chapter 4 noise measurement and instrumentation
TRANSCRIPT
1
Noise Measurement and Instrumentation
Topics:• Introduction• Sound Measurement• Sound Intensity Mapping & source
identification• Instrumentation• Sound Power• Sound Power Measurement
2
IntroductionSound is a sensation of acoustic waves (disturbance/pressure fluctuations setup in a medium).
Unpleasant, unwanted, disturbing sound is generally treated as Noise and is a highly subjective feeling.
3
Introduction
4
Noise Measurement and Instrumentation
Topics:• Introduction• Sound Measurement• Sound Intensity• Instrumentation• Intensity Spectrum Level• Sound Power• Sound Power Measurement
5
Sound Measurement
• Provides definite quantities that describe and rate sound.
• Permit precise, scientific analysis of annoying sound (objective means for comparison).
• Help estimate Damage to Hearing. • Powerful diagnostic tool for noise reduction
program: Airports, Factories, Homes, Recording studios, Highways, etc.
6
Sound MeasurementQuantifying Sound
Root Mean Square Value (RMS) of Sound Pressure
Mean energy associated with sound waves is its fundamental feature
energy is proportional to square of amplitude
12
2
0
1 [ ( )]T
p p t dtT⎡ ⎤
= ⎢ ⎥⎣ ⎦∫
0.707p a=
Acoustic Variables: Pressure and Particle Velocity
(for harmonic sound waves)
7
Sound Measurement
Range of RMS pressure fluctuations that a human ear can detect extends from
0.00002 N/m2 (threshold of hearing)
to
20 N/m2 (sensation of pain) 1000000 times larger
Atmospheric Pressure is 105N/m2
so the peak pressure associated with loudest sound is 5000 times smaller than atmospheric pressure
The large range of associated pressure is one of the reasons we need alternate scale
Range of Pressure
8
Noise Measurement and Instrumentation
Topics:• Introduction• Sound Measurement• Sound Intensity• Instrumentation• Intensity Spectrum Level• Sound Power• Sound Power Measurement
9
Sound Intensity
Sound Intensity
10
Sound IntensityA plane progressive sound wave traveling in a medium (say along a tube) contains energy and
Rate of transfer of energy per unit cross-sectional area is defined as Sound Intensity
0
1 T
I p u dtT
= ∫2
0
PIcρ
=
1010ref
IIL LogI
=
21 01
10 10 20
/( )20 102 5 (2 5) /( )
p cpSPL Log dB Log dBe e c
ρρ
= =− −
12 12
10 10 1012 2 20 0
10 1010 10 1010 (2 5) /( ) (2 5) /( )ref
I ISPL Log dB Log Loge c I e cρ ρ
− −
−= = +− −
For air, ρ0c ≈ 415Ns/m3 so that 0.16 dBSPL IL= +
Holds true also for spherical waves far away from source
11
Sound Intensity Measurement
12
Sound Intensity Measurement
13
Noise Measurement and Instrumentation
Topics:• Introduction• Sound Measurement• Sound Intensity• Instrumentation• Intensity Spectrum Level• Sound Power• Sound Power Measurement
14
Instrumentation
Instruments for analysing Noise
Constant Bandwidth Devices
Proportional Bandwidth Devices
2U
L
ff
= c U Lf f f≈
Absolute Bandwidth = fU - fL = fL
% Relative Bandwidth = (fU-fL / fc) = 70.7%
If we divide each octave into three geometrically equal subsections, i.e.,
1/32U
L
ff
=
These bands are thus called 1/3rd octave bands with % relative bandwidth of 23.1%
1/102U
L
ff
=For 1/10th Octave filters, 5.1% relative bandwidth
2nU
L
ff
=
n=1 for octave,
n=1/3 for 1/3rd octave
15
Octave Band Filters
Octave and 1/3rd Octave band filters
mostly to analyse relatively smooth varying spectra
If tones are present,
1/10th Octave or Narrow-band filter be used
16
Instrumentation- Microphones
Measurement transducer to measure noise
• Condenser Microphone• Dynamic Microphone• Ceramic Microphone
Condenser Microphone
17
Instrumentation- Condenser Microphone
• Can be used in extreme condition
• Insensitive to vibrations
• Very expensive • Sensitive to humidity &
moisture
•Measurements range can be from 0.01 Hz to 140 KHz•Dynamic range up to 140 dB
18
Instrumentation- Dynamic Microphone
• Generation of the electrical signal in a moving coil in a magnetic field. The moving coil is connected to the diaphragm that deflects under pressure fluctuations of the sound.
• Excellent sensitivity characteristics.• Relatively insensitive to extreme variation in the humidity.• Cheaper than condenser microphone.
But:-Can not be used in places where strong magnetic fields are present.Lower frequency response than condenser microphone.
19
Instrumentation- Ceramic Microphone
• Sensing element is the piezoelectric crystal.• High frequency response.• High dynamic range.• Very cheap & can often be custom built.• Common for research application as size is also small.
But:-• These are sensitive to the vibration and pressure fluctuation.
Ceramic Microphone is also called as “Piezoelectric” microphone
20
Noise Measurement and Instrumentation
Topics:• Introduction• Sound Measurement• Sound Intensity• Instrumentation• Intensity Spectrum Level• Sound Power• Sound Power Measurement
21
Intensity Spectrum Level
DeciBel measure of ℑ is the Intensity Spectrum Level (ISL)
.110logref
HzISLI
⎛ ⎞ℑ= ⎜ ⎟⎜ ⎟
⎝ ⎠If the intensity is constant over the frequency bandwidth w (= f2- f1),
then total intensity is just I= ℑ w and
and Intensity Level for the band is
1 .1
wI HzHz
= ℑ×
10logIL ISL w= +If the ISL has variation within the frequency band (w), each band is subdivided into smaller bands so that in each band ISL changes by no more than 1-2dB
22
Intensity Spectrum Level
IL is calculated and converted to Intensities Ii and then total intensity level ILtotal is
10logi
itotal
ref
IIL
I
⎡ ⎤⎛ ⎞⎢ ⎥⎜ ⎟⎝ ⎠⎢ ⎥=⎢ ⎥⎢ ⎥⎣ ⎦
∑10logi i iIL ISL w= +
as SPL and IL are numerically same, 10logSPL PSL w= +
PSL (Pressure Spectrum Level) is defined over a 1Hz interval – so the SPL of a tone is same as its PSL
101010log 10
iIL
totali
IL⎡ ⎤
= ⎢ ⎥⎣ ⎦∑10log
ii
totalref
IIL
I
⎡ ⎤⎛ ⎞⎢ ⎥⎜ ⎟⎝ ⎠⎢ ⎥=⎢ ⎥⎢ ⎥⎣ ⎦
∑ Can be written as
Thus, when intensity level in each band is known, total intensity level can be estimated
23
Noise Measurement and Instrumentation
Topics:• Introduction• Sound Measurement• Sound Intensity• Instrumentation• Intensity Spectrum Level• Sound Power• Sound Power Measurement
24
Sound Power
Intensity : Average Rate of energy transfer per unit area
22 W/m
4WI
rπ=
22 2
0
4 4 Watt pW r I rc
π πρ
= =
Sound Power Level: 1010 logref
WSWLW
=
Reference Power Wref =10-12 Watt
dB
Peak Power output:
Female voice – 0.002W, Male voice – 0.004W, A soft whisper – 10-9W, An average shout – 0.001W Large orchestra – 10-70W, Large Jet at takeoff – 100,000W
15,000,000 speakers speaking simultaneously generate 1HP
25
Noise Measurement and Instrumentation
Topics:• Introduction• Sound Measurement• Sound Intensity• Instrumentation• Intensity Spectrum Level• Sound Power• Sound Power Measurement
26
From pressure measurement for varioussound fields
1.Free field technique
2.Reverberant field technique
3.Semi reverberant technique
Estimation of sound power
27
Free Field Condition Diffuse Field
I=0
Uniform sound energy density
2828
• Used for measuring the sound power of any m/cproducing soundo that contains prominent discreet frequency component or narrow
band spectra.
• Can also be used when ‘directional nature’ of the soundradiation are required.
Free Field techniques
Anechoic Chamber
29
Sound Power Measurement• Free field technique
– Sound power of machines which is having discrete frequency spectrum
– Carried out in an anechoic chamber
• Test procedure
– Making no. of measurements on an imaginary surface of hemisphere/sphere with machine at centre.
– No. of microphone positions required depends upon degree of directionality of sound field
– Sound power once average sound pressure level is established then it is integrated over the surface area .
30
• Test procedure involves making number of SPLmeasurements on the surface of an imaginary averagesphere/hemisphere distance from the source is found.
Continues…
Free Field techniques
Finding sound power (ISO 3745)
The measurement can be made in
a large anechoic chamber or can
made in a free field above a
reflecting plane.
31
24I rπΠ = ×2
12 1210log 10log 10log 4 10log10 10I rπ− −
Π = +
11 20logIL L rΠ = + +
20log 11PL L r dBΠ = + + _ I Pwith L L≈
20log 8IL L r dBΠ = + +
For hemispherical surface
Sound power of the source is then computed using the following equations
Free Field techniques continues…
32
Sound Power Measurement
• Reverberant field technique
– Carried out in a reverberation room– Complete diffused sound field sound pressure is independent of
distance from the source
• Sound power can be calculated from
i) The acoustic characteristics of the roomii) The sound pressure level in the room
• Applicability
– Source which does not produce discrete frequencies – And narrow spectrum
33
Reverberant Field Techniques
Reverberant Chamber
• In a completely reverberant (diffused) field,o sound waves are continuously being
reflected from bounding surfaces.o sound pressure field is essentially
independent of distance from source.o the flow of the energy is uniform in all
directions and the sound energy densityis uniform.
• The sound power of a source is reverberant sound field can beobtained from
o The acoustic characteristics of the room ando Sound pressure level in the room.
34
Sound Power Measurement
• Principle– Consider directional sound source of total power
– Sound intensity because of direct field
where and
• Average sound absorption coefficient of the room is
∏
2 20/ / 4p c Q rθ θρ π= ∏
/ sQ I Iθ θ= 20; /I p cθ θ ρ= 2/ 4sI rπ= ∏
1 1 2 2
1 2
............
n navg
n
S S SS S S
α α αα + + +=
+ + +
α s are absorption coefficients of different materials
S are surface area of different absorbing materials in the chamber
35
Sound Power Measurement
• The energy which is reflected back is
• Upon making required substitutions
– Sound power level is given by
(1 )rev avgα∏ =∏ −
10 2
410log { }4pQL L
r Rθ
π∏ = − +
pL Is the sound pressure level in chamber
R is room the constant given by /(1 )avg avgR Sα α= −
3636
Consider a directional source (total sound power Π) placed inthe centre of the reverberation room. The contribution of thedirect (un reflected) field to the sound intensity in the room is
2
20 4p Q
C rθ θ
ρ πΠ
= ;s
IQIθ
θ =2
0
;p
ICθ
θ ρ= 2_
4sand IrπΠ
=Where
Considerationso the source does not produce any prominent discrete frequency
component or narrow band spectrao If such sound field exits, a rotating diffuser should be used ando The lowest discrete frequency which can be reliably is measured
about 200 Hz. The free field techniques is recommended fordiscrete noise source bellow 200 Hz .
Reverberant Field Techniques
3737
The sound field produced by the reflected sound has now got tobe determined. For the purpose we needed to introduce theconcept sound absorption
Sound Absorption Coeff
Sound Transmission CoeffT
i
II
ς =
a
i
II
α =
i R T DΠ = Π +Π +Π
Absorbed sound intensity
Incident sound intensity
Reflected sound energy
Transmitted sound energy
Dissipated within the surface.
Where
a
i
R
T
D
IIΠΠΠ
Reverberant Field Techniques
3838
Now all the energy which is not reflected is absorbed. (its either
transmitted through material or dissipated in the material as
heat via flow constriction and vibrational motion of the fibers in
the material. )
Hence the absorbed sound energy is given equation .
A D TΠ = Π +Π
Open Window has α = ?
Reverberant Field Techniques
3939
• When sound field is neither free nor complete diffuse.
• Use calibrated sound source with known power spectrum.
• No. of microphones position
Q N1 202 124 68 3
•When sound field is neither free nor complete diffuse.
Semi-Reverberant Field Techniques
4040
Let
Then
o For semi reverberant field, small αt of room, room size belarge so that measurement are made in free field.
o Make no. of measurements (Lpi) or spherical on hemispherical area at radius r1.
Semi-Reverberant Field Techniques
4141
o For reverberant field measurement b/g noise < 10dB of
the sound source level.
o No valid measurements if b/g noise difference < 4 dB
o When large room size can not exist. Near field
measurements may be necessary.
o Test surface should be within 1m from radiating surface.
o After averaging out SPL measurements
S : Surface area of measuringsurface
Semi-Reverberant Field Techniques
4242
Room type V/S (m)Room without highly
reflecting surface 20 - 50 50 - 90 90 - 3000 >3000
Room with highly reflecting surface 50-100 100 - 200 200-600 >600
∆ 3 2 1 0
A correction factor to account for absorption andreflection from nearby areas.
Semi-Reverberant Field Techniques
43
Noise Metrics
44
Sound level measurements
• IEC International Standard 651 ”Sound Level Meters”
• Tolerances per frequency band defined for 4 classes of accuracy– Type 0: precision laboratory use– Type 1: general purpose– Type 2: low price– Type 3: not used in practice (too wide
tolerances)
45
Calibration of Sound Level meters
46
The acoustic signal is a very small compared to
the atmospheric pressure
Sound Pressure Level dB scale
And the pressure amplitude varies over a very wide range Sound Pressure Level dB
scale.
47
Sound Pressure Level dB scale
48
The acoustic pressure is very small compared to the atmospheric pressure
SPL (dB scale)
What sound level meter will do, pick up N samples over a period T
2
0
1 ( )T
rmsp p t dtT
= ∫
2
1
1 N
rms ip pN
= ∑
49
How much this N should be ? Most of sound level meters offer two options
• Fast averaging • 125 m sec of averaging
time (slow)• Fast varying signals• Impulse Averaging (I)• 35 ms of averaging time,
for impacts
• Slow Averaging• Approx 1 second of time
averaging • Slow varying signals• When we are interested
in representative values
SPL (dB scale)
50
51
Equivalent Level Leq
52
Leq ValueEquivalent constant level that wouldgive the same sound exposure
Why Integrating sound level meters?
][)(1100
2
2
, dBdtp
tpT
LogLT
refTeq ⎟
⎟⎠
⎞⎜⎜⎝
⎛= ∫
][10110 10, dBt
TLogL
jL
jTeq ⎟⎟⎠
⎞⎜⎜⎝
⎛= ∑
53
Sound Intensity
54
Sound Intensity
Instantaneous sound intensity
I r t p r t u r t( , ) ( , ) ( , )=
I rT
p r t u r t dtT
( ) ( , ) ( , )= ∫1
0
We are normally interested in the time average of the intensity, which gives the “active” intensity, corresponding to a net transport of sound energy
_ on the variable represents it a vector quantity
55
Measurement of sound intensity
The pressure is approximated by the average of the two pressure measurements
2)()()( tptptp BA +
=
56
Measurement of Sound Intensity
tu
rp r
∂∂
ρ∂∂
0−=
The zero mean flow momentum equation in the r direction
gives
u t pr
dr
p p dr
t
B A
t
( ) ( ( ) ( ))≈ − = − −−∞ −∞∫ ∫
1 1
0 0ρτ
ρτ τ τ
ΔΔ Δ
approximated by
u t pr
dr
t
( ) = −−∞∫
1
0ρ∂∂
τ
57
The direct method
dtdpptptpTr
It
AB
T
BAr ∫∫∞−
−+Δ
−= τττρ
))()(())()((12
1
00
58
The indirect or FFT method
59
Approximate measure of Sound Intensity
Intensity measured in dB with reference as 10-12 W/ m2
For a localized source general, intensity is a directional quantity
In order to capture this directional effect, special intensity probes are used which measure the correlated signal from two microphones. Aligned in a directional line.
6060
Source Intensity ( I )Decibel intensity
level
Multiple of TOH
intensityThreshold of Hearing (TOH) ITOH = 10−12 W/m2 0 dB 100
Rustling leaves 10−11 W/m2 10 dB 101
Whisper 10−10 W/m2 20 dB 102
Normal conversation 10−6 W/m2 60 dB 106
Busy street traffic 10−5 W/m2 70 dB 107
Vacuum cleaner 10−4 W/m2 80 dB 108
Large orchestra 6.3*10−3 W/m2 98 dB 109.8
iPod at maximum volume level 10−2 W/m2 100 dB 1010
Front rows of a rock concert 10−1 W/m2 110 dB 1011
Threshold of pain 101 W/m2 130 dB 1013
Military jet takeoff 102 W/m2 140 dB 1014
Instant perforation of eardrum 104 W/m2 160 dB 1016
Common sounds with estimates of intensity and decibel level22
TOH10 m
Wmeterwattsin measured is
areaunit powerintensity wherelog10levelintensity Decibel ==⎟
⎟⎠
⎞⎜⎜⎝
⎛= I
II
<http://www.glenbrook.k12.il.us/gbssci/Phys/Class/sound/u11l2b.html>
61
Filtering and Weighting Filters
6262
Frequency components present in a general noise source
Sum of 3 harmonics (based on http://zone.ni.com/cms/images/devzone/tut/a/8c34be30580.gif)
Nice demo to listen to Fourier series harmonics: http://www.jhu.edu/~signals/listen-new/listen-newindex.htm
63
Filtering
64
Weighting Filters
Source: Sound and Vibration Book, MWL, KTH, Sweden
65
Weighting Filters
66
Human hearing frequency response
For subjective responses in special cases there are B-, C- and D-weighting curves•very high or low level•special noise, e.g., of aircraft
A-weighting curve
67
Octave Analysis
68
Octave Analysis
220 Hz 440 Hz 880 Hz
A AA
• Analysis performed through a parallel bank of bandpass filters
• One octave corresponds to the doubling of the frequency
• Reference frequency is 1 kHz (audio domain)
69
Octave AnalysisOctave analysis gives log-spaced frequency information.
Similar to human perception of sound1/1, 1/3, 1/12, and 1/24 octave analysisFFT gives linearly-spaced frequency information.
Applications
•noise emissions testing
•acoustic intensity measurement
•sound power measurement
•audio equalization
70
71
72
Source Localizationn
73
Localization: Beamforming
),(1 θωY
),(2 θωY)(1 ωF
)(2 ωF
)(ωmF ),( θωmY
)(ωMF),( θωMY
)(ωS
Σ ),( θωZ
θcosmd
md
θ
74
64 Microphone Array
75
Beamforming Data model:• Microphone signals are delayed versions of S(ω)
Stack all microphone signals in a vector
d is `steering vector’
• Output signal Z(ω,θ) is
)]([][ θτ mm ksky −=s
mm f
cd θθτ cos)( =
[ ]Tjj Mee )()(21),( θωτθωτθω −−= Kd
∑=
⋅==M
m
Hmm YFZ
1
* ),()(),()(),( θωωθωωθω YF
)(.),( )( ωθω θωτ SeY mjm
−=
)().,(),( ωθωθω SdY =
76
Beamforming
• Spatial directivity pattern: `transfer function’ for source at angle θ
• Fixed Beamforming– Delay-and-sum beamforming– Weighted-sum beamforming– Near-field beamforming
∑=
− ⋅===M
m
Hjm
meFS
ZH1
)(* ),()()()(),(),( θωωω
ωθωθω θωτ dF
77
Delay-and-sum beamforming• M=5 microphones
• d=3 cm inter-microphone distance
• ψ=60° steering angle
• fs=5 kHz sampling frequency
-20
-10
0
90
270
180 0
Spatial directivity pattern for f=5000 Hz
78
Weighted-Sum beamforming• Sensor-dependent complex weight + delay• Weights added to allow for better beam shaping
ψcos)1( dm−
Σ d
d2Δ
mΔ
1Δ
ψ
1w
2w
mw
∑=
Δ+=M
mmmm kywkz
1
][.][
79
Noise Source Location in an Engine
Fillip et al (2007)