Grup de Física de les Radiacions 24 th International Conference on Nuclear Tracks in Solids CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER Measurements in quasi- monoenergetic neutron beams at the EC-IRMM Van der Graaf accelerator C. Domingo. M.J. García-Fusté. K. Amgarou. E. Morales and J. Castelo

Upload: badrani

Post on 08-Jan-2016

47 views

Category:

Documents


1 download

DESCRIPTION

CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER. Measurements in quasi-monoenergetic neutron beams at the EC-IRMM Van der Graaf accelerator. C. Domingo. M.J. García-Fusté. K. Amgarou. E. Morales and J. Castelo. Overview. The UAB PADC neutron dosimetry system Dosemeter response - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

Grup de Física de les Radiacions24th International Conference on

Nuclear Tracks in Solids

CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

Measurements in quasi-monoenergetic neutron beams at the EC-IRMM Van der Graaf accelerator

C. Domingo. M.J. García-Fusté. K. Amgarou. E. Morales and J. Castelo

Page 2: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS2

0.

1.

2.

3.

4.

5.

Overview

1. The UAB PADC neutron dosimetry system

2. Dosemeter response

3. Measurements at EC-IRMM Van der Graaf facility

4. Results

5. Conclusions

Page 3: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS3

0.

1.

2.

3.

4.

5.

The UAB PADC neutron dosimetric system

PADC neutron detector: CR-39

Converters: Rich in H. fast neutron

interaction Nylon rich in N. high

cross section for thermal neutrons

Wrapped in Climafol pouches

Dosemeter configuration

6 cm

6 cm

Methacrylate (5 mm)

Makrofol (300 μm)

Polyethylene (3 mm)

CR-39 (500 μm)

Nylon (100 μm)

Page 4: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS4

0.

1.

2.

3.

4.

5.

The UAB PADC neutron dosimetric system

Process after irradiation

6 cm

6 c

m

Cutting

Electrochemical Etching

Page 5: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS5

0.

1.

2.

3.

4.

5.

The UAB PADC neutron dosimetric system

Typical track diameters of 100 μm

Semi-automatic counting system Sony camera ImageJ analysis

software Scanned area: 0.53 cm2

Track density determination

Page 6: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS6

0.

1.

2.

3.

4.

5.

Dosemeter response Expected behaviour

NRPB Report W25 R.J. Tanner et al.. Radit. Meas. 34 (2001) 457-461

Page 7: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS7

0.

1.

2.

3.

4.

5.

Dosemeter response

Calibration with ISO neutron sources and realistic fields at IRSN Cadarache (France)

SourceType of

field

Mean

EnergyRΦ

(neutron-1)

RHp(10)

(cm-2·mSv-1)

241Am-Be ISO 4.15 MeV (1.82 ± 0.13)×10-4 450 ± 33

Bare 252Cf ISO 2.13 MeV (2.01 ± 0.13)×10-4 591 ± 39

(252Cf + D2O)/Cd ISO 539 keV (5.55 ± 0.27)×10-5 499 ± 24252Cf + D2O ISO 539 keV (5.17 ± 0.23)×10-5 518 ± 23

SIGMA Realistic 79 keV (1.422 ± 0.064)×10-5 575 ± 26

Page 8: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS8

0.

1.

2.

3.

4.

5.

Dosemeter fluence response

Page 9: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS9

0.

1.

2.

3.

4.

5.

Dosemeter Hp(10.0º) response

Page 10: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS10

0.

1.

2.

3.

4.

5.

Measurements at EC-IRMM Van der Graaf facility

EC NUDAME program

Quasi-monoenergetic neutron fields from few hundreds keV to 20 MeV

New energies for dosemeter calibration

Page 11: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS11

0.

1.

2.

3.

4.

5.

Measurements at EC-IRMM Van der Graaf facility

Neutron energies used:

300 keV. 800 keV.

7.5 MeV and 16 MeV Methacrylate phantom

used BSS spectrometry for

determination of neutron spectrum and H*(10) reference value at each irradiation point

Neutron beams and methods

Page 12: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS12

0.

1.

2.

3.

4.

5.

Measurements at EC-IRMM Van der Graaf facility

Page 13: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS13

0.

1.

2.

3.

4.

5.

Results

4 dosemeters per energy Background determined with 4 extra non-irradiated dosemeters

Nominal

EnergyTarget

Mean Energy

with BSSRΦ

(neutron-1)

RHp(10)

(cm-2·mSv-1)

300 keV T(p.n)3He 269 keV (4.13 ± 0.53)×10-5 211 ± 27

800 keV T(p.n)3He 662 keV (1.24 ± 0.16)×10-4 371 ± 48

7.5 MeV D(d.n)3He 7.53 MeV (8.95 ± 0.16)×10-5 128 ± 24

16 MeV T(d.n)4He 12.66 MeV (1.19 ± 0.15)×10-4 198 ± 25

Page 14: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS14

0.

1.

2.

3.

4.

5.

Results – Fluence response

Page 15: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS15

0.

1.

2.

3.

4.

5.

Results – Dose response

Page 16: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS16

0.

1.

2.

3.

4.

5.

Conclusions

The knowledge of the dosemeter response have been expanded with new energies

An important decrease of the response in the epithermal region has been confirmed

… to be done New values at thermal and low epithermal

regions should be obtained to complete the dosemeter response function

Page 17: CALIBRATION OF THE UAB PADC BASED NEUTRON DOSEMETER

24th ICNTS17

0.

1.

2.

3.

4.

5.

Thanks for your attention!