chapter 9 telmesartan - shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/8225/16/16_chapter...
TRANSCRIPT
CHAPTER 9
TELMESARTAN
242
9.1. DRUG PROFILE
Telmisartan is a member of a family of drugs called angiotensin receptor
blockers (ARBs), which includes losartan (Cozaar), valsartan (Diovan), irbesartan
(Avapro), and candesartan (Atacand). Telmisartan was approved by the FDA in
November of 2000. Telmisartan is used alone or in combination with other
medications to treat high blood pressure.
Telmisartan is an angiotensin II receptor blocker that shows high affinity for
the angiotensin II receptor type 1 (AT1), with a binding affinity 3000 times greater for
AT1 than AT2. It has the longest half-life of any ARB (24 hours) [1-2]
and the largest
volume of distribution]. In addition to blocking the RAs, Telmisartan acts as a
selective modulator of peroxisome proliferator-activated receptor gamma (PPAR-γ), a
central regulator of insulin and glucose metabolism. It is believed that Telmisartan’s
dual mode of action may provide protective benefits against the vascular and renal
damage caused by diabetes and cardiovascular disease (CVD).[2]
It is used to treat
high blood pressure (hypertension [1]
). High blood pressure reduction helps prevent
strokes, heart attacks, and kidney problems. This drug works by blocking the hormone
angiotensin thereby relaxing blood vessels, causing them to widen. This drug may
also be used to treat congestive heart failure and to help protect the kidneys from
damage due to diabetes.
Figure: 9.A: Structure of Telmisartan
243
Systematic (IUPAC) name : 2-(4-{[4-methyl-6-(1-methyl-1H-1,3-benzodiazol-
2-yl)- 2-propyl-1H-1,3-benzodiazol-1 yl] methyl}
phenyl) benzoic acid
Formula : C33H30N4O2
Mol. Mass : 514.617 g/mol
Routes : Oral
Excretion : Faecal 97%
This drug can cause serious (possibly fatal) harm to an unborn baby if used
during pregnancy. Therefore, it is important to prevent pregnancy while taking this
medication. Other Side effects are similar to other angiotensin II receptor antagonists
and include tachycardia and bradycardia (fast or slow heartbeat), hypotension (low
blood pressure), edema (swelling of arms, legs, lips, tongue, or throat, the latter
leading to breathing problems), and allergic reactions.[3]
The ONTARGET Trial
Programme (The Ongoing Telmisartan Alone and in combination with Ramipril
Global Endpoint Trial) was one of the largest ARB clinical study programmes ever
undertaken.[4]
25,620 patients from 733 centers in 41 countries were randomized for
5.5 years of treatment of either Telmisartan the ACE inhibitor ramipril or a
combination of the two.
List of brand names of Telmisartan:
S.No. BRAND NAME
FORMULATIO
N
COMBINATION AVAILABLE
STRENGTH
Mg
MANUFACTURER
1 ARBITEL-AV Tab Atorvastatin 50 MICRO
2 CRESAR 80 Tab 80 CIPLA
3 CRESAR AM Tab Amlodipine 50 CIPLA
4 CRESLIP Tab Atorvastatin 50 CIPLA
5 INDITEL Tab 20 ZYDUS ALIDAC
6 INDITEL AM Tab Amlodipine 50 ZYDUS ALIDAC
244
7 LIPISAR Tab Atorvastatin 50 INTAS
8 LIPISAR 40 Tab Atorvastatin 50 INTAS
9 STAMLO-T Tab Amlodipine 50 DR. REDDY'S
LABS
10 TALY-40 Tab 40 GENETIC PHARMA
11 TALY-H Tab Hydrochlorothiazide 60 GENETIC PHARMA
12 TARGIT H Tab Hydrochlorothiazide 60 PFIZER
13 TARGIT Tab 20 PFIZER
14 TARGIT Tab 80 PFIZER
15 TELAST Tab 20 INTAS
16 TELDAY AV Tab Atorvastatin 50 TORRENT
17 TELEACT AM Tab Amlodipine 50 RANBAXY
18 TELIPRIL Tab Ramipril 50 SUN PHARMA
19 TELIPRIL Tab Ramipril 50 SUN PHARMA
20 TELISTA 20 Tab 20 LUPIN
21 TELISTA-H Tab Hydrochlorothiazide 100 LUPIN
22 TELMA Tab 20 ZOLTAN
(GLENMARK)
23 TELMIKIND Tab 80 MANKIND
24 TELMIKIND-
AM
Tab Amlodipine 50 MANKIND
25 TELMIKIND-H Tab Hydrochlorothiazide 100 MANKIND
26 TELMINORM Tab 20 IPCA
27 TELMISAT 40 H Tab Hydrochlorothiazide 60 BIOCON
28 TELMISAT 80 Tab Hydrochlorothiazide 100 BIOCON
29 TELMITOP HC Tab Hydrochlorothiazide 60 ELLIFE (ELDER)
30 TELMITOP Tab 20 ELLIFE (ELDER)
31 TELMITOP Tab 40 ELLIFE (ELDER)
32 TELPRES Tab 20 NICHOLAS
33 TELSAR Tab 40 UNISEARCH
245
Table 9.1
34 TELSARTAN Tab 20 DR. REDDY'S
LABS
35 TELSARTAN-R Tab Ramipril 50 DR. REDDY'S
LABS
36 TELSITE Tab 20 AVENTIS PHARMA
37 TETAN Tab 80 ALEMBIC
38 TELVAS H Tab Hydrochlorothiazide 100 ARISTO
39 TEMAX Tab 20 WOCKHARDT
40 TEMAX Tab 40 WOCKHARDT
41 TETAN Tab 80 ALEMBIC
42 TETAN AV Tab Atorvastatin 50 ALEMBIC
43 TSARTAN Tab 40 OCTANE BIOTECH
44 TSARTAN-H Tab Hydrochlorothiazide 60 OCTANE BIOTECH
45 VASIZITEL-H Tab Hydrochlorothiazide 60 SYNTONIC LIFE
SCIENCES
46 ZITELMI Tab 40 FDC
47 ZITELMI-H tab Hydrochlorothiazide 60 FDC
48 CRESAR TAB 20 CIPLA
246
9.2. LITREATURE SURVEY
Several analytical methods have been reported for the determination of
Telmisartan in pure drug, pharmaceutical dosage forms and in biological samples
using spcetrophotometry liquid chromatography, electro kinetic chromatography high
performance thin layer chromatography either in single or in combined forms.
Sujana K et al [5]
has developed and validated a simple, selective, precise and
stability indicating RP high Performance Liquid Chromatographic (HPCL) method of
an analysis of Telmisartan in pure and Pharmaceutical dosage form. The
chromatographic conditions comprised of a reversed phase C8 column (4.6 x 150mm,
3.5m, Make: XTerra), with a mobile phase Composed of Buffer and Methanol
(40:60v/v, Adjusted the pH to 3.0 with ortho Phosphoric acid).
Flow rate was 0.5 mL / min. Detection was carried out at 230 nm. The retention time
of Telmisartan was 2.6 min. The linear regression analysis data for the calibratin plots
showed good linear relationship in the concentration range 20-100. The values of
correclation coefficient, slope and intercept were, 0.9998, 2.326 and -6.708,
repectively. The method was successfully validated in accordance to
ICH guidelines acceptance criteria for specificity, linearity, precision, recovery, rugge
dnessand robustness. The drug undergoes degradation under acidic, basic, Peroxide a
nd thermal degradation conditions. All the peaks of degraded product were resol
ved from the active pharmaceutical ingredient with significantly different retent
ion time. As the method could effectively separate the drug from its degradation
product, it can be employed as a stability-indicating one.
R. Vijayamirtharaj et al [6]
has developed a RP-HPLC method for the
determination of Telmisartan and Atorvastatin calcium in bulk and in formulation
using UV detector. Selected mobile phase was a combination of Acetonitrile: Buffer
(0.01M Potassium dihydrogen phosphate) 65:35 PH 4.00 (adjusted with
Orthophosphoric acid) and the wavelength selected was 250nm. The flow rate was
kept at 2.0 ml/min, and the injection volume was 10µl. The separation was performed
at ambient temperature. Retention time of Telmisartan and Atorvastatin calcium was
found to be 3.72 and 6.14 minutes respectively. Linearity of the method was found to
be 319-480 µg/ml for Telmisartan and 86-130 µg/ml for Atorvastatin calcium. The
correlation co-efficient of Telmisartan was found to be 0.9998 and the correlation co-
efficient of Atorvastatin calcium was found to be 0.9999. Accuracy of the method
247
was determined through recovery studies by adding known quantities of standard drug
to the pre analyzed test solution and was found to be 98.92-100.02 for Telmisartan
and 99.93-100.96 for Atorvastatin calcium respectively The system suitability
parameters such as theoretical plates and tailing factor were found to be 6347, 1.652
and 9720, 1.394 respectively for Telmisartan and Atorvastatin calcium. This method
was validated according to ICH guidelines.
V.A. Patel et al [7]
has developed and validated thin layer liquid
chromatography (TLC) method for the simultaneous estimation of Telmisartan and
ramipril in a combined dosage form. Procedure does not require prior separation of
components from the sample. Telmisartan and Ramipril were determined by High
Performance Thin Layer chromatography method (HPTLC) in tablet dosage form.
The method was carried out in TLC Precoated silica gel on aluminum plate 60 F 254,
(10 cm ×10 cm, prewashed by methanol and activated at 60° C for 5 min prior to
chromatography). The solvent system was Acetone: Benzene: Ethyl acetate: Glacial
acetic acid in the proportion of 5:3:2:0.03, (v/v/v/v) with R Value for Telmisartan and
ramipril was 0.673 and 0.353 respectively. The linearity regression analysis for
calibration showed 0.996 and 0.998 for Telmisartan and ramipril with respect to peak
area and height in the concentration range of 100- 1800 ng/spot and 300-1800 ng/spot
respectively. The method developed can be used for routine analysis of drugs content
in tablet dosage form.
V. P. Kurade et al [8]
described a rapid high performance liquid
chromatographic method has for the estimation of ramipril and Telmisartan
simultaneously in combined dosage form. A Genesis C18 column having dimensions
of 4.6×250 mm and particle size of 5μm in isocratic mode, with mobile phase
containing a mixture of 0.01 M potassium dihydrogen phosphate buffer (adjusted to
pH 3.4 using orthophosphoric acid): methanol: acetonitrile (15:15:70 v/v/v) was used.
The mobile phase was pumped at a flow rate of 1.0 ml/min and the eluents were
monitored at 210 nm. The selected chromatographic conditions were found to
effectively separate ramipril (Rt: 3.68 min) and Telmisartan (Rt: 4.98 min) having a
resolution of 3.84. The method was validated in terms of linearity, accuracy,
precision, and specificity, limit of detection and limit of quantitation. Linearity for
ramipril and Telmisartan were found in the range of 3.5-6.5 μg/ml and 28.0-52.0
μg/ml, respectively. The percentage recoveries for ramipril and Telmisartan ranged
248
from 99.09-101.64% and 99.45-100.99%, respectively. The limit of detection and the
limit of quantitation for ramipril were found to be 0.5μg/ml and 1.5 μg/ml
respectively and for Telmisartan were found to be 1.5 μg/ml and 3.0 μg/ml,
respectively. The method was found to be robust and can be successfully used to
determine the drug content of marketed formulations.
SB Wankhede et al [9]
has described a validated reverse phase high
performance liquid chromatographic method for simultaneous estimation of
Telmisartan and hydrochlorothiazide in tablet formulation. Chromatography was
performed on a ODS Hypersil C18 (25 cm×4.6 mm I.D) column from thermo in
isocratic mode with mobile phase containing acetonitrile: 0.05 M KH2PO4 pH 3.0
(60:40). The flow rate was 1.0 ml/min and the eluent was monitored at 271 nm. The
selected chromatographic conditions were found to effectively separate Telmisartan
(RT- 5.19 min) and hydrochlorothiazide (RT- 2.97 min). Linearity for Telmisartan
and hydrochlorothiazide were found in the range of 4.1-20.48 µg/ml and 1.28-6.4
µg/ml, respectively. The proposed method was found to be accurate, precise,
reproducible and specific and can be used for simultaneous analysis of these drugs in
tablet formulation.
A. Gupta et al [10]
has developed and validated a simple, rapid, precise,
sensitive and reproducible reverse phase high performance liquid chromatographic
(RP-HPLC) method for determination of Telmisartan (TELM) in tablet dosage form.
Chromatographic separation was achieved on a 250 × 4.6 mm, 5μ, Waters symmetry
column in gradient mode, with mobile phase consisting of a mixture of solution (10
mM potassium dihydrogen phosphate, pH 3.5 ± 0.01): acetonitrile (64:40) was used.
The quantitation performed at flow rate of 1.0 mL/min at 230 nm and run time was 12
min. The analytical method was validated as per ICH guideline for linearity, accuracy,
precision, specificity, limit of detection, limit of quantification, robustness and
stability and method can be extended to the analysis of TELM in tablet formulations.
The relative standard deviation values for precision was less than 2%, and % recovery
was greater than 98% for TELM. The drug undergoes oxidative degradation, thermal
degradation and in alkali medium.
249
Pengfei Li et al [11]
has developed a rapid, selective and sensitive method for
the determination of the angiotensin II receptor antagonist, Telmisartan, in human
plasma. Telmisartan and the internal standard, diphenhydramine, were extracted from
plasma using diethyl ether–dichloromethane (60:40, v/v), and separated on a Zorbax
extend C18 column using methanol–10 mM ammonium acetate (85:15, v/v) adjusted
to pH 4.5 after mixing with formic acid as mobile phase. Detection was carried out by
multiple reactions monitoring on a Q-trap™ LC–MS/MS system with an ESI
interface. The assay was linear over the range 0.5–600.0 ng/ml with a limit of
quantitation of 0.5 ng/ml and a limit of detection of 0.05 ng/ml. Intra- and inter-day
precision were <6.7% and <8.1%, respectively, and the accuracy was in the range
88.9–111.0%. The assay was applied to a pharmacokinetic study of Telmisartan given
as a single oral dose (80 mg) to healthy volunteers.
Zhang H et al [12]
has been developed a rapid HPLC method using a
monolithic column with fluorescence detection has been for determination of
Telmisartan in human plasma. Sample preparation was done by protein precipitation
with acetonitrile and naproxen was used as IS. The compounds were detected by
fluorescence detection, using an excitation wavelength of 300 nm and emission
wavelength of 385 nm. Calibration curves of Telmisartan were linear in the range of
1-200 ng/mL. The assay was high throughput, sensitive and precise, and it was
successfully applied to a bioequivalence study of two formulations of Telmisartan.
S. V. Londhe et al [13]
developed a sensitive and reproducible HPLC method
for quantitative analysis of Telmisartan. The drug was separated from its degradation
products on a C 18 column at ambient temperature with methanol-water 80:20 (v / v ),
pH 4.0 (adjusted by addition of orthophosphoric acid), as mobile phase at a flow rate
of 1.0 mL min −1
. Under these conditions the retention time of Telmisartan was 4.85
± 0.05 min. Quantification on the basis of peak area was achieved by UV detection at
225 nm; calibration plots were linear in the concentration range 10–60 μg mL −1
.
When the method was applied to a pharmaceutical formulation there was no
chromatographic interference from tablet excipients. The method was validated for
precision, robustness, recovery, and limits of detection and quantification. The drug
was subjected to acidic and alkaline hydrolysis, and oxidising, dry heat, wet heat, and
250
photodegrading conditions. Because the method could effectively separate the drug
from its degradation products, it can be regarded as stability indicating.
L. R. Bhat et al [14]
has been developed a simple, selective, and precise
reverse phase high performance liquid chromatographic method for the simultaneous
determination of Telmisartan and hydrochlorothiazide from pharmaceutical
formulation. The mobile phase consisted of methanol and acetonitrile (70: 30 v/v) at a
flow rate of 1 mL/min and the wavelength of detection was 270 nm. Rabeprazole was
used as an internal standard. The retention times of Telmisartan, hydrochlorothiazide
and rabeprazole were 1.79 ± 0.01, 2.80 ± 0.01, and 3.19 ± 0.01 minutes, respectively.
The developed method was validated according to ICH guidelines. The proposed
method can be used for determination of these drugs in combined dosage forms.
251
9.3. EXPERIMENTAL
9.3.1. Instrumentation
Peak HPLC containing LC 20AT pump and variable wavelength
programmable UV-Visible detector and Rheodyne injector was employed for
investigation. The chromatographic analysis was performed on a Chromosil C18
column (250 mm × 4.6 mm, 5µm). Degassing of the mobile phase was done using a
Loba ultrasonic bath sonicator. A Denwar Analytical balance was used for weighing
the materials.
9.3.2. Chemicals and Solvents
The reference sample of Telmisartan (API) was obtained from Cipla, Mumbai.
The Formulation CRESAR (Telmisartan) was procured from the local market.
Methanol, Acetonitrile used was of HPLC grade and purchased from Merck
Specialities Private Limited, Mumbai, India. And orthophosphoric acid used was AR
grade purchased from local market.
9.3.3. The mobile phase
A mixture of Methanol: 0.1% orthophosphoric acid :Acetonitrile in the ratio of
40: 50:10 v/v/v was prepared and used as mobile phase.
9.3.4. The buffer solution
About 1.0 mL of orthophosphoric acid was diluted to 1000 mL with water and
filtered through 0.45μ nylon filter.
9.3.5. Standard solution of the drug
For analysis 100 ppm standard solution was prepared, required concentrations
were obtained from 100 ppm solution by appropriate dilution.
9.3.6. Sample (tablet) solution
The formulation tablets of Telmisartan (CRESAR - 20 mg) were crushed to
give finely powdered material. From the Powder prepared a 12 ppm solution in
mobile phase and then filtered through Ultipor N66 Nylon 6, 6 membrane sample filter
paper.
252
9.4. METHOD DEVELOPMENT
For developing the method (as described in Chapter 1 and 2), a systematic
study of the effect of various factors was undertaken by varying one parameter at a
time and keeping all other conditions constant. Method development consists of
selecting the appropriate wave length and choice of stationary and mobile phases. The
following studies were conducted for this purpose.
9.4.1. Detection wavelength
The spectrum of 10ppm solution of the Telmisartan in methanol was recorded
separately on UV spectrophotometer. The peak of maximum absorbance wavelength
was observed. The spectra of Telmisartan were showed maximum absorbance at
256nm.
9.4.2. Choice of stationary phase
Preliminary development trials have performed with octadecyl columns with
different types, configurations and from different manufacturers. Finally the expected
separation and peak shapes were obtained on Chromosil C18 (250 mm x 4.6 mm,
5μm) column.
9.4.3. Selection of the mobile phase
In order to get sharp peak, low tailing factor and base line separation of the
separation of the components, a number of experiments were carried out by varying
the composition of various solvents and flow rate. To have an ideal separation of the
drug under isocratic conditions, mixtures of solvents like methanol, water and
Acetonitrile with or without different buffers indifferent combinations were tested as
mobile phases on a Chromosil C18 column. A mixture of Methanol:0.1%
orthophosphoric acid :Acetonitrile in the ratio of 80:05:15 v/v/v was proved to be the
most suitable of all the combinations since the chromatographic peak obtained was
better defined and resolved and almost free from tailing.
253
9.4.4. Flow rate
Flow rates of the mobile phase were changed from 0.5 – 1.5 mL/min for
optimum separation. A minimum flow rate as well as minimum run time gives the
maximum saving on the usage of solvents. It was found from the experiments that 1.5
mL/min flow rate was ideal for the successful elution of the analyte.
9.4.5. Optimized chromatographic conditions
Chromatographic conditions as optimized above were shown in Table.9.2.
These optimized conditions were followed for the determination of Telmisartan in
bulk samples and in its Formulations. The chromatogram of standard (4ppm) shown
in Figure 9.B.
Mobile phase MEOH: 0.1 % OPA: ACN: 80:05:15 v/v/v
Pump mode Isocratic
Mobile phase PH
5.8
Diluent
Mobile phase
Column
chromosil C18 column (250 mm x 4.6 mm, 5μ)
Column Temp
Ambient
Wavelength
256 nm
Injection Volume
20 μl
Flow rate 1.5 mL/min
Run time 6 min
Retention Time 2.7 min
Table 9.2: Optimized chromatographic conditions for estimation of Telmisartan
254
Figure 9.B: Chromatogram of standard solution
255
9.5. VALIDATION OF THE PROPOSED METHOD
The proposed method was validated as per ICH guidelines (as described in
Chapter 1 and 2). The parameters studied for validation were specificity, linearity,
precision, accuracy, robustness, system suitability, limit of detection, limit of
quantification, and solution stability.
9.5.1. Specificity
The specificity of method was performed by comparing the chromatograms of
blank, standard and sample (Prepared from Formulation). It was found that there is no
interference due to excipients in the tablet formulation and also found good
correlation between the retention times of standard and sample. The specificity results
are shown in Table 9.3.
NAME OF THE SOLUTION Retention Time in Min
Blank NO PEAKS
Telmisartan 2.790
Table 9.3: Specificity study
9.5.2 Linearity
Linearity was performed by preparing mixed standard solutions of Telmisartan
at different concentration levels including working concentration mentioned in
experimental condition i.e. 12 ppm. Twenty micro liters of each concentration was
injected in duplicate into the HPLC system. The response was read at 256 nm and the
corresponding chromatograms were recorded. From these chromatograms, the mean
peak areas were calculated and linearity plots of concentration over the mean peak
areas were constructed individually. The regressions of the plots were computed by
least square regression method. Linearity results were presented in Table 9.4.
256
Table 9.4: Linearity results
Figure 9.C: On X axis concentration of sample, On Y axis peak area response
Level Concentration of Telmisartan in ppm Mean peak
area
Level -1 2 85426
Level -2 4 160991.9
Level -3 6 247225.6
Level -4 8 353178.8
Level -5 10 437519.9
Level -6 12 534746.3
Range: 2-12ppm
Slope
Intercept
Correlation coefficient
45459.12
-150323.5
0.999
-100000
0
100000
200000
300000
400000
500000
600000
0 5 10 15
Series1
Linear (Series1)
257
9.5.3. Precision
Precision is the degree of repeatability of an analytical method under normal
Operational conditions. Precision of the method was performed as intraday precision,
Inter day precision.
9.5.3.1. Intraday precision
To study the intraday precision, six replicate standard solutions (12ppm) of
Telmisartan were injected. The percent relative standard deviation (% RSD) was
calculated and it was found to be 0.039, which are well within the acceptable criteria
of not more than 2.0. Results of system precision studies are shown in Table 9.5.
SAMPLE
CONC(PPM) INJECTION
No.
PEAKS
AREA
R.S.D
(Acceptance
criteria ≤
2.0%)
Telmisartan
12
1 588925.5
0.039
2 588289.1
3 588470.8
4 588815.8
5 588548.2
6 588557.4
Table 9.5: System Precision (Intra Day)
9.5.3.2. Inter Day precision
To study the interday precision, six replicate standard solution of Telmisartan
was injected on third day of sample preparation. The percent relative standard
deviation (% RSD) was calculated and it was found to be 0.373, which are well within
the acceptable criteria of not more than 2.0. Results of system precision studies are
shown in Table 9.6.
258
SAMPLE
CONC(PPM) INJECTION
No.
PEAKS
AREA
R.S.D
(Acceptance
criteria ≤ 2.0%)
Telmisartan
12
1 586338.7
0.123
2 585740.8
3 582157.1
4 582307.0
5 581967.1
6 586231.3
Table 9.6: System Precision (Inter Day)
9.5.4. Accuracy
The accuracy of the method was determined by standard addition method. A
known amount of standard drug was added to the fixed amount of pre-analyzed tablet
solution. Percent recovery was calculated by comparing the area before and after the
addition of the standard drug. The standard addition method was performed at 50%,
100% and 150% level of 12ppm. The solutions were analyzed in triplicate at each
level as per the proposed method. The percent recovery and % RSD was calculated
and results are presented in Table 9.7. Satisfactory recoveries ranging from 99.0 to
102.0 were obtained by the proposed method. This indicates that the proposed method
was accurate.
259
Level Amount of
Telmisartan
spiked (ppm)
Amount of
Telmisartan
recovered(ppm)
% Recovery
%RSD
50 %
6
6
6
5.98
5.95
5.95
99.66
99.16
99.16
0.289
100%
8
8
8
7.98
7.94
7.93
99.75
99.75
99.12
0.370
150%
10
10
10
9.98
9.91
9.92
99.8
99.1
99.2
0.380
Mean % of
recovery 99.41
Mean
RSD =
0.346
Table 9.7: Percentage Recovery and % RSD
9.5.5. Robustness
The robustness study was performed by slight modification in flow rate of
Mobile phase, pH of the buffer and composition of the mobile phase. Telmisartan at 6
ppm concentration was analyzed under these changed experimental conditions. It was
observed that there were no marked changes in chromatograms, which demonstrated
that the developed method was robust in nature. The results of robustness study are
shown in Table 9.8.
260
Condition Mean area % assay % difference
Unaltered 590802.8 100.0 0.0
Flow rate at 1.4 mL/min
Flow rate at 1.6mL/min
585143.0
586252.8
99.04
99.22
0.96
0.72
Mobile phase:
MEOH: ACN: 0.1 % OPA
75% 20% 05%
85% 10 % 05%
589314.9
596219.5
99.74
100.9
0.26
0.10
pH of mobile phase at 5.6 592326.2 100.25 0.25
pH of mobile phase at 6.0 593572.1 100.4 0.4
Table 9.8: Robustness
9.5.6. System suitability
System suitability was studied under each validation parameters by injecting
six replicates of the standard solution 2 ppm). The results obtained were within
acceptable limits (Tailing factor ≤2 and Theoretical plate’s ≥2000) and are
represented in Table 9.9. Thus, the system meets suitable criteria.
Parameter Tailing factor Theoretical plates
Specificity study 1.71 6638
Linearity study 1.20 7694.97
Precision study 1.71 6705.46
Table 9.9: System Suitability
261
9.5.7. Limit of detection and Limit of quantification
Limit of detection (LOD) is defined as the lowest concentration of analyte that
gives a detectable response. Limit of quantification (LOQ) is defined as the lowest
Concentration that can be quantified reliably with a specified level of accuracy and
Precision. For this sample was dissolved by using Mobile Phase and injected until
peak was disappeared. After 15ng/ml dilution, Peak was not clearly observed. So it
confirms that 15ng is limit of Detection and 50ng dilution is Limit of Quantification.
For this study six replicates of the analyte at lowest concentration were Measured and
quantified. The LOD and LOQ of Telmisartan are given in Table 9.10.
Table 9.10: LOQ and LOD
Formulation:
The proposed method has been applied to the assay of commercial
tablets containing Telmisartan. Sample was analyzed for five times after
extracting the drug as mentioned in assay sample preparation of the
experimental section. After analysis test result assay of Telmisartan in Tablet
is18.57% and is very close to the labeled amount.
Parameter Measured volume
Limit of Quantification 50ng
Limit of Detection 15ng
262
9.6. DISCUSSION ON THE RESULT
To optimize the RP-HPLC parameters, several mobile phase
compositions were tried. A satisfactory separation and good peak symmetry was
found in a mixture of Acetonitril: Methanol: Acetonitrile: 0.1%OPA in the ratio of
80:15:05 v/v and 1.5 mL/min flow rate proved to be better than the other mixtures in
terms of resolution and peak shape. The optimum wavelength for detection was set at
256nm at which much better detector responses for drug was obtained. As it was
shown in Figure 9.B the retention times were 2.7 min for Telmisartan. The number of
theoretical plates was found to be 6638.20, which indicates efficient performance of
the column. A system suitability test was applied to representative chromatograms for
various parameters. The results obtained were within acceptable limits and are
represented in Table 9.9. Thus, the system meets suitable criteria.
The calibration curve was obtained for a series of concentration in the
range of 2-12µg/ml and it was found to be linear. Seven points graphs was
constructed covering a concentration range 2-12µg/ml. The standard deviation of the
slope and intercept were low. The data of regression analysis of the calibration curves
are shown in Table. Calibration curve found to be linear with r2=0.999, Intercept (-
150323.5) and Slope (45459.12) respectively. The results obtained were within
acceptable limits where capacity factor >2.0, tailing factor ≤2.0 and theoretical plates
>2000 13. In all cases, the relative standard deviation (R.S.D) for the analytic peak
area for two consecutive injections was < 2.0%. The data of regression analysis of the
calibration curve was shown in Table 9.4.
Precision was evaluated by carrying out six independent sample preparation
of a single lot of formulation. Low values of standard deviation denoted very good
repeatability of the measurement. Thus it was showing that the equipment used for the
study was correct and hence the developed analytical method is highly repetitive.
RSD of intraday precision was found to 0.039. For the interday precision a study
carried out on consecutive days indicated a RSD of 0.373. This indicates good method
precision. Results are shown in Table 9.5, 9.6.
Standard addition method at 50%, 100% and 150% to the proposed
HPLC method is carried out to find the Accuracy of the Telmisartan. The results
showed good recoveries ranging from 99.00 to 101.45%. The mean recovery data
263
obtained for each level as well as for all levels combined (Table 9.7) were within
2.0% of the label claim for the active substance with an R.S.D. < 2.0%, which
satisfied the acceptance criteria set for the study.
The proposed method has been applied to the assay of commercial
tablets containing Telmisartan. Sample was analyzed for five times after extracting
the drug as mentioned in assay sample preparation of the experimental section. The
results (18.57%) presented good agreement with the labeled content. Low values of
standard deviation denoted very good repeatability of the measurement.
The statistical evaluation of the proposed method was revealed its
good linearity, reproducibility and its validation for different parameters and let us to
the conclusion that it could be used for the rapid and reliable determination of
Telmisartan in tablet formulation. All these factors lead to the conclusion that the
proposed method is accurate, precise, simple, sensitive and rapid and can be applied
successfully for the estimation of Telmisartan in bulk and in pharmaceutical
formulations without interference and with good sensitivity.
264
9.7. BIBILOGRAPHY
1. Pritor prescribing information.
2. Benson, S. C.; Pershadsingh, H.; Ho, C.; Chittiboyina, A.; Desai, P.; Pravenec,
M.; Qi, N.; Wang, J. et al; "Identification of Telmisartan as a Unique
Angiotensin II Receptor Antagonist with Selective PPAR -Modulating
Activity". Hypertension 2004, 43 (5): 993.
3. Drugs.com: Micardis.
4. Ontarget, I.; Yusuf, S.; Teo, K.; Pogue, J.; Dyal, L.; Copland, I.; Schumacher,
H; Dagenais, G. et al; "Telmisartan, Ramipril, or Both in Patients at High Risk
for Vascular Events"; New England Journal of Medicine; 2008, 358 (15):
1547.
5. ―Stability indicating rphplc method for the determination of Telmisartan in
pure and pharmaceutical formulation‖; International Journal of Pharmacy and
Pharmaceutical Sciences; 2011; 3(2).
6. ―Development and validation of RP-HPLC method for the simultaneous
estimation of Telmisartan and atorvastatin calcium in tablet dosage forms‖;
Pharmacie Globale (IJCP); 2010, 4 (03).
7. V.A. Patel et al; International Journal on Pharmaceutical and Biological
Research; 2010; 1(1): 18-24.
8. ―RP-HPLC Estimation of Ramipril and Telmisartan in Tablets‖; Indian J
Pharm Sci; 2009; 71(2): 148–151.
9. ―RP-HPLC method for simultaneous estimation of Telmisartan and
hydrochlorothiazide in tablet dosage form‖; Indian Journal of Pharmaceutical
science; 2007; 69(2): 298-300.
10. ―Determination of Telmisartan and forced degradation behavior by RP-HPLC
in tablet dosage form‖; Journal of Pharmacy Research; 2011; 4(4): 1270-1273.
11. ―Determination of Telmisartan in human plasma by liquid chromatography–
tandem mass spectrometry‖; Journal of Chromatography B; 2005; 828(1-2),
15, 126-129.
12. ―Rapid determination of Telmisartan in human plasma by HPLC using a
monolithic column with fluorescence detection and its application to a
bioequivalence study‖; 2009; 877(29): 3729-33.
265
13. ―Stability-indicating RP-HPLC method for analysis of Telmisartan in the bulk
drug and in formulations‖; Acta Chromatographica,; 2010; 22(4): 539-548.
14. ―Validated RP-HPLC method for simultaneous determination of Telmisartan
and hydrochlorothiazide in pharmaceutical formulation‖; J. Liq. Chromatogr.
Relat. Technol; 2007; 30(17-20): 3059-3067.
15. ―Q2A: Text on Validation of Analytical Procedures‖; International Conference
on Harmonization; Federal Register; 1995; 60(40): 11260–11262.
16. ―Q2B: Validation of Analytical Procedures: Methodology; Availability‖;
International Conference on Harmonization; Federal Register; 1997; 62(96):
27463–27467.
17. "Analytical Procedures and Methods Validation: Chemistry, Manufacturing
and Controls Documentation; Availability"; FDA; Federal Register (Notices);
2000; 65(169): 52776 – 52777.
18. www.fda.gov/cder/guidance/cmc3.pdf.
19. USP 25–NF 20; ―Validation of Compendial Methods Section (1225) (United
States Pharmacopeal Convention, Rockville, Maryland, USA, 2002)‖; 2256.
20. G.A. Shabir; ―Validation of HPLC Chromatography Methods for
Pharmaceutical Analysis‖; Understanding the Differences and Similarities
between Validation Requirements of FDA, the US Pharmacopeia and the ICH;
J. Chromatogr. A; 2003; 987(1-2): 57-66.
21. C.E. Wood; "Medicare Program; Changes to the Hospital Outpatient
Prospective"; Med. J. Aust; 1996; 165: 510–514.
22. A. Prentice; "Medical Management of Menorrhagia": Br. Med. J; 1999; 319,
1343–1345.
23. D.T. Baired and A.F. Glasier; "Hormonal Contraception"; New Engl. J. Med;
1993; 328: 1543–1549.
24. P.E. Belchetz; "Hormonal Treatment of Postmenopausal Women"; New Engl.
J. Med; 1994; 330: 1062–1071.
25. International Conference on Harmonization (ICH) of Technical Requirements
for the Registration of Pharmaceuticals for Human Use; ―Validation of
analytical procedures: definitions and terminology‖; Geneva (1996).
266
26. U.S. FDA; Title 21 of the U.S. Code of Federal Regulations:
21 CFR 211—Current good manufacturing practice for finished
pharmaceuticals.
27. U.S. FDA - Guidance for Industry (draft) Analytical Procedures and Methods
Validation: Chemistry, Manufacturing, and Controls and Documentation,
2000.
28. ISO/IEC 17025, General requirements for the competence of testing and
calibration laboratories, 2005.
29. International Conference on Harmonization (ICH) of Technical Requirements
for the Registration of Pharmaceuticals for Human Use, Validation of
analytical procedures: Methodology, adopted in 1996, Geneva.
30. U.S. EPA, Guidance for methods development and methods validation for the
Resource Conservation and Recovery Act (RCRA) Program, Washington,
D.C. (1995).http://www.epa.gov/sw-846/pdfs/methdev.pdf.
31. General Chapter 1225, Validation of compendial methods, United States
Pharmacopeia 30, National Formulary 25, Rockville, Md., USA, The United
States Pharmacopeial Convention, Inc., (2007).
32. U.S. FDA - Guidance for Industry, Bioanalytical Method Validatio.n
33. G. C. Hokanson, A life cycle approach to the validation of analytical methods
during pharmaceutical product development, Part I: The initial validation
process, Pharm. Tech., Sept. 1994, pp. 118–130.