catalogo odournet 2014-10-31
DESCRIPTION
Catalogo Odournet year 2014TRANSCRIPT
Odour measuring technology
from the leader
2014
PRODUCT
CATALOGUE
Index
1. About Odournet
Welcome to the world of Odournet 2
Working Groups with Odournet involvement 3
Odournet - Scopes of accreditation 4
TO-series labs around the world – since 1996 5
Exemplary of an Olfactometry Laboratory 6
2. Odour testing equipment and accessories
Olfactometer TO-series 7
Perceived odour intensity - acetone 11
Laboratory accessories 12
Air supply 17
3. Odour sampling equipment
Sampling devices 18
Sampling hoods 20
On site pre dilution 22
Liquid to gas vaporizer 23
Sample bags 24
4. Software products
TO-series controlling software 25
Odourmap® - odour complaint management 26
iNose – Odour Dispersion Monitoring System 27
1
The specially designed, unique technology is used by more than 250 laboratories, research
institutes, universities, and companies world wide. Reliable, high-quality measuring technology
consists primarily of the robust measurement equipment, well-engineered software as well as
customer service and support. At Odournet, these elements interact in perfect harmony. We are
your one-stop provider of odour measurement technology solutions.
Odournet was not only instrumental in standardising olfactometry, and chaired the technical
working group that established the EN 13725 standard, Odournet members were also key players
in all other relevant odour working and standardisation groups since 1993, which is illustrated on
the following page.
Odournet has also pioneered ISO 17025 accreditation. An overview of our current accreditation
scopes can be found in the second next page.
The expertise of our employees gives us the best knowledge to develop the best customised odour
measurement and sampling equipment in the world. Don’t miss the chance to benefit from this
knowledge and expertise - Be part of it!
All our products come with a 24 months warranty and the first calibration according to EN
13725 and ISO 17025 after 12 months for free.
We are at your service to set up your professional odour laboratory!
Best regards
Marc Andresen
Head of Consumer Relations – Odournet GmbH
Odournet GmbH Fraunhoferstr.13
24118 Kiel
Deutschland
Tel.: +49 (0)431-22012-0
Fax: +49 (0)431-22012-17
Odournet GmbH ∙ Fraunhoferstr.13 ∙ 24118 Kiel
Commerzbank AG
Kto. 0119 942 000 ∙ BLZ 21080050
IBAN DE 44210800500119942000
SWIFT-BIC DRESDEFF 210
USt-Idnr. DE 812100460
Amtsgericht Kiel ∙ HRB 13106 KI
Geschäftsführer:
Bettina Mannebeck
Björn Maxeiner
Simon Rützel-Grünberg
Welcome to Odournet
from experts for experts
Marc Andresen
Dear Madam, dear Sir,
We combine the expertise, experience and innovative
capacity of the two foremost pioneers in developing
modern olfactometry over the past 40 years: ecoma and
Odournet.
Odournet is the world leading supplier for odour
laboratory technologies and odour measurement
equipment, including olfactometers, sampling tools,
software, calibration services, and more.
Most scientific research papers and applied environmental
olfactometry research of the past 20 years are based on
data obtained using our odour measurement equipment.
3
Working Groups with Odournet involvement
1993 - 1995 NVN 2820
Air Quality – Sensory odour measurement using an olfactometer (Netherlands
Normalisation Institute)
1995 - 2003
2012 - now
CEN/TC264/WG2
'Olfactometry‚
EN13725 Air Quality – Determination of odour concentration by dynamic
olfactometry with Ton van Harreveld (CEO Odournet Group) as Convenor of CEN
standardisation working group TC264/WG2 ‘Odours’
2002 - 2005 NVN 2818: 2005 Writing Dutch standard for hedonic tone measurements
since 2003 VDI 3894 - Requested
continuation of team
Emissions an immissions from animal husbandries - Housing systems and
emissions - Pigs, cattle, poultry, horses
2004 - 2011 VDI 3880 Olfactometry – Static sampling
since 2006 CEN/TC 264/WG 27 Air quality - Determination of odour exposure in ambient air by using field inspection
2008 - 2011 ISO/TC 34/SC 12 Sensory
analysis - NEN committee (NL) Dutch standardisation committee on sensory analysis (general)
since 2008 mirror commitee to
ISO/TC 146/SC 6/WG 14
Determination of odour emissions from building products using test chambers ( ISO
16000-28 and ISO 16000-30)
since 2008 VDI 4302 Sensory testing of indoor air and determination of odour emissions from building
products
2009 - 2013 VDI 3884 Olfactometry - Determination of odour concentration by dynamic olfactometry -
Supplementary instructions for application of DIN EN 13725
n/a NF X-43-103 Evolution of field inspection french method, intensity method
n/a AFNOR comittee on EN-13725 French group discussion about EN 13725 revision
2010 - 2013 VDI 3883 Effects and assessment of odours - Conflict management in air pollution abatement
- Fundamentals and application to ambient odour
since 2010 VDI-WG 3885/1 Odour measurements of liquids
since 2011 VDI 3883-4, NA 134-03-10-08 Neighborhood complaints of odor
since 2012 NEN - odours (geurmetingen) Dutch standardization committee for odour measurements (general)
since 2012 VDI NA 134-02-01-12 Calculation of plume rise
since 2012 VDI 4221 Quality assurance - Requirements on interlaboratory tests in immission control -
Requirements on providers of interlaboratory tests
since 2013 NVN 2818 Revision of Dutch standard for hedonic tone measurements
since 2014 VDI 3788 / 2 Environmental meteorology - Odour dispersion modelling - source term
determination via backward modelling -
4
Odournet - Scopes of accreditation
France Cofrac accreditation nº 1-1964.
Germany
DAkkS accreditation nº D-PL-17433-01 and D-EP-17433-01.
In addition, Odournet GmbH is an accredited inspection body for performance tests
and interlaboratory tests according to EN 13725 and for the automobile guideline VDA
270.
Netherlands
RVA accreditations K077 and L403.
In addition, our Amsterdam laboratory has a calibration accreditation for calibrating
dilution apparatus (olfactometers) according to EN 13725 and certification of reference
gases of n-butanol and SF6 in nitrogen according to ISO 6143.
Spain
ENAC accreditation nº 1046/LE2024.
Odournet SL is also an Authorized Control Body (OCA) in the Balearic Islands
according to Decree 104/2010 of 22 September, for the determination of odours.
United
Kingdom
UKAS accreditation nº 2430. ISO 9001 certification, certificate number CA13725.
Odournet UK are also accredited for bioaerosol sampling of stack emissions
in accordance with VDI 4257 Part 2 as per the Environment Agency Technical
Guidance Note M2.
Within Odournet we have pioneered ISO 17025 accreditation, obtaining the first
accreditation certificate on 28 September 1993 for our Amsterdam laboratory.
Currently we run 6 laboratories for olfactometry according to EN 13725 under 'testing'
accreditations, in the Netherlands, Spain, Germany, the United Kingdom and France.
Being accredited by European Accreditation Collaboration members implies that our
laboratory and sampling team accreditations are recognised in all EU countries.
For the up-to-date accreditation scopes, please see the table below:
5
TO-series olfactometer labs around the world – since 1996
Germany 95
Italy 23
Switzerland 8
Czech Republic 8
Spain 6
Austria 7
Poland 5
France 5
Turkey 5
England 3
Belgium 3
Chile 3
Norway 2
Latvia 1
Estonia 1
China 1
Argentina 1
Bangladesh 1
Lithuania 1
Israel 1
South Africa 1
USA 1
Hungary 3
Russia 3
Denmark 3
Sweden 3
Hong Kong 3
Korea 2
Mexico 2
Portugal 2
Ireland 2
Australia 2
Finland 2
Singapore 1
Exemplary Floor Plan of an Olfactometry Laboratory
6
Required air supply 115 l/min, oil-free
Dilution range 22 (1:4) - 216 (1:65,536)
Interval steps Factor 2
Dilution accuracy < 5%
Response/
setting time
< 0.1 sec
Required
sample volume
High concentrations: 1-2 liter
Low concentrations: 6-8 liter
Maximum calibration
interval
12 months
(according to
EN 13725:2003)
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Scope of supply:
• Filter system
• Robust transport box
• Heating system to prevent contamination
• Full version software:
Dynamic olfactometry and intensity/hedonic tone
on USB stick
• 5 m compressed air connecting PTFE pipeline
• Compressed air connection (8mm external)
• 50 x 8l bags
• Calibration certificate according to ISO 17025
Article no.: TO810000
The olfactometer TO8 meets all requirements
of a dilution device for olfactometric
measurements according to EN 13725:2003. It
uses a unique and special designed dilution
system for odour samples.
The main application of the TO8 is to measure
odour concentration, providing a full record of
the data obtained, according to EN 13725:2003.
The measurement process is fully automated.
You can also assess hedonic tone or intensity
(VDI 3882).
The compact size and transportability allows
you to measure on site in a mobile laboratory, if
required. This option allows for reducing sample
storage time to a minimum.
Back to content 7
Olfactometer TO8
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
a) assembled
b) transport size
a) 650 x 650 x 470 mm
b) 650 x 650 x 270 mm
Weight: 17 kg
Dilution principle Gas jet pumps and
calibrated precision orifices
for sample dosage
Vent for surplus
diluted odour
Outlet air via active carbon
filter
Number of panel
members working
simultaneously
4
Maximum panel
size
(software setting)
16
Analyses time for
one sample
(EN13725:2003)
5 min
Fully automatic, 100 % compliant to EN 13725:2003
Unique & special designed dilution system
Software includes project & panel database
Yes-/No-Method, Intensity & Hedonic Tone
Most sold olfactometer worldwide
Fastest analysing time, Fastest setup time
Mobile & plug and play technology
Heated dilution system to prevent contamination
Article no.: TO890000
The olfactometer TO9 meets all prerequisites
of a dilution device for olfactometric
measurements following EN 13725:2003. Using
the TO9 permits a complete description of an
odour sample.
The TO9 meets all functions and benefits of the
TO8 plus the Dual-Forced-Choice-Method (D-F-
C). The max. dilution expands to 217 which
allows you to measure higher odour
concentrations.
8
Olfactometer TO9
All functions and
benefits of the TO8
Plus Dual-Forced-
Choice-Method (D-F-C)
Fastest analysis (Y/N-Method about 5 min, D-F-C-Method about 7 min)
Max. dilution 217
Scope of supply:
• Filter system
• Robust transport box
• Heating system to prevent contamination
• Full version software:
Dynamic olfactometry and intensity/hedonic tone
on USB stick
• 5 m compressed air connecting PTFE pipeline
• Compressed air connection (8mm external)
• 50 x 10l bags
• Calibration certificate according to ISO 17025
Back to content
Required air supply D-F-C:180 l/min, oil-free
Dilution range (D-F-C) 2³ ( 1:8) - 217 ( 1:131,072)
Interval steps Factor 2
Dilution accuracy < 5%
Response/
setting time
< 0.1 sec
Required
sample volume
High concentrations: 2-3 liter
Low concentrations: 8-10 liter
Maximum calibration
interval
12 months
(according to
EN 13725:2003)
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
a) assembled
b) transport size
a) 650 x 650 x 470 mm
b) 650 x 650 x 270 mm
Weight: 18 kg
Dilution principle Gas jet pumps and
calibrated precision orifices
for sample dosage
Vent for surplus
diluted odour
Outlet air via active carbon
filter
Number of panel
members working
simultaneously
4
Maximum panel
size
(software setting)
16
Analyses time for
one sample
(EN13725:2003)
5 min
Article no.: TO880000
The olfactometer TO8-8 meets all prerequisites
of a dilution device for olfactometric
measurements following EN 13725:2003. Using
the TO8-8 permits a complete description of an
odour sample..
The TO8-8 offers the possibility to work with 8
panel members, allowing a higher accuracy in
the measurement results. The whole
measurement program is computer controlled
and runs automatically. It includes several
measurement sequences, recovery breaks for
panel members and clean air flushing functions
for the dilution unit.
9
Olfactometer TO8-8
Yes/No method
according to
EN 13725:2003
No Intensity and
Hedonic Tone
Up to 8 panel
members working
simultaneously
Laboratory use for studies
with high number of panel
members, higher statistic
values
Scope of supply:
• Filter system
• Robust transport box
• Heating system to prevent contamination
• Full version software:
Dynamic olfactometry and intensity/hedonic tone
on USB stick
• 5 m compressed air connecting PTFE pipeline
• Compressed air connection (8mm external)
• 50 x 10l bags
• Calibration certificate according to ISO 17025
Back to content
Required air supply D-F-C:180 l/min, oil-free
Dilution range (D-F-C) 2³ ( 1:8) - 217 ( 1:131,072)
Interval steps Factor 2
Dilution accuracy < 5%
Response/
setting time
< 0.1 sec
Required
sample volume
High concentrations: 2-3 liter
Low concentrations: 8-10 liter
Maximum calibration
interval
12 months
(according to
EN 13725:2003)
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
a) assembled
b) transport size
a) 900 x 900 x 470 mm,
b) 650 x 650 x 270 mm
Weight: 22 kg
Dilution principle Gas jet pumps and
calibrated precision orifices
for sample dosage
Vent for surplus
diluted odour
Outlet air via active carbon
filter
Number of panel
members working
simultaneously
8
Maximum panel
size
(software setting)
16
Analyses time for
one sample
(EN13725:2003)
5 min
Article no.: TO820000
The olfactometer TO8s meets all prerequisites
of a dilution device for olfactometric
measurements following EN 13725:2003. Using
the TO8s permits a complete description of an
odour sample.
The TO8s meets all functions of the TO8. It is
specially designed for travelling around the
world. This makes it possible to perform
measurements around the world and to
undertake international projects. Weighing only
21 kg (packed in a trolley) it is also permitted on
most flights. The TO8s is the smallest 4-place
olfactometer in the world.
10
Olfactometer TO8s
All functions and
benefits of the TO8
Extra small and light
olfactometer
(about 9 kg)
Easy to build up in 10
min on site
Specially designed for
travelling and
measurements on site
in a mobile lab
Scope of supply:
• Filter system
• Robust transport box
• Heating system to prevent contamination
• Full version software:
Dynamic olfactometry and intensity/hedonic tone
on USB stick
• 5 m compressed air connecting PTFE pipeline
• Compressed air connection (8mm external)
• 50 x 8l bags
• Calibration certificate according to ISO 17025
Back to content
Required air supply 115 l/min, oil-free
Dilution range 22 (1:4) - 216 (1:65,536)
Interval steps Factor 2
Dilution accuracy < 5%
Response/
setting time
< 0.1 sec
Required
sample volume
High concentrations: 1-2 liter
Low concentrations: 6-8 liter
Maximum calibration
interval
12 months
(according to
EN 13725:2003)
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
a) assembled
b) transport size
a) 420 x 420 x 410 mm,
b) 400 x 400 x 220 mm
Weight: 8.5 kg
Dilution principle Gas jet pumps and
calibrated precision orifices
for sample dosage
Vent for surplus
diluted odour
Outlet air via active carbon
filter
Number of panel
members working
simultaneously
4
Maximum panel
size
(software setting)
16
Analyses time for
one sample
(EN13725:2003)
5 min
Air supply: 3 - 5 bar cleaned compressed
air
Accessories (optional): External compressor (cleaned
compressed air)
Filter system (three stages)
Test gas (e.g. acetone;
concentration 2500 mg/m3)
PureSniff XL device
Sample bags
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Article no.: AC000001
The Acet-ON allows the comparison of the
intensity of a sample with different specified
intensities of the reference substance acetone,
and hence, the determination of the perceived
intensity Π (pi).
The use of comparative sources ensures that
all panel members evaluate the perceived
intensity based on the same reference scale.
The Acet-ON is fully compliant to the
requirements of the standard ISO 16000-
28:2012 as well as the guideline VDI 4302
(draft). The device allows the presentation of 6
different acetone concentrations ranging from 0
pi to 15 pi, as well as 1 blank (odourless
compressed air), and makes it easier for the
panel members to assess the perceived
intensity of unknown samples.
The use of CRM gas allows traceability of the
presented concentrations, which results in a
high accuracy of acetone concentrations.
Another advantage of the Acet-ON is the use of
only one nose mask, which allows gas release
only when required meaning room air
contamination is kept to a minimum.
11
Acet-ON
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
425 x 275 x 170 mm,
Weight: 4 kg
Dilution principle Gas jet pumps and
calibrated precision orifices
Volume flow: 0.7 l /sec (pre-set by
manufacturer)
100% compliant to
ISO 16000-28:2012
and VDI 4302 (draft)
High accuracy of
acetone concentrations
Low room air
contamination
6 different pi-
concentrations and
1 blank
Compact/transportable
Low gas consumption
The Acet-ON in combination with the
direct sniffing device PureSniff XL
Scope of supply:
• 1 nose funnel
• 5 m compressed air connecting PTFE pipeline
Optional:
• Filter system
• PureSniff XL
Back to content
Volume flow 0.7 l /sec (pre-set by
manufacturer)
Air supply 2 – 5 bar
Power supply External power supply (12V,
65W max)
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Article no.: AC000002
The PureSniff XL allows the standardised
presentation of a sample:
As the operator triggers a volume flow, the
smell sample streams out of the sample bag
and passes through a funnel (diffuser) to the
human nose. That way, each olfactory
tester/examiner gets identical sample air with a
standardised volume flow and a constant
presentation time.
Additionally the PureSniff XL can be used for
indoor air sampling and is fully compliant to the
requirements of the standard ISO 16000-
28:2012 as well as the guideline VDI 4302
(draft).
The PureSniff XL consists of a sealed tube
made of clear PVC for visual checks.
The scope of supply includes the PureSniff XL,
a diffuser and a separate power-pack. Ready to
use sample bags (60l), with PTFE adapter and
corks can be purchased as additional extras.
12
Direct Sniffing PureSniff XL
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
420 x 425 x 1200 mm
Weight: 12.8 kg
Sample volume 60 l max
Duration of
presentation:
2.5 sec
Constant flow
Standardised sample
presentation
PTFE and stainless
steel (EN 13725:2003)
100% compliant to
ISO 16000-28:2012
and VDI 4302 (draft)
Easy to use and
connect
Intensity/hedonic
tone/liking
Direct evaluation of an undiluted sample using
the PureSniff XL
Scope of supply:
• PureSniff with 1 cylinder
• 1 nose funnel
• Remote control
Optional:
• Sampling function
• Power pack
Back to content
Duration of presentation 2.5 sec
Volume flow 20 l/min
Air supply 2 – 5 bar
Power supply External power supply (12V,
65W max)
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Article no.: TO890004
PureSniff consists of 2 sealed tubes, which can
be provided with sample bags independently of
each other. The device works as follows:
As the operator triggers a volume flow, the
smell sample is pushed out of the sample bag
and passes through a funnel (diffuser) to the
human nose. That way, each olfactory
tester/examiner gets an identical sample with a
standardized volume flow and a constant
provision time.
As default setting, the sample is offered for
about 2.5 sec at a volume flow of 20l/min. Using
an 10l bag filled to approx. 70%, one sample
serves for about 10 presentations. For visual
checks the cylinder is made of clear PVC.
13
Direct Sniffing PureSniff II
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
274 x 430 x 860 mm
Weight: 9 kg
Sample volume 10 l maximum
Constant flow
Standardised sample
presentation
PTFE and stainless
steel (EN 13725)
Easy to use and
connect
Intensity / hedonic
tone / liking
Ready for comparable
tests (Sample A
against Sample B)
Direct evaluation of an undiluted sample
using the PureSniff II
Scope of supply:
• PureSniff with 2 cylinders
• 2 nose funnels
• Remote control
Back to content
Duration of presentation 2.5 sec
Volume flow 20 l/min
Air supply 2 – 5 bar
Power supply External power supply (12V,
65W max)
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Article no.: TO900001
PureSniff III consists of 3 sealed tubes, which
can be provided with sample bags
independently of each other. The device works
as follows:
As the operator triggers a volume flow, the
smell sample is pushed out of the sample bag
and passes through a funnel (diffuser) to the
human nose. That way, each olfactory
tester/examiner gets an identical sample with a
standardized volume flow and a constant
provision time.
As default setting, the sample is offered for
about 2.5 sec at a volume flow of 20l/min. Using
an 10l bag filled to approx. 70%, one sample
serves for about 10 presentations. For visual
checks the cylinder is made of clear PVC.
14
Direct Sniffing PureSniff III
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
275 x 660 x 860 mm
Weight: 12 kg
Sample volume 10 l maximum
Constant flow
Standardised sample
presentation
PTFE and stainless
steel (EN 13725)
Easy to use and
connect
Intensity / hedonic
tone / liking
Ready for tripple
testing (Sample A vs
Sample B vs Sample C)
Direct evaluation of an undiluted sample using
the PureSniff III
Scope of supply:
• PureSniff with 3 cylinders
• 3 nose funnels
• Remote control
Back to content
Standard deviation of
the dilutions
< 5%
Response time < 1 s
Set-up time < 1 s
Maintenance interval 12 months (according to
EN 13725:2003)
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Scope of supply:
• 2 x 3m compressed air connecting PTFE pipeline
• 1 T-unit 8 mm
Optional:
• Sample bag (8 l) with PTFE adapter and PTFE
sealings.
• Underpressure sampling device (20,000 Pa).
Article no.: EPD10000
The EPD is an easy to use pre-dilution system
for highly-concentrated odour samples. It has
been developed to increase the measuring
range at high odorant concentrations for
olfactometers.
In addition, it protects the olfactometer from
self-contamination.
The EPD is designed for use in olfactometry
laboratories and gives the operator the
opportunity, together with a TO-series
olfactometer, to process odour samples with a
concentration of up to 6 x 106 OUE/m3 with just
one additional mixing process.
Due to the dilution, which precedes the
olfactometric measurement, virtually all known
odour emissions from environmental protection
and industrial markets can be olfactometrically
measured.
15
Pre Dilution Device EPD
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
45 x 105 x 250 mm,
Weight: 2kg
Dilution principle Gas jet pumps
Control
mechanism of
volumetric flow
Gas jet pumps
Dilution ranges Standard: 1:10/1:100, 1:2
to 1:200 also possible
Manual pre dilution
system for laboratory
use
Portable
Low noise level
PTFE and stainless
steel (EN 13725:2003)
Pre-set dilutions 1:10
and 1:100
Easy to use and
connect
Fast pre dilution
Low air and sample
consumption
Back to content
Filling materials Silica gel, activated carbon,
cotton, micro-fine filter
Maintenance interval Depending on supply air
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Scope of supply:
• Filter System
• 3 m PA tube 8x6x1
• 2 m PTFE tube 8x6x1
• 1 pressurized air gun
• 1 T-unit 8 mm
Optional:
• Filter material exchange set
Article no.: F0010000
The filter system consists of a PCA tube
(Makrolon®) with a stainless steel lid and base.
This unit is available with four feet for mobile
application or with a wall bracket for use in
laboratories. The filter system connects an oil-
free compressor or a bottle of compressed gas
and the olfactometer. It is filled with silica gel
(3), activated carbon (4), filter cotton (5) and a
microfine filter (6) for the separation of moisture,
organic compounds and dust. The aim is to
achieve uncontaminated reference-air / zero-air.
The inlet (2) and outlet (7) for the air are 8 mm
compressed air quick connectors.
The system has been tested to 5 bar. The
pressure limit valve (1) limits the working
pressure of the TO series olfactometer to 4 bar
which secures the system.
16
Filter System
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
510 x 340 x 135 mm
Weight: 4 kg
Pressure max. 5 bar, pressure limit
valve
Filter system for
laboratory use
Portable
Easy to use
Low maintenance
Connection via oil free
compressor or
compressed gas
possible
Separation of
moisture, organic
compounds and dust
Back to content
For running an olfactometry laboratory an oil-
free compressor is necessary to provide
pressurized air to the olfactometer.
Odournet offers medical air compressors for
the olfactometers of the TO-series.
17
Compressors
Compressors for TO-
series olfactometer
Type Volt/Ph dB (A) m³/h bar l/min Watt Kg Olfactometer
Dr. Sonic 320-
50V-ES-3
230/1 62 14 8.2 235 2200 72 TO8, TO8s,
TO9, TO8-8
Back to content
Article no.: Z0020038
Filling time 30, 15, 10 or 5 minutes
Suction pressure 50 mbar
Battery NiCd 6V 5Ah
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Scope of supply:
• Calibrated nozzle for 30-minutes sampling
• Battery charger
• PTFE flexible probe tube 2 m
• Countdown timer
• 5 x 10l sample bags
Accessories (optional):
• Stainless steel probe
• Calibrated nozzle for different volumetric flows
(i.e. filling times)
• Sample bags 10l
Article no.: CSD10000
The CSD30 is a sampling device for collecting
air (odour) grab samples in sampling bags, with
a default sampling duration of 30 minutes.
Using the lung principle, the sample flows into
the bag through the sampling tube only, without
passing through pumps or other equipment. The
materials used for ‘wetted’ parts being in
contact with the sample flow are PTFE and
stainless steel, in compliance with EN 13725.
The sample flow is regulated using a calibrated
nozzle integrated in the sampling device. Other
nozzles, flows and filling times are available
upon request. The sampling rate is independent
of the static pressure conditions at the sampling
point if the pressure difference is less than 150
mbar. The sampling flow is closed automatically
once the bag is full. Alternatively, the user can
select a shorter sampling duration using a
countdown timer. The CSD30 sampling device
operates from its integrated rechargeable
NiCd and battery does not require an AC power
connection. The CSD30 unit is robust, and is
easily operated by a technician.
18
Sampling Device CSD30
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
700 x 190 x 260 mm
‚
Weight: 4.8 kg
Sample volume
(bag size)
10 litres
Micro processor
controlled pump, flow
controlled by orifice
disc
5 min, 10 min, 15 min or
30 min continuous
sampling
VDI 3880 and
EN 13725 compliant
Easily portable
Stops automatically
High quality rechargeable batteries for about 16 x 30 min samples
Back to content
Maximum vacuum
output
300 mbar
Recommended range of
application
+/- 150 mbar
Lead-gel batteries 2 x 6 V
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Scope of supply:
• Charger
• PTFE probe tube 2 m
Accessories (optional):
• 10 litre sample bags
Article no.: S0010000
The Sampling Device Standard consists of a
vacuum container, which is evacuated by a
vacuum pump. The sampling point and the
standard sampler are connected by a probe.
Due to the evacuation in the sampling device,
the sample bag, inside the device, sucks in
sample air via the probe. During this process,
none of its components come into contact with
the sample air due to the construction of the
sampling device.
The range of application of the Sampling Device
Standard is, at pressure conditions of 150
mbar, at the peak of the probe (in the air flow
cross section). The filling time of an 10 litre bag
is 30 to 60 seconds depending on the pressure
conditions. The cylinder is made of transparent
PVC in order to aid the optical check of the
filling and the condition of the probe.
The Sampling Device Standard is robust, light
and easy to use. Due to its battery-operation the
sampling pump can be used independently of
the power supply.
19
Sampling Device Standard
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
730 x 160 x mm
Weight: 4.5 kg
Sample volume
(bag size)
10 litres
Filling time 30 – 60 s
Quick, short sampling
Easily portable
Easy to use
High quality
rechargeable batteries
Back to content
Flow approx. [m³/(m²x
h)]
30 / 15 / 7.5
Power supply Lithium polymer battery 11.1 V,
5.5 Ah with a
special charger
Sampling time with
CSD30
30, 15, 10 or 5 minutes
Power consumption Approx. 1 A
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Scope of supply:
• Ventilated sampling hood
• Floater for liquid surfaces
• High-quality Lithium polymer battery
• Charger power pack for use on 230V connections
• Connection cables
• Remote control
Optional:
• Sampling device CSD30
Article no.: BPH10000
The ventilated sampling hood supplements the
30-minute sampling system CSD 30 and is
primarily used for sampling non-ventilated,
odour emitting solid or liquid area sources.
When sampling liquids, floats are attached at
the sides of the hood. The system consists of a
hood, which is automatically ventilated by two
radial fans with a constant volumetric flow. The
inlet fan sucks in outside air via an
exchangeable orifice and transports the air from
the top into a corner of the hood via an active
carbon filter. From there the air is
distributed across the whole surface of the
hood, it absorbs the emitted odourants, and is
extracted to the opposite corner by the second
radial fan via another exchangeable orifice with
the same volumetric flow. The sample for
olfactometer tests (primarily using the CSD 30
sampling system) is taken from the suction
nozzles. The same volumetric flow is provided
on the inlet side and the suction side.
20
Ventilated Sampling Hood
Manufacturer Odournet GmbH, Germany
Dimensions
L / W / H [mm]
Ready for use approx.
1030 / 530 / 250
Weight Depending on the design
approx. 10 to 13 kg
Material Al Mg 3 19 chromatised
Adjusting the
volumetric flow
Exchangeable orifice on
the suction side at the inlet
and suction fan
For sampling on solid
or liquid surfaces
Light and easy to use
CSD30 mountable
VDI 3880 and
EN 13725 compliant
High quality batteries,
for about
6 samples/charge
Back to content
Diameter of the chimney 150 cm
Total height of the hood: 165 cm
Weight 8 kg
Material Al Mg 3 19 chromatised
Warranty 24 months (warranty extension
up to 5 years through service
contracts possible)
Scope of supply:
• Hood made of aluminum (1m2 with carry
handles)
• Chimney Ø 150 mm with wind collar
Optional:
• Cover plate with sampling connection for
measuring saturation concentrations on
passive sources rubber skirting running round
the base of the hood
Article no.: UBH10000
This hood allows defined sampling on flow area
sources with very low discharge speeds (e.g.
bio filter, ventilated compost clamp).
The volume of the exhaust air flow is guided
through the hood and the chimney, and is
consolidated. A wind collar reduces the
susceptibility to side-winds at the chimney
outlet and thereby keeps the flow conditions in
the chimney constant. The sampling hood has a
base surface area of 1m2 and a chimney with a
diameter of
150 mm. Due to the surface ratio between the
hood base surface and the chimney cross
section no relevant dynamic pressure is
exerted when the hood is applied. The chimney
can be disassembled from the body of the hood
by loosening the four knurled nuts. There are
carry handles on the body of the hood, so that it
can be transported easily. As the susceptibility
of the sampling hood to side-winds is
particularly a problem at the chimney outlet, a
wind collar is used here which shields it from
the wind and thereby keeps the flow conditions
in the hood constant.
21
Non Ventilated Sampling Hood
Manufacturer Odournet GmbH, Germany
Base surface 1 m²
Length 100 cm
Width 100 cm
For sampling on
ventilated surfaces
e.g. biofilter
Dismountable
VDI 3880 and
EN 13725 compliant
Back to content
Nitrogen or synthetic air Two-level bottle pressure
reducer
N2 or synthetic air
consumption
prox. 100l/h
Sampling time with
CSD30
30 or 10 minutes
Pressure at the
sampling point
-200 hPa bis +40 hPa
Max. Temperature at
the sampling point
240˚ C
Warranty 24 months
Scope of supply:
• Stainless steel pre-dilution probe,
with sapphire screen for approx. 1:10
• Two-level bottle pressure reducer
• Connection tubes
• Calibration datasheet
The following is required:
• N2, compressed air cylinder
• Sampling device
Accessories (optional):
• Additional sapphire screens for different
dilution ratios of approx. 1:5 or 1:20
Article no.: Z0020022
The GSP can supplement the sampling system
CSD 30 and dilutes the sample directly at the
sampling point. Here the pre-dilution can be
performed with nitrogen or synthetic air, whereby
the pre-dilution ratio is fixed by the manufacturer
(standard 1:10 with individual calibration data).
The pre-dilution ratio can, however, be set to
different exhaust air conditions on request by
exchanging a sapphire screen accessible from the
outside (e.g. dilution 1:5 or 1:20).
The pre-dilution system provides the sample air
flow with a constant supply of nitrogen (N2) or
synthetic air directly into the tip of the sampling
probe. The moist-warm sample is thereby diluted
immediately at the point of extraction with dry N2
or synthetic air so that when it cools down to the
transportation and laboratory temperature no
condensation forms in the probe, in the tube or in
the sample bag. As a result, no additional heating
is necessary.
Furthermore the GSP has a built-in particle filter to
protect against contamination as well as a micro
Prandtl tube, which can be used to monitor the
process.
22
On Site Pre Dilution Device GSP
Manufacturer Odournet GmbH, Germany
Dimensions
L / W / H [mm]
1800/150/30, can be
expanded by 800 mm
using an extension probe
Weight
1.5 kg
Control
mechanism of the
volume flow
Gas jet pumps
Sample air Calibrated sapphire screen
Pre dilution at the
source
Avoid losses due to
condensation
VDI 3880 compliant
Incl. a Prandtl tube for
air speed monitoring
Back to content
Air supply 3-5 bar clean compressed
air
Volume flow outlet 75 l/h
Power supply 230V DC
Weight Control unit: ca. 4kg
Heating block: ca. 2kg
Warranty 24 months (warranty
extension up to 5 years
through service contracts
possible)
Scope of supply
• 1 x LiGaVa incl.:
• 1 x Temperature control unit
• 1 x Vaporization unit (heat ing block)
Accessories:
• Heating power supply
• Sample bags made of Nalophane®
• Additional heating blocks
Article no.: PM000182
The LiGaVa system vaporizes any kind of liquid
into the gas phase. Since time, volume stream,
temperature, as well as the amount of liquid, are
all known variables, it is a highly standardized
sampling system. A set volume of the liquid
sample is injected into the LiGaVa, which is
designed to vaporise the sample. In order to
achieve the correct vapour pressure for your
particular liquid sample, the evaporation
temperature can be individually set (up to
220°C). As the gas is formed it streams into a
connected sample bag at a constant volume flow
rate. The liquid has now been transferred into
gaseous form ready for various analyses:
Odour concentration
Odour intensity & Liking
Molecular Profile
23
Liquid to gas vaporizer - LiGaVa
Manufacturer Odournet GmbH, Germany
Dimension
(L x W x H)
Control unit: 350mm x 150mm x
260mm (without grip)
Heating block: 120mm x 120mm
x 80mm
Adjustable
temperature range
5°C above room temperature to
220°C
Determine the
detection threshold
Optimise your
substance-perception-
ratio (linear vs.
logarithmical scale)
Learn how your
fragrance is perceived
Compare fragrances
Determine the
molecular profile
Optimise your formula
Back to content
Sample bags have been designed especially
to respond to the quality requirements of
EN 13725:2003, to minimize the risk of
contamination of odour samples. The bags
have an approximate capacity of 8l, 10l, 40l or
60l depending which sampling device is used.
24
Sample bags
Ready to use bags
Special sizes at request
Material: Nalophan®
Type Amount Connectors Sealings Sampling Device Article no.
8 litre bags 5, 10, 25,
50, 100
PTFE PTFE Sampling Device
Standard (model
until 2014)
Z0020003
10 litre bags 5, 10, 25,
50, 100
PTFE PTFE CSD30, Sampling
Device Standard
(model since 2014)
Z0020004
40 litre bags 5, 10, 25,
50
Swagelock Swagelock Vessel ON000015
60 litre bags 5, 10, 25,
50
PTFE PTFE PureSniff XL –
sampling function
Z0020039
Back to content
Scope of supply:
• USB stick with software (odour threshold
measurement)
Accessories (not supplied):
• Plug-in for intensity and hedonics
• USB cable
Article no.: SW000002
The control software offers easy, intuitive operation.
Integrated test routines check the correct functioning of
your olfactometer. A comfortable tester management
system provides extensive functions to manage the
applied testers as per EN 13725:2003.
The archiving of the measuring data produced such as
n-butanol allows the quality control of your measuring
results. The creation of additional databases for
reference substance checks is possible with up to 16
databases. The detailed recording and evaluation of
your measurement results corresponds to all the
requirements of EN 13725:2003.
This also includes the requirements of the accreditation
process as per ISO/IEC 17025:2005 on the
documentation of the raw data collected including the
saving of data which is not used for the calculation of the
end result (Discarding passages).
In order to guarantee the fastest possible support in
case of defects, the software provides automatic error
reports.
The software supports the measurement of individual
samples by up to 16 testers in 4 teams. Measurements
already completed can be continued at any time.
25
TO-series controlling software
Processor Pentium III+
System memory 128 MB RAM
Available hard disc
space
10 MB
Connecting via USB
Direct control of
hardware and
measurement
procedure
Storage and calculation
of results
Interaction and collection of data from panel members
Running with TO7, TO8,
TO9, TO8-8, TO8s
Panel management up
to 16 databases for
different reference
materials
Upgrade for Intensity
and Hedonic Tone
Fully EN 13725
compliant
USB port
Windows based
operating system
Windows 2000 / XP / Vista /
Windows 7
TO series olfactometer TO8 / TO8-S / TO8-8 / TO9
Back to content
Article no.: OMP00001
Odourmap is an innovative communication
platform on which plant operators, citizens and
authorities can be equally integrated into the
odour management process. Odourmap
records and visualises a variety of data that is
relevant for dealing with odour problems and
citizen complaints, providing a multilateral,
easy-to-use platform, where all parties involved
can access the information relevant to them.
Watching our Explainer Film you can discover
Odourmap and its benefits in just 90 seconds.
26
Odourmap Web-based Tool for Odour Complaint Management
Online Citizen Monitoring Collection and registration of citizen observations and complaints
Smartphone App Collection and registration of citizen observations and complaints
Grid Inspections Data input from systematic Grid Inspections using citizen panels, professional
observers (according to VDI 3940 Part 2) or facility site staff.
Meteo Data Online display of current Meteo Data and optionally also the modelled current
impact caused by known, regulated emission sources. This can help in plausibility
checks of observations, and in assigning observations to the most plausible source
under the current wind flow conditions.
Facility Module Mapping of odour sources, the setting of facility-specific odour types, the
assessment of the odour potential of odour sources and the input of limit values. In
addition, the contact details of facility operators can be registered.
Communication and complaints management tool
Promotes engagement between operators, regulators and citizens
Complaint validation and prediction of future impact risk
Operator of platform retains full access control
Integrates site based data with community observations
Historic analysis and visualisation of odour events
Expandable according to client requirements
Odourmap is designed to also integrate other data streams like dispersion modelling as well as sensor
data (e-noses) and emission data and odour critical operational data.
As a modular platform Odourmap is continually being enhanced and the range of features extended.
For further details and a guided Odourmap tour in a TeamViewer Live Presentation to experience the
overall look and feel of Odourmap and its easy-to-use functions and features at first hand, please contact
Marc Andresen (T: +49 431 22012-14; E: [email protected] ).
Back to content
Article no.: GRD00048
iNose offers a complete solution to odour
control:
iNose provides a clear insight into the
dispersion of gas and odour based on current
and reliable weather data. That way iNose
allows operators to plan production or to adjust
production settings based on the (forthcoming)
meteorological circumstances.
This prevents discontent and improves the
relationship with the surrounding community.
27
iNose – Odour Dispersion Monitoring System
Real-time display of emissions
Display of effects of adjustments in the operation process
Immediate response to changes in current meteorological data and weather forecasts
Reliable data for complaint validation
Always up-to-date. Easy and fast in use.
Little to no maintenance. Accessible from anywhere at any time.
iNose automatically imports the most current meteorological data from the database administered by the
leading MeteoGroup Europe. The most common dispersion computer model 'Plume' calculates the
dispersions. Google Maps provides maps and satellite images from various providers. The substance
properties are taken from our Odournet library of Odour Thresholds, based on extensive literature review
and on in-house odour thresholds gained from thousands of studies in all odour relevant sectors in the
past 30 years, or may be determined by our environmental consultants at your site.
With iNose, a company can obtain a clear insight into their (future) odour and gas dispersion over the
surroundings. This is made possible by the permanent connection with the most current weather
forecasts and readings.
Prevent penalties or even recall of permit to operate and favourable position when negotiating new permits.
Minimize chances of complaints and thus improved relation with the immediate surroundings by building good
reputation and image.
Validate complaints.
Benfit from the automatic alert management in case of (future) excessive emissions.
Back to content
Sensory Expertise delivered worldwide
The ODOURNET Group
Odournet GmbH · Fraunhoferstr. 13 · 24118 Kiel · Germany · T +49 431 22012-0 · F +49 431 22012-17
[email protected] · www.odournet.com