chapter 4 accelerometer testing - information and...

29
86 Chapter 4 Accelerometer Testing 4.1 Introduction The accelerometer designed and realized so far is intended for an aerospace application. Detailed testing and analysis needs to be conducted to qualify the product for the end use and establish the suitability of the product for the intended use. These tests can take many different forms including chip level probing for electrical characterization, sensitivity estimation, bias stability, performance evaluation under thermal and dynamic environments. Details of performance testing procedure of commercially available accelerometers are presented by number of researchers [47-49]. MEMS pendulous accelerometers have already demonstrated good performance in automobile and other commercial applications. The challenge of using of this technology in aerospace inertial navigation is about, significantly improving the bias stability, cross-axis sensitivity, temperature sensitivity and detailed measured performance demonstration for the intended application. The accelerometer was tested using procedure and data analysis methods similar to the practices in defence departments which are

Upload: lamque

Post on 04-Apr-2018

219 views

Category:

Documents


6 download

TRANSCRIPT

Page 1: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

86

Chapter 4

Accelerometer Testing

4.1 Introduction

The accelerometer designed and realized so far is intended for an

aerospace application. Detailed testing and analysis needs to be

conducted to qualify the product for the end use and establish the

suitability of the product for the intended use. These tests can take many

different forms including chip level probing for electrical characterization,

sensitivity estimation, bias stability, performance evaluation under

thermal and dynamic environments. Details of performance testing

procedure of commercially available accelerometers are presented by

number of researchers [47-49].

MEMS pendulous accelerometers have already demonstrated good

performance in automobile and other commercial applications. The

challenge of using of this technology in aerospace inertial navigation is

about, significantly improving the bias stability, cross-axis sensitivity,

temperature sensitivity and detailed measured performance

demonstration for the intended application.

The accelerometer was tested using procedure and data analysis

methods similar to the practices in defence departments which are

Page 2: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

87

broadly adopted from IEEE STD 337-1972 [50]. This describes the test

procedures for linear, single axis pendulous analog, torque balance

accelerometer.

The test procedure consists of observing the output of the test

device to input acceleration using Earth‟s gravitation field or an external

excitation source.

The objective of the test is to characterize the accelerometer

Bias stability

Linearity

Hysteresis

Cross- axis sensitivity

Temperature sensitivity

Bandwidth.

Shock response

It was presumed that physical sources of the errors described in the

reference [50] remain valid for the micromachined accelerometer also. In

both micromachined and conventional accelerometers, the temperature

sensitivity of performance is an important parameter. However, due to

the small scale of MEMS device, temperature is expected to play a larger

role.

The error model components selected for micromachined

accelerometers are bias, scale factor, non-linearity, cross-axis sensitivity

and temperature. The effect of misalignment of the input axis is not

investigated.

Page 3: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

88

The equation relating different error model components to the final

overall error expressed in % of FSO is as follows.

Average output = E1 + E2 + E3 + E4 x ∆T + E5.

E1 – Bias drift

E2 – Non linearity

E3 – Cross- axis sensitivity

E4 – Temperature sensitivity of bias drift

E5 – Misalignment of input axis.

∆T – change in device temperature during testing.

4.2 Chip level probing

The primary aim of the chip level probing is to validate the wafer

fabrication process and to certify the device functionality for further

packaging and integration operations.

Fig 4.1 Test set-up for electrical probing.

A standard Cascade make probe station is integrated with Agilent LCR

meter and is used for measuring the chip level electrical parameters. This

Page 4: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

89

set-up can measure capacitance from 10 atto Farads to 10 Farads with

an accuracy of 0.05%.

Ten number of accelerometers are selected for probing to verify the

consistency of the results. The average nominal capacitance values of

Co1 (capacitance between top electrode and proof-mass) and Co2

(capacitance between bottom electrode and proof-mass) are tabulated in

table - 4.1

Sl.No Average nominal

capacitance [Co1]pF

Average nominal capacitance

[Co2]pF

1 3.0102 3.1124

Table 4.1 Measured values of average nominal capacitance.

Observations and comments:-

1. The average nominal capacitance between electrodes and proof-

mass measured is around 3 pF which is very close to the simulated

value of 2.5 pF. The discrepancy in nominal capacitance may be

due to stray capacitance caused by electrical routing & pads and

also due to fine difference in air gap on both sides.

2. However the variation in nominal capacitances Co1 and Co2 is

small hence can be balanced using internal trim capacitors of the

ASIC during electronic integration.

Page 5: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

90

3. Finally the probing results proves that the device is successfully

realized as per the design without any stiction or shorts and the

gap between proof–mass and electrodes is same on both sides and

is cleared for further electronic integration and testing.

4.3 Packaging & electronics integration

The fabricated wafers are diced and the individual chips are die

bonded on a ceramic carrier in LCC package as shown in plate 4.1. Gold

layout is patterned on the ceramic substrate in such a way so as to

accommodate both the sensor chip and all other required electronic

components. The chip is wire bonded with 1 mil gold wire. The wire

bond is tested for its pull strength to ensure quality wire bond.

The interface circuit for converting the variation in capacitance to

voltage is implemented using a standard capacitance to voltage

conversion ASIC, MS3110 from Irvine Sensors. MS3110 is a general

purpose, ultra noise CMOS IC that requires only a single +5V DC supply

and some decoupling components. It has gain & DC offset trim functions

and on chip EEPROM for storage of program coefficients. The circuit

gives a DC bias of 2.5V at 0 „g‟, which is also equal to the reference

voltage of the ASIC.

The fabricated accelerometer chip shows a deviation in nominal

capacitance values from the designed values and hence the capacitance

Page 6: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

91

bridge is not balanced at zero „g‟. As a result there is an initial offset in

the output voltage when at zero „g‟. The offset is nullified by using the

internal capacitances in the ASIC. In case where the offset is much more

than the limit of the internal capacitances, provision is made to add an

external capacitor of suitable value in parallel with the lower capacitance

in the bridge. Provision is made in the interface circuit board for the

tuning of the ASIC coefficients after the final assembly of the components

to cater for packaging effects also.

Plate 4.1 Packaged sensor with electronics.

4.4 Scale factor test

Scale factor or sensitivity of an accelerometer is the ratio of the

sensor electrical output to mechanical input typically rated in mV /g.

This is the fundamental parameter to specify a sensor and forms the

Page 7: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

92

basis for further detailed performance testing. The test method block

diagram is shown in fig 4.2

Fig 4.2 Test set-up for scale factor measurement.

The accelerometer sensor is mounted on an automatic

accelerometer calibrator and connected to power supply and the output

is monitored using precision digital oscilloscope. The calibrator generates

a physical excitation signal of magnitude ±1 „g‟ at a frequency of 159Hz.

The screen shot of the oscilloscope captured during testing is presented

in fig 4.3

Page 8: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

93

Fig 4.3 Sensitivity of accelerometer

It can be seen from the oscilloscope output that the sensitivity of the

accelerometer measured is 62 mV/g in both directions.

Observations and remarks:-

The sensitivity of the accelerometer is 62 mV/g at room

temperature. Change in sensitivity can be obtained by programming the

gain of the ASIC. The sensor is programmed to have an offset voltage of

2.5 V at zero „g‟. The maximum output at +30 „g‟ is 4.5V and at -30 „g‟ it

is 0.5V. Hence the maximum sensitivity that can be obtained is

66mV/g. With measured sensitivity of 62mV/g the sensor yields 3.1 mV

for 50 milli „g‟ resolution, which can be detected easily.

Page 9: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

94

4.5 Hysteresis test

A sensor should be capable of following the changes of the input

parameter regardless of which direction the change is made, hysteresis is

the measure of this property. An example of hysteresis within an

accelerometer is the presence of residual deflection/strain within the

sensor's spring after acceleration has been applied and then removed. In

the presence of hysteresis, an accelerometer will not be able to

successfully repeat its null position; this will lead to unstable bias.

Hysteresis is expressed as % of FSO.

Fig 4.4 Block diagram of Hysteresis testing

As shown in the fig 4.4, the sensor is mounted on a centrifuge in such

a way that the sensing axis is in radial direction. The sensor is suitably

rotated at different speeds to obtain upto + 30 „g‟ acceleration

in steps. The output measured at different accelerations

Page 10: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

95

applied from 0 „g‟ to 30 „g‟ and while returning from 30 „g‟ to 0 „g‟ are

plotted in fig 4.5 and the output values are given in table 4.2.

Fig. 4.5 Hysteresis test result

Table - 4.2 Accelerometer output at different ‘g’

0.00

500.00

1000.00

1500.00

2000.00

2500.00

3000.00

3500.00

4000.00

4500.00

-30 -20 -10 0 10 20 30

Ou

tpu

t V (

mV

)

Acceleration g

Hysterisis

Up

Down

Acceleration (‘g’) UP (mV) Down (mV)

-30 320.00 322.00

-18 1063.00 1064.00

-6 1800.00 1802.00

0 2181.00 2180.00

6 2548.00 2550.00

18 3294.00 3297.00

30 4041.00 4041.00

Page 11: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

96

Observation and remarks:-

From the readings presented it can be seen that the maximum deviation

in output is at 18 „g‟. The hysteresis is given as

(∆y)/2 X 100 = 0.08 % of FSO

FSO The measured value of hysteresis is 0.08%, which is much less than the

specified value of 0.15 % of FSO.

4.6 Bias stability or drift

Bias stability is specified as a percentage of FSO at constant temperature

over a specified time period.

Fig 4.6 Test set-up for bias stability measurement.

The accelerometer is positioned in such a way that its sensitive axis (Z) is

perpendicular to the earth gravitation vector. In this way the sensor is

not subjected to any acceleration. The sensor output is connected to a

data logger. The offset variation is measured using a digital multi meter

(DMM) and logged over a period of two hours and plotted as in Fig 4.7.

Page 12: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

97

Fig 4.7 Bias stability with time.

Observations and comments

Long term bias stability measurements have demonstrated an

overall measured value of ±0.5mV over a period of 120minutes. This

works out to be 0.025% of FSO, which is well within the requirement of

0.15% of FSO.

4.7 Linearity test

The transfer function of the sensor (input/output relationship) is

not perfectly linear. Non-linearity is expressed as the ratio of maximum

deviation of output voltage from a best fit straight line to full scale output

of the device. This is expressed as a percentage of FSO and the equation

is given below.

2.1700

2.1750

2.1800

2.1850

2.1900

0 20 40 60 80 100 120

Ou

tpu

t(V

)

time(Min)

Drift Testing of SE03 for 2 Hours

Page 13: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

98

Non linearity = Maximum Deviation(Volts) X 100 %

Full Scale Output(Volts)

Non-linearity is one of the major sources of error in aerospace class of

accelerometers and shall be limited to less than 1% of the FSO.

Fig 4.8 Block diagram of linearity test

To conduct the test, Modalshop make automatic accelerometer

calibration workstation is used. The system uses back to back

comparison calibration method as per ISO 16063-21 [51] and generates

test reports automatically. It can apply a peak acceleration of ± 20 „g‟ at

a reference frequency of 100Hz. Hence the test range is limited to ± 20 „g‟

instead of full range of ± 30 „g‟.

Page 14: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

99

The Accelerometer is mounted on the shaker with its sensitive axis

(Z- axis) along the shaker excitation axis. The sensor is subjected to „g‟

sweep from ±1 „g‟ to ±20 „g‟ progressively at a reference frequency of

100 Hz. The „g‟ output from the sensor at different „g‟ values is compared

with that of a standard acceleration sensor and report generated

automatically as shown in Fig 4.9.

Page 15: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

100

Fig 4.9 Linearity test result (calibrator output)

Page 16: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

101

Observations and remarks

From fig 4.9 it can be seen that the linearity error of the sensor is

0.32% of FSO, which is well within the specified value of 1%.

4.8 Cross-axis sensitivity test

Aerospace systems experience acceleration forces along all three axes

i.e. pitch, roll and yaw. Accelerometer with its sense axis mounted along

a particular direction shall sense acceleration in that direction only and

shall be immune to the accelerations applied on other axes.

Cross-axis sensitivity is the output that is obtained on the sensing

axis for an acceleration applied on a perpendicular axis. This is

expressed as a percentage of the full scale output sensitivity. The sensor

has two cross-axis sensitivities 𝑆𝑍𝑌 and 𝑆𝑍𝑋 . The first subscript is the

sense axis and the second subscript is the off-axis direction. Cross-axis

sensitivity is given by

𝑆𝑍 𝐶𝑟𝑜𝑠𝑠 =𝑆𝑧𝑥

𝑆𝑧× 100

𝑆𝑍 𝐶𝑟𝑜𝑠𝑠 =𝑆𝑧𝑦

𝑆𝑧× 100

Page 17: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

102

Fig 4.10 Cross-axis testing block diagram

The sensor is mounted on a precision centrifuge in such a way that

its sensitive axis (Z) is along the Earth‟s gravitational vector. This method

of mounting eliminates both radial and tangential components of

acceleration acting on the sense direction. The sensor output for 1‟g‟

acting due to gravity is nullified and set to zero in the DMM. Now by

suitably rotating the centrifuge at appropriate speed the required cross

axis acceleration is applied on the accelerometer and output of the

sensor is recorded through a data logger.

Page 18: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

103

Fig 4.11 Cross-axis sensitivity plot

Cross axis

acceleration Output (V)

0 2.1812

5 2.1823

10 2.1834

15 2.1845

20 2.1855

25 2.1866

30 2..1874

Table 4.3 Cross-axis sensitivity output

2.1600

2.1650

2.1700

2.1750

2.1800

2.1850

2.1900

0 5 10 15 20 25 30 35

sen

sor

ou

tpu

t (

Vo

lts)

cross-axis acceleration applied (g)

Cross-axis sensivitty

Page 19: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

104

Observations and remarks

Since the sensor is symmetrical along X-axis and Y-Axis the cross-

axis test is done along one direction only.

The measured value of cross-axis sensitivity is 0.313% of FSO,

which is well within the specified value of 1% of FSO. However it is

more than the simulated value of 0.01% of FSO. This may be due

to the initial deflection present in the accelerometer because of the

Earth‟s gravity, fabrication error in positioning the beams at the

centre and due to sensor mounting misalignment in the package.

4.9 Temperature sensitivity test

Aerospace systems, during their operational period are exposed to

harsh environmental conditions, which includes vibration and wide

operational range of temperatures. The two most important performance

parameters that need to be studied for their temperature effects are bias

stability and offset variation. The temperature sensitivity of the

accelerometer is the sensitivity of a given performance characteristic to

operating temperature. It is expressed as the change of the characteristic

per degree of temperature change, typically in ppm/°C for scale factor

and mg/°C for bias. This figure is useful for the estimation of maximum

sensor error with temperature as a variable while modelling.

Page 20: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

105

4.9.1 Temperature sensitivity of offset value:

Temperature sensitivity of zero „g‟ offset is, the variation in the zero

„g‟ offset value over the operating temperature range. The offset variation

is measured by placing the accelerometer in a thermal chamber fig 4.12

and subjecting it to different operating temperatures. The accelerometer

is mounted in such a way that its sense axis is perpendicular to the

Earth‟s gravitational axis. The output is noted down using a precision

DMM which in turn is connected to a data logger.

Fig 4.12 Temperature sensitivity of offset test block diagram

Fig 4.13 Temperature sensitivity of offset

2.1760

2.1780

2.1800

2.1820

2.1840

2.1860

2.1880

-20 0 20 40 60 80

off

set

var

iati

on

(V

olt

s)

Temperature deg C

Page 21: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

106

Temp (ºC) Nominal output

(V)

-20 2.1773

0 2.1786

20 2.1812

40 2.1837

60 2.1860

80 2.1871

Table 4.4 Temperature sensitivity of offset

The zero „g‟ offset voltage is measured at -20, 0, 20, 40, 60, 80°C.

The zero „g‟ offset voltage at -20°C is subtracted from the value obtained

at 80°C. The resulting value obtained is divided by the accelerometer's

FSO to express the change in output in terms of % of FSO or alternately

it can be expressed as ppm also.

Observations and remarks:

From Table 4.4, it can be seen that over the operating temperature

range - 20°C to + 80°C the maximum change in zero „g‟ output is

9.8mV. Hence the temperature sensitivity of zero „g‟ error works

out to be 0.52% of FSO (or) alternately this can be expressed as

1.58mg/°C (at 62mV/g sensitivity).

Page 22: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

107

The offset variation is fairly linear with temperature, hence by

implementing suitable temperature compensation techniques, the

effect can be reduced considerably.

4.9.2 Temperature sensitivity of the scale factor:

Temperature sensitivity of the scale factor is the change in the

sensitivity of the accelerometer from the room temperature sensitivity as

the temperature changes. The variation is measured using special test

set-up in which accelerometer is placed on one end of an arm which is

inside the thermal chamber and the other end is outside and is

connected to a precision rotary table. Now the accelerometer is subjected

to the temperatures -20, 0, 20, 40, 60, 80°C. By rotating the rotary table

the accelerometer is subjected to ± 1 „g‟ acceleration and the scale factor

value is noted down at different temperatures as shown in fig 4.14. After

the testing is complete, the data is analyzed. The sensitivity at 25°C is

subtracted from each of the measurements. The resulting maximum

change in sensitivity is divided by the accelerometer's sensitivity at 25°C

to express the change in output in terms of ppm change in scale factor.

Page 23: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

108

Fig 4.14 Temperature sensitivity of the scale factor

Temp (ºC) Sensitivity (mV/ ‘g’)

-20 61.906

0 61.923

20 61.948

40 61.965

60 61.989

80 62.006

Table 4.5 Temperature sensitivity of scale factor

61.9

61.92

61.94

61.96

61.98

62

62.02

62.04

-20 0 20 40 60 80

sen

siti

vit

y

mv/g

'

temperature C

Page 24: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

109

Observations and remarks:

From Table 4.5 it can be seen that the scale factor variation over

the temperature range is a maximum of 0.1mV. Hence the scale

factor stability is 1612 ppm. This stability figure is adequate for

control class aerospace applications.

By adopting closed loop control techniques, scale factor variation

can be reduced considerably and more precise accelerometers can

be realized.

Page 25: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

110

4.10. Bandwidth test

The bandwidth is defined as the useful frequency range, in which

the output of the sensor is within ±3dB of the nominal value. The test

set-up block diagram is shown in fig 4.15.

Fig 4.15 Test set-up for bandwidth measurement.

The accelerometer is mounted on Modalshop make automatic

dynamic shaker. In this system, the output of the accelerometer under

test is compared with an inbuilt reference accelerometer output and the

performance is compared. A sine sweep signal of 1 „g‟ magnitude is

applied from a frequency of 10 Hz to 10000 Hz taking the output at 100

Hz as the reference value. The deviation in amplitude response as a

function of frequency is shown in fig 4.16.

Page 26: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

111

Fig 4.16 Frequency sweep output of accelerometer

Observations and remarks

From fig 4.16 amplitude response vs. frequency plot it can be seen that,

±3dB deviation in output is occurring at 800Hz. Hence the sensor meets

the operational bandwidth of 100Hz.

Page 27: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

112

4.11 Shock test

Accelerometers used in aerospace applications are subjected to high

shocks during the operation. The sensor shall withstand the shock and

exhibit normal performance after the shock is withdrawn.

Fig 4.17 Block diagram of shock testing

As shown in the fig 4.17 the sensor is mounted on a shock tester in

such a way that the sensing axis is along the shock input axis. A half

sine shock signal of 50 „g‟ magnitude is applied for a duration of

11msec. Fig. 4.18 gives the input shock spectrum and fig. 4.19 is the

response shock spectrum of the accelerometer, the graph plots the

response of three different accelerometers with different sensitivities. Also

the shock response spectrum provides a measure of response time of the

sensor.

Power

Supply Accelerometer Data

Logger

Shock

Tester

Page 28: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

113

Fig 4.18 Input shock spectrum

Fig 4.19 Response shock spectrum

- Sensor 1

- Sensor 2

- Sensor 3

Page 29: Chapter 4 Accelerometer Testing - Information and …shodhganga.inflibnet.ac.in/bitstream/10603/2272/10/10...The accelerometer was tested using procedure and data analysis methods

114

Observations and remarks:-

From the test results, it can be seen that, all the three

accelerometers have responded in a similar way to the applied

shock and the response is along the lines of input shock.

It can be seen that the accelerometers have very fast response of

less than 1msec.

4.12 Results & discussion

The sensitivity of the sensor measured is 62 mV/g.

The sensor demonstrated linearity error and cross-axis sensitivity

less than 1% of FSO as designed.

The hysteresis and bias stability values measured are less than

0.15% of FSO and meet the sensor specifications.

The temperature sensitivity of zero „g‟ error or offset error is 0.52%

of FSO (or) 1.58mg/°C (at 62mV/g sensitivity) and is fairly linear

over the temperature range.

The scale factor stability over the operational temperature range is

1612 ppm.

The operational bandwidth of the sensor is >100Hz as designed.

Sensor response time measured from shock test is less than

one msec.