1 Project no. 610886 Project acronym: ClinicIMPPACT Clinical Intervention Modelling, Planning and Proof for Ablation Cancer Treatment Work programme objective: ICT 2011.5.2 Virtual Physiological Human Instrument: STREP Deliverable D6.8 Scientific publication on evaluation of the clinical study results Due date of deliverable: 31/07/2017 Actual submission date: 08/09/2017 Organisation name of lead contractor for this deliverable: MUL Start date of project: 01/02/2014 Duration: 42 months

Upload: others

Post on 19-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

1

Project no. 610886

Project acronym: ClinicIMPPACT

Clinical Intervention Modelling, Planning and Proof for

Ablation Cancer Treatment

Work programme objective: ICT 2011.5.2 Virtual Physiological Human

Instrument: STREP

Deliverable D6.8

Scientific publication on evaluation of the clinical study results

Due date of deliverable: 31/07/2017

Actual submission date: 08/09/2017

Organisation name of lead contractor for this deliverable: MUL

Start date of project: 01/02/2014 Duration: 42 months

Page 2: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

2

Scientific publication below will be submitted to the Journal of Radiology

Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment

(ClinicIMPPACT) Trial – Planning and real-time simulation of hepatic RFA procedures

Michael Moche1, Camila Andrea Hinestrosa Arteta1, Daniel Seider1, Marina Kolesnik2, Jurgen

Fuetterer3, Sjoerd Jenniskens3, Roberto Blanco4, Martin Eibisberger5, Philip Voglreiter6, Horst

Ruppert Portugaller5, Mika Pollari, Ronan Flanagan7, Panchatcharam Mariappan7, Harald Busse1

and Martin Reinhardt1

1 Department of Diagnostic and Interventional Radiology, University Hospital Leipzig,

Germany

2 Fraunhofer Institute for Applied Information Technology FIT, Germany

³ Department of Diagnostic and Interventional Radiology, University Hospital Nijmegen,

Netherlands 4 Medical Imaging Centre of Southwest Finland, Turku University Hospital, Finland

5 University Clinic of Radiology Graz, Graz, Austria

6 Graz University of Technology, Institute of Computer Graphics and Vision, Austria 7 NUMA Engineering Services Ltd., Louth, Ireland

Corresponding author:

Martin Reinhardt, MD, Department of Diagnostic and Interventional Radiology, University Hospital

of Leipzig, Liebigstraße 20, 04103 Leipzig, Germany.

Tel: +49-341-9716987; Fax: +49-341-9717409;

E-mail: [email protected]

Funding information:

This research was funded by the Seventh Framework Program of the European Union (grant number

610886)

Type of manuscript: Original Research. This manuscript was not presented at any RSNA meeting

and has not been submitted to any future meeting.

Keywords: RFA; liver; lesion prediction; segmentation; perfusion CT

Page 3: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

3

Abstract

Purpose:

Patient-specific factors as tumor proximity to vessels or tissue perfusion can affect lesion size during

radiofrequency ablation (RFA) of liver tumors and lead to both, over- and under treatment. We aimed

to clinically validate a comprehensive software tool designed to optimize the planning and simulation

of RFA procedures.

Methods and Materials:

In this prospective clinical trial, we included 47 hepatic lesions from 44 patients with different liver

malignancies, who underwent multi-phase computed tomography (including quantification of liver

perfusion) prior to RFA. The actual RFA-needle position was assessed by intraprocedural CT-scans

and registered to a 3D-model created prior RFA. Image computations and thermal simulation allowed

real-time processing during the RFA procedure. Size, shape and position of the simulated lesions

were compared to the true ablation zones determined by follow-up CT one month after RFA.

Results:

There was a strong correlation (r=0.71, ρ<0.0001) and no significant difference (Wilcoxon p=0.49)

between the simulated- and segmented lesion volumes. Surface- and volume deviation between the

simulated- and segmented lesions were 3.8 mm±2.0 SD (absolute average error, AAE), and

1.1 %±1.2 SD (relative volume deviation). AAE was positively correlated to the duration of the

ablation (r=0.29, p<=0.05) and the duration of the ablation was an independent predictor of AAE

(β=0.027±0.01 minutes, p=0.02), after adjustments for tissue perfusion and presence or absence of

liver cirrhosis.

Conclusion:

Our software application provided good accuracy for predictions of lesions resulting from

radiofrequency ablations in the liver. Implementation into clinical routine may help to improve the

safety and success of radiofrequency ablations in the liver.

Page 4: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

4

Introduction

Local thermal ablation techniques such as radiofrequency ablation (RFA) are considered first

line treatment options for inoperable patients with early-stage hepatocellular cancer (HCC) (1). With

comparable overall survival RFA is also a viable therapeutic alternative for early metastatic liver

cancer in inoperable patients (2), and in the presence of poor liver function (3,4).

The percutaneously applied RFA probe induces cell death via coagulative necrosis. Despite

precise image guided needle placement (e.g. using computed tomography) RFA can be challenging

in terms of sufficiency. This is mainly owed to the complex relationship between applied thermal

energy and variable local tissue perfusion, more specifically due to the presence or absence of

neighboring cooling venous and portal-venous vessels, which divert heat from the ablation zone (heat

sink effect).

Reliable real-time therapy monitoring remains the missing link which would likely reduce

discrepancies between the expected and the real lesion, which can only be sufficiently assessed by

follow-up imaging once post interventional tissue swelling disappeared but lesion shrinkage has not

yet set in, which is usually the case 1 month after RFA. The intraprocedural uncertainty about the

volume and shape of the resulting ablation zone leads to over- (up to 9% of major complications) (5),

or under treatment (with up to 40% local recurrence) (5,6).

Different approaches to simulate and predict the resulting ablation zone have been described

(7–10), however only few groups incorporated important patient specific parameters, i.e. liver

perfusion into their models (10,11). Preclinical studies included extensive experiments on cell

cultures and animals to create a heat dependent cellular death model, which was implemented into

the current simulation algorithm (12,13). The RFA-Guardian software provides a simulator for

interventional radiologists (IRs) to plan, review and optimize RFA procedures.

After generating a 3D-model of the liver, the actual needle position is assessed by

intraprocedural CT-scans, registered to the 3D-model to simulate the thermal lesion. Image

computations and thermal simulation allow real-time processing during the RFA procedure. Size,

Page 5: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

5

shape and position of the simulated lesions are then compared to the true ablation zone as determined

by follow-up CT one month after RFA.

In this trial, we collected pilot clinical data to test how precise, as measured by volume and

surface deviation, and how efficient, compared to a retrospective control group, newly developed

software can provide real-time predictions of resulting lesion sizes during RFA of liver malignancies.

Materials and Methods

Subjects and study design

In this multicenter prospective clinical trial, we invited N=44 (N=14 Turku, Finland; N=3

Graz, Austria; N=12 Nijmegen, Netherlands; and N=15 Leipzig, Germany) patients with pre-

diagnosed primary (hepatocellular carcinoma, HCC) or secondary malignant liver tumors (e.g.

metastases). One patient from Turku died, within the four weeks between RFA treatment and follow

up visit, of reasons unrelated to the treatment and was therefore excluded from the analysis. Four

patients presented with N=2 lesions each, which led to N=47 lesions included in this trial. Diseases

were diagnosed in two independent imaging modalities or alternatively histologically proven.

Decisions on optimal therapies were made, according to AASDL guidelines, by each local tumor

board (consisting of radiologists, oncologists and hepatobiliary surgeons), which evaluated options

such as resection, liver transplantation, adjuvant or neoadjuvant chemotherapy, or image-guided

procedures such as RFA. Resection or other treatment options were not suitable for these patients,

who presented with a maximum of 3 lesions and a maximum diameter of 3 cm per lesion.

Furthermore, sufficient coagulation parameters according to European School of

Interventional Radiology - guidelines were required. Participants were at least 18 years old and

fulfilled standard eligibility criteria for undergoing pre-interventional diagnostic imaging, such as

absence of pregnancy or current nursing, dysfunction of the kidney or thyroid gland, and known

anaphylactic reaction to iodine/contrast agent. Concurrent participation in other interventional trials

as well as splenectomy were exclusion criteria.

Page 6: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

6

Patients gave both verbal and written informed consent and approvals for the trial were

granted by the institutional ethics committees at all four clinical sites and the German Federal Office

for Radiation Protection.

The study design of this clinical trial is described in great detail elsewhere (14). The main

components of the study design are shown in Figure 1 and described in the sections below.

Preliminary results with focus on technical aspects have been published earlier (13).

Figure 1. Study design overview

Figure 1. Flow chart of the ClinicIMPPACT Trial

Page 7: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

7

Imaging

Planning Imaging

Multiphase diagnostic CT imaging of the abdomen was carried out as part of the clinical

routine, comprising the parameters shown in Table 1. Patients were fasted at least 4 hours prior to

the examination. No oral contrast agent was used. First a non-enhanced scan was performed.

Intravenous contrast agent was injected at a flow rate of 3ml/s (contrast agent ≥300mlg/ml) and CT

scans were conducted in the arterial, portal venous, and venous contrast phase. Total contrast volume

depended on individual body weight. All images were acquired during breath hold in expiration, with

arms elevated above the head. Based on the portal venous data set multiple segmentation processes

are carried out to segment the liver, vessel trees, and the tumor, and to ultimately create a complete

3D model (10).

Parameters Non-enhanced Arterial phase Portal venous phase Venous phase

Covered Area Liver Liver Liver to Symphysis Liver

Delay (s) - 15 45 60 Collimation (mm) (256) × 0.625 (256) × 0.625 (256) × 0.625 (256) × 0.625 Rotation time (s) 0.75 0.75 0.75 0.75 mAs 180 220 250 250 kV 120 120 120 120 Reconstruction (mm) 2 2 2 2 Increment (mm) 0.5 0.5 0.5 0.5

Table 1. Scan parameters for planning CT imaging

Perfusion CT of the liver

Perfusion is a powerful source of tissue cooling and has the potential to affect heat transfer

substantially and ultimately influence the size of the induced thermal lesion (15). For example, in

liver cirrhosis hepatic artery flow is higher, compared to healthy liver tissue, while portal venous

flow is strongly reduced which leads to a lower overall perfusion (16,17). Dynamic CT measurements

after contrast administration were used to quantify tissue perfusion in the liver (18).

Page 8: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

8

The IMPPACT project showed that inclusion of patient specific perfusion values in the heat

transfer calculation can substantially improve the accuracy of the simulated lesions´ size (19). A

dedicated imaging protocol was developed to combine low radiation exposure with adequate image

quality (14).

CT perfusion analysis

The clinical partners of the ClinicIMPPACT project used CT scanners from three different

manufacturers (Siemens, Philips and Toshiba). The development of a common software tool for

standardized CT data analysis was therefore essential (19). Arterial and portal venous perfusions,

both measured in ml/min/100 ml, were identified as the most important perfusion parameters related

to the microcirculation in non-tumorous parenchyma. The maximum slope (MS) method was chosen

due to several specific advantages, e.g. short acquisition time and reduced radiation dose.

Regions of interest (ROI) were visually placed into the according hepatic regions and the

spleen, whose peak enhancement generally serves to separate arterial and portal-venous perfusion

from hepatic perfusion. Further details about CT perfusion analysis can be found here (19).

CT-guided Radiofrequency ablation

After acquisition of a non-enhanced CT scan, with the patient under full anesthesia, to

localize the tumor and plan optimal needle access, the IR inserted the probe using CT guidance. Once

the probe (RITA Starburst® XL, AngioDynamics, Latham, NY, USA) was in its final location the

hooks were completely deployed and a non-enhanced CT scan (under temporary apnea) was

performed. Once the IR confirmed the deployed needles’ proper position the CT data was then

transferred to an external PC for fast needle registration, which semi-automatically registers the

needle into the 3D model created prior to the intervention (10).

The RFA and computer simulation of the lesion (13) started simultaneously. The employed

standard- or modified ablation protocols can be found is provided in table 2. Optional to a standard

non-enhanced CT-scan after the ablation, a control CT with contrast agent was performed to evaluate

potential residual tumor. If the lesion was not sufficiently treated, needle adjustment and additional

Page 9: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

9

ablation was performed under CT-guidance. If the ablation result seemed sufficient, the needle was

removed under track ablation (14).

Deploy to

Target temperature (°C)

Power (Watt)

Duration (min) Procedure

2 cm 105 150 15 Heat until target temperature is reached, then deploy to 3 cm

3 cm 105 150 14.5 Heat until target temperature is reached, then deploy to 4 cm

4 cm 105 150 14 7 min after target temperature is reached, then deploy to 5 cm

5 cm 105 150 7 7 min after target temperature is reached, then cool down

Table 2. Ablation protocol for a 5 cm ablation using the RITA Starburst®-probe (RITA Starburst®

XL, AngioDynamics, Latham, NY, USA)

Real-time RFA lesion prediction

Because the 3D model of the liver was generated pre-operatively it was required to register

the heat source (i.e., the needle) to the precomputed model during the RFA procedure. We previously

designed a needle-registration method that offers both fast computing (< 2 minutes) and accuracy

(average absolute error < 5.0 mm).

In the recent IMPPACT project a prototype for RFA simulation was developed (20), that

provides state of the art visualization. Compared to others (21,22), the prototype was the first tool to

consider cell death instead of simple heat approximations for its simulation (23). The improved,

current version of the tool – the RFA Guardian performs all real time computations by using a modern

graphics processing unit (GPU) (24,25). A detailed description of the algorithm can be found

elsewhere (13).

Page 10: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

10

Follow-up imaging and postprocessing

Follow up imaging was performed 1 month after RFA via multiphase CT (non-enhanced, arterial,

portal venous, venous) for HCC lesions or using mono-phase CT (portal venous) for metastases, both

under breathold in expiration. All CT images, whether pre-, peri- or post-interventional (i.e. follow-

up) were evaluated by radiologists with more than 10 years of experience in abdominal imaging, at

each participating clinical site.

In a post-processing step the simulated lesion from the intra-procedural CT scan (acquired

in apnea under full anesthesia) was co-registered to the semi-automatically segmented true ablation

lesion from the follow up CT scan (acquired under breathold in expiration). Then volume- and

surface deviations between the simulated and the segmented lesion were computed (Figure 2). The

volumes of the simulated and true RFA lesion were determined by counting the number of matching

voxels, i.e. the voxels of simulation and segmented data, sharing the space coordinates, and dividing

by the sum of the voxels of the simulated and real lesions (10).

Page 11: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

11

Figure 2. Lesion shift due to registration error of simulated lesion (red, from intraproceduaral scan)

and segmented lesion (blue, from 1 month post RFA scan).

Statistical analysis

Normally distributed data were described using means and standard deviations, skewed

variables (simulated- and segmented lesion volumes) were described with medians and

interquartile ranges. Initial correlations between simulated- and segmented lesion volumes were

determined using Spearman rank correlations (ρ). Non-parametric One-Way ANOVA (Wilcoxon

Signed-Ranks Test) was used to assess differences between simulated- and segmented lesion

volumes.

To analyze primary outcome variables, we employed the following established scores to

represent volume- and surface deviation for the registered simulation and simulation alone,

respectively: Relative volume deviation (RVD), a ratio between the absolute volume difference to

the volume of the actual lesion. Surface deviation is calculated based on the Hausdorff distance

formula and represented by the average absolute error (AAE). The secondary outcome variable

(workflow efficiency of the intra-procedural software application) was investigated by comparing

the durations of the RFA procedures between a sub sample of study population (N=31) and a

retrospective control group (N=31), pair-matched for tumor volume. Both, durations and tumor

volumes were compared using Student´s t-test. The beginning and end of the RFA procedure were

defined as the “time stamp” on the planning- and post procedural CT scans.

Because tissue perfusion can potentially influence the resulting lesion size and because the

duration of the ablation can be an indicator of the complexity of the ablation we correlated these

variables with AAE and RVD using Pearson rank correlation coefficients (r). Significant correlation

between these variables were investigated further using multivariate linear regression with ablation

time, tissue perfusion and presence or absence of liver cirrhosis included as quantitative variables

and AAE and RVD as dependent variables, respectively. Cirrhotic tissue can acts as thermal

Page 12: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

12

insulation, preventing the applied heat to spread beyond the target area and may therefore alter the

ablation results (26). Effect sizes of multivariate linear regression models are reported as parameter

estimates with [95% confidence limits]. Statistical data analyses were performed using the SPSS

software (SPSS Inc., Chicago, Illinois, USA, version 18.0).

Results

Summary measures for continuous variables of the study population and lesions, and the

retrospective control group are shown in table 1. In the study group twenty seven patients presented

with HCC, seventeen with colorectal carcinoma metastasis, and three with other metastases (from

cholangiocellular-, ovarian-, and gastric carcinoma). Liver cirrhosis was present in twenty three

patients of the study group. Transarterial chemoembolization (TACE) was performed in seven

lesions of the study group prior to RFA.

There was a strong correlation between the simulated- and segmented lesion volume (r =

0.71, ρ<0.0001, Figure 3) and a Wilcoxon Signed-Ranks Test indicated that the median simulated-

and segmented lesion volumes did not differ significantly (Wilcoxon p=0.49, table 3). When

comparing the surface- and volume deviation (AAE and RVD, respectively) between the

simulated- and segmented lesion, based on the registration of the intra-procedural CT-scan to the

one month post ablation control scan we found an AAE of 6.1 mm (± 3.2 SD) and RVD of 1.3 %

(± 3.4 SD). By excluding the error that goes along with co-registration through translation of both

lesions into the same coordinate system, AAE improved significantly to 3.8 mm ((± 2.0 SD;

p<0.0001); RVD: 1.1 % (± 1.2 SD); p=0.59) when simulated and segmented lesion were compared

directly. This actual AAE (aAAE) was positively correlated to the duration of the ablation (r =

0.29, p<=0.05, Figure 5), but not with tissue perfusion (r = −0.18, p=0.22). RVD was not

associated with either tissue perfusion (r = −0.23, p=0.11), or with the duration of the ablation (r =

Page 13: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

13

0.07, p=0.62). In a linear model the duration of the ablation was an independent predictor of aAAE

(β=0.027±0.01 minutes, p=0.02).

Table 3. Demographic and anthropometric characteristics of the study population. Median

simulated versus segmented volumes did not differ significantly (Wilcocxon p=.49). The duration

of the treatment did differ significantly between the volume matched study group and the

retrospective control group (student´s t-test: P=0.04). BMI = body mass index. NA = non

applicable.

Variables Lesions in study group

(N=47; F=9; M=35)

Volume matched study group

(F=7; M=24)

Volume matched retrospective control group

(F=7; M=24)

Age (years) 68.2 ± 10.1 67.5 ± 11.2 64.8 ± 9.9

BMI (kg/m2) 27.2 ± 4.8 26.6 ± 4.3 26.7 ± 3.6

Tissue perfusion (ml/min/100 ml) 61.0 ± 24.9 NA NA

Tumor perfusion (ml/min/100 ml) 39.9 ± 30.9 NA NA

Tumor volume (cm3) 5.46 ± 5. 6.06 ± 5.6 6.05 ± 5.5

Simulated lesion volume (cm3) 17.28 [9.09 – 22.2] NA NA

Segmented lesion volume (cm3) 18.09 [7.77 – 22.97] NA NA

Duration of treatment (min) 93.0 ± 43.1 91.4 ± 31.4 77.4 ± 22.2

Duration of ablation (minutes) 39.4 ± 20.4 NA NA

Page 14: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

14

Figure 3 shows the correlation between simulated and segmented RFA – lesion volumes (r = 0.71,

ρ<0.0001, Figure 2). Excluding the two outliers (top right and bottom far right) did not change the

results significantly (r = 0.71, vs. r=0.67, both ρ<0.0001).

0102030405060708090

0 20 40 60 80

Sim

ulat

ed v

olum

e (c

m3 )

Segmented volume (cm3)

Page 15: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

15

Figure 4: Bland-Altman plot showing the limits of agreement (mean and 95% confidence interval)

between paired values for simulated and segmented RFA - lesion volumes.

Figure 5: The AAE was positively correlated to the duration of the ablation (r = 0.32, p=0.03), after

exclusion of the registration error.

-60-50-40-30-20-10

010203040

0 20 40 60 80

Sim

ulat

ed -

segm

ente

d vo

lum

e (c

m3 )

Mean of simulated and segmented volume (cm3)

0

20

40

60

80

100

120

0 2 4 6 8

Abla

tion

dura

tion

(min

utes

)

aAAE (mm)

Page 16: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

16

Discussion

In this first prospective clinical trial for the evaluation of the precision and efficacy of a

software application, developed to predict RFA induced hepatic tissue necrosis in patients with

liver malignancies, we found that the application predicted lesions with good accuracy and at a

high speed, suitable for clinical routine.

We found a rather large registration error, which is best explained by a variation in the

position and shape of the liver in the intra-procedural CT- scan (used for the simulation) and the 1

month – control scan (used for true lesion segmentation) (Figure 2). An animal study with 12

ablations in porcine livers found a mean deviation (point-to-mesh error) between predicted and true

ablation of 5.3 ± 3.6 mm. The problem is well known in the fields of radiotherapy and nuclear

medicine. In stereotactic body radiation therapy, this problem is conquered by the use of implanted

markers which act as localizing and tracking devices to optimize tumor targeting (27). In our cases

the differences in the position and shape of the liver, as a result of longitudinal or rotational

movements related to variations in diaphragm position, are severe because the intra-procedural

scan is acquired in end-expiration apnea with the patient under full anesthesia. This includes

complete relaxation of the diaphragm, while the 1 month control scan is performed in active

expiration with breathold in the conscious patient. Hence, it was not surprising to find a reduction

in surface deviation (AAE) of 37.5% and a reduction in volume deviation of 20.6 % (RVD) after

translation of the lesions into the same coordinate system. However, because the simulation is

computed based on the real intra-procedural CT scan with the RFA – probe in place this issue is

only relevant for this evaluation study but irrelevant towards the accuracy of the simulation

algorithm itself.

In a preliminary analysis we found that the surface deviation decreased by 5-10% when

including measured perfusion values, rather than estimated CT-perfusion values (13). The

importance of hepatic blood flow and the proximity of the tumor to blood vessels is further

highlighted when taking a closer look at two outliers indicated in Figure 3. In one case the

Page 17: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

17

segmented (true) lesion was underestimated by ca. 50 cm3, in another case overestimated by

roughly 30 cm3. The explanation for the significant underestimation of the resulting ablation zone

was due to unpredicted arterial damage in the first case which led to non-perfusion and finally

necrosis of a large (unpredicted) tissue areal adjacent to the ablation site. The reason for the

significant overestimation of the ablation in the second example is unclear, but most likely due to

miscalculations based on the provided perfusion values of the tissue. Especially the first example

displays one limitation of the simulation in regards of predicting vessel damage in a highly

perfused organ.

Interestingly, duration of the ablation was an independent predictor of surface deviation. A

prolongation of the ablation of 37 minutes would lead to one additional millimeter of surface

deviation. Consequently, one could argue that lager lesions and lesions in proximity to large

vessels, which require longer ablation phases might be less suitable for adequate lesions prediction.

The Bland–Altman plot further underlines the fading precision of the lesion prediction with larger

volumes.

In regards to efficacy we identified data transfer to be one of the most limiting factors for

sufficient implementation of the software application into the clinical workflow. In this study, the

software was not connected to the local data networks, due to data security concerns. This led to

time consuming data transfer from the CT scanner to the software workstation, via external data

drive. This fact explains the average difference in procedure duration of 14 minutes between the

trial and a matched retrospective control group. It would therefore be desirable to integrate the

system into the local hospital network (in this specific case after FDA approval as medical device)

and therefore eliminate this confounder.

In summary, our results indicate good precision and efficacy of a dedicated software

solution to accurately predict lesion size and shape after radiofrequency ablation of liver tumors.

As mispredictions of true ablation zones can lead to increased tumor recurrence due to under

treatment or complications due to overtreatment this tool might be of interest to the interventional

Page 18: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

18

radiology community. It potentially helps treating tumors in close proximity to larger vessels and

might support less experienced IR´s to perform RFA with confidence. Further randomized clinical

phase II trials are needed to test whether the use of this tool improves patient outcome.

Conflict of interest

The authors declare no conflict of interests.

Acknowledgments

The research leading to these results has received funding from the European Community's Seventh

Framework Program under grant agreement no. 610886, project ClinicIMPPACT and grant

agreement no. 600641, project GoSmart. Furthermore we would like to thank and name all associated

investigators: Martin van Amerongen, Jan Egger, Jukka Ilari, Philipp Stiegler, Horst Portugaller,

Dieter Schmalstieg, Mark Dokter, Phil Weir, Nikita Garnov, Tuomas Alhonnoro, Miko Lilja and

Bianca Schmerböck.

Abbreviations: BF: blood flow; BV: blood volume; CT: computed tomography; EU: European

Union; HCC: hepatocellular carcinoma; IMPPACT: Intervention Modelling, Planning and Proof for

Ablation Cancer Treatment; RFA: radio frequency ablation; US: ultrasound.

Page 19: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

19

References

1. European Association For The Study Of The Liver, European Organisation For Research

And Treatment Of Cancer. EASL-EORTC clinical practice guidelines: management of

hepatocellular carcinoma. J Hepatol. 2012 Apr;56(4):908–43.

2. Crocetti L, de Baere T, Lencioni R. Quality improvement guidelines for radiofrequency

ablation of liver tumours. Cardiovasc Intervent Radiol. 2010 Feb;33(1):11–7.

3. Bruix J, Sherman M, Practice Guidelines Committee, American Association for the Study

of Liver Diseases. Management of hepatocellular carcinoma. Hepatol Baltim Md. 2005

Nov;42(5):1208–36.

4. Thandassery RB, Goenka U, Goenka MK. Role of local ablative therapy for hepatocellular

carcinoma. J Clin Exp Hepatol. 2014 Aug;4(Suppl 3):S104-111.

5. Wong SL, Mangu PB, Choti MA, Crocenzi TS, Dodd GD, Dorfman GS, et al. American

Society of Clinical Oncology 2009 clinical evidence review on radiofrequency ablation of hepatic

metastases from colorectal cancer. J Clin Oncol Off J Am Soc Clin Oncol. 2010 Jan 20;28(3):493–

508.

6. Fukuhara T, Aikata H, Hyogo H, Honda Y, Morio K, Morio R, et al. Efficacy of

radiofrequency ablation for initial recurrent hepatocellular carcinoma after curative treatment:

Comparison with primary cases. Eur J Radiol. 2015 Aug;84(8):1540–5.

7. Chang IA, Nguyen UD. Thermal modeling of lesion growth with radiofrequency ablation

devices. Biomed Eng Online. 2004 Aug 6;3(1):27.

8. Altrogge, Inga, et al. Sensitivity analysis for the optimization of radiofrequency ablation in

the presence of material parameter uncertainty. Int J Uncertain Quantif. 2.3 (2012).

9. Audigier, Chloé, et al. Efficient lattice Boltzmann solver for patient-specific

radiofrequency ablation of hepatic tumors. IEEE Trans Med Imaging. 34.7 (2015): 1576-1589.

Page 20: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

20

10. Payne S, Flanagan R, Pollari M, Alhonnoro T, Bost C, O’Neill D, et al. Image-based multi-

scale modelling and validation of radio-frequency ablation in liver tumours. Philos Transact A Math

Phys Eng Sci. 2011 Nov 13;369(1954):4233–54.

11. Audigier C, Mansi T, Delingette H, Rapaka S, Passerini T, Mihalef V, et al. Comprehensive

preclinical evaluation of a multi-physics model of liver tumor radiofrequency ablation. Int J

Comput Assist Radiol Surg. 2017 Jan 17;

12. Voglreiter P., Mariappan P. , Alhonnoro, T. Busse H., Weir P. , M.Pollari, R.Flanagan, M.

Hofmann, D. Seider, P. Brandmaier, M.J. van Amerongen, R. Rautio, S. Jenniskens, R. Blanco

Sequeiros, R.H. Portugaller, P. Stiegler, J. Futterer,, D. Schmalstieg, M. Kolesnik, M. Moche. RFA

guardian: comprehensive simulation of the clinical workflow for patient specific planning,

guidance and validation of RFA treatment of liver tumors. International Journal of Computer

Assisted Radiology and Surgery June 2016, Volume 11, Supplement 1, pp 1–286.

13. Mariappan P, Weir P, Flanagan R, Voglreiter P, Alhonnoro T, Pollari M, et al. GPU-based

RFA simulation for minimally invasive cancer treatment of liver tumours. Int J Comput Assist

Radiol Surg. 2017 Jan;12(1):59–68.

14. Reinhardt M, Brandmaier P, Seider D, Kolesnik M, Jenniskens S, Sequeiros RB, et al. A

prospective development study of software-guided radio-frequency ablation of primary and

secondary liver tumors: Clinical intervention modelling, planning and proof for ablation cancer

treatment (ClinicIMPPACT). Contemp Clin Trials Commun. 2017 Dec 1;8:25–32.

15. Wu D, Tan M, Zhou M, Sun H, Ji Y, Chen L, et al. Liver computed tomographic perfusion

in the assessment of microvascular invasion in patients with small hepatocellular carcinoma. Invest

Radiol. 2015 Apr;50(4):188–94.

16. Chen L, Zhang L, Liang M, Bao J, Zhang J, Xia Y, et al. Magnetic resonance imaging with

gadoxetic acid disodium for the detection of hepatocellular carcinoma: a meta-analysis of 18

studies. Acad Radiol. 2014 Dec;21(12):1603–13.

Page 21: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

21

17. Lu X, Kikuchi H, Hirooka K, Isobe Y, Watanabe H, Kobayashi Y, et al. Method for

estimating the temperature distribution associated with the vessel cooling effect in radio frequency

ablation. Conf Proc Annu Int Conf IEEE Eng Med Biol Soc IEEE Eng Med Biol Soc Annu Conf.

2015 Aug;2015:4836–9.

18. Kim SH, Kamaya A, Willmann JK. CT perfusion of the liver: principles and applications in

oncology. Radiology. 2014 Aug;272(2):322–44.

19. van Amerongen M. J., Garnov N., Jenniskens, S. F. M. Alhonnoro T., Pollari M. Liver CT

perfusion values for colorectal liver metastases – a protocol for future use in radiofrequency

simulation software. Poster presented at: European Congress of Radiology 2016;

20. Kerbl, Bernhard, et al. Intervention planning of hepatocellular carcinoma radio-frequency

ablations. Springer Berl Heidelb 2012.

21. Villard C, Soler L, Gangi A. Radiofrequency ablation of hepatic tumors: simulation,

planning, and contribution of virtual reality and haptics. Comput Methods Biomech Biomed Engin.

2005 Aug;8(4):215–27.

22. Schumann C, Rieder C, Haase S, Teichert K, Süss P, Isfort P, et al. Interactive multi-criteria

planning for radiofrequency ablation. Int J Comput Assist Radiol Surg. 2015 Jun;10(6):879–89.

23. Mayrhauser U, Stiegler P, Stadlbauer V, Koestenbauer S, Leber B, Konrad K, et al. Cell to

cell interactions influence sensitivity of liver cell lines during hyperthermia. Anticancer Res. 2011

Nov;31(11):3713–7.

24. Rieder, Christian, et al. A Shader Framework for Rapid Prototyping of GPU�Based

Volume Rendering. Computer Graphics Forum Vol. 30. No. 3. Blackwell Publishing Ltd, 2011.;

25. Steinberger, Markus, et al. Softshell: dynamic scheduling on GPUs.

26. Livraghi T, Goldberg SN, Lazzaroni S, Meloni F, Solbiati L, Gazelle GS. Small

Hepatocellular Carcinoma: Treatment with Radio-frequency Ablation versus Ethanol Injection.

Radiology. 1999 Mar 1;210(3):655–61.

Page 22: Clinical Intervention Modelling, Planning and Proof for ... · Clinical Intervention, Modeling, Planning and Proof for Ablation Cancer Treatment ... to clinically validate a comprehensive

22

27. Dawood O, Mahadevan A, Goodman KA. Stereotactic body radiation therapy for liver

metastases. Eur J Cancer Oxf Engl 1990. 2009 Nov;45(17):2947–59.