cardiosoft specifications tender
DESCRIPTION
topTRANSCRIPT
-
GE Healthcare
CardioSoft* Diagnostic System Tender SpecificationsUser Provided Hardware
TheclinicalacquisitionsystemPCsmustrunWindowsXPProfessional orWindows7Professional
Processor:Pentium4ClassProcessor1.6GHz(WindowsXPProfessional),Pentium4ClassProfessional,2GHz(Windows7)
ThesystemsupportslaserprintersbothLetterandA4papersizes
Thesystemhasminimumof20GBharddrivewith4GBoffreeHDDspace
Systemhasatleast512MBRAM,1GBisrequiredforWindows7Professional
ThesystemmustuseWindowsoperatingsystemcompatiblemonitors,printers,mouseinterfacesandexternalstoragedevices,CD-RW,SDcard andUSBexternalharddrives
ThesystemmustsupportUSBandserialcompatibleperipheraldevices
Graphicsadapter:SVGA1024x768minimum(SXGA1280x1024recommended)
Connectivity
Thesystemmustbeinterfacedtoelectronicmedicalrecordssystems (EMRs)usingeitherEMRGateway(HL7/XML)oradirect,embeddedinterface
Thesystemmustprovideabi-directionalinterfacetotheGEHealthcareMUSE* Cardiology Information System
ThesystemmustsupportDICOMformattedreportexport
Thesystemmustsupportbi-directionalDICOMmodalityworklists
Thesystemmustsupportclientworkstationsandshareddatabase capability.(WindowsXPProfessionalorWindows7Professional)
ThesystemmustsupportclientworkstationoperationinCitrix networkenvironments
ThesystemmustprovidepatientproceduredatastorageonaPCnetwork
ThesystemmustprovideaWebviewable(HTML)reportandthe associatedWeb-portalstructure
ThesystemmustexportastructuredXMLdataexport
-
ThesystemmustprovideaconfigurableTTLsynchronizationpulseforimagingandotherexternaldevices(CAMUSBA/T)
Thesystemmustbeabletoswitchbetweenashared databaseandstand-alonedatabasemode
Thesystemmustallowfordatabasesynchronizationwhenre-connectedtoashareddatabaseconfiguration
Thesystemmustallowautomaticandmanualproceduretransfer(sharedDB,DICOM,MUSE)
ThesystemmustacceptECGdatafromGEcardiographs:MAC*800,MAC1200,MAC1600,MAC3500,MAC5000, MAC5500
Thesystemmustprovideaconfigurabletimeouttool toreleaseunusedworkstationlicenses
Thesystemmustsupporteditingandprocedurereview inCitrixnetworkenvironments
Thesystemsupportsasuctionelectrodesystem(KISS)
ThesystemmustinterfacewiththeTango+AutomatedBPsystem(requiresCAMUSBA/T)
Thesystemmustprovideametaboliccartcommunicationinterfaceconfiguration
System Compatibility
Thesystemmusthavecustomizablefinalreport
Thefinalreportmustbeconfigurablewithalogoorimagefor display and printing
Thesystemmustallowconfigurablecolorschemes todisplayreal-timedata
Thesystemmustallowconfigurablecolorschemesfor final report printing to color printers
Thesystemmustprovidebi-directionaldatainterfaceviaBDT/GDTV2.1
FinalreportsmustbeexportablefromthesysteminWord,PDForDICOM
Comprehensiveproceduredatamustbeexportablefrom thesysteminXML,ExcelandHL7
ThesystemmustprovideinterfacestoHolterprograms toallowreportviewing
Theusersystemcontrolsmustbeavailableasicons orastranslatedtext
TheECGtracedisplayspeedmustbeadjustableinrealtime
Thesystemdatabasemustbesortablebyprocedure type,patientname,dateofprocedure,patientIDor confirmedstatus
Duringprocedures,usermustbeabletoopenaprevious procedureforthesamepatienttoassessbaseline,peak andmaximumSTdepressionpoints
Duringtheprocedure,theusermustbeabletoreview procedureandpatientinformationfromtheactive ECG screen
Thesystemmustprovidemulti-levelandmulti-personpasswordprotectedsecurity
Clinical Application Software
Thesystemmustsupport3-,6-,12-and15-leadECG acquisition,displayandreporting.
Allleadsconfigurationscanbedisplayedon-screen and printed in final reports
Thesystemmustprovidepop-upmediansinpost-testreviewtoviewindividualQRScomplexestimesynchronizedtoleadsviewedinthetrendgraphs
Thesystemmustprovideinterpretativetoolsforresting ECG(12SL*)
Thesystemmustprovideinterpretativetoolsforstress ECG(XTI)
ThesystemmustprovideriskassessmenttoolsCAD,StrokeandDukeTreadmillScore
ThesystemmustprovideriskassessmenttoolsT-WaveAlternans(TWA-MMAmethod)
ThesystemmustallowforECGdisplayspeedand amplitudeadjustment
ThesystemmustprovideaudioECGduringreview toconveyarrhythmias
ThesystemmustprovideECGfreeze-frameduring real-timedisplay
Thesystemmuststoreanddisplayforreviewup to60minutesofrestingECGdata
Musthaveoneormoreofthefollowingregulatoryclearances:510(k),CEmark,SFDA,KFDA,CSA
Thesystemmustallowforcomparisontoprevious proceduresdataincludingECGswithmeasurement ability(calipers)ofalltraces
Thesystemmustprovidereportingfunctionalitytoincluderesultsfromuptotwentypriorprocedures.Proceduresmaybeofdifferentmodalities
ThesystemmustprovideBPmeasurementsinbothmmHgandkPA
SystemmustbeabletofreezeECGtracingduring procedurewhilereal-timeECGisdisplayed
ThesystemmustbeabletoreanalyzeECGproceduredata,resetmeasurementpointsandrecalculateS-Tsegmentvaluesbasedonnewpoints
-
ThesystemmustbeabletoscorepatienttestresultsusingDukeTreadmillscore
ThesystemmustprovidefulldisclosureECGandallowdetailedreviewandmeasurementoffulldisclosureECG
FulldisclosureECGmustbeexportableasanXMLdataset
Thesystemmustallowfulldisclosurestripstobeappendedto final report
Thesystemmustallow10-secondin-teststrip tobeappendedtofinalreport
Thesystemmustbeconfigurableforpaperless documentationstorage/archiving
Thesystemmustsupportremotetechnicalsupport viatheInternet
Thesystemmustsupportviewingoffinalreportsusingnon-proprietary,freesoftware
Thesystemmustsupportsavingfinalreports innon-proprietaryformat
Thesystemmusthavetheabilitytosimulateastresstest forqualityassuranceandtraining
Thesystemmusthavetheabilitytochangeprotocol toanyotherprogrammedprotocolsormanuallycontrol thesystemduringtheprocedure
Thesystemmustsupportautomatedramped,staged or pharmacological protocols
Thesystemmusthavetheabilitytoreviewfinalreport andin-testreportsonscreen
Thesystemmusthavetheabilitytoview3-,6-,12-or15- ECGleadsonscreenduringexam
ThesystemmusthavetheabilitytoeditfinalreportwhileviewingECGinrecovery
Thesystemmusthavetheabilitytopermituser tocustomizeprotocols
Thesystemmusthavetheabilitytocustomizeusersetups
Thesystemmusthavetheabilitytomodifyfiducialpoint,suchasS-Tslopeandlevelmeasurementpoint
Thesystemmusthavevisualandaudiblenotification foruser-definedparameters
Thesystemmusthaveuser-customizablescreenlayouts ofvitalsigndata
MeetsorexceedsperformancerequirementsofANSI/AAMIEC11forgainaccuracy,frequencyresponse,CMRR,systemnoise,dynamicrange,inputimpedance
Usercontrolledfilters:FRF,cubicspline,linefiltersfor50Hzand60Hz,andlowpassfiltersfrom20Hzto150Hz
Thesystemmustprovideimpedancechecktosupport qualityofelectrodepreparation
Thesystemmustprovideon-lineusermanualwith indexedhyperlinks
On-screenleadviewmustbeselectableandchangeableduringprocedures
Thesystemmustprovidereportsusingexercisespirometrydata from the ES option
ThesystemmustinterfacewiththeGEHealthcareT2100,T2000treadmills
ThesystemmustinterfacewithGEHealthcareeBike uprightandsupineergometers
ThesystemmustprovideadetachablepatientECG acquisitionmodulewithreplaceablepatientleadwires
Thesystemmustprovidedetachable,exchangeable connectors from pinch to clip to snap
Thesystemmustsupportspirometryassessment
SpirometryassessmentwillincludeFVC,MVV,FEV6,FEV1,FEV1/FVC,FEV1/VC,FEV1/FEV6,MEF25,(FEF75),MEF50(FEF50),MEF75(FEF25),MEF25%-75%,PEF,FET,FIVC,PIF, IVC,ERV,IRV,Pre-Post%Variation,LungAge
Thespirometryassessmentmustincludeadult-predictedNHANES-III,Knudson_83,Knudson_76,Crapo,Morris,ERS(ECCS/EGKS),Forche(Austria),Sapaldia(Switzerland),Roca(Spain),Berglund,Gulsvik,Hedenstrm,Gore,Cherniak(onlyforMVV)
Thespirometryassessmentmustincludepediatric- predictedDockery(Harvard),Hsu,Zapletal,Polgar,Hibbert
Thesystemmustsupportambulatorybloodpressure assessment
Bloodpressureassessmentwillincludepulsepressure,averages,diurnalhighandlowvaluesandgraphs
**Notallhardwareorsoftwareoptionsandfunctionsareavailableineverymarketorcountry. ConsultwithyourlocalGErepresentativetodetermineifspecifichardwareoroptionsare forsaleinyourmarket.
-
2012GeneralElectricCompanyAllrightsreserved.
GeneralElectricCompanyreservestherighttomakechangesinspecificationsandfeaturesshownherein,ordiscontinuetheproductdescribedatanytimewithoutnoticeorobligation.
*GE,GEMonogram,12SL,CardioSoft,MACandMUSEare trademarksofGeneralElectricCompany.
GEHealthcare,adivisionofGeneralElectricCompany.
CitrixisatrademarkofCitrixSystems,Inc.
Microsoft,Excel,Word,Windows,andWindowsServer aretrademarksofMicrosoftCorporation
PentiumisatrademarkofIntel.
About GE Healthcare
GEHealthcareprovidestransformationalmedicaltechnologiesandservicesthat areshapinganewageofpatientcare.Ourbroadexpertiseinmedicalimagingand informationtechnologies,medicaldiagnostics,patientmonitoringsystems,drug discovery,biopharmaceuticalmanufacturingtechnologies,performanceimprovement andperformancesolutionsserviceshelpsourcustomerstodeliverbettercaretomore peoplearoundtheworldatalowercost.Inaddition,wepartnerwithhealthcareleaders,strivingtoleveragetheglobalpolicychangenecessarytoimplementasuccessfulshift tosustainablehealthcaresystems.
Ourhealthymaginationvisionforthefutureinvitestheworldtojoinusonourjourney aswecontinuouslydevelopinnovationsfocusedonreducingcosts,increasingaccess,andimprovingqualityaroundtheworld.HeadquarteredintheUnitedKingdom,GEHealthcareisaunitofGeneralElectricCompany(NYSE:GE).Worldwide,GEHealthcareemployeesarecommittedtoservinghealthcareprofessionalsandtheirpatientsinmorethan100countries.FormoreinformationaboutGEHealthcare,visitourwebsiteatwww.gehealthcare.com
EMEA DOC1047317 5/12
GE HealthcareP.O.Box900,FIN-00031GE,FinlandGEDirectUnitedKingdom:+44(0)8000329201 www.gehealthcare.com