citsens bio technical data sheet - c-cit sensors ag...
TRANSCRIPT
Technical Data Sheet CITSens Bio / MeMo
Content CITSens Bio........................................................................................................................................2 CITSens MeMo...................................................................................................................................3 Sensortypes..........................................................................................................................................4
PG 13.5 Sensor................................................................................................................................4Polymer Coated Sensor....................................................................................................................5Cap Sensor.......................................................................................................................................6
Page 1/6 C-CIT Sensors AG TechSheetBio_V1
CITSens Bio
Available Sensors
Screen Printed Electrode (SPE) with Immobilized Enzyme
Delivery Gamma irradiated (25 kGy), double packed
Dimension (SPE) Length: 230 mm or 330 mmTip Diameter: 8 mmThickness: 0.15 mm
Analytes GlucoseMeasuring range 0 - 30 mmol/L 0 - 5.4 g/LPrecision +/- 0.1 mmol/LResolution 0.1 mmol/L
LactateMeasuring range 0 - 15 mmol/L 0 – 1.35 g/LPrecision +/- 0.1 mmol/LResolution 0.1 mmol/L
Glutamate On request
Page 2/6 C-CIT Sensors AG TechSheetBio_V1
CITSens MeMo
Available Sensors
Screen Printed Electrode (SPE)
Delivery Gamma irradiated (25 kGy), double packed
Dimension (SPE) Length: 230 mm or 330 mmTip Diameter: 8 mmThickness: 0.15 mm
Analytes Glucose / Lactate combined
GlucoseMeasuring range 0 - 30 mmol/L 0 - 5.4 g/LPrecision +/- 0.1 mmol/LResolution 0.1 mmol/L
LactateMeasuring range 0 - 15 mmol/L 0 – 1.35 g/LPrecision +/- 0.1 mmol/LResolution 0.1 mmol/L
Glucose / Glutamate combined
GlucoseMeasuring range 0 - 30 mmol/L 0 - 5.4 g/LPrecision +/- 0.1 mmol/LResolution 0.1 mmol/L
Glutamate On request
pH
pHMeasuring range 4 – 9 pHPrecision +/- 0.1 pHResolution 0.05 pH
Page 3/6 C-CIT Sensors AG TechSheetBio_V1
Available Winter 2
017 / 2018
Sensortypes
PG 13.5 Sensor for SUB
The PG 13.5 Sensor is the best choice for any stirred bioreactors.
Through the shape based on a standard pH-probe the PG 13.5 Sensor is easily integradeable to any stirred bioreactor via lid or side port.
Using PEEK as main in process material the PG 13.5 sensor provides a high biosafety grade to not affect your culture's behaviour.
Sensor specifications:
Length: 40 – 300 mm Material: PEEK (USP Class VI)
In Process Time:Glucose: 2 – 3 WeeksLactate: 2 WeeksGlutamate: On request
Product Number Description Pcs / Unit Delivery time
3002-511.1 CITSens Bio Glucose Sensor, PG 13.5, short (-160mm) 12 3 Weeks
3002-511.2 CITSens Bio Glucose Sensor, PG 13.5, long (-300mm) 6 3 Weeks
3002-512.1 CITSens Bio Lactate Sensor, PG 13.5, short (-160mm) 12 3 Weeks
3002-512.2 CITSens Bio Lactate Sensor, PG 13.5, long (-300mm) 6 3 Weeks
3002-521.1 CITSens MeMo Glu / Lac Combi Sensor, PG 13.5, short (-160mm) 12 3 Weeks
3002-521.1 CITSens MeMo Glu / Lac Combi Sensor, PG 13.5, long (-300mm) 6 3 Weeks
+41 79 935 74 [email protected]
Page 4/6 C-CIT Sensors AG TechSheetBio_V1
Polymer Coated Sensor
The Polymere Coated Sensor is the best choice for any static cultivation.
Using a polymere coating instead of PEEK materials makes the sensor capable to measure in small voluminas in almost every kind of reactor.
Through the elasticity of the screen printed electrode in combination with the polymer coating makes the sensor toour first choice when developing a customized solution.
Sensor specifications:
Length: 40 – 300 mm Material: Polymer (USP Class VI)
In Process Time:Glucose: 2 WeeksLactate: 2 WeeksGlutamate: On request
Product Number Description Pcs / Unit Delivery time
3002-511.1 CITSens Bio Glucose Sensor, PG 13.5, short (-160mm) 24 3 Weeks
3002-511.2 CITSens Bio Glucose Sensor, PG 13.5, long (-300mm) 24 3 Weeks
3002-512.1 CITSens Bio Lactate Sensor, PG 13.5, short (-160mm) 24 3 Weeks
3002-512.2 CITSens Bio Lactate Sensor, PG 13.5, long (-300mm) 24 3 Weeks
3002-521.1 CITSens MeMo Glu / Lac Combi Sensor, PG 13.5, short (-160mm) 24 3 Weeks
3002-521.1 CITSens MeMo Glu / Lac Combi Sensor, PG 13.5, long (-300mm) 24 3 Weeks
+41 79 935 74 [email protected]
Page 5/6 C-CIT Sensors AG TechSheetBio_V1
Cap Sensor
The Cap Sensor is the best choice for long therm monitoring in agitated environments where the PG 13.5 Sensor is not suitable.
Using PEEK as main in process material the Cap Sensor provides a high biosafety grade to not affect your culture's behaviour.
Sensor specifications:
Length: 40 – 300 mm Material: Polymer (USP Class VI)
In Process Time:Glucose: 2-3 WeeksLactate: 2 WeeksGlutamate: On request
Product Number Description Pcs / Unit Delivery time
3002-501.1 CITSens Bio Glucose Sensor, Cap Sensor, short (-160mm) 24 3 Weeks
3002-501.2 CITSens Bio Glucose Sensor, Cap Sensor, long (-300mm) 24 3 Weeks
3002-502.1 CITSens Bio Lactate Sensor, Cap Sensor, short (-160mm) 24 3 Weeks
3002-502.2 CITSens Bio Lactate Sensor, Cap Sensor, long (-300mm) 24 3 Weeks
3002-521.1 CITSens MeMo Glu / Lac Combi Sensor, Cap Sensor, short (-160mm) 24 3 Weeks
3002-521.1 CITSens MeMo Glu / Lac Combi Sensor, Cap Sensor, long (-300mm) 24 3 Weeks
+41 79 935 74 [email protected]
Page 6/6 C-CIT Sensors AG TechSheetBio_V1