Calibration of the VA2
B Pal,R Mountain,
M Artuso6/29/2010
Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection
![Page 2: Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection](https://reader036.vdocuments.us/reader036/viewer/2022062516/56649d2d5503460f94a030a4/html5/thumbnails/2.jpg)
Calibration fixture is replaced by single channel fixture
![Page 3: Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection](https://reader036.vdocuments.us/reader036/viewer/2022062516/56649d2d5503460f94a030a4/html5/thumbnails/3.jpg)
Single channel fixture schematic
1.0 pF
![Page 4: Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection](https://reader036.vdocuments.us/reader036/viewer/2022062516/56649d2d5503460f94a030a4/html5/thumbnails/4.jpg)
Channel Gain Channel 141 previously used in diamond measurements
![Page 5: Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection](https://reader036.vdocuments.us/reader036/viewer/2022062516/56649d2d5503460f94a030a4/html5/thumbnails/5.jpg)
Input Bias
Nominal gain from specs is 30 mV/fC, which is ½ of the gain we measured in ch 141.
Nominal value value Used
Vfp -0.2 to 0.5 V -0.4 V
Vfs 700 mV 700 mV
Pre_Bias 100 mV 99.8 mV
Sha_Bias 22 mV 22.3 mV
i_buf 14 mV 13.8 mV
![Page 6: Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection](https://reader036.vdocuments.us/reader036/viewer/2022062516/56649d2d5503460f94a030a4/html5/thumbnails/6.jpg)
Gain vs. I_buf
![Page 7: Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection](https://reader036.vdocuments.us/reader036/viewer/2022062516/56649d2d5503460f94a030a4/html5/thumbnails/7.jpg)
To do