Department of Electronics & Tele-communication EngineeringREPORT OF CE-LAB
1) CONCISED REPORT OF EQUIPMENTS :
Sr.
noEquipments
Total No.
Completely Ok
Having minor
Faults &Repairable
Not inUse&
Repairable
Not in use & Not
repairableRemarks
1 Voltage Power Supply 6 3 3 - -Minor Repairing is
required2 Single Probes 24 22 - - 2 -3 Dual Probes 5 5 - - - -4 Bread Boards 16 16 - - - -5 Spectrum Analyzer 6 6 - - - -6 Function Generators 7 2 2 3 - -7 CRO 6 2 4 - - -
8 Micro Ammeters 5 3 1 1 -Minor Repairing is
required
9 DSO 1 1 - - - -10(i)
DMM (Meco 801) 4 1 2 - 1 -
10(ii)
DMM (SINETECH 4022) 8 4 3 1 - -
11 Trainee kits 45 42 3 - -Minor Repairing in
1 kit is required12 Pulse Generators 4 3 1 - - -13 Cutter 2 1 - - 1 Extra required
2) REQUREMENTS :
Sr.No.
Components / Equipments Description No. Remark
1 OP-AMP 741 200 Urgent2 Diodes 1N400x 300 Urgent3 LCR Meter - 10 Urgent4 IC Tester - 10 Urgent
5 Resistors 100-100K ohm100
Individual Urgent
6 Capacitors 1PF-100Micro F100
Individual Urgent
7 Inductors 1mH-100mH100
Individual Urgent
3) FULL REPORT :
Sr.No Equipment
Total
No.
ModelNo
Serial NoStatu
sDescription / Faults Remark
1Aplab regulated dc
power supply6
7115 1 390/10 CE/12
OK NO FAULTS OK
LQ6324
2 02
3 03
4 04 Faulty4.5-5.5 Vtg knob I not
workingRepairable
5 05 Faulty4.5-5.5 Vtg knob I not
working
6 01 Faulty0-30 Vtg knob I not
workingRepairable
2.A
CRO20 MHz
Oscilloscope 4
ST201 1 100214105 OK No Faults OK
ST201
2 100214104 FaultyCHANNEL-II is not
workingRepairable
3 100214106 Faulty Repairable
4 100214107 Faulty
CH-I Knob is not working and LINE, INV CH-II Knob is
missing
Repairable
2.B
CRO 30 MHz Oscilloscope
2 Caddo 8011 02124626
OK - OK2 02124627
3 Digital Multimeter 12
MECO 801
1 206599 FaultyDc current meter is
not working.Repairable
2 206596 Faulty Display problem Repairable
3 206578 Faulty Totally Faulty Repairable
4 206564 OK - OK
SINETECH
4022
1 12
OK - OK2 14
3 15
4 16
5 13Faulty
Dc current meter is not working.
Repairable6 11
7 17Faulty Connectivity problem Repairable
8 19
410MHz AM/FM
Function Generator
4
ST4063 1 0706711 FaultyModulation frequency knob is not working
Repairable
1 0706710
OK - OK2 0706712
3 0706713
5TDM/PCMRECEIVER TRAINER
1 ST2104 1 0107862 OK - OK
6PAM/PPMPWM
DEMODULATION TRAINER
2 ST21101 0107862
OK - OK2 0107863
7 DSB/SSB/AM MODULATION TRAINER
2 ST2202 20207836
OK - OK0207837
8DIGITAL
COMMUNICATION TRAINER
3 ST2103
1 0207911
OK - OK2 0207912
3 0106760
9DIGITAL
COMUNICATION TRAINER
2 ST21021 1206585
OK - OK2 1206586
10
FREQUENCY MODULATION/
DEMODULATIONTRAINER
2 ST22031 0507816
OK - OK2 0507817
11DSB/SSB AM
TRANSMITTER RECEIVER
1 ST2201 1 0907864 OK - OK
12TDM/PCMRECEIVER TRAINER
1 ST2104 1 1205713 OK - OK
13 1 ST2101 1 1205574 OK - OK
14DIGITAL
COMUNICATION TRAINER
1 ST2105 1 0106512 OK - OK
15PAM/PPM/PWM
TRAINER1 ST2116 1 0107543 Faulty
PPM block is not working
Repairable
16MOBILE PHONE
TRAINER2 ST2132
1 060692OK - OK
2 060693
17ANTENNA AND
RADAR TRAINER2 ST2261
1 104304OK - OK
2 104305
18VCD PLAYER
TRAINER1 ST2662 1 120750 OK - OK
19SATELLITE
COMM. TRAINER UPLINK
1 ST2272 1 070647(1) OK - OK
20
SATELLITE COMM. TRAINER TRANSPONDER
1 ST2272 1 070647(2) OK - OK
21SATELLITE
COMM. TRAINER DOWNLINK
1 ST2272 1 070647(3) OK - OK
22 ISDN EMULATOR 1 - 1 AF03C13753 OK - OK
23SPECTRUM ANALYZER
4 FS300
1 01
OK - OK2 023 034 04
24DUAL ARBITARY
GENERATOR1 AM-300 1 01 OK - OK
25SIGNAL
GENERATOR1 SM-300 1 01 OK - OK
26RF SIGNAL
GENERATOR2 -
1 050658OK - OK
2 050659
27SCIENTECH
1MHZ FUNCTION GENERATOR
4 ST4046
1 10021618 FaultyOUT-DATED
MAY BE REPAIRABLE
2 10021619 Faulty
3 10021620 OK - OK
4 10021621 Faulty Triangular
27 DSO 1 TDS2024 1 039537 OK - OK
Most of the multimeters are not in use because they lack batteries, fuses. So they are needed urgently (Problem in current measurement).
DSO should be in use. All the faulty CRO, Function Generators have faults in buttons, which can be easily repaired. Number of Micro-ammeters, & cutters are less than required. LCR meter, IC tester, & Logic gate ICs should be available in lab.
Report prepared by,
Mr.Gajbhare Sandeep. (89836625475)Mr. Patil Satish G. (9860387401)
Head of Department Lab InchargeDr. S.V. Bonde Prof. A.I.Tamboli