    1 - Introduction

    2 Troubleshooting

    3 -Parts Lists

    4 -Removal & Replacement

    5 -Verification Procedures

    CELL-DYN3200

    SERVICE

  • 8/10/2019 CD 3200 Service Manual

    2/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 1

    4 0 ,QWURGXFWLRQ Proprietary Information

    4 0 ,QWURGXFWLRQ

    Table of Contents

    1.1 Proprietary Information . . . . . . . . . . . . . . . . . . . . . . . . . 1

    1.2 Revision Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

    1.3 About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

    Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

    Service Manual Section Descriptions . . . . . . . . . . . . . 3

    Section 1: Introduction & General Data . . . . . . . . . . . . 3

    Section 2: Troubleshooting . . . . . . . . . . . . . . . . . . . . . . 3

    Section 3: Parts Lists . . . . . . . . . . . . . . . . . . . . . . . . . . 3

    Section 4: Removal & Replacement . . . . . . . . . . . . . . . 3

    Section 5: Verification Procedures . . . . . . . . . . . . . . . . 3

    Section 6: Preventive Maintenance & Total Call. . . . . . 3

    1.4 CELL-DYN 3200 Operators Manual . . . . . . . . . . . . . . 3

    1.5 Warranty Information . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    1.6 System Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    The CELL-DYN 3200 Hematology System . . . . . . . . . . 4

    System External Components . . . . . . . . . . . . . . . . . . . 4

    System Internal Components . . . . . . . . . . . . . . . . . . . . 5

    1.7 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Signal Words and Graphic Warning Conventions . . . 5Signal Words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Graphic Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Biosafety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    Laser Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    1.8 CELL-DYN 3200 Hematology Parameters. . . . . . . . . . 9

    1.9 System Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 10

    1.1 Proprietary InformationThis CELL- DYN 3200 System Service Manual was developedsolely for use with Abbott Laboratories equipment as specified

    in the operating instructions. The Manual is copyrighted byAbbott Laboratories. No part of this document may bereproduced, stored, or transmitted in any form or by any meanselectronic, mechanical, photocopied, recorded, or otherwisewithout the prior written permission of Abbott Laboratories.

    This CELL- DYN 3200 System Service Manual is published byAbbott Diagnostics, a Division of Abbott Laboratories, AbbottPark, IL 60064-3500, U. S. A. Please direct all inquiriesconcerning information in this manual to the foregoing address.

    CELL- DYNis a registered trademark of Abbott Laboratories.

    Vacutaineris a registered trademark of Becton-Dickinson andCompany.

    This service manual was developed to be used by trainedAbbott Laboratories Field Service Personnel. The revisionstatus of this manual is the responsibility of the manual holder.

    In no event shall Abbott Laboratories or its subsidiaries be

    liable for any damages incurred in connection with or arisingfrom the use of this manual by persons not fully trained byAbbott Laboratories.

    1998 by Abbott Laboratories. All rights reserved.

  • 8/10/2019 CD 3200 Service Manual

    3/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 2

    4 0 ,QWURGXFWLRQ 1.2 Revision Log

    1.2 Revision Log

    The Revision Status of this manual is indicated below. Be surethat the manual contains the latest revision number of allpages.

    Table 1-1 Revision Log

    Revision Date Pages Revised/Added TSBs Incorporated ISAs Incorporated

    A August Complete Revision 109 - 001A, 003 109 - 003, 004, 005, 006

  • 8/10/2019 CD 3200 Service Manual

    4/334CELL-DYN3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 3

    4 0 ,QWURGXFWLRQ 1.3 About This Manual

    1.3 About This Manual

    Introduction

    This Service Manual contains the information necessary toalign, troubleshoot, repair, and maintain the CELL-DYN 3200

    Automated Hematology System.

    Service Manual Section Descriptions

    The manual is composed of five sections. The following aredescriptions of the contents of each of the sections.

    Section 1: Introduction

    This section contains an introduction to the system, information

    on the use of the manual, warranty information, safetyconsiderations, hematology parameters, and systemspecifications.

    Section 2: Troubleshooting

    This section contains procedures to be used by the FSR inidentifying and isolating problems. The procedures arepresented in a flowchart format and are accompanied byillustrations for added clarity. A detailed explanation of the use

    of the flowcharts is given at the beginning of the section.

    Section 3: Parts Lists

    This section contains parts lists and graphics showing thelocations of FRUs. This section is used by the FSR inconjunction with Troubleshooting and Removal & Replacement.

    Section 4: Removal & Replacement

    This section contains the procedures to be used by the FSR forthe removal and replacement of subassemblies andcomponents. The procedures are presented in a paragraphformat and are accompanied by illustrations for added clarity.

    Section 5: Verification Procedures

    This section contains procedures that enable the FSR to verifythe integrity of various instrument functions. These proceduresmay be used to facilitate troubleshooting, or be dictated byTroubleshooting and Removal and Replacement.

    This section also contains procedures such as adjustments,alignments, and calibrations used to ensure proper instrumentperformance. These procedures may be performed on a routine

    maintenance basis, or be dictated by Removal andReplacement.

    1.4 CELL-DYN 3200 Operators ManualOperating instructions and routine maintenance procedures forthe CELL-DYN 3200 are included in the Operators Manualprovided with each system. It is essential that the serviceengineer read and understand the Operators Manual, and beable to correctly perform all routine operating functions before

    attempting to troubleshoot and repair the system.

  • 8/10/2019 CD 3200 Service Manual

    5/334CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 4

    4 0 ,QWURGXFWLRQ 1.5 Warranty Information

    1.5 Warranty InformationComponents comprising the CELL-DYN 3200 System havebeen designed by Abbott Laboratories for optimal performance.Substitution of reagents, accessories, or system components

    may adversely affect performance and may invalidate anywarranty agreements.

    Abbott Laboratories cannot accept responsibility for any clinicalresults produced by use of reagents, calibrators, controls,

    equipment, or supplies either not manufactured by or notapproved by Abbott Laboratories.

    1.6 System IntroductionEach CELL-DYN 3200 System may be comprised of three orfour major hardware components: Analyzer, Video Display,Keyboard, and optional Sample Loader.

    The CELL-DYN 3200 Hematology System

    When approaching a hematology system from a serviceperspective, the word system must be emphasized becausethere are both internal components (system hardware andsoftware) and external components (reagents, controls, etc.)that comprise the complete system. The Service

    Representative must realize that both of these types ofcomponents can adversely affect overall system performance.Therefore, it follows that all external components andconditions such as reagents, environment, integrity of samplesand controls be checked and verified as correct beforeperforming service on the internal components. In theinvestigation of any non-specific or data related complaint, the

    system hardware and software should be the last componentsof the system to be suspected as the cause of the problem.

    System External Components

    The following is a list of the CELL-DYN 3200 externalcomponents and conditions:

    Operator Technique (Maintenance)

    System Reagents

    Isotonic Diluent

    Diluent/Sheath

    Hemoglobin Lyse

    WBC Lyse

    System Controls & Calibrators

    CELL-DYN 3200 Whole Blood Control

    OPTI-CAL

    Latex Beads

    Whole Blood Specimens

    System Environmental Integrity (Temperature, RFI, ESD)

    Power Line Integrity

  • 8/10/2019 CD 3200 Service Manual

    6/334CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 5

    4 0 ,QWURGXFWLRQ 1.7 Safety

    System Internal Components

    The following is a list of the CELL-DYN 3200 internalcomponents:

    CELL-DYN 3200 System Components

    Flow Panel

    Optics Bench

    Power Supplies

    Measurement Electronics

    Computer Electronics

    A/D & D/A Electronics

    Air Supply

    Personal Computer Module

    Tube Manipulation Hardware

    1.7 SafetyOperation, maintenance, and servicing of hematology systemsmay expose individuals to potential safety and health hazards.All work must be performed in accordance with proceduresdescribed in the Abbott Operators and Service Manuals, or asdirected by an Abbott Technical Support Representative.

    Warnings are inserted in this manual to alert personnel to

    potential hazards. The standard warning conventions includingsignal words (e.g. Caution) and symbols are described below.

    Signal Words and Graphic Warning

    Conventions

    Signal Words

    '$1*(5= Denotes an immediate hazard which, if not

    avoided, could result in serious injury or death. This signal wordrepresents the highest level of any hazardous situation.

    :$51,1*= Denotes a hazard which, if not avoided, couldresult in moderate to serious injury.

    &$87,21= Denotes potential hazards that could result inminor injury. Also, used for conditions or activities that could

    threaten equipment performance.

    Graphic Conventions

    The general hazard icon identifies an activity or areathat may present a hazard to personnel or equipment.

    The electrical hazard icon alerts personnel to thepossibility of electrical shock if procedural or

    engineering controls are not observed.

  • 8/10/2019 CD 3200 Service Manual

    7/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 6

    4 0 ,QWURGXFWLRQ 1.7 Safety

    Biosafety

    Clean-up spills of potentially infectious materials in accordancewith established biosafety practices. A generally acceptedprocedure for cleaning such spills is to absorb the spill withtoweling or other absorbent material, wipe the area with a

    detergent solution, and then wipe the area with an appropriatetuberculocidal disinfectant such as 10% chlorine bleachsolution.

    Prior to service or maintenance, the instrument should be

    decontaminated in accordance with the following:

    a. Remove and dispose of contaminated disposables in aregulated medical waste container.

    b. Clean and decontaminate the exterior of theinstrument using a detergent solution followed by a10% chlorine bleach solution or other tuberculocidal

    disinfectant. Flush the fluid pathway with chlorinebleach solution as specified in the Operators Manual.

    Dispose of waste in accordance with local, state and federalregulations. Probes, needles, broken glass, slides, and othersharps that are contaminated with potentially infectioussubstances should be collected in a sharps container fordisposal as regulated medical waste. Contaminated gloves,wipes, and other disposables should be placed in a standardmedical waste container.

    The biohazard icon identifies an activity or area wherepersonnel may be exposed to infectious substances.

    The laser hazard icon identifies an activity or areawhere personnel will be exposed to an eye hazard ifprocedural or engineering controls are not observed.

    The electrostatic discharge icon identifies an activity orarea where personnel must wear a ground strap whileservicing the system.

    :$51,1*= 327(17,$/%,2+$=$5'Considerall clinical specimens, reagent controls, surfaces, or

    components that contain or have contacted humanblood or serum as potentially infectious. Wear gloves,lab coats and safety glasses, and follow otherbiosafety practices as specified in the OSHA

    Bloodborne Pathogen Rule (29CFR Part 1910.1030)or other equivalent biosafety procedures.

    :$51,1*= 327(17,$/%,2+$=$5'Theaspiration needle/probe is sharp and potentiallycontaminated with infectious materials. Avoid contactwith the tip of the needle/probe.

  • 8/10/2019 CD 3200 Service Manual

    8/334

    CELL-DYN3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 7

    4 0 ,QWURGXFWLRQ 1.7 Safety

    Electrical Safety

    Laser Safety

    The Class 1 Laser Product Label (Abbott PN 9230702A) isshown below (Figure 1-1). The label consists of black letteringagainst a white background. The label is located on thebackside of the instrument and positioned at a clearly visible

    location.

    The inner protective cover laser warning labels must not be

    removed and are to remain legible. The protective housinglabel (Abbott PN 9230701D) is shown in Figure 1-2. The label

    consists of black lettering against a yellow background andappears on the protective cover within theoptics bench and onthe flow panel (see Figure 1-3and Figure 1-4).

    Figure 1-1 Laser Product Label (PN 9230702A)

    :$51,1*= (/(&75,&$/+$=$5' When theinstrument is powered and protective covers areremoved, there are exposed electrical systems thatcould startle or seriously injure personnel upon

    contact. Use appropriate safety precautions to preventbody and/or tool contact with live electricalcomponents, especially power supplies.

    DANGER: LASER LIGHT Avoid direct exposure tobeam. Do not look directly into the laser beam or anyreflection of the beam from a mirror-like surface.

    When the access door, instrument top cover, or otherinner protective cover is removed, (Helium-Neon)laser power up to 10 mW continuous wave at 632.8nm in a beam with 1 mR divergence could beaccessible in the interior of the optics bench. Thisamount of energy, with insignificant attenuation withdistance, is sufficient to cause eye damage.

    &$87,21: Use of controls or adjustments orperformance of procedures other than those specifiedherein may result in hazardous laser light exposure. Ifthe instrument is used or modified in a manner notspecified by the manufacturer, the protection providedby the instrument may be impaired.

    32109

  • 8/10/2019 CD 3200 Service Manual

    9/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 8

    4 0 ,QWURGXFWLRQ 1.7 Safety

    Figure 1-2 Laser Warning Label (PN9230702D)

    Figure 1-3 Top Laser Warning Label Location

    For beam alignment and other open beam configurations,follow instructions provided by the Service Manual, Advisories,and Bulletins regarding the requirement for using laser safety

    eyewear. If required, assure laser eyewear are not damagedand have an optical density of 1- 2 at a wavelength of 632.8nm. During open beam configurations, assure that the beam isconfined to the laser bench area and only personnel with propereye protection are present.

    .CUGT

    9CTPKPI

    .CDGN

    4 , W G WL

  • 8/10/2019 CD 3200 Service Manual

    10/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 9

    4 0 ,QWURGXFWLRQ 1.8 CELL-DYN 3200 Hematology Parameters

    Figure 1-4 Front Laser Warning Label Location

    1.8 CELL-DYN 3200 HematologyParameters

    The following tables list the CELL-DYN 3200 Hematology

    Parameters and their abbreviations.

    .CUGT

    9CTPKPI

    .CDGN

    Table 1-2 Red Cell Parameters

    Full Name Abbreviation

    Red Blood Cell Concentration RBC

    Hemoglobin Concentration HGB

    Hematocrit HCT

    Mean Cell Volume MCV

    Mean Cell Hemoglobin MCH

    Mean Cell Hemoglobin Concentration MCHC

    Red Cell Distribution Width RDW

    Table 1-3 White Cell Parameters

    Full Name Abbreviation

    White Blood Cell Concentration WBC

    Neutrophil Concentration NEUNeutrophil Percentage %N

    Lymphocyte Concentration LYM

    Lymphocyte Percentage %L

    Monocyte Concentration MONO

    Monocyte Percentage %M

    Eosinophil Concentration EOS

    4 , W G WL 1 9 S t S ifi ti

  • 8/10/2019 CD 3200 Service Manual

    11/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 10

    4 0 ,QWURGXFWLRQ 1.9 System Specifications

    1.9 System SpecificationsThe following tables contain the specification for the CELL-DYN3200 System.

    Eosinophil Percentage %E

    Basophil Concentration BASO

    Basophil Percentage %B

    Table 1-4 Platelet Parameters

    Full Name Abbreviation

    Platelet Concentration PLT

    Mean Platelet Volume MPV

    Platelet-crit PCT*

    Platelet Distribution Width PDW*

    * Clinical significance has not been established for these

    parameters. Therefore they are not reportable in the US.

    They are provided for laboratory use only.

    Table 1-3 White Cell Parameters (Continued)

    Full Name Abbreviation

    Table 1-5 Physical Specifications (Unpackaged)

    CS Model SL Model GraphicsPrinter

    Height 19.0 (48.2 cm) 19.0 (48.2 cm) 7.2 (18.3 cm)

    Width 32.8 (83.3 cm) 32.8 (83.3 cm) 16.1 (40.9 cm)

    Depth 23.3 (59.2 cm) 29.3 (74.4 cm) 9.9 (25.1 cm)

    Weight 214 lb (97.3 kg) 232 lb (105.5 kg) 9.9 lb (4.5 kg)

    Table 1-6 Physical Specifications (Packaged)

    CS Model SL ModelGraphics

    Printer

    Height 38 (96.5 cm) 38 (96.5 cm) 12 (30.5 cm)

    Width 34 (86.3 cm) 34 (86.3 cm) 19.5 (49.5 cm)

    Depth 31 (78.7 cm) 31 (78.7 cm) 14.8 (37.6 cm)

    Weight 255 lb (116 kg) 273 lb (124 kg) 16.4 lb (7.4 kg)

    4 ,QWURGXFWLRQ 1 9 System Specifications

  • 8/10/2019 CD 3200 Service Manual

    12/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 11

    4 0 ,QWURGXFWLRQ 1.9 System Specifications

    Table 1-7 Analyzer Input Requirements

    Module Voltage FrequencyMax

    Current

    Analyzer

    110 VAC + 10 % 50/60 3 Hz 6 amps

    120 VAC + 10 % 50/60 3 Hz 6 amps

    220 VAC + 10 % 50/60 3 Hz 3 amps

    240 VAC + 10 % 50/60 3 Hz 3 amps

    Analyzer &

    Sample

    Loader

    110 VAC + 10 % 50/60 3 Hz 6 amps

    120 VAC + 10 % 50/60 3 Hz 6 amps

    220 VAC + 10 % 50/60 3 Hz 3 amps

    240 VAC + 10 % 50/60 3 Hz 3 amps

    Monitor 100 - 240 VAC 50/60 1.3 amps

    Printer 120 VAC 10% 50/60 0.5 amps220 - 260 VAC

    Table 1-8 Performance Specifications

    Background Precision CorrelationLinearRange

    WOC 0.10K/L 2.7% 0.99 0.0-250 K/ L

    NOC 0.10 K/L 2.7% 0.99 0.0-250 K/ L

    RBC 0.20 M/L 1.5% 0.98 0.0-8.0 M/ L

    HGB 0.1g/dl 1.0% 0.98 0.0-25.0g/dL

    MCV N/A 1.0% 0.98 35-180 fL

    RDW N/A 5.0% 0.91 N/A

    PLT 5.0K/L 4.0% 0.980.0 - 1750

    K/L

    MPV N/A 5.0% 0.72 2.0-18.0 fL

    4 0 ,QWURGXFWLRQ 1 9 System Specifications

  • 8/10/2019 CD 3200 Service Manual

    13/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300August 1998 1 - 12

    4 0 ,QWURGXFWLRQ 1.9 System Specifications

    Table 1-9 Operational Specifications

    Operating Environment

    Temperature 15o- 30oC (59o- 86oF)

    Relative Humidity 20% to 80%, RHNC

    Complete Cycle TimesStartup (From Power Off) ~7 minutes

    Startup (From Standby) ~5 minutes

    Count Cycle, Patient ~52 seconds

    Count Cycle + FrgWBC ~100 seconds

    Count Cycle + ResRBC ~100 seconds

    Shutdown (To Standby) ~130 seconds

    Shutdown (To Power Off)~130 seconds

    Throughput

    CS Model

    Open 80 Samples/Hour

    Closed 68 Samples/Hour

    SL Model

    Open 80 Samples/Hour

    Closed 68 Samples/Hour

    Whole Blood Aspiration Volume

    Open Mode 125 L

    Closed Mode 250 L

    Specimen Tube Dimensions

    Diameter 13 mm

    Length 75 mm

    Minimum Specimen Volumes

    Closed and Sample Loader Mode 1 mLOpen Mode 500 L

    Batch Size

    1 to 50 tubes per batch

    Bar Code Formats

    Interleaved 2-of-5 code

    Codabar

    Code 39Code 128 (check digit only)

    Bar Code Specifications

    Printed on good quality label stock

    0.25-inch quiet zone on each end

    0.01-inch minimum narrow bar width

    0.5-inch minimum bar length

    2-inch maximum label length

    1.25-inch maximum label width

    5 7URXEOHVKRRWLQJ Table of Contents TOC

  • 8/10/2019 CD 3200 Service Manual

    14/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300-August 1998 2 - 1

    5 7URXEOHVKRRWLQJ Table of Contents

    5 7URXEOHVKRRWLQJ

    Table of Contents

    2.1 Summary. .. . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

    2.2 Troubleshooting Approach . . . . . . . . . . . . . 2-2

    2.3 Error Messages . . . . . . . . . . . . . . . . . . . . . . . 2-3

    IP-1 Fluid Leaks, Open Mode Probe or Wash Block . . . 2-4IP-2 Fluid Leaks, Closed Mode Needle or Wash Block .2-6IP-3 Incomplete Aspiration . . . . . . . . . . . . . . . . . . . . . . . 2-8IP-4 Flow Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11IP-5 RBC/HGB Diluent Overpressure . . . . . . . . . . . . .2-13IP-6 Optical Parameter Count Problems . . . . . . . . . . .2-15

    IP-7 Reagent Empty . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26IP-8 Optical Backgrounds Too High . . . . . . . . . . . . . . .2-31IP-9 Histogram Problems . . . . . . . . . . . . . . . . . . . . . . .2-32IP-10 Controls Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-33IP-11 Vacuum Accumulator #1 Wet . . . . . . . . . . . . . . . .2-35IP-12 Vacuum Accumulator #2 Wet . . . . . . . . . . . . . . . .2-37IP-13 HGB Problems . . . . . . . . . . . . . . . . . . . . . . . . . . .2-39IP-14 X-B Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-41IP-15 Optical Parameters Imprecision . . . . . . . . . . . . . .2-44IP-16 HGB Imprecision . . . . . . . . . . . . . . . . . . . . . . . . .2-47

    IP-17 Primary Vacuum/Pressure Isolation, . . . . . . . . . .2-48IP-18 Pressure Leak Isolation . . . . . . . . . . . . . . . . . . . .2-57IP-19 Vacuum Leak Isolation . . . . . . . . . . . . . . . . . . . . .2-64IP-20 Shear Valve Position Fault . . . . . . . . . . . . . . . . . .2-70IP-21 Flags . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-72IP-22 External Waste Full . . . . . . . . . . . . . . . . . . . . . . .2-74IP-23 Blank Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-76IP-24 HSSL Failures/Message Reception Time-Out . . . 2-77

    IP-25 Hard Drive Errors . . . . . . . . . . . . . . . . . . . . . . . . . 2-78

    IP-26 Printer Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-79IP-27 Sample Loader & Aspiration Tower

    Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . 2-81Appendix 2A -Cable Connection Diagram . . . . . . . . . . . . 2A-1Appendix 2B -System Block Diagrams . . . . . . . . . . . . . . 2B-1Appendix 2C -System Flow Diagrams . . . . . . . . . . . . . . . 2C-1

    TOC

    5 0 7URXEOHVKRRWLQJ 2.1 Summary

    http://trblsh.pdf/http://sec2trshoot/trblsh.pdf
  • 8/10/2019 CD 3200 Service Manual

    15/334

    CELL-DYN3200 SYSTEM SERVICE MANUAL Part Number 9140300- August 1998 2 - 2

    5 7URXEOHVKRRWLQJ 2.1 Summary

    2.1 Summary

    This chapter contains the following information for CELL-DYN3200:

    Error Messages

    Isolation Procedures (IPs)

    Block Diagrams

    Normal Scatter/Histograms

    Service Commands

    2.2 Troubleshooting Approach

    Troubleshooting is required whenever an Observed Error or aDisplayed Error occurs. Displayed Errors can be viewed on the

    Bulletin Line or status box of the Data Station screen. (TheBulletin Line is down about 3/4 way on the screen.) Detailedinformation can be found in the fault report screen of thediagnostics submenu.

    When a malfunction occurs, follow these recommendedprocedures:

    1. Use this Troubleshooting Guide as follows:

    a. If possible, begin by verifying the reported error.Perform the exact procedure that produced the errorcondition.

    b. When an error occurs, either observed or displayed,look up the error in:

    the Error Messagestable or IP 27, Sample Loader & Aspiration Tower Error

    Messagestable

    c. To refer to the appropriate IP or VP, click on the redlink.

    d. Begin at the START box and follow the directions in theIP.

    e. If a part is adjusted or replaced, go toSection 5.2,Verification Matrix.

    f. Find the part in left column.

    g. Perform VPs in order given (i.e., 1, 2, 3, 4 etc.)

    2. Perform Total Call.

    5 0 7URXEOHVKRRWLQJ 2.3 Error Messages

  • 8/10/2019 CD 3200 Service Manual

    16/334

    CELL-DYN3200 SYSTEM SERVICE MANUAL Part Number 9140300- August 1998 2 - 3

    J g

    2.3 Error Messages

    Error Corrective Action

    Background, HGB Too High IP-13

    Background, RBC/ PLT Too High IP-8

    Background, WBC Too High IP-8Bar Code Reader Failure IP-27

    C:\ Hard Drive Errors IP-25

    Controls Out IP-10

    Diluent/Sheath Empty IP-7

    Erratic Results, HGB IP-13

    Erratic Results, Optical

    IP-6Ensure Blood Sam

    ple is Fresh. Differential

    count is not reliable

    before 30 minutes or after

    12 hours after sample is

    drawn.

    External Waste Full IP-22

    Flags IP-21

    Flow Errors IP-4

    Fluid Leaks, Closed Mode Probe IP-2

    Fluid Leaks, Open Mode Probe IP-1

    HCT Errors IP-14

    HGB Errors IP-13

    Histogram Errors IP-9

    HSSL Com Failures IP-24

    Incomplete Aspiration IP-3

    Invalid Tube Height IP-27

    Leaks, Pressure IP-18

    Leaks, Vacuum IP-19

    Load Zone Empty Detector

    Malfunction IP-27

    MCH Errors IP-14

    MCHC Errors IP-14

    MCV Errors IP-14

    Message Reception Timeout IP-24

    Mix Head Not At Top Position IP-27

    Mix Head Stuck At Top Position IP-27

    Mix Head Stuck In Angle Position IP-27Mix Head Stuck In Vertical

    PositionIP-27

    Mix Zone Rack Position Error IP-27

    No Tube Present IP-27

    Overpressure, Diluent Syringe IP-5

    Precision Problems HGB IP-16,OPT IP-15

    Pressure Errors IP-17

    Printer Errors IP-26

    Probe Stuck At Home Position IP-27

    Probe Unable To Reach Home

    PositionIP-27

    Rack Bar Code Read Failure IP-27

    Shear Valve Position Fault IP-20

    Error Corrective Action

    5 0 7URXEOHVKRRWLQJ 2.3 Error Messages

  • 8/10/2019 CD 3200 Service Manual

    17/334

    CELL-DYN3200 SYSTEM SERVICE MANUAL Part Number 9140300- August 1998 2 - 3A

    J g

    Tube Dropped During Mixing IP-27

    Tube Lost Moving From Position

    3 To 4IP-27

    Tube Position Bar Code Read

    FailureIP-27

    Tube Stuck In Position 3 IP-27

    Tube Stuck In Position 4 IP-27

    Unexpected Tube In Position 4

    After Rack AdvanceIP-27

    Unload Area Hardware

    MalfunctionIP-27

    Unload Zone Full Detector

    Malfunction IP-27

    Unload Zone Nearly Full Detector

    MalfunctionIP-27

    Vacuum Accumulator 1 Wet IP-11

    Vacuum Accumulator 2 Wet IP-12

    Vacuum Errors IP-17

    WBC Imprecision

    IP-15Ensure Blood Sam-

    ple is Fresh. Differential

    count is not reliable

    before 30 minutes or after

    12 hours after sample is

    drawn.

    WBC/HGB Lyse Empty IP-7

    X-B Errors IP-14

    Error Corrective Action

    5 7URXEOHVKRRWLQJ IP-1 Fluid Leaks, Open Mode Probe or Wash Block

    http://cd3200toc.pdf/http://cd3200toc.pdf/http://cd3200toc.pdf/http://cd3200toc.pdf/http://cd3200toc.pdf/http://cd3200toc.pdf/
  • 8/10/2019 CD 3200 Service Manual

    18/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300-August 1998 2 - 4

    IP-1 Fluid Leaks, Open Mode Probe or Wash Block (Page 1 of 2)

    Start

    Probe LeakingIn Ready Mode

    ?

    - Clamp tubing betweenUltrasonic sensor andShear valve

    Wash Block LeakingIn Ready Mode

    ?

    N O

    YES

    Leaking

    Stops?

    - Verify connections onUltrasonic Sensor- Verify and tighten fitt ingson Y-Valve- Verify connection on top

    of Probe- Replace Y-Valve(PL-8)

    YES

    - Clamp tubing at the top of

    Valve 11

    N O

    Leaking

    Stops

    ?

    YES

    - Verify pinch tubing in

    Valve 11- Verify and replaceValve 11 (PL-4.1)

    - Verify Shear valve

    fitt ings- Clean Shear Valve(VP-48)- Replace Shear Valve(PL-9.1)

    N O

    1

    YES

    2

    N O

    5 7URXEOHVKRRWLQJ IP-1 Fluid Leaks, Open Mode Probe or Wash Block

  • 8/10/2019 CD 3200 Service Manual

    19/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 5

    IP-1 Fluid Leaks, Open Mode Probe or Wash Block (Page 2 of 2)

    1

    - Verify pinch tubing in

    Valve 15- Verify and replace Valve15 (PL-4.1)

    2

    - Run a several CountCycles while observing thebottom of the Wash Block

    Liquid

    Leaking

    ?

    - Ensure Vacuum #1 and Pressure #3are within specifications(VP-16)- Ensure bottom f i t t ing on Wash Block

    is not restricted- Ensure tubing between Wash Blockand Valve 14 is not restricted- Verify operation of Valve 14 andreplace if necessary(VP-10), (PL-4.1)

    - Replace pinch tubing in Valve 14(TM-1)- Ensure tubing between Valve 14 andWC #4 is not restr icted- Verify operation of right Vent Trap

    Check Valve and replace i f necessary(TM-1)- Ensure tubng between WC #4 andValve 31 is not restricted- Verify operation of Valve 31 and

    replace if necessary(VP-10), (PL-4.1)- Replace pinch tubing in Valve 31(TM-1)

    - Ensure f i t t ings on Wash

    block are not leaking

    N O

    YES

    5 7URXEOHVKRRWLQJ IP-2 Fluid Leaks, Closed Mode Needle or Wash Block

  • 8/10/2019 CD 3200 Service Manual

    20/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 6

    IP-2 Fluid Leaks, Closed Mode Needle or Wash Block (Page 1 of 2)

    Start

    Probe LeakingIn Ready Mode

    ?

    - Clamp tubing betweenUltrasonic sensor andShear valve

    Wash Block Leaking

    In Ready Mode

    ?

    N O

    YES

    Leaking

    Stops

    ?

    - Verify connections onUltrasonic Sensor

    - Verify and tighten fitt ingson Y-Valve- Verify connection on top

    of Probe- Replace Y-Valve(PL-8)

    YES

    - Clamp tubing at the top of

    Valve 11

    N O

    LeakingStops

    ?

    YES- Verify pinch tubing inValve 11

    - Verify and replaceValve 11(PL-4.1)

    - Verify Shear valvefitt ings- Clean Shear Valve(VP-48)- Replace Shear Valve

    (PL-9.1)

    N O

    1

    YES

    2

    N O

    Leave tubing connected, andremove Needle from Wash Block

    5 7URXEOHVKRRWLQJ IP-2 Fluid Leaks, Closed Mode Needle or Wash Block

  • 8/10/2019 CD 3200 Service Manual

    21/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 7

    IP-2 Fluid Leaks, Closed Mode Needle or Wash Block (Page 2 of 2)

    1

    - Verify pinch tubing in

    Valve 34- Verify and replace Valve34 (PL-4.1)

    2

    - Run a several CountCycles while observing thebottom of the Wash Block

    Liquid

    Leaking

    ?

    - Ensure Vacuum #1 and Pressure #3 arewithin specifications(VP-16)- Ensure bottom f i t t ing on Wash Block is notrestricted

    - Ensure tubing between Wash Block andValve 37 is not restricted

    - Verify operation ofValve 37and replace ifnecessary (VP-10), (PL-4.1)- Replace pinch tubing in Valve 37(TM-1)

    - Ensure tubing between Valve 37 and WC #4is not restricted- Verify operation of right Vent Trap CheckValve and replace if necessary(TM-1)- Ensure tubng between WC #4 and Valve 31 is

    not restricted- Verify operation ofValve 31 and replace ifnecessary (VP-10),(PL-4.1)- Replace pinch tubing in Valve 31 (TM-1)

    - Ensure f i t t ings on Wash

    block are not leaking

    N O

    YES

    5 7URXEOHVKRRWLQJ IP-3 Incomplete Aspiration

  • 8/10/2019 CD 3200 Service Manual

    22/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 8

    IP-3 Incomplete Aspiration (Page 1 of 3)

    Start

    - Aspirate NormalControl in both open

    and closed modes.

    Mode(s)

    Affected

    ?Open Only

    Closed Only

    - Verify Y-Valve is rotating

    properly (VP-7).

    Rotating

    Properly

    ?

    - Clean Shear Valve (VP-48).

    Problem

    Corrected

    ?

    Total Call

    YES

    NO

    - Replace Y-Valve Assy (PL-8) .

    - Replace MDM (Pl-9.2)

    - Replace SDM/CDM (PL-15).

    NO

    YES

    2 4

    Both

    1

    5 7URXEOHVKRRWLQJ IP-3 Incomplete Aspiration

  • 8/10/2019 CD 3200 Service Manual

    23/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 9

    IP-3 Incomplete Aspiration (Page 2 of 3)

    1

    - Run an open cycle andt ime the blood movementto the Ultrasonic Sensor.

    Start when Valve 12opens and stop whenblood reaches sensor.

    - Clean Open Probe(VP-50).

    - Ensure there are norestr ict ions between Open

    Probe and Y-Valve.

    - Replace Open Probe (PL-8).

    < 3.5 Seconds

    ?

    N O

    - Place a blood sampleunder Open Probe andmanual ly open Valve 12until the blood is justpast the Y-Valve.

    - Verify an air gap ofapproximately 1 inchbetween the diluent andblood columns.

    - If no air gap is present,

    check the Open Probeplumbing for air leaks.

    YES

    2

    - Run a closed cycle and

    time the bloodmovement to theUltrasonic Sensor. Startwhen Valve 12 opens

    and stop when bloodreaches sensor.

    - Clean Closed Probe(VP-51) .

    - Ensure there are no

    restr ict ions between ClosedProbe and Y-Valve.

    - Replace Closed Probe(PL-8).

    - Verify Valve 38(VP-10)an d

    ensure Vent Trap plumbing isfree of restrictions.

    < 7.0 Seconds

    ?

    N O

    - Manually place a blood

    sample under Closed

    Probe.- From Main Menu, press

    D IAGNOSTICS

    MORE

    TOWER TEST

    NEEDLE DOWN

    - Manually open Valve 12until the blood is justpast the Y-Valve.

    - Verify an air gap ofapproximately 3 inchesbetween the diluent andblood columns.

    - Check the Closed

    Probe plumbing for airleaks.

    Air GapPresent

    ?

    3

    YES

    NO

    YES

    5 7URXEOHVKRRWLQJ IP-3 Incomplete Aspiration

  • 8/10/2019 CD 3200 Service Manual

    24/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 10

    IP-3 Incomplete Aspiration (Page 3 of 3)

    - INITIALIZE and PRIMEinstrument.

    - Run a closed cycle andverify that the bloodcolumn stops just pastSample Sensor #2.

    - If blood does not stop:

    - Perform Sample Sensor

    adjustment(VP-12).

    - Re-seat sensor cable.- Replace Sample

    Sensor(PL-9.1).

    - Replace SampleSensor Board(PL-9.2).

    3 4

    - Verify Vac 2 level(VP-16) . Troubleshoot as necessary.

    - Ensure there are no restrictions between Y-Valve andValve 12

    -Verify Valve 12 and tubing (VP-10). Replace as necessary(PL-4.1),(TM-1) .

    - Run an open cycle andt ime the bloodmovement to theUltrasonic Sensor. Startwhen Valve 12 opensand stop when blood

    reaches sensor.

    < 3.5 Seconds

    ?

    NO

    - Run an open cycle andverify that blood column

    stops just pastUltrasonic Sensor. Ib lood does not stop:

    - Re-seat Ultrasonic

    Sensor Cable.

    - Replace UltrasonicSensor(PL-8).

    - Replace UltrasonicSensor Board (PL-9.2) .

    YES

    5 7URXEOHVKRRWLQJ IP-4 Flow Errors

  • 8/10/2019 CD 3200 Service Manual

    25/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 11

    IP-4 Flow Errors (Page 1 of 2)

    Start

    68

    WC 1

    SampleStagingPump

    Delivery Syringe

    From Main Menu:D IAGNOSTICS

    MORE

    MOTOR OPERATION

    EXERCISE MOTORS

    - Type A and press Enter.

    - Observe the Delivery Syringeand ver i fy smooth upwardmovement (See Figures).

    DrippingO K

    ?

    1

    N O

    YES

    - Run a Count Cycle and observethe Del ivery Syr inge and WC1.Verify that when the syringe ismoving up, l iquid is dripping intoWC1 at a fast and constant rate

    (See Figures).

    SmoothMovement

    ?

    3

    N O

    2

    YES

    5 7URXEOHVKRRWLQJ IP-4 Flow Errors

  • 8/10/2019 CD 3200 Service Manual

    26/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 12

    IP-4 Flow Errors (Page 2 of 2)

    1

    - Verify level of 9 psi (VP-16)

    - Verify tubing and operation ofValves 64, 65, 91, and 56 and

    replace as necessary(TM-1) ,(PL-4.1).

    - Perform Optical Flow Cell

    Cleaning Procedure(VP-44).

    - Replace In-Line Sheath Filter(PL-4.2).

    3

    - Remove and clean Del iverySyringe.

    - Replace Delivery Syringe

    (PL-7).

    - Replace Syringe DriverAssembly (PL-7).

    - Replace MPM(PL-11).

    2

    - Check Laser for stableoutput (VP-46).

    OutputStable

    ?

    - Replace Laser (PL-10).

    - Replace Laser Power Supply(PL-2.2).

    N O

    - Remove, inspect, and cleanFlow Cell Nozzle.

    - Replace Flow Cel l Assy. (PL-10).

    YES

    5 7URXEOHVKRRWLQJ IP-5 RBC/HGB Diluent Overpressure

  • 8/10/2019 CD 3200 Service Manual

    27/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 13

    IP-5 RBC/HGB Diluent Overpressure (Page 1 of 2)

    Start

    - Check resistance ofOverpressure Sensor

    SensorShorted

    ?

    YES

    - Replace Sensor (PL-4.2)

    - Remove, inspect, and clean

    Shear Valve(VP-48)

    - Ensure di luent paths fromsyringe to RBC and HGBMixing Chambers are free of

    restrictions

    - Verify operation of Valves 27and 26 (V P-10) and replace if

    necessary (PL-4.1)

    N O

    Prob lemCorrected

    ?

    YES

    - Verify Shear Valve rotation time

    is

  • 8/10/2019 CD 3200 Service Manual

    28/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 14

    IP-5 RBC/HGB Diluent Overpressure (Page 2 of 2)

    1 2

    Total Call

    In Ready Mode, try to manuallyrotate the front and back of the

    Shear Valve.

    I f there is movement:

    - Replace defective section:

    Front (PL-9.1)

    Rear (PL-9.1)If problem persists:

    - Replace Shear Valve Driver(PL-9.1)

    - Lubricate Shear Valve Driver(VP-42)

    If problem persists:

    - Replace Shear Valve Driver(PL-9.1)

    Total Call

    32002

    Rim NotchFaces Right

    Shear Valve Sect ions

    Should Not Move

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP 6 O ti l P t C t P bl (P 1 f 11)

  • 8/10/2019 CD 3200 Service Manual

    29/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 15

    IP-6 Optical Parameter Count Problems (Page 1 of 11)

    Start

    Note: This procedure assumes that HGB Results are OK in Open and Closed Modes, and there are no hardware or software Error Messages.

    I t is designed to f ind and correct gross WBC, NOC, RBC, and PLT problems. For Optical Precis ion Problems see IP-15.

    All " 0"

    ?

    From Main Menu, press

    RUN

    SPECIMEN TYPE

    FRAGILE WBC

    - Run a Normal Control f ive t imes andrecord the optical values.

    - Compare the results to the Data Sheet.

    RBC & PLTOnly " 0"

    ?

    N O

    3YES

    N O

    WBC Only " 0"

    ?

    4YES

    5YES

    RBC Only " 0"

    ?

    PLT Only "0"

    ?

    NO

    NOC Only " 0"

    ?

    6YES

    8

    7YES

    YES

    1

    NO

    NONO

    - Verify proper Reagents and ExpirationDate.

    - Perform Background Count Veri f icat ion(VP-56).

    BackgroundsO K

    ?

    YES

    Go to(IP-8),correct andreturn here.

    N O

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP 6 O ti l P t C t P bl (P g 2 f 11)

  • 8/10/2019 CD 3200 Service Manual

    30/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 16

    IP-6 Optical Parameter Count Problems (Page 2 of 11)

    1

    All Low

    ?

    RBC & PLT

    Only Low?

    N O

    9YES

    N O

    WBC OnlyLow

    ?

    10YES

    YES

    NOC OnlyLow

    ?

    12YES WBC Only

    High?

    15YES

    All High

    ?

    13YES

    N O

    NOC OnlyHigh

    ?

    16YES

    RBC & PLT

    Only High

    ?

    14YES

    N O

    11

    2

    N O

    NO

    N O

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP 6 Optical Parameter Count Problems (Page 3 of 11)

  • 8/10/2019 CD 3200 Service Manual

    31/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 17

    IP-6 Optical Parameter Count Problems (Page 3 of 11)

    2

    All Erratic

    ?

    RBC & PLT

    Only Erratic

    ?

    N O

    17YESWBC Only

    Erratic

    ?

    18YES

    YES

    NOC OnlyErratic

    ?

    20YES

    19

    Total Call

    N O

    N O

    N O

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP-6 Optical Parameter Count Problems (Page 4 of 11)

  • 8/10/2019 CD 3200 Service Manual

    32/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 18

    IP-6 Optical Parameter Count Problems (Page 4 of 11)

    3

    - Veri fy samples are being staged toOptical Flow Cell by Peri-pump.

    Troubleshoot as necessary.

    - Verify Sheath Reservoir is fi l l ingproperly.

    - Verify Pressure 2 (VP-16)an dproper Sheath f low to Optical FlowCel l . Troubleshoot as necessary.

    - Replace In-Line Filter(PL-4.2).

    - Verify operation of Samp leInjection Syringe(VP-6),(VP-7).

    Troubleshoot as necessary.

    Prob lemCorrected

    ?

    - Verify laser power(VP-46).

    N O

    Total Call

    YES

    Power OK

    ?

    - Connect new Laser Head to

    Laser Power Supply andveri fy power (see Warning).I f good, replace Laser Head(PL-10). If not, replaceLaserPower Supply (PL-2).

    NO

    N O

    - Veri fy beam al ignment(VP-20), (VP-21).

    - If not able to get data onthe Extended Count

    Screen, replace in orderthe Optical Flow Cell,MAM , SPM, and CPU/DCM (PL-10), (PL-12),(PL-13).

    YES

    'DQJHU'DQJHU'DQJHU'DQJHU=+LJK

    9ROWDJH

    00$VVXUH

    WKH

    SRZHU

    WR

    WKH

    LQVWUXPHQW

    LV

    RII1 $KLJKYROWDJHFKDUJHPD\UHPDLQRQWKHODVHUFRQQHFWRUZLWKWKHSRZHURII1 'RQRWWRXFKWKHPDOHSLQV1 6KRUWPDOHSLQV

    WR&KDVVLV1

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP-6 Optical Parameter Count Problems (Page 5 of 11)

  • 8/10/2019 CD 3200 Service Manual

    33/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 19

    IP 6 Optical Parameter Count Problems (Page 5 of 11)

    4

    - Veri fy RBC/PLT Sample is beingstaged to Optical Flow Cell.

    - Verify operation of Valve 54 (VP-

    10) and tubing. Replace asnecessary(PL-4.1),(TM-1).

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1SET POINT ENTRY

    - Veri fy RBC/PLT Threshold is550.

    - Reseat Preamp, MAM, SPM,CPU/DCM Cables .

    - Replace MAM (PL-12).

    - Replace 10 oPreamp(PL-10).

    - Replace SPM(PL-10).

    - Replace CPU/DCM(PL-13).

    5

    - Veri fy WBC Sample is being

    staged to Optical Flow Cell.

    - Verify operation of Valve 53and 55(VP-10)and tubing.Replace as necessary(PL-

    4.1) ,(TM-1) .

    - Veri fy operat ion of WBCSyringe(VP-6), (VP-7).

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy WBC Threshold is

    360.

    - Reseat Preamp, MAM, SPM,CPU/DCM Cables .

    - Replace MAM (PL-12).- Replace 0oPreamp(PL-10).

    - Replace SPM(PL-10).

    - Replace CPU/DCM(PL-13).

    - Replace MAM(PL-12).- Replace SPM(PL-10).

    - Replace CPU/DCM (PL-13).

    6

    - Replace MAM(PL-12).- Replace SPM(PL-10).

    - Replace CPU/DCM(PL-13).

    7

    8

    - Veri fy NOC Sample is

    being staged to OpticalFlow Cell.

    - Verify operation of Valve

    41(VP-10)and tubing.Replaceas necessary(PL-4.1),(TM-1) .

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1SET POINT ENTRY

    - Veri fy WBC Threshold is400.

    - Perform NOC GainAdjustment(VP-25).

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP-6 Optical Parameter Count Problems (Page 6 of 11)

  • 8/10/2019 CD 3200 Service Manual

    34/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 20

    IP 6 Optical Parameter Count Problems (Page 6 of 11)

    9

    -- Verify Sheath Reservoir is fi l l ingproperly.

    - Verify Pressure 2 (VP-16)an d

    proper Sheath flow to Optical FlowCel l . Troubleshoot as necessary.

    - Replace In-Line Filter(PL-4.2).

    - Verify operation of SampleInjection Syringe(VP-6),(VP-7).Troubleshoot as necessary.

    - Replace Vent Accumulator Fi l ter(PL-4.2) .

    Prob lemCorrected

    ?

    - Verify laser power (VP-46).

    N O

    Total Call

    YES

    Power OK

    ?

    - Connect new Laser Head to

    Laser Power Supply andveri fy power (See Warning).I f good, replace Laser Head(PL-10).If not, replaceLaser Power Supply (PL-2).

    N ON O

    - Clean Laser Bench(VP-43).

    - Clean, debubble,and wetOptical Flow Cell(VP-44),(VP-45).

    - Verify beam alignment (VP-20 ),(VP-21).

    - Replace Optical Flow Cell (PL-

    10 ).

    - Replace MAM(PL-12).

    - Replace CPU/DCM(PL-13).

    YES

    'DQJHU'DQJHU'DQJHU'DQJHU= +LJK9ROWDJH00 $VVXUHWKHSRZHUWRWKHLQVWUXPHQWLVRII1 $KLJKYROWDJHFKDUJHPD\UHPDLQRQWKHODVHU

    FRQQHFWRUZLWKWKHSRZHURII1 'RQRWWRXFKWKHPDOHSLQV16KRUWPDOHSLQVWR&KDVVLV1

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP-6 Optical Parameter Count Problems (Page 7 of 11)

  • 8/10/2019 CD 3200 Service Manual

    35/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 21

    p ( g )

    - Veri fy RBC/PLT Sample isbeing staged to OpticalFlow Cell.

    - Verify operation of Valve 54

    (VP-10) and tubing.Replace as necessary (PL-4.1),(TM-1).

    - Clean Shear Valve(VP-48).

    10

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy RBC/PLT Thresholdis 550.

    - Verify 0oand 10oL ogAmplifiers (VP-28).

    - Verify 0oand 10oGain(VP-26 ),(VP-27).

    - Replace MAM (PL-12).

    - Replace SPM (PL-10).

    - Replace CPU/DCM(PL-13).

    Prob lemCorrected

    ?

    Total Call

    YES

    N O

    - Veri fy WBC Sample isbeing staged to OpticalFlow Cell.

    - Verify operation of Valve 53and 55(VP-10)and tubing.Replace as necessary (PL-4.1) ,(TM-1) .

    - Veri fy operat ion of WBCSyringe (VP-6), (VP-7).

    - Clean Shear Valve(VP-48).

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy WBC Threshold is

    360.

    - Veri fy WBC Gain(VP-23).

    - Replace MAM(PL-12).- Replace SPM(PL-10).

    - Replace CPU/DCM(PL-13).

    11

    Prob lemCorrected

    ?

    Total Call

    YES

    N O

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP-6 Optical Parameter Count Problems (Page 8 of 11)

  • 8/10/2019 CD 3200 Service Manual

    36/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 22

    p ( g )

    12 13

    - Clean Laser Bench (VP-43).

    - Clean, debubble, and wetOptical Flow Cell (VP-44),

    (VP-45).

    - Verify beam alignment (VP-20 ),(VP-21).

    - Replace Optical Flow Cell(PL-10).

    - Replace MAM(PL-12).

    - Replace CPU/DCM(PL-13).

    14

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy RBC/PLT Threshold is

    550.

    - Verify 0oand 10o Log Amplifiers

    (VP-28).- Verify 0oand 10oGain(VP-26),

    (VP-27).

    - Replace MAM(PL-12).

    - Replace SPM (PL-10).

    - Replace CPU/DCM(PL-13).

    - Veri fy NOC Sample is beingstaged to Optical Flow Cell.

    - Verify operation of Valve 41(VP-10)and tubing. Replace asnecessary(PL-4.1), (TM-1) .

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy NOC Threshold is 400.

    - Perform NOC Gain Adjustment(VP-25).

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP-6 Optical Parameter Count Problems (Page 9 of 11)

  • 8/10/2019 CD 3200 Service Manual

    37/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 23

    16

    - Veri fy operat ion of WBCSyringe(VP-6),(VP-7).

    - Clean Shear Valve(VP-48).

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy WBC Threshold is360.

    - Veri fy WBC Gain(VP-23).

    - Replace MAM(PL-12).

    - Replace SPM(PL-10).- Replace CPU/DCM(PL-13).

    15

    Prob lemCorrected

    ?

    Total Call

    YES

    N O

    - From Main Menu, pressD IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy NOC Threshold is400.

    - Perform NOC Gain

    Adjustment(VP-25).

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP-6 Optical Parameter Count Problems (Page 10 of 11)

  • 8/10/2019 CD 3200 Service Manual

    38/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 24

    17

    - Veri fy samples are being staged toOptical Flow Cell by Peri-pump.Troubleshoot as necessary.

    - Verify Sheath Reservoir is fi l l ingproperly.

    - Verify Pressure 2 (VP-16)an dproper Sheath flow to Optical FlowCel l . Troubleshoot as necessary.

    - Replace In-Line Filter(PL-4.2).

    - Replace Vent Accumulator Fi l ter(PL-4.2).

    - Veri fy operat ion of SampleInjection Syringe(VP-6),(VP-7).Troubleshoot as necessary.

    Prob lemCorrected

    ?

    N O

    Total Call

    YES

    - Verify stable laser power

    (VP-46).

    Power Stable

    ?

    - Connect new Laser Head toLaser Power Supply andveri fy power (See Warning).I f good, replace Laser Head

    (PL-10). If not, replace LaserPower Supply(PL-2).

    N O

    - Clean Laser Bench(VP-43).

    - Clean, debubble, and wet

    Optical Flow Cell(VP-44),(VP-45).

    - Replace Optical Flow Cell

    (PL-10).

    - Replace MAM(PL-12).

    - Replace SPM(PL-12).

    - Replace CPU/DCM(PL-13).

    YES

    N O

    'DQJHU'DQJHU'DQJHU'DQJHU= +LJK9ROWDJH00 $VVXUHWKHSRZHUWRWKHLQVWUXPHQWLVRII1 $KLJKYROWDJHFKDUJHPD\UHPDLQRQWKHODVHUFRQQHFWRUZLWKWKHSRZHURII1 'RQRWWRXFKWKH

    PDOHSLQV1 6KRUWPDOHSLQVWR&KDVVLV1

    5 7URXEOHVKRRWLQJ IP-6 Optical Parameter Count Problems

    IP-6 Optical Parameter Count Problems (Page 11 of 11)

  • 8/10/2019 CD 3200 Service Manual

    39/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 25

    - Verify RBC/PLT Sample is being

    staged to Optical Flow Cell.

    - Verify operation of Valve 54 (VP-10)

    and tubing. Replace as necessary

    (PL-4.1) ,(TM-1) .

    - Clean Shear Valve (VP-48) .

    18

    - From Main Menu,

    press

    DIAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Verify RBC/PLT

    Threshold is 550.

    Problem

    Corrected

    ?

    Total Call

    YES

    NO

    - Verify operation of WBC

    Syringe (VP-6) ,(VP-7) .

    - Clean Shear Valve (VP-48).

    - From Main Menu, press

    DIAGNOSTICS

    F 12F 1

    SET POINT ENTRY

    - Verify WBC Threshold is

    360.

    - Replace MAM (PL-12) .

    - Replace SPM (PL-10) .

    - Replace CPU/DCM (PL-13) .

    19

    Problem

    Corrected?

    Total Call

    YES

    NO

    20

    - Verify NOC Sample is being staged

    to Optical Flow Cell.

    - Verify operation of Valve 41 (VP-10)

    and tubing. Replace as necessary

    (PL-4.1) ,(TM-1).

    - From Main Menu, press

    DIAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Verify NOC Threshold is 400.

    5 7URXEOHVKRRWLQJ IP-7 Reagent Empty

    IP-7 Reagent Empty (Page 1 of 5)

  • 8/10/2019 CD 3200 Service Manual

    40/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 26

    Start

    - Verify reagents, type andexpiration.

    - Ensure reagent

    containers are connected

    to the correct tubing andinput ports.

    - Ensure sinkers are on

    lines and lines are at

    bottom of containers.

    - Ensure reagent lines arenot crimped or clogged.

    - INITIALIZE and PRIME

    instrument.

    HGB LyseEmpty

    ?

    WBC Lyse

    Empty

    ?

    NO

    Diluent/Sheath

    Empty

    ?

    NO

    Total Call

    NO

    - Open Right Front Door

    and observe HGB/

    WBC Syringe.

    - Open Left Front Door

    and observe Diluent andSheath Reservoirs.

    YES

    YES

    YES

    2

    1

    5 7URXEOHVKRRWLQJ IP-7 Reagent Empty

    IP-7 Reagent Empty (Page 2 of 5)

  • 8/10/2019 CD 3200 Service Manual

    41/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 27

    1

    From Main Menu,press

    RUN

    CLEAR FAULT

    SyringeMoving

    Properly?

    SyringeFill ing

    Properly

    ?YES

    - Clean/replace syringe(PL-7).

    - Verify Syringe Drive roperat ion(VP-6),(VP-7).

    - Replace Syringe DriverAssembly (PL-7),(RR-11) .

    - Replace MPM(PL-11).

    N O

    - Verify operation and tubingin valves 28 for HGB, 23for WBC. (VP-10)S eeFigure. Replace as

    necessary(PL-4.1),(TM-1) .- Replace SDM2(PL-15)

    - Verify operation and tubing invalves 24 for HGB, 25 for WBC.(VP-10)See Figure. Replace as

    necessary(PL-4.1),(TM-1).- Replace SDM2(PL-15).

    N O

    - Remove Top Cover andobserve Reagent Sensor

    Board. See Figure

    - Inspect and seatconnectors on J1 ofReagent Sensor Board

    and J26 of FCM.

    - Replace FCM(PL-14.1).

    YES

    Floss TubingUp and DownEnsure Tubingis Seated in Slot

    32098

    5 7URXEOHVKRRWLQJ IP-7 Reagent Empty

    IP-7 Reagent Empty (Page 3 of 5)

  • 8/10/2019 CD 3200 Service Manual

    42/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 28

    2

    From Main Menu, pressRUN

    CLEAR FAULT

    - Observe the Di luent and SheathReservoirs and verify that both fi l l tothe Sensor Electrodes andimmediately stop. See Figure

    DiluentNot Fil l ingProperly

    ?

    Sheath

    Not Fil l ingProperly

    ?

    Diluent Fil l ingPast Electrode

    ?

    N O

    N O

    Sheath Fil l ingPast Electrode

    ?

    N O

    5

    Note: The "Di luent/Sheath Empty"message can be generated by the In- l ine

    Di luent/Sheath Sensor, the Di luent

    Reservoir Level Sensor, or the SheathReservoir Level Sensor.

    YES

    6YES

    7YES

    BothFill ing

    Properly?

    4

    Neither

    Fill ingProperly

    ?

    N O

    YES

    N O

    3YES

    YES

    32099

    Fill

    ToHere

    5 7URXEOHVKRRWLQJ IP-7 Reagent Empty

    IP-7 Reagent Empty (Page 4 of 5)

  • 8/10/2019 CD 3200 Service Manual

    43/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 29

    3

    - Remove Top Cover andobserve Reagent SensorBoard. See Figure

    - Inspect and seatconnectors on J1 ofReagent Sensor Boardand J26 of FCM.

    - Replace FCM(PL-14.1).

    4

    - Ensure Di luent/Sheath plumbing

    is free of restrictions.

    - Ensure there is no saline bridge atreservoir connectors.

    - Verify the Vac 1 level(VP-16)and verify Vac 1 retention (VP-15 ).

    - Ensure Vac 1 lines to reservoirs isfree of restrictions.

    - Verify operation of Valves 61, 62,63, and 64(VP-10).Replace asnecessary(PL-4.1).

    - Verify operation of Check Valves.

    Replace as necessary (TM-1).

    - Verify Reservoirs are not crackedor leaking. Replace as necessary

    (PL-4.2).

    - Replace FCM(PL-14.1).

    5

    - Ensure Vac 1 line to reservoiris free of restrictions.

    - Ensure there is no salinebridge at reservoir connector.

    - Verify operation of Valves 62and 63 (VP-10). Replace asnecessary (PL-4.1)

    - Verify operation of CheckValve Replace as necessary(TM-1) .

    - Verify Reservoir is not

    cracked or leaking Replaceas necessary(PL-4.2).

    - Replace FCM(PL-14.1).

    6

    - Ensure Vac 1 line toreservoir is free ofrestrictions.

    - Ensure there is no salinebridge at reservoirconnector.

    - Verify operation of Valves 61

    and 64(VP-10).Replace asnecessary(PL-4.1).

    - Verify operation of CheckValve. Replace asnecessary(TM-1).

    - Verify Reservoir is notcracked or leaking. Replace

    as necessary (PL-4.2).

    - Replace FCM(PL-14.1).

    5 7URXEOHVKRRWLQJ IP-7 Reagent Empty

    IP-7 Reagent Empty (Page 5 of 5)

  • 8/10/2019 CD 3200 Service Manual

    44/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 30

    7

    32100

    - Clean sensor contacts

    (Electrodes andconnector) at top of

    Reservoir .

    - Remove pins 1 and 4and clean and cr imp

    metal contacts so theyare touching. See Figure

    - If pins 1 and 4 are

    damaged, replace withpins 2 and 3.

    - Reseat Sensor Cables(CDM 2 J8 for Diluent, or

    CDM 2 J10 for Sheath.

    - Replace FCM (PL-14.1).

    5 7URXEOHVKRRWLQJ IP-8 Optical Backgrounds Too High (Page 1 of 1)

    IP-8 Optical Backgrounds Too High (Page 1 of 1)

  • 8/10/2019 CD 3200 Service Manual

    45/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 31

    Start

    WBC OnlyToo High

    ?

    - Veri fy Background Counts

    (VP-56).

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy WBC Threshold is 360.

    - Replace WBC Lyse Container.

    - Clean/replace WBC Syringe(PL-7).

    - Ensure WBC Mixing Chamber isbeing rinsed and drained properly.

    - INITIALIZE and PRIME Instrument.

    YES

    - From Main Menu, press

    D IAGNOSTICS

    F 12F 1

    SET POINT ENTRY

    - Veri fy RBC/PLT Threshold is 550.

    - From Main Menu, press

    SPECIAL PROTOCOLS

    MORE

    REAGENT RESERVOIR

    - Follow on-screen instructions and press

    EMPTY DIL/SHEATH

    - Replace Di luent Container and press

    FILL DIL/SHEATH

    - Run multiple backgrounds and verify.

    Backgrounds

    O K

    ?

    Total Call

    YES

    N O

    - Perform the following in order and verifybackgrounds after each act ion taken.

    - Perform Extended Autoclean.

    - Clean Shear Valve(VP-48).

    - Clean and debubble Optical Flow Cel l(VP-44).

    - Clean/replace Di luent and Sample

    Injection Syringes(PL-7).

    - Replace In-line Sheath Filter(PL-4.2).

    - Ensure RBC/PLT Mixing Chamber isbeing rinsed and drained properly.

    - Clean Laser Bench(VP-43).

    N O

    5 7URXEOHVKRRWLQJ IP-9 Histogram Problems (Page 1 of 1)

    IP-9 Histogram Problems (Page 1 of 1)

  • 8/10/2019 CD 3200 Service Manual

    46/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 32

    Start

    - Verify reagents; type andexpiration date.

    - Ensure reagents are atroom temperature.

    - Veri fy Backgrounds(VP-56).

    Prob lem

    Corrected

    ?

    Total Call

    YES

    - Perform the following in order

    and verify histograms aftereach act ion taken.

    - Ensure RBC/PLT and/orWBC Mixing Chambers are

    being r insed and drainedproperly.

    - Perform Extended Autoclean.

    - Clean Shear Valve (VP-48).

    - Clean, debubble,and wet

    Optical Flow Cell (VP-44),(VP-45).

    - Fine align Optical Flow Cell

    (VP-20).

    - Clean Laser Bench(VP-43).

    - Clean/replace Di luent, WBCLyse, and Sample Inject ion

    Syringes (PL-7).

    - Replace In-line Sheath Filter(PL-4.2).

    WBC Only

    ?

    N O

    Prob lemCorrected

    ?

    Total Call

    N O

    - Verify 0oand 10oL ogAmplifiers (VP-28).

    - Verify 0oand 10oGains(VP-26),(VP-27).

    - Replace WBC Lyse for WBC

    problem or Di luent for RBC/PLT problem.

    BackgroundsOK

    ?

    YES

    Go to(IP-8)

    N O N O

    - Veri fy WBC Gains(VP-23).

    YES

    YES

    5 7URXEOHVKRRWLQJ IP-10 Controls Out

    IP-10 Controls Out (Page 1 of 2)

  • 8/10/2019 CD 3200 Service Manual

    47/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 33

    Start

    - Verify reagents; type andexpiration date.

    - Ensure reagents are at roomtemperature and have not

    been previously frozen.

    - Ensure Controls andCal ibrators have not been

    frozen or heat stressed.

    - Check Control Lot #

    - Use current Control InsertSheet to ver i fy Q.C means/l imits are correct if customerhas not establ ished their ownmeans.

    - From Main Menu, press

    SETUP

    Q.C. SETUP MENUQ.C. LIMITS

    - Edit as necessary.

    - Run Q.C.

    Q.C. OK

    ?

    - Perform Autoclean.

    - From Main Menu, press

    SPECIAL PROTOCOLS

    MORE

    AUTOCLEAN

    - Veri fy Backgrounds(VP-56).

    Total Call

    - Open fresh CommercialControl , and run Q.C.

    Al lBackgrounds

    O K?

    Go to(IP-8) .

    ResultsAcceptable

    ?

    Total Call

    1

    N O

    YES

    YES

    N O

    YES

    N O

    5 7URXEOHVKRRWLQJ IP-10 Controls Out

    IP-10 Controls Out (Page 2 of 2)

  • 8/10/2019 CD 3200 Service Manual

    48/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 34

    1

    - Clean Shear Valve(VP-48).

    - Recalibrate out-of-

    range parameter(s)(VP-58).

    - Verify System Precision

    - From Main Menu, press

    RUN

    SPECIMEN TYPE

    - Select empty Q.C. Fi le

    Q.C. SPECIMEN

    - Review CV%(VP-57).

    Precis ionIn Spec

    ?

    YES

    - Determine which

    parameter(s) are out-of-range, and go to(IP-15)or ( IP-16) for HGB andcorrect.

    - For indices (MCH or

    MCHC), determine whichparameter (HGB, MCV,or RBC is imprecise andtroubleshoot accordingly.

    MCH = HGB X10

    RBC

    MCHC = HGB X 100

    MCV X RBC

    NO

    5 7URXEOHVKRRWLQJ IP-11 Vacuum Accumulator #1 Wet

    IP-11 Vacuum Accumulator #1 Wet (Page 1 of 2)

  • 8/10/2019 CD 3200 Service Manual

    49/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 35

    Start

    - Reinitialize AnalyzerFrom Main Menu:

    D IAGNOSTICS

    MORE

    INITIALIZATION

    - Veri fy that Waste Chamber 2 isnot overflowing. If overflowing,verify Pressure Valve 18(VP-10).

    Replace as necessary (PL-4.1).Check output Check Valve.Replace as necessary (TM-1) .

    - Ensure operat ion of WasteChamber 2 Trap and Trap OutputCheck Valve. Replace asnecessary (PL-4.2)and (TM-1) .

    Liquid ExitingAnalyzer

    ?

    Press:

    MORE

    - Observe the Waste Line intoExternal Waste Container andpress:

    DRAIN ACCUMULATORS

    1

    - Perform Vacuum

    Accumulator #2 RinsingProcedure (VP-60)

    N O

    Prob lemCorrected

    ?

    Total Call

    YES

    N O

    YES

    - Perform VacuumAccumulator #2 RinsingProcedure (VP-60)

    5 7URXEOHVKRRWLQJ IP-11 Vacuum Accumulator #1 Wet

    IP-11 Vacuum Accumulator #1 Wet (Page 2 of 2)

  • 8/10/2019 CD 3200 Service Manual

    50/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 36

    1

    - Remove connector fromJ11 on VPM and checkresistance betweenconnector pins. Must be

    greater than 20Megohms. See Figure

    > 20 Megohms

    ?

    - Replace VPM(PL-5)

    - Replace FCM(PL-14.1)

    YES

    - Check Cable and replace

    if necessary(PL-5)

    - Replace Accumulator(PL-5)

    NO

    V P MB D

    J1 0Vac 1 Acc

    J1 1Vac 2 Ac

    32003

    5 7URXEOHVKRRWLQJ IP-12 Vacuum Accumulator #2 Wet

    IP-12 Vacuum Accumulator #2 Wet (Page 1 of 2)

  • 8/10/2019 CD 3200 Service Manual

    51/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 37

    Start

    - Reinitialize Analyzer

    From Main Menu:

    D IAGNOSTICS

    MORE

    INITIALIZATION

    - Veri fy that Waste Chamber 2 is

    not overflowing. If overflowing,verify Pressure Valve 18(VP-10).Replace as necessary(PL-4.1).Check output Check Valve.Replace as necessary (TM-1) .

    - Ensure operat ion of WasteChamber 2 Trap and Trap OutputCheck Valve. Replace as

    necessary(PL-4.2)an d(TM-1) .

    Liquid ExitingAnalyzer

    ?

    Press:

    MORE

    - Observe the Waste Line intoExternal Waste Container and

    press:

    DRAIN ACCUMULATORS

    1

    - Perform VacuumAccumulator #2 RinsingProcedure (VP-60)

    N O

    Prob lemCorrected

    ?

    Total Call

    YES

    N O

    YES

    - Perform VacuumAccumulator #2 Rinsing

    Procedure (VP-60)

    5 7URXEOHVKRRWLQJ IP-12 Vacuum Accumulator #2 Wet

    IP-12 Vacuum Accumulator #2 Wet (Page 2 of 2)

  • 8/10/2019 CD 3200 Service Manual

    52/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 38

    1

    - Remove connector fromJ11 on VPM and checkresistance betweenconnector pins. Must be

    greater than 20Megohms. See Figure

    > 20 Megohms

    ?

    - Replace VPM(PL-5)

    - Replace FCM(PL-14.1)

    YES

    - Check Cable and replace

    if necessary (PL-5)

    - Replace Accumulator(PL-5)

    NO

    V P MB D

    J1 0Vac 1 Acc

    J1 1Vac 2 Ac

    32003

    5 7URXEOHVKRRWLQJ IP-13 HGB Problems

    IP-13 HGB Problems (Page 1 of 2)

  • 8/10/2019 CD 3200 Service Manual

    53/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 39

    Start

    - Veri fy proper Reagents andExpiration Date.

    From Main Menu, Press:

    RUN

    SPECIMEN TYPE

    FRAGILE WBC

    - Run a Normal Control fivet imes and record the NOC

    and HGB values.

    - Compare the NOC and HGBresults to the Data Sheet.

    "0", Low, High, orErratic

    ?

    - Remove, inspect, and clean Shear

    Valve.

    - Perform Vacuum & Pressure Leve lVerification/Adjustment (VP-16).

    - Verify tubing and operation of Valves51, 28, 24, 33, 93, and 26(VP-10)and replace as necessary (PL-4.1).

    - Verify operation of HGB Lyse andRBC Di luent Syr inges(VP-6),(VP-7)and replace as necessary (PL-7).

    - Perform HGB CleaningProcedure(VP-49).

    From Main Menu, Press:

    D IAGNOSTICS

    RAW DATA SUMMARY

    - Run f ive background countsand observe the HGB Sampleand Reference Readings.

    Both

    HGB Only

    1

    5 7URXEOHVKRRWLQJ IP-13 HGB Problems

    IP-13 HGB Problems (Page 2 of 2)

  • 8/10/2019 CD 3200 Service Manual

    54/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 40

    1 2

    Total Call

    - Perform HGB Adjustment

    (VP-5)

    3

    4

    - Verify tubing, operation (VP-10),and f low through Valve 94 andreplace as necessary (TM-1) ,(PL-4.1).

    5

    - Verify connector at J18 of FCM.

    - Replace HGB Flow Cel l (PL-4.2).

    - Replace FCM (PL-14.1).

    - Replace CPU/DCM (PL-13).

    Select path from Chart below.

    Reference Sample

    1 95 0- 20 50 1 95 0- 20 50

    < 1950 < 1950

    > 2050 >2050

    Errat ic 1950-2050

    1600 1950-2050

    Errat ic Erratic

    0 0

    2

    3

    3

    4

    5

    4

    5

    5 7URXEOHVKRRWLQJ IP-14 X-B Errors

    IP-14 X-B Errors (Page 1 of 3)

  • 8/10/2019 CD 3200 Service Manual

    55/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 41

    RBC RBC MC VM CV HG B HGB

    NA NA NANA

    NANA

    X-B Parameter

    MCV

    M CH

    M C H C

    Effect Of RBC, MCV, & HGB On X-B ParametersStart

    - Clean Fan Filters and verify all Fans are runningproperly.

    - Ensure there is at least 6" of clearance behind andon the left side of Instrument.

    - Ensure Instrument is not subject to drastictemperature variations (direct sunlight, heating/airconditioning ducts).

    - If Fan Filters were clogged and then cleaned, allow a15 minute temperature stabilization period.

    - Review patient result in Datalog, and look for groups( 3 or more) of abnormal MCV, RBC, or HGB.

    -I f MCV shif ts, ask Customer for possible causes,such as Oncology or Geriatric Patients.

    I f MCH or MCHC errors:

    - See table to determine which parameter (HGB,RBC, MCV) is bad.

    Note: HCT = RBC X MCV/10

    MCH = HGB X 10/RBC

    MCHC = (HGB X 100)/(RBC X MCV)

    - Review Q.C. Logs for all 3 Control Levels.

    - Veri fy RBC, HGB, and MCV Control Levels are notat the high or low end of range.

    Controls at orNear 2 SD

    of mean

    ?

    Go to (IP-10)

    Controls Out.

    YES

    - Review Customer Maintenance Log.

    - Review Maintenance Procedures andDates.

    - Review X-B Results.

    - Determine i f Maintenance wasperformed just before X-B Shift.

    N O

    1

    5 7URXEOHVKRRWLQJ IP-14 X-B Errors

    IP-14 X-B Errors (Page 2 of 3)

  • 8/10/2019 CD 3200 Service Manual

    56/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 42

    1

    Shift

    After

    Maintenance

    ?

    - Repeat Maintenance

    Proceduresperformed before X-B

    Shift.

    - Monitor X-B for 5 DataPoints (100 Samples).

    YES

    X-B In

    Range

    ?

    Total Call

    YES

    - Verify Reagent, Type and

    Expiration Date.

    Reagent

    Change

    Required?

    2

    NO

    - Monitor X-B for 5

    Data Points (100

    Samples).

    X-B In

    Range

    ?

    YES

    Total Call

    YES 2

    NO

    5 7URXEOHVKRRWLQJ IP-14 X-B Errors

    IP-14 X-B Errors (Page 3 of 3)

  • 8/10/2019 CD 3200 Service Manual

    57/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 43

    2

    - Verify System Precision

    - From Main Menu, press

    R UN

    SPECIMEN TYPE

    - Select empty Q.C. File

    Q.C. SPECIMEN

    - Run one Patient 10

    times.

    - Review CV% (VP-57) .

    Al l

    Parameters

    OK?

    - Verify Calibrator Insert

    Sheet matches Current

    Calibrators.

    - Perform SystemCalibration (VP-58) .

    - Determine which

    parameter(s) are out-of-

    range, and go to (IP-15) or

    (IP-16) for HGB and correct.

    NO

    YES

    5 7URXEOHVKRRWLQJ IP-15 Optical Parameters Imprecision

    IP-15 Optical Parameters Imprecision (Page 1 of 3)

  • 8/10/2019 CD 3200 Service Manual

    58/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 44

    Start

    - Clean Fan Filters and verify all Fans are

    running properly.

    - Ensure there is at least 6" of clearancebehind and on the left side of Instrument.

    - Ensure Instrument is not subject to drastictemperature variations (direct sunlight,heating/air conditioning ducts).

    - If Fan Filters were clogged and thencleaned, al low a 15 minute temperaturestabilization period.

    - Perform Autoclean

    From Main Menu, press

    SPECIAL PROTOCOLS

    MORE

    AUTOCLEAN

    - Clean Shear Valve(VP-48).- Manually clean, debubble, and wet Optical

    Flow Cell (VP-44),(VP45) .

    - Clean Optics Bench(VP-43).

    - Replace In-line Sheath Filter (PL-4.2).

    - Verify System Precision.

    - From Main Menu, press

    RUN

    SPECIMEN TYPE

    - Select empty Q.C. File

    Q.C. SPECIMEN

    - Run one Pat ient 10 t imes.

    - Review CV%(VP-57).

    Prob lem

    Corrected?

    Total Call

    YES

    Al l

    Imprecise

    ?

    N O

    3

    1

    Backgrounds

    OK

    ?

    - Veri fy Background Counts

    (VP-56).

    YES

    Go to(IP-8).N O

    2

    RBC/PLTOnly

    ?

    N O

    N O

    YES

    YES

    5 7URXEOHVKRRWLQJ IP-15 Optical Parameters Imprecision

    IP-15 Optical Parameters Imprecision (Page 2 of 3)

  • 8/10/2019 CD 3200 Service Manual

    59/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 45

    1

    W B C

    Only

    ?

    NOC

    Only

    ?

    NO

    - Ensure HGB Mixing

    Chamber is being r insed

    and drained properly.

    Total Call

    NO

    YES

    - Verify samples are being staged to

    Optical Flow Cell by Peri-pump.Troubleshoot as necessary.

    - Verify Sheath Reservoir is fil l ing

    properly.

    - Verify Pressure 2 (VP-16)an d

    proper Sheath flow to Optical FlowCell. Troubleshoot as necessary.

    - Verify operation of Sample

    Injection Syringe(VP-6) ,(VP-7).

    Troubleshoot as necessary.- Replace Optical Flow Cell (PL-10) .

    - Replace Shear Valve Sections(PL-9.1) .

    - Replace MAM (PL-12).

    - Replace SPM (PL-10).

    - Replace CPU/DCM (PL-13).

    - Replace WBC Lyse Container.

    - From Main Menu, press

    DIAGNOSTICSF 12

    F 1

    SET POINT ENTRY

    - Verify WBC Threshold is 360.

    - Ensure WBC Mixing Chamber is

    being rinsed and drainedproperly.

    - Verify operation of WBC Lyse

    Syringe (VP-6) ,(VP-7) .Troubleshoot as necessary.

    - Verify/adjust WBC Gain (VP-23).

    - Replace Shear Valve Sections

    (PL-9.1) .

    - Replace MAM (PL-12).

    - Replace SPM (PL-10).

    - Replace CPU/DCM (PL-13).

    YE S

    2

    5 7URXEOHVKRRWLQJ IP-15 Optical Parameters Imprecision

    IP-15 Optical Parameters Imprecision (Page 3 of 3)

    3

  • 8/10/2019 CD 3200 Service Manual

    60/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 46

    3

    - Replace Diluent Container.

    - From Main Menu, press

    D IAGNOSTICS

    F 12

    F 1

    SET POINT ENTRY

    - Veri fy RBC/PLT Threshold is 550.

    - Ensure RBC/PLT Mixing Chamber isbeing rinsed and drained properly.

    - Verify 0oand 10oLog Amplifiers(VP-28).

    - Verify 0oand 10oGain(VP-26),(VP-27).

    - Replace Shear Valve Sect ions (PL-9.1).

    - Replace MAM (PL-12).

    - Replace SPM (PL-10).

    - Replace CPU/DCM(PL-13).

    5 7URXEOHVKRRWLQJ IP-16 HGB Imprecision (Page 1 of 1)

    IP-16 HGB Imprecision (Page 1 of 1)

  • 8/10/2019 CD 3200 Service Manual

    61/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300

    August 19982 - 47

    Start

    - Clean Fan Filters and verify all Fans are

    running properly.- Ensure there is at least 6" of clearance

    behind and on the left side ofInstrument.

    - Ensure Instrument is not subject todrastic temperature variations (direct

    sunlight, heating/air conditioning ducts).

    - If Fan Filters were clogged and thencleaned, al low a 15 minute temperature

    stabilization period.

    - Clean Shear Valve(VP-48).

    - Manually clean HGB Flow Cell (VP-49).

    - Verify System Precision.

    - From Main Menu, press

    RUN

    SPECIMEN TYPE

    - Select empty Q.C. File

    Q.C. SPECIMEN

    - Run one Patient 10 times.

    - Review HGB CV%(VP-57).

    HGB CVsOK

    ?

    - Ensure HGB Flow Cel l is being

    rinsed, drained, and ventedproperly. Verify Valves 51, 94, 93,(VP-10)and tubing . Replace asnecessary (PL-4),(TM-1) .

    - Clean HGB Lyse Syr inge.

    - Verify HGB Lyse Syringe (VP-6),

    (VP-7). Replace as necessary(PL-7).

    - Inspect and re-seat Cable at J18of FCM(PL-4.2).

    - Replace HGB Flow Cel l (PL-4.2).

    - Replace FCM(PL-14.1).

    - Replace CPU/DCM(PL-13).

    Total Call

    YES

    - Replace HGB Lyse

    Container>

    HGB CVsOK

    ?

    Total Call

    N O

    YES

    N O

    N O

    57URXEOHVKRRWLQJ IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 1 of 9)

    Start

  • 8/10/2019 CD 3200 Service Manual

    62/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300

    August 1998 2 - 48

    Start

    - Reinitialize Analyzer

    From Main Menu:

    D IAGNOSTICS

    MORE INITIALIZATION

    - Remove Left Side Cover and

    extend Air Supply(RR-7) toobserve pumps. (Vac-Front,Pres-Rear)

    Both

    ConstantlyRunning

    ?

    4

    Pres

    Constant lyRunning

    ?

    N O

    1

    VacConstant ly

    Running

    ?

    N O

    N O

    - Disconnect tubing frompositions 6 and 7 on AirSupply Manifold, and

    observe Pumps and VPM

    Pump LEDs. See Figures

    5

    6

    - Observe LEDs. SeeFigure

    YES

    YES

    YES

    1

    2

    3

    4

    5

    6

    7

    12PSI

    9PSI

    5PSI12"Hg8

    "Hg

    12PSI

    12"Hg

    TO

    FLOW

    PANEL

    To S L

    ToS L

    32004

    32005

    57URXEOHVKRRWLQJ IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 2 of 9)

  • 8/10/2019 CD 3200 Service Manual

    63/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300

    August 1998 2 - 49

    1

    Both LEDsO N

    ?

    - Replace Pump RelayBoard(PL-5)

    N O

    - Reinitialize Analyzer

    From Main Menu:

    D IAGNOSTICS

    MORE

    INITIALIZATION

    - Observe Flow Panel

    valves

    YE S

    All ValvesClose Then

    Open

    ?

    - Check +12V UR and Fuse.

    See Figure- Replace if necessary(PL-1)

    YE S

    Prob lem

    Corrected

    ?

    Total Call

    Y ES

    - Replace Power Supply(PL-1)

    N O

    FuseBlown

    ?

    Check fuse again

    N O 8

    2

    N O

    Y ES

    +12 VNRFuse (Top)

    32006

    57URXEOHVKRRWLQJ IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 3 of 9)

    2

  • 8/10/2019 CD 3200 Service Manual

    64/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 50

    2

    LEDs BrightThen Dim When

    Activated

    ?

    - Whi le observing VPMSolenoid LEDs (SeeFigure) press:

    INITIALIZATION

    3

    YES

    - Check +24V UR (28.9V-37.7V) and Fuse

    (Bottom Fuse)

    - Replace Fuse ifnecessary (PL-1)

    N O

    Prob lem

    Corrected

    ?

    Total Call

    YES

    - Check Fuse again

    FuseBlown

    ?

    NO

    - Replace Power Supply(PL-1)

    N O 8

    YES

    Note: If the +24 UR is approximatelythe same as the +12 UR, the +24VFuse is blown or the +24V Power

    Supply is defective.

    32007

    57URXEOHVKRRWLQJ IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 4 of 9)

    7H

    gToS L

  • 8/10/2019 CD 3200 Service Manual

    65/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 51

    3

    - Pinch tubing as shown inFigure

    PumpsStop

    ?

    - Inspect 12 psi and 12" Hgplumbing andAccumulators. Replaceas necessary(PL-5),

    (TM-2)

    - Inspect Vent Valves 1 and4. Replace as necessary(PL-5)

    - Replace VPM (PL-5)

    N O

    - Unpinch tubing

    - Pinch tubing at positions6 and 7 that supply 12psi and 12" Hg to SampleLoader

    YES

    PumpsStop

    ?

    - Inspect Sample Loaderplumbing and 3-WayValves. Replace asnecessary (PL-18.3),

    (TM-3)

    YES- Ensure Valves 31 and 32,

    97 and 98, 17 and 18 arenot open at the same

    time.

    - Inspect Flow Panel 12 psiand 12" Hg plumbing for

    leaks.

    N O

    Pinch

    1

    2

    3

    4

    5

    6

    12PSI

    9PSI

    5PSI12"Hg

    8"Hg

    12PSI12"H

    TO

    FLOW

    PANEL

    o S

    To S L

    32008

    57URXEOHVKRRWLQJ IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 5 of 9)

    4

  • 8/10/2019 CD 3200 Service Manual

    66/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 52

    - Observe Pressure PumpLED. See Figure Page 1

    LEDON

    ?

    - Replace Pump Relay

    Board (PL-5)

    N O

    - Pinch tubing See Figurethis page

    PumpStops

    ?

    NO

    - Inspect 12 psi plumbingand Accumulator.Replace as necessary(PL-5),(TM-2)

    - Inspect Vent Valve 1.Replace as necessary(PL-5)

    - Replace VPM(PL-5)

    - Unpinch tubing and pinchat position 6 on Manifold.

    YES

    PumpStops

    ?

    - Inspect Sample Loader

    plumbing and 3-WayValves. Replace asnecessary (PL-18.3),(TM-3)

    - Inspect Flow Panel 12 psiplumbing for leaks.

    N O

    YES

    YES

    Pinch

    1

    2

    3

    4

    5

    6

    7

    12PSI

    9PSI

    5P

    SI12"Hg

    8"Hg12PSI12"Hg

    TO

    FLOW

    PANEL

    ToS L

    ToS L

    32009

    57URXEOHVKRRWLQJ IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 6 of 9)

    5

  • 8/10/2019 CD 3200 Service Manual

    67/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 53

    - Observe Vacuum PumpLED. See Figure Page 1

    LEDON

    ?

    - Replace Pump RelayBoard(PL-5)

    NO

    - Pinch tubing. See Figure

    PumpStops

    ?

    N O

    - Inspect 12" Hg plumbingand Accumulator.Replace as necessary(PL-5), (TM-2)

    - Inspect Vent Valve 4.Replace as necessary(PL-5)

    - Replace VPM(PL-5)

    - Unpinch tubing and pinchat position 7 on Manifold.

    YES

    PumpStops

    ?

    - Inspect Sample Loaderplumbing and 3-Way

    Valves. Replace asnecessary (PL-18.3),(TM-3)

    - Inspect Flow Panel 12"Hg plumbing for leaks.

    N O

    YES

    YES

    Pinch

    1

    2

    3

    4

    5

    6

    7

    12PSI

    9PSI

    5PSI12"Hg

    8"Hg

    12PSI12"Hg

    TO

    FLOW

    PANEL

    ToS L

    ToS L

    32010

    57URXEOHVKRRWLQJ

    IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 7 of 9)

    6- Replace VPM

  • 8/10/2019 CD 3200 Service Manual

    68/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 54

    NeitherPumpRuns

    ?

    VAC

    Doesn'tRun

    ?

    Pres

    Doesn'tRun

    ?

    N O

    N O

    LEDsO N

    ?

    - Reseat connectors VPM J18, PRMJ1, J2, J3, J4

    - Check Fuse on PRM. Replace i f

    necessary(PL-5)

    - Veri fy 115VAC to PRM

    - Replace PRM (PL-xx)

    YES

    Replace VPM(PL-5)

    - Reseat connectors VPM J18, PRM

    J1, J3- Check 115VAC across Pins 1 and 2

    on PRM P3. IF present, replace VacPump. I f not, replace PRM(PL-5)

    VAC LED

    O N

    ?

    YES

    Pres LED

    O N

    ?

    YES

    - Reseat connectors VPM J18,PRM J1, J2

    - Check 115VAC across Pins 1 and2 on PRM P2. IF present, replace

    Pres Pump. I f not, replace PRM(PL-5)

    7

    - Replace VPM(PL-5)

    - Replace VPM(PL-5)

    YES

    YES

    YES

    N O

    N O

    N O

    5 7URXEOHVKRRWLQJ IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 8 of 9)

    7

  • 8/10/2019 CD 3200 Service Manual

    69/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 55

    - Re-connect tubing topositions 6 and 7 (6-Red,7-Green)

    - Observe Vacuum andPressure Pumps andt ime the durat ion betweenrun cycles.

    Pres

    Cycles< 2 min

    ?

    Va c

    Cycles< 2 min

    ?

    See Pressure Leaks(IP-18) See Vacuum Leaks (IP-19)

    - Perform Vacuum andPressure LevelVeri f ication/Adjustment(VP-16)

    - Perform Vacuum andPressure RecoveryVerification(VP-17)

    BothOK

    ?

    YES

    - Replace Pump Head

    - If problem persists,

    replace Pump AssemblyPump Head:

    Pressure (PL-5)

    Vacuum (PL-5)

    Pump Assy:

    Pressure (PL-5)

    Vacuum (PL-5)

    YES

    YES

    Total Call

    N O

    N O

    N O

    5 7URXEOHVKRRWLQJ IP-17 Primary Vacuum/Pressure Isolation,

    IP-17 Primary Vacuum/Pressure Isolation, (Page 9 of 9)

    8

  • 8/10/2019 CD 3200 Service Manual

    70/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 56

    - If the Fuse continues to blow,assemblies on the +12V UR or

    +24V UR must be disconnectedone at a time until the shortedassembly is found andreplaced. (See Figures)

    - Power-down instrument and

    connect Meter between TP2-+12V or TP1-+24V and TP4 on

    PSM and remove connectorsuntil short is found

    - Replace defect ive assembly and

    verify proper operation

    CDM 2 & S DM/CDM Power Connec tors

    J3 J1+24V UR

    +24V UR

    +12V UR

    +12V UR

    +24V UR

    +24V UR

    +24V UR

    +12V UR

    +12V UR

    UR RE T

    5

    4

    3

    1

    2

    3

    1

    2

    3

    1

    FromP.S.

    J1/J2

    ToS DM A

    1

    2

    3

    1

    2

    3

    J2

    ToS DM B

    J4

    ToS DM C

    J5

    ToS DM D

    BackJ1

    J2

    J3

    J4

    UR RE T

    UR RE T

    UR RE T

    1

    345

    13

    12

    4

    123

    5

    RET

    PS M

    FC M

    To

    To

    To

    To

    CDM2-J3FCM-J5

    SDM/CDM-J3

    Shear

    ValveDriver-J3

    SDM9-J10

    Note: J4 and J5 are not onS DM /CDM

    32011

    5 7URXEOHVKRRWLQJ IP-18 Pressure Leak Isolation

    IP-18 Pressure Leak Isolation (Page 1 of 7)

    Note: This procedure is designed to isolate leaks in the pressure plumbing. Refer to IP-17 Primary Vacuum/Pressure Isolation for troubleshooting gross pressure failures.

  • 8/10/2019 CD 3200 Service Manual

    71/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 57

    Start

    - Reinitialize AnalyzerFrom Main Menu:

    D IAGNOSTICS

    MORE

    INITIALIZATION

    - Remove Left Side Cover andextend Air Supply (RR-7) toobserve Pres Pump. (Vac-Front, Pres-Rear)

    - Observe Pressure Pumpand time the duration

    between run cycles.

    PumpCycles< 2 min

    ?

    Total Call

    N O

    - Adjust Pressure Levels ifnecessary(VP-16)

    1

    YES

    5 7URXEOHVKRRWLQJ IP-18 Pressure Leak Isolation

    IP-18 Pressure Leak Isolation (Page 2 of 7)

    1

  • 8/10/2019 CD 3200 Service Manual

    72/334

    CELL-DYN 3200 SYSTEM SERVICE MANUAL Part Number 9140300 August 1998 2 - 58

    LEDs BrightThen Dim When

    Activated

    ?

    - Whi le observing