Bioelectric Fat Analyzer
Presented by
P.Sandeep Reddy
B.Santhosh Naik
K.Venkata Rajesh
Guide: Mr.Vennala Kishore Assistant Professor
(Dept of ECE) BVRIT
![Page 2: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/2.jpg)
Presentation Outline
Introduction Hardware used Flow Chart Block diagram Working Existing methods Implementation and Result Future Scope Conclusion
![Page 3: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/3.jpg)
Introduction
• Weight alone is not a clear indicator of good health because it does not distinguish fat and body mass.
• Obesity means an excess of body fat regardless of weight.
![Page 4: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/4.jpg)
Principle
Bioelectric Impedance Analysis
![Page 5: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/5.jpg)
Hardware Procured
• AT89C52• ADC0809• 16*2 Character LCD• Step Down Transformer (Centre Tapped)• External Clock Generator• 12 MHz Crystal Oscillator • Power Supply Circuit• Keypad 4*3• Copper Electrodes
![Page 6: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/6.jpg)
F Flow ChartStart
Accept Inputs (Age, weight)
Adjust Body Peripheral Length
Set electrodes
Apply constant current electrodes
Calculate fat free mass
Calculate body fat by substituting weight , age ,
voltage in regression equation
Set body length by replacing each pair of
electrodes
Find the potential Difference
Calculate body percentage
Display result on LCD Screen
EndNo
Yes
![Page 7: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/7.jpg)
Block Diagram
![Page 8: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/8.jpg)
Working
The measurement is taken when the electrodes are placed at the wrist and the contra lateral ankle.
By injecting an AC current of the order 10uA through one’s body and measuring the voltage across the electrodes and finding instantaneous impedance.
![Page 9: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/9.jpg)
Enter the Age and Weight of the user. Percentage of the Body’s Fat is Calculated by
regression equations.
Males: Bodyfat = 0.0923*weight + 0.1605*age - 0.0263*voltageFemales: Bodyfat = 0.1871*weight + 0.5800*age – 0.0920*voltage
![Page 10: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/10.jpg)
Existing Methods
DEXA (Dual-energy X-ray Absorptiometry) Scan
Underwater Weighing Bod Pod (Whole Body Plethysmography) Skinfold Measurements
![Page 11: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/11.jpg)
Implementation and Result
Step 1: Switch ON the Power Supply Circuit
![Page 12: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/12.jpg)
Step 2: The message “Hold Electrodes” on the 16*2 Character LCD gets displayed.
![Page 13: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/13.jpg)
Step 3:: On seeing the message “Hold Electrodes” on the LCD display, hold it with two figures of your hand as shown in below picture
![Page 14: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/14.jpg)
Step 4: Enter the weight and age of the user using 4*3 keypad present in the Kit and press Go key. After a message “Please enter 1 for MALE or 2 for FEMALE gets displayed” on the display, Enter 1 if the user is 1 or enter 2 if female and press Go key.
![Page 15: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/15.jpg)
Step 5: Fat percentage of the User gets displayed on the LCD screen.
![Page 16: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/16.jpg)
Advantages
Measuring body’s fat with this method is much easier than compared to any other method.
Can be used in any environmental conditions.
Much of reasonable cost.
![Page 17: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/17.jpg)
Future Scope
One extension would be to improve accuracy by implementing multi-frequency bioimpedance analysis.
Another possible extension would be miniaturize the device and making it portable.
![Page 18: Bioelectric fat analyser](https://reader035.vdocuments.us/reader035/viewer/2022081420/55b9a819bb61ebc5228b475c/html5/thumbnails/18.jpg)
Conclusion
For our accuracy goal, we were able to achieve body fat prediction within a 10% error margin for subjects matching our test samples, but we had poor accuracy outside that fairly specific range of weights and body compositions
The device operated at a safe current and did not harm any of our test subjects. Our input signal to the user is approximately 12 μA, which is quite close to our goal of 10 μA, and is still safe to pass through a user’s body mass.