    Z540.3 Compliance, and thegeneral topic of Accreditation,

    Whats an accredited cal?

    Bob Stern (707) 577-2886SSU - AmericasAgilent NCSLI member delegateVoting member, 174 Stds Writing CommitteeMember, 171 Z540.3 Handbook CommitteeMember, Agilent Cal Cert Working Group (CCWG)

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    2/23

    NCSLI TutorialNCSLI TutorialImplementing ANSI/NCSL Z540.3Implementing ANSI/NCSL Z540.3NCSLI TutorialNCSLI TutorialImplementing ANSI/NCSL Z540.3Implementing ANSI/NCSL Z540.3

    July 2009 @ NCSLI

    Mark your calendar for the Wed 24 Mar 2010 tutorial atMeasurement Science Conference in Pasadena

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    3/23

    New Z540.3 Web Portal

    http://www.agilent.com/find/Z540.3

    ,

    , ,

    articles, etc.List of models with Z540.3 calibration compliance from Agilent Technologies

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    4/23

    Did you Know?

    Guidelines on Assessment and Reporting of Compliance with Specification ILAC-G8:1996

    Document can be downloaded free: http://www.ilac.org/

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    5/23

    ILAC-G8 Conformance

    Upper acceptance limit

    Upper specified tolerance

    ??

    ??

    ??

    Fail

    Nominal

    Lower specified tolerance

    Pass

    Pass

    95% expanded uncertainty

    Guidelines on Assessment and Reporting of Compliance with Specification ILAC-G8:1996Document can be downloaded free: http://www.ilac.org/

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    6/23

    ILAC-G8

    ILAC G8, Appendix A, page 10, drives Euro-ABs (like UKAS)to require Guard bands for K=2 (95% confidence interval)

    Source: Guidelines on Assessment and Reporting of Compliance with Specification ILAC-G8:1996

    Pass #,Indeterminate

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    7/23

    Z540.3 Handbook Compliance Method #5

    Upper acceptance limit

    Upper specified tolerance95 % expanded

    uncertainty guard

    band

    Pass

    Fail

    Fail

    Fail

    Fail

    Fail

    Adjust if overAcceptance limit

    Nominal

    Lower acceptance limit

    Lower specified tolerance

    Pass

    95% expanded uncertainty

    95 % expandeduncertainty guard

    band

    Guidelines on Assessment and Reporting of Compliance with Specification ILAC-G8:1996Document can be downloaded free: http://www.ilac.org/

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    8/23

    Pragmatic Method for Pass/Fail Conformance

    Reporting

    Upper acceptance limit

    Upper specified tolerance95 % expanded

    uncertainty guard

    band

    Fail1

    Fail1Fail1

    Fail

    95% expanded uncertainty

    Adjust if overAcceptance limit

    Nominal

    Lower acceptance limit

    Lower specified tolerance

    Pass

    Fail1Measured value

    95 % expanded

    uncertainty guard

    band

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    9/23

    NCSLI Measure Magazine, March issue

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    10/23

    Z540.3 Requirements for outside

    calibration suppliers

    Source: Z540.3, paragraph 5.5 Outside Suppliers

    Accreditation?

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    11/23

    A2LA now accredits to Z540.3

    Source: http://www.a2la.org/documents/searchdocuments.cfm

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    12/23

    Excerpt: A2LA C207 audit checklist for Z540.3

    Source: http://www.a2la.org/documents/searchdocuments.cfm

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    13/23

    Confused about Accreditation??

    1. What exactly is accreditation?

    2. How can I use it to compare Calibration Lab A to Lab B?

    . a o ge or er an ccre e a

    4. How is an accredited cal different (or better) from a standard calibration?

    5. Is Accredited cal better than ISO 17025?

    6. Can I expect to see Z540.3 accredited calibration service?

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    14/23

    How to compare cal lab quality check for

    Accreditation, right?

    ABC XYZ

    Source : http://www.a2la.org/dirsearchnew/newsearch.cfm

    A2LA Cert No. 2546.01A2LA Cert No. 1395.05

    Is this all you need to know?

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    15/23

    Scope of Accreditation

    1. The calibration lab tells the A2LA auditor which parameters it seeksaccreditation.

    2. For each requested parameter the A2LA auditor then:i. Spends time with the technicians assessing their competency/measurement technique

    to measure this particular parameter.

    ii. Audit the measurement uncertainty documentation.

    How it works:

    . u t tracea ty pat to a suc as or

    All Americas calibrations are performed by a lab withthe scope-of-accreditation for the relevant parameters.In rare cases, instruments are transhipped to Roseville, CA if needed.

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    16/23

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    17/23

    Authorized Use of A2LA Symbol general guidance

    Source: P101 Reference to A2LA Accredited Status A2LA Advertising Policy

    A2LA can only act when their symbolappears on a calibration certificate

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    18/23

    Allowed Use of A2LA Symbol on Cal Certificates

    Next year

    943 models

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    19/23

    Coming Soon What is the benefit to youIf your supplier includes theirAccreditation body symbol onthe Cal cert for your instrument?

    . . . it means they welcome close auditorscrutiny of that particular procedure.

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    20/23

    Scope of Accreditation The Backbone

    Parameter/Equipment Range Best Uncertainty () Comments

    DC Voltage Generate Up to 0.22 V 7.5 V/V + 0.4 V Fluke 5720A/5725A

    Up to 2.2 V 5 V/V + 0.7 V

    Up to 11 V 3.5 V/V + 2.5 V

    Up to 22 V 3.5 V/V + 4 V

    Up to 220 V 5 V/V + 40 V

    Up to 1100 V 6.5 V/V + 400 V

    AC Current Measure

    Group/Presentation TitleAgilent Restricted

    Month ##, 200X

    Resistance Measure

    Capacitance GenerateRF Absolute Power Measure

    Attenuation Generate

    Digital Modulation Measure

    Etc, Etc, Etc

    The scope of accreditation is a listing of all of the parameters that a lab is capable

    of making accredited measurements in. The lab decides which parameters itwants to represent, but the accrediting body assesses the capability.

    The customers focus: What are youcapable of measuring, in what ranges,and how accurate are you?

    Source : Scott Arrants

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    21/23

    Your measurement uncertainty may vary . . .

    The fine print at the end of any lab accreditation certificate . . .

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    22/23

    Z540.3 Compliance its boththe Scope andthe individual procedure

    3458A

    Meas. Uncertainty

    Part of BMC

    User 5 digit DMMto be calibrated

    Josephson Junction atAgilents Loveland Primary Stds. Lab

    ABC voltmetercal system

    1. Meas. Uncertainty ofABC system?

    2. Compliance methodfor 5 digit DMM calprocedure??

  • 7/31/2019 Bob-Stern Roseville NCSLI 1410

    23/23

    Accreditation Evaluation Summary Points

    1. Look past the front page!

    2. Make sure the scope covers all the parameters of your instrument.

    3. Make sure the calibration measurement capability (aka. best uncertainty)can measure your instrument.

    4. Check what lab standards are used for our instrument in the trace info

    on a sample cal certificate.5. Z540.3 Accreditation ask for it A2LA is open for business

    6. Look for the accreditation symbol on the Cal certificate.