Cerebral Oximetry
Wesam Farid MousaAssist prof Anesthesia &
ICUDammam University
![Page 2: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/2.jpg)
Experts have created helmets to protect the brain of motorcyclists
![Page 3: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/3.jpg)
Under general Anesthesia, the human brain can be in danger and the anesthesiologists are
sometimes faced with neuronal dysfunction/death during anesthesia
![Page 4: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/4.jpg)
NEURODETECTION & NEUROPROTECTION
• No “real time” bioassays to detect the onset of brain injury
•CT, ECHO, EEG & MRI are late indicators of brain injury
![Page 5: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/5.jpg)
• The need is critical: Cerebral ischemia is the
leading cause of compromised neurocognitive outcomes
The duration of reduced oxygenation has a direct impact on brain function
![Page 6: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/6.jpg)
• THE EVIDENCES Approximately 465,000
cardiopulmonary bypass surgeries performed in the U.S. each year
Cerebral oxygen desaturation occurs in 17%-23% of cardiopulmonary bypass surgeries (or about 79,000-107,000 cases per year) Most desaturation events cannot be detected by common monitoring devices
![Page 7: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/7.jpg)
• THE EVIDENCES• Adverse neurologic outcomes have been
reported in 6% to 53% of patients undergoing cardiopulmonary bypass surgeries
• Neurological complications vary in
severity; some are transient and some are permanent
• Adverse cerebral outcomes include focal injury, stupor, coma, seizures, memory deficit or deterioration in intellectual function
![Page 8: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/8.jpg)
Incidence of Cerebral Desaturation Events
![Page 9: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/9.jpg)
WHY CEREBRAL OXIMETRY?
• NIRS cerebral oximetry is non-invasive and continuous “real-time” detection that guides clinicians in their interventions
![Page 10: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/10.jpg)
WHY CEREBRAL OXIMETRY?
• The brain: Complex and fragile system The brain weighs about 1200 to 1500
grams, accounts for only 2% of body weight, but receives 15% of total cardiac output and utilizes 20% of the oxygen
Elapsed time is critical in desaturation events
![Page 11: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/11.jpg)
Why is Cerebral Oximetry different from other parameters?
° Cerebral Oximetry (SctO2) measures regionalmetabolism and the balance of local
tissue oxygen supply and demand.
° SctO2 is a combined saturation measurementof arterial blood and venous blood.
° other monitoring modalities are invasive
![Page 12: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/12.jpg)
Cerebral Oxygen Saturation Consensus Level
![Page 13: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/13.jpg)
From Pulse Oximetry to Cerebral Oximetry•Both are based on Beer-Lambert’s Law that describes light absorption
• Pulse Oximetry: Only Two wavelengths! Using light loss (Systole – Diastole) to calculate only signals due to pulsation of arterial blood to derive arterial O2 saturation.
• Cerebral Oximetry: Looking into entire nonpulsating field to derive tissue O2 saturation. It requires precise light sources and more wavelengths to account for light lost from other elements -as melatonin- in addition to Oxy- and Deoxy-Hemoglobin.
Light lost due to pulsation of arterial blood
Light lost due to non-pulsating arterial blood
Light lost due to venous blood
PulseOximetry(2 λ)
CerebralOximetry(multiple λ)
Light lost from the Background:• Light lost due to other molecules• Light scattering lost
time
Inci
dent
Lig
ht In
tens
ity
Ligh
t Los
t Sys
tole
tim
eLi
ght L
ost D
iast
ole
![Page 14: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/14.jpg)
Arterial Pulse Oximetry
• Tells how much oxygen blood is carrying in arterial system (oxygenated Hb)
• Unable to determine how much oxygen is used (venous system)
• Unable to determine how different organs use oxygen (global vs regional) Accuracy about 2% (not accurate below 85%)
![Page 15: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/15.jpg)
SpO2 is100% arterial blood
i.e., all pulsing
SctO2 is mix of:• 30% arterial • 70% venous
• no pulse needed
![Page 16: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/16.jpg)
VeinsArteries
Tissue (Arterioles, capillaries, venules)
Pulse Oximetry SpO2 : Arterial
Oxygen Saturation Normal range:
90-100%
Cerebral OximetrySctO2: Regional TissueOxygen SaturationNormal Range: 60-80%
SjvO2: Jugular VenousOxygen SaturationNormal Range: 50-70%
![Page 17: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/17.jpg)
sats95%
VBAG65%
NIRS
![Page 18: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/18.jpg)
The Scientific Basis For CerebralOximetry NIRS
The Spectrum of Light The NIRS “Window” From 650nm to 900nm The Modified Beer-Lambert Law The Elliptical Photon Pathway In The Brain Chromophores In Cerebral Tissue Random Background Scattering Of Light More Wavelengths = Greater Accuracy Depth Of Penetration
![Page 19: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/19.jpg)
The Spectrum of Light
• The term “spectrum” for the colors produced by a prism,
• colors in sunlight can be recombined to make white light.
Sir Isaac Newton1642-1727
![Page 20: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/20.jpg)
Microwave Infrared Near-infrared Visible Ultraviolet X-ray Gamma ray
Window for NIRS Between 650 and 950 Nanometers
![Page 21: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/21.jpg)
![Page 22: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/22.jpg)
Fundamental Principle of Oximetry: Beer – Lambert Law
![Page 23: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/23.jpg)
Emitter-detector pairs showing the banana-shaped paths of light.
![Page 24: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/24.jpg)
How Does Light Penetrate Bone?
Near infrared light and some wavelengths of visible red light easily penetrate human tissue,including bone Light is used to noninvasively interrogate gray matter in the frontal portion of the cerebral cortexThe depth of the light path is based on the separation of light source to detector and the strength of the light source
![Page 25: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/25.jpg)
Chromophores In Cerebral Tissue
NIRS is based on specific chromophores absorption of near infra red light for various wavelenghths. In case of cerebral tissues, these chromophores are: oxygenated hemoglobin, deoxygenated hemoglobin and oxidized cytochrome (Caa3)
![Page 26: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/26.jpg)
Other Chromophores
• There are a number of chromophores which absorb light in the same region as oxy-Hb and deoxy-Hb
• Interference caused by these chromophores must be detected and accounted for to get accurate SctO2 measurements
wavelengths for CO-, Met-, total-Hb, Skin pigmentation
![Page 27: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/27.jpg)
CADAVERS: NIRS Oxygenated Hb Showed a 90% DecreaseWithin 7h of Death, Continued Decrease in from 7 to 96 h.
![Page 28: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/28.jpg)
Absorption spectra for oxygenated haemoglobin (HbO2), deoxygenated haemoglobin (Hb), Caa3, melanin, and water (H2O) over wavelengths in NIR range. Note the relatively low peak for Caa3.
![Page 29: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/29.jpg)
Commercial cerebral NIRS devices currently utilize wavelengths in the 700–850 nm range to maximize separation between Hb and HbO2. The presence of melanin as found in human hair can significantly attenuate Hb, HbO2, and Caa3 signals.
Abs
orpt
ion
![Page 30: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/30.jpg)
Light Absorbance of Hemoglobin
![Page 31: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/31.jpg)
![Page 32: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/32.jpg)
Arterial Pulse Oximetry
![Page 33: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/33.jpg)
• Coronary artery bypass grafting (CABG) on/off pump
• Isolated valve replacement (AVR, MVR)• Combined valve replacement and CABG• Aortic arch replacement with DHCA• Ventricular assist device (LVAD, RVAD,
BiVAD)• Cardiac transplantation
Clinical Applications Of Cerebral Oximetry
![Page 34: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/34.jpg)
• Liver transplantation• ECMO• Carotid endarterectomy• “Beach Chair” positioning surgery (e.g.,
shoulder surgery)• Neonatal cyanotic heart defects• High-frequency ventilation• Bowel Ischemia (Neonates)
![Page 35: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/35.jpg)
Sizes are: Neonate, pediatric, adult
![Page 36: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/36.jpg)
Somatic Oximetry
![Page 37: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/37.jpg)
![Page 38: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/38.jpg)
Brain Areas Accessible by NIRSLIMITATIONS: What Part of the Brain is Involved
![Page 39: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/39.jpg)
LIMITATIONS: What Part of the Brain is Involved
![Page 40: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/40.jpg)
LIMITATIONS: What is the cause
Know how Different Organs Behave Under Different Conditons
Ischemia (Poor Flow)
Vs
Hypoxia (No Oxygen)
Vs
Venous Congestion
![Page 41: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/41.jpg)
Old Technology uses two wave lengths, with no absolute normal values and theirAlgorithms are only designated for adults
![Page 42: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/42.jpg)
AccurateImprecise AccuratePrecise InaccuratePrecise InaccurateImprecise
Precision and Accuracy
Standard Finger Oxygen Sat =2%Old technology NIRS: Two channels 9.72%Recent technology NIRS : Five channels 3%
![Page 43: Cerebral oximetry](https://reader036.vdocuments.us/reader036/viewer/2022081507/58880c5b1a28ab554e8b6c33/html5/thumbnails/43.jpg)
Thank You