blood glucose and ketone monitoring system

2
APEXBIO Taiwan Glucose and ketone testing 0.5 μL for glucose 0.8 μL for ketone Large memory PC link Blood Glucose and Ketone Monitoring System Bluetooth 15197:2015 Autocoding Glucose and Ketone testing

Upload: others

Post on 16-Oct-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

APEXBIOT a i w a n

Glucose andketone testing

0.5 µL for glucose0.8 µL for ketone

Large memory PC link

Blood Glucose and Ketone Monitoring System

Bluetooth

15197:2015

Autocoding

Glucose andKetone testing

Blood Glucose and Ketone Monitoring System

Ketones are formed when there is not enough sugar or glucose to supply the body’s fuel needs, the body turns to fat for energy. When fat is broken down, ketone bodies are made and can accumulate in the body(1). In an individual with diabetes, ketones are more likely to show up when there is not enough insulin in the body(1). Untreated high blood glucose with positive ketones can lead to a life-threatening condition called diabetic ketoacidosis (DKA)(2). The MultiSure GK Blood Glucose and Ketone Monitoring System o�ers a quick and easy way to test for both glucose and ketone bodies in blood using separate test strips and determine the immediate risk for DKA. Blood glucose testing requires 0.5 μL of blood and result can be obtained in 5 seconds. Ketone testing requires 0.8μL of blood and result can be obtained in 8 seconds.

(1) Ketone (n.d.). University of California, San Francisco, Diabetes Education Online. Retrieved Oct. 23, 2014, from http://dtc.ucsf.edu/types-of-diabetes(2) Erika Gebel (2013). How to Avoid DKA, American Diabetes Association. Retrieved from http://www.diabetesforecast.org/2013/mar/how-to-avoid-dka.html

Specifications

Measurement Range

Sample Volume

Quick Testing Time

Hematocrit

Operating Temperature

Memory

Additional Features

Glucose: 20~600 mg/dLKetone: 0.1~8.0 mmol/L

Glucose: ≥ 0.5 μL / Ketone: ≥ 0.8 μL

Glucose: 5 seconds / Ketone: 8 seconds

Glucose: 10%~70% / Ketone: 20%~60%

5°C~45°C

Glucose: 500 test results / Ketone: 100 test results

PC link, Bluetooth (optional)

Rev.11/18

No.7, Li-Hsin Rd. V, Hsinchu Science Park, Hsinchu, Taiwan, R.O.C. TEL : 886-3-5641952 / FAX : 886-3-5678302 / E-Mail : [email protected] / Website : www.apexbio.com

APEXBIOT a i w a n

ISO 15197 : 2015Sample < 100 mg/dL

Sample ≥100 mg/dL

±10 mg/dL196 / 198

99%

±10%393 / 402

98%

±15 mg/dL198 / 198100.0%

±5 mg/dL146 / 198

74%

±5%317 / 402

79%

±15%401 /402

99.8%

Glucose Testing Performance

y = 1.0054x -1.734R² = 0.9959

0

50

100

150

200

250

300

350

400

450

500

550

0 50 100 150 200 250 300 350 400 450 500 550

Mul

tiSur

e G

K M

eter

Glu

cose

read

ing

(mg/

dL)

YSI mean value (mg/dL)

ABCDE

A

B

C

D

Ketone Testing Performance

y = 0.9885x + 0.2066R² = 0.9908

0.01.02.03.04.05.06.07.08.09.0

10.0

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0Reference (mM)

Mul

tiSur

e G

K M

eter

Ket

one

read

ing

(mM

)

Mean.0.72.64.16.87.5

S.D.0.050.110.170.240.26

C.V.--

4.04.13.53.4

Lot 1