c. thomas mcelroy et al- the ace-maestro instrument on scisat: description, performance, and...

Upload: m4m4da

Post on 06-Apr-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    1/16

    The ACE-MAESTRO instrument on SCISAT: description,

    performance, and preliminary results

    C. Thomas McElroy,1,2,* Caroline R. Nowlan,2 James R. Drummond,2,3 Peter F. Bernath,4,5

    David V. Barton,1 Denis G. Dufour,2 Clive Midwinter,1,2 Robert B. Hall,1 Akira Ogyu,1

    Aaron Ullberg,1 David I. Wardle,1 Jay Kar,2 Jason Zou,2 Florian Nichitiu,2

    Chris D. Boone,4 Kaley A. Walker,2,4 and Neil Rowlands6

    1Environment Canada, 4905 Dufferin Street, Toronto, Ontario, Canada M3H 5T42Department of Physics, University of Toronto, 60 St. George Street, Toronto, Ontario, Canada M5S 1A7

    3Department of Physics and Atmospheric Science, Dalhousie University, Halifax, Nova Scotia, Canada B3H 3J54Department of Chemistry, University of Waterloo, 200 University Avenue West, Waterloo, Ontario, Canada N2L 3G1

    5Department of Chemistry, University of York, Heslington, York, UK YO10 5DD6COM DEV Ltd., 303 Terry Fox Drive, Ottawa, Ontario, Canada K2K 3J1

    *Corresponding author: [email protected]

    Received 6 November 2006; revised 20 February 2007; accepted 21 February 2007;

    posted 21 February 2007 (Doc. ID 76743); published 20 June 2007

    The Measurement of Aerosol Extinction in the Stratosphere and Troposphere Retrieved by Occultation(MAESTRO) instrument on the SCISAT satellite is a simple, compact spectrophotometer for the mea-surement of atmospheric extinction, ozone, nitrogen dioxide, and other trace gases in the stratosphereand upper troposphere as part of the Atmospheric Chemistry Experiment (ACE) mission. We provide anoverview of the instrument from requirements to realization, including optical design, prelaunch andon-orbit performance, and a preliminary examination of retrievals of ozone and NO 2. 2007 OpticalSociety of America

    OCIS codes: 010.1280, 010.4950.

    1. Introduction

    The Atmospheric Chemistry Experiment (ACE) mis-sion on the SCISAT satellite is designed to exam-ine the composition of the Earths atmospherefrom space, with special emphasis on the middle-atmosphere ozone distribution and related tracegases in the Arctic. SCISAT is Canadas first scien-tific satellite for atmospheric sounding in over 30

    years and was launched on 12 August 2003. TheACE payload comprises two instruments: the Fou-rier transform spectrometer (ACE-FTS; hereinafterreferred to as the FTS), described in Bernath et al.[1], which operates in the infrared (IR) from 750 to4400 cm1 2.313.3 m with an unapodized spec-tral resolution of 0.02 cm1, and the Measurement of

    Aerosol Extinction in the Stratosphere and Tropo-

    sphere Retrieved by Occultation instrument (ACE-MAESTRO; hereinafter referred to as MAESTRO),which is a dual, diode-array spectrometer measuringin the UV-visible-near-infrared (UV-VIS-NIR) spectralregions with a nominal wavelength range between 285and 1015 nm (400 and 1010 nm for its primary mea-surement mode of solar occultation). The FTS instru-ment also contains two solar imagers consisting of

    128 128 pixel arrays [2]. The VIS imager arrayoperates at 527.1 nm with a full width at half-maximum (FWHM) filter bandwidth of 13.3 nm, andthe NIR imager operates at 1020.6 nm with a band-width of 19.4 nm.

    Both the FTS and MAESTRO make measurementsin the solar occultation mode and share a commonsuntracker and optical boresight. Therefore, they ex-amine almost the same slant column of air, albeit withdifferent fields of view (FOVs), and with only a slightdifference resulting from differential refractive effectsand a relatively small 0.02 (1 km tangent altitude)

    0003-6935/07/204341-16$15.00/0 2007 Optical Society of America

    10 July 2007 Vol. 46, No. 20 APPLIED OPTICS 4341

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    2/16

    pointing alignment offset between the centers of theMAESTRO and FTS FOVs. A solar occultation mea-surement occurs at the point during the orbit when thesatelliteSun vector intersects the atmosphere, andspectral measurements of the extinction of solar radi-ation can be made for determining profiles of atmo-spheric constituents. Solar occultation provides a highsignal-to-noise ratio and high-vertical resolution. Thehigh-vertical resolution is possible due to the largegeometric weighting of the absorption in the tangentlayer (the layer of the solar rays closest approach tothe Earths surface) relative to that of the layers abovein a spherical atmosphere. Approximately 15 sunriseand 15 sunset measurements are possible each dayfrom SCISATs circular 650 km and 74 inclinationorbit. In addition to the primary solar occultation mea-surement mode, MAESTRO also has the capability ofmaking measurements of solar radiation scatteredback into space by the Earths atmosphere.

    This paper describes the MAESTRO instrument,beginning with the science requirements, and thendiscusses the design, testing, and characterization ofthe instrument. The current operational data pro-

    cessing algorithms are also presented, and sampleretrieved ozone and NO2 profiles from solar occulta-tion are shown.

    2. Scientific Contribution

    A full investigation of the global ozone budget re-quires detailed species measurements (including O3,H2O, NO, NO2, N2O5, HNO3, HNO4, HCl, ClNO3, andClO) and measurements of aerosol extinction in the

    VIS, NIR, and mid-IR in order to derive the composi-tion, size, and density of aerosols and polar strato-spheric clouds (see, for example, World MeteorologicalOrganization [3]). In combination with atmosphericmodels, measurements of these species can be used toprovide valuable insights into ozone chemistry and dy-namics, differences between the Arctic and Antarcticozone distributions, the chlorine budget, and winter-time denitrification. Although the FTS is able to mea-sure aerosol extinction in the mid-IR and as many as33 trace gases, when the FTS was originally selectedfor the SCISAT payload in 1998, it was recognizedthat the capabilities of the ACE mission could begreatly enhanced by the addition of a UV-VIS-NIRspectrometer that would extend the wavelengthrange of the ACE measurements, particularly foraerosol extinction measurements, as well as provid-ing some scientific and mission redundancy. With

    this in mind, MAESTRO was selected as an addi-tion to ACE in early 2000, just over 3 years beforethe launch.

    The overall mission objectives for the ACE mission,as described by Bernath et al. [1], are to (1) betterquantify the chemical and dynamical processes affect-ing the stratospheric and upper tropospheric ozonedistribution, particularly in the Arctic; (2) explore therelationship between atmospheric chemistry and cli-mate change; (3) examine the effects of biomass burn-ing on the free troposphere; and (4) determine aerosoland cloud properties to reduce uncertainties in quan-

    tifying their effects on the global energy balance. Ex-amining the ozone budget is the highest priority for the

    ACE mission.The primary scientific enhancements to the ACE

    mission resulting from the addition of MAESTRO are(1) the provision of moderate-resolution, wavelength-dependent aerosol extinction data in the 4001010 nmwavelength range; (2) the determination of ozone andNO2 concentration profiles with higher precision and

    vertical resolution than the FTS data can provide;and (3) improved traceability of ozone measure-ments to other UV-VIS-NIR satellite measurements,such as those from the Stratospheric Aerosol andGas Experiment III (SAGE III) [4] and Polar Ozoneand Aerosol Measurement III (POAM III) [5] instru-ments.

    These enhancements are made possible fromMAESTRO solar occultation spectral measurementsin the UV-VIS-NIR, where there are strong absorp-tion features due to ozone, NO2, H2O, and O2, as wellas scattering by aerosols and molecules (Rayleighscattering). Other weakly absorbing molecules suchas the O2O2 dimer also absorb in the MAESTRO

    measurement range. From these extinction data, mo-lecular species concentrations can be retrieved, andthe residual extinction remaining after accountingfor the molecular absorption and scattering can betaken as representative of aerosol attenuation. ThisUV-VIS-NIR aerosol extinction is required to prop-erly quantify the properties of atmospheric aerosols(e.g., particle densities and size distribution) and toestimate the aerosol surface area in the same atmo-spheric regions where composition measurementsare made.

    MAESTRO also has a secondary measurementmode for observing sunlight scattered back into spacefrom the atmosphere below the satellite, permittingthe global mapping of ozone and NO2 total columnamounts. Depending on the season, atmospheric con-ditions, and location, SO2, OClO, and BrO may alsobe observable in this mode.

    The specific science goals for MAESTRO are es-sentially derived from the original desire of the

    ACE team to extend the wavelength range of themission for aerosol extinction measurements inthe UV-VIS-NIR. By virtue of these additional mea-surements in this wavelength range and the instru-ments design, MAESTRO is also able to contributein other ways to the ACE science goals. Mission andscientific redundancy is provided in occultation

    measurements of ozone and several other species.MAESTRO also provides measurements with en-hanced vertical resolution 12 km over the FTS

    vertical resolution 4 km, which is important forimproving the modeling of chemistry and dynamics ofair parcels, and in particular for studying polar strato-spheric clouds, which are highly structured in the

    vertical. These requirements andor contributions (ex-tended wavelength coverage for aerosol extinction, sci-entific andor mission redundancy, and higher-verticalresolution) established the following measurementgoals: High-vertical-resolution measurements of aero-

    4342 APPLIED OPTICS Vol. 46, No. 20 10 July 2007

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    3/16

    sol optical depth (OD) with accuracies of 0.01 OD810 km and 0.001 OD 1050 km, and aerosol ex-tinction wavelength-dependence accuracy of 0.005 ODper 100 nm; high-vertical-resolution ozone profileswith accuracies of 15% 810 km, 10% 1020 km,and 3% 2050 km (assuming a midlatitude profile);and high-vertical-resolution NO2 profiles with ac-curacies of 25% 810 km, 15% 1020 km, 10%2040 km, and 15% 4050 km (assuming a mid-latitude profile). Additional species (H

    2O, SO

    2, OClO,

    BrO) may be observable in occultation or backscatterin addition to the primary products of ozone, NO2,and aerosol extinction.

    To our knowledge, this is the first time an instru-ment measuring in the UV-VIS-NIR and an instru-ment measuring in the mid-IR have been placed onthe same spacecraft to view a nearly identical beam ofsolar radiation from space. As such, the direct com-parison of results from MAESTRO and the FTS is ofparticular scientific interest. These simultaneous ob-servations should lead to improvements in the under-standing of the spectroscopy of molecules measuredby both instruments and the retrieval methodologies

    employed. This has direct application to understand-ing the relationships between measurements fromother instruments made using different spectroscopictechniques, such as between the HALogen Occulta-tion Experiment [6] (HALOE), which measures in theIR, and SAGE III, which measures in the UV-VIS-NIR. In addition to providing colocated trace gas andaerosol measurements, this coalignment also allowsMAESTRO data analysis to use the tangent pointingand pressure-temperature profiles derived by theFTS CO2 retrievals [7].

    3. MAESTRO Instrument

    A. Spectrometer

    MAESTRO comprises two independent diode-arrayspectrometers. One spectrometer measures UV and

    visible spectra from 285 to 565 nm with a spectral

    resolution of approximately 1.5 nm, while the sec-ond measures VIS and NIR spectra from 515 to1015 nm with a resolution of approximately 2 nm.These spectrometers will be referred to as the UV and

    VIS spectrometers, respectively. The main featuresof MAESTRO, including the properties of the twospectrometers, are summarized in Table 1. Whilethe nominal wavelength range of measurements is2851015 nm, the wavelength range that can be cali-

    brated on-orbit in occultation by the matching of solarFraunhofer features in observed spectra to theirknown wavelengths is smaller (400545 nm for theUV spectrometer and 5201010 nm for the VIS) andis also listed in Table 1. These wavelength-calibrationlimits depend on the availability of sufficient signal tonoise as a function of wavelength, and the occurrenceof Fraunhofer lines that can be used for wavelength-pixel matching. Wavelengths past these limits areextrapolated. The above limits are for occultationspectra; backscatter spectra wavelength calibrationin the UV can be extended below 300 nm.

    Each individual spectrometer is relatively simple

    and small, consisting of only a lens, slit, concave ho-lographic grating, and photodiode array detector. Insolar occultation mode, the two spectrometers areilluminated by light from two separate, folded three-mirror telescopes coupled by beam splitters. The en-tire MAESTRO instrument is approximately a 0.5 mcube, weighs 8 kg, and consumes 14 W peak power.Figure 1 shows a photograph of the MAESTRO in-strument, labeled with the locations of its solar andbackscatter ports. A schematic of the instrumentsoptical design is shown in Fig. 2. The spectrometerdesign draws heavily on heritage from the Meteorolog-ical Service of Canadas SunPhotoSpectrometer (SPS),which was developed for flight on the Space Shuttle in1992 and used extensively as part of the NASA ER-2stratospheric chemistry research program in the 1990sas the Composition and Photodissociative Flux Mea-surement (CPFM) instrument [8,9].

    Table 1. Main Characteristics of the ACE-MAESTRO Instrument

    Spectrometer

    UV VIS

    Nominal wavelength range 285565 nm 5151015 nm

    Calibrated wavelength range in occultation 400545 nm (pixels 463954) 5201010 nm (pixels 211010)

    Spectral resolution 1.5 nm 2 nm

    Pixel spacing 0.3 nm 0.5 nmMain sources of extinction O3, NO2, aerosol and

    molecular scatteringO3, O2, O2O2, H2O, aerosol

    and molecular scattering

    Secondary sources of extinction SO2, OClO, BrO NO2SCISAT orbit circular, 650 km altitude, 74 inclination

    Instrument FOV 1.2 km (elevation) 25 km (azimuth) (at 22 km tangent)

    Observation modes Solar occultation

    Nadir backscatter

    Detectors 1024 pixel, Reticon photodiode array

    Gratings Concave holographic, 94 mm focal length

    Signal to noise 10003000 (high Sun)

    10 July 2007 Vol. 46, No. 20 APPLIED OPTICS 4343

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    4/16

    During an occultation, 7% of the beam from theFTS primary pointing mirror is directed through aquartz window to the MAESTRO input optics using apickoff mirror inside the FTS enclosure. The beam isthen split between the two MAESTRO telescopes us-ing two beam splitters situated just inside the solarinput port. The first polka-dot beam splitter directsapproximately two-thirds of the beam power to theUV spectrometer to help match the different signalsin the UV and VIS regions, which result from thegreater atmospheric scattering and absorption occur-ring in the UV spectral range. Most of the remaininglight is passed to the VIS spectrometer by the secondbeam splitter and a small amount 10% of residuallight is directed through an exit port, which was alsoused for alignment during prelaunch activities. Tele-scopes on each spectrometer provide an image mag-

    nification of 7.The FOV of each spectrometer is designed to in-

    clude both the solar and the backscatter views bycombining them with beam splitters. In solar occul-tation mode, the UV and VIS spectrometers use lightfrom the suntracker entering through a common viewport, but in backscatter mode, each spectrometer hasits own input port. An 8020 beam splitter situated atthe entrance port of each spectrometer (after the so-lar telescope) allows the transmission of 80% of inputradiation in backscattered view, but only 20% of theradiation if the instrument is observing the Sun.

    Each spectrometer, in spite of its small size, has antendue (A 1.5 105 cm2 sr) similar to thatof the Global Ozone Monitoring Experiment [10](GOME) instrument by virtue of the spectrometerslarge aperture and is therefore much more sensitivethan needed for solar occultation observations; as aresult, the loss of signal through the solar optics is ofno consequence.

    The entire satellite bus is pointed toward the Sunwith modest pointing accuracy 1 using magnetictorque rods. An internal high-speed, two-axis sun-tracker pointing mirror, driven by a signal from aquadrant sensor operating at 1.55 m, is used to di-rect the FTS FOV toward the radiometric center of thesolar image. A pickoff mirror directs solar radiationfrom this suntracker to the MAESTRO spectrome-ters during occultation. Both the solar and backscat-ter ports are permanently open. To make backscattermeasurements, the MAESTRO solar and backscat-ter FOVs are oriented 160 apart, with the backscat-ter FOV defined relative to the satellite coordinatesat a fixed 20 off the antisolar vector, toward the

    nadir direction. When the instrument makes solaroccultation observations at a tangent altitude of100 km (essentially exoatmospheric), the backscatterFOV points into dark space. When the solar FOV istangent at the satellite altitude (650 km), the back-scatter FOV will point at the ground aft of the satel-lite at sunrise and ahead of the satellite at sunset ata point where the local solar zenith angle is approx-imately 80. Once the local solar zenith angle at thesatellite reaches 90, the suntracker mirror is pointedoff the Sun so that the solar input will be zero, andthe backscatter signals will dominate the input.

    Fig. 1. Photograph of MAESTRO instrument, showing occulta-tion and backscatter viewing ports.

    Fig. 2. Schematic of MAESTRO optics.

    4344 APPLIED OPTICS Vol. 46, No. 20 10 July 2007

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    5/16

    After passing through the 8020 beam splitter,each beam is fully linearly polarized by a GlanTaylor polarizing prism. This prevents anomalousdiffraction features of the diffraction gratings (Woodsanomalies [11]) from interacting with the scene in-tensity so that polarization will not influence themeasurements during the backscatter mode. The an-gle between the polarizer and the spacecraft coordi-nates is fixed; therefore, the plane of polarization isalways the same relative to the SunEarth-line-of-sight angle. The polarization for the backscatter port,determined as the direction of the electric vector, isperpendicular to the plane of the Earth, Sun, andsatellite. The VIS beam also passes through an order-sorting filter, consisting of OG515 glass, to preventlight from higher orders of diffraction contributing tothe measured signal. The UV spectrometer has nosecond-order diffraction problemsthanks to the Planckfalloff at short wavelengths for solar radiation.

    The 50 m 2.5 mm entrance slits of the two spec-trometers are projected to infinity by UV-corrected,22 mm focal-length achromatic lenses. After passingthrough each slit, light is diffracted by a 5 cm

    5 cm, glass-substrate, 94 mm focal-length concaveholographic diffraction grating, custom made by

    American Holographic Ltd.The two independent MAESTRO spectrometers are

    fabricated on a common mechanical structure made ofa 1 mm thick aluminum sheet. This spectrometer as-sembly contains the 22 mm lenses, slits, gratings,and detectors and is mounted inside the spectrometerhousing denoted in Fig. 1. A single mount carries thediffraction gratings of both halves of the instrument.The grating mount is attached to the spectrometerstructure by crossed-spring, wire flexure pivots and atorsion element. The angle of the gratings, the re-maining degree of freedom of the grating mount, canbe adjusted using a thermally scanned push rod toexamine the sub-pixel-scale structure of the solarspectrum as observed by the instrument. The designof the grating mount includes temperature compen-sation of the focus of the spectrum so that changes intemperature do not cause changes in the instrumentline-shape function (ILS). Figure 3 is a photograph ofan early prototype of the MAESTRO spectrometerassembly, showing the kinematic sheet-metal struc-ture that ensures a very stable and lightweight in-strument.

    The slit-lens-telescope combination of MAESTROis designed to have a instrument FOV equivalent to

    roughly 1 km from SCISATs 650 km orbit. A narrowvertical FOV is achieved by aligning a narrow 0.022angular-extent axis of the entrance slit with theEarth radial direction using the orientation of thespacecraft. In the perpendicular (horizontal) direc-tion, the slits larger angular extent (0.65 on the UVand 0.89 on the VIS, as measured during preflightcharacterization) is designed to be centered on theSun and larger than the Suns width 0.5 to re-duce the instruments sensitivity to small horizontalsatellite pointing motions (yaw), to capture as muchsignal as possible, and to permit averaging over in-

    tensity variations across the width of the Sun. ThisFOV also results in an instantaneous backscatterfootprint of approximately 1.7 km 70 km fromSCISATs 650 km orbit, although the long detectorintegration times used in backscatter measurementsresult in an effective footprint as large as 70 km 70 km.

    Two Reticon 1024-element, randomly address-able photodiode-array detectors, of the same typeused in the GOME satellite instrument, detect thediffracted spectra. The detectors are intimately in-terconnected with a soft-programmable, XILINXfield-programmable gate array-based (FPGA) mi-croprocessor system, which selects the hardwarereadout algorithm and controls the timing of theobservations made by the detector as well as coad-ding spectra and communicating with the instru-ment control computer. There is one computersystem for each detector mounted on the readoutelectronics boards and one for the instrument con-trol and communications functions. A fully synchro-

    nous state machine is embedded in the readoutelectronics FPGA as a peripheral of the 8 bit micro-computer. This state machine controls the observa-tion process and orchestrates the collection andcoadding of spectra during a period when the mi-crocomputer is actually turned off to avoid the gen-eration of asynchronous noise.

    Occultation measurements span a very large dy-namic range, both as a function of tangent altitudeand as a function of wavelength; observation tangentheights range from the surface to 100 km over an at-mosphere whose density changes exponentially with

    Fig. 3. Photograph of prototype MAESTRO spectrometer assem-bly. The corner of one grating is just visible in the lower left-handcorner of the diagram.

    10 July 2007 Vol. 46, No. 20 APPLIED OPTICS 4345

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    6/16

    altitude, while the strong HartleyHuggins ozone ab-sorption bands 350 nm, in combination with theRayleigh scattering that increases as 4, remove mostradiation in the UV. This large dynamic range pre-sents a challenge in terms of the design of theMAESTRO detection electronics if optimal measure-ments are to be made. To address this issue, theMAESTRO detector electronics have been designedto operate in various measurement modes through

    the use of a randomly addressable detector, whereparticular pixels of the detector can be integratedand read out independently from other pixels in thearray.

    The design of the Reticon photodiode array permitsits use in various ways. Internal transmission gatesallow individual pixels to be either isolated from acommon video line or connected to it. This selectioncan be done in one of two ways: individual gates canbe turned on using an internal decoder that allows a10 bit address word to select a particular pixel to beread and reset, or a separate input can be used toconnect all pixels simultaneously to the video line,causing all pixels to be simultaneously reset to anexternal reference potential of 2.5 V. In both modes,the current to reset the detector elements is suppliedby an electrometer amplifier tied to a precision volt-age reference. The measurement of light flux is per-formed by determining the amount of charge lostbecause of exposure to light when the pixels are reset.

    A high-resistance feedback resistor 300 k in theelectrometer circuit produces a measurable signal,which is amplified and then digitized by a radiation-hardened, 16 bit analog-to-digital (AD) converter.

    MAESTRO measurements use two main detectorreadout modes. The first mode, the multiple resetmode, can be used for very bright observations. At

    very high signal levels, the short integration times10150 s are comparable with the time for an in-dividual pixel readout 14 s, so that it may bethat only one pixel can be read between simultaneousresets of the array pixels before other pixels saturate.This type of observation begins with the startingpixel address placed on the array address bus and thearray held in the reset state. In this state, the elec-trometer resistor is shorted by a transmission gate,and the pixels are all continuously recharged via thereset circuitry. The pixel is read by removing thereset, waiting for a short integration time (e.g.,10150 s), and then asserting the address strobe to

    read out the selected pixel. The process of reset, in-tegrate, and read is repeated for each pixel separatelyuntil the entire array has been interrogated.

    The second mode is the parallel integration mode.For low light levels and long detector integrationtimes, all pixels on the array are integrated simulta-neously, and then each pixel is read out in sequence,with no pixel resets between reads. The observationbegins with the array reset control line holding alldiodes in reset state, and after a predefined integra-tion time, the address strobe is asserted. This gatesthe first, addressed pixel onto the video line where

    the reset charge is measured and digitized. Subse-quently, the address strobe is held unchanged, andeach of the photodiodes is read out, in turn, by chang-ing the pixel address applied to the array followed by

    AD reads. Changing the address automatically iso-lates the already reset, selected photodiode pixel andcharge, which then begins integrating for the nextscan. The majority of MAESTRO measurements inoccultation are made with midrange integrationtimes, which require a variation of the parallel inte-gration mode readout system. In this mode, the arrayis divided into a number of equal-sized readoutgroups consisting of 8512 pixels. After being reset,the entire array is integrated, all pixels within agroup are read out, and then the entire array is againreset and integrated, and the next group of pixelsread out until the full spectrum is assembled.

    The array is always read from the long-wavelengthto the short-wavelength end of the detector to ensurethat the pixels with the highest light levels (when theSun is attenuated by the atmosphere) have theshorter total integration times. Each spectral mea-surement uses the same base integration time for

    every pixel on the array (determined from the ex-pected signal at each tangent height), but the timerequired for signal readout of each pixel means apixel continues to integrate while waiting for otherpixels to be read. Therefore, the effective integrationtime of a pixel is always longer than the base inte-gration time and depends on its position on the arrayrelative to the first pixel in the readout queue.

    Each occultation observation is based around ameasurement cycle of 330 ms, chosen so that it ispossible for measurements to be made at 1 km alti-tude intervals in the case where the beta angle (theangle between the satellite orbital plane and theEarthSun vector) is zero. Each spectral measure-ment consists of a preread measurement in the mul-tiple reset mode with a fixed integration time for allaltitudes and another measurement optimized bymode and integration time for the expected light levelat the particular tangent altitude. These optimizedmeasurements usually consist of two to three indi-

    vidual spectra that are coadded to maximize thesignal-to-noise ratio within the 330 ms measurementtime. The number of pixels in a group and the inte-gration time are chosen according to the expectedsignal level at a given tangent height based on aradiative transfer model of the atmosphere. Com-mand tables defining the readout parameters are up-

    loaded to the satellite and stored in the instrumentmemory. They are produced for different beta anglesto match the observation timing to the timing of theoccultation. Figure 4 describes a sample series ofmeasurement parameters used for a single occulta-tion. Based on the expected solar occultation signal atevery tangent height, each observation is assigned (1)a pixel integration time, (2) the number of pixels tointegrate in a group before resetting the detector toprevent saturation during detector readout for thatintegration time, (3) the resulting number of readoutgroups of pixels across the 1024-pixel array, and (4)

    4346 APPLIED OPTICS Vol. 46, No. 20 10 July 2007

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    7/16

    the number of coadded spectra, n, required to com-plete a measurement cycle of roughly 330 ms. Theselection of the number of pixels in a group is depen-dent on the integration time, and the same groupcount (but different n) may sometimes apply toslightly different integration times, as evidenced bythe step features in the integration time in the16 pixels 64 groups n, 128 pixels 8 groups n, and 1024 pixels 1 groups n bands.

    B. Preflight and On-Orbit Characterization

    The MAESTRO and FTS instruments were charac-terized together at the Instrument CharacterizationFacility at the University of Toronto during Februaryand March 2003. The goal of the characterizationtests was to assess the performance of the instru-ments in a simulated orbital environment. To achievethis aim, the instruments were mounted on a replicaof the satellite baseplate and placed in a vacuumchamber for the duration of the characterization pe-riod. A cylindrical liquid-nitrogen-cooled shroud waspositioned around the instruments to simulate thethermal environment of space, and a liquid-helium

    target positioned adjacent to the FTS detector radi-ator cone was used to simulate the FTSs deep-spaceradiator view. Fluid at various temperatures waspumped through a coiled pipe in thermal contact withthe bottom of the baseplate to simulate the thermalfluctuations encountered on orbit due to periodicchanges in thermal loading from the Earth and Sun.

    A suite of characterization tests was repeated fordifferent baseplate temperatures ranging from 25to 25 C.

    Light from various sources was directed through awindow into the vacuum tank to the FTS pointing

    mirror and, in turn, a portion of this light was passedon to MAESTRO as in an occultation measurement.

    A solar simulator source was built to simulate an oc-cultation measurement for both instruments by com-bining IR light from a high-temperature 3000 Kblackbody source with VIS light from a 1000 W quartz-halogen lamp using a germanium beam combiner. Bycomparing MAESTRO laboratory spectra with solarspectra taken on orbit, the irradiance produced at theinstrument by the quartz-halogen lamp was deter-mined to be roughly 1

    100of that of the Sun. The com-

    bined beam was directed through a gas cell andfocused on a field stop with a circular aperture equiv-alent to the angular extent of the Sun. A slowly movingknife-edge shutter was placed next to the field stop tosimulate the effect of the solar disk changing shapeduring satellite sunrises and sunsets. The system wasused to perform several instrument characterizationtests, and also to make ozone and NO2 gas cell mea-surements simultaneously with the FTS [12,13]. Du-four et al. [12] describe the design of the opticalsystem in detail, as applied to the ozone gas cell tests.In addition to tests performed with the main optical

    test system and solar simulator, several bench testswere performed on MAESTRO using lasers and emis-sion lamps to examine instrument line shape andwavelength-pixel dispersion.

    In combination with preflight calibration, on-orbittests and data analysis allow full and updated instru-ment characterization. The following sections sum-marize the characterization of the instrument fromthe combination of preflight and on-orbit tests.

    1. Instrument Line ShapeDiode lasers at wavelengths of 532, 670, and 830 nm,as well as a mercury emission lamp, were used pre-flight to examine the instrument line shape. TheFWHM of the line shape is used as a measure of theinstruments spectral resolution. The FWHM changessomewhat across both detectors but was found to beapproximately 1.5 nm in the UV, and between 1.9 nmand 2.2 nm in the VIS instrument. Figure 5 showsthe VIS instrument line-shape function determinedfrom a laser measurement at 830 nm, as well asadditional line shapes determined at nearby wave-lengths on-orbit. The on-orbit VIS line shapes weredetermined by deconvolution from observed on-orbitspectra and modeled high-spectral-resolution absorp-tion spectra. The high-resolution absorption of molec-ular oxygen was modeled for a set of observations in

    the O2 A band (at 762 nm) and the B band (at690 nm) using a line-by-line radiative transfer modelwith O2 line parameters from HITRAN 2004 [14] andthe ACE a priori temperature profiles described inBoone et al. [7], which at altitudes below 30 km arederived from the operational analyses of the GlobalEnvironmental Multiscale model of the Canadian Me-teorological Centre. These theoretical high-resolutionspectra were used to deconvolve an average line shapefrom observed MAESTRO spectra. TheA- andB-band-determined line shapes are similar in shape to eachother and differ mainly in their FWHMs, which is to be

    Fig. 4. Schematic of sample detector measurement parameters(number of pixels read in a group before detector reset numberof resulting pixel groups across detector number of coadded

    spectra n), and pixel integration times as a function of tangentheight for one occultation sequence.

    10 July 2007 Vol. 46, No. 20 APPLIED OPTICS 4347

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    8/16

    expected as the spectral resolution changes across thedetector. They suggest that the on-orbit instrumentline shape is more asymmetric than the one originallymeasured in the laboratory. The difference in theasymmetry of the line shape between the prelaunchtests and the on-orbit retrievals is most likely due tochanges in the intensity distribution of light across theinstrument aperture. While a good effort was made touniformly fill the FOV during ground tests, there mayhave been a difference between the distribution of lightin the tests as compared with the filling of the FOV onorbit.

    2. Signal-to-Noise RatioThe analysis of high-Sun spectra indicates that boththe UV and VIS spectrometer pixels have an averageof4.1 AD counts of readout noise 1. This is animprovement over the approximately seven countsof readout noise measured in the laboratory duringcharacterization. At maximum levels, the signal pro-

    vides an additional random (shot) noise component of

    nearly four counts. While the 16 bit AD converterallows 65,536 counts on the detector, in practice a fullwell readout from the detector is designed to be ap-proximately half-scale. At half-scale on the detector,the signal-to-noise ratio can be as high as 3000. How-ever, observations are generally limited to 10,000counts to avoid any risk of saturation or possiblenonlinearities at higher count levels, and most high-Sun measurements are designed to operate at peaksignal-to-noise ratios of approximately 1500.

    Figure 6(a) shows the signal and estimated noiseas a function of wavelength and corresponding pixel

    number for a typical high-Sun measurement col-lected during sunset occultation ss4043 (13 May 200409:48:56UT at 65.3 N, 75.2 E, referenced to the 30km subtangent location). The signals in the figure areshown in detector counts per microsecond after cor-rection for each pixels effective integration time.Their corresponding noise values are shown in thesame units. These spectra are uncorrected for abso-lute response, and therefore, the relative magnitudesof the UV and VIS signals are not expected to appearconsistent. The noise is a function of readout noise,shot noise (and therefore total signal), and the num-ber of coadded spectra (in this particular case, the UV

    signal results from only one spectrum, but the VIS isa product of onboard coadding of three spectra). Fig-ure 6(b) shows the resulting signal-to-noise ratio ofthis high-Sun measurement. The VIS spectrometerhas been read out in the 8 pixel per group parallelintegration mode; the individual pixel readout times14 s are comparable with the individual basepixel integration times 308 s, and therefore thelast pixel read in a group experiences a noticeablylonger effective integration time than the first pixelas it waits for readout. This difference in signalacross a group of eight pixels is quite apparent in the

    Fig. 5. Instrument line shape for the VIS spectrometer, showingthe ILS measured preflight with an 830 nm diode laser, as well astwo ILS functions determined by deconvolution of the ILS fromoccultation measurements in the O2 A and B bands.

    Fig. 6. (a) Signal and noise levels for a sample high-Sun mea-surement collected at the beginning of sunset occultation ss4043(13 May 2004 09:48:56UT at 65.3 N, 75.2 E). Both the signal andnoise have been corrected for effective integration times and areshown in detector counts per microsecond. (b) The measurementsresulting signal-to-noise ratio. The curves representing noise andsignal to noise on the VIS spectrometer show structure in eightpixel groups resulting from the use of the Parallel Integrationreadout mode using 8 pixels per readout group.

    4348 APPLIED OPTICS Vol. 46, No. 20 10 July 2007

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    9/16

    noise and signal-to-noise spectra, which cycle in am-plitude every eight pixels (however, not in the signalspectra, which have been corrected for this effect).The UV spectrum was collected in the multiple resetmode where each pixel is integrated separately, andthen the entire array is reset after each read, so thatall effective pixel integration times are the same, andits noise and signal-to-noise do not show featuresfrom pixel grouping.

    3. Absolute ResponseThe absolute response of the spectrometer was mea-sured in the laboratory using a Spectralon Lambertiandiffuser plate illuminated by the quartz-halogen lamp.The solar UV spectra showed a maximum absoluteresponse of 1.3 109 countssW m2 nm1, whilethe peak sensitivity of the VIS spectra was 6.5 108 countssW m2 nm1. Knowledge of the ab-solute sensitivity is not strictly necessary for occul-tation measurements, which require only the solarreference spectrum measured at high Sun, but itallows a proper comparison of instrument responsebefore and after launch with comparisons of high-Sun

    measurements on orbit. As can be observed in Fig. 6,on orbit, all solar radiation at wavelengths less than400 nm is virtually extinguished from the solar spec-tra. This was not observed in quartz-halogen lampspectra collected on the ground where the responsiv-ity across the UV detector was strong for all wave-lengths. The cause of this reduction in intensitycannot be decisively determined with the instrumentin orbit. Only the occultation spectra are affected; thebackscatter spectra contain the short wavelength sig-nals below 400 nm. As a result, molecules absorbingat wavelengths less than 400 nm are unfortunatelynot observable in occultation, including BrO, OClO,

    and SO2, and the HartleyHuggins bands of O3. For-tunately, the primary data products of O3 (measuredin the Chappuis band by the VIS spectrometer, andthe short-wavelength edge of the Chappuis band bythe UV spectrometer), NO2, and aerosols, as well asH2O and O2, are not affected.

    4. Field of ViewDuring preflight characterization, FOV maps of thespectrometer slits were obtained by deflecting thebeam of an incoming laser with wavelength 532 nmto a variety of angles using the suntracker mirror andmeasuring the resulting signal intensities on bothspectrometers. The FOV coincidence of the two slits

    was confirmed, and it was determined that the longaxes of the slits were parallel with each other to 0.4.The elevation extent of both slits was determined tobe 0.022, defined by the FWHM of the response. Theazimuthal extents were measured at 0.65 UV and0.89 VIS. The FWHM slit size results in an instru-ment FOV of 1.2 km in the vertical and approxi-mately 35 UV and 45 VIS km in the horizontal fora 22 km tangent altitude. During an occultation thesignal comes only from the solar disk and the sig-nal extent in the horizontal is then 25 km. Afterlaunch, scans across the solar disk using the sun-

    tracker mirror corroborated the FOV knowledge onorbit.

    4. Measurements

    SCISAT was launched under a cooperative launchagreement with NASA on 12 August 2003 by a Pe-gasus Launch Vehicle released from a Lockheed 1011aircraft off the coast of California, near Vandenberg

    Air Force Base. As soon as the satellite was in orbit,the FTS and MAESTRO were turned on. An exten-sive engineering commissioning phase ensued, lead-ing up to full instrument operations in February2004. Engineering tests, beginning in October 2003,involved collecting test spectra to determine instru-ment health. These included occultation observa-tions, exercising the thermal scan rod, measuring theinternal photodiode test light sources, and a trialbackscatter observation. MAESTRO began regularoperations on 21 February 2004. As of mid-January2007, MAESTRO had collected 7175 sunrise, 8110sunset, and 1787 backscatter observations.

    An occultation sunrise or sunset measurement se-quence typically takes 13 min, depending on the

    angle of the Sun relative to the orbital plane, andconsists of approximately 60 spectra collected at tan-gent altitudes between 0 and 100 km (however, mostuseful measurements do not extend below 510 kmdue to the presence of clouds) and 20 spectra collectedbetween 100 and 150 km for calculation of a high-Sunreference spectrum. A series of 80 spectra are alsocollected using identical detector commands, eitherimmediately before (sunrise) or after (sunset) the oc-cultation when the Sun is below the horizon, to use inthe calculation of detector dark current. Measure-ments from the beginning of the mission to August2005 tend to be densely spaced in the vertical attangent altitudes below 50 km, with a spacing be-tween spectra of300 m to 2 km, while spectra werecollected only every 5 km above a 50 km tangent. Themeasurement command tables were modified on 10

    August 2005 to allow a 2 km vertical spacing betweenmeasurements at altitudes above 50 km. Occulta-tions collected after this date contain between 60 and80 spectra collected at tangent heights below 100 km.

    Backscatter measurements are a secondary mea-surement mode and are only collected as opportunitypresents. The availability of satellite command andcontrol resources primarily determines the frequencyof backscatter measurements. Each backscatter mea-surement command results in the collection of 160

    spectra in sequence, cycling between three differentintegration times to ensure spectra are neither satu-rated nor too weak with changing albedo below thesatellite.

    5. Data Processing: Version 1.2 Retrieval Algorithm

    The MAESTRO retrieval algorithm is still under de- velopment. Here we present the algorithm for theversion 1.2 release of MAESTRO data. The process-ing of MAESTRO occultation data involves two majorsteps. The first is the conversion of raw data (Level 0)to spectra calibrated in wavelength and corrected for

    10 July 2007 Vol. 46, No. 20 APPLIED OPTICS 4349

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    10/16

    various instrument parameters, such as pixel inte-gration times and dark current, as well as instrumentstray light (Level 1), and the second is the conversionof calibrated spectra to extinction, slant column den-sities, and vertical profiles (Level 2). In this section,the data processing is described with reference to theprimary occultation measurement mode. Althoughthe initial Level 0 to Level 1 processing and spectralfitting methods also apply to backscatter measure-ments, backscatter measurements are a secondarymeasurement goal and have not yet been examinedextensively, and no results will be presented in thispaper. Version 1.2 MAESTRO data provide ozone andNO2 profiles, as well as total wavelength-dependentatmospheric optical depth extinction. The analysis ofozone and NO2 are presented here. The developmentof the aerosol product is still ongoing and results arenot discussed in detail in this paper.

    A. Level 0 to Level 1 Processing: Calibration

    The data from the SCISAT spacecraft are downlinkedto several ground stations, assembled at the CanadianSpace Agency (CSA) Mission Operations Centre in St.

    Hubert, and transferred to the Science OperationsCentre (SOC) computers at the University of Water-loo, where the mission archive resides. MAESTROdata are also archived separately at the University ofToronto and Environment Canada in Toronto, On-tario. The Level 0 MAESTRO data are archived asbinary files, and contain science data from the instru-ment as well as instrument engineering data collectedfrom the main SOC database. Geolocation data arealso provided by the SOC.

    The conversion of raw spectra to calibrated spectrahas two major steps: correcting raw spectra and assign-ing wavelengths to the spectra. The signal analyzedin an occultation measurement is the wavelength-dependent transmission of the atmosphere, which isdefined by the ratio of the intensity of the spectrumcollected during the occultation, when the atmosphereis absorbing and scattering, to a high-Sun referencespectrum, which can be collected at any point in theorbit when the satelliteSun vector does not intersectthe atmosphere. In practice the high-Sun spectra aretaken as close in time to the atmospheric spectra aspossible. As long as the instrument can measure areliable high-Sun reference spectrum from orbit, thedata do not require absolute calibration using the in-strument responsivity. However, the Level 0 to Level1 processing must apply a number of corrections to

    the raw data to remove the effects of the variablepixel integration times across the detector arrayand the different readout modes between spectra,and to correct for instrumental stray light, elec-tronic fixed-pattern offset, dark current, and chargecarryover caused by resetting the detector veryquickly.

    The on-orbit wavelength calibration uses Fraun-hofer solar absorption features to determine the re-lationship between each pixel and its correspondingwavelength. The pixel-wavelength calibration fits thewavelengths of an observed MAESTRO high-Sun ref-

    erence spectrum from each occultation to those fromthe higher spectral resolution (0.5 nm) ATLAS-1 com-posite solar reference spectrum of Thuillieret al. [15],degraded to the MAESTRO resolution by convolutionwith the MAESTRO instrument line-shape function.While the ATLAS-1 solar reference spectrum is cali-brated in absolute intensity, the observed MAESTROsolar spectrum is not. In addition, the observed spec-trum is contaminated with a cyclic change in sen-sitivity because of etaloning on the detector. Toproperly align the spectral features, a high-pass filteris applied to each spectrum so that only the high-frequency spectral information remains. These high-frequency spectra are then normalized to each otherin intensity for the purpose of the wavelength-calibration procedure.

    An iterative, least-squares minimization is used tofit the wavelengths of the observed high-pass-filteredspectrum to the ATLAS-1 spectrum by determiningcoefficients for a polynomial that describes the wave-lengths as a function of pixel number on the array. Itwas determined empirically that the use of a seventh-order polynomial provides an acceptable match over

    the spectral fitting window.Figure 7 shows all MAESTRO occultation spectra in

    detector counts per microsecond for occultation num-ber ss4043, after the detector and wavelength calibra-tions described above have been applied. For clarity,all the VIS spectra have been scaled by one constantfactor to match the UV at the long-wavelength end ofthe UV spectrum. The most intense spectrum is a solarreference spectrum collected just before sunset, at atangent height of approximately 150 km. As thesespectra are not calibrated in absolute intensity, theinstrument responsivity is still visible in the falloff ofintensity at short wavelengths in the VIS spectra(caused by the OG515 order filter), and there is visible

    Fig. 7. MAESTRO UV (blue) and VIS (red) spectra in detectorcounts per microsecond from occultation ss4043, after scaling fordetector integration time, fixed-pattern offset, dark count, andgrating stray light. The VIS spectra are scaled by a constant factorto align with the intensity of the UV spectra at the UVs long-wavelength end.

    4350 APPLIED OPTICS Vol. 46, No. 20 10 July 2007

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    11/16

    evidence of detector etaloning in the oscillatorychanges in sensitivity as a function of wavelength (e.g.,the broad peak features between 450 and 550 nm inthe UV and 600 and 750 nm in the VIS spectra).Several prominent atmospheric absorption featuresare immediately visible in these calibrated spectra,including the O3 Chappuis absorption band near600 nm, water vapor near 950 nm, and the O2 A and

    B bands near 762 and 690 nm, respectively.

    B. Level 1 to Level 2 Processing: Optical Depths, SlantColumn Densities, and Vertical Profiles

    To derive constituent amounts from the Level 1 spec-tra, a nonlinear, least-squares spectral fitting code isused to fit forward-modeled effective optical depths(at MAESTRO resolution) to those apparent opticaldepths observed during the occultation. The opticaldepth spectra are calculated by

    p ln IpI0p, (1)where p is the optical depth at each pixel with

    wavelength p, Ip is the occultation spectrum, andI0p is the reference spectrum at high Sun for wave-length p. As long as the instrument responsivityremains constant between the time the solar refer-ence spectrum, I0, is taken and the occultation spec-trum, I, is measured, the instrument responsivityfunction is removed in the calculation of opticaldepth, and therefore its knowledge is not required.Figure 8 shows optical depth spectra for several sam-ple tangent heights derived from the ss4043 occulta-tion and high-Sun spectra shown in Fig. 7.

    The v1.2 optical depth calculation assumes thatMAESTRO views the same part of the Sun during

    exoatmospheric and occultation measurements andthat inhomogenieties across the Suns surface may beignored. MAESTRO observations use light integratedacross the full width of the solar disk to minimize theimpact of spatial variations of intensity across the

    Suns surface. The suntracker is designed to refer-ence the center of the Sun, and on-orbit tests con-firm its performance. There will be some variation inthe FOV location on the Sun due to the effectivemovement of the FOV in the Earths radial direction,especially at lower tangent heights 20 km, as re-fraction and attenuation distort the solar image andshift the pointing direction upward. These effects areunder study using scans done across the solar diskwhen collecting exoatmospheric spectra. Appropriatecorrections to future versions of the retrieval algo-rithm will be made if required. Data comparisonsdone to date suggest that these effects are not likelyto be large [16].

    A global-fitting algorithm, similar to that describedby Carlotti [17] for IR limb sounding, has been devel-oped for the MAESTRO analysis, but has not yet beenimplemented operationally. A global-fitting code al-lows the retrieval of constituent profiles from a si-multaneous fit of all MAESTRO spectra during anoccultation. The current data release (version 1.2)uses a classic two-step method to derive vertical vol-ume mixing ratio profiles. Spectral fitting is used to

    derive a slant column density (SCD) for each spec-trum, and these slant column densities are subse-quently used to calculate a profile. The retrievalprocess does not make use of any a priori informationor other auxiliary constraints such as smoothing. Thespectral-fitting code was originally developed for fit-ting spectra from MAESTROs precursor instrument,the CPFM, used on the ER-2 aircraft [8], and re-trieves the apparent SCD of a trace gas for eachindividual spectrum along the line of sight of thatobservation. For radiation traveling along a path oflength L, the optical depth is defined by

    0

    L

    , lnldl, (2)

    where , l is the cross section representing theabsorption features of a particular absorber and maybe dependent on local pressure or temperature alongthe path, and nl is the number density of the con-stituent in segment dl of the path. The SCD, X, rep-resents the total amount of the constituent along theline of sight of the observation (units of molecules perarea), 0

    L nldl. In terms of SCD, the total monochro-matic apparent optical depth (the sum of opticaldepths of all individual constituents) is written as

    O3 NO2 Rayleigh aerosols,

    O3XO3 NO2XNO2 RayleighXRayleigh cXc lXl sXsaerosols. (3)

    In addition to retrieving trace-gas amounts, eachspectral retrieval also determines the magnitudes ofindividual constant Xc, linear Xl, and quadraticXs SCD offsets that are used to account for back-ground aerosol absorption by using effective cross

    Fig. 8. MAESTRO UV (blue) and VIS (red) optical depth spectrafor four sample tangent altitudes from occultation ss4043.

    10 July 2007 Vol. 46, No. 20 APPLIED OPTICS 4351

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    12/16

    sections for to represent the wavelength depen-dency of the offsets. Wavelength shift and stretchterms are also fit to fine-tune the pixel-wavelengthassignments to within 0.01 nm. Rayleigh scatteringis accounted for by retrieving an air column densityusing a Rayleigh scattering cross-section vector.

    The various main terms fit and modeled in eachspectrum are described in Table 1 as main sources ofextinction. The spectral fits are performed across awide range of wavelengths, from 420 to 545 nm in theUV and 530 to 755 nm in the VIS, and are modeled ata wavelength spacing of 0.1 nm. O3 and NO2 are fitusing cross sections from the GOME flight model[18,19], collected at 202 and 221 K, respectively. BothO3 and NO2 cross sections have some temperaturedependency. The use of the largely temperature-independent Chappuis band for O3 retrievals meansthat the uncertainty associated with the use of crosssections collected at different temperatures is mini-mal 1%. The temperature dependency of NO2 islarger, and the uncertainty associated with using across section at a fixed temperature is estimated to bebetween 5% and 10%. All trace gases and spectral

    parameters are fit using a direct spectral fit, with theexception of NO2, which is fit using a differentialoptical absorption spectroscopy method. The NO2cross section is divided into high- and low-frequencycomponents using a smoothing filter, and the high-frequency component of the cross section is fit inde-pendently to determine the retrieved NO2 SCD. Anamount of absorption from the smoothed cross sectionis also included, based on the same SCD of NO2.

    Although not currently operational data products,H2O, O2, and O2O2 are also forward modeled and fitin the retrieval to account for their spectral contribu-tions. O2O2 column density is determined usingcross sections from Greenblatt et al. [20]. Currentlythe spectra are corrected for H2OandO2B and bandabsorption by retrieving approximate SCDs using amoderate-resolution effective cross section deter-mined from a line-by-line radiative transfer calcula-tion through one representative 20 km tangent pathusing line parameters from HITRAN 2004 [14].

    The SCDs of O3 and NO2 retrieved for occultationss4043 are shown in Fig. 9 as a function of timeelapsed since the beginning of the occultation obser-

    vation sequence and as a function of correspondingtangent altitude below 100 km. These figures showthe SCDs derived for NO2 from the UV channel, andozone from the UV and VIS channels. For compari-

    son, the FTS SCDs are also plotted for ozone andNO2. (The FTS data processing does not retrieveSCDs but rather inverts spectra directly to profiles.Therefore, the FTS SCDs plotted in Fig. 9 are not v2.2products from the FTS but are from integration alongthe MAESTRO line of sight using the FTSs ozoneand NO2 profiles.)

    Due to the nature of ozone absorption in the UVand VIS spectrometer wavelength regions, the strato-spheric VIS retrieval of ozone is generally consider-ably more accurate than that only performed fromthe UV spectrometer, using spectra between 420 and

    545 nm. The retrieval of ozone in these spectral re-gions uses only the short-wavelength edge of theChappuis band, where there are other strong broad-band spectral features from aerosols and Rayleighscattering whose retrieval may interfere with that ofozone. In the troposphere and lower stratosphere,where the peak of Chappuis ozone removes a consid-erable amount of light in the short-wavelength end ofthe VIS detector and Rayleigh scattering is intense,

    the VIS integration times must still remain relativelyshort to ensure that the detector does not saturate atits longer wavelengths. In this case, ozone retrievedusing the UV spectrometer is comparable in reliabil-ity with the retrieval from the VIS spectrometer (see,for example, the relatively large noise in optical depthspectra collected at 6.4 and 10.7 km in Fig. 8 in thepeak of the Chappuis band).

    Vertical profiles of trace gases are derived fromMAESTRO using a nonlinear Chahine relaxation in-

    version [21]. Figure 10 shows sample ozone and NO2profiles for three occultations, including ss4043,

    Fig. 9. Slant column densities for occultation ss4043 of (a) ozonederived from the UV and VIS spectrometers and FTS profiles and(b) NO2 derived from the UV spectrometer and FTS profiles. Theslant column densities are plotted as a function of time elapsed

    since the beginning of the occultation sequence, as well as theircorresponding tangent heights. See the text for a discussion ofuncertainties.

    4352 APPLIED OPTICS Vol. 46, No. 20 10 July 2007

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    13/16

    which was derived from the SCDs pictured in Fig. 9.Figure 10 also shows the FTS v2.2 ozone and NO2

    profiles derived at the University of Waterloo ScienceOperations Centre for the same occultations, aswell as ozone profiles from ozonesonde launchesthat coincided with ACE observations. The latitudelongitude location of each occultation is referenced toits 30 km subtangent point. The data points in theprofiles denote the retrieved volume mixing ratiosretrieved at the tangent heights from either theMAESTRO or FTS occultation. The solid curves rep-resent the profiles interpolated onto equally spacedgrids. The spacing of these altitude grids is 0.5 km forMAESTRO and 1 km for the FTS. Both the tangent

    and interpolated profiles are available in the MAE-STRO v1.2 and FTS v2.2 data releases.

    A full validation of the preliminary v1.2 MAESTROozone and NO2 products has been performed by Kar

    et al. [16], including comparisons with the FTS andcoincident profiles from ozonesondes, SAGE III, andPOAM III. As a result, a detailed assessment of re-trieval performance with respect to the FTS andother instruments is not presented here. We can,however, give typical estimates of retrieval uncer-tainties for individual profiles. Table 2 lists precisionestimates from retrieval noise at different altitudesfor a typical Arctic spring retrieval of VIS ozone andUV NO2 profiles and for the slant columns tangent at

    Fig. 10. Volume mixing ratio profiles of ozone (in parts per 106 by volume) and NO2 (in parts per 109 by volume) retrieved from

    MAESTRO and the FTS for sunset occultations ss2856 (Arctic winter) and ss4043 (Antarctic fallwinter), and sunrise occultation sr4333(midlatitude summer), as well as ozone mixing ratios from coincident ozonesonde flights. The site of the ozonesonde launch is noted in eachlegend, as well as itsdistance d from the 30 km subtangent point of the ACEoccultation and the time in hours elapsed since the occultation,t. The data points signify the retrieved amount at each tangent point, while the solid curves represent the profiles interpolated ontogrids equally spaced in altitude (every 1 km for the FTS and 0.5 km for MAESTRO). See the text for a discussion of uncertainties.

    10 July 2007 Vol. 46, No. 20 APPLIED OPTICS 4353

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    14/16

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    15/16

    MAESTRO data releases will implement global fit-ting retrievals, as well as provide results for H2O- andO2-derived pressure and temperature [23], and aero-sol information separated from the total wavelengthdependent optical depth. Eventually, the ACE teamhopes to provide a single ozone product for the mis-sion derived by combining information from the FTSand MAESTRO.

    This first flight of a UV-VIS-NIR instrument(MAESTRO) and a mid-IR instrument (ACE-FTS)measuring along the same line-of-sight promises toimprove our understanding of ozone and NO2 retriev-als in the UV-VIS-NIR and mid-IR, and provide syn-ergistic opportunities for science. In addition, withthe recent decommissioning of SAGE II, SAGE III,and POAM III, MAESTRO remains the only primar-ily solar occultation UV-VIS-NIR instrument in orbit,and is helping to ensure the continuity of importantUV-VIS-NIR solar occultation measurements.

    The Atmospheric Chemistry Experiment (ACE)is a Canadian-led mission mainly supported by theCanadian Space Agency. The MAESTRO instrument

    was developed with additional financial support fromEnvironment Canada, the Canadian Foundation forClimate and Atmospheric Sciences, and the NaturalSciences and Engineering Research Council of Can-ada. Launch services were provided by NASA. Theinstrument was developed and built by EnvironmentCanada and EMS Technologies. We thank the mem-bers of the MAESTRO test team, particularly RonIrvine, Bob Hum, Paul Chen, and Oleg Mikhailov, fortheir work during instrument development and pre-flight characterization.

    References

    1. P. F. Bernath, C. T. McElroy, M. C. Abrams, C. D. Boone, M.Butler, C. Camy-Peyret, M. Carleer, C. Clerbaux, P.-F. Coheur,R. Colin, P. DeCola, M. De Mazire, J. R. Drummond,D. Dufour,W. F. J. Evans, H. Fast, D. Fussen, K. Gilbert, D. E. Jennings,E. J. Llewellyn, R. P. Lowe, E. Mahieu, J. C. McConnell, M.McHugh, S. D. McLeod, R. Michaud,C. Midwinter, R. Nassar, F.Nichitiu, C. Nowlan, C. P. Rinsland, Y. J. Rochon, N. Rowlands,K. Semeniuk, P. Simon, R. Skelton, J. J. Sloan, M.-A. Soucy, K.Strong, P. Tremblay, D. Turnbull, K. A. Walker, I. Walkty, D. A.Wardle, V. Wehrle, R. Zander, and J. Zou, Atmospheric Chem-istry Experiment (ACE): mission overview, Geophys. Res. Lett.32, L15S01, doi:10.1029/2005GL022386 (2005).

    2. K. L. Gilbert, D. N. Tumbull, K. A. Walker, C. D. Boone, S. D.McLeod, M. Butler, R. Skelton, P. F. Bernath, F. Chateauneuf,and M.-A.Soucy, The on-board imagersfor the Canadian ACE

    SCISAT-1 mission (submitted to J. Geophys. Res.).3. World Meteorological Organization, Scientific Assessment of

    Ozone Depletion: 2002, Global Ozone and Research MonitoringProject, Report 47 (World Meteorological Organization, 2003).

    4. L. W. Thomason and G. Taha, SAGE III aerosol extinctionmeasurements: initial results, Geophys. Res. Lett. 30, 1631,doi:10.1029/2003GL017317 (2003).

    5. R. L. Lucke, D. R. Korwan, R. M. Bevilacqua, J. S. Hornstein,E. P. Shettle, D. T. Chen, M. Daehler, J. D. Lumpe, M. D.Fromm, D. Debrestian, B. Neff, M. Squire, G. Knig-Langlo,and J. Davies, The Polar Ozone and Aerosol Measurement(POAM) III instrument and early validation results, J. Geo-phys. Res. 104, 1878518799 (1999).

    6. J. M. Russell III, L. L. Gordley, J. H. Park, S. R. Drayson,W. D. Hesketh, R. J. Cicerone, A. F. Tuck, J. E. Frederick, J. E.Harries, and P. J. Crutzen, The Halogen Occultation Exper-iment, J. Geophys. Res. 98, 1077710797 (1993).

    7. C. D. Boone, R. Nassar, K. A. Walker, Y. Rochon, S. D. McLeod,C. P. Rinsland, and P. F. Bernath, Retrievals for the atmo-spheric chemistry experiment Fourier-transform spectrome-ter, Appl. Opt. 44, 72187231 (2005).

    8. C. T. McElroy, A spectroradiometer for the measurement ofdirect and scattered solar spectral irradiance from on-board

    the NASA ER-2 high-altitude research aircraft, Geophys. Res.Lett. 22, 13611364 (1995).9. C. T. McElroy, C. Midwinter, D. V. Barton, and R. B. Hall, A

    comparison of J-values from the composition and photodisso-ciative flux measurement with model calculations, Geophys.Res. Lett. 22, 13651368 (1995).

    10. J. P. Burrows, M. Weber, M. Buchwitz, V. Rozanov, A.Ladsttter-Weissenmayer, A. Richter, R. DeBeek, R. Hoogen,K. Bramstedt, K.-U. Eichmann, M. Eisinger, and D. Perner,The Global Ozone Monitoring Experiment (GOME): missionconcept and first scientific results, J. Atmos. Sci. 56, 151175(1999).

    11. R. W. Wood, Anomalous diffraction gratings, Phys. Rev. 48,928936 (1935).

    12. D. G. Dufour, J. R. Drummond, C. T. McElroy, C. Midwinter,

    P. F. Bernath, K. A. Walker, W. F. J. Evans, E. Puckrin, andC. Nowlan, Intercomparison of simultaneously obtained in-frared (4.8 m) and visible (515715 nm) ozone spectra using

    ACE-FTS and MAESTRO, J. Phys. Chem. A 109, 87608764(2005).

    13. D. G. Dufour, J. R. Drummond, C. T. McElroy, C. Midwinter,P. F. Bernath, K. A. Walker, and C. Nowlan, Simultaneousmeasurements of visible (400700 nm) and infrared (3.4 m)NO2 absorption, J. Phys. Chem. A 110, 1241412418 (2006).

    14. L. S. Rothman, D. Jacquemart, A. Barbe, D. C. Benner, M.Birk, L. R. Brown, M. R. Carleer, C. Chackerian Jr., K. Chance,L. H. Coudert, V. Dana, V. M. Devi, J.-M. Flaud, R. R. Gama-che, A. Goldman, J.-M. Hartmann, K. W. Jucks, A. G. Maki,J.-Y. Mandin, S. T. Massie, J. Orphal, A. Perrin, C. P. Rins-land, M. A. H. Smith, J. Tennyson, R. N. Tolchenov, R. A. Toth,J. Vander Auwera,P. Varanasi, and G. Wagner, The HITRAN2004 molecular spectroscopic database, J. Quant. Spectrosc.Radiat. Transfer 96, 139204 (2005).

    15. G. L. Thuillier, L. Floyd, T. N. Woods, R. Cebula, E.Hilsenrath, M. Hers, and D. Labs, Solar irradiance referencespectra, in Solar Variability and its Effects on Climate, Geo-

    physical Monograph 141, J. M. Pap and P. Fox, eds. (AmericanGeophysical Union, 2004), pp. 171194.

    16. J. Kar, C. T. McElroy, J. R. Drummond, J. Zou, F. Nichitiu,K. A. Walker, C. E. Randall, C. R. Nowlan, D. G. Dufour, C. D.Boone, P. F. Bernath, C. R. Trepte, L. W. Thomason, and C.McLinden, Initial comparison of ozone and NO2 profiles from

    ACE-MAESTRO with balloon and satellite data (submitted toJ. Geophys. Res.).

    17. M. Carlotti, Global-fit approach to the analysis of limb-scanning atmospheric measurements, Appl. Opt. 27, 32503254 (1988).

    18. J. P. Burrows, A. Richter, A. Dehn, B. Deters, S. Himmelmann,S. Voigt, and J. Orphal, Atmospheric remote-sensing refer-ence data from GOME2. Temperature-dependent absorptioncross sections of O3 in the 231794 nm range, J. Quant. Spec-trosc. Radiat. Transfer 61, 509517 (1999).

    19. J. P. Burrows, A. Dehn, B. Deters, S. Himmelmann, A. Richter,S. Voigt, and J. Orphal, Atmospheric remote-sensing refer-ence data from GOME: Part 1. Temperature-dependent ab-sorption cross-sections of NO2 in the 231794 nm range, J.Quant. Spectrosc. Radiat. Transfer 60, 10251031 (1998).

    10 July 2007 Vol. 46, No. 20 APPLIED OPTICS 4355

  • 8/2/2019 C. Thomas McElroy et al- The ACE-MAESTRO instrument on SCISAT: description, performance, and preliminary results

    16/16