Cutting Accessories • Burs • Blades • K-Wires • Steinman Pins • Twist Drills
CLEANING & STERILIZATION INSTRUCTIONS
2
EXPLANATION OF SYMBOLS
Authorized representative in the European community.
Batch code.
Catalog number.
Conforms with the essential requirements of the European community directives.
Consult instructions for use.
Date of manufacture.
Do not re-sterilize.
Do not reuse.
Do not use if package is damaged.
Manufacturer.
Non-sterile.
Sterilized using irradiation.
Use by.
PghCiNBDLMd
IK H
3
CLEANING RECOMMENDATIONS
CAUTIONS & WARNINGS:
• Donotre-useorreprocessusedcuttingaccessories.• Followtheinstructionsandwarningsissuedbythesuppliersofanycleaninganddisinfectionagentsandequipmentused.• Followuniversalprecautionsandprotectiveapparelwhenhandlingandcleaningcontaminatedcuttingaccessories.• Disposeofallcuttingaccessoriesproperlyaftereachuse.• Undercertainclassificationsofrisk, theWorldHealthOrganization(WHO),or localregulatoryauthoritiesrecommendspecialCJD(Creutzfeldt-JakobDisease)inactivationprocessingprocedures.ConsultWHOandlocalregulationsforfurtherinformation.
• Donotcleanorre-sterilizecannulateddrills,toolsteel,440steel,laserweldedorbrazedtwopieceblades.• Donotprocessaccessoriesinautomatedwasher-sterilizer.
General to all Cutting Accessories: • Onlyusewithappropriatepowersystem.Seeapplicablepowersysteminstructionsforuse. • Donotusecuttingaccessoryatspeedsexceedingthehandpiecemanufacturer’sspecifications. • Disposeofproductorrecycleinaccordancewithlocallawsandregulations. Burs,RaspsandTwistDrills: • Avoidapplicationsofexcessivelateralforce. • Useproperburguardwithburperthehandpiecemanufacturer’sspecifications. • Donotusecarbidebursfortwistingorbendingcuts,breakagemayoccur.
4
Saw Blades: • Donotuseexcessiveforcetotwistorbendblades. • Cuttinginmetalguidesoraccidentalcontactwithmetalretractorsmaydamagethebladeandmaynecessitateitsreplacement. • Insertsawbladeintoalignmentguideorcuttingfixturepriortoactivatinghandpiece. • Sawbladesbecomehotfromfriction.Irrigationofbladesisrecommendedduringuseandisrequiredwhenusinganalignmentguideorcuttingfixture.
K-WiresandPins: • Immobilizethepatient’sareaoffixationtopreventexcessiveforcesontheK-Wireorpin. • ExcessiveforcesontheK-Wireorpinmaycauselooseningormetalfatigueandsubsequentdevicefracture. • K-Wiresorpinshouldberemovedassoonasadequatehealinghasoccurred.
5
CLEANING PROCEDURES:
• Itisalwaysrecommendedtouseunusedsterilecuttingaccessoriesineachprocedure.• Unusedcuttingaccessoryremovedfromoriginalnon-sterilepackage: 1. Sterilize as directed on the table on page 8.• Unusedcuttingaccessorycontaminatedduringsurgery: 1. Removecuttingaccessoriesfromhandpieceandorattachment. 2. Clean cutting accessories as soon as possible to avoid drying of contaminants. Disinfecting the cutting accessories may be needed to protect personnel
handlingtheinstrumentbeforecleaningandsterilization.Ifcleaningmustbedelayed,immersethecuttingaccessoryinaneutral-pHcleaningsolutiontoprevent drying of contaminants on the instrument.
SawBladesandK-Wires: 1. Rinsesoiledcuttingaccessorythoroughlywithalargeamountoflukewarm(targetlowerendofthe72-110°F/22-43°Ctemperaturerange)runningtap
waterforaminimumofone(1)minute. 2. MakeanenzymaticcleaningsolutionsuchasSTERIS®Prolystica™2xConcentrateEnzymaticPresoakandCleaner[1/8ounceperone(1)gallonofwarm
tapwater](targetlowerendofthe91-109°F/33-43°Ctemperaturerange). 3. Soakthecuttingaccessoryintheenzymaticcleaningsolutionfortwo(2)minutes,makingsurethecuttingaccessoryiscompletelyimmersedtoprevent
aerosolization of contaminants. 4. Manuallycleanthecuttingaccessoryusingasoft-bristledbrushforone(1)minutetoremovesoilanddebris. 5. Rinsesoiledcuttingaccessorybyimmersingthecuttingaccessoryinasufficientvolume(enoughtocoverthedevice)oflukewarm,tapwater(targetlower
endofthe72-110°F/22-43°Ctemperaturerange)forone(1)minute. 6. RepeatStep5two(2)moretimesforatotalofthree(3)rinses.Visuallyinspectcuttingaccessoriesforthepresenceorabsenceofresidualsoiland/orcleaner. 7. Dryimmediatelywithacleanlint-freesoftcloth. 8. Inspectfordamageandreplacewithnewcuttingaccessoryifdamaged. 9. Sterilizeasdirectedonthetableonpage8.
6
Burs - Recommended Manual Cleaning Procedure (Rotary Instruments): 1. Rinsethebursinrunningcooltapwaterforone(1)minute. 2. Prepareafreshneutral-pH,enzymaticcleaningsolution(suchasEnzol)perManufacturer’sworst-caseinstructionsbyusingone(1)ouncepergallonof
warm(86-109°F/30-43°C)tapwater. 3. Immersethebursinthecleaningsolutionandsoakinthepreparedsolutionforten(10)minutes. 4. Usingastiff-bristledbrush,cleaneachburindividuallyforaminimumofone(1)minutetoremovecontaminants.Careshouldbetakentocleancrevices
and other hard to reach areas thoroughly. 5. Thoroughlyrinsethebursunderrunningwarmtapwater(temperaturetonotexceed113°F/45°C)foraminimumofone(1)minuteanduntilvisiblyclean. 6. PrepareafreshneutralpHenzymaticsolution(seeStep3)inasonicatorbath.Completelysubmergethebursinthesolutionandsonicateforfifteen(15)
minutes. Separate burs from contacting and damaging each other. 7. Rinsedeviceusingrunningtapwater(temperaturetonotexceed113°F/45°C)foraminimumofone(1)minute. 8. Visuallyinspectdevicesforresidualsoil. 9. Dryimmediatelywithacleanlint-freesoftcloth. 10. Inspectfordamageandreplacewithnewcuttingaccessoryifdamaged. 11. Sterilize as directed on the table on page 8.
7
STERILIZATION RECOMMENDATIONS
WARNINGS:
• Theuseofdisinfectingsolutionsforanexteriorinstrumentwipewillnotsterilizethecuttingaccessory.• Inspectsterilizedcuttingaccessories.Ifdiscolorationorrustisvisibledonotuse.
CAUTIONS:
• Donotprocessinautomatedwasher-sterilizer.• Donotre-useorreprocessusedcuttingaccessories.• Donotusecoldsterilizationsolutionsasthesecontainoxidizingagentswhichmaycausedamagetothedevice.• Donotsterilizewhileconnectedtohandpieceandorattachment.• WraptheinstrumentsperAAMIrecognizedguidelines.• Followthesterilizermanufacturer’swritteninstructionsforcycleparameters,loadconfigurationandAAMIguidelinesforsteamsterilization.• Donotcleanorre-sterilizecannulateddrills,toolsteel,440steel,laserweldedorbrazedtwopieceblades.
8
PARAMETERS FOR STERILIZATION
Blades/K-Wires/SteinmanPins/TwistDrills
Burs–StainlessSteel,CarbideandDiamond
Note:Theseprocesseshavebeenvalidatedforaone-timesterilizationasbeingcapableofcleaningandsterilizingBurs/Blades/K-Wires/SteinmanPins/TwistDrills.
SteamSterilizationType MinimumTemperature MinimumExposureTime MinimumDryTimePre-Vacuum(Wrapped) 270°F(132°C) 4minutes 30minutesGravity(Wrapped) 270°F(132°C) 15 minutes 30minutes
SteamSterilizationType MinimumTemperature MinimumExposureTime MinimumDryTimePre-Vacuum(Pouch) 273°F(134°C) 3minutes 30minutesGravity(Pouch) 275°F(135°C) 10minutes 30minutes
9
RETURN GOODS POLICY
Contact your distributor regarding returned goods policy.
10
PRODUCT DISPOSAL
Dispose of product or recycle in accordance with local laws and regulations.
11
BrasselerU.S.A.Medical,LLCOneBrasselerBoulevard•Savannah,GA31419800-535-6638Ext.7050•912-921-7578(fax)BUSAMedical.com
Emergo EuropeMolenstraat 15
2513BH,TheHagueTheNetherlands
Tel:+31(0)70-345-8570Fax:+31(0)70-346-7299
MP
T-3881-REV1-08.13-XXM-XXX©2013BrasselerU.S.A.Medical,LLC•Allrightsreserved•PrintedintheU.S.A.
C 0470
BrasselerU.S.A.Medical,LLC,hasimplementedaqualitymanagementsystemthatiscertifiedunderISO13485:2003.STERIS®andProlystica™aretrademarksofSTERISCorporation.BrasselerU.S.A.Medical,LLC,isnotaffiliatedwiththeaboveandmakesnoclaimtocopyrightsortrademarkswhicharethepropertyofthiscompany.BUSA®isatrademarkofPeterBrasselerHoldings,LLC,oritsaffiliates.Allothertrademarksaretrademarksoftheirrespectiveownersorholders.Colors,specificationsandproductavailabilitysubjecttochange.BUSA®SurgicalPower&AccessoriesproductsaresoldbyBrasselerU.S.A.Medical,LLC,OneBrasselerBoulevard,Savannah,Georgia31419,UnitedStates.