    Continuous EmissionsMonitoring System II

    Gasmet Technologies, April 2010

  • 7/29/2019 CEM II

    2/27

    INTRODUCTION

    Continuous Emissions Monitoring System II

  • 7/29/2019 CEM II

    3/27

    Background

    First Gasmet CEMS was introduced in2001

    System and software has been modified

    during last years but not dramatically Over 160 CEMS and 60 sub systems has

    been supplied until today world wide

    Continuous improvement is required toremain competitive

    CEMS must be easier to use and maintain Legislation demands vary in different countries

    Price competition

    More FTIR competitors

  • 7/29/2019 CEM II

    4/27

    CEM II Introduction New sampling system

    Controller and heated pump/filterpackage in one 19 enclosure

    TCP/IP connection to computer

    New I/O approach Beckhoff K-bus modules for analog

    and digital I/O

    PCI expansion cards are not needed

    Smaller cabinet 600x600x2150mm

    Air conditioning unit included inabove dimensions

    Complete cabinet can be shipped on apallet (CX4000 in separate box)

  • 7/29/2019 CEM II

    5/27

    CEM II & CEMS Comparison

    600 600 mm footprint

    Entire sampling unit in 19 enclosure, 4 U high

    Glass door, display visible when door closed

    Height with air conditioning unit 2150 mm

    800 800 mm footprint

    Separate sampling unit controller and hot box

    Door must be opened to see display

    Height with air conditioning unit 2450 mm

    CEM IICEMS

  • 7/29/2019 CEM II

    6/27

    600 600 mm footprint

    Access space required for door opening:1800 600 mm

    Clean side panels, no side clearancerequired

    Two CEM II systems often fit in thespace required for one old CEM system

    800 800 mm footprint

    Access space required for door opening:2400 800 mm

    Additional side clearance required formain switch and optional side-mounted airconditioning unit

    Height with top mounted air conditioning

    unit 2450 mm may be a problem

    CEM II & CEMS Comparison

  • 7/29/2019 CEM II

    7/27

    BASIC STRUCTURE

    Continuous Emissions Monitoring System II

  • 7/29/2019 CEM II

    8/27

    CEM II Basic Structure

    Computer

    Sampling System

    ZrO2 analyzer

    Keyboard

    Gasmet CX-4000 FTIROption with TDL O2 laser

    TFT Display

    Cooling Unit

    Sample Gas In

    Sample Gas Outlet:

    left or right side wall

    3 x 16 A, 230 VAC

    (opt. 115 VAC)

    Analog inputs / outputs

    Digital / Relay outputs

    Fieldbus

    Zero Gas In (N2)

    Span Gas In

    Instrument air In

    Cabinet equipped with socle or wheels (option)

    Drawer(s) for spare parts and manuals

  • 7/29/2019 CEM II

    9/27

    GASMET Cx-4000 FTIR Gas

    Analyzer Gasmet GICCOR interferometer

    Excellent temperature andvibration stability

    High signal to noise ratio

    Peltier cooled MCT detector

    Heated sample cell 180 C

    Option with TDL O2 laser

    Optical paths 2.5m, 5m or 9.8m

    Measuring ranges from low ppm up tohigh % level

    RS-232 output to industrial computer

  • 7/29/2019 CEM II

    10/27

    ZrO2 Oxygen Analyzer

    ZrO2 technology

    TV approved

    Detection limit < 1 ppm

    4-20 mA output

    Automated single point

    calibration possible with CEM IIsampling system

  • 7/29/2019 CEM II

    11/27

    GASMET Industrial Computer Specifications:

    2,4 GHz Pentium IV processor

    512 MB memory

    2 x 40 GB hard discs with rack

    RAID I Disk operating system

    1 x 40 GB empty spare disk

    Windows Xp operating system

    DVD-RW

    Modem & Network card + PC

    Anywhere Hardware watchdog support

  • 7/29/2019 CEM II

    12/27

    Sampling Unit One 19 rack mounted unit including all

    valves, flow switch, pump, filter, flowmeters and a PLC controller

    Sample flow and instrument air pressurealarms

    PLC controller in sampling unit removesthe need for PCI slots in computer. All I/Ofrom computer is routed through the PLCin the sampling system (except RS-232)

    Temperature controllers for heated sample

    lines and oven are controlled by theCalcmet software.

    Connected to Gasmet computer viaEthernet cable. Maintenance switch infront panel

    Separate models for 100-115 VAC and 230VAC / 50 - 60 Hz

  • 7/29/2019 CEM II

    13/27

    Sample Probe M&C SP2000-H

    Weather protection shield

    Mounting flange DN 65 PN 6B

    Operating temperature 0-180 C,

    Dust loadings < 2 g/m3

    Sample tube SP2000-SS

    Material SS316

    Lenght 1 m

    Sample gas temperature 600 C

    Heated probe tube recommended

    in wet applications Higher temperatures and dust

    loadings on request

  • 7/29/2019 CEM II

    14/27

    Heated Lines Specifications

    Operating temperature up to 180 C

    Teflon (PTFE) inner core, replaceable

    6 mm outer diameter (4 mm ID)

    Nylon braid external jacket

    230 VAC power supply

    Options

    Rigid plastic jacket

    115 VAC power supply

    PFA tubing instead of PTFE

    1/4 outer diameter tubing Integrated power and alarm cables for probe

    Calibration gas tube for performing span and zero checks via probe

    Color coding

    230 VAC versions - red stripe

    115 VAC versions - yellow stripe

  • 7/29/2019 CEM II

    15/27

    Cabinet (Standard)

    Cabinet size600x600x2150mm

    Cooling unit on the top

    Front door and back door

    Front door with window Standard door opening

    directions shown

    Other directions availablefor extra cost

    Plinth or wheels below the

    cabinet Connector plate behind the

    TFT display

    Easy access to terminals

    1. Computer2. Keyboard3. Sampling Unit4. Oxygen analyser5. CX4000 FTIR analyser6. Air conditioning unit7. TFT display & Connector plate

    8. Drawer for Spare parts

  • 7/29/2019 CEM II

    16/27

    Cabinet (Option)

    Cabinet size600x800x2100mm

    Cooling unit on the back door

    Front door and back door

    Front door with window Standard door opening

    directions shown

    Other directions withextra charge

    Wheeled cabinet not available

    Additional rack space for e.g.FID analyser

    Deeper cabinet (800 mm)ensures that larger analyserswith more connectors in therear can be easily installed

    1. Computer2. Keyboard3. Sampling Unit4. O2 analyzer5. CX40006. Air conditioning unit (rear door)7. TFT display & Connector plate

    8. Drawer for Spare parts

  • 7/29/2019 CEM II

    17/27

    SAMPLING SYSTEM

    Continuous Emissions Monitoring System II

  • 7/29/2019 CEM II

    18/27

    Sampling System Flow Diagram

    Sample gas in

    0.5 lpm to oxygenanalyzer4 lpm to FTIRanalyzer

    Instrument air in Spangas in

    Zero gas inTypically N2 (quality 5.0) or Synthetic air (N2 80%,O2 20%)

    Instrument air backflush and spangas feed to probe are standard

    features Instrument air must be connected

    in order for the system to startmeasurement

    Multipoint sampling and multiplespan gases are available as

    external valve manifolds

    White = NO, valve normally openBlack = NC, valve normally closed

    Shuttle valve

    Back blow

    valves

  • 7/29/2019 CEM II

    19/27

    Sampling Unit

    Sampling unit contains

    Heated pump (180 C, PTFE membrane) Particle filter (sub-micron, sintered 316L steel)

    Flow alarm sensors (FTIR & O2)

    Solenoid and shuttle valves for controlling sample and test gas flow

    Rotameters for adjusting flow of test and purge gases

    Programmable controller with I/O interface to computer running Calcmet

  • 7/29/2019 CEM II

    20/27

    Sampling Unit Gas Connectors

    GAS INPUTS

    Sample gas (heated)

    Span gas

    Zero gas

    Instrument air

    GAS OUTPUTS

    Sample out 1 (heated), FTIR

    Sample out 2 (heated), O2

    Span gas to probe

    Interferometer purge air

    O2 reference air

    All gas connections are made using6 mm Swagelok fittings located onback panel

    Line holders for heated lines (notshown)

  • 7/29/2019 CEM II

    21/27

    Sampling Unit Pump & Filter

    FILTER

    Bonded microfibre or sintered steel

    filtration of particulates 2 m or 0,1 m

    Temperature 180C

    PUMP

    Material ss-316

    PTFE diaphragm

    Flow 4 l/min

    Temperature 180C

    Sample gas pressure 1,5 bar max

  • 7/29/2019 CEM II

    22/27

    Sampling Unit Valves

    INSTRUMENT AIR VALVES

    Sample gas / backflush

    SOLENOID VALVES

    Span gas via sample out

    Span gas via probe Zero gas

    Safety air

    BACK BLOW VALVES

    Sample gas / backflush

    Maintenance gases

    Cracking pressure 0.14 bar

    Automatic solenoid valves for

    gas inlets and back blow valves

    to avoid sample gas flow to

    unheated side.

    Sample gas/ flush gas valve isoperated using instrument air

    Instrument air pressure range 0

    - 10 bar (max.)

  • 7/29/2019 CEM II

    23/27

    Sampling Unit Flow meters

    FLOW 4 l/min

    Safety air

    Zero gas

    Span gas to sample out

    FLOW 0.5 l/min

    Interferometer purge

    O2 analyser reference air

    Adjustable flow meters for

    unheated gases

    Located on front panel

  • 7/29/2019 CEM II

    24/27

    Sampling Unit Temperature

    Controllers TEMPERATURE CONTROLLERS Sampling unit includes

    thermocouple analog inputs for

    heated module and for three

    sample lines.

    Range 0 180 C

    Type TcK

    Sample line 1 connector has 24

    VDC control for external solid

    state relay.

    Temperature display in Calcmetsoftware

    Temperature control circuit in

    Beckhoff BC9000 controller

    inside sampling system

  • 7/29/2019 CEM II

    25/27

    Sampling Unit I/ODIGITAL OUTPUTS (24 DVC)

    Alarm outputs are displayed onLED lights and unit hascorresponding digital outputs.

    Alarm outputs are:

    System alarm,

    Service request,

    Maintenance

    (can be set also manually),

    Concentration alarm,

    Result valid Combination for each alarm can be

    set in Calcmet.

    Unit has digital output commandto start Oxygen analysercalibration

    DIGITAL INPUTS

    Digital inputs are:

    Cabinet air condition unitcooling alarm,

    Cabinet temperature,

    Probe temperature,

    Zero gas pressure,

    Oxygen analyser alarm

    It is also possible to activate spangas test via probe or System

    standbyby digital input.

  • 7/29/2019 CEM II

    26/27

    Sampling Unit I/O to CalcmetCOUNTERS

    Pump membrane age in hours

    All valve ages in times

    DIGITAL INPUTS

    Cabinet air condition unit coolingalarm,

    Cabinet temperature,

    Probe temperature,

    Zero gas pressure,

    Oxygen analyser alarm

    SENSORS

    Sample out 1 flow low

    Sample out 2 flow low

    Instrument air pressure low

    Maintenance switchTEMPERATURES

    Sample Line 1 temp & alarm

    Sample Line 2 temp & alarm

    Sample Line 3 temp & alarm

    Heated module temp & alarm