Photodynamic Diagnosis (PDD)
DETECTION Equipment Overview
Blue Light Flexible Cystoscopy with Cysview®
Detection: Flexible Blue Light Cystoscopy with Cysview®
1) Equipment Overview
• PDD Flexible Video Cystoscope
• Suction Assembly
• PDD CCU – TELECAM SL II FI
• PDD Light Source – D-Light C
• Suction Pump – ENDOMAT® LC
2) Procedure
• Equipment Start-Up Sequence
• Troubleshooting
3) Reprocessing: Cleaning and Sterilization
Detection with Flexible PDD: Equipment Overview
Component Name Part # Notes
Flexible
Cystoscope
Flexible
PDD Video
Cystoscope
11272VNIUAK/
11272VNIAK
Can be used in-office without
general anesthesia. Connecting
cable 20213070 is needed and
sold separately.
Camera
Control Unit
(CCU)
TELECAM
SL II FI
20213011U1 Flexible PDD-specific sensitivity
and image brightness
Light Source D-LIGHT C 20133620-134 Same light source used for both
Detection and Resection
Suction Pump ENDOMAT®
LC
20330320-1 Used to provide a clear surgical
view during procedures. Cysview
combined with urine can cause a
murky green effect. Wall suction
can be used as an alternative
Documentation
Unit
AIDATM HD
Connect
20205501-140 Record clinical images and
videos of each Blue Light case.
Flexible PDD Cystoscope
Light Source
Connection Leakage Tester
Connection
Deflection
Lever Handle
Suction
Assembly
Video Adapter
Cable (20213070)
Video connection
from Flexible PDD
Cystoscope to CCU
Flexible
Sheath
Working Channel
Flexible PDD Cystoscope
Working Channel Instruments
• 27425ZG Forceps, Biopsy, Flexible,
Double, 5Fr, 60cm
• 27425FG Forceps, Grasping,
F/Stone (Stent Removal), 5Fr,
60cm
• 27723T Electrode, Bugbee,
Monopolar, Ball, 73cm, 4Fr
• 26006M HF Cord, Monopolar,
Bugbee, 8mm plug, 300cm
Best Practice: When using a bugbee,
plug generator into a different outlet
from the Flex PDD tower to minimize
risk of video interference
Video Adapter Cable (20213070)
1. Unscrew the protection cap on PDD Flexible Cystoscope and plug in the Video Adapter Cable (20213070) by aligning white dots
2. Next, insert the PDD Video Cystoscope into the Light Source Connection of the D-Light C until it clicks.
3. Plug the other end of the Video Adapter Cable into the CCU. The top of the video adapter cable connector is labeled “UP”.
Flexible PDD Cystoscope
Button 3
• Quick Press: Menu
• Long Press: N/A
Button 1
• Quick Press: Toggle Blue Light On/Off
• Long Press: Brightness (“PEAK” setting is recommended)
Button 2
• Quick Press: Image Capture
• Long Press: Image Archive
Flexible PDD Cystoscope – Suction Assembly
Suction Valve 27100MV
(Single-Use)
Valve Intake 11900VA
Connecting Tube 27100MS
(Single-Use)
Forceps Unit/ Luer Lock 11272AZK
Luer Connector w/Stopcock 27502
Flexible PDD Cystoscope – Suction Assembly
1) Attach the Forceps Unit and Valve Intake to each end of the Connecting Tube.
2) Insert the Valve Intake into the channel in the scope handle
3) Attach the Forceps Unit to the instrument channel port in scope handle.
4) Screw the Suction Valve onto the Valve Intake.
Flexible PDD CCU – TELECAM SLII FI
Power
Switch
White Balance
Required before each
procedure
Video Adapter
Cable (20213070)
Video connection
from Flexible PDD
Cystoscope to
CCU
Connection for PDD
Flexible Cystoscope
Use connecting cable
(20213070) from scope
to CCU
Flexible PDD CCU – TELECAM SLII FI
SCB Cable Connection
Ensure the SCB cable is
connected between CCU
and light source
Flexible PDD Light Source – D-Light C
Power
Switch
Standby
Avoid thermal
damage when
cystoscope is not
in use
Brightness
Adjustment
Set to 100%
Blue Light On/Off
Toggle between White
Light and Blue Light.
Function duplicated on
Cystoscope Button #1
Light Cable
Connection
Mode
Always use MANUAL setting
Flexible PDD Light Source – D-Light C
Lamp Life
Warning Light
Turns on at 350
hours of service.
Call your KARL
STORZ Rep if
warning light
appears
Total Hours of Usage
Appears for 3 seconds
when unit is powered up
Note: The model number for
D-Light C is the same for Rigid
PDD and Flex PDD
Units with “-E” at the end of
the serial number are
compatible with both
Rigid and Flex PDD
Units without “-E” in the
serial number are compatible
with Rigid PDD only
Power
Switch Suction Rate
Recommended setting:
400 mL/min
Suction
Start/Stop
Mode
Select “Normal”
(0-1000mL/min)
Pump Head
Connection for
Suction Tubing
(031247-10)
ENDOMAT® LC
ENDOMAT ® LC
Inflow
Outflow
Pump Reel
Lever
1) Lower the Pump Reel Lever until it is horizontal
2) Insert the tubing into the pump head with the inflow on
the bottom and outflow on the top
– Note: The suction tubing is directional and will only fit
properly when inserted in the correct direction
3) Raise pump reel lever back to the vertical position
4) Connect the free end of the inflow tube to the Valve
Intake (11900VA) of the Suction Assembly on the PDD
Flexible Cystoscope.
5) Connect the free end of the outflow tube to the
collection container for suction waste.
Flexible PDD Cystoscope – ENDOMAT ® LC Setup
Luer Connector 27500/3
Suction Tubing 031247-10
(Single-Use)
Inflow Outflow
Inflow
Outflow
NOTE: The shorter piece of
tubing included with the suction
tubing set is not needed for the
Flex PDD procedure and can be
discarded.
Discard
Procedure: Flexible Blue Light Cystoscopy
1) Cysview® is instilled directly into the patient’s bladder via catheter one hour before Blue Light Cystoscopy, which should be performed within 1-3 hours after instillation.
2) Flexible PDD Cystoscope is inserted
3) Bladder is completely emptied using the suction feature on the Flexible PDD Video Cystoscope and refilled again.
4) Bladder is examined and mapped using both white light and blue light
Start-Up Sequence
1) Plug the PDD Video Cystoscope into the Camera Control Unit (CCU) and the D-Light C
2) Start Up Sequence:
3) Power up remaining components (ENDOMAT® LC, etc.)
Monitor CCU D-Light C AIDATM
Troubleshooting Problem: No Light at distal end of the scope
Solution:
1) Check the SCB cable connection between CCU and D-Light C
2) Check the connection between Video Adapter Cable (20213070) and the PDD Video Cystoscope
3) Check connection between Video Adapter Cable and CCU
4) Shut down the system and reboot according to Start-Up Sequence:
Monitor CCU D-Light C AIDATM
Reprocessing
Sterilization
Flexible PDD Video Cystoscope
• STERIS System 1E (SS1E) sterilization is validated for the PDD Video Cystoscope
• For High Level Disinfection (HLD), Resert XL HLD (2.0% Hydrogen Peroxide) is approved
Suction Assembly
• Steam Sterilization is approved for reusable components
• The Suction Valve and Connecting Tube are single-use and should be discarded
Suction Assembly
• Disassemble the Suction Assembly
• Discard Suction Valve and Connecting Tube
• The Valve Intake and Forceps Unit can be reused after they have been cleaned and steam sterilized.
• Important: Double check that the base and double luer lock of the Forceps Unit remain screwed in tightly together after steam sterilization. If not, leakage can occur during a case.
Reprocessing
Reprocessing Post-Procedure
1) Wipe down the Flexible PDD Cystoscope with a lint free, disposable cloth, moistened with water
2) Detach the Video Adapter Cable
3) Close the cap on the Flexible PDD Cystoscope for the video cable connector and ensure it is screwed on tightly.
Leakage Test
1) Ensure that the Leakage Tester Connecting Tube and Leakage Tester Connection on the scope are completely dry
2) Follow directions in IFU
3) Use the grey button on Leakage Tester to release remaining pressure and detach from scope
Reprocessing
PDD Flexible Video Cystoscope
• Immerse in cleaning solution
• Rinse and brush working channel using Cleaning Brush (110910-50) and Irrigation Adapter (11900RA)
• Attach Irrigation Adapter by pressing and twisting simultaneously to ensure a tight fit
• Attach syringe to proximal end of Irrigation Adapter to rinse working channel
• Clean distal lens
• Inspect for visual damages
Reprocessing
STERIS System 1E (SS1E)
For SS1E parts, order directly from Steris: (800) 548-4873
STERIS Quick Connect
QKC1734E
STERIS Universal Flexible Processing Tray
C1160E
In case of technical difficulties please call:
• Contact: (800) 421-0837; Press 2
• Support email: [email protected]
• Hours of Operation : 6:00am to 3:00pm PST Monday-Friday
• Medical Emergency After Hour Support 24/7/365
Technical Support & Repair
Cysview Risk & Safety Information