GE Healthcare CardioSoft * Diagnostic System Tender Specifications User Provided Hardware • The clinical acquisition system PCs must run Windows ® XP Professional or Windows 7 Professional • Processor: “Pentium 4 Class” Processor ≥ 1.6 GHz (Windows XP Professional), “Pentium 4 Class” Professional, ≥ 2 GHz (Windows 7) • The system supports laser printers both Letter and A4 paper sizes • The system has minimum of 20 GB hard drive with 4 GB of free HDD space • System has at least 512 MB RAM, 1 GB is required for Windows 7 Professional • The system must use Windows operating system compatible monitors, printers, mouse interfaces and external storage devices, CD-RW, SD card and USB external hard drives • The system must support USB and serial compatible peripheral devices • Graphics adapter: SVGA 1024 x 768 minimum (SXGA 1280 x 1024 recommended) Connectivity • The system must be interfaced to electronic medical records systems (EMRs) using either EMR Gateway (HL7/XML) or a direct, embedded interface • The system must provide a bi-directional interface to the GE Healthcare MUSE * Cardiology Information System • The system must support DICOM formatted report export • The system must support bi-directional DICOM modality worklists • The system must support client workstations and shared database capability. (Windows XP Professional or Windows 7 Professional) • The system must support client workstation operation in Citrix ® network environments • The system must provide patient procedure data storage on a PC network • The system must provide a Web viewable (HTML) report and the associated Web-portal structure • The system must export a structured XML data export

Upload: chris-hartoyo

Post on 17-Dec-2015

5 views

Category:

Documents


0 download

DESCRIPTION

top

TRANSCRIPT

  • GE Healthcare

    CardioSoft* Diagnostic System Tender SpecificationsUser Provided Hardware

    TheclinicalacquisitionsystemPCsmustrunWindowsXPProfessional orWindows7Professional

    Processor:Pentium4ClassProcessor1.6GHz(WindowsXPProfessional),Pentium4ClassProfessional,2GHz(Windows7)

    ThesystemsupportslaserprintersbothLetterandA4papersizes

    Thesystemhasminimumof20GBharddrivewith4GBoffreeHDDspace

    Systemhasatleast512MBRAM,1GBisrequiredforWindows7Professional

    ThesystemmustuseWindowsoperatingsystemcompatiblemonitors,printers,mouseinterfacesandexternalstoragedevices,CD-RW,SDcard andUSBexternalharddrives

    ThesystemmustsupportUSBandserialcompatibleperipheraldevices

    Graphicsadapter:SVGA1024x768minimum(SXGA1280x1024recommended)

    Connectivity

    Thesystemmustbeinterfacedtoelectronicmedicalrecordssystems (EMRs)usingeitherEMRGateway(HL7/XML)oradirect,embeddedinterface

    Thesystemmustprovideabi-directionalinterfacetotheGEHealthcareMUSE* Cardiology Information System

    ThesystemmustsupportDICOMformattedreportexport

    Thesystemmustsupportbi-directionalDICOMmodalityworklists

    Thesystemmustsupportclientworkstationsandshareddatabase capability.(WindowsXPProfessionalorWindows7Professional)

    ThesystemmustsupportclientworkstationoperationinCitrix networkenvironments

    ThesystemmustprovidepatientproceduredatastorageonaPCnetwork

    ThesystemmustprovideaWebviewable(HTML)reportandthe associatedWeb-portalstructure

    ThesystemmustexportastructuredXMLdataexport

  • ThesystemmustprovideaconfigurableTTLsynchronizationpulseforimagingandotherexternaldevices(CAMUSBA/T)

    Thesystemmustbeabletoswitchbetweenashared databaseandstand-alonedatabasemode

    Thesystemmustallowfordatabasesynchronizationwhenre-connectedtoashareddatabaseconfiguration

    Thesystemmustallowautomaticandmanualproceduretransfer(sharedDB,DICOM,MUSE)

    ThesystemmustacceptECGdatafromGEcardiographs:MAC*800,MAC1200,MAC1600,MAC3500,MAC5000, MAC5500

    Thesystemmustprovideaconfigurabletimeouttool toreleaseunusedworkstationlicenses

    Thesystemmustsupporteditingandprocedurereview inCitrixnetworkenvironments

    Thesystemsupportsasuctionelectrodesystem(KISS)

    ThesystemmustinterfacewiththeTango+AutomatedBPsystem(requiresCAMUSBA/T)

    Thesystemmustprovideametaboliccartcommunicationinterfaceconfiguration

    System Compatibility

    Thesystemmusthavecustomizablefinalreport

    Thefinalreportmustbeconfigurablewithalogoorimagefor display and printing

    Thesystemmustallowconfigurablecolorschemes todisplayreal-timedata

    Thesystemmustallowconfigurablecolorschemesfor final report printing to color printers

    Thesystemmustprovidebi-directionaldatainterfaceviaBDT/GDTV2.1

    FinalreportsmustbeexportablefromthesysteminWord,PDForDICOM

    Comprehensiveproceduredatamustbeexportablefrom thesysteminXML,ExcelandHL7

    ThesystemmustprovideinterfacestoHolterprograms toallowreportviewing

    Theusersystemcontrolsmustbeavailableasicons orastranslatedtext

    TheECGtracedisplayspeedmustbeadjustableinrealtime

    Thesystemdatabasemustbesortablebyprocedure type,patientname,dateofprocedure,patientIDor confirmedstatus

    Duringprocedures,usermustbeabletoopenaprevious procedureforthesamepatienttoassessbaseline,peak andmaximumSTdepressionpoints

    Duringtheprocedure,theusermustbeabletoreview procedureandpatientinformationfromtheactive ECG screen

    Thesystemmustprovidemulti-levelandmulti-personpasswordprotectedsecurity

    Clinical Application Software

    Thesystemmustsupport3-,6-,12-and15-leadECG acquisition,displayandreporting.

    Allleadsconfigurationscanbedisplayedon-screen and printed in final reports

    Thesystemmustprovidepop-upmediansinpost-testreviewtoviewindividualQRScomplexestimesynchronizedtoleadsviewedinthetrendgraphs

    Thesystemmustprovideinterpretativetoolsforresting ECG(12SL*)

    Thesystemmustprovideinterpretativetoolsforstress ECG(XTI)

    ThesystemmustprovideriskassessmenttoolsCAD,StrokeandDukeTreadmillScore

    ThesystemmustprovideriskassessmenttoolsT-WaveAlternans(TWA-MMAmethod)

    ThesystemmustallowforECGdisplayspeedand amplitudeadjustment

    ThesystemmustprovideaudioECGduringreview toconveyarrhythmias

    ThesystemmustprovideECGfreeze-frameduring real-timedisplay

    Thesystemmuststoreanddisplayforreviewup to60minutesofrestingECGdata

    Musthaveoneormoreofthefollowingregulatoryclearances:510(k),CEmark,SFDA,KFDA,CSA

    Thesystemmustallowforcomparisontoprevious proceduresdataincludingECGswithmeasurement ability(calipers)ofalltraces

    Thesystemmustprovidereportingfunctionalitytoincluderesultsfromuptotwentypriorprocedures.Proceduresmaybeofdifferentmodalities

    ThesystemmustprovideBPmeasurementsinbothmmHgandkPA

    SystemmustbeabletofreezeECGtracingduring procedurewhilereal-timeECGisdisplayed

    ThesystemmustbeabletoreanalyzeECGproceduredata,resetmeasurementpointsandrecalculateS-Tsegmentvaluesbasedonnewpoints

  • ThesystemmustbeabletoscorepatienttestresultsusingDukeTreadmillscore

    ThesystemmustprovidefulldisclosureECGandallowdetailedreviewandmeasurementoffulldisclosureECG

    FulldisclosureECGmustbeexportableasanXMLdataset

    Thesystemmustallowfulldisclosurestripstobeappendedto final report

    Thesystemmustallow10-secondin-teststrip tobeappendedtofinalreport

    Thesystemmustbeconfigurableforpaperless documentationstorage/archiving

    Thesystemmustsupportremotetechnicalsupport viatheInternet

    Thesystemmustsupportviewingoffinalreportsusingnon-proprietary,freesoftware

    Thesystemmustsupportsavingfinalreports innon-proprietaryformat

    Thesystemmusthavetheabilitytosimulateastresstest forqualityassuranceandtraining

    Thesystemmusthavetheabilitytochangeprotocol toanyotherprogrammedprotocolsormanuallycontrol thesystemduringtheprocedure

    Thesystemmustsupportautomatedramped,staged or pharmacological protocols

    Thesystemmusthavetheabilitytoreviewfinalreport andin-testreportsonscreen

    Thesystemmusthavetheabilitytoview3-,6-,12-or15- ECGleadsonscreenduringexam

    ThesystemmusthavetheabilitytoeditfinalreportwhileviewingECGinrecovery

    Thesystemmusthavetheabilitytopermituser tocustomizeprotocols

    Thesystemmusthavetheabilitytocustomizeusersetups

    Thesystemmusthavetheabilitytomodifyfiducialpoint,suchasS-Tslopeandlevelmeasurementpoint

    Thesystemmusthavevisualandaudiblenotification foruser-definedparameters

    Thesystemmusthaveuser-customizablescreenlayouts ofvitalsigndata

    MeetsorexceedsperformancerequirementsofANSI/AAMIEC11forgainaccuracy,frequencyresponse,CMRR,systemnoise,dynamicrange,inputimpedance

    Usercontrolledfilters:FRF,cubicspline,linefiltersfor50Hzand60Hz,andlowpassfiltersfrom20Hzto150Hz

    Thesystemmustprovideimpedancechecktosupport qualityofelectrodepreparation

    Thesystemmustprovideon-lineusermanualwith indexedhyperlinks

    On-screenleadviewmustbeselectableandchangeableduringprocedures

    Thesystemmustprovidereportsusingexercisespirometrydata from the ES option

    ThesystemmustinterfacewiththeGEHealthcareT2100,T2000treadmills

    ThesystemmustinterfacewithGEHealthcareeBike uprightandsupineergometers

    ThesystemmustprovideadetachablepatientECG acquisitionmodulewithreplaceablepatientleadwires

    Thesystemmustprovidedetachable,exchangeable connectors from pinch to clip to snap

    Thesystemmustsupportspirometryassessment

    SpirometryassessmentwillincludeFVC,MVV,FEV6,FEV1,FEV1/FVC,FEV1/VC,FEV1/FEV6,MEF25,(FEF75),MEF50(FEF50),MEF75(FEF25),MEF25%-75%,PEF,FET,FIVC,PIF, IVC,ERV,IRV,Pre-Post%Variation,LungAge

    Thespirometryassessmentmustincludeadult-predictedNHANES-III,Knudson_83,Knudson_76,Crapo,Morris,ERS(ECCS/EGKS),Forche(Austria),Sapaldia(Switzerland),Roca(Spain),Berglund,Gulsvik,Hedenstrm,Gore,Cherniak(onlyforMVV)

    Thespirometryassessmentmustincludepediatric- predictedDockery(Harvard),Hsu,Zapletal,Polgar,Hibbert

    Thesystemmustsupportambulatorybloodpressure assessment

    Bloodpressureassessmentwillincludepulsepressure,averages,diurnalhighandlowvaluesandgraphs

    **Notallhardwareorsoftwareoptionsandfunctionsareavailableineverymarketorcountry. ConsultwithyourlocalGErepresentativetodetermineifspecifichardwareoroptionsare forsaleinyourmarket.

  • 2012GeneralElectricCompanyAllrightsreserved.

    GeneralElectricCompanyreservestherighttomakechangesinspecificationsandfeaturesshownherein,ordiscontinuetheproductdescribedatanytimewithoutnoticeorobligation.

    *GE,GEMonogram,12SL,CardioSoft,MACandMUSEare trademarksofGeneralElectricCompany.

    GEHealthcare,adivisionofGeneralElectricCompany.

    CitrixisatrademarkofCitrixSystems,Inc.

    Microsoft,Excel,Word,Windows,andWindowsServer aretrademarksofMicrosoftCorporation

    PentiumisatrademarkofIntel.

    About GE Healthcare

    GEHealthcareprovidestransformationalmedicaltechnologiesandservicesthat areshapinganewageofpatientcare.Ourbroadexpertiseinmedicalimagingand informationtechnologies,medicaldiagnostics,patientmonitoringsystems,drug discovery,biopharmaceuticalmanufacturingtechnologies,performanceimprovement andperformancesolutionsserviceshelpsourcustomerstodeliverbettercaretomore peoplearoundtheworldatalowercost.Inaddition,wepartnerwithhealthcareleaders,strivingtoleveragetheglobalpolicychangenecessarytoimplementasuccessfulshift tosustainablehealthcaresystems.

    Ourhealthymaginationvisionforthefutureinvitestheworldtojoinusonourjourney aswecontinuouslydevelopinnovationsfocusedonreducingcosts,increasingaccess,andimprovingqualityaroundtheworld.HeadquarteredintheUnitedKingdom,GEHealthcareisaunitofGeneralElectricCompany(NYSE:GE).Worldwide,GEHealthcareemployeesarecommittedtoservinghealthcareprofessionalsandtheirpatientsinmorethan100countries.FormoreinformationaboutGEHealthcare,visitourwebsiteatwww.gehealthcare.com