Bioprosthetic Valve Fracture To Optimize Results Following VIV TAVR Keith B. Allen, MD Clinical Associate Professor of Surgery, UMKC Co Director of Structural Heart St. Luke’s Mid America Heart Institute Kansas City, Missouri

Upload: phamthuy

Post on 30-Jun-2019

212 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Bioprosthetic Valve Fracture To Optimize Results Following VIV TAVR

Keith B. Allen, MDClinical Associate Professor of Surgery, UMKC

Co Director of Structural HeartSt. Luke’s Mid America Heart Institute

Kansas City, Missouri

Page 2: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

DisclosuresEdwards (Research, Proctor/Speaker’s Bureau)

Medtronic (Research, Speaker’s Bureau)Abbott (Research, Consulting)

Page 3: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

The smaller the surgical valve, the higher the mortality!

1-Year Mortality After Valve-in Valve TAVRValve-in-Valve International Data (VIVID) Registry Dvir, et al. JAMA 2014

Page 4: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Impact of Residual Gradient on 1-Year Mortality

Webb J, et al. JACC 2017; 69:2253-62

PARTNER ViV Study

Page 5: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Explanted 23 S3 VIV in a 21 Magna Ease after only two years

Durability of VIV TAVR UNKNOWN!

Transcatheter valves placed VIV are by definition constrained and not optimally expanded and the effect on durability is unknown

Page 6: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

‘Tricks’ to Optimize Results with VIV TAVR

Page 7: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Impact of Implantation Depth on Minimizing Residual Gradients

During VIV TAVRSimonato, et al. EuroIntervention 2016;12:909-917

Implant Depth

Page 8: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Optimal Implantation Depth: CoreValveSimonato, et al. EuroIntervention 2016;12:909-917

1-2 mm deep is the sweet spot

Page 9: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Optimal Implantation Depth: Sapien XT/S3Simonato, et al. EuroIntervention 2016;12:909-917

Sweet Spot is as Shallow as Comfort Allows

Page 10: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Rated for Mature Audiences Only

Page 11: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

➢ Bench testing demonstrated BVF, when performed with a non-compliant balloon and high-pressure inflation, can fracture the surgical valve and allow further expansion of the THV

➢In early clinical experience, BVF resultedin reduced residual gradients and increased valve EOA following VIV TAVR

Page 12: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

US Surgical Tissue Valves that can beFractured

Allen KB, Chhatriwalla A, et al. Ann Thorac Surg 2017;104:1501–8

Page 13: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Set up for BVF

3 41

2

Hand Inflation

Saxon JT, Allen KB, Cohen DJ, Chhatriwalla A. Bioprosthetic Valve Fracture (BVF): Bench to Bedside. Interventional Card Rev 2018;13(1):20–6.

Page 14: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Confirmation of FractureAny of the following

– The indeflator pressure drops (in the absence of balloon rupture)

– An audible snap

– The balloon waist releases – may be subtle

– Visual confirmation of the fractured ring may be difficult to detect in real time

Page 15: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

BVF Balloon Position ImportantSevere AI related to detached leaflet from CoreValve frame

can occur with incorrect balloon position

Constrained Area

Constrained Area20 mm

22 mm 23 mm

BVF balloon should not be more then 2 mm larger then CV constrained area

Allen KB, Chhatriwalla A, et al. Ann Thorac Surg 2017;104:1501–8

Page 16: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Some Valves Can Be Modified

Stretched

Page 17: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Some Valves Can Be Modified

Bent

Page 18: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

PORTICO VALVE IN A REMODELED TRIFECTA GT

Page 19: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Presented at HVS Meeting 2018

Remodeled Trifecta following VIV with Corevalve

Page 20: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Some Valves Cannot Be Fractured or Modified

Page 21: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Richard Lee and Michael LimSaint Louis University

St. Louis, MO

Brian W. Hummel Lee Memorial Hospital

Fort Myers, FL

Mark J. Russo, Bruce J. Haik Newark Beth Israel Medical Center

Newark, NJ

Pranav Loyalka, Tom C. NguyenMemorial Herman – Texas Medical Center

Houston, TX

Keith B. Allen and Adnan ChhatriwallaSt. Luke’s Mid America Heart Institute

Kansas City, MO

Josh RovinMorton Plant Hospital

Tampa Bay, FL

Juhana KarhaAustin Heart

Austin, TX

Dennis J. Gory Peace Health Medical Group

Eugene, Oregon

Anthony BavryThomas BeaverAshkan Karimi

University of Florida Gainsville, FL

Brian WhisenantIntermountain Medical Group

Sandy, UT

Ed GarrettMemphis Baptist

Memphis, TN

Vinod ThouraniWashington Med Star Hospital

Washington DC

John Webb, Danny DvirUniversity of British Columbia

Vancouver, British Columbia, Canada

DW ParkAsian Medical Center

Seoul, South Korea

Pradeep YadavPenn State Hershey Medical Center

Hershey, PA

BVF performed in 75 patients at 21 centersMarch 2016 to April 2018

Alan Yeung Stanford University

Palo Alto, CA

Adam Greenbaum MD and William O’Neill MDHenry Ford Detroit Medical Center

Detroit, MI

Alex Pak, Zafir Hawa, James Mitchell North Kansas City Hospital

North Kansas City, MO:

Jason Ricci McLaren Northern MI Hospital

Petoskey, MI

Deutsches HerzzentrumBerlin, GermanyAxel Unbehaun

Page 22: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Largest BVF Series To Date➢ Surgical valves were categorized by their true internal diameter (ID) as

- Small (< 18.5 mm)- Intermediate (>18.5 to < 21 mm)- Large (> 21 mm)

➢ THV’s were categorized as either “right sized” or “up sized” based on whether the heart team selected a THV size equal to, or larger than, that recommended by the VIV App

➢ Non-compliant balloons used to perform fracture were categorized based on their size in comparison to true ID of the surgical valve being fractured

- Small if they were < 3 mm larger and large if they were >3 mm larger

Page 23: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

METHODSInvasive hemodynamic measurements and calculation of the valve EOA were performed at

In patients that had BVF first only baseline and final gradients were obtained

Page 24: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Timing of BVF

Page 25: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained
Page 26: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

BVF Complications

Page 27: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

BVF Complications

◼ No coronary occlusions

◼ No annular ruptures (clinical or subclinical)

◼ No new pacemakers

◼ 95% (71/75) no AI

◼ 2 minor strokes → no residual

◼ 1 chordal tear → moderate-severe MR treated with MitraClip

◼ 2 severe AI from disruption of TAVR valve → treated successfully

with second valve-in-valve

Page 28: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Residual Gradient BVF after VIV TAVR (n=66)

0

20

40

60

80

40.6 ± 16 p<0.001

8.1 ± 4.8

Baseline After VIV TAVR After BVF

19.0 ± 8.8

p<0.001

Mea

n R

esid

ua

l G

rad

ien

t (m

m H

g)

Page 29: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Final Mean EOABVF after VIV TAVR (n=66)

0

0.5

1

1.5

2

2.5

0.8 ± 0.3

p<0.001

2.1 ± 0.8

Baseline After VIV TAVR After BVF

1.4 ± 0.8

p<0.001

Mea

n V

alv

e E

OA

(cm

2)

Page 30: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Univariate Predictors of Best HemodynamicsUnivariate Associations

Variable Beta Weight (95% CI) p-value

THV Type 0.15 (-2.86, 3.16) 0.92

THV Right vs Up Sized -0.77 (-3.75, 2.21) 0.61

Surgical Valve True ID 3.23 (-3.40, 10.86) 0.40

Baseline Mean Gradient 0.06 (-0.04, 0.15) 0.24

Mode of Valve Failure -2.35 (-8.28, 3.58) 0.43

Large BVF Balloon Size 4.44 (0.57, 8.32) 0.03

Timing of BVF (BVF after VIV TAVR) 8.76 (4.74, 12.79) <0.001

Page 31: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Multivariable Predictors of Best HemodynamicsMultivariable Associations

Variable Beta Weight (95% CI) p-value

THV Type 0.53 (-2.78, 3.84) 0.75

THV Right vs Up Sized 0.64 (-3.07, 4.35) 0.73

Surgical Valve True ID 3.73 (-3.62, 11.08) 0.31

Baseline Mean Gradient 0.10 (-.003, .203) 0.052

Mode of Valve Failure -0.58 (-6.44, 5.28) 0.84

Large BVF Balloon Size 4.94 (1.09, 8.80) 0.013

Timing of BVF (BVF after VIV TAVR) 8.91 (4.64, 13.18) <0.0001

Page 32: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

➢ Heavily calcified roots

➢ Intra rather then supra-annularly positioned surgical valves

19 mm Sorin Mitroflow

Page 33: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

➢ Heavily calcified roots

➢ Intra rather then supra-annularly positioned surgicalvalves

➢ Chimney roots with effaced or absent sinus that don’t have room for valve expansion

➢ Coronary occlusion

Page 34: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

CONCLUSIONS

➢BVF can be performed safelyand results in reducedresidual gradients andincreased valve area

➢ Lots still to learn

Page 35: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained
Page 36: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Univariate Predictors of Best Hemodynamics

To Be Continued

Page 37: Bioprosthetic Valve Fracture To Optimize Results Following ... · Severe AI related to detached leaflet from CoreValve frame can occur with incorrect balloon position Constrained

Multivariable Predictors of Best HemodynamicsMultivariable Associations

Variable Beta Weight (95% CI) p-value

THV Type 0.53 (-2.78, 3.84) 0.75

THV Right vs Up Sized 0.64 (-3.07, 4.35) 0.73

Surgical Valve True ID 3.73 (-3.62, 11.08) 0.31

Baseline Mean Gradient 0.10 (-.003, .203) 0.052

Mode of Valve Failure -0.58 (-6.44, 5.28) 0.84

Large BVF Balloon Size 4.94 (1.09, 8.80) 0.013

To Be Continued