were used
for each combination to help account for the potential variability in the positioning
of the Gastec tube within the flow of air exiting from the air filtration system.
5. Recommendations
The experiment undertaken used active sampling to measure the ozone levels
emitting from the surface of the AirManager Air Filtration Unit. There may be
merit in repeating the experiment with the operational modes of (i) korona on and
filter in and (ii) korona on and out, using passive sampling. Passive sampling does
not use a pump and allows the air to naturally interact with the contents of the
Gastec tube. This would involve placing a number of passive samplers around the
vicinity of the Unit whilst it was in operation. The running time used should be
typical of situations where the Unit would be functioning. This experiment may
provide a better representation of the ozone levels emitted during a normal
working phase of the AirManager Air Filtration Unit.
………………………………………….
Dr. Alexis J. Holden
………………………………………….
Date
FINAL REPORT
Assessment of ozone leakage from an
air filtration unit: passive sampling
FOR
David Hallam
Quest International (UK) Limited
Unit 8F
Kayley Industrial Estate
Richmond Street
Ashton-u-Lyne
Lancashire
OL7 0AU
Written by
Dr. Alexis J. Holden BSc PhD
CChem MRSC
Senior Lecturer, Analytical Science
University of Central Lancashire
Faculty of Science
Preston
PR1 2HE
Tel. No: 01772 893521
Fax No: 01772 892903
Email: [email protected]
13th
January 2004
Contents Page
1 Instructions from customer 2
2 Test performed 2
3 Results 3
4 Discussion and conclusions 4
Appendix 1: Principle of operation of the passive sampling cards 5 Appendix 2: Details of the sensitivities of each passive sampling card
used. 6
Appendix 3a: Indication of the different positions of the passive sampling
cards. 7
Appendix 3b: Indicative plan of the different ceiling positions of the
passive sampling cards. 8
Appendix 3c: Indicative plan of the different positions of the passive
sampling cards. 9
1. Instructions from customer
The company, Quest International Limited, supplied to UCLan, 1 AirManagerAir
Filtration Unit (AM4). The company required tests to be undertaken which
determined if a significant concentration of ozone was leaking from the air
filtration system when it was in operation over an 8 hour period. The analysis was
to be passively sampled.
2. Tests performed
A room of approximately 36.75m3 (3.5m (h) x 3.5m x 3.0m) was selected. It had a
small window on one wall and a normal sized door. One wall (containing the
window) was an external wall the other 3 walls were not exposed to the outside.
The room was dark with fluorescent lighting and received minimal natural light.
To measure the ozone levels passively a number of different sampling cards were
used:
(i) ChromAir ozone cards;
(ii) ChromAir nitrogen cards;
(iii) SafeAir ozone cards;
(iv) SafeAir nitrogen dioxide cards.
The cards were purchased from AFC International Inc. (USA). Nitrogen dioxide is
a potential positive interferent beyond 0.3 ppm with both of the ozone sampling
cards. The experiment monitored this substance alongside the ozone levels. The
principle of operation of the cards is given in Appendix 1.
The experiment was performed on the 4th and 5th
November 2003. The
temperature of the room at an average of 19 ºC.
A sample of each card was placed at random positions within the room. Cards
were placed on the floor, walls and suspended from the ceiling. An indication of
the positions is given in Appendix 2.
Two experiments were conducted, the first, the control experiment where the
cards were monitored for 8 hours whilst the air filtration system was switched off.
In the second experiment the air filtration system was switched on with the corona
on and filter in. The sampler readings were recorded every 15 minutes for the first
hour and then read after a further 7 hours for both experiments.
Originally the experiment was designed to record any change in ozone level
detected by the card every hour. This was eventually rejected as it would have
meant disturbing the air flow (by opening and closing the door) when the recorder
entered the room each hour.
3. Results
An example of a SafeAir ozone qualitative card for a control analysis and a result
from the ‘system on’ experiment is shown in Figure 1.
Control experiment:
(i) All ChromAir ozone cards gave a reading of 0.08 ppm/ hr*
- therefore for 8 hour sampling, 0.08/8 = 0.01
ppm
(ii) No change observed with the SafeAir ozone cards
(iii) No change observed in any of the nitrogen dioxide cards.
Figure 1. Examples from the ozone detection for the (a) blank and (b) ‘system on’
experiments.
* This is the lowest recordable concentration available with the ChromAir ozone
cards; this is the background reading of the cards.
‘System on’ experiment:
(i) All ChromAir ozone cards gave a reading of 0.40 ppm/ 8hr
- therefore for 8 hour sampling, 0.40/8 = 0.05
ppm
(ii) All SafeAir ozone cards showed a qualitative change indicating the
presence of ozone.
(iii) No change observed in any of the nitrogen dioxide cards.
Overall levels of ozone = [‘System on’ experiment] – [Control experiment]
= [0.05 ppm] – [0.01 ppm]
= 0.04 ppm
Therefore, the time weighted average for the concentration emitted over an 8
hour period is 0.04 ppm (40 ppb).
4. Discussion and conclusions
The HSE occupational exposure limits (OEL) for ozone over an 8 hour period is
0.2 ppm (200 ppb). This experiment produced a recordable value of 0.04 ppm (40
ppb) over an 8 hour period. HSE also states a limit of exposure for 15 minutes of
0.4 ppm (400 ppb), the experiment conducted for this reported produced no
observable changes in the passive samplers within the first hour of the experiment.
It should be noted that the actual observation of any change in colour on the
sampling cards is quite subjective. The assessment of the recorded value is added
by a colour comparator chart but this was extremely rudimentary in accuracy. All
readings were observed by 1 person (Ms June Gardner and the results of a
proportion of the samples confirmed by Dr. Alexis Holden).
……………………………………………………
Dr. Alexis J. Holden
……………………………………………………
Date
Appendix 1: Principle of operation of the passive sampling cards
ChromAir Ozone passive samplers (Part #380010-10)
The ChromAir passive monitor is a patented direct read autogenic exposimeter. The
device is constructed from six cells attached on one side to a flat indicator layer and
on the other side to a series of different diffusive resistances. Ozone gas diffuses to
the cells through the different diffusive resistances and reacts with the indicator layer,
producing colour change from blue to light blue and finally to white upon high
exposure. The colour produced on the indicator layer is a direct measure of the
exposure dose. Visual colour comparison is achieved by observing the formation of
the light blue threshold colour on the individual cell and reading the corresponding
exposure dose.
SafeAir ozone passive samplers (Part#382004)
The SafeAir ozone badge is a monitoring system designed to indicate the presence of
ozone at concentrations below the permissible exposure limit. The SafeAir ozone
badge detects the presence of ozone by forming a colour change in the shape of an
exclamation mark inside the triangle. This indication is produced by a colour forming
reaction which occurs when ozone reacts with a flat indicator layer. The colour
change is from blue to a lighter blue.
ChromAir Nitrogen Dioxide passive samplers (Part #380006)
The ChromAir passive monitor is a patented direct read autogenic exposimeter. The
device is constructed from six cells attached on one side to a flat indicator layer and
on the other side to a series of different diffusive resistances. Nitrogen dioxide gas
diffuses to the cells through the different diffusive resistances and reacts with the
indicator layer, producing colour change from yellow to beige to brown. The colour
produced on the indicator layer is a direct measure of the exposure dose. Visual
colour comparison is achieved by observing the formation of the beige threshold
colour on the individual cell and reading the corresponding exposure dose.
SafeAir nitrogen dioxide passive samplers (Part#382013)
The SafeAir nitrogen dioxide badge is a monitoring system designed to indicate the
presence of nitrogen dioxide at concentrations below the permissible exposure limit.
The SafeAir nitrogen dioxide badge detects the presence of nitrogen dioxide by
forming a colour change in the shape of an exclamation mark inside the triangle. This
indication is produced by a colour forming reaction which occurs when nitrogen
dioxide reacts with a flat indicator layer. The colour change is from white to pale
yellow.
Appendix 2: Details of the sensitivities of each passive sampling card used.
ChromAir system card
Ozone cards Nitrogen dioxide
Physical specifications
Dimensions 10.5 cm x 5.5 cm x 0.25
cm
10.5 cm x 5.5 cm x
0.25 cm
Weight 11 g 11 g
Refrigerated shelf life 1 year 1 year
Colour change Blue to white Yellow to beige to
brown
Sampling parameters
Exposure range for:
Badge 0.08 – 1.6 ppm.hr 0.5 - > 13 ppm.hr
Badge with colour comparator 0.08 – 2.6 ppm.hr 0.35 – 40 ppm.hr
Max. recommended sampling
time
10 hours 2 days
Min. recommended sampling
time
5 minutes 15 minutes
Temperature range 16 – 30 ºC 10 – 40 ºC
SafeAir system card
Ozone cards Nitrogen dioxide
Physical specifications
Dimensions 7.4 cm x 4.1 cm x 0.1
cm
7.4 cm x 4.1 cm x 0.1
cm
Weight 1.5 g 1.5 g
Refrigerated shelf life 1 year 1 year
Colour change Blue to white Yellow to brown
Sampling parameters
Exposure level 0.05 ppm.hr 1.0 ppm.hr
Minimum detectable limit (8
hours)
0.006 ppm 0.125 ppm
Max. recommended sampling
time
2 days 10 hours
Min. recommended sampling
time
15 minutes 15 minutes
Temperature range 16 – 33 ºC 15 – 40 ºC
Appendix 3a: Indication of the different positions of the passive sampling cards.
Label Distance from filtration system
(via floor)
Distance from filtration system
(via air)
F1 4 ft 2 in. -
F2 2 ft 11 in. -
F3 2 ft 6 in. -
F4 1 ft 10 in. -
F5 2 ft 4 in. -
F6 3 ft 3 in. -
F7 4 ft 0 in. -
F8 3 ft 2 in. -
F9 5 ft 0 in. -
F10 4 ft 4 in. -
F11 2 ft 1 in. -
F12 4 ft 9 in. -
F13 3 ft 7 in. -
F14 2 ft 0 in. -
F15 2 ft 6 in. -
F16 3 ft 7 in. -
F17 4 ft 4 in. -
F18 2 ft 4 in. -
F19 2 ft 7 in. -
W1 6 ft 2 in. 8 ft 3in.
W2 3 ft 7 in. 1 ft 2 in
W3 5 ft 10 in. 7 ft 11 in.
W4 4 ft 4 in. 7 ft 9 in.
W5 6 ft 7 in. 9 ft 0 in.
W6 5 ft 10 in. 8 ft 4 in.
W7 3 ft 9 in. 6 ft 7 in.
W8 5 ft 0 in. 6 ft 9 in.
W9 1 ft 10 in. 6 ft 4 in.
W10 7 ft 9 in. 10 ft 0 in.
C1 4 ft 5 in. 6 ft 4 in.
C2 6 ft 5 in. 6 ft 4 in.
C3 4 ft 0in. 5 ft 9 in.
C5 5 ft 0 in. 4 ft 8 in.
C6 6 ft 6 in. 6 ft 7in.
C7 6 ft 8in. 5 ft 0in.
C8 7 ft 4in. 6 ft 9 in.
C9 7 ft 7 in. 7 ft 10in.
C10 7 ft 4 in. 8 ft 2 in.
Appendix 3b: Indicative plan of the different ceiling positions of the passive
sampling cards.
DOORDOOR
WINDOW
AIR
FILTRATION
SYSTEM
C9 C3
C7
C6
C2
C1
C5
C10
Appendix 3c: Indicative plan of the different positions of the passive sampling
cards.
WINDOW
AIR
FILTRATION
SYSTEM
W5
W4
F12
W9
F13
F10
F9
W3
F11
F14
F15
F16 W6
F17
F18
F19 W7
F8
W2
F7
F6
W1 W8
W10
F1
F2
F4
F3
F5
DOOR
Page 38 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
CCFRA Technology Limited
Chipping Campden
Gloucestershire
GL55 6LD, UK
Tel: +44 (0)1386 842000
Fax: +44 (0)1386 842100
www.campden.co.uk
CAMPDEN & CHORLEYWOOD FOOD RESEARCH ASSOCIATION
CHIPPING CAMPDEN, GLOS. GL55 6LD.
INVESTIGATION OF EFFECTIVENESS OF AIR MANAGER UNIT
CONFIDENTIAL TO: David Hallam
Quest International Air Manager Division
Unit 8F
Kayley Industrial Estate
Richmond Street
Ashton-under-Lyne
Greater Manchester
OL7 0AU
REPORT NO. FH83479/1
AUTHORS: Dr. K. L. Brown,
Chloe Stripp
May 2005
Report Approved By : Report Checked By :
Name : :
Information emanating from this company is given after the exercise of all reasonable care and skill in its compilation, preparation
and issue, but is provided without liability in its application and use. Campden & Chorleywood Food Research Association.
Company limited by guarantee. Registered No. 510618 England
CCFRA Technology Limited. Registered No. 3836922 England
CCFRA Group Services. Registered No. 3841905 England
Registered Office: Chipping Campden, Glos. GL55 6LD
Information emanating from the CCFRA Group is given after the exercise
of all reasonable care and skill in its compilation, preparation and issue.
but is provided without liability in its application and use.
ISO 9001: 2000
Certified
Page 39 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Introduction and background information
CCFRA were approached by David Hallam of Quest International to evaluate the
effectiveness of an Air Manager unit against airborne challenge tests in a controlled
environmental room. Initial tests were done using BVM190 units in a room of
volume 39.6m3. Later experiments were done using an AM4 unit. The test organisms
were chosen either because they were typical environmental contaminants or because
they were standard test strains.
Summary of experiments
Units tested: BVM190. This unit can treat 165m3/h.
AM4. This unit can treat 190m3/h
Room volume: 39.6m3 (3.5m wide, 4.1m long and 3m high minus airlock 1m wide,
1.14m long and 3 m high). Four fans (12x12cm) positioned on the floor of the room
were used to ensure the aerosols were mixed adequately in the air. Each fan is rated
at a nominal 30l/s throughput.
Parameters monitored: Air temperature and relative humidity.
Test organisms:
Pseudomonas aeruginosa (NCIMB 10421) (Test strain for disinfectant testing)
Staphylococcus aureus (NCIMB 9518) (Test strain for disinfectant testing)
Penicillium chrysogenum (CABI 024314)
Bacillus subtilis var globigii B17 (culture from Nottingham University) (Used as a
test organism for packaging and air sterilisation)
Neurospora (Chrysonilia) sitophila (CABI 021944) (At request of client)
Preparation of spore suspensions and cultures:
Bacillus subtilis var. globigii spores
A stock culture of Bacillus subtilis var. globigii was inoculated into Nutrient Broth
(Oxoid CM1) (NB) and incubated overnight at 30°C in a shaking incubator. Nutrient
Agar (Oxoid CM3) (NA) plates were then seeded with the culture and incubated agar
side down for 6 days at 30°C. The cultures were checked microscopically to ensure
that spores had been produced. The spores were removed using a metal scraper into
sterile distilled water. The initial count was determined to be 5.7 x 109/ml. The spore
crop was diluted to give a concentration of approx. 107/ml.
Page 40 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Penicillium chrysogenum and Neurospora (Chrysonilia) sitophila spores
Penicillium chrysogenum and Neurospora (Chrysonilia) sitophila spores from a
previous spore crop were spread onto the surface of pre-poured 150mm diameter Petri
dishes containing Malt Extract Agar (Oxoid CM59). These were incubated for 2 days
at room temperature until the fungal growth had covered the surface of the plates.
The temperature was then increased to 25oC for 3 days to induce spore production.
The spores were scraped from the surface of the agar in a suspension of 0.1% Tween
80. The resulting suspension was filtered through sterile glass wool, spun down twice
at 2000rpm for 20 min and resuspended in Maximum Recovery Diluent (MRD, Oxoid
CM733). The concentration of spores was estimated using a Haemocytometer.
Recovery medium for experiments using mould was Malt Extract Agar (Oxoid
CM59).
Pseudomonas aeruginosa and Staphylococcus aureus
Ps. aeruginosa and S. aureus were grown overnight ready for trials the next day on
TSA (Oxoid CM131) at 30 and 37oC respectively. Growth was removed and
resuspended in Maximum Recovery Diluent (MRD, Oxoid CM733). The resulting
suspension contained approximately 109 cells/ml (determined using a
spectrophotometer). The concentration was then adjusted with MRD to give
approximately 107 cells/ml. The recovery medium for S. aureus: Baird Parker Agar
(Oxoid CM275) plus supplement (Oxoid SR54). The recovery medium for Ps.
aeruginosa: Pseudomonas Agar base (Oxoid CM559).
Spraying procedure for test organisms:
A Collison nebuliser was used to create an aerosol at approximately 1m above the
floor at one side of the room. The nebuliser releases approximately 0.2ml/min
aerosol. Aerosol times were adjusted in line with the concentration of the bacteria in
the suspension to give a starting level of approximately 105 – 10
6/m
3 in the room air.
This represents very heavily contaminated air and the aim was also to have countable
numbers in the air samplers.
Air sampling:
The 2 sampling ports in one wall of the room were connected to 2 Oxoid MAQS air
samplers. A Mattson-Garvin air sampler that sampled air over one hour continuously
was connected to the room and used to give a picture of the airborne counts over the
time frame of the experiment.
Theory:
The room is 39.6m3 and the rate of throughput in the BVM190 is 165m
3/h.
Let V1 = throughput of unit
Let VT = total room volume
Let Cin = Concentration of bacteria/m3 entering unit at any point in time
Let Cout = Concentration of bacteria/m3 leaving the unit
Let CT = Concentration at a point in time
Page 41 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Reduction in concentration of bacteria, p = (Cin-Cout)/Cin
Also dCt/dt = -(V1/VT).(p.CT)
For first order reaction, Ct = C0e-kt
Where C0 = starting concentration and k = -(V1/VT).(p)
Plotting log CT against time (t) should give a straight line with slope –(V1/VT).p/2.303
Assuming 100% efficiency of the unit in removing airborne bacteria, p becomes equal
to 1. In approximately 15 minutes the unit has processed a volume of air equivalent to
the room volume so V1/VT after 15 minutes is approximately 1. Dividing 15 by 2.303
gives approximately 7. Therefore we would expect to see one log reduction in
airborne count in around 7 minutes (assuming 100% efficiency). The AM4 unit
should give 1 log reduction in approximately 5.4 min using the same calculations.
The control run with the unit switched off was estimated to take at least 4 hours based
on previous work. A ranging trial using the spores of B. globigii was done first to
determine the sampling times for subsequent experiments.
Experimental protocol:
Suspensions of B. globigii or mould spores or suspensions of overnight bacterial
cultures were produced as appropriate. The Collison nebuliser was filled with the
suspension (one organism tested at a time).
The room circulation fans and the Quest unit were switched on.
The nebuliser was operated for a fixed time interval to create the aerosol.
A continuous sample was taken using a Mattson Garvin air sampler and discrete
samples were taken every 15 min using two Oxoid MAQS samplers (100 litre or 200
litre sample size).
At the end of sampling, any remaining airborne micro-organisms were flushed out
using the room extract system and the floor was mopped with disinfectant.
Each trial was repeated with the Air Manager unit switched off.
The airborne counts were compared with and without the Air Manager unit operating
to estimate the performance of the unit.
Pag
e 4
2 o
f 1
6
Hygie
ne:
\20
05
\co
ntr
act\
Air
manag
er/F
H8
34
79
/1
Ta
ble
1 O
per
ati
ng
pa
ram
eter
s d
uri
ng
tri
als
Ex
pt.
No.
Air
Man
ager
Unit
Org
anis
m
Num
ber
neb
uli
sed
Unit
on/o
ff
Dura
tion
(h)
Tem
p (
oC
) R
H %
N
ebuli
ser
tim
e (m
in)
Neb
uli
ser
vol
(ml)
Neb
uli
ser
pre
ssure
(bar
)
1
BV
M190
Ser
.3682
B.s
ub
tili
s 7.5
x10
7
On
1
20.0
35
5
2.5
2
2
��
7.5
x10
7
Off
1
20.0
35
5
1.7
2
3
��
5.1
x10
7
Off
1
18.5
35
5
1.3
6
1.5
4
��
4.0
x10
7
On
2
16.5
52
4
1.3
6
1.5
5
��
4.0
x10
7
Off
2
18.4
62
4
1.3
6
1.5
6
BV
M190
Ser
.3698
�
4.0
x10
7
On
2
21.3
63
4
1.3
6
1.5
7
��
4.0
x10
7
Off
2
21.5
49
4
1.3
6
1.5
8
��
4.0
x10
7
On
4
17.5
56
4
1.3
6
1.5
9
��
4.0
x10
7
Off
4
18.2
55
4
1.3
6
1.5
10
AM
4
Ser
.3656
P.c
hry
sog
enu
m
6.9
x10
7
On
4
18.5
53
4
1.3
6
1.5
11
��
6.9
x10
7
Off
4
17.9
45
4
1.3
6
1.5
12
�B
.su
bti
lis
4.0
x10
7
On
4
17.2
55
4
1.3
6
1.5
13
��
4.0
x10
7
Off
4
17.2
49
4
1.3
6
1.5
14
�S
.au
reu
s 3.4
x10
7
On
4
18.2
56
4
1.3
6
1.5
15
��
5.3
x10
7
Off
4
17.6
58
4
1.3
6
1.5
16
�P
.aeru
gin
osa
2.6
x10
7
On
4
18.1
62
4
1.3
6
1.5
17
��
2.9
x10
7
Off
4
18.4
64
4
1.3
6
1.5
18
�N
.sit
op
hil
a
1.4
x10
7
On
4
18.1
76
10
3.4
1.5
19
��
1.4
x10
7
Off
4
18.6
55
10
3.4
1.5
The
MA
QS
sam
ple
rs w
ere
set
to s
ample
100 l
itre
s fo
r ex
per
imen
ts 1
-3, an
d w
ere
set
to s
ample
200 l
itre
s fo
r th
e re
mai
nder
of
the
work
.
Page 43 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Figure 1 Layout of test room for experiments 1 and 2
Figure 2 Layout of test room for experiments 3 to 19
Key:
Neb = Collison nebuliser used to produce aerosol positioned approximately 1m above the floor.
Fans were small 15W fans (12x12cm) moving air at approximately 3m/s and positioned on the floor.
MAQS samplers sampled through 2 ports in the wall.
Arrows indicate direction of air flow. In layout 2, the three small fans were pointing upwards to
circulate the air.
The Air Manager was on a lab stool.
Air
Manager Fan Fan
Fan Fan
Mattson
sampler Neb
MAQS
samplers
Fan
up
Fan
up
Fan
up
Mattson
sampler
MAQS
samplers
Air
Manager
Neb
Fan
Page 44 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Results
Experiments 1 – 5 were done using the first BVM190 unit that can treat up to
165m3/h. Results from these first trials using spores of Bacillus subtilis var globigii
showed little difference with unit switched on or off (Figures 3 and 4). Because there
was no visible means of determining whether this unit was working properly or not, a
new unit that had been tested for correct functioning was supplied. This second
BVM190 unit was used for experiments 6 to 9 and was only marginally better than the
first (Figures 5 and 6). It was then decided to use the larger capacity AM4 unit that
can treat 190m3/h. This was used for experiments 10 to 19
Results with the AM4 unit for P. chrysogenum spores are shown in Figures 7 and 8.
After 4 hours there was approximately a 3-log difference in airborne count between
experiments with the unit on versus off. There was approximately 1 log difference
after approximately 112 min.
Results for B. subtilis spores showed approximately 1 log difference after 4 hours
with the AM4 unit compared to approximately ¾ of a log reduction with the BVM190
unit. (Figures 9 and 6 respectively).
Staphylococcus aureus results (Figure 10) showed approximately 1 log difference
after 65 minutes. This is also shown in the photograph (Figure 11).
Ps aeruginosa gave approximately 1¾-log difference after approximately 80 min.
(Figure 12).
Neurospora (Chrysonilia) sitophila colonies were very difficult to count because the
organism grows rapidly, covering the Petri dish and even growing out of the dish very
quickly. (Figure 16). Results were therefore expressed as presence or absence of
growth (Tables 3 and 4). The airborne spores had been removed by the AM4 unit
within 90 min as shown by the MAQS samples and between 1 and 2 hours as shown
by the Mattson results (Figures 13 and 14). By comparison, with the AM4 switched
off, airborne spores were still detectable between 3 and 4 hours. Figure 15 shows
MAQS sample plates after 3h and 3h 15min with the unit switched off. This was the
approximate time when the spores had settled naturally from the air. (One MAQS
plate was positive after 3h 30min).
Page 45 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Figure 3 Results using BVM190 (Ser No 3682)
B. subtilis spores: Experiment 1 (unit on) and experiment 2 (unit off) MAQS results
0
0.2
0.4
0.6
0.8
1
1.2
0 10 20 30 40 50 60 70
Time (min)
Lo
g N
/ lo
g N
o p
er
m3
mean on
mean off
Figure 4 Mattson results for Bacillus subtilis from 1 – 2 hours during
experiments 1 and 2. Plate on the left shows results with Airmanager BVM190
off and on the right with Airmanager BVM190 on.
Page 46 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Figure 5 Results using BVM190 (Ser No 3698)
B. subtilis spores: Experiment 8 (unit on) and experiment 9 (unit off) - MAQS results
-0.20
0.00
0.20
0.40
0.60
0.80
1.00
1.20
0 50 100 150 200 250 300
Time (min)
Lo
g N
/ lo
g N
o p
er
m3
mean off
mean on
Figure 6 Results using BVM190 (Ser No 3698)
B. subtilis spores: Experiment 8 (unit on) and experiment 9 (unit off) - Mattson results
0
0.5
1
1.5
2
2.5
3
3.5
4
0 50 100 150 200 250
Time (min)
Lo
g c
ou
nt
ov
er
5 m
in
unit on
unit off
Page 47 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Figure 7 Results using AM4 (Ser No 3656)
P. chrysogenum spores: Experiment 10 (unit on) and experiment 11 (unit off) - Mattson results
-1
-0.5
0
0.5
1
1.5
2
2.5
3
3.5
0 50 100 150 200 250
Time (Min)
Lo
g c
ou
nt
over
5 m
in
unit on
unit off
Figure 8 Mattson results for Penicillium chrysogenum from 1 – 2 hours during
experiments 10 and 11. Plate on the left shows results with Airmanager AM4 off
and on the right with Airmanager AM4 on.
Page 48 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Figure 9 Results using AM4 (Ser No 3656)
B. subtilis spores: Experiment 12 (unit on) and experiment 13 (unit off) - Mattson results
0
0.5
1
1.5
2
2.5
3
3.5
4
0 50 100 150 200 250
Time (min)
Lo
g C
ou
nt
ov
er
5 m
in
Unit On
Unit Off
Figure 10 Results using AM4 (Ser No 3656)
Staph aureus: Experiment 14 (unit on) and experiment 15 (unit off) - Mattson results
0
0.5
1
1.5
2
2.5
3
3.5
4
0 50 100 150 200 250
Time (min)
Lo
g c
ou
nt
over
5 m
in
unit on
unit off
Page 49 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Figure 11 Mattson results for Staphylococcus aureus from 1 – 2 hours during
experiments 14 and 15. Plate on the left shows results with Air Manager AM4 on
and on the right with Airmanager AM4 off.
Figure 12 Results using AM4 (Ser No 3656)
Ps aeruginosa: Experiment 16 (unit on) and experiment 17 (unit off) - Mattson results
-1
-0.5
0
0.5
1
1.5
2
2.5
3
3.5
4
0 50 100 150 200 250
Time (min)
Lo
g c
ou
nt
over
5 m
in
Unit on
Unit off
Page 50 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Table 2 Results from MAQS samplers (200 litre samples) for AM4 unit operating against
airborne Neurospora (Chrysonilia) sitophila spores
Time (min) Unit AM4 on Experiment 18 Unit AM4 off Experiment 19
MAQS 1 MAQS 2 MAQS 1 MAQS 2
5 + + + +
15 + + + +
30 + + + +
45 + + + +
60 + + + +
75 - - + +
90 - + + +
105 - - + +
120 - - + +
135 - - + +
150 - - - +
165 - - + -
180 - - + +
195 - - - -
210 - - - -
225 - - - -
240 - - - -
Key: Since Neurospora colonies spread very rapidly over the plates it was not possible
to count individual colonies so results were expressed as:
(+) = growth over plate, (-) = no growth
Table 3 Results from Mattson sampler for AM4 unit operating against airborne
Neurospora (Chrysonilia) sitophila spores
Time period
(h)
Unit AM4 on. Experiment 18 Unit AM4 off. Experiment 19
0-1 + +
1-2 + +
2-3 - +
3-4 - +
Key: Since Neurospora colonies spread very rapidly over the plates it was not possible
to count individual colonies so results were expressed as:
(+) = growth over plate, (-) = no growth
Page 51 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Figure 13 Mattson results for Neurospora (Chrysonilia) sitophila from 0-1h on
the left and 1-2h on the right with AM4 unit on. Airborne Neurospora has been
removed between 1-2 h. (Experiment 18)
Figure 14 Mattson results for Neurospora (Chrysonilia) sitophila from 0-1h on
the left and 2-3h on the right with AM4 unit on. Airborne Neurospora has been
totally removed by 2h. (Experiment 18)
Page 52 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Figure 15 Results from MAQS plates for Neurospora (Chrysonilia)
sitophila with AM4 unit (experiment 19). Samples taken after 3h on
the left with unit off and after 3h 15m on the right with unit on.
Figure 16 Results from MAQS plates for Neurospora (Chrysonilia) sitophila with
AM4 unit off (experiment 19). Samples taken after 5 minutes showing how this
organism spreads over the whole of the agar surface and beyond the edge of the
plate.
Page 53 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Conclusions
The AM4 unit was effective in removing aerosols of the test organisms: Pseudomonas
aeruginosa (NCIMB 10421), Staphylococcus aureus (NCIMB 9518), Penicillium
chrysogenum (CABI 024314), Bacillus subtilis var globigii B17 (culture from
Nottingham University) and Neurospora (Chrysonilia) sitophila (CABI 021944).
Within the time frame of these experiments (up to 4 hours) log reductions achieved
were typically between 1 and 2 log reductions compared to results with the unit
switched off.
Page 54 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Mountainheath Services Ltd. Environmental Analysis & Consultancy
Analytical Report
Quest International (UK) Ltd Report No: N1487 Addendum
Unit 8F, Kayley Industrial Estate, Date Received: 12 August 2005
Richmond Street Date Tested: 17/08/05-09/09/05
Ashton Under Lyne Date Issued: 09 September 2005
OL7 0AU Page: 1 of 1
For the attention of: David Hallam (Director) By e-mail (original by post)
In order to better assess the efficiency of the filter at removing oil pyrolysis products, results
were calculated for a wider range of organic compounds as well as for total organics removal.
These are presented in the table below.
Laboratory reference 94436 94434 94433 94432
Client referenceConc. µg/m
3
no filter
Conc. µg/m3
3 passes
% removal
Conc. µg/m3
9 passes
% removal
Conc. µg/m3
ambient air
hexanal 140 4.8 99 1.7 99 0.484-methyloctane 160 14 97 9.7 98 7.1
N,N-dimethylacetamide 13000 590 98 400 100 370phenol 730 120 98 45 100 87
n-dodecane 59 1.9 99 0.90 102 2.2dodecamethylcyclohexasiloxane 260 13 100 6.2 100 6.2
total organics 45000 960 99 760 100 760
They indicate a better performance than previously calculated from the N,N-
dimethylacetamide figures. The erroneous result was caused by a computer integrator error as
it was not able to cope with the very high absolute values involved, i.e.: they were out of
range. Results have been recalculated taking this into consideration. The results show a
consistent performance of 97% to 100% removal after three passes of the filter considering
the ambient air concentration as the base line.
The total ion current traces were re-examined in order to investigate the occurrence of
unchanged oil additives. This was achieved by constructing ion chromatograms of these
compounds using relevant mass fragments in their mass spectra. These compounds, i.e.:
tricresyl phosphate, N-phenyl-2-naphthalenamine, 1-naphthalenamine, 2-naphthalenamine
and 4-octyl-N-(4-octylphenyl)benzenamine, could not be detected at a concentration greater
than approximately 0.10µg/m3. This is not conclusive evidence that the compounds were not
present in the rig as they are relatively involatile and not therefore suitable for Tedlar bag
sampling. The results from the filter sampling would be more indicative of the absence of
these compounds as this might be expected to be a more appropriate method for their
recovery. The absence of any response in the traces obtained from the filter samples would
indicate their absence above a concentration of approximately 1.0 µg/m3.
Pamela M Howarth Sally Paynter Alex Waggott
Director Quality Manager Director
Page 55 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Page 56 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Page 57 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Page 58 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Page 59 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Page 60 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Page 61 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1
Page 62 of 16 Hygiene:\2005\contract\Airmanager/FH83479/1