    BIOCHEMICAL OXYEGEN DEMAND(BOD)

    GAYATHRI S MOHAN

    CPET 712

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    2/47

    BIOCHEMICAL OXYGEN DEMAND

    The biochemical oxygen demand (BOD) determination is an empirical test in which

    standardized laboratory procedures are used to determine the relative oxygen

    requirements of wastewaters, effluents, and polluted waters.

    Applications are in measuring waste loadings to treatment plants and in evaluating

    the BOD-removal efficiency of such treatment systems.

    The test measures the

    Oxygen utilized during a specified incubation period for the biochemical

    degradation of organic material (carbonaceous demand)

    Oxygen used to oxidize reduced forms of nitrogen (nitrogenous demand) unlesstheir oxidation is prevented by an inhibitor.

    Oxygen used to oxidize inorganic material such as sulfides and ferrous iron.

    The seeding and dilution procedures provide an estimate of the BOD at pH 6.5 to

    7.5.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    3/47

    IntroductionInterference: Oxidation of reduced forms of nitrogen, such as ammonia and organic

    nitrogen, by microorganisms.

    The interference from nitrogenous demand can be prevented by an inhibitory

    chemical.

    Ultimate BOD (UBOD)

    The UBODmeasures the oxygen required for the total degradation of organic material

    (ultimate carbonaceous demand) and/or the oxygen to oxidize reduced nitrogen

    compounds(ultimate nitrogenous demand).

    UCBOD + UNBOD UBOD

    Report results as carbonaceous biochemical oxygen demand (CBOD5) when

    inhibiting the nitrogenous oxygen demand. When nitrification is not inhibited,

    report results as BOD5

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    4/47

    Dilution Requirements

    The BOD concentration in most wastewaters exceeds the concentration of

    dissolved oxygen (DO) available in an air-saturated sample. Therefore, it is

    necessary to dilute the sample before incubation to bring the oxygen demand and

    supply into appropriate balance.

    Because bacterial growth requires nutrients such as nitrogen, phosphorus, andtrace metals, these are added to the dilution water, which is buffered to ensure

    that the pH of the incubated sample remains in a range of 6.5 to 7.5, suitable for

    bacterial growth.

    The source of dilution water is not restricted and may be distilled, tap, orreceiving-stream water free of biodegradable organics and bioinhibitory

    substances such as chlorine or heavy metals.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    5/47

    5-Day BOD TestPrinciple

    The method consists of filling with sample, to overflowing, an airtight bottle of the

    specified size and incubating it at the specified temperature for 5 d. Dissolved

    oxygen is measured initially and after incubation, and the BOD is computed from

    the difference between initial and final DO.

    Sampling and storage

    1) Grab samplesStore the sample at or below 4C from the time of collection, if

    analysis is not begun within 2 h of collection.

    Begin analysis within 6 h of collection; when this is not possible because the

    sampling site is distant from the laboratory and report length and temperature of

    storage with the results.

    In no case start analysis more than 24 h after grab sample collection.

    2) Composite samplesKeep samples at or below 4C during compositing. Limit

    compositing period to 24 h.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    6/47

    Apparatus

    Incubation bottles

    Use glass bottles having 60 mL or greater capacity (300-mL bottles having a ground-

    glass stopper and a flared mouth are preferred).

    Clean bottles with a detergent, rinse thoroughly, and drain before use.

    Use a water seal to prevent the drawing of air in to the bottle.

    Place a paper or plastic cup or foil cap over flared mouth of bottle to reduce

    evaporation of the water seal during incubation.

    Air incubator or water bath

    thermostatically controlled at 20 1C.

    Exclude all light to prevent possibility of photosynthetic production of DO.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    7/47

    Reagentsa. Phosphate buffer solution

    b. Magnesium sulfate solution

    c. Calcium chloride solution

    d. Ferric chloride solution

    e. Acid and alkali solutions,1N

    f. Sodium sulfite solution

    g. Nitrification inhibitor,2-chloro-6-(trichloromethyl) pyridine

    h. Glucose-glutamic acid solution

    i. Ammonium chloride solution

    j. Dilution water

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    8/47

    Procedure

    a. Preparation of dilution water

    Place desired volume of water in a suitable bottle and add 1 mL each of

    phosphate buffer, MgSO4, CaCl2, and FeCl3 Solution, Seed dilution water, if

    desired.

    Before use bring dilution water temperature to 20 3C.

    Saturate with DO by shaking in a partially filled bottle or by aerating with

    organic-free filtered air. Alternatively, store in cotton-plugged bottles long enough

    for water to become saturated with DO.

    b. Dilution water storage

    Source water may be stored before use as long as the prepared dilution water

    meets quality control criteria in the dilution water blank. Discard stored source water if dilution water blank shows more than 0.2 mg/L DO

    depletion in 5 d.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    9/47

    Procedure

    c. Glucose-glutamic acid check To check the dilution water quality.

    Periodically check dilution water quality, seed effectiveness, and analytical

    technique by making BOD measurements on a mixture of 150 mg glucose/L and

    150 mg glutamic acid/L as a standard check solution. Glucose has an exceptionally high and variable oxidation rate but when it is used

    with glutamic acid, the oxidation rate is stabilized and is similar to that obtained

    with many municipal wastes.

    Determine the 5-d 20C BOD of a 2% dilution of the glucose-glutamic acid standard

    check solution

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    10/47

    Procedure

    d. Seeding

    Seed source: Some samples do not contain a sufficient microbial population (forexample, some untreated industrial wastes, disinfected wastes, high-temperaturewastes, or wastes with extreme pH values). For such wastes seed the dilutionwater or sample by adding a population of microorganisms.

    The preferred seed is effluent or mixed liquor from a biological treatment systemprocessing the waste.

    Where such seed is not available, use supernatant from domestic wastewater aftersettling at room temperature for at least 1 h but no longer than 36 h.

    When effluent or mixed liquor from a biological treatment process is used,inhibition of nitrification is recommended.

    Some samples may contain materials not degraded at normal rates by themicroorganisms. In such cases develop an adapted seed in the laboratory bycontinuously aerating a sample of settled domestic wastewater and adding smalldaily increments of waste.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    11/47

    ProcedureSeed control Determine BOD of the seeding material as for any other sample. This is

    the seed control. To determine DO uptake for a test bottle, subtract DO uptake attributable to the

    seed from total DO uptake

    e. Sample pretreatment

    Check pH of all samples before testing

    1) Samples containing caustic alkalinity (pH >8.5) or acidity (pH

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    12/47

    Sample treatment In some samples chlorine will dissipate within 1 to 2 h of standing in the light. If

    residual chlorine is present, dechlorinate sample and seed the dilution water using

    Na2SO

    3.

    Determine required volume of Na2SO3 solution on a 100- to1000-mL portion of

    neutralized sample by adding 10mL of 1 + 1 acetic acid or 1 + 50 H2SO4, 10 mL

    potassium iodide (KI) solution (10 g/100 mL) per 1000 mL portion, and titrating

    with Na2SO3solution to the starch-iodine end point for residual.

    Samples containing other toxic substancesCertain industrial wastes, for example,

    plating wastes, contain toxic metals. Such samples often require special study and

    treatment.

    Sample temperature adjustmentBring samples to 20 1C before making

    dilutions.

    Nitrification inhibitionIf nitrification inhibition is desired add 3 mg of 2-chloro-6-

    (trichloro methyl) pyridine (TCMP) to each 300-mL bottle before capping or addsufficient amounts to the dilution water to make a final concentration. Note the

    use of nitrogen inhibition in reporting results.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    13/47

    Dilution technique

    Make several dilutions of sample that will result in a residual DO of

    atleast 1 mg/L and a DO uptake of at least 2 mg/L after a 5-d

    incubation.

    Five dilutions are recommended unless experience with a particular

    sample shows that use of a smaller number of dilutions produces atleast two bottles giving acceptable minimum DO depletion and

    residual limits.

    In the absence of prior knowledge, use the following dilutions: 0.0 to 1.0% for

    strong industrial wastes, 1 to 5% for raw and settled wastewater, 5 to25% for

    biologically treated effluent, and 25 to 100% for polluted river waters. When using graduated cylinders or volumetric flasks to prepare dilutions, and

    when seeding is necessary, add seed either directly to dilution water or to

    individual cylinders or flasks before dilution.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    14/47

    Dilution technique Seeding of individual cylinders or flasks avoids a declining ratio of seed to sample

    as increasing dilutions are made. When dilutions are prepared directly in BOD bottles and when seeding is

    necessary, add seed directly to dilution water or directly to the BOD bottles.

    Using a wide-tip volumetric pipet, add the desired sample volume to individual BODbottles

    Add appropriate amounts of seed material either to the individual BOD bottles orto the dilution water

    Fill bottles with enough dilution water, seeded if necessary

    Determine initial DO on one bottle

    Stopper tightly, water-seal, and incubate for 5 d at 20C.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    15/47

    Determination of initial DO If the sample contains materials that react rapidly with DO, determine initial DO

    immediately after filling BOD bottle with diluted sample.

    If rapid initial DO uptake is insignificant, the time period between preparingdilution and measuring initial DO is not critical but should not exceed 30 min.

    Dilution water blank: Use a dilution water blank as a rough check on quality ofunseeded dilution water and cleanliness of incubation bottles.

    Together with each batch of samples incubate a bottle of unseeded dilution water.Determine initial and final DO. The DO uptake should not be more than 0.2 mg/Land preferably not more than 0.1 mg/L

    Incubation: Incubate at 20C 1C BOD bottles containing desired dilutions, seedcontrols, dilution water blanks, and glucose-glutamic acid checks. Water-sealbottles.

    Determination of final DO: After 5 d incubation determine DO in sample dilutions,

    Blanks.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    16/47

    Methods to find DOTwo methods for DO analysis are

    The Winkler or iodometric method and its modifications -titrimetric

    procedure based on the oxidizing property of DO

    Electrometric method using membrane electrodes-rate of diffusion ofmolecular oxygen across a membrane

    The choice of procedure depends on the interferences present, the

    accuracy desired, and, in some cases, convenience or expedience.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    17/47

    Iodometric MethodsPrinciple

    It is based on the addition of divalent manganese solution, followed by strongalkali, to the sample in a glass-stoppered bottle.

    DO rapidly oxidizes an equivalent amount of the dispersed divalent manganous

    hydroxide precipitate to oxides of higher valence state.

    In the presence of iodide ions (I-) in an acidic solution, the oxidized manganese

    reverts to the divalent state, with the liberation of iodine (I2)equivalent to theoriginal DO content

    The released I2 can then be titrated with standard solution of sodium thiosulfate

    (Na2S2O3) using a starch indicator

    The titration end point can be detected visually, with a starch indicator, or

    electrometrically, with potentiometric or dead-stop techniques.

    The liberated iodine also can be determined directly by simple absorption

    spectrophotometers.

    Interference

    Certain oxidizing agents liberate iodine from iodides (positive interference) and some

    reducing agents reduce iodine to iodide (negative interference).

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    18/47

    Modifications of Iodometric

    method azide modification-removes interference caused by nitrite

    the permanganate modification-removes the interference of ferrous iron

    the alum flocculation modification-suspended solids interference

    the copper sulfate-sulfamic acid flocculation modification-activated-sludge

    mixed liquor interference.

    The choice is based on the effect of interferences, particularly oxidizing or

    reducing materials that may be present in the sample.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    19/47

    Azide ModificationReagents

    a. Manganous sulfate solutionb. Alkali-iodide-azide reagent

    c. Sulfuric acid

    d. Starch

    e. Standard sodium thiosulfate titrant

    f. Standard potassium bi-iodate solution

    StandardizationDissolve approximately 2 g KI in 100 to 150 mL distilled water

    few drops of conc H2SO4+ 20.00 mL standard bi-iodate solution

    Dilute to 200 mL

    Titrate the liberated iodine against Na2S2O3Starch indicator, pale straw color end point

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    20/47

    Procedure

    a. To the sample collected in a 250- to 300-mL bottle

    1 ml of MnSO4

    1 mL alkali-iodide-azide reagent

    Stopper carefully to exclude air bubbles and mix by inverting bottle a few times.

    When precipitate has settled sufficiently add 1.0 mL conc H2SO4

    Restopper and mix by inverting several times until dissolution is complete.

    Titrate a volume corresponding to 200mL original sample after correction forsample loss by displacement with reagents. Thus, for a total of 2mL (1 mL each) ofMnSO4 and alkali-iodide-azide reagents in a 300-mL bottle, titrate

    200 300/(300 2) = 201 mL.

    b. Titrate with 0.025M Na2S2O3solution to a pale straw color. Add a few drops ofstarch solution and continue titration to first disappearance of blue color.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    21/47

    Permanganate Modification

    only on samples containing ferrous iron

    ineffective for oxidation of sulfite, thiosulfate, polythionate, or the organic matter

    in wastewater.

    Reagents

    a. Potassium permanganate solution

    b. Potassium oxalate solution

    c. Potassium fluoride solution

    All the reagents required for the azide modification

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    22/47

    Procedure To a sample collected in a 250- to 300-mL bottle

    0.70 mL conc H2SO4+ 1 mL KMnO4solution + 1 mL KF solution

    Stopper and mix by inverting

    Remove permanganate color completely by adding 0.5 to 1.0 mL K2C2O4solution

    Mix well and allow to stand in the dark to facilitate

    the reaction.

    1 ml of MnSO4+ 1 mL alkali-iodide-azide reagent

    Stopper, mix, and let precipitate settle for short time

    Acidify with 2 ml H2SO4

    take 200 300/(300 7.7) = 205 mL for titration

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    23/47

    Alum Flocculation Modification

    The interference due to solidsmay be removed by alum flocculation.

    Reagents

    a. Alum solution

    b. Ammonium hydroxide

    c. Manganous sulfate solution

    d. Alkali-iodide-azide reagent

    e. Sulfuric acid

    f. Starchg. Standard sodium thiosulfate titrant

    All the reagents required for the azide modification

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    24/47

    Procedure

    To a sample collected in a 250- to 300-mL bottle

    10 mL alum solution + 1 to 2 mL conc NH4OH

    Stopper and invert gently for about 1 min settle for about 10 min

    Continue sample treatment as same of azide modification.

    C S lf t S lf i A id

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    25/47

    Copper Sulfate-Sulfamic Acid

    Flocculation Modification Used for biological flocs such as activated sludge mixtures, which have high oxygen

    utilization rates.Reagents

    a. Copper sulfate-sulfamic acid inhibitor solution

    b. All the reagents required for the azide modification

    ProcedureAdd 10 mL CuSO4-NH2SO2OH inhibitor to a 1-L glass-stoppered bottle.

    Insert bottle in a special sampler designed so that bottle fills from a tube near bottomand overflows only 25 to50% of bottle capacity.

    Collect sample, stopper, and mix by inverting

    Let suspended solids settle and siphon relatively clear supernatant liquor into a 250-to 300-mL DO bottle

    Continue as per azide modification

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    26/47

    Membrane Electrode Method

    Being completely submersible, membrane electrodes are suited for analysis in situ.

    Principle

    Oxygen-sensitive membrane electrodes of the polarographic or galvanic type are

    composed of two solid metal electrodes in contact with supporting electrolyte

    separated from the test solution by a selective membrane.

    In polarographic, electrode reaction is spontaneous

    In galvanic types external source of applied voltage is needed to polarize the

    indicator electrode.

    Membranes like Polyethylene and fluorocarbon membranes are normally used asthey are permeable to molecular oxygen and are relatively rugged.

    In the device diffusion current is linearly proportional to the concentration of

    molecular oxygen.

    The current can be converted to concentration units by calibration.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    27/47

    Apparatus

    Oxygen-sensitive membrane electrode, polarographic or galvanic, with appropriate

    meter.

    Procedure

    a. Calibration :calibrate membrane electrodes by reading against air or a

    sample of known DO concentration (determined by iodometric method) as well as in asample with zero DO.

    b. Sample measurement: Follow all precautions recommended by manufacturer to

    insure acceptable results.

    c. Validation of temperature effect: Check frequently one or two points to verify

    temperature correction data

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    28/47

    Calculation Membrane electrodes exhibit a relatively high temperature coefficient largely due

    to changes in the membrane permeability

    The effect of temperature on the electrode sensitivity, (microamperes per

    milligram per liter)

    log = 0.43 mt+ b

    where:

    t= temperature, C,

    m= constant that depends on the membrane material, and

    b= constant that largely depends on membrane thickness.

    sensitivity at any desired temperature ( and t)=

    log = log 0+ 0.43 m(t t0)

    Temperature correction can be found from the nomographic charts

    Electrode sensitivity varies with salt concentration according to the following

    relationship:

    log s= 0.43 mSCs+ log 0

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    29/47

    Calculation

    For each test bottle meeting the 2.0-mg/L minimum DO depletion and the

    1.0-mg/L residual DO, calculate BOD5 as follows:

    When dilution water is not seeded:

    where:

    D1= DO of diluted sample immediately after preparation, mg/L,

    D2= DO of diluted sample after 5 d incubation at 20C, mg/L,

    P= decimal volumetric fraction of sample used

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    30/47

    CalculationBOD (mg/L)= [(D1D2)f(B1B2)]/P

    where:

    D1= DO of diluted sample immediately after preparation, mg/L,D2= DO of diluted sample after 5 d incubation at 20C, mg/L,

    P= decimal volumetric fraction of sample used,

    B1= DO of seed control before incubation, mg/L,

    B2= DO of seed control after incubation mg/L, and

    f = ratio of seed in diluted sample to seed in seed control = (%seed in diluted sample)/(% seed in seed control).

    If seed material is added directly to sample or to seed control

    bottles:

    f = (volume of seed in diluted sample)/(volume of seed in seed control)

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    31/47

    Ultimate BOD Test

    An extension of 5-day BOD test.

    Principle

    Placing a single sample dilution in full, airtight bottles and incubating underspecified conditions for an extended period depending on wastewatereffluent, river, or estuary quality.

    Dissolved oxygen (DO) is measured (with probes) initially and intermittentlyduring the test. From the DO versus time series, UBOD is calculated.

    Temp: 20C

    When DO concentrations approach 2 mg/L, the sample should be reaerated.

    If the result is being used to estimate the rate of oxidation of naturallyoccurring surface waters, addition of nutrients and seed probably acceleratesthe decay rate and produces misleading results.

    If only UBOD is desired, it may be advantageous to add supplemental nutrientsthat accelerate decay and reduce the test duration. When nutrients are used,they also should be used in the dilution water blank.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    32/47

    Apparatus

    Incubation bottles: Glass bottles with ground-glass stoppers 2-L (or larger)

    capacity clean bottles with a detergent and wash with dilute HCl (3N) to

    remove surface films and precipitated inorganic salts; rinse thoroughly

    with DI water before use.

    To prevent drawing air into the sample bottle during incubation, use a

    water seal.

    Place a clean magnetic stirring bar in each bottle to mix contents before

    making DO

    measurement or taking a subsample.

    Reservoir bottle:4-L or larger glass bottle. Close with screw plastic cap ornon-rubber plug.

    Incubator or water bath, thermostatically controlled at 20 1C. Exclude

    all light to prevent the possibility of photosynthetic production of DO.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    33/47

    Procedure: River water samples Fill large BOD bottle with sample at 20C. Add no nutrients, seed, or

    nitrification inhibitor if in-bottle decay rates will be used to estimate in-stream rates.

    Measure DO in each bottle, stopper, and make an airtight seal. Incubate at

    20C in the dark.

    Measure DO in each bottle at intervals of at least 2 to 5 d over a period of

    30 to 60 d(minimum of 6 to 8 readings) or longer under specialcircumstances.

    To avoid oxygen depletion in samples containing NH3-N, measure DO

    more frequently until nitrification has taken place. If DO falls to about 2

    mg/L, reaerate .

    Pour a small amount of sample into a clean vessel and reaerate theremainder directly in the bottle by vigorous shaking or bubbling with

    purified air (medical grade). Refill bottle from the storage reservoir and

    measure DO. This concentration becomes the initial DO for the next

    measurement. If using 300-mL BOD bottles, pour all of the sample from

    the several bottles used into a clean vessel, reaerate, and refill the smallbottles.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    34/47

    Analyze for nitrate plus nitrite nitrogen on Days 0, 5, 10, 15, 20, and 30.Alternatively, determine NO2-N andNO3-N each time DO is determined,thereby producing corresponding BOD and nitrogen determinations. If theultimate demand occurs at a time greater than 30 d, make additionalanalyses at 30-d intervals.

    If the purpose of the UBOD test is to assess the UBOD and not to provide

    data for rate calculations, measure nitrate nitrogen concentration only atDay 0 and on the last day of the test.

    When using a dilution water blank, subtract DO uptake of the blank fromthe total DO consumed.

    High-quality reagent water without nutrients typically will consume amaximum of 1mg DO/L in a 30- to 90-d period. If DO uptake of the dilution

    water is greater than 0.5 mg/L for a 20-d period, or 1 mg/L for a 90-dperiod, report the magnitude of the correction and try to obtain higher-quality dilution water for use with subsequent UBOD tests.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    35/47

    Wastewater treatment plant samples Use high-quality reagent water for dilution water.

    Add no nitrification inhibitors if decay rates are desired.

    If seed and nutrients are necessary, add the same amounts of each to the dilution

    water blank the ultimate BOD of the diluted sample should be in the range of 20 to

    30 mg/L. Dilution to this level probably will require two or three sample reaerations during

    the incubation period to avoid having dissolved oxygen concentrations fall below 2

    mg/L.

    Fill a BOD bottle with dilution water to serve as a dilution water blank

    Use 2-L or larger BOD bottles (alternatively, multiple 300-mL BOD bottles) for each

    dilution. Add desired volume of sample to each bottle and fill with dilution water.

    Treat blank the same as all samples.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    36/47

    Calculations

    UBOD can be estimated by using a first-order model described

    as follows:

    BODt= UBOD(1 e-Kt)

    where:

    BODt= oxygen uptake measured at time t, mg/L, and

    k= first-order oxygen uptake rate.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    37/47

    Respirometric MethodRespirometric methods provide direct measurement of the oxygen consumed by

    microorganisms from an air or oxygen-enriched environment in a closed vessel

    under conditions of constant temperature and agitation.

    Respirometry measures oxygen uptake more or less continuously over time.

    Respirometric methods are useful for assessing

    biodegradation of specific chemicals

    treatability of organic industrial wastes

    the effect of known amounts of toxic compounds on the oxygen-

    uptake reaction of a test wastewater or organic chemical

    the concentration at which a pollutant or a wastewater

    measurably inhibits biological degradation

    the effect of various treatments such as disinfection, nutrient

    addition, and pH adjustment on oxidation rates

    the oxygen requirement for essentially complete oxidation ofbiologically oxidizable matter

    the need for using adapted seed in other biochemical oxygen

    uptake measurements, such as the dilution BOD test

    Respirometric data typically will be used comparatively, that is, in a direct

    comparison between oxygen uptakes from two test samples or from a test sampleand a control.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    38/47

    Types of Respirometers Manometric respirometers relate oxygen uptake to the change in pressure

    caused by oxygen consumption while maintaining a constant volume.

    Volumetric respirometers measure oxygen uptake in incremental changesin gas volume while maintaining a constant pressure at the time of

    reading.

    Electrolytic respirometers monitor the amount of oxygen produced by

    electrolysis of water to maintain a constant oxygen pressure within the

    reaction vessel. Direct-input respirometers deliver oxygen to the sample from a pure

    oxygen supply through metering on demand as detected by minute

    pressure differences.

    Reaction-vessel contents are mixed by using a magnetic or mechanical

    stirring device or by bubbling the gaseous phase within the reaction vesselthrough the liquid phase.

    All respirometers remove carbon dioxide produced during biological

    growth by suspending a concentrated adsorbent (granular or solution)

    within the closed reaction chamber or by recirculating the gas phase

    through an external scrubber.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    39/47

    Interferences Evolution of gases other than CO2may introduce errors in pressure or volume

    measurements; this is uncommon in the presence of dissolved oxygen. Incomplete CO2absorption will introduce errors if appropriate amounts and

    concentrations of alkaline absorbent are not used.

    Temperature fluctuations or inadequate mixing.

    Fluctuations in barometric pressure

    Most commercial respirometers can detect oxygen demand in increments as smallas 0.1 mg.

    Upper limits of oxygen uptake rate are determined by the ability to transfer oxygen

    into the solution from the gas phase, which typically is related to mixing intensity.

    The continuous readout of oxygen consumption in respirometric measurements

    indicates lag, toxicity, or any abnormalities in the biodegradation reaction. The

    change in the normal shape of an oxygen-uptake curve in the first few hours may

    help to identify the effect of toxic or unusual wastes entering a treatment plant in

    time to make operating corrections.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    40/47

    Apparatus

    Respirometer system

    Incubator or water bath

    Reagentsa. Distilled water

    b. Phosphate buffer solution

    c. Ammonium chloride solution

    d. Calcium chloride solutione. Magnesium sulfate solution,

    f. Ferric chloride solution

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    41/47

    Reagents

    Potassium hydroxide solution,6N

    Acid solutions,1N

    Alkali solution,1N

    Nitrification inhibitor: Reagent-grade 2-chloro-6

    (trichloromethyl) pyridine (TCMP) or equivalent. Glucose-glutamic acid solution

    Electrolyte solution

    Sodium sulfite

    Trace element solution Yeast extract solution

    Nutrient solution

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    42/47

    Procedure Instrument operation: Follow respirometer manufacturers instructions for

    assembly, testing, calibration, and operation of the instrument. Sample volume: Small volumes or low concentrations may be required for high-

    strength wastes. Large volumes may be required for low-strength wastes to

    improve accuracy.

    Data recording interval: Set instrument to give data readings at suitable intervals.

    Intervals of 15 min to 6 h typically are used.

    Sample preparation

    HomogenizationIf sample contains large settleable or floatable solids,

    homogenize it with a blender and transfer representative test portions while all

    solids are in suspension. If there is a concern for changing sample characteristics,

    skip this step.

    pH adjustmentNeutralize samples to pH 7.0with H2SO4or NaOH of such strength

    that reagent that reagent quantity does not dilute the sample more than 0.5%.

    DechlorinationAvoid analyzing samples containing residual chlorine by collecting

    the samples ahead of chlorination processes. If residual chlorine is present, aerate

    or let stand in light for 1 to 2 h. If a chlorine residual persists, add Na2SO3solution.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    43/47

    Procedure

    Initial oxygen concentrationIf samples contain dissolved oxygen concentrations

    above or below the desired concentration, agitate or aerate with clean and filtered

    compressed air for about 1 h immediately before testing.

    Temperature adjustmentBring samples and dilution water to desired test

    temperature (1C) before making dilutions or transferring to test vessels.

    Sample dilution: Use distilled water or water from other appropriate sources freeof organic matter for dilution.

    Nutrients, minerals, and buffer : Add sufficient ammonia nitrogen to provide a

    COD:N:P ratio of 100:5:1 or a TOC:N:P ratio of 30:5:1.

    Add 2 mL each of calcium, magnesium, ferric chloride, and trace mineral solutions

    to each liter of diluted sample unless sufficient amounts of these minerals are

    present in the original sample.

    Phosphorus requirements will be met by the phosphate buffer if it is used (1

    mL/50 mg/L COD or ultimate BOD of diluted sample usually is sufficient to

    maintain pH between 6.8 and 7.2).

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    44/47

    Nitrification inhibition: If nitrification inhibition is desired, add 10 mg 2-

    chloro-6-(trichloromethyl) pyridine (TCMP)/L sample in the test vessel.

    Seeding: Use sufficient amounts of seed culture to prevent major lags in

    the oxygen uptake reaction but not so much that the oxygen uptake of the

    seed exceeds about 10% of the oxygen uptake of the seeded sample.

    Determine the oxygen uptake of the seeding material as for any other sample.

    This is the seed control.

    Incubation: Incubate samples at 20C or other suitable temperature 1.0C.

  • 8/11/2019 Biological Oxygen Demand Test by APHA

    45/47

    Calculations To convert instrument readings to oxygen uptake, refer to

    manufacturers procedures.

    Correct oxygen uptake for seed and dilution by the following equation:

    C= *A B(SA/SB)](1000/NA)

    where:

    C= corrected oxygen uptake of sample, mg/L,

    A= measured oxygen uptake in seeded sample, mg,

    B= measured oxygen uptake in seed control, mg,