bio-columns products, applications, and principles … · bio-columns products, applications, and...
TRANSCRIPT
Bio-Columns Products,
Applications, and Principles
Paul Dinsmoor
BioHPLC Columns Specialist
Sep. 27, 2012
BioChromatography
2012 EMEA Seminar Series
2
Size Exclusion
BioHPLC Columns
Define the Separation Goals
For SEC, what is the molecular weight range of the proteins you
wish to separate?
The answer to this questions helps you determine the
appropriate pore size to achieve the best resolution of your
protein.
Pore volume is a key factor in separation.
Size Exclusion Process
13-17 February 2012
Bio-applications Training Program
5
Larger molecules spend less time
in the pores and elute sooner.
Smaller molecules spend longer
in the pores and elute later.
Mechanism of SEC Separation – Pore Size
Determines Linear Separation Range
Agilent Bio SEC-5 Column Characteristics – Pore Size/ Separation Range
Some General Guidelines for SEC
1. When methods are to be validated, test for ruggedness
with several different column lots, mobile phase
preparations, and operators.
2. As a rule of thumb, SEC will only provide baseline
separation of molecules with more than a 2 fold difference
in MW.
3. Sample volume should be limited to below 5% of the total
column volume. Max resolution .5 – 2% CV.
Loaded Sample Volume / Resolution
AU
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
1.60
1.80
2.00
Min
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0
Injection volume: 10 L
BSA Concentration
(from low to high):
1 mg/mL
2 mg/mL
5 mg/mL
10 mg/mL
20 mg/mL
50 mg/mL
Column: Bio SEC-3, 150Å, 7.8x300mm
Mobile phase: 150 mM Phosphate, pH 7.0
Flow Rate: 1.0 mL/min
Injection volume: 10 µL
Detector: 214 nm
Column Selection Criteria
Pore size
depends on molecular weight range of sample
avoid exclusion of sample components
maximize pore volume in required separation region
Particle size
smaller particles for higher resolution
Number of columns/ length
compromise between resolution and analysis time
Column ID
smaller column ID can be used for reduced solvent
consumption and smaller injection volume
What does Exclusion look like ?
Exclusion: very sharp
peak with almost
vertical take-off !
Normal: somewhat
broad, generally
gaussian in shape
11
Pore Size Comparison
September 30, 2012
Agilent Confidential
12
Eluent: 50mM NaH2PO4 + 0.15M NaCl, pH6.8
Columns: Agilent Bio SEC-3 ,4.6x300mm
Flow: 0.35ml/min
Detector: UV@220nm
System: Agilent 1260 Infinity Bio-Inert LC System
Sample: BioRad Gel Filtration Standards Mix
Mouse IgG
1. Thyroglobulin Aggregates
2. Thyroglobulin
3. IgA
4. g-globulin
5. Ovalbumin
6. Myoglobin
7. Vitamin B12
300Å
150Å
100Å
1
2
3
4
5 6
7
300Å
150Å
100Å
1
2
3
4
5
1. Dimer
2. Monomer
3. Monomer Fragment
4. Azide
5. Retained Molecule
Peptides Proteins Globular
proteins
Column Choice: Resolving Ranges
100D 1kD 10kD 100kD 1,000kD 10,000kD
Agilent Bio SEC 100Å
Agilent Bio SEC 150Å
Agilent Bio SEC 300Å
ProSEC 300S
Zorbax GF-250
Th
yro
glo
bu
lin
IgG
Ova
lbu
min
B
SA
Myo
glo
bin
Zorbax GF-450
Agilent Bio SEC 500Å
Agilent Bio SEC 1000Å
Agilent Bio SEC 2000Å
Insu
lin
0 2 4 6 8 10 12 14
Time (min)
020
040
0
mAu
214
nm
0 2 4 6 8 10 12 14
Time (min)
020
040
0
mAU
214
nm
Agilent Bio SEC-5
150 Å
TSK
G2000SWxl 1. thyroglobulin, 5.64 min
2. BSA dimer, 6.23 min
3. BSA monomer, 7.02 min
4. ribonuclease A, 9.22 min
5. poly-DL-alanine 10.02 min
6. uracil, 11.81 min
1. thyroglobulin, 5.43 min
2. BSA dimer, 6.19 min
3. BSA monomer, 6.93 min
4. ribonuclease A, 8.74 min
5. poly-DL-alanine 9.90 min
6. uracil, 12.13 min
Resolution is affect by pore volume
14
Effects of Particle Size on Resolution
Decreasing the Particle Size of the GPC media in a column improves
the resolution
The smaller the particle the more particles that can be packed into a
column. Therefore the number of pores increases
The smaller the particles, the faster equilibrium is reached in the
column between the stationary phase (the pore volume) and the
mobile phase (interstitial volume)
Improved Efficiency With Smaller Particles
Column: Bio SEC-3 300Å and Bio SEC-5 300Å
Buffer: 150 mM Phosphate buffer, pH 7
Flow rate: 1.0 mL/min for 7.8x300 mm
Temperature: Ambient (~23° C)
Detection: UV 214nm
Injection: 10 µL (3 L for 4.6x300 mm)
Sample: 1) Thyroglobulin (1.0 mg/mL), 670 kD; 2) BSA
dimer, 132 kD; 3) BSA (1.0 mg/mL), 66 kD; 4)
Ribonuclease A (1.0 mg/mL), 13.7 kD, and 5) Uracil
(2.5 g/mL), 120D.
Min 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Agilent Bio SEC-3, 300Å,
7.8x300mm
Agilent Bio SEC-5, 300Å,
7.8x300mm
Column Pressure
Bio SEC-3
(7.8x300mm) 93 bar
Bio SEC-5
(7.8x300mm) 45 bar
Peak Protein SEC-3, 300Å
(7.8x300mm)
SEC-5, 300Å
(7.8x300mm)
1 Thyroglobulin 2460 1120
2 BSA Dimer 5100 2720
3 BSA 13090 6590
4 Ribonuclease A 22000 11160
5 Uracil 38500 27860
Determine Optimal Mobile Phase Conditions
What conditions give you the best results?
What additives may be required to reduce non-specific
interactions?
What is the optimal flow rate?
1. 150mM Sodium phosphate buffer
2. 150mM Sodium phosphate buffer + 50mM NaCl
3. 150mM Sodium phosphate buffer + 100mM NaCl
4. 150mM Sodium phosphate buffer + 300mM NaCl
Agilent Bio SEC-3, 300A 7.8 x 300mm MAb separation, Sodium phosphate buffer and different NaCl concentrations
Size Exclusion Chromatography
2012 EMEA Seminar Series
19
• Impurity testing, aggregation analysis (separation of monomer/dimer/aggregates)
• Molecular weight characterization – over wide MW range (1000 – 10M) possible
• Expression and folding studies
• Separation of Reaction Components and Products, (conjugates)
• Purification
• Desalting and salt exchange
• Collection of fractions under non-denaturing conditions
Non-Denaturing Conditions
Separations Based on Size in Solution
NEW Size Exclusion Columns
• 5m Particle
• 100Å, 150Å, 300Å, 500Å, 1000Å,
2000Å pore sizes
• High stability and long lifetime
• Great reproducibility
20
• Unique, 3m particle
• 100Å, 150Å, 300Å pore sizes
• Highest resolution
• Highest efficiency
• Faster SEC separations
Agilent Biopharma Training, AFO NPT 2010
Aggregation Studies
2012 EMEA Seminar Series
21
ICH (Q6B):
• “The category of aggregates include dimers and higher multiples of
the desired product. These are generally resolved from the desired
product and product-related substances, and quantitated by appropriate
analytical procedures (e.g., size exclusion chromatography, capillary
electrophoresis)” (ICH Q6B)
• “Multimers and aggregates should also be appropriately characterized
using a combination of methods. The formation of aggregates, sub-visible
and visible particulates in the drug product is important and should be
investigated and closely monitored on batch release and during stability
studies.
Agilent Bio SEC-5 Monoclonal Antibody Aggregation Monitoring
Columns: Bio SEC-5, 300Å, 7.8x300mm
Buffer: 150 mM sodium phosphate, pH 7
Flow Rate: 1.0 mL/min
Sample: Mab (2.5 mg/mL)
Injection: 10 L
Concentration: 10, 2.5, 0.5 mg/mL
(from high to low)
Temperature: Ambient
AU
-0.04
-0.02
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
Minutes
1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00
Dimer
Monomer
Bio SEC-5 columns can easily handle higher flow rates, this separation
was done at 1 mL/min, resolution between dimer and monomer can be
improved by decreasing the flow rate
Agilent Bio SEC-5 Monoclonal Antibody Aggregation Monitoring
MAb Dimer
Mab Aggregates
Columns: Bio SEC-5, 300Å, 7.8x300mm
Buffer: 150 mM Phosphate, pH 7
Flow Rate: 1.0 mL/min
Sample: Mab
Temperature: Ambient
AU
-0.008
-0.006
-0.004
-0.002
0.000
0.002
0.004
0.006
0.008
0.010
0.012
0.014
0.016
0.018
0.020
Min 1.0 2.00 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0
Buffer Peak
Low molecular
weight forms
Monitoring Aggregation and Impurities
2012 EMEA Seminar Series
24
AU
-0.005
0.000
0.005
0.010
0.015
0.020
0.025
0.030
0.035
Minutes
1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00
Column: Bio SEC-3 300Å, 7.8x300mm
Mobile phase: 150 mM Phosphate, pH 7
Flow rate: 1.0 mL/min
Temperature: Ambient
Sample: Monoclonal antibody (10 L, 5 mg/mL)
Buffer/Excipients
MAb
Dimer MAb
Aggregates
Low Molecular
Weight Impurities
Phosphate buffer to control pH and provide ionic strength
Optimized to eliminate interactions – for this MAb
Agilent Bio SEC-5 High Molecular Weight Proteins
Proteins:
1.Thyroglobulin (670kD)
2. Uracil (120D)
Column: Bio SEC-5, 1000Å, 4.6x300mm
Buffer:150 mM phosphate buffer, pH 7.0
Flow rate: 0.35 mL/min
Detection: 214 nm
Temperature: Ambient
1
Possible Dimer
AU
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0.14
0.16
0.18
0.20
0.22
Min
0 2 4 6 8 10 12 14
2
Min 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Columns: Bio SEC-5, 7.8x300mm
Buffer: 150 mM Phosphate, pH 7.0
Flow Rate: 1.0 mL/min
Detector: 214 nm
Injection: 10 µL (2.5 mg/mL)
SEC-5, 1000Å
SEC-5, 500Å
SEC-5, 300Å
SEC-5, 150Å
Agilent Bio SEC-5 Size Exclusion Separations of E. coli Lysate
Columns: Bio SEC-5, 1000Å, 7.8x300mm + Bio SEC-5, 150Å, 7.8x300mm
Buffer: 150 mM Phosphate, pH 7.0
Flow Rate: 1.0 mL/min
Detector: 214 nm
Injection: 10 µL (2.5 mg/mL)
AU
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
Minutes
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Agilent Bio SEC-5 Size Exclusion Separations of E. coli Lysate
(Two Columns In Series)
Column: Column: Bio SEC-3 150Å, 7.8x300mm & Bio SEC-3 300Å, 7.8x300mm
Mobile phase: 0.15 M Phosphate, pH 7.0
Flow rate: 1.0 mL/min
Detection: UV 214 nm
Temperature: Ambient
Injection: 10 µL
Sample: E. coli lysate (2.5 mg/mL)
Min 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0
Bio SEC-3 150Å, 7.8x300mm
Bio SEC-3 300Å, 7.8x300mm
Agilent Bio SEC-5
Separations of E. coli Lysate
AU
0.00
0.10
0.20
0.30
0.40
0.50
0.60
0.70
0.80
0.90
1.00
1.10
1.20
Min 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0
3
1
2
Column: Bio SEC-3, 150Å (3µm particle, 7.8x300mm)
Sample: 1, Protein with MW 70KD (from an undisclosed drug company)
2, Protein 1 attached with 1 PEG (MW 40KD) molecule
3, Protein 1 attached with 2 PEG (MW 40KD) molecules
Mobile phase: 150 mM Phosphate pH=7.0; Flow rate: 1.0 mL/min
Wavelength: 214 nm; Temperature: Ambient
Sample concentration: 2.0 mg/mL; Injection volume: 20 µL
Column backpressure: 110 bar
Size Exclusion Separation of Protein and
PEGylated Proteins
30
Fast separation of dimer from monomer
using the Agilent Bio SEC-3 300Å
7.8 x 150 mm column
31
Column: Agilent Bio SEC-3, 7.8 x150mm
Sample: mAb (2mg/ml)
Injection: 5ul
Flow rate: 1.0, 1.5 and 2ml/min (56 bar , 75 bar, 105 bar)
Eluent: 150mM sodium phosphate
Detection: 220nm
Flow Rate Resolution
Monomer/Dimer
Monomer
Efficiency
Percentage
Dimer
1.0ml/min 1.53 3,510 0.64
1.5ml/min 1.43 2,502 0. 47
2.0ml/min 1.13 1,917 0.64
monomer
dimer
2.0 1.5 1.0
ml/m ml/min ml/min
Fast separation of dimer from monomer using the
Agilent Bio SEC-3 300 Å 7.8 x 150 mm column
Fast SEC – TB5990-8613EN
Calibration-free Molecular Weight Determination ? Agilent 1260 Infinity Multi Detector Suite for Protein SEC
September 30, 2012 32
Helps to differentiate aggregates of monoclonal antibodies
Gamma Globulin (MW 150kD) Agilent Bio SEC-5 500Å, 7.8 x 300 mm
2012 EMEA Seminar Series
33
Monomer
Dimer High MW
aggregates
RI imbalance
(fresh eluent) UV, 254nm
LS 90
RI
Ion Exchange
BioHPLC Columns
Ion Exchange Chromatography Separation based on charge
Ion Exchange Chromatography of Proteins
Ion-exchange chromatography (IEC) discriminates between proteins on the basis of accessible surface charges and their corresponding electrostatic interaction with the column’s stationary phase.
The degree of protein retention is dependent on the strength and number of interactions.
The 3-D structure of the protein determines which surface residues will be available to contact the column’s stationary phase.
Secondary interactions/ non-specific interactions (particularly hydrophobic interactions) can also impact chromatography
Some Guidelines for IEX
1.The General Rule for choosing a Bio IEX column
– Acidic proteins: SAX or WAX
– Basic proteins: SCX or WCX
2. Consider the isoelectric point (pI) of your protein when choosing the pH of your mobile phase:
– if pH>pI, your protein will have a net negative charge
– If pH<pI, your protein will have a net positive charge
3. The pH of your starting buffer should be 0.5 to 1 pH unit
– above pI for Anion Exchange
– below pI for Cation Exchange
Proteins have a net charge, dependant on pH
+
+
+
+
+ -
-
-
-
-
+
+
+ +
At their isoelectric point, pI, the net charge is zero and the molecule will not be
retained on an ion-exchange column.
If pH < pI, the net charge will be positive and the molecule can be analyzed by
cation-exchange. If pH > pI, the net charge will be negative and the molecule can
be analyzed by anion-exchange.
13-17 February 2012
Bio-applications Training Program
38
Protein Isoelectric Point - Charge
2012 EMEA Seminar Series
39
Pro
tein
Ne
t C
ha
rge
+
_
pH 4 6 8 10 12
Iso
ele
ctirc
Po
int (p
I)
+
- - - - - -
- - - - - - - -
- -
Cation
Exchan
ge
Particle
pH lower than
the protein pI
pH greater than
the protein pI
I
Basics of Bio-Chromatography Ion Exchange Chromatography
• Sample is injected in a mobile phase buffer with a low
salt – this binds proteins to the column
• Analytes are typically eluted at constant pH with
increasing salt (ionic strength) to displace proteins
from stationary phase
• Higher charge proteins bind more strongly, more salt is
needed to elute them
• A typical mobile phase will contain NaCl
40
Ion Exchange Chromatography How it works
Bio IEX Columns Target Applications
• Antibodies
• Peptides in volatile buffers
• Glycans (glycosylated bio-molecules, not sugar analysis)
• Proteins
• Polynucleotides
• Cell lysates
• 2D separations
Ion Exchange Columns
• Non-porous PS/DVB particles
• Uniform polymeric coating and WCX
layer, specifically designed for antibody
separations
• Available in 10 µm, 5 µm, 3 µm, 1.7 µm
particle sizes
Agilent Bio MAb Column Characteristics
44
Charge Variant Analysis
2012 EMEA Seminar Series
45
Light Chain
Fc
Fab
Antigen
binding
Hinge
Glycosylation
site Truncation
(lysine)
Pyroglutamate Heavy Chain
Disulfide
shuffling
Deamidation/ Oxidation
Charge Variants (charge heterogeneity) are translational modifications
Sites where charge variants can occur
Ion Exchange Chromatography Charge Isoform Analysis of Monoclonal Antibodies
AU
0.00
0.05
0.10
0.15
0.20
0.25
0.30
0.35
0.40
0.45
0.50
0.55
0.60
Minutes
0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 32.00 34.00
Column: Agilent Bio MAb, NP5, 4.6mm x 250mm
Buffer A: 10 mM Sodium Phosphate, pH 7.50
Buffer B: A + 100 mM NaCl, pH 7.50
Gradient: 15-95% B in 60 min
Flow rate: 0.8 mL/min.
Sample: 5 L, 5 mg/mL, mAb
(Acidic Isoforms)
weakly bound
(Basic Isoforms)
Strongly bound
Min 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
A
B
C
D
E
Optimization of method conditions for the isoform characterization of a
monoclonal antibody. Changes in the buffer conditions, pH and gradient
conditions sharpen peaks and increase resolution of acidic and basic isoforms.
Ion Exchange (Method) Charge Isoform Analysis of Monoclonal Antibodies
Columns: Agilent Bio MAb, NP10, 4.6x250 mm
Mobile phase: A, 10 mM phosphate, pH 7.5
B, A + 0.1M NaCl
Gradient: A) 15-75%B in 30 min
B) 15-65%B in 30 min
C) 15-55%B in 30 min
D) 15-47.5%B in 30 min
E) 15-40%B in 30 min
Flow rate: 0.8 mL/min
Sample: Monoclonal Antibody
Injection: 10 L (1.5 mg/mL)
Temperature: 25 oC
Detection: UV 214 nm
NEW Ion Exchange Columns
• Non-porous PS/DVB particles
(polystyrene divinylbenzene)
• Uniform polymeric coating with SCX,
WCX, SAX, WAX layers, designed for
protein and peptide separations
• Available in 10 µm, 5 µm, 3 µm, 1.7 µm
particle sizes
• High surface area
• High capacity
Agilent Bio IEX Columns Comparing Separations on Each Particle Size
Min
0 2 4 6 8 10 12 14 16 18 20 22
1.7 m
3 m
5 m
10 m
Column: Bio WCX-NP, 4.6x50mm
Buffer A: 20 mM PBS
Buffer B: A+1.0 M NaCl
Gradient: 0-100%B (20 min)
Flow rate: 1.0 mL/min for NP10, NP5, NP3
0.75 mL/min for NP1.7
Sample
1) Ribonuclease A
2) Cytochrome C
3) Lysozyme
Concentration: 1.0 mg/mL
Detector: 280 nm
Average N ~80,000 for WCX-NP1.7
Peak N
Peak N
Agilent Bio IEX High Resolution Protein Separations
min 0 2 4 6 8 10 12 14
BSA
Ovalbumin Column : Bio SAX, NP3 , 4.6x50mm
Buffer: 20 mM Tris, pH 8.0
Gradient: 0-0.3 M NaCl (15 min)
Flow Rate: 0.5 mL/min
Backpressure: 110 bar
Detector: 214nm
Agilent Bio IEX Columns Peptide Separations
ELSD
@280nm
C1 C2
C3
C4
Column: Bio SCX, NP3, 4.6x50mm
Buffer A: 5 mM CH3COONH4
Buffer B: 0.5 M CH3COONH4:ACN=4:1 (v/v)
Gradient: 0-50% B (20 min)
Flow rate: 0.6 mL/min
Detector: ELSD & UV 280 nm
Injection volume: 5 µL (0.1 mg/mL)
Peptide Sequence Net Charge
C1 Ac-Gly-Gly-Gly-Leu-Gly-Gly-Ala-Gly-Gly-Leu-Lys-amide + 1
C2 Ac-Lys-Tyr-Gly-Leu-Gly-Gly-Ala-Gly-Gly-Leu-Lys-amide + 2
C3 Ac-Gly-Gly-Ala-Leu-Lys-Ala-Leu-Lys-Gly-Leu-Lys-amide + 3
C4 Ac-Lys-Tyr-Ala-Leu-Lys-Ala-Leu-Lys-Gly-Leu-Lys-amide + 4
Agilent Bio IEX Columns High Resolution Glycan Separations
min 0 10 20 30 40 50 60 70
LU
0
100
200
300
400
500
G0F G1F
G2F SA1F (NGNA)
SA2F (NGNA)
2-AA (anthranilic acid) labeled N-linked oligosaccharide profiling of an IgG1 sample
G0F: asialo, agalacto, core-fucosylated biantennary glycan
G1F: asialo, mono-galacto, core-fucosylated biantennary glycan
G2F: asialo, di-galacto, core-fucosylated biantennary glycan
SA1F: mono-sialylated, galactosylated, core-fucosylated biantennary glycan
SA2F: di-sialylated, galactosylated, core-fucolylated biantennary glycan
NGNA: N-glycolylneuraminic acid
Column: Bio SAX, NP5, 4.6x150mm
Buffer A: 2.5% (v/v) acetic acid, 0.5% TEA in H2O
Buffer B: 0.5% acetic acid in ACN
Gradient: 0-100%B (60 min)
Flow Rate: 0.3 mL/min
Fluorescence detector: Ex/Em=360/425nm,
2-AA labeled N-linked oligosaccharide profiling of an IgG1 sample
min 20 30 40 50 60
LU
0
100
200
300
400
Isomer of G1F Isomers of SA1F (NGNA)
G0F G1F
G2F SA2F
(NGNA)
min 20 30 40 50 60 70 80
0
50
100
150
200
250
No Isomer of G1F
Separated Isomers of SA1F (NGNA)
Not Separated
G0F G1F
G2F SA2F
(NGNA)
Bio SAX, NP5, 2.1x150mm
Asahipak Amino NH2P-50 2D,
5 µm, 2.1x150 mm
Agilent Bio SAX Comparison to Column Commonly Used For Glycans
min 20 22.5 25 27.5 30 32.5 35 37.5 40
Bio SAX, NP5
Competitor SAX Column
Sample: 50-mer Oligo
Buffer A: 0.025 M TRIS pH9
Buffer B: A+1.0 M LiCl
Gradient: 30 to 75%B in 45min
Flow Rate: 1mL/min
System: Agilent 1100
Injection: 50 μL
Detection: UV260nm
Agilent Bio IEX Columns Oligonucleotides, Comparison of Non-Porous vs. Porous
min 0 5 10 15 20 25 30 35 40
mAU
0 5
10 15 20 25
Bio SAX, NP5
min 0 5 10 15 20 25 30 35 40
mAU
0 2.5
5 7.5 10
12.5 15
17.5 Competitor SAX Column
Min 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34
Bio SAX, NP10
4.6x50 mm
Peak Capacity > 40
Bio SAX, NP5
4.6x50 mm
Peak Capacity > 60
Bio SAX, NP3
4.6x50 mm
Peak Capacity > 70
Buffer A: 20 mM Tris, pH 9.0
Buffer B: A + 0.5 M NaCl
Gradient: 0-100%B (30 min)
Flow Rate: 0.5 mL/min
Injection: 10 µL/min (2.5 mg/mL)
Agilent Bio IEX Columns E. Coli Lysate Separation:
Particle Size Effect on Resolution/Efficiency
Confidentiality Labelwww.sepax-tech.com
10 11 12 13 14 15 16 17 18 19 20 21 22
Bio SAX, NP10
Bio SAX, NP5
Bio SAX, NP3
Agilent Bio IEX Column E. Coli Lysate Separation:
Particle Size Effect on Peak Capacities
Agilent Bio IEX Columns E. Coli Lysate Separation:
Particle Size Effect on Peak Capacities
Min 0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34
Bio SAX, NP3, 4.6x50mm
E. coli Lysate (2.5 mg/mL)
Min 18 19 20 21
Injection
50 µL
20 µL
10 µL
The Agilent 1260 Infinity Bio-inert Quaternary LC
2012 EMEA Seminar Series
58
100% Bio-inert Precious sample does not touch metal surfaces
pH range 1-13 (shortterm 14)
2 M salt, 8 M urea
No stainless steel inmobile phase flow path
New capillary technology
UHPLC capability 600 bar
Superior Ease of Use and Robustness Highly corrosion resistant
Active seal wash
Quaternary buffer mixing
Agilent proven quality
Agilent warranty and service quality
Superior Bio-HPLC columns for biotherapeutic
characterization
The choice for both, bioanalytical and biopurification up to 10 ml/min
Metal Free
Sample Flow Path
New Capillary Design
Metal cladded PEEK capillary
Newest fitting design: Hybrid technique
New capillary technolgy enables
600 bar AND is completely
metal-free!!
Buffer Advisor Software Quaternary Buffer Mixing with Buffer Advisor and 1260
Infinity Bio-inert LC System
September 30, 2012
Introduction to Agilent Buffer Advisor
60
• Recommends buffer stock solutions
• Shows recipes for stock solutions
• Calculates optimized gradient steps
• Enables automation and unified methods throughout a company
Instead of setting up different
bottles with different pH…
…you use a quaternary Bio-
inert LC with Buffer
Advisor software to
instantly run salt- and pH
gradients
…Saving Time and Resources….
September 30, 2012
Introduction to Agilent Buffer Advisor
61
Agilent Bio WCX: Data Analysis
2012 EMEA Seminar Series
62
Step 1
• pH 7.0 to 5.0
• In 0.2 unit increments
Step 2
• Buffer concentration
• 10, 15, 20, 25 mM
Step 3
• Gradient slope
• Rate of [salt] increase
Robust IEX Method
11 pH values
4 concentrations
at each pH
3 gradients
For each chromatogram calculate resolution between peaks 1 and 2
1
2
Reverse Phase
BioHPLC Columns
Characterization of Primary Structure
2012 EMEA Seminar Series
64
Levels of Characterization
Improve
accuracy
and
resolution
Intact MAb
Reduction/Alkylation
Light and Heavy Chain
Enzymatic Digestion
Peptide Map
Basics of Bio-Chromatography Strategy for RP-HPLC Method Development of
Proteins and Peptides
• Start at low pH (acidic mobile phase) and choose the initial column and conditions
• Initial selection parameters include: pore size (300Å), mobile phase, bonded phase,
particle size, column length and internal diameter
• Obtain best resolution by optimizing:
• Gradient steepness, bonded phase, temperature, column configuration
• Obtain best recovery by optimizing:
• Bonded phase, temperature, sample solubility
• Evaluate alternate columns/technologies for improved selectivity and efficiency
• Try ZORBAX 300Extend-C18 at higher pH for different selectivity; LC/MS
• Try Poroshell 300SB for improved efficiency and shorter analysis times; LC/MS
February 2010
Agilent Confidential 65
300Å Reverse Phase Columns Reverse Phase of Proteins
Requirements
• Wide pore - 300Å for unrestricted
access to bonded phase
• Multiple bonded phases for
selectivity optimization
• Bonded phases with different pH
ranges for LC/MS and method
confirmation (WCBP)
• Many configurations for LC/MS
compatibility, small sample sizes
and proteomics
Columns available
• 300StableBond – C18, C8, C3, CN
• Poroshell 300SB – C18, C8, C3
• 300 Extend-C18
• Configurations from nano (0.075
mm i.d.) to prep (21.2 mm i.d.)
Basics of Bio-Chromatography Strategy for RP-HPLC Method Development of
Proteins and Peptides
• Initial selection parameters include:
- pore size (300Å), mobile phase, bonded phase, particle size, column length and ID
• Obtain best resolution by optimizing:
- Gradient steepness, bonded phase, temperature, column configuration
• Obtain best recovery by optimizing:
- Bonded phase, temperature, sample solubility
• Evaluate alternate columns/technologies for improved selectivity and efficiency
- Try ZORBAX 300Extend-C18 at higher pH for different selectivity; LC/MS
- Try Poroshell 300SB for improved efficiency and shorter analysis times; LC/MS
68
StableBond 300Å
• Four different bonded-phases,
300SB-C18, C8, CN, and C3 for
selectivity optimization
• Extremely stable at low pH
• Use with TFA, Formate and
Acetate mobile phases
• Stable at high temperature – up
to 80 - 90°C
• Particle Sizes: 3.5, 5, and 7
O
R RSi
OH
SiO
SiR R R
OOH
R
R1 R1 R1
300Å Reverse Phase Columns ZORBAX 300SB
69
• Uses unique bidentate-C18 bonded phase for long lifetime at high pH
• Double endcapped
• Can also be used at low pH
• Ammonium hydroxide mobile phase good for high pH LC and LC/MS
Si
O
Si
O
Silica Support
C18C18
ZORBAX 300Extend-C18 Columns for the Analysis of Proteins and Peptides at High pH
Comparison of Small Peptide Selectivity
Differences on 300SB Bonded Phases
Conditions:
Columns: ZORBAX 300SB, 4.6 x 150 mm, 5 mm
Mobile Phase: Gradient, 0 - 26% B in 30min.
A = 0.1% TFA in Water
B = 0.1% TFA in Acetonitrile
Temperature: 40°C
Sample: 2 µg of each peptide
Flow Rate: 1.0 mL / min.
Detection: UV-210nm
300SB-C18
300SB-C8
300SB-C3
300SB-CN
NEW! Zorbax 300 SB RRHD for Proteins and Peptides!
•Stablebond 300 silica
•C-18, C-8, C-3, and diphenyl bonded phase
•1.8 um particle size
•1200 Bar pressure limit for uHPLC
•Other phases to follow
•2.1 x 50 mm and 2.1 x 100 mm
Four Rapid Resolution High Definition Phases
Reversed-Phase RRHD
72
C18 C8 C3 Diphenyl
Peptide Maps Increasing protein size/hydrophobicity
Small Proteins
Ligand Application
C18 Small intact proteins/peptide maps
C8 Intact proteins
C3 Larger /hydrophobic proteins, including MAbs
Diphenyl Unique selectivity
ZORBAX
300Å, 1.8 µm
Agilent 1290 Infinity
Family of Four Reversed-Phase Ligands
Peptide Mapping
Reversed-Phase RRHD
73
min 0 2.5 5 7.5 10 12.5 15 17.5
mAU
0
10
20
30
40
50
60
min 0 2 4 6 8 10 12 14 16 18
mAU
0
5
10
15
20
25
30
35
TFA
formic acid
5990-8244EN
Monoclonal Antibody Tryptic Digest
ZORBAX RRHD 300SB-C18
Fast Separation of Reduced and Alkylated MAb
Reversed-Phase RRHD
74
min 1 2 3 4 5
mAU
0
20
40
60
80
100
120
0.4
54
0
.50
1
0.5
69
0.7
66
3.8
97
4.0
10
Light chain
Heavy chain 1 Heavy chain 2
Columns: ZORBAX RRHD 300SB-C3, 2.1 x 100 mm, 1.8 µm
Sample: Reduced MAb (IgG1) (1.0 mg/mL)- BioCreative IgG1
Sample injection: 2 µL
Mobile phase A: 0.1% TFA in water
Mobile phase B: 80% n-propyl alcohol, 10% ACN, 9.9% water and 0.1% TFA
Temperature: 74 °C
Flow rate: 0.5 mL/min
Detection: UV, 280
Conditions optimized for the ZORBAX RRHD 300SB-C3 Column
CHO Cell Derived MAb
5990-9667EN
Time
(min)
%B
0 1
2 20
5 50
6 90
6.1 1
Comparison C3 and Diphenyl Phases
Reversed-Phase RRHD
75
min 1 1.2 1.4 1.6 1.8 2 2.2 2.4
mAU
2
4
6
8
10
12
1.3
37
1.5
87
1.6
66
1.7
54
1.8
11
min 1 1.2 1.4 1.6 1.8 2 2.2 2.4
mAU
2
4
6
8
10
12 1.3
45
1.5
64
1.6
46
1.7
87
1.8
46
Column: 2.1 x 50 mm columns
Sample injection: 1 µL
Mobile phase A: 0.1% TFA in d. water
Mobile phase B: 0.01% TFA in ACN
Flow rate: 1 mL/min (~700bar)
HPLC instrument: 1290 Infinity Series
Detection: UV, 280
ZORBAX RRHD 300SB-C3, 1.8 µm
ZORBAX RRHD 300-Diphenyl, 1.8 µm
Ribonuclease A, Cytochrome C and Lysozyme (3 mg/mL)
Time
(min)
%B
0 10
2.5 70
Arrows indicate better separation
resolution of ZORBAX RRHD
300-Diphenyl
Poroshell 300
300 A pore size
Stablebond chemistry
available in C-3, C-8,C-18, and C-18 Extend
5 um particle size
Porous Shell
SOLID CORE
5 m
Poroshell Particles Provide More Efficient Peaks
for Peptides and Proteins
Poroshell 300 is for the separation of large peptides and proteins
Poroshell particles have a solid silica core with a superficially porous surface with a 300Å pore size
This results in more efficient mass transfer and sharper peaks with large proteins and peptides
Comparison of Diffusion Distance Totally porous silica vs. superficially porous silica
0.25 µm
5 µm
Totally Porous Particle
2.5 µm Required diffusion distance
for a macromolecule
reduced 10 fold !
5 µm
Superficially Porous Particle
Flow Rates for Poroshell Columns
Column Internal Diameter Porous Particle
Flow Rate Range
Poroshell
Flow Rate Range
2.1 mm 0.1 – 0.3 mL/min 0.3 – 3 mL/min
1.0 mm 30 – 60 L/min 0.08 - 0.75 mL/min
• Very high flow rates can be used effectively with Poroshell columns
1 2 3
4
Sustained Efficiency and Resolution
Poroshell 300 Reverse Phase Effect of Increasing Flow Rate in Protein Analysis with
Poroshell
1 2 3
4
Sustained Efficiency and Resolution
Poroshell 300 Reverse Phase Effect of Increasing Flow Rate in Protein Analysis with
Poroshell
1 2 3 4
Effect of Increasing Flow Rate in Protein
Analysis using Totally Porous Silica
Lost Efficiency and Resolution
1 2 3 4
Lost Efficiency and Resolution
Effect of Increasing Flow Rate in Protein
Analysis using Totally Porous Silica
• Poroshell technology facilitates ultra-fast HPLC analysis of very large protein subunits. Note the improved peak shape and recovery on Poroshell, in <2 min. Dramatic example of the lower surface area advantage.
Improved Recovery with Poroshell Analysis of High-Molecular Weight Proteins - Thyroglobulin 660 kDa
High Flow Rates with 2.1 mm ID Poroshell for High
Resolution and Fast Separations
1
2
34
5
67
8
0 0.5 1.0
Time (min)
Columns: Poroshell 300SB-C18
2.1 x 75 mm, 5 mm
Mobile Phase: A: 0.1% TFA
B: 0.07% TFA in ACN
Gradient: 5 – 100% B in 1.0 min.
Flow Rate: 3.0 mL/min.
Temperature: 70°C
Pressure: 250 bar
Detection: UV 215 nm
Sample:
1. Angiotensin II
2. Neurotensin
3. Rnase
4. Insulin
5. Lysozyme
6. Myoglobin
7.Carbonic Anhydrase
8.Ovalbumin
Reversed Phase
Heavy and light chain analytical characterization
Use of 1200 LC, Poroshell 300SB
columns and UV detection to
characterize antibodies
February 2010
Agilent Confidential 86
More Poroshell Bonded Phases Provide Selectivity
Options to Enhance Resolution:
min0 0.5 1 1.5 2 2.5
mAU
0
100
200
300
400
DAD1 A, Sig=215,16 Ref=360,100 (05080223.D)
Poroshell SB-C18, 2.1 x 75 mm
1
2 3 4
5, 6
7
8
9 0.9
76
1.0
77
1.2
89
1.3
95
1.4
63
1.5
23
1.7
41
1.8
60
3.0
12
min0 0.5 1 1.5 2 2.5
mAU
0
100
200
300
400
DAD1 A, Sig=215,16 Ref=360,100 (05080225.D)
51
23
4
6
7
8
9
Poroshell SB-C3, 2.1 x 75 mm
0.9
01
1.0
36
1.2
64
1.3
54 1.4
30
1.4
49
1.4
97
1.7
39
1.8
37
Column: Agilent Poroshell (2.1 x 75 mm); Temp.: 70 0C; Flow: 0.5 mL/min; Det: UV 215 nm
Mobile Phase: A= 0.1% TFA/H2O, B= 0.07% TFA/ACN; Gradient: 5-100% B in 3.0 min
1.5
5
4 6
7
1.5
5
4 6
7
1.5
4
5, 6
7
1.5
4
5, 6
7
Samples:
1. Angiotensin II
2. Neurotensin
3. RNase A
4. Insulin B Chain
5. Insulin
6. Cytochrome C
7. Lysozyme
8. Myoglobin
9. Carbonic
Anhydrase
Poroshell 300SB-C18
2.1 x 75 mm
Poroshell 300SB-C3
2.1 x 75 mm
•Changing from SB-C18 to SB-C3, within the Poroshell family results in resolution of peaks 5 and 6, still in 3 min!
PLRP-S
Polystyrene divinylbenzene bead
Available in 100, 300, 1000, and 4000A pore sizes
Particle sizes 3, 5, 8, 10, and higher
Various geometries from Microbore to Preparative
PLRP-S Continued:
Features
pH 1-14
Extreme buffer concentrations
High temperature stability
Durable and Resilient
Inherently hydrophobic so does not
require a bonded alkyl chain to confer
hydrophobicity
Benefits
Acid and base cleanup
Typically < 8M
200°C
Long Lifetimes
Avoids typical silica problems of
silanol group
Analysis of very large biomolecules
or high speed separations 1000A & 4000A pores
PLRP-S & RP-Silica Comparison
Eluent A :1% ACN (0.1% TFA) Flow rate: 1.0 mL/min
Eluent B :99% ACN (0.1% TFA) Detector: UV, 210 nm
Gradient :35-80% B, 0-20 mins
Key
1 H-(Phe)1-OH
2 H-(Phe)2-OH
3 H-(Phe)3-OH
4 H-(Phe)4-OH
5 H-(Phe)5-OH
PLRP-S 300Å 5 µm RP-Silica C18 300Å 5 µm
mins
1
2
3 4
5
0 20 mins
1
2 3
4
5
0 20
25-bp Ladder Double Stranded DNA Ladder
1
1 1
12% PAGE
25
25
50
25
25
50
50
100
100
100
100
50
1 1
1 1
time
HPLC
PLRP-S 100Å
PLRP-S 300Å
PLRP-S 1000Å
PLRP-S 4000Å
Column: PLRP-S 150x2.1mm ID
Eluent A: 100mM TEAA
Eluent B: 100mM TEAA, 50% ACN
Gradient: 12.5%-50%B in 150 mins
Flow Rate: 200µl/min
Pore Size Resolving Range
100Å 50-70 bp
300Å Up to 250-300 bp
1000Å Up to 400-450 bp
4000Å >500 bp
In Summary
Agilent has many options for the analysis of bio-molecules and
will continue to develop new products to address this growing
market.
New BioHPLC SEC New Zorbax SB 300A
and IEX Brochure 1.8um RRHD Brochure
#5990-5195 #5990-8124EN
BioHPLC Columns Navigator Poster #5990-5526EN
94
New BioHPLC Column
Selection Guide
#5990-9384
95
New LC Handbook
#5990-7595
September 30, 2012
Agilent Bio-inert LC Seminar
96
Biomolecule Purification
Brochure
#5990-8335
New Protein Identification
And Impurity Profiling
#5991-0625EN
New Applications on the bio-inert LC and
Bio-IEX Columns: Usage of Buffer Advisor
# 5990-9628EN # 5990-9629EN
Further Information: App Notes
Reversed-Phase RRHD
98
Pub Number Title
5990-9628EN Optimizing Protein Separations with Weak Cation-Exchange Columns
5990-9629EN pH Gradient Elution for Improved Separation of Monoclonal Antibody Charge Variants
5990-9894EN Optimum Pore Size for Characterizing Biomolecules with Agilent Bio-SEC Columns
5990-9931EN Fast Separations Using Agilent Weak Cation Exchange Columns
5990-9631EN Rapid UHPLC analysis of reduced monoclonal antibodies using Agilent ZORBAX Rapid Resolution High Definition (RRHD)
300SB-C8 column
5990-9667EN Reversed-phase optimization for ultra fast profiling of intact and reduced monoclonal antibodies using Agilent Rapid Resolution
High Definition 300SB-C3 column
5990-9668EN Ultra high speed and high resolution separations of reduced and intact monoclonal antibodies with Agilent ZORBAX RRHD
sub-2 µm 300 Diphenyl UHPLC column
Further Information: App Notes & Primer
Reversed-Phase RRHD
99
Pub
Number
Title
5990-
7989EN
Agilent ZORBAX 300SB-C18 1.8 µm Rapid
Resolution High Definition columns for proteins
5990-
8244EN
Analyze MAb and BSA digests by UHPLC with UV
detection and Agilent ZORBAX RRHD 300SB-C18
5990-
7988EN
Analysis of oxidized insulin chains using reversed-
phase Agilent ZORBAX RRHD 300SB-C18
5990-
9248EN
Fast separation of recombinant human erythropoietin
using reversed-phase Agilent ZORBAX RRHD
300SB-C18, 1.8 µm
5990-
9016EN
Reversed-phase separation of Intact Monoclonal
Antibodies using Agilent Zorbax RRHD 300SB
5990-
9544EN
Separation of recombinant Human Erythropoietin
(rEPO) using Agilent Bio-SEC 3
5990-
9270EN
Separation of Protein Standards on Agilent 3um Ion
Exchange Columns by Cation Exchange
Chromatography
New Primer: Recombinant
Protein Characterization
#5990-8561
Further Information: App Notes & Posters
Reversed-Phase RRHD
100
Pub
Number
Title
5989-
9899EN
Fast Protein Separations Using Agilent Poroshell 300
5990-
6414EN
Characterization of monoclonal antibodies on Agilent
1260 Bio-inert Quaternary LC by SEC using Bio-SEC
columns
5990-
6192EN
Physicochemical characterization of a therapeutic
protein by peptide mapping, SEC and IEX using Agilent
1260 LC
5990-
8613EN
Fast Separation of Monoclonal Antibody Monomer and
Dimer by SEC with Agilent Bio-SEC
5990-
8895EN
Defining the optimum parameters for efficient size
separations of proteins
5989-
6840EN
Comparison of Zorbax Stablebond 300A columns to
optimize selectivity for antibody separations using
LC/MS
5989-
0604EN
Optimization of rapid HPLC analysis of monoclonal
Intact Antibodies. Light Chains, and Heavy Chains
using Zorbax Poroshell
Further Information: App Notes & Posters
PUB # Title
5989-
0030EN
Rapid HPLC Analysis of Monoclonal Antibody IgG1
Light Chains Using Zorbax Poroshell 300SB-C8
5990-
9940EN
Charge Variant Analysis of Monoclonal Antibodies by
pH Gradient Separation on Cation-exchange Columns
5990-
0691EN
Increasing Throughput: The Advantages of pH
Gradients in High Speed Ion-Exchange Analysis of
Proteins
5990-
0680EN
Protein Aggregation: A Multiple Detector Investigation
into Size Exclusion Chromatography
6989-
9733EN
Rapid Human Polyclonal IgG Quantification using the
Agilent Bio-Monolith HPLC Column
5989-
9674EN
Rapid IgM Quantification in Cell Culture Production
and Purification Process Monitoring Using the Agilent
Bio-Monolith QA Column
5990-
3247EN
Fast Monitoring of Bacteriophage Production During
Fermentation Using the Agilent Biomonolith HPLC
Column
5990-
5524EN
Rapid Analysis of Adenovirus Type 5 Particles with
Bio-Monolith Anion-Exchange HPLC Columns to
Support the Development of a High Titre
Manufacturing Platform
101
Agilent Biopharma Training, AFO NPT 2010
BioHPLC Columns on the
Agilent Website
103
NEED ASSISTANCE?
LC Column help desk
1-800-227-9770
• orders
• customer service
(option 1,1)
• technical support
• applications assistance
(option 3,3,2)
Thank you !
Questions?