boschetti&iacumin2005 rcms

8
RAPID COMMUNICATIONS IN MASS SPECTROMETRY Rapid Commun. Mass Spectrom. 2005; 19: 3007–3014 Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/rcm.2161 Continuous-flow d 18 O measurements: new approach to standardization, high-temperature thermodynamic and sulfate analysis Tiziano Boschetti* and Paola Iacumin* Dipartimento di Scienze della Terra, Universita ` degli Studi di Parma, Parco Area delle Scienze 157/A, 43100 Parma, Italy Received 26 July 2005; Revised 26 August 2005; Accepted 26 August 2005 The continuous-flow method for the determination of the O-isotope composition of solid samples has significant advantages over off-line extraction methods, but the problem has arisen of the stan- dardization of results that has only partially been resolved by the use of water standards. We pro- pose a new approach to standardization that uses carbothermic reduction of calcium carbonate standards catalyzed by high-purity AgCl. Analytical accuracy, precision, variance homogeneity and long-term stability are proven. Preliminary data on the barium sulfate d 18 O analyses are reported, with a closer look at the different results obtained by off-line and on-line methods on intercomparison standards NBS-127 and MSS3. Copyright # 2005 John Wiley & Sons, Ltd. Determination of d 18 O composition in organic and inorganic compounds by thermal conversion—continuous flow isoto- pic ratio mass spectrometry (TCCFIRMS) is hindered by the shortage of internationally distributed solid calibration stan- dards. New standards are continually being prepared by institutions such as IAEA and NIST, but the requirements for more and more specificity and the sectionalism of scienti- fic research have far exceeded the present capabilities to produce appropriate standards, both in the case of the bulk stable isotope analysis and in case of the compound-specific isotope analysis. Many laboratories use water standards for the 18 O/ 16 O measurement normalization 1 because water is an interna- tionally distributed material and many samples are subject to intercomparison exercises using an off-line validated method. 2 Use of water as standard for solid analysis in TCCFIRMS is not free from problems due to: (a) evaporation and isotopic fractionation of the small volume of water standard (0.1 mL); (b) calibration line too wide or with non- equispaced standards (non-homogeneous variance); and (c) different reactivity of standards and investigated samples. Water samples can be measured with helium dilution, but the helium dilution offset cannot be compared with the absolute solid sample isotopic results since they are not measured in the same manner. 3 Calcium carbonates have a validated off- line method of analysis 4 and are distributed internationally as reference materials, but their use in the TCCFIRMS standar- dization has been avoided because of the incomplete yield of reaction. 3 The usefulness of halide salt addition in the thermal and carbothermal chemical analysis of many O-containing compounds is also widely known. 5–7 Most recently, non- halogenated and halogenated catalysis were tested for the continuous-flow isotope analysis of oxygen from carbonates and silicates. 8,9 In this study, we corroborate the positive results of the catalytic carbothermic reduction of calcium carbonate using AgCl and demonstrate its applicability to a new standardization approach and to the determination of the 18 O/ 16 O isotope ratio analysis of barium sulfate. EXPERIMENTAL The apparatus used in this study (Fig. 1) consists of a thermal conversion—elemental analyzer unit (TCEA, Thermo Finnigan) composed of an outer reaction tube made of Al 2 O 3 ceramic (17 mm o.d., 14 mm i.d., length 470 mm) and an inner glassy carbon tube (12 mm o.d., 7 mm i.d., length 355 mm) assembled following the instructions reported in the ThermoFinnigan TCEA manual. 10 A graphite crucible is inserted in the hottest zone of the reaction furnace, and heated to 1400–14208C. At these temperatures, carbon from the inner tube and glassy carbon grit, which partially fill the reactor, primes the reduction of the analyzed compounds fol- lowing the reaction O-compound ðsÞ þ C ðsÞ ¼ CO ðgÞ þ½residualsĸ ðs;g;lÞ ð1Þ where subscripts s, g and l represent solid, gaseous and liquid by-products, respectively. The space between the tubes is continuously flushed with high purity He (5.5 Sapio, Monza, Milan, Italy) to prevent undesirable oxida- tion and to carry the CO gas to a gas chromatography (GC) column downstream of the reactor. We used two con- figurations: high gas configuration, where the GC column was held at 908C and the system flushed with He carrier gas pressure at 1.0 bar; and low gas configuration, where the GC column was held at 708C and the system flushed with He gas at 0.85 bar. The combination of low tempera- ture and gas pressure guarantees a good GC separation between the 12 C 16 O and 14 N 2 peaks when N-compounds Copyright # 2005 John Wiley & Sons, Ltd. *Correspondence to: T. Boschetti or P. Iacumin, Dipartimento di Scienze della Terra, Universita ` degli Studi di Parma, Parco Area delle Scienze 157/A, 43100 Parma, Italy. E-mail: [email protected]; [email protected]

Upload: freetime8334

Post on 18-Dec-2015

218 views

Category:

Documents


0 download

DESCRIPTION

Continuous-flow d18O measurements: new approach to standardization, high-temperature thermodynamic andsulfate analysis

TRANSCRIPT

  • RAPID COMMUNICATIONS IN MASS SPECTROMETRY

    Rapid Commun. Mass Spectrom. 2005; 19: 30073014

    Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/rcm.2161

    Continuous-flow d18O measurements: new approach to

    standardization, high-temperature thermodynamic and

    sulfate analysis

    Tiziano Boschetti* and Paola Iacumin*Dipartimento di Scienze della Terra, Universita` degli Studi di Parma, Parco Area delle Scienze 157/A, 43100 Parma, Italy

    Received 26 July 2005; Revised 26 August 2005; Accepted 26 August 2005

    The continuous-flow method for the determination of the O-isotope composition of solid samples

    has significant advantages over off-line extraction methods, but the problem has arisen of the stan-

    dardization of results that has only partially been resolved by the use of water standards. We pro-

    pose a new approach to standardization that uses carbothermic reduction of calcium carbonate

    standards catalyzed by high-purity AgCl. Analytical accuracy, precision, variance homogeneity

    and long-term stability are proven. Preliminary data on the barium sulfate d18O analyses are

    reported, with a closer look at the different results obtained by off-line and on-line methods on

    intercomparison standards NBS-127 and MSS3. Copyright # 2005 John Wiley & Sons, Ltd.

    Determination of d18O composition in organic and inorganiccompounds by thermal conversioncontinuous flow isoto-

    pic ratio mass spectrometry (TCCFIRMS) is hindered by the

    shortage of internationally distributed solid calibration stan-

    dards. New standards are continually being prepared by

    institutions such as IAEA and NIST, but the requirements

    for more and more specificity and the sectionalism of scienti-

    fic research have far exceeded the present capabilities to

    produce appropriate standards, both in the case of the bulk

    stable isotope analysis and in case of the compound-specific

    isotope analysis.

    Many laboratories use water standards for the 18O/16O

    measurement normalization1 because water is an interna-

    tionally distributed material and many samples are subject to

    intercomparison exercises using an off-line validated

    method.2 Use of water as standard for solid analysis in

    TCCFIRMS is not free from problems due to: (a) evaporation

    and isotopic fractionation of the small volume of water

    standard (0.1mL); (b) calibration line too wide or with non-equispaced standards (non-homogeneous variance); and (c)

    different reactivity of standards and investigated samples.

    Water samples can be measured with helium dilution, but the

    helium dilution offset cannot be compared with the absolute

    solid sample isotopic results since they are not measured in

    the same manner.3 Calcium carbonates have a validated off-

    linemethod of analysis4 and are distributed internationally as

    reference materials, but their use in the TCCFIRMS standar-

    dization has been avoided because of the incomplete yield of

    reaction.3 The usefulness of halide salt addition in the thermal

    and carbothermal chemical analysis of many O-containing

    compounds is also widely known.57 Most recently, non-

    halogenated and halogenated catalysis were tested for the

    continuous-flow isotope analysis of oxygen from carbonates

    and silicates.8,9 In this study, we corroborate the positive

    results of the catalytic carbothermic reduction of calcium

    carbonate using AgCl and demonstrate its applicability to a

    new standardization approach and to the determination of

    the 18O/16O isotope ratio analysis of barium sulfate.

    EXPERIMENTAL

    The apparatus used in this study (Fig. 1) consists of a thermal

    conversionelemental analyzer unit (TCEA, Thermo

    Finnigan) composed of an outer reaction tube made of

    Al2O3 ceramic (17 mm o.d., 14 mm i.d., length 470 mm) and

    an inner glassy carbon tube (12 mm o.d., 7 mm i.d., length

    355 mm) assembled following the instructions reported in

    the ThermoFinnigan TCEA manual.10 A graphite crucible is

    inserted in the hottest zone of the reaction furnace, and

    heated to 140014208C. At these temperatures, carbon fromthe inner tube and glassy carbon grit, which partially fill the

    reactor, primes the reduction of the analyzed compounds fol-

    lowing the reaction

    O-compounds Cs COg residualss;g;l 1

    where subscripts s, g and l represent solid, gaseous and

    liquid by-products, respectively. The space between the

    tubes is continuously flushed with high purity He (5.5

    Sapio, Monza, Milan, Italy) to prevent undesirable oxida-

    tion and to carry the CO gas to a gas chromatography

    (GC) column downstream of the reactor. We used two con-

    figurations: high gas configuration, where the GC column

    was held at 908C and the system flushed with He carriergas pressure at 1.0 bar; and low gas configuration, where

    the GC column was held at 708C and the system flushedwith He gas at 0.85 bar. The combination of low tempera-

    ture and gas pressure guarantees a good GC separation

    between the 12C16O and 14N2 peaks when N-compounds

    Copyright # 2005 John Wiley & Sons, Ltd.

    *Correspondence to: T. Boschetti or P. Iacumin, Dipartimento diScienze della Terra, Universita` degli Studi di Parma, ParcoArea delle Scienze 157/A, 43100 Parma, Italy.E-mail: [email protected]; [email protected]

  • are analyzed. A trap, filled with 2/3 AscariteII1 20-30 mesh

    (Thomas Scientific, Swedesboro, NJ, USA) plus 1/3 anhy-

    drous Mg(ClO4)2 (Anhydrone1, CEInstrumentsThermo

    Quest, Milan, Italy), is interposed between the reactor and

    the GC column, thus cutting off acid gases (e.g. chlorine)

    and water. The TCEA unit is coupled to a Delta Plus XP

    spectrometer (ThermoFinnigan) through two open-splits

    ConFloIII interface11 (Fig. 1). The mass spectrometer is oper-

    ated in linear mode (extraction lens at 2.7 kV). To calculatethe yield of oxygen, variable amounts of benzoic acid were

    used (Hekatech, Wedberg, Germany). Solid samples and

    standards, weighed into silver capsules (3.2 4 mm,0.02 mL; Santis Analytical AG, Teufen, Switzerland), are cali-

    brated against pure CO reference gas peaks (Messer 4.7;

    1.5 bar at ConFloIII panel) in order to have comparable

    area and signal (Table 1). Comparison with calibration,

    intercomparison or laboratory standard materials16 is necessary

    prior to every new batch of samples as signal variations are

    possible due to d18O variation of the reference CO gas fromdifferent tanks (from 0.0 to ca. 5.5% V-SMOW), as is refo-cusing of the mass spectrometer after switching between

    other units (e.g. CHN) and replacement of the ion source.

    Table 1. Weighted amounts of some compounds analyzed by TCCFIMRS, proportionally calculated in comparison with the

    relative area and signal (V) of standard CO reference gas peaks

    Compound Melting pointa (8C) Min-max weightb (mg) Total Oc (%)

    C7H6O2 benzoic acid 122 0.0740.154 26.202C12H22O11 sucrose 186 0.0490.080 51.415C6H10O5 cellulose 180290

    d 0.0240.080 49.338KNO3 K nitrate 334 0.0470.081 47.474Ag3PO4 Ag phosphate 849 0.2090.329 15.289BaSO4 Ba sulfate 1580 0.1060.152 27.421CaCO3 Ca carbonate 8991339

    e 0.0970.113 47.956

    a Bale et al.;12 Bayley et al.;13 Kirk-Othmer.14b Resulting signals were compared with the reference peak of CO& 3.3V& 60000 area (1.5 bar at the ConFloIII unit).c Element atomic weight from Loss.15d Decomposition at 1808C, combustion at 2908C.e Decomposition at 8998C; melts at high pressure: 13398C at 100 bar.

    Figure 1. Schematic diagram of the thermal combustion/elemental analyzer (TC/EA) continuous-flow (CF)

    isotopic ratio mass spectrometer (IRMS) (modified from Gehre and Strauch9). Original Thermo Finnigan

    configuration was modified with an as/an trap*: 2/3 AscariteII1 1/3 Anhydrone1 trap.

    Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 30073014

    3008 T. Boschetti and P. Iacumin

  • Aware of the lack of certified standards with an appro-

    priate low 18O/16O ratio for CO reference gas standardiza-

    tion, in this work we have always considered d18O(CO) 0%(V-SMOW) in every CO tank, so differences between known

    and measured values of the standards after least-squares

    standard calibration are fundamentally due to d18O varia-tions of CO from different tanks. In the case of substances

    with high melting points, principally calcium carbonate and

    barium sulfate, we have tested the effectiveness of some

    analytical grade additives such as silver chloride (99.9%

    AgCl; Strem Chemicals, Newburyport, MA, USA) and

    hexamethylenetetramine (99% C6H12N4, A.C.S. reagent;Aldrich, St. Louis, MO, USA). These materials were preferred

    to other catalysts due to the high yield of the previous

    experiments,8 the relatively low causticity in comparison

    with the fluorinated residual gas of fluoride salts (e.g. KF),

    and the limited presence of other elements that could

    invalidate the analysis by the formation of undesirable

    residuals like oxides and carbides. Standards, samples and

    additives were dried in a vacuum oven for 12 h at 1208C;afterwards, materials and silver capsules were stored in

    vacuum driers before use.

    Any CO2 by-product is avoided by the high temperature of

    the reactor (>9808C) that displaces the Boudouard reaction:

    CO2g Cs , 2COg 2completely toward CO.13

    The theoretical evolutions of reactions reported through-

    out the text were predicted with FACTSAGE-Web thermo-

    dynamic data specific for high temperatures,12 assuming

    solid activities and gases partial pressures equal to unity.

    These assumptions are reasonable considering the high

    purity of the samples and standards and the solids weight

    calibration in order to bring a PCO of &1.01.5 bar. Themelting of silver capsules (960.88C) was not considered inthe modelling because the phase change Ag(s) to Ag(l) does

    not involve a Gibbs free energy variation. In addition, the

    formation of silver-containing residuals (e.g. silver carbide)

    after carbothermic reduction is highly improbable.

    RESULTS AND DISCUSSION

    Calcium carbonate analysisMany authors have advised against the use of calcium carbo-

    nate in TCCFIRMS standardization.1,3 In fact, at the operative

    temperature of the reactor (from 133014508C), only two ofthree oxygen atoms react with C to form CO during the car-

    bothermic reduction of the calcium carbonate:

    CaCO3s Cs , CaOs 2COgGr1400C 198:3 kJ; Gr1600C 262:76 kJ

    3Preliminary results on several calcium carbonate samples

    from this study at standard configuration (furnace at 14008C,GC column at 908C, He carrier flow pressure at 1 bar) confirmthe yield of oxygen as CO at about 67%. Problems arising

    from the incomplete yield of reaction are: (a) isotopic

    fractionations17,18 and (b) memory isotopic effects related to

    the presence of solid residual CaO(s) which is quite reactive

    and can exchange with CO(g).19,20

    Kornexl et al.8 observed that temperatures higher than

    16008C are necessary for the quantitative release of oxygenfrom calcium carbonate. This temperature is unattainable for

    a TCEA unit using an Al2O3 ceramic tube (classified as a

    synthetic sintered mullite); other than, according to a recent

    study,21 the higher the temperature, the higher the back-

    ground on mass 28 (12C16O) due to diffusion of (light) oxygen

    from the ceramic tube. In fact, in the range 133014508C, wehave found a significant linear correlation between tempera-

    ture and the mass 28 signal (N 6; r 0.9944):mass 28 3:7549 0:1998T 4872 277

    where T is the temperature in 8C and mass 28 is the signal inmV for the ion at m/z 28.

    The use of high temperatures for long periods reduces the

    life of reactor tubes, the furnace and the thermocouple. In

    addition, thermodynamic calculations reveal that at the temp-

    eratures of 1400 and 16008C the Gibbs free energy of reaction(3) is lower than those of reactions (4) and (5), giving support

    for the formation of calcium carbide or elemental calcium

    (*note that calcium is liquid at 8428C and gaseous at 15038C):

    CaCO3s 4Cs , CaC2s 3COgGr1400C 103:75 kJ; Gr1600C 212:75 kJ

    4CaCO3s 2Cs , 3COg CaGr1400C 4:35 kJ; Gr1600C 108:73 kJ

    5The non-total yield problems could be solved by using

    catalysts.8,9 In this study, CO peaks were detected during the

    blanks analysis of hexamethylenetetramine catalyst, prob-

    ably due to the high hygroscopicity of amines and the

    consequent hydration during the weighing routine and/or

    the permanence of the samples in the autosampler before

    analysis. It is also possible that the high nitrogen content of

    this compound could pollute the oxygen isotope analysis due

    to the mass interference generated by accumulation of N2 and

    NO gas in the ion source of the mass spectrometer.

    While on-line carbon reduction plus fluorination does not

    give satisfactory results,22 catalytic carbothermic reduction

    using AgCl as an additive is the recommended method for

    the on-line analysis of carbonates.9 We obtained CO free

    blanks using approximately 500mg of AgCl plus 300 mg ofglassy carbon powder (obtained from grinding of glassy

    carbon grit used for reactor assembling) placed into the silver

    capsule. This result is due to the low hygroscopicity and the

    high purity of AgCl and of the mixture. Using these amounts,

    the procedure was applied to calcium carbonate laboratory

    standards, previously analyzed isotopically by the H3PO4decomposition method,4 and the results were used for

    TCCFIRMS calibration at 14208C and in the low gas config-uration (Table 2). Remarkably, mean yields of 100% were

    obtained. Most probably, the reactions involved are:

    CaCO3s AgCls 4Cs , 3COg Agl CaC2s ClgGr1420C 24:57 kJ 6

    CaCO3s 2AgCls 2Cs , 3COg CaCl2l AglGr1420C 438:13 kJ 7

    High-temperature continuous-flow 18O measurements of sulfates 3009

    Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 30073014

  • Table 2. Analytical d18O (CO3) results of calcium carbonate laboratory standards. Two standards are commercial reagentsgrade (TCS2, CE-CO3) and two natural carbonates (TCS1, Carrara). Known and Measured columns report the IRMS values

    measured on the CO2 (off-line acid decomposition method4) and on the CO (on-line AgCl catalyzed carbotermic reduction),

    respectively. Equation of least-squares linear regression calculated on these data is reported

    Code DescriptionWeight

    (mg)Signal

    (V at mass 28)CO

    (rel. area) O-yield*Known**

    (% vs. V-SMOW)Measured(% vs. CO) Mean Std.dev.

    TCS1 CaCO3 Sardinian 0.122 4.3 157280 105 23.0 28.19 28.3 0.2stalagmite 0.155 5.2 186750 102 28.25

    0.156 5.1 185143 102 28.54

    TCS2 CaCO3 Merck ACS-ISO-F 0.144 5.6 202793 108 10.9 17.54 17.8 0.299% reagent grade 0.128 4.7 172545 103 17.87

    0.148 5.5 200902 104 17.96

    CE-CO3 CaCO3 Carlo Erba RPE 0.160 5.7 209184 102 3.3 10.16 10.6 0.499% reagent grade 0.170 6.7 245484 101 10.79

    0.160 6.3 231538 102 10.910.138 3.9 143327 103 10.35

    Carrara CaCO3 Carrara marble 0.121 4.6 167782 106 29.0 34.68 34.4 0.30.148 5.0 180347 103 34.120.141 4.3 152711 103 34.40

    Least-squares regression parameters (ymeasured; xknown)Intercept: 7.5742 0.1768Slope: 0.9177 0.0095Pearsons correlation coefficient (r) 0.9994Standard error: 0.35

    * In comparison with benzoic acid; yield excess due to weighting error.** d18O (vs. VSMOW) [1.03092 * d18O (vs. VPDB)] 30.92.49

    Table 3. Analytical results and statistical tests relative to least-squares linear regression of international (IAEA-CH-6, NBS-18,

    NBS-19) and laboratory standards (CE-CO3, TCS1, TCS2). Linearity (Fishers test calculated value is greater than the tabulated

    values) and variance homogeneity (Hartley and Cochrans test calculated values are smaller than the tabulated values) are

    confirmed. Least-squares regression line described in this table is parallel to least-squares regression line of Table 2, as

    confirmed statistically by Students t-test on the slopes

    CE-CO3 IAEA-CH-6 TCS1 TCS2 NBS-19 NBS-18

    Known values (% vs.V-SMOW) 3.3 36.4 23.0 10.9 28.6 7.2

    7.62 38.56 25.77 14.61 31.44 10.76Measured (% vs. CO) 7.35 37.74 25.71 14.86 30.41 11.11

    7.14 38.12 26.01 14.49 30.40 11.46 S(sum)

    Std.dev. 0.24 0.41 0.16 0.19 0.60 0.35 1.95Variance 0.057 0.168 0.025 0.036 0.360 0.124 0.77Mean 7.37 38.14 25.83 14.65 30.75 11.11 127.85Sy 22.11 114.41 77.50 43.95 92.26 33.33 383.56Sy2 163.07 4363.86 2002.13 644.03 2837.71 370.54 10381.34Least-squares regression parameters (ymeasured; xknown)Intercept: 4.4354 0.1417Slope: 0.9254 0.0065Pearsons correlation coefficient (r) 0.9996Standard error: 0.33

    ANOVA Deviance Degree of freedom (DF) Variance

    Total 2208.020 17Between (B) 2206.480 5 441.296Within (W) 1.53997 12 0.12833

    Variancetests

    DF(num/den)

    Calculatedvalues

    Tabulated values(significance level)(0.05) (0.01)

    Fisher test (B/W) F(5/12) 3438.745 3.11 5.06Hartley test F(6/2) 14.3003 19.5 99.3Cochran test 0.46691 0.61 0.72

    3010 T. Boschetti and P. Iacumin

    Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 30073014

  • The accuracy, precision, linearity and variance homo-

    geneity were checked by the analysis of three laboratory

    standards and three IAEA internationally distributed

    materials:16 NBS-19 calcium carbonate calibration standard

    and two intercomparison standards, NBS-18 calcium carbo-

    nate and IAEA-CH-6 sucrose (Table 3). For the sucrose

    standard, internationally agreed d18O measures are not yetavailable, but a value of 36.4% (V-SMOW) is commonlyreported.1,23 The high standard deviation on NBS-19 is

    probably due to the large grain size of this reference material

    (a poor feature of the material when related to the small

    amount weighted for the TCCFIRMS analysis). The linearity

    and variance homogeneity of regression were evaluated and

    confirmed statistically (Fishers, Hartleys and Cochrans

    tests; Table 3). In comparison with the least-squares

    regression line data of Table 2, the difference between the

    intercepts (i.e. measured d18O value) is mainly due to thevariation of the d18O composition of the CO reference gas;the parallelism between the two regression lines was

    confirmed statistically by a Students t-test at high prob-

    ability values (P> 0.1). Long-term stability is less than the

    calibration error (0.3%) and was evaluated by continuousanalysis throughout 12 h of keratinic samples having d18Ovalues lower than the lowest one of the calibration standards

    (3.3%). The deviations from the two calibrations performedat the start and the end of the data acquisition are shown

    in Fig. 2.

    Barium sulfate analysisAs with calcium carbonate, the use of barium sulfate stan-

    dards in TCCFIRMS oxygen isotope analyses was also

    recently advised due to the memory effects related to oxygen

    scavenging by Ba2 ions resulting from carbothermic reduc-tion of BaSO4.

    24 In agreement with the literature,8 our preli-

    minary analysis on barium sulfate samples by TCEA in

    standard configuration, at 14008C and without additives,gives yields of 9497%. Tests on five barium sulfate samples

    at 14208C and at a variable height of the graphite crucible inthe reactor give a mean yield of 75%. From these yield data,

    we infer that two competitive reactions control the carbother-

    mic reduction of barium sulfate:

    BaSO4s 4Cs , BaSs 4COgGr1400C 614:60 kJ; Gr1420C 628:21 kJ

    8

    BaSO4s 3Cs Ags , BaOs 3COg AgSgGr1400C 215:70 kJ; Gr1420C 229:34 kJ

    9Based on these reactions, oxygen isotopic fractionation and

    memory effects of BaO residuals should be expected during

    analyses.In this study, we have investigated the use of AgClC

    additives and TCEA low gas configuration on BaSO4

    Figure 2. Long-term stability test on 54 keratinic samples (13 h of analysis). The calculated

    d18O values (0.3% vs. V-SMOW) of the samples plotted on the dashed line represent thebest fit of the two standard calibrations (d18Ofirst calibration/d

    18Osecond calibration 1).

    High-temperature continuous-flow 18O measurements of sulfates 3011

    Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 30073014

  • intercomparison standards and we have obtained an average

    yield of 100%, probably following the reaction:

    BaSO4s AgCls 4Cs , 4COg Agl BaSs ClgGr1420C 539:19 kJ 10

    Table 4 shows some results of the effectiveness of the ana-

    lyses applied to two barium sulfate standards and using cal-

    cium carbonate as calibration standards. Certified values

    were obtained by other authors using an off-line sulfate reduc-

    tion: MSS-3 (10.5% vs. V-SMOW, Okayama University stan-dard)25 and NBS-127 (9.3% vs. V-SMOW, seawater sulfatefrom Monterey Bay, California, distributed from IAEA and

    NIST with the codes NBS-127 and RM8557, respectively).16

    Our NBS-127 d18O value of 8.5% is comparable with thosefrom other TCCFIRMS analyses, but not with the results

    obtained by the off-line reduction technique. In Table 5, the

    published and communicated d18O(SO4) values of NBS-127from on-line and off-line techniques are summarized, the

    difference being statistically significant according to

    Students t-test.

    The true oxygen isotope composition of seawater sulfate is

    an unresolved problem. In the past decades, the proposed

    d18O values obtained by off-line methods were 8.6%,31,32

    8.7%,33 9.39.6%,3436 and 9.9%.37 Curiously, the lowestvalues agree with the present on-line NBS-127 determina-

    tions. Off-line carbothermic reduction of BaSO4 has been

    carried out by heating the BaSO4-graphite mixture with

    conventional furnaces,38 electrical resistance,25,39,40 and

    radio-frequency induction.34,4144 Reaction temperatures of

    the various methods were variable from 850 to 12008C. In thereduction, CO and CO2 form via the reaction:

    BaSO4s 3Cs !8501200C

    BaSs 2COg CO2g 11

    The gas products CO2CO are expelled from the reactionchamber to a vacuum line where CO2 is collected in a liquid

    nitrogen trap and the remaining CO is subjected to a high-

    voltage electrical discharge from 1 up to 6 kV for COCO2conversion (inverse of reaction (2)) within a quartz or Pyrex

    tube. Explanations of the differences related to d18O of theseawater sulfate were:

    1. isotope effects during conversion of CO into CO2;

    2. oxygen isotopic exchange between CO and quartz/Pyrex

    walls in the conversion tube (particularly if the tube is not

    cooled);

    3. isotopic alteration during long storage of the seawater

    samples;

    4. contamination of the BaSO4 precipitated by impurities like

    organic materials, carbonate and phosphate.

    Table 5. Summary of the published and communicated d18O(SO42) values of NBS-127 from on-line and off-line techniques

    (differences being statistically significant according to Students t-test)

    d18O(SO4) seawater sulfate standard (NBS-127, RM 8557)

    On-line method Off-line method

    Hall et al.27 9.2 0.2Kornexl et al.8 8.7 0.2 Gonfiantini et al.16 9.3 0.3Bohlke et al.26 8.6 0.2 Bottcher et al.28 9.5This work 8.5 0.2 S. Halas (pers. comm.)a 9.69 0.06

    Palmer et al.29 8.9 0.2

    Mean 8.6 Mean 9.3

    Median 8.6 Median 9.3Std.dev. 0.1 Std.dev. 0.3

    aAt the time of writing, this value was corrected to 9.91% (Peryt et al.30).

    Table 4. Some results of the effectiveness of the analyses applied to two barium sulfate standards and using calcium carbonate

    as calibration standards

    Weight (mg) O-yield (%)Signal

    (V at mass 28)Measured

    (% vs. CO)Calculated

    (% vs. V-SMOW)Off-line value

    (% vs. V-SMOW) dcalc doff-lineNBS-127 0.164 104 3.6 15.5 8.6 9.3 0.7

    0.122 90 3.3 15.3 8.5 0.80.175 107 4.0 15.4 8.5 0.80.159 102 3.7 15.2 8.3 1.0

    Mean 101 8.5 0.8Std.dev. 8 0.2

    MSS-3 0.175 110 4.0 16.4 9.7 10.5 0.80.162 102 3.4 16.9 10.1 0.40.128 101 2.7 17.1 10.4 0.10.126 91 2.2 16.8 10.0 0.5

    Mean 101 10.1 0.4Std.dev. 8 0.3

    3012 T. Boschetti and P. Iacumin

    Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 30073014

  • Only the possible role of process 2 was experimentally

    assessed.25,39 Process 1 is subject to a kinetic fractionation:17,25

    kineticCO-CO2 k1k2 1:02 12

    where k1 and k2 are, respectively, the rates constants of the

    C16O and C18O molecules in the forward reaction. Thus, if a

    yield less than 100% of CO2 is obtained from BaSO4 reduction

    (and if the yield of CO2 from CO is from 90 to 98 percent),25,34

    sequential electrical decomposition of CO leads to an enrich-

    ment in 18O in the final CO2 gas.

    To a first approximation, the CO electrical decomposition

    is analogous to a Rayleigh distillation in a closed system

    because there is a complete mixing of the product CO2; then:

    pt rt;0 1000 1 fr

    1 fr 1000 13

    Supposing that drt,0 8.6% (seawater SO42) and calculat-ing the product composition (dpt) at fr 0.1 to 0.0 (fraction ofCO unconverted), the oxygen composition of the total CO2from BaSO4 reduction should be between 8.9 and 11.3%(Table 6). Fractionation connected to CO glow discharge

    processes has been studied recently at voltages between 3 and

    9 kV, at 50 Hz,45 confirming the enrichment of 18O in CO2 and

    indicating the presence of a brown-black deposit on the

    reactor wall connected to the formation of a mixture of

    polymeric forms of malonic anhydride C3O2 (also noted

    by some authors during the setting of the method for

    BaSO4 processing; G. Cortecci and A. Longinelli, personal

    communication).

    Effects 3 and 4 can be rendered negligible provided that

    pure BaSO4 is analyzed.46 Other problems could derive from

    defective mixing of BaSO4 with graphite, because the CO

    generated could diffuse and react with uncoated BaSO4particles:47

    BaSO4 4CO , BaS 4CO2 14CO2 could diffuse back into carbon to generate more CO

    according to the Boudouard reaction, but at the temperature

    of the method the rate of reaction (14) is higher than the

    reaction rate of the Boudouard reaction. This problem,

    however, was minimized by mixing BaSO4 with a large

    excess of graphite.23

    CONCLUSIONS

    Continuous-flow determination of d18O composition inorganic/inorganic materials is an already established

    technique, reducing sample size and times of analyses in

    comparison with an off-line technique. Our method of stan-

    dardization extends these advantages by working with only

    solid standards. Our method can analyze samples in the

    range 36.4 d18O 3.3 (% vs. V-SMOW), but long-term sta-bility was experimentally confirmed up to 6%. Eventually,a carbonate standard with a value of d18O&20% could besynthesized using Kim and ONeils equilibration method48

    with a water sample of d18O&50% (e.g. nival precipita-tions or Artic ice). Considering the small amount of sample

    required for on-line analyses, solid standards with homoge-

    neous small grains are recommended. Our determinations on

    different standard matrices (calcium carbonate and barium

    sulfate) were successfully performed using AgCl which (i)

    obtains the complete reduction of the compounds and (ii)

    acts on the kinetic of reactions primed by the gaseous chlor-

    ine. Differences between off-line and on-line reduction techni-

    ques in the d18O determination of barium sulfate (0.6%) arerelated to some analytical problems in the former method,

    Table 6. Calculated isotopic composition of total CO2 resulting from BaSO4 carbon reduction off-line method considering

    fractionation effects due to the disproportionation of CO to CO2. The CO2 values were calculated as [(8.6 y) (dpt x)]. Boldcontoured values are the most probable isotopic values considering the molar ratio of x (CO2) and y (CO2) obtainable by the

    method. In the calculations, we have assumed no significant difference between the first formed gases from BaSO4 carbon

    reduction,25 so drt,0 d(CO) d(CO2) 8.6%

    x = 1-y 1-y 1-y 1-y 1-y 1-y 1-y 1-y 1-y % of CO2 from BaSO4y = (0.9*(1-fr)) (0.8*(1-fr)) (0.7*(1-fr)) (0.6*(1-fr)) (0.5*(1-fr)) (0.4*(1-fr)) (0.3*(1-fr)) (0.2*(1-fr)) (0.1*(1-fr)) % of CO2 from CO

    fr pt

    0.10 13.64 12.69 12.23 11.78 11.32 10.87 10.42 9.96 9.51 9.050.09 13.29 12.44 12.01 11.59 11.16 10.73 10.31 9.88 9.45 9.030.08 12.92 12.18 11.78 11.38 10.98 10.59 10.19 9.79 9.39 9.000.07 12.53 11.89 11.53 11.16 10.79 10.43 10.06 9.70 9.33 8.970.06 12.12 11.58 11.25 10.92 10.59 10.26 9.92 9.59 9.26 8.930.05 11.69 11.24 10.95 10.65 10.36 10.07 9.77 9.48 9.19 8.890.04 11.22 10.86 10.61 10.36 10.11 9.86 9.61 9.35 9.10 8.850.03 10.71 10.44 10.24 10.03 9.83 9.62 9.42 9.21 9.01 8.800.02 10.15 9.97 9.81 9.66 9.51 9.36 9.21 9.06 8.90 8.750.01 9.50 9.40 9.31 9.22 9.13 9.04 8.95 8.87 8.78 8.690.00 8.60 8.60 8.60 8.60 8.60 8.60 8.60 8.60 8.60 8.60

    dptmean isotopic composition of the CO2 product reservoir at the time t following Rayleigh distillation under closed system conditions (see textfor details); fr remaining fraction of CO reactant reservoir at the time t.

    High-temperature continuous-flow 18O measurements of sulfates 3013

    Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 30073014

  • such as isotopic exchange of the gases with the quartz cham-

    ber, a kinetic fractionaction factor connected with the incom-

    plete reaction yields and possible memory effect caused by

    polymers of malonic anhydride. In the light of this and con-

    sidering the higher temperature and the single step of the

    reduction reaction to produce the measuring gas in the

    TCCFIRMS method, we think that the on-line high-tempera-

    ture carbothermic reduction using the AgCl additive is

    the best way to produce 18O/16O ratios of barium sulfate

    samples.

    AcknowledgementsWe are grateful to Gianni Cortecci (Institute of Geosciences

    and Earth Resources, CNR, Pisa, Italy) who improved the

    manuscript. G.C. also provided us with MSS3 standard.

    Many thanks are due to also Antonio Longinelli, Lorenzo

    Toscani and Giampiero Venturelli (University of Parma,

    Italy) for useful comments and suggestions. CaCO3 off-line

    analyses were performed by Maura Pellegrini and Enrico

    Selmo at Earth Sciences Department, University of Parma.

    We are indebted to Prof. Stan Halas (Mass Spectrometry

    Lab., Institute of Physics, UMCS, Lublin, Poland) for permis-

    sion to publish his results on the NBS-127 standard. Finally,

    special thanks are extended to Guido Giazzi and Roberto

    Manzoni (Thermo Electron, Rodano, Milan, Italy) for support

    and technical assistance on the TCEA unit. The valuable com-

    ments of Prof. John J. Monaghan (University of Edinburgh)

    and of an anonymous referee helped to improve the clarity

    of the manuscript.

    REFERENCES

    1. Kornexl BE, Werner RA, Gehre M. Rapid Commun. MassSpectrom. 1999; 13: 1248.

    2. Epstein S, Mayeda T. Geochim. Cosmochim. Acta 1953; 4: 213.3. Showers WJ, Barrick R, Genna B. Anal. Chem. 2002; 74: 143A.4. McCrea JM. J. Chem. Phys. 1950; 18: 849.5. Guichon G. Ann. Chim. 1960; 5: 295.6. Merz W. Anal. Chim. Acta 1970; 50: 305.7. Panda PK, Mariappan L, Kannan TS. Ceramics Int. 1999; 25:

    467.8. Kornexl BE, Gehre M, Hofling R, Werner RA. Rapid

    Commun. Mass Spectrom. 1999; 13: 1685.9. Gehre M, Strauch G.Rapid Commun.Mass Spectrom. 2003; 17:

    1497.10. Thermo Finnigan, High Temperature Conversion Elemental

    Analyzer (TC/EA)Operating Manual, Issue 9/2003.11. Werner RA, Bruch BA, Brand WA. Rapid Commun. Mass

    Spectrom. 1999; 13: 1237.12. Bale CW, Chartrand P, Degterov SA, Eriksson G, Hack K,

    Ben Mahfoud R, Melancon J, Pelton AD, Petersen S. Calphad2002; 26: 189.

    13. Bailey JE, Brinker CJ, Cornils B. Ulmanns Encyclopedia ofIndustrial Chemistry (6th edn). John Wiley: Chichester, 2002.

    14. Kirk-Othmer. Encyclopedia of Chemical Technology (4th edn).John Wiley: Chichester, 2001.

    15. Loss RD. Pure Appl. Chem. 2001; 75: 1107.16. Gonfinatini R, Stichler W, Rozanski K. In References and

    Intercomparison Materials for Stable Isotopes of Light Elements;IAEA-TECDOC-825: Vienna, 1995; 1329.

    17. Ustinov VI, Grinenko VA. Russ. J. Phys. Chem. 1982; 56:1677.

    18. Koziet J. J. Mass Spectrom. 1997; 32: 103.19. Babaeva M, Bystrov DS, Kovalgin AYu, Tsyganenko AA.

    J. Catal. 1990; 123: 396.20. Yanagisawa Y, Kashima S. Surf. Sci. 2000; 454/456: 379.21. Gehre M, Geilmann H, Richter J, Werner RA, Brand WA.

    Rapid Commun. Mass Spectrom. 2004, 18: 2650.22. Werner RA. Isot. Environ. Health Stud. 2003; 39: 85.23. Farquar GD, Henry BK, Styles JM. Rapid Commun. Mass

    Spectrom. 1997; 11: 1554.24. Straight WH, Karr JD, Cox JE, Barrick RE. Rapid Commun.

    Mass Spectrom. 2004; 18: 2897.25. Sakai H, Krouse HR. Earth Planet. Sci. Lett. 1971; 11:

    369.26. Bohlke JK, Mroczkowski SJ, Coplen TB. Rapid Commun.

    Mass Spectrom. 2003; 17: 1835.27. Hall AJ, Boyce AJ, Fallick AE, Hamilton PJ.Chem. Geol. 1991;

    87: 99.28. Bottcher ME, Bernasconi SM, Brumsack HJ. In Proc. Ocean

    Drilling Program, Scientific Result, vol. 161, Zahn R, ComasMC, Klaus A (eds). 1999; 413421.

    29. Palmer MR, Helvac C, Fallick AE. Chem. Geol. 2004; 209:341.

    30. Peryt TM, Halas S, Kovalevych VM, Petrychenko OY,Dzhinoridze NM. Geol. Quat. 2005; 49: 235.

    31. Holser WT, Kaplan IR, Sakai H, Zak I. Chem. Geol. 1979;25: 1.

    32. Claypool GE, Holser WT, Kaplan IR, Sakai H, Zak I. Chem.Geol. 1980; 28: 199.

    33. Zak I, Sakai H, Kaplan IR. In Isotope Marine Chemistry,Golberg ED, Horibe Y, Sakurashi K (eds). Rokakuho: Tokyo,1980; 339373.

    34. Longinelli A, Craig H. Science 1967; 156: 56.35. Cortecci G, Molcard R, Noto P. Nature 1974; 250:

    134.36. Leone G, Ricchiuto TE, Longinelli A. In Studies on Sulphur

    Isotope Variations in NatureProceedings of an Advisory GroupMeeting; IAEA: Vienna, 1987; 514.

    37. Rafter TA, Mizutani Y. N. Z. J. Sci. 1967; 10: 816.38. Rafter TA. N. Z. J. Sci. 1967; 10: 493.39. Mizutani Y. Geochem. J. 1971; 5: 69.40. Shakur MA. PhD thesis, Department of Physics, University

    of Calgary, Calgary, Alberta, Canada, 1982.41. Lloyd RM. J. Geophys. Res. 1968; 73: 6099.42. Longinelli A, Cortecci G. Rend. Soc. Ital. Min. Pet. 1970; 26:

    733.43. Holt BD, Engelkemeir AG. Anal. Chem. 1970; 42:

    1451.44. Holt BD. Anal. Chem. 1977; 49: 1664.45. Mori S, Akatsuka H, Suzuki M. J. Nucl. Sci. Technol. 2001; 38:

    850.46. Longinelli A. In Handbook of Environmental Isotope Geochem-

    istry, vol. 3, The Marine Environment, A, Fritz P, Fontes JCh(eds). Elsevier: New York, 1989; 219255.

    47. Jagtap SB, Pande AR, Gokarn AN. Ind. Eng. Chem. Res. 1990;29: 795.

    48. Kim SA, ONeil JR. Geochim. Cosmochim. Acta 1997; 61: 3461.49. Available:http://www.cstl.nist.gov/div837/837.01/outputs/

    standards/algorithm/background.htm.

    3014 T. Boschetti and P. Iacumin

    Copyright # 2005 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2005; 19: 30073014