bloodhound™ logging system infrared™ hydrocarbon analyzer system 0-100% auto ranging total gas...
TRANSCRIPT
![Page 1: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/1.jpg)
Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System• 0-100% auto ranging • Total gas with Chromatograph model- C1-C4 iso and C4
normal • Accurate and Sensitive with Auto Ranging • Low end +/ - .0001% = one unit (on 10,000 unit full range) • High end +/- 1 % full scale • Reliability
– No filament, flame or pellistor to fail – No saturation – No in field calibration (set and stable for 18 months) – No drift from filament degradation – No drift from carbonization
• CO2 - included• O2 - included • H2S - included
DATA • Triple redundant data records • Real time (6 second new data capture cycle) data viewing at
rig site or anywhere world wide via web site • WITS compatible (can output or collect and display incoming
drilling data)
AUTOMATION • Remote systems diagnostic capabilities • Multiple function alarm system • Automatically generated real time mudlog (just add lithology) • Automatic sample flow monitor and adjustment
COMMUNICATION• Hardwire and wireless communication capabilities • Built in GPRS data modem eliminates satellite dish and rental costs • Alarm notification via cell phone text message world wide
POWER
• Low operating current design (DC low voltage extends life and eliminates shock danger)
• Functional from 12v car electrical system • Up to 120 min on board backup power supply (with monitor)
MONITORING• Remote diagnostics on sample flow, sample line vacuum pressure,
total hydrocarbons and other functions • Multiple function automatic alarms based on custom high low
parameters • Automatic alarm notification via cell text message • 24/7 remote manned alarm monitoring service packages
http://www.bloodhoundrentals.com
![Page 2: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/2.jpg)
Bloodhound™ Logging System Side View Sample Gas inlet
Port has regulated flow and vacuum
pressure
Exhaust port. ½” size keeps the line
from freezing in cold weather
Geolograph switch input
GPRS Radio antenna
Ethernet RJ-45 to Internet
Audio Alarm
Positive Pressure vent air
supply. Inside case under
positive pressure with filtered air.
Analog Out Relay contacts
for Depth & Alarms
DC power in/out Runs from DC car adaptor
or powers DC devices e.g. iBall Cavitator
Universal AC power From 90-250VAC
power and 45-65 Hz
WITSPason
compatible RS422
interface
WITSNull Modem Connection(9-pin FtoF)
RS232 Straight-thru Personal Computer
or External Data logger
USB-B Serial (COM)
PC connection (PC Driver Req’d)
![Page 3: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/3.jpg)
Gas Charting Functions
•Calculates Lag•Set PC Alarms•Edit Depth•Zoom on Chromats•Data File Export
Communication Network
http://www.bloodhoundrentals.com
Internet
64.19.xx.xx:xxxx
Bloodhound™Data Archive
Web Remote Gas ChartWeb Browser
Well Status via Mobile Device Accesse.g. Voice Response – Text - Email
GPRS Cell RadioAT&T Data Service
Field Data NetworkEthernet connection(e.g. Sat. Internet)
Bloodhound™ Logging System• Total Gas• Gas Chromatography• CO2• O2• H2S• Local Instrument Alarms
Serial ConnectionInstrument to PC
At Well Site
WITS
WITs to EDR
Gas Chart - Local
Well Site PC
Gas Chart - Remote
Remote PC
Gas Chart - Remote
Remote PC
![Page 4: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/4.jpg)
Data Logger SoftwareUsed both for Local & Remote Logging
• Application Logging software used Locally or Remotely
• Real-Time Data– Table (6 sec update)– Flexible Charting capability
• Configuration– Lag– Depth editing– Scale adjustments– Export LAS and XLS data
file for importing to any Mudlogging software
http://www.bloodhoundrentals.com
![Page 5: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/5.jpg)
Web Remote Data Logger
• Accessible from any Internet connected browser
• Real-Time Data– Table
• 10 sec update
– Chart Image• 1 min update
• Comment Editing Capable
http://www.bloodhoundrentals.com
![Page 6: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/6.jpg)
Mobile Device Access
• Mobile Phones based on Windows Mobile
http://www.bloodhoundrentals.com
• IPhone
![Page 7: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/7.jpg)
Real-Time Data Viewer
Sample Line Vacuum
Pressure Great for
troubleshooting leaks, blockages
etc.
Flow Rate Automatically
controlled
Variable Speed Sample Pump controls Flow
Rate to Set Point
RSSIReceived
SignalStrength Indicator
Monitors Cell network signal
http://www.bloodhoundrentals.com
![Page 8: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/8.jpg)
Real-Time Data Viewerwith WITS Stream Viewer
![Page 9: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/9.jpg)
Multiple Well Real-Time Data Viewer
![Page 10: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/10.jpg)
Website Multiple Instrument Viewerhttp://www.iballmultiview.com
• Accessible from any Internet connected browser
• Multiple Well Snapshot View– Real-Time Data
– Table• 10 sec update
![Page 11: Bloodhound™ Logging System Infrared™ Hydrocarbon Analyzer System 0-100% auto ranging Total gas with Chromatograph model- C1-C4 iso and C4 normal Accurate](https://reader036.vdocuments.us/reader036/viewer/2022062806/56649e715503460f94b70750/html5/thumbnails/11.jpg)
24/7 Call Center Support(303-459-4660)
• Technical Support– Tier 1 call center agents available 24 hours a day, 7 days a week
• Interactive Voice Response (IVR) w/Text & Email– Voice Response - Available 24/7 for retrieving a voice response
that recites individual Bloodhound Hole Depth, Bit Depth and Total Gas
– Text Message - Available 24/7 for retrieving Hole Depth, Bit Depth, Total Gas, Sample line flow rate, Vacuum pressure and ROP in min/ft and ft/hr
– Email – Available 24/7 by request with all Bloodhound diagnostic information
• Sales– Agents available during business hours