clinical applications and maintenance proceedures for laboratory equipment
TRANSCRIPT
CLINICAL APPLICATIONS AND MAINTENANCE PROCEEDURES FOR
LABORATORY EQUIPMENT
Without reference, identify at least four out of six basic facts about the clinical applications of a centrifuge.
LABORATORY CENTRIFUGE
Purpose Machine used to separate substances of different
densities that are suspended in a fluid Centrifuging only speeds up what will eventually
happen with the force of gravity You are, therefore, creating “artificial gravity”
LABORATORY CENTRIFUGE
Works because heavier particles move farther from the center of rotation and collect at the bottom of the container
Some common applications: Separating red blood cells from whole blood Separating plasma from whole blood Separating platelets from plasma Separating sediment from urine
LABORATORY CENTRIFUGE
Centrifuge components: An electric motor to produce rotation An apparatus called a rotor (or head) to hold the
samples Come in many different sizes depending upon the
centrifuge and application Hold varying amounts and/or sizes of sample
containers (usually tubes)
LABORATORY CENTRIFUGE
Two basic rotor designs:Fixed position (called fixed angle or vertical rotor) –
samples remain in an upright, vertical position during rotation
Swing-bucket rotors – tube holders are moveable and swing into a horizontal position during rotation
Rotors must always be balanced for proper and safe rotation (the higher the rotation speed the more exact the balance must be)
LABORATORY CENTRIFUGE
Controls for speed and time
Some units (primarily small units) have a set speed (cannot be changed)
Many units have variable speeds
Nearly all units have variable timers
LABORATORY CENTRIFUGE
Safety features (should be mandatory on all units purchased) Lid locks to prevent lid from being opened while
rotor is spinning Out-of-balance detector to stop rotation if rotor is
not balanced properly Optional features such as refrigeration or
heating system can be added
LABORATORY CENTRIFUGE
Terms Centrifugal force – the force that makes rotating
bodies move away from the center of rotation Centrifugation – the process by which substances
of different densities are separated from one another by centrifugal force
LABORATORY CENTRIFUGE
G-Force (or G’s) – unit of measurement for samples undergoing centrifugation (a force of 1g is equal to the force of gravity at the earth’s surface)
Relative centrifugal force (RCF) – the force (measured in G’s) that a sample undergoes and depends upon:
LABORATORY CENTRIFUGE
Speed of rotation (measured in revolutions per minute [RPM])
The rotating radius (the distance measured from the center axis of the rotor to the extreme tip of the tube
Formula for calculating the RCF using the RPM’s from a centrifuge display
RCF = (1.119 x 10-5) x RPM² x r• 1.119 x 10-5 is a conversion factor• r = radius of the rotor in centimeter
LABORATORY CENTRIFUGE
Basic factors that a laboratory technician considers when spinning a sample
Speed of rotation This factor, of course, determines the RCF that the
sample undergoes Usually displayed as RPM on the centrifuge display Some units can calculate and display the actual RCF
LABORATORY CENTRIFUGE
Amount of time the sample should spin (centrifugation time)• The time of centrifugation includes the time of
acceleration, but not deceleration• The higher the force of gravity, the less time it will
take for separation to occur• The time required is therefore inverse to the
gravities generated by the centrifuge• Amount of time will also depend upon what is being
centrifuged (some materials take longer to separate than others)
LABORATORY CENTRIFUGE
Types of laboratory centrifuges Common laboratory centrifuges
Operate at speeds of about 3,000 RPM, but can attain speeds as high as 7,000 RPM
Some specialized centrifuges are equipped with heating or refrigeration systems to provide controlled temperatures
Two common configurations: Floor standing (larger, hold more/bigger samples) Tabletop (smaller units)
LABORATORY CENTRIFUGE
Microhematocrit centrifuges Very small units capable of high speeds in the range of
10,000 – 11,000 RPM Used with tiny capillary tubes that hold a very small
whole blood sample Used for quickly estimating red blood cell counts during
patient screening and blood donation procedures
LABORATORY CENTRIFUGE
Micro centrifuges• Small, tabletop, high-speed units with maximum
speeds around 13,000 – 15,000 RPM• Used for a wide variety of molecular biology
studies and various procedures that require only a short spin cycle
LABORATORY CENTRIFUGE
Ultracentrifuges Large, high-cost, specialized machines Used mostly in research labs for separation of
proteins and nucleic acids Capable of speeds in the range of 20,000 –
100,000 RPM All ultracentrifuges have refrigeration systems to
compensate for the heat generated by the high rotation speeds
LABORATORY CENTRIFUGE
LABORATORY CENTRIFUGE
General Maintenance Aside from the necessary preventive maintenance
checks, routine verification of RPM and time are mandatory
RPM is normally determined using• Handheld photo-tachometer• Strobe light• Vibrating reed tachometer
Motor brushes (on units that utilize them) should be checked at regular intervals
LABORATORY CENTRIFUGE
Without reference, answer questions with at least a 70% accuracy about the laboratory refrigerator.
LABORATORY REFRIGERATOR
Most refrigeration systems are based on a vapor compression cycle for cooling.
This is a cycle where a refrigerant is forced to change from liquid to vapor and back which causes a strong cooling reaction.
This is explained in thenext few slides
LABORATORY REFRIGERATOR
There are four main components to a refrigeration system Compressor – Refrigerant enters compressor as a
vapor and is compressed to a high pressure Condenser – cools the vapor until it starts
condensing
LABORATORY REFRIGERATOR
Expansion Valve (also called throttle valve) – liquid refrigerant goes through which causes the pressure to abruptly decrease causing flash evaporation and auto-refrigeration. This remaining mixture is part liquid and part vapor at a low temperature and pressure.
LABORATORY REFRIGERATOR
Evaporator – The refrigerant then goes through the evaporator coils or tubes where a fan circulates air from within the refrigerator across the coils, thus cooling the air in the refrigerator
LABORATORY REFRIGERATOR
Calibration and Maintenance Temperature check• Use calibrated thermometer to verify temperature is
within tolerance Clean dust and dirt off of any fans as well as off of
the evaporator coils or tubes.
LABORATORY REFRIGERATOR
Without reference, answer questions with at least a 70% accuracy about the laboratory microscope.
LABORATORY MICROSCOPE
Definition – A microscope is an instrument used to look at objects to small to see with the naked unaided eye.
The most common type of laboratory microscope is an optical microscope. This is usually in the form of a light microscope. A light microscope uses a series of lenses to allow
you to view a lighted object with up to 1500 times magnification.
LABORATORY MICROSCOPE
There are a few components of note in a light microscope. Eyepieces – this is the part that you look directly
through. They normally contain a lens with a small magnification amount – usually around 10 times.
Objectives – these are the main lenses in the microscope and are responsible for the largest portion of magnification
Stage and stage controls – the stage is the surface on which your specimen is placed. The stage can be moved a small amount using the stage controls in most microscopes. The controls are normally separated into coarse and fine adjustments.
LABORATORY MICROSCOPE
Diaphragm and light source – The light source is usually located belowthe stage with the diaphragm between them to control the amount and intensity of the light that reachesthe specimen.
LABORATORY MICROSCOPE
Common Maintenance The most common thing that will need to be done
with a microscope is the replacement of the light source bulb.• This is a very simple procedure in most units
Other that bulb replacement you may be required to clean the lenses or clean or repair the coarse and fine adjustment mechanisms. Anything beyond this normally requires sending the microscope out for service, but luckily problems of this nature are extremely rare.
LABORATORY MICROSCOPE
• Coarse and fine adjustment mechanismso The coarse and fine adjustment mechanisms are based
on a series of gears attached to two different adjustment knobs. As each of these knobs are turned the stage is moved. The coarse adjustment causes the stage to move quickly while the fine adjustment moves it slowly. This causes the specimen to move which helps to focus the image.
LABORATORY MICROSCOPE
o There is one thing to keep in mind about the maintenance of the coarse and fine adjustment mechanisms.From time to time the gears that control these
adjustments will become dirty or clogged with dirt and/or other debris.
You should know when this is the case because you will have difficulty adjusting the microscope.
Stage either won’t move at all or will move with short jerky motions. In this situation you must disassemble and clean the
adjustment mechanisms. When you do this you must be extremely careful to keep track of all of the small pieces within the microscope. Nothing is worse than going through all of the work to fix something and then it won’t work because you lost a piece.
LABORATORY MICROSCOPE
LABORATORY MICROSCOPE This is not the way to fix a microscope!