Central Reverse Osmosis Units Used For HEMODIALYSIS 2020 2436 3046 Revised Feb, 2013 Better Water LLC Central R.O. System Operation Central R.O. System Operator Manual Man Central RO Operator's Rev Dec 2014

Upload: trancong

Post on 21-May-2018

215 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Central Reverse Osmosis Units

Used For HEMODIALYSIS

2020 2436 3046

Revised

Feb, 2013

Better Water LLC

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 2: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 3: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Our Company…………………………………………… 01

Important Operator Warnings and Cautions…………..... 03

Introduction to the Central R.O. System…..…………… 05

System Pre-requisites.………………..………………… 07

Description…………………..……………………….… 09

Feed water Requirements……….………………….….. 11

Membrane specifications………………………….….. 13

R.O. Pump Air cooled……………………………….…. 15

Pressure Transmitter……………………………….…… 17

Digital Flush Timer…………………………………...… 19

Miscellaneous features…………………………………. 21

Pre-treatment interlock fault light…………………...… 23

.03 Pyrogen final filter……………………..................... 25

Variable Frequency drive……….…………..………...... 29

R.O. remote alarm box……………..…………………… 31

Disinfect tank…………………. ………………………. 33

System operation……………………………………...... 35

Timed operate button…………………….…………….. 37

Startup…………………………………………………. 39

Initial operational values………………………………. 41

Unit Adjustments……………………………………… 43

VFD Operation………………………………………… 49

Cleaning and disinfecting membranes………………… 53

Sanitizing the sanitary sample ports…………………… 59

Long term R.O. storage…………………………….….. 61

Shutdown……………………………………………… 63

System Maintenance………………………………….. 65

Central R.O. System Operation

Table of Contents

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 4: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Consumable Parts ………………………………………… 71 Change out Membranes…………………………………… 73 Factory Reset……………………………………………… 77 Optional assemblies………………………………………. 79 System trouble shooting………………………………….. 81 Example checklist………………………………………… 85 Limited Warranty…………………………………………. 89 Appendix 1 Technical Service Bulletins………………….. 91 Unit Test Data…………………………………………..… 99

Central R.O. System Operation Table of Contents

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 5: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

2020 CENTRAL R.O. Models And Specifications

Model EQRM-203K EQRM-205K

Capacity GPM/GPD 2.5/3600 3.75/5400

# of Membranes 2 3

Size of Membranes 4” x 40” Full fit

Control Voltage 24vac

Electrical Requirement 120vac/20 amp circuit, 208/330vac 3ph

Dimensions Approximately 20”x20” x 72” (includes pre-filters & junction boxes)

Incoming Water 6 gpm @30 psi 8 gpm @ 30 psi

Feed Water Requirements Temp:50f-92F,total chlorine <0.1ppm,SDI <3, TOC <2 ppm ,Operating pH range: 6-10,Bacterial Count

<50 cfu

Gauges RO Pump, Membranes, Reject

Flow meters Product, Reject, Recirculation

Water Quality Monitor Digitally Displays: % Rejection, Feed TDS, Product TDS, Alarm Set-Point

Standard Alarms Poor Water Quality, Low Feed Pressure, High Pump Pressure, High Membrane Pressure, High

Product Pressure and High Feed Water Pressure

Other Standard features

Disinfect Tank include , Renalin / Minncare Compatible , Product Water Divert to Drain , Silkscreened

Labeling , Rugged Stainless Steel Frame Construction , Baked on Powder Coat Finish , Polyamide

Membranes , Extremely Quiet Air-cooled pump with Power savings V F D

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014

Page 6: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

3046 CENTRAL R.O. Models And Specifications

Model EQRMA-3014K EQRMA-3016K EQRMA-3018K EQRMA-3020K

Capacity GPM/GPD 10.00/14,400 11.25/16,200 12.5/18,000 13.75/19,800

# of Membranes 8 9 10 11

Size of Membranes 4”x 40” full fit

Control Voltage 24vac

Electrical Requirement 208/230/406vac 3ph

Dimensions Approximately 30”x46” x 72” (includes pre-filters & junction boxes)

Incoming Water 8 gpm @ 30 psi

Feed Water Requirements Temp:50f-92F,total chlorine <0.1ppm,SDI <3, TOC <2 ppm ,Operating pH range: 6-10,

Bacterial Count <50 cfu

Gauges RO Pump, Membranes, Reject

Flow meters Product, Reject, Recirculation

Water Quality Monitor Digitally Displays: % Rejection, Feed TDS, Product TDS, Alarm Set-Point

Standard Alarms Poor Water Quality, Low Feed Pressure, High Pump Pressure, High Membrane Pressure, High Product

Pressure and High Feed Water Pressure

Other Standard features

Disinfect Tank include , Renalin / Minncare Compatible , Product Water Divert to Drain , Silkscreened Label ing ,

Rugged Stainless Steel Frame Construction , Baked on Powder Coat Finish , Polyamide Membranes , Extremely Quiet

Air-cooled pump with Power savings V F D

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014

Page 7: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

2436 CENTRAL R.O.

Models and Specifications

Model EQRMA -245K EQRMA -247K EQRMA -249K EQRMA -24102K EQRMA -2412K

Capacity GPM / GPD 3. 75 / 5400 5. 0 / 7200 6. 25 /9000 7. 5 / 10800 8.75/12600

# of Membranes 3 4 5 6 7

Size of Membranes 4” x 4 0” full fit

Control Voltage 24vac

Electrical Requirements 208/230/450 3ph

Dimensions Approximately 39" W x 42” D x 77” H (includes pre-filters & junction boxes)

Incoming Water 8 gpm 12 gpm 14 gpm 16 gpm 19 gpm

Feed Water requirements Temp. : 50f-92f, Total Chlorine < 0. 1p p m, T O C < 2p p m, S D I < 3, Operating pH range : 610, 30

psi inlet pressure , Bacterial Count < 50cfu

Gauges Pre-Filter, Membrane , Reject , Product

Flow meters Product , Reject

Water Quality monitor Digitally Displays : % Rejection , Feed TDS , Product TDS , Alarm Set -Point

Standard Alarms

Poor Water Quality , Low Feed Pressure , High Membrane Pressure , High Product Pressure , & High

Fee Water

Temperature

Other Standard features

Disinfect Tank include , Renalin / Minncare Compatible , Product Water Divert to Drain , Silkscreened

Label ing , Rugged Stainless Steel Frame Construction , Baked on Powder Coat Finish , Polyamide

Membranes , Extremely Quiet Air-cooled pump with Power savings V F D

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014

Page 8: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014

Page 9: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Our Company

User Assistance

Information

Contact Us

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Better Water is a leading integrated manufacturer of water

treatment machines, components and equipment for the

industrial, commercial and institutional markets.

Located in Smyrna, Tennessee,

Better Water continues its history

of manufacturing and worldwide

distribution of equipment

specifically designed for the renal

dialysis market.

Founded in 1971, Better Water

has built a reputation for solving

our customers' toughest problems

with high-quality products and

unmatched service.

Support is available regarding all BW LLC Systems

24 hours-7 Days.

Normal business hours are Monday through Friday (excluding

holidays) 8:00 am to 3:30 pm Central Time call:

615-355-6063 Emergency Assistance is available after normal business hours

(including holidays)

BEFORE calling for emergency assistance:

• Check the Operator’s Trouble-shooting guide

• Check the electrical power connections, fuses and/or

circuit breakers

• Check all valves to ensure each is in the correct position.

If this fails to correct the problem and an emergency

hemodialysis treatment situation exists, call:

615-355-6063 Your call will be automatically forwarded to the on-call cell phone

directory. Please copy all the numbers and select the most

appropriate number for your emergency.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 1 of 99

Page 10: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014

Page 2 of 99

Page 11: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Warnings and Cautions

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

WARNINGS and CAUTIONS WARNING: It is unsafe to operate this device without

first reading and understanding the entire Operator's

Instruction Manual.

WARNING: Misuse or improper operation of this device

could result in serious injury, death, or other serious

reactions to a patient undergoing hemodialysis treatment.

WARNING: Misuse, improper use or handling of

chemical disinfectants and chemical cleaning solutions

could result in serious injury or even death. You must

comply with the information contained in the Material

Safety Data Sheet (MSDS) for the chemical being used.

CAUTION: Improper operation of this device could result

in a low or no flow alarm on the hemodialysis machine.

CAUTION: Misuse or improper operation of this device

will void any warranty.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 3 of 99

Page 12: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

USER

CAUTIONS

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

User Cautions

1. When used as a medical device, federal law restricts this

device to sale by or on the authority of a physician. Per

CFR 801.109 (b)(1).

2. Improper operation of any of these devices could result

in a low or no-flow alarm on the dialysis machines.

3. Misuse or improper operation of these devices will void

any warranty.

4. Where water is mentioned, unless otherwise noted, it

must be AAMI standard quality water.

5. Electrical and plumbing connections must adhere to

local statutes and any facility codes. Connect this device to

a proper ground connection in accordance with the

National Electrical Code. Do not remove the ground wire

or ground plug. Do not use an extension cord with this

equipment.

6. Do not remove any Caution, Warning or any other

descriptive labels from the device.

7. Do not operate this device in an explosive environment

or in the presence of flammable materials. Do not use this

device to store, mix or transfer flammable liquids.

8. Movement or vibrations during shipment may cause

connections to loosen.

9. Do not operate this unit in an environment where

temperatures may be below 50˚ F or above 92˚ F.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 4 of 99

Page 13: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Introduction

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

The Reverse Osmosis (R.O.) Machine is a multiple-

membrane, device that is pressurized by a stainless steel

pump and motor. The Reverse Osmosis process uses a

semi-permeable membrane to separate and remove

dissolved solids, organics, Pyrogen, submicron colloidal

matter from the water. The process is called "reverse"

osmosis since it requires pressure to force pure water

across a membrane, leaving the impurities behind. Reverse

Osmosis is capable of removing 95%-99% of the total

dissolved solids (TDS) thus providing safe, pure water.

Based upon your facility’s specifications and information

about tap water data, the R.O. was designed and built to

exacting standards and Good Manufacturing Practices as

outlined by the FDA. The R.O. is a device that uses a

membrane separation process for removing solvent

(contaminants) from solution (tap water).

The R.O. is the most important and costly component in

the water treatment system. With appropriate pretreatment

of the tap water, proper cleaning and disinfection, the R.O.

membrane life can be prolonged for several years.

Central R.O. System Operation

Central R.O. System Operator Manual

OSMOSIS EQUILIBRIUM REVERSE OSMOSIS

Osmonic

Pressure

Water flows from the lower

concentration of salts

across a permeable

membrane to the higher

concentration.

Osmotic Pressure is

the pressure required

to stop water flow and

reach equilibrium.

By applying pressure greater

than the osmotic pressure, the

flow of water is reversed. The

water flows from a higher

concentration of salts to a lower

concentration.

Man Central RO Operator's Rev Dec 2014

Page 5 of 99

Page 14: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014

Page 6 of 99

Page 15: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

System

Prerequisites

Space

Water supply

Drain

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

System

Prerequisites

The system should be level and located as close as possible

to the point of use.

Minimum space required 60” Deep x 40” to 102” Wide

(depending upon model and configuration) x 72” Tall.

The floor must be capable of support up to 500 pounds.

Space considerations should include adequate Operator

access.

The system requires a 1 inch female NPT threaded water

connection with an adjacent shut off valve.

The system requires a sanitary drain capable of discharging

20 gallons per minute (GPM) or better.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 7 of 99

Page 16: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014

Page 8 of 99

Page 17: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Description

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Description

The RO system fluid contact components are constructed

entirely of polyethylene, 300 series stainless steel or

other AAMI recognized materials.

The Reverse Osmosis process uses a semi-permeable

membrane to separate the dissolved solids, organics,

pyrogens, submicron colloidal matter from the water. The

process is called "reverse" osmosis since it requires

pressure to force pure water across a membrane, leaving

the impurities behind. Reverse Osmosis is capable of

removing 95%-99% of the total dissolved solids (TDS)

thus providing safe, pure water.

Example of an R.O. Membrane

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014

Page 9 of 99

Page 18: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 10 of 99

Page 19: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Feed Water

Requirements

Pressure

Flow Rate

Silt Density

Index

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Feed Water

Requirements

The importance of monitoring and controlling the

municipal tap water can not be underestimated. The R.O.

feed water must meet the following parameters:

40 psi (Optimum) to 90 psi (Maximum). The minimum

pressure must be maintained with water flowing at the

maximum required flow-rate.

The Requirement for Feed Water will be determined by the

size and model of the specific R.O.

Refer to the unit’s Specification Sheet for specific flow

values. Chlorine is commonly used as a disinfecting agent

in municipal systems. Disinfection byproducts can form

when disinfectants, such as chlorine, react with naturally

present compounds in the water. Chlorine/Chloramines in

the RO unit feed water must be less than 0.1 ppm.

Silt Density Index (SDI) is a test to measure of the level of

suspended solids in feed water. High SDI values can lead

to membrane fouling. A SDI of less than 3 is considered

acceptable for the R.O. system.

Turbidity is the amount of small particles of solid matter

suspended in water as measured by the amount of

scattering and absorption of light rays caused by the

particles. Feed water turbidity must be less than 1

nephelometric turbidity units (NTU).

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 11 of 99

Page 20: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Feed Water

Requirements

Con’t

Hardness

Water

Temperature

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Hardness is a characteristic of feed water due to the

presence of dissolved calcium and magnesium; water

hardness is responsible for most scale formation and can

form insoluble residue in pipes and other water contact

surfaces. Hardness is usually expressed in grains per gallon

or parts per million, all as calcium carbonate equivalent.

Hardness level of feed water for the R.O. system must be

less than 2 gpg (grains per-gallon) or 34.2 ppm (parts-per-

million).

Feed water temperature must be between 70 and 92ºF

(Optimum: 77ºF). The R.O. will shut-down in a high feed

temperature alarm at 93ºF or as set by the installer.

If high temperature alarm conditions are causing frequent

shut downs, contact Better Water for assistance.

Man Central RO Operator's Rev Dec 2014 Page 12 of 99

Page 21: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Membrane

Specifications

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Membrane Array Flow Rate Specifications

The flow rates for the arrays listed below are rated at 77˚f

feed water, + or – 20% and at 50% recovery. *Note: The

2436 Series RO has no Re-Circulate adjustment and

was set at the factory. No user adjustment is required.

2020 RO Series Product Flow Reject Flow Re-Circulate Flow

2 Membrane Array 2.5 GPM 2.5 GPM 3.5 GPM

3 Membrane Array 3.75 GPM 3.75 GPM 3.5 GPM

2436 RO Series Product Flow Reject Flow Re-Circulate Flow*

3 Membrane Array 3.75 GPM 3.75 GPM 3.5 GPM*

4 Membrane Array 5.0 GPM 5.0 GPM 2.5 GPM*

5 Membrane Array 6.25 GPM 6.25 GPM 2.5 GPM*

6 Membrane Array 7.5GPM 7.5 GPM 5.0 GPM*

7 Membrane Array 8.75 GPM 8.75GPM 5.0 GPM*

3046 RO Series Product Flow Reject Flow Re-Circulate Flow

8 Membrane Array 10.0 GPM 10.0 GPM 5.0 GPM

9 Membrane Array 11.25 GPM 11.25 GPM 4.0 GPM

10 Membrane Array 12.5 GPM 12.5 GPM 1.5 GPM

11 Membrane Array 13.75 GPM 13.75 GPM 1.5 GPM

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 13 of 99

Page 22: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 14 of 99

Page 23: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

R.O. Pump –

Air Cooled

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

R.O. Pump – Air Cooled

A stainless steel, multi-stage, air-cooled centrifugal pump

designed for continuous duty service. The pump is

powered by a 3 or 5 hp motor. When a no-water situation

occurs, the pump will automatically shut-down to prevent

damage from running dry.

There are no daily startup requirements for the pump.

The performance is monitored daily by the user to ensure

design pressure is being maintained.

If this device should fail, AAMI quality water will not be

produced by the R.O. and could cause harm to the patients

undergoing hemodialysis treatments.

Other than routine cleaning of the exterior surfaces, there

is no required cleaning or maintenance schedule or

procedure.

Central R.O. System Operation

Central R.O. System Operator Manual

Example of

Air-Cooled

Pump

Man Central RO Operator's Rev Dec 2014 Page 15 of 99

Page 24: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 16 of 99

Page 25: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Pressure

Transmitter

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Pressure Transmitter

The measurement of water pressure during the operation of

the system will be accomplished using a Digital Pressure

Transmitter.

The pressure sensor uses either piezo-resistive or metallic

thin film to convert input mechanical pressure into an

output electrical signal.

This device is automatic and requires no operator start-up

procedures.

The performance is monitored daily by the user to ensure

design pressure is being maintained.

If the Pressure Transmitter fails, the RO Pump will operate

continuously at maximum pressure.

Other than routine cleaning of the exterior surfaces, there

is no required cleaning or maintenance schedule or

procedure.

Central R.O. System Operation

Central R.O. System Operator Manual

Example of Pressure

Transmitter

Man Central RO Operator's Rev Dec 2014 Page 17 of 99

Page 26: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 18 of 99

Page 27: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Digital Flush

Timer

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Digital Flush Timer

The Digital Flush Timer is an important component of the

Reverse Osmosis Machine. The flush timer is a 24-hour

adjustable 24 volt Digital mechanism that controls the

frequency and duration of the Flush Cycle.

The Flush timer serves two purposes: It is used to set the

time of day, and the frequency and duration of the Flush

Cycle(s).

The digital flush timer is equipped with a battery back-up,

so loss of power to the R.O. will not affect the timer

operation. However, the actual life of the battery is not

known so the timer should be checked monthly for correct

time of day setting.

The Flush Timer was preset set by the system installers on

the day of the installation. Once the Frequency and the

Length of the Flush Cycle have been set, that setting will

remain until manually changed.

For details regarding the Flush Timer, see the section

Maintenance – Flush Timer Adjustment.

Example of

Digital Flush Timer

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 19 of 99

Page 28: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 20 of 99

Page 29: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Miscellaneous

Features

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Miscellaneous Features

The Low Pressure Switch is located on the suction side

of the pump. The switch, via the main controller board,

acts to shut down the R.O. in the event of insufficient

pressure to the RO.

A keyed switch is an additional safety feature and

prevents unauthorized activation of the disinfect cycle. A

key must be inserted to place the unit into the disinfect

mode.

A red alarm light will turn on when any alarm conditions

occur. The light will remain on until the alarms conditions

are cleared and the alarms are reset by the RESET button

on the Control Panel.

Central R.O. System Operation

Central R.O. System Operator Manual

NOTICE If the alarm condition is caused by poor water quality in the first 3

minutes, the alarm light will be on, the audio alarm will not sound,

and the alarm condition will not be sent to the nurse’s station. The

light will go out when the water quality meets the requirement(s).

After 3 minutes, if the water quality alarm condition persists, both

the light and audio alarm will be on. The alarm condition will be sent

to the nurse’s station.

Man Central RO Operator's Rev Dec 2014 Page 21 of 99

Page 30: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 22 of 99

Page 31: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Pre-Treatment

Interlock Fault

Light

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Pre-Treatment Interlock Fault Light The Pre-Treatment Interlock Fault Light is located on the

face of the Control Box (on the front of the R.O.). This

light will illuminate and alert the operator that the

interlock system has been interupted. This is normal

during the Backwash and/or regeneration of any pre-

treatment backwashing filters or softeners. This light

will also illuminate when other fault conditions occur,

such as interlock wires being disconnected, interlock

relay failure of Main Control Board problems. The R.O.

will not run when this light is illuminated. If this light is

illuminated and no filter or softener is in backwash or

regeneration mode, call for assistance.

Pre-Treatment

Interlock Fault Light

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 23 of 99

Page 32: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 24 of 99

Page 33: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

.03 Pyrogen

Final Filter

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

.03 Pyrogen Final Filter

The .03 final Pyrogen filter is located on the left

side of the R.O. on the Product Line. There is a

Sanitary Sampling Port after the filter.

On 2436 and 3046 R.O. units this filter uses a

standard 20” Filter Housing and on 2020 R.O.

units the filter is a cartridge.

The .03 filter should be changed every 6 months.

The Product Pressure Gauge on the front of the

R.O. will show the pressure before this filter.

On a Tank Feed System, the product hose from

the R.O. is connected to the reservoir via a 3 way

valve. There is no pressure gauge on the outlet

side of the filter.

On a Direct Feed System, an additional outlet

pressure gauge is installed on this filter. The

Product Pressure Gauge on the front of the R.O.

will display the pressure on the inlet side of the

filter. Both of these gauges should be monitored

and the filter should be changed if the ∆P reaches

or exceeds 20psi.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 25 of 99

Page 34: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

.03 Pyrogen

Final Filter

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

The Replacement Filter is part number:

SUCAPE00575 through Better Water LLC Sanitary Sample Ports

Product Connection.

Hosed to reservoir

(Tank Feed System)

or

Hosed to Distribution

Loop (Direct Feed

System)

Man Central RO Operator's Rev Dec 2014 Page 26 of 99

Page 35: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

.03 Pyrogen

Final Filter

And Sample

port for 2020

R.O.

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

The Replacement Filter is part number:

SUCAPE00547 through Better Water LLC

2020 R.O. Sample Ports

Man Central RO Operator's Rev Dec 2014 Page 27 of 99

Page 36: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 28 of 99

Page 37: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Variable

Frequency

Drive

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Example of Variable

Frequency Drive

Controller

Variable Frequency Drive

Variable-frequency drives are reliable, easy to operate,

increase the degree of flow control, and reduce pump

noise. A variable-frequency drive is an electronic

controller that adjusts the speed of an electric (R.O. Pump)

motor by modulating the power being delivered. Variable

frequency drives provide continuous control matching

motor speed to the specific demands of the work being

performed.

The performance is monitored daily by the user to ensure

design pressure is being maintained.

The probability of failure is remote; however should the

variable frequency drive fail, the R.O. pump will not

operate properly.

Other than routine cleaning of the exterior surfaces, there

is no required cleaning or maintenance schedule or

procedure.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 29 of 99

Page 38: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 30 of 99

Page 39: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

R.O. Remote

Alarm Box

When May Be

Omitted

Daily Startup

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

The R.O. Remote Alarm Box is a molded plastic box,

usually located on the patient floor, in a position where it

can be easily seen by clinic personnel during normal work

duties. The box is equipped with audible and visual alarms

that monitor the R.O. and Reservoir water level. The

Remote Alarm Monitoring Box requires no external power

supply, but receives 24vac power and signals from the

R.O. which it is monitoring. This box has 2 RED lights

that will illuminate when the R.O. goes into an alarm

condition or when the water level in the reservoir falls

below the bottom float, and 1 AMBER light that will

illuminate and flash when the R.O. is in Disinfect Mode.

AAMI standards require that the R.O. alarms be audible in

the patient area. If the R.O. is located close enough to be

heard in the patient area, the remote alarm may be omitted

This device is automatic and requires no daily startup

procedures.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 31 of 99

Page 40: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

R.O. Remote

Alarm Box

Con’t Monitoring

Requirements

Consequences of

Failure

Maintenance

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

This is a monitoring device; therefore it will monitor the

R.O. constantly, as to the state of the water.

The probability of failure is remote; however, should the

Remote Alarm Box fail, the clinic personnel my not be

aware of the changes in water quality.

See the Section – Maintenance.

Typical RO Remote

Alarm Monitoring

Box

Man Central RO Operator's Rev Dec 2014 Page 32 of 99

Page 41: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Disinfect Tank

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

The Disinfect Tank is made of a non corrosive, molded

plastic with a lid. The tank is equipped with 2 ports on the

top (side) and a single port on the bottom (side), and all the

necessary hoses, valves and fittings. (Some assembly

required) The hoses supplied with the Disinfect Tank will

connect to the top (side) ports on the Disinfect Tank and

the other ends will connect to the R.O. Product and Drain.

The third hose will connect the bottom (side) port to the

Clean/Disinfect Valve on the R.O. (See Disinfect

Procedures).

The storage tank 3-way Valve

Each R.O. is supplied with a Disinfect Tank

to speed the process of disinfection

when multiple R.O. units are used.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 33 of 99

Page 42: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

DISINFECT TANK ASSEMBLY

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 34 of 99

Page 43: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

System

Operation

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

System Operation

Before you start using this device, Operators must

read and understand this manual in its entirety. This

manual of Operator's Instructions describes in

considerable detail all of the steps and procedures

required to safely operate this device. With proper

operation, maintenance and care, this device should

give you years of reliable service.

It is unsafe to operate this device without a basic

understanding of water treatment and a thorough

understanding of the contents of this manual.

The R.O. was designed and built to your facility’s

specifications and information regarding the current

tap water conditions at your site. There is not a

Reverse Osmosis Machine on the market that is a

panacea for all water treatment requirements. The

R.O. cannot do the job alone. It is imperative to

monitor the tap water and feed water conditions.

Incoming tap water contaminants, temperature,

pH, pressure and flow-rates have a direct impact

on the quality and quantity of the Reverse

Osmosis Machine output.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 35 of 99

Page 44: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

System

Operation

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

The Operator must be aware of changing tap water

conditions. This can be easily accomplished with

good, two-way communications with the local

municipal water supplier and with routine testing of

the tap water.

To emphasize the importance of water treatment and

proper use of water treatment equipment used for

hemodialysis, the following is quoted from Health

and Human Services Publication FDA 89-4234,

"Numerous reports have documented that use of inadequately

treated water for hemodialysis poses a severe threat to the

health and safety of the hemodialysis patient. Despite this,

water treatment and water quality are often neglected areas of

hemodialysis. A major reason for this neglect is that water

treatment is a Technically complex subject which is not

generally a part of the education and training of clinical staff

in hemodialysis facilities."

Man Central RO Operator's Rev Dec 2014 Page 36 of 99

Page 45: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Timed Operate

Button

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Timed Operate Button The Timed Operate Button is located on the face of the

Control Box (on the front of the R.O. Its purpose is to

allow the R.O. to run for 30 minutes prior to the operator

to performing their daily checks (such as Chlorine,

hardness, etc.)

The Timed Operate Mode runs for 30 minutes and

cannot be altered or changed. The TANK–OFF–

DIRECT switch must be in the TANK position before

starting the Timed Operate. (It can be activated when in

the STB position but is not recommended) The Timed

Operate will not operate in DIRECT mode. When this

button is pressed, the R.O. will run for 30 minutes and

will turn off automatically. (NOTE: During this timed

operate cycle, the storage tank R.O. ON & R.O. OFF

floats are inactive. When the timed operate cycle is

completed, the storage tank floats are reactivated for

normal operation.)

Timed Operate

Button

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 37 of 99

Page 46: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 38 of 99

Page 47: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Startup

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Startup Check the RO CONTROL PANEL:

a) All alarm lights should be OFF and NO audible alarm

should be sounding. If any one or more of the alarm

lights are ON and/or an audible alarm is sounding,

PRESS the Alarm RESET button. All alarm lights

should go out and audible alarms should stop sounding.

b) The keyed, OPERATE-DISINFECT switch

should be in the OPERATE position. If this

switch is in the DISINFECT position, you

MUST verify that the RO does NOT contain any

disinfecting solution BEFORE proceeding to the

next step.

c) The OPERATE-DISINFECT-OFF switch

should be in the OPERATE position. If the

switch is in the DISINFECT or OFF position,

see step b and the warning above.

d) If you must reset the switches listed above,

the alarm lights will illuminate and the audible alarm will

sound, Press the ALARM RESET button to return the

alarms to normal.

Central R.O. System Operation

Central R.O. System Operator Manual

ALARM RESET

WARNING

Disinfectants can cause serious injury or death to patients undergoing

hemodialysis treatment.

OPER. DISINFECT OFF

OPER. DISINFECT

Man Central RO Operator's Rev Dec 2014 Page 39 of 99

Page 48: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Startup

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

To start the RO: Simply move the OPERATE-FLUSH

switch from FLUSH to the OPERATE

position. Press the Timed Operate

button.

NOTE: Should the Storage Tank level

have dropped to the R.O. START

Proximity sensor, the R.O. would have automatically started

upon turning the switch from FLUSH to OPERATE.

The Quality Purge cycle will commence. An air purge cycle

will run for approximately 30 seconds. Then the Product

Purge Valve will open and run to drain until the water

quality is above the set point.

After the Purge Cycles have completed, the R.O. should be

operating within the parameters as listed below:

i) Membrane pressure should read: 75-250 psi.

ii) There is no standard for Reject Pressure.

iii) Product Pressure should read: less than 25 psi on tank

feed systems and less than 70 psi on direct feed systems.

iv) The water quality should be above the set-point (90%).

OPER. FLUSH

NOTICE:

After approximately 30 seconds the RO pump will start, and at the same

time, the Water Quality Alarm light will illuminate. This is a normal

function to advise you that the first water produced by the RO is being

routed to drain until the water quality equals or surpasses the set-point

on the Water Quality Monitor. This purge cycles will take approximately

two minutes.

Man Central RO Operator's Rev Dec 2014 Page 40 of 99

Page 49: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Initial

Operational

Values

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Initial Operational Values After the Purge Cycles have completed, the R.O. should be

operating within the parameters as listed below:

i) Membrane pressure should read: 75-250 psi.

ii) There is no standard for Reject Pressure.

iii) Product Pressure should read: less than 25 psi on tank

feed systems and less than 70 psi on direct feed systems.

iv) The water quality should be above the set-point (90%).

If the RO is not within the design specifications listed

above, go to the section titled: Adjustment Procedures.

Perform a thorough quality assurance check of your entire

water treatment system including pretreatment, primary

treatment, post treatment, and distribution.

Central R.O. System Operation

Central R.O. System Operator Manual

NOTICE:

Refer to the Product Data Sheet for detailed information regarding

product flow meter, reject flow meter and recirculation flow meter

readings specific for this RO machine.

Man Central RO Operator's Rev Dec 2014 Page 41 of 99

Page 50: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 42 of 99

Page 51: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Unit

Adjustments

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Unit Adjustments On occasion, the R.O. may require adjustments that can be

performed by qualified operators. There are many factors

that can affect the performance of the R.O. If minor

adjustments to the R.O. do not produce the desired results,

you must investigate changes in the water feeding the R.O.

Water pH, temperature, pressure, TDS, and flow changes

may cause a drastic reduction in the R.O. performance. In

worst case situations, changes to the water can create

conditions the R.O. cannot handle without supplemental

processing.

PREPARATION: Before making any adjustments,

read this entire section and pay close attention to

cautions, notes, and items marked important.

Central R.O. System Operation

Central R.O. System Operator Manual

CAUTION:

Only qualified RO operators should make adjustments to the R.O.

CAUTION:

While making adjustments, the RO must be running with no alarm

conditions indicated.

Man Central RO Operator's Rev Dec 2014 Page 43 of 99

Page 52: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Unit

Adjustments

Con’t

Product

flow

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Unit Adjustments

1. On the front panel of the R.O., locate the "T-

Handle" on the Reject/Product Regulating

Valve.

2. Slowly, turn the "T-Handle" on the

Reject/Product Regulating Valve clockwise or

counterclockwise, as required, to set (as close as

possible) the Reject and the Product Flows. (i.e.

[50%recovery: 5 gpm product and 5 gpm reject flow].

3. After the Reject and Product Flows have been balanced,

perform a Quality Assurance Check on the R.O. to

insure all pressures and flows are within design

specifications.

CAUTION:

The inlet water temperature must be 70°F - 92°F with 77°F being

optimum. The temperature must be measured while the R.O. is running.

The optimum R.O. inlet water pressure is >40 psi while flowing at a

minimum rate appropriate for the size of the R.O. See Product Data

Sheet for specific details about this unit.

Man Central RO Operator's Rev Dec 2014 Page 44 of 99

Page 53: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Unit

Adjustments

Con’t

Reject Flow

Recirculation

Flow

Membrane

Pressure

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

The Reject Flow Adjustment is achieved with the Product

Flow Adjustment. See Product Flow Adjustment

There is no recirculation flow adjustment on the 2436 R.O.

machine. This is controlled by the built-in factory-set flow

controller

There are no user adjustments to the Membrane Pressure.

The Membrane Pressure is adjusted automatically via the

Variable Frequency Drive on the R.O. pump, based on the

designed flow rate of the R.O.

Should adjustments appear to be needed, contact Better

Water for assistance.

Man Central RO Operator's Rev Dec 2014 Page 45 of 99

Page 54: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Unit

Adjustments

Con’t

Product Pressure

Adjustment (Tank

Feed)

Product Pressure

Adjustment

(Direct-Feed)

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

There are no requirements to adjust the Product Pressure on

a Tank Feed system. The Product Pressure Gauge is for

measuring any possible system changes in the future.

On Direct Feed Systems, the product pressure should be

adjusted only when there are no requirements for R.O.

water. Locate the pressure T-handle on the pressure

regulating valve at the end of the distribution loop (return to

R.O.).

To increase product pressure, turn T-handle clockwise.

Maximum product pressure is 70 psi.

To decrease product pressure, turn the T-handle counter-

clockwise. The minimum pressure is 20 psi.

If you are unsure of any adjusting procedure, please call the

Assistance Phone Number for consultation.

NOTICE:

Product pressure must be 10 psi greater than feed water/Pre-Filter out

pressure to the R.O. Adjustments on feed water pressure may be required

before attempting to adjust product pressure.

Man Central RO Operator's Rev Dec 2014 Page 46 of 99

Page 55: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Unit

Adjustments

Con’t

Flush Timer

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Normally, the Reverse Osmosis Machine Membranes will

be flushed multiple times per day and for only 15 minutes

per flush. If your operation is considered less than normal

(less than 6 days a week operation), the FREQUENCY of

the FLUSH CYCLE should be set to occur more often. In

this case, we recommend that the FREQUENCY of the

FLUSH CYCLE should be set to occur (at a minimum) once

every four hours for 15 minutes per flush cycle.

If you should elect to operate the clinic (water treatment

system) on an unscheduled basis, there would be no

problem, because the OPERATE-FLUSH switch controls

the Flush Timer. In other words, the Flush Timer would be

inactivated when the OPERATE-FLUSH switch is in the

OPERATE position, and when the OPERATE-FLUSH

switch is returned to the FLUSH position, the FLUSH

TIMER would be reactivated. In either position of the

OPERATE-FLUSH switch, the FLUSH TIMER CLOCK

would continue to keep the correct time of the day; however,

the position of the OPERATE-FLUSH switch would

determine if a flush of the membranes would occur or not.

CAUTION

The time that a FLUSH CYCLE would occur should not coincide with the

backwash or regeneration of a pre-treatment component.

Man Central RO Operator's Rev Dec 2014 Page 47 of 99

Page 56: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 48 of 99

Page 57: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

VFD Control

Panel

Operation

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

VFD Control Panel Operation

This section does not apply to a 2020 series R.O.

The VFD is pre-programmed with operational parameters

set by Better Water and locked out to the end user.

When trouble shooting VFD issues Call Better Water

for technical assistance.

Located on the VFD front panel are 2 arrow keys; an up

arrow key and a down arrow key. Also located on the

Front panel display is a RUN and a STOP button (which

are inactive during operation), a MODE and a ENT

(ENTER) button. The VFD set point and Product/Reject

flow is set by adjusting the up or down arrow key. To

properly set the correct set point and Product/Reject of the

R.O., the R.O. must be on and running.

Run the R.O. 5 minutes before setting the set point. The

set point relates to the desired product and reject flow of

the R.O. Example: A 4 membrane array R.O. will be set at

5 GPM Product and 5 GPM Reject at 50% recovery. To

calculate the flow, count how many membranes that are in

the array then multiply by 1.25 GPM. 4 membranes X

1.25= 5.0 GPM Rated Product Flow plus or minus 20%. At

50% recovery the Reject flow will be 5 GPM also. When

setting the correct flow, the Post Filter pressure should be

at 40 psi on the R.O. when setting. If not, adjust the Post

City Boost Pressure on the Pre-Treatment. Push the up or

down arrow key to set the desired Product flow and set

point. Once set, then push the ENT (ENTER) button.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 49 of 99

Page 58: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

VFD Control

Panel

Operation

Con’t

Breaker

VFD

Transducer

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

FC will appear briefly in the display, and then the display

will read actual frequency the R.O. Pump is running at. If

the ENT (ENTER) is not pushed the VFD will revert back to

the original set point. On the front panel of the R.O., locate

the “T-Handle” on the Reject/Product Regulating Valve.

Slowly turn the “T-Handle” on the Reject/Product

Regulating Valve clockwise or counterclockwise as

required, to set (as close as possible) the Reject and the

Product as balanced as possible. (i.e. [50% Recovery: 5

GPM Product and 5 GPM Reject]. Continue further

adjustment as necessary utilizing the VFD and “T-Handle to

adjust to proper designed flow rate.

The breaker is located in Junction Box. If the breaker has

tripped the R.O. Pump and VFD will not operate. This

breaker will shut off power to the VFD to prevent damage to

the VFD and the R.O. PUMP.

The transducer is threaded into a tee located post R.O.

Pump/pre membrane. The transducer communicates with the

VFD to maintain the set point by analyzing pressure

differentials pre and post R.O. Pump. Example: If the pre or

post R.O. Pump pressure drops the transducer transmits that

signal to the VFD. The VFD will increase the frequency to

compensate for the pressure loss and will try to maintain set

point and Product/Reject flows. The same is true if the pre

or post R.O. Pump pressures increases the transducer signals

the VFD to decrease frequency.

Man Central RO Operator's Rev Dec 2014 Page 50 of 99

Page 59: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

VFD Control

Panel

Operation

Con’t

Troubleshooting

VFD Errors.

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

On occasion power interruptions due to storms, power

outages and other electrical phenomena can possibly cause a

VFD error to occur. Check the Breaker in the R.O. Junction

first. If it is tripped reset it. If not proceed to the VFD

control box and open the box.

If the VFD is flashing a code: OC1, OC2, OC3, EPHI,

EPHO, OP1, OP2, OP3, OL1 or OL2. This usually means

the VFD tripped.

If OC1, OC2 or OC3, this means there has been a over

current event that has occurred.

EPHI or EPHO means there has either been a phase input or

phase output event.

OP1, OP2 or OP3 means there has been a over voltage

event.

OL1 or OL2 means there has been a overload event.

If LStP is flashing in the display that means the VFD is idle

and below minimum programmed set point (This set point is

6 Hz and is pre-programmed and is locked out to the end

user.) If any of these error codes are displayed with the

exception of LStP, turn Operate switch to OFF then unplug

or disconnect main power. Wait for 10 to 15 seconds and for

the VFD display to power down. Then reconnect or plug

back in main power then turn Operate switch back to

Operate. If the error persists of will not reset call Better

Water LLC for further technical assistance.

Remember, this will not apply to a 2020 series R.O.

Man Central RO Operator's Rev Dec 2014 Page 51 of 99

Page 60: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 52 of 99

Page 61: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Cleaning and

Disinfecting

Membranes

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Cleaning and Disinfecting Membranes

The procedure below applies to the Low pH cleaning, High

pH cleaning, and disinfection.

It is recommended that a Low pH cleaning and

Disinfection be performed Monthly and a Low ph

clean, High pH clean, and Disinfection be performed

Quarterly.

1. Before beginning cleaning or disinfection rinse the

Disinfect Tank with R.O. water and drain

thoroughly.

2. Prepare the Disinfect Tank for hook up to R.O.

Connect the Product disinfect hose from the

storage tank 3 way valve and connect to the Park

fitting on the disinfect tank. Make sure the feed

valve at the bottom of tank is closed. Set the

Operate-Disinfect-Off Switch to Operate,

Operate-Disinfect Key Switch to Operate, and

Operate-Flush Switch to Operate. Put the R.O.

into Direct Mode via the toggle switch inside the

control box. Then push the Alarm Reset Button to

start the R.O. Fill tank with 15 gallons

(approximately 14 inches) of Product water for the

Low and High pH cleaning. For Disinfection fill

tank with 20 gallons (approximately 18 inches) of

Product Water.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 53 of 99

Page 62: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Cleaning and

Disinfecting

Membranes

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

3. When Disinfect tank has been filled with Product

water connect the supplied ¾” Reject hose from the

Disinfect tank to the R.O. Drain. Next connect the

supplied ¾” Feed hose from the Disinfect tank to the

Clean/Disinfect ball valve located at the base of the

pump. Before completely tightening the Parker fitting

to Clean/Disinfect ball valve open the Disinfect tank

ball valve to prime the feed hose. When water starts

trickling out of the Parker fitting then tighten the

Parker Fitting. Note: If the Disinfect tank feed hose

hasn’t been properly primed the pump will air lock

and could cause damage to the pump.

4. Before adding the Cleaners or Disinfectant start the

R.O. in Disinfect. Set the Operate-Flush Switch to

Operate, Operate-Disinfect-Off Switch to Disinfect,

then turn Operate-Disinfect Key Switch to Disinfect.

The R.O. will start up immediately. Verify flow by

looking at the Product and Reject Flow Meters and

visually see if water flowing back in to the Disinfect

tank through the Product and Drain hoses.

5. Before beginning cleaning with Low or High pH

cleaners it is recommended that a baseline pH

reading be taken form the Product returning into

the Disinfect tank.

6. Cleaners. If Cleaning add 1lb or 454 grams of either

Low pH or High pH cleaner to the Disinfect tank.

During circulation while cleaning the operator should

verify the pH of the cleaner used by checking the

Product returning back to Disinfect tank. For Low pH

cleaner the pH should read less than 3 ph units. For

High pH cleaner the ph should read greater than 9 pH

units.

Man Central RO Operator's Rev Dec 2014 Page 54 of 99

Page 63: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Cleaning and

Disinfecting

Membranes

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

7. Disinfection. If Disinfecting add 750 mls of either

Renalin or MinnCare to the Disinfect tank. Circulate

the cleaners or disinfectant for 15 minutes then shut

down R.O.

8. During circulation while disinfecting with Renalin or

MinnCare, a 1% Peracetic acid test strip should be

used to verify that the Product returning back to the

Disinfect tank is at 1%.

9. Cleaner Dwell Times. For Cleaners there is no

recommended dwell time.

10. Disinfectants Dwell Times. For Disinfectants such

as Renalin or MinnCare a 2 hour dwell is

recommended.

11. Rinsing Cleaners from the R.O. Next shut off R.O.

and connect the Storage Tank Disinfect hose to drain.

Disconnect the supplied ¾’ Drain hose from R.O. to

the Disinfect tank and reconnect the R.O. drain hose.

Close the Disinfect tank ball valve. Remove the ¾”

Disinfect tank Feed hose from the Clean/Disinfect ball

valve and close the Clean/Disinfect ball valve on the

pump. Start the R.O. in normal Operate mode.

Operate-Disinfect-Off Switch to Operate, Operate-

Disinfect Key Switch to Operate, and Operate-

Flush Switch to Operate. Then push the Alarm

Reset Button to start the R.O. Run the R.O. until the

pH from the Product returns to baseline pH that was

observed before cleaning.

NOTE: The R.O. must be in the Direct mode to run.

Man Central RO Operator's Rev Dec 2014 Page 55 of 99

Page 64: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Cleaning and

Disinfecting

Membranes

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

12. Rinsing Disinfectant from the RO. Next shut off

R.O. and connect the Storage Tank Product hose to

drain. Disconnect the supplied ¾’ Drain hose from

R.O. to the Disinfect tank and reconnect the R.O.

drain hose. Close the Disinfect tank ball valve.

Remove the ¾” Disinfect tank Feed hose from the

Clean/Disinfect ball valve and close the

Clean/Disinfect ball valve on the pump. Start the

R.O. in normal Operate mode.

NOTE: R.O. must be in direct mode in order to

operate.

Operate-Disinfect-Off Switch to Operate, Operate

-Disinfect Key Switch to Operate, and Operate-

Flush Switch to Operate. Then push the Alarm

Reset Button to start the R.O. Run the R.O. until the

Disinfectant has been rinsed from the R.O. To verify

that the disinfectant has been rinsed out from the

Product a Residual Peracetic acid test should be used.

If strip shows a positive continue rinsing to drain until

a negative reading is obtained. When a negative

reading has been obtained then turn off the R.O. for a

Rebound Break.

13. The Rebound Break is for 15 to 20 minutes. This is

highly recommended as residual disinfectant could

still be present in the membranes even if a negative

result was obtained during rinsing. After the Rebound

break time has been completed restart the R.O. in

normal Operate again and recheck Residual again. If

the Residual is positive or negative continue rinsing

an additional 15 minutes before hooking Product line

back to the Storage Tank. Turn 3-way valve on the

storage tank back to tank fill.

14. Direct Feed cleaning and disinfection. Leave the

Three-Way loop return valve on to RO. Follow the

above steps. When cleaning or disinfection is

completed turn the Three-Way loop return valve to

rinse cleaners and disinfectants to drain from the RO

and Loop. When rinsing is completed and RO is clear

turn the Three-Way loop return valve back on to RO.

Man Central RO Operator's Rev Dec 2014 Page 56 of 99

Page 65: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Cleaning And

Disinfecting

Agents For RO

Membranes Con’t

Minncare

BWI Acid Cleaner

1000

BWI Alkaline

Cleaner 2000

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

MINNCARE: DISINFECTING AGENT (Replaces Renalin or Actril)

• Application: R.O. and system disinfectant. MUST use Acid

Cleaner 1000 BEFORE using Minncare. For routine

maintenance, there should be a dwell time of 2-4 hours. If a

highly contaminated condition exists, dwell time can be

increased to 12 hours.

• Dilution: 750 cc’s to 20 gallons of R.O. water. • Minncare Part Number : SUMCOO00575

• 1% Test Strips Part Number : SUMCOO00577

• Residual Test Strips used to confirm rinsing – Part

Number : SUMCOO00576

BWI ACID CLEANER 1000: (Replaces Minnclean AC & Citric Acid)

• Cleaner may be used with brass fittings.

• Application: For removing mineral scale in membrane

applications. MUST first use Acid Cleaner 1000 BEFORE

using: Minncare, Alkaline Cleaner 2000.

• Dilution: 1 pound of Acid Cleaner 1000 to 15 gallons of

R.O. water.

• Acid Cleaner Part Number : SUMCOO00572

BWI ALKALINE CLEANER 2000: (Replaces Minnclean TF &

Organoclean)

• Cleaner may be used with brass fittings.

• Application: For removing oil, grease, biological

matter and grime on Thin Film Composite

membranes. MUST first use Acid Cleaner 1000

BEFORE using Alkaline Cleaner 2000.

• Dilution: 1 pound of Alkaline Cleaner 2000 to 15

gallons of R.O. water.

• Alkaline Cleaner Part Number: SUMCOO00571

NOTE: The label on all Minncare and BWI products will have mixing

directions.

WARNING:

All Cleaning And Disinfecting Agents Used With RO Units May Be

Hazardous If Handled Improperly. Proper Protective Equipment Must

Be Used

CAUTION:

Can not be used on R.O. with brass fittings;

R.O. Unit must have stainless steel fittings.

Man Central RO Operator's Rev Dec 2014 Page 57 of 99

Page 66: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 58 of 99

Page 67: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Sanitizing the

Sanitary Sample

Ports

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Sanitizing the Sanitary Sample Ports

Sanitize the Sanitary Sampling Port immediately before

each sampling operation.

1. Turn the port stem to the left, and allow a full flow of

liquid to pass out of the port for 1-2 minutes. Then

close the port by turning the stem to the right.

2. Fill a 20ml plastic polypropylene syringe with at least

10ml of 70% ethanol, 90% isopropyl alcohol, or 3%

hydrogen peroxide solution. Attach the port needle to

the syringe.

3. Insert the needle all the way into the port through the

stem opening, and express the sanitant into the port.

Allow a few milliliters to flow out of the port outlet.

4. As the needle is removed from the opening, squirt a

few milliliters of the sanitant over the outer surface of

the stem.

5. When ready to sample, open the port and rinse

product water 1 to 2 minutes before sampling. This

rinses the residuals and any endotoxins from the

sample port.

Central RO System Operation

Central RO System Operation

Man Central RO Operator's Rev Dec 2014 Page 59 of 99

Page 68: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 60 of 99

Page 69: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Long Term R.O.

Storage

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Long Term R.O. Storage Before preserving and storing the R.O. It is recommended

that the R.O. be cleaned with a Low pH cleaner first, High

pH cleaner second, and disinfect third. Sample as per your

procedures for colony count (cultures), endotoxins (LAL),

and AAMI.

MEMSTOR preservation: Prepare 15 gallons of RO or

D.I. water in clean/disinfect tank. Connect Product and

Reject hoses to Disinfect Tank. Connect the Disinfect Feed

hose to the R.O. Clean/Disinfect ball valve. Prime the

Disinfect Feed hose before starting. Failure to do so will

cause the pump to air lock. Turn OPERATE DISINFECT

OFF to DISINFECT and turn the OPERATE

DISINFECT key switch to DISINFECT and verify

circulation before adding the BWI 3000 MEMSTOR

(PART # SUMCOO00574). Weigh 2.5 lbs (1350 grams)

or using a 1000 ml graduated cylinder pour to 1000 ml

mark then pour that in Clean/Disinfect Tank then pour

another 350mls in graduated cylinder and pour that in

Clean/Disinfect Tank for total of 1350mls. Circulate for 15

minutes. After the 15 minute circulation is completed shut

R.O. off and disconnect the Clean/Disinfect Tank and

discard the MEMSTOR solution. Cap the Product and

Reject Ports.

(Male dust cap Reject or Drain EQTAFIO01865

Female dust cap EQTAFIO01864 for Pruduct ).

Preservation is good is good for 6 months. If longer than 6

month preservation is required the R.O. will have to be

rinsed for 2 hours then repacked again with BWI 3000

MEMSTOR for each 6 month period.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 61 of 99

Page 70: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 62 of 99

Page 71: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Shutdown

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Shutdown After the hemodialysis day is over, the R.O. may be shut-

down by changing the setting on only one switch.

To shut-down the Reverse Osmosis Machine (R.O.):

1. Ensure there are no further requirements for water for

hemodialysis or other systems used in your facility.

2. Move the OPERATE-FLUSH switch from the

OPERATE to the FLUSH position, and the RO will

enter into an automatic and scheduled Flush program.

The amber, Flush light will come on when the R.O.

actually starts-up and goes into flush at the pre-determined

time, or when manually cycled into the flush mode by the

24 hr. timer.

Central R.O. System Operation

Central R.O. System Operator Manual

NOTICE:

The R.O. is not completely turned-off. It is placed into an idle or

standby mode that is called the FLUSH Mode.

OPER. FLUSH

Man Central RO Operator's Rev Dec 2014 Page 63 of 99

Page 72: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 64 of 99

Page 73: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

SYSTEM

MAINTENANCE

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Maintenance Task Frequency

(More Often If Needed)

Notes

Check the System For Leaks

Daily

Visual Inspection

Monitor the System For Unusual

Sounds

Daily

Visual/Auditory

Clean External Surfaces

Weekly

Use Soft Damp Towel

or Sponge.

(DO NOT USE

BLEACH)

Record Operational Values

(i.e. Flows, Pressures,

Temperature, etc.)

Daily or more often as

required by facility

Record on Daily (Shift)

Checklist

Change Pre-Filter

Every 30 Days or when

∆P reaches or exceeds 15

psi

Disinfect The System

(membranes)

Every 30 Days

See Operator’s Manual

Disinfection and

Cleaning

.03 Pyrogen Final Filter

Every 6 months

Perform Chemical, Microbial and

Endotoxin Testing On Feed and

Product Water As Per AAMI

Requirements

Every 12 Months

(Chemical)- 30 Days

(Microbial) or more often

as required.

Submit Samples To A

Qualified Testing

Laboratory

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 65 of 99

Page 74: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

SYSTEM

MAINTENANCE

Con’t

Flush timer

adjustment

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

DIGITAL FLUSH TIMER

It is recommended that operators

verify the time and day at least

monthly. The digital flush timer is

equipped with a Lithium CR2032

battery. If the battery fails, it should

be replaced.

To replace the battery perform the following procedure:

1. Remove the 4 amp fuse (in black fuse holder)

inside the Control Box. {This will disconnect all

24 volt power from the Control Box.

2. Remove the Timer from the Control Box Back

Plate by removing the white plastic retainers

threaded onto the threaded studs in the corners of

the timer.

3. It is not necessary or recommended that the wires

be disconnected from the timer.

4. Gently pull the timer straight out, without putting

undue stain on the wires. Gently turn the timer so

the back is facing out.

5. The battery is located on the back of the timer.

6. With a coin or screwdriver, open the battery

compartment and remove the old battery.

7. Replace the battery (Lithium CR2032), taking

care to insert the new battery properly.

8. Replace the battery cover.

9. Gently put the timer back on the 4 retaining

threaded studs and replace the white plastic

threaded retainers.

10. Replace the 4 amp fuse.

WARNING

Electrical Hazard

The RO Control Box has 24 VAC

Man Central RO Operator's Rev Dec 2014 Page 66 of 99

Page 75: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

SYSTEM

MAINTENANCE

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

To set the digital timer to the correct day and time, perform

the following procedure:

1. Push and HOLD the CLOCK button,

2. Select the Day desired

3. Select the Hour desired

4. Select the minute desired

5. When set to correct day, hour, and minute,

Release the CLOCK Button.

6. NOTE: This Timer Does Not Have A Daylight

Savings Time feature. If you are in an area that

recognizes the time change twice a year, you

will need to reset the timer at each time change.

TO SET THE TIME FOR FLUSH MODE, perform the

following procedure:

1. Select the day or days for the RO to Flush

2. Push the TIMER button. The display should read

1 ON --:--

3. Options are single and multiple days

a) 7 days, Mo thru Su, (recommended

setting) b) Single Day, (any day of the week)

c) 5 days, Mo thru Su

d) 2 day week-end, Sa & Su

e) 6 days, Mo thru Sa

f) 3 days, Mo thru We

g) 3 days, Th thru Sa

h) Every other day, Mo, We, Fr

Man Central RO Operator's Rev Dec 2014 Page 67 of 99

Page 76: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

SYSTEM

MAINTENANCE

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

It is recommended that the timer be set to 7 days, Mo thru

Su.

Next Select hour and minute for the time the R.O is to Flush.

1. Push the TIMER button. The display should read

1 OFF --:--

2. Set this the same as above.

3. THE recommended MINIMUM Flush Time

should be set for no less than 15 minutes.

4. EXAMPLE:

1 ON 8:00am Mo Tu We Th FR SA SU

1 OFF 8:15am Mo Tu We Th FR SA SU

In this example, the R.O. is set to OPERATE/FLUSH mode.

At 8:00am Monday thru Sunday, the R.O. will go into Flush

cycle. At 8:15am Monday thru Sunday, the R.O. will shut

off the Flush cycle.

It is recommended that the R.O. should go into a Flush

cycle 6 times per day. (This covers 3 day/week operations

and 6 day/week operations when the R.O. is not in regular

use.)

Up to 8 time intervals can be set, but 6 is recommended.

It is also recommended that the R.O. not be set to flush

during regularly scheduled Backwash or regeneration of

Filters and Softeners. If the timer is set during these times,

the Interlock System will prevent the R.O. from running, (or

will shut down the R.O.). If the R.O. does not get a proper

and regular FLUSH, fouling of the R.O. membranes or Low

Pressure Alarms can occur.

Man Central RO Operator's Rev Dec 2014 Page 68 of 99

Page 77: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

SYSTEM

MAINTENANCE

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Manual Flush

This timer has a Manual Flush Feature. To activate the

manual Flush Feature:

1. The RO must be in FLUSH MODE (switch on

front of control box)

2. Push the MANUAL button on the flush timer

(with the R.O. set to FLUSH, on the outside/front

of the Control Panel)

3. The R.O. will start a Flush cycle. (This will be

indicated on the timer display and RED indicator

light on the timer)

4. The timer below the display will be: ON-AUTO-

OFF with a bar in the LCD display indicating

which mode is activated. When in Manual

Mode, the bar in the LCD will be above ON.

5. To terminate the Manual Flush Mode, push the

MANUAL button twice and the bar in the LCD

display will move to above the OFF.

6. Push the MANUAL button once more to set the

bar in the LCD display above AUTO.

7. (NOTE: if the bar in the LCD display is not above

AUTO, the R.O. will not flush automatically in

FLUSH MODE.

Man Central RO Operator's Rev Dec 2014 Page 69 of 99

Page 78: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

SYSTEM

MAINTENANCE

Con’t

Cleaning The RO

Membranes

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

On occasion, your R.O. Membranes will require cleaning

and disinfecting with chemical cleaners and disinfectants to

regain maximum performance.

Membrane fouling is indicated when:

• The Product Flow decreases and the Reject Flow

increases, and the two cannot be adjusted to design

specifications.

• The Pump Pressure increases, the Membrane Pressure

increases, and the Reject Pressure decreases.

• The Quality Monitor indicates a continuous decline in

water quality.

• Monthly disinfection is recommended.

This procedure covers the mechanical steps required to clean

and disinfect the Reverse Osmosis Machine Membranes.

The preparation of the chemical cleaning solution must be in

accordance with the specifications established for the

selected cleaning chemical.

The cleaning chemical must be handled in accordance with

its Material Safety Data Sheet (MSDS).

NOTICE:

Other factors, such as changes in the tap water pH, TDS, temperature,

and/or pressure can cause drastic changes in the overall performance of

the RO.

WARNING:

Chemical cleaners can cause serious injury or death.

CAUTION: This procedure should be performed by trained and qualified

people

Man Central RO Operator's Rev Dec 2014 Page 70 of 99

Page 79: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better Water LLC

CONSUMABLE PRODUCTS

The following is a list of consumable products that are associated with the

Central Reverse Osmosis Machine.

Please call (615) 355-6063 for current pricing.

20” Big Blue Filter SUCAOO00553

4x40 Membrane SUMEM008

.03 Filter, 20" SUCAPEC00575

.03 Filter (2020 R.O. only) SUCAPE00547

Minncare SUMCOO00575

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

www.betterwater.com

Man Central RO Operator's Rev Dec 2014 Page 71 of 99

Page 80: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better Water LLC

1% Test Strips SUMCOO00577

Residual Test Strips SUMCOO00576 (used to confirm rinsing)

Bw Acid Cleaner 1000 SUMCOO00572

BW ALKALINE CLEANER 2000 SUMCOO00571

NOTE: The label on all Minncare and BW products will have mixing directions.

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

www.betterwater.com

Man Central RO Operator's Rev Dec 2014 Page 72 of 99

Page 81: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Membrane Change Procedure

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Membrane Change Procedure Note: failure to follow these installation instructions can result in voiding the membrane warranty. 1: Remove the ⅜” Blue product tubing from the ⅜” x ½” Fast-n-tite. 2. Remove the ⅜” x ½” Fast-n-tite. 3: Loosen but do not remove the 2 screws from the 2 Stainless steel plates. 4: Remove the 2 Stainless steel plates. 5. It is suggested that a “T’ type tool be fabricated as illustrated to facilitate membrane removal. 6. Thread the tool into the end-cap until snug (do not over tighten) and pull up. 7. Remove the membrane. Use channel locks or a similar tool to grasp the membrane. Be sure to wear sterile gloves when handling new membranes.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 73 of 99

Page 82: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Membrane

Change

Procedure

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Replacing the O-rings.

It is recommended that when replacing the membrane, the

O-rings get replaced also. Follow the previous procedure to

remove the lower end-cap with the exception there is a ½”

plug that must be unscrewed from the end-cap. It is

recommended that medical grade glycerin be used to

lubricate all O-rings due to its water solubility. Simply

reverse the process to finish installing the O-rings and

membranes.

8. Re-tape the ⅜” x ½” Fast-n-tite with about 4 wraps of

Teflon tape.

9. Insert the ⅜” Blue product tubing into the ⅜” x ½”

Fast-n-tite.

10. Re-tape the ½” plug with about 4 wraps of Teflon tape.

11. Start up the R.O. and run to product to drain for 2 hours

before connecting the R.O. to the storage tank.

Man Central RO Operator's Rev Dec 2014 Page 74 of 99

Page 83: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Membrane

Change

Procedure

Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

EQFHOO00908

Stainless steel Plates Do not remove 2 screws

EQFHOO00906

End Cap

EQFHOO00907

Outer O-ring

PLFISS00868

⅜” x ½” Fast-n-tite

PLFI003

Hose barb

EQFHOO00910

Adapter

EQFHOO00911

Inner o-rings

Membrane

Call for P/N EQFHOO00905

Housing

Same as the

Top

Assembly

½” plug

Man Central RO Operator's Rev Dec 2014 Page 75 of 99

Page 84: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 76 of 99

Page 85: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Factory Reset

For The R.O.

Control Board

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Factory Re-Set

for the Reverse Osmosis Machine Control

Board

1. Turn off main power switch.

2. Open the Control Box and locate the 2 small black

square push button switches, labeled “CAL UP” &

“CAL DOWN”

3. Push in both black buttons simultaneously and hold

them in for 5 seconds, while holding both buttons in,

turn on the RO and hit the re-set button.

4. Release both black buttons, (after 5 seconds)

5. Turn off the R.O. and wait 5-10 seconds.

6. Turn on R.O.

7. This will reset the control board to Factory Default.

8. Close Control Box.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 77 of 99

Page 86: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 78 of 99

Page 87: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

OPTIONAL ASSEMBLY

At this time there are no Optional Assemblies available for the Central

Reverse Osmosis units.

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 79 of 99

Page 88: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 80 of 99

Page 89: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

System Trouble

Shooting

No Power To The RO

Power To RO But RO

Will Not Start

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

The information presented in this document is intended to

serve as a guide only for qualified operators. It is not all

inclusive of problems that may be encountered. This guide

should aid operators with reminders and routine trouble-

shooting tasks.

For any problem outside the confines of this guide, call for

technical assistance.

Possible Causes Possible Solutions Main Breaker Tripped

(In Building Electrical

Panel)

Reset Breaker

RO Not Plugged Into

Electrical

Receptacle.

Plug RO Power Cord Into Receptacle.

(The RO Is Equipped With A Hospital-Grade Twist-And-

Lock Plug. Make Sure That The Plug Is Seated And Locked

Into The Receptacle.)

Motor Starter Protector

Tripped.

Reset Motor Starter Protector In RO Junction Box.

Interlock Circuit

Interrupted Verify If Any Pre-Treatment Equipment (i.e. Carbon,

Softener) are in Backwash Or Regeneration. (The Interlock

Fault Indicator light will be illuminated.) Reservoir Water Level Is

Above Middle Float

Switch

This Is Normal. The RO Will Start When The Water Level In

The Reservoir Falls Below The Middle Float.

RO Is In An Alarm

Condition a. Identify The Specific Alarm By Observing The

Illuminated Alarm Light On The Front Of The RO

Control Panel.

b. Press The Red Reset Button On The RO Control Panel.

c. Reference Alarm Conditions Listed Below.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 81 of 99

Page 90: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

System Trouble

Shooting

Con’t

Power To RO But

RO Will Not Start

Continued

RO Runs

Intermittently

Incoming RO Feed

Water characteristics

Out Of Design

Specifications

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Possible Causes Possible Solutions

Switches Are In Incorrect Positions

For Normal Operation, Verify As Follows:

a. OPERATE-FLUSH Switch In The OPERATE Position

b. OPERATE-DISINFECT Keyed Switch In The

OPERATE Position

c. OPERATE-DISINFECT-OFF Switch In The

OPERATE Position

d. TANK-OFF-DIRECT Switch (Inside The RO Control

Box) In Either TANK Or DIRECT Position.

Faulty Interlock Circuit Connections

Check All Interlock Wiring And Connections.

Pre-Treatment Equipment Not

Operating To Design Specifications.

Check All Pre-Treatment Equipment To Verify Proper

Operation.

Source Water Has Changed From

Initial Design Water Analysis.

Have Source Water Analyzed By Qualified Laboratory And

Compare Results To The Initial Design Water Analysis.

If Source Water Characteristics Have Changed Significantly,

Changes May Be Required To Pre-Treatment Equipment.

Contact Technical Assistance.

Low Pressure Alarm Verify The City Booster Pump Is Operating Properly And

Providing Sufficient Pressure For RO Operation.

High Feed Temperature Alarm Verify The Blending Valve Is Set To 77o F At The RO. Adjust

As Needed.

High Membrane Pressure Alarm Press The Red ALARM RESET On The Control Panel.

If the alarm condition does not clear, call Better Water for

assistance.

High Product Alarm On Tank Feed Systems – Check The Product Hose For Kinks

Or Obstructions. Clear As Needed.

On Direct Feed Systems – Check The Pressure Regulating

Valve At The End Of The Loop (RO Return). Adjust As

Needed.

Poor Water Quality Alarm The Water Quality Has Fallen Below 90% Rejection (Pre-Set

At Factory)

When The Water Quality Reaches 95% Or More, The Alarm

Will Silence.

Man Central RO Operator's Rev Dec 2014 Page 82 of 99

Page 91: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

System Trouble

Shooting

Con’t

Alarm Conditions

RO Shuts Down In

Alarm Condition

(With Alarm Lights)

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Possible Causes Possible Solutions

An Alarm Condition Has Occurred.

Note The Alarm Indicated On The Panel

Perform The Corrective Action For The Specific Cause

Of The Alarm Condition Listed Above.

Press The Red ALARM RESET On The Control Panel.

If the alarm condition does not clear, call Better Water

for assistance.

Loss Of Power See Sections Related To Power Above

Motor Starter Protector Has Tripped Reset Motor Starter Protector In RO Junction Box.

Reservoir Is Full This Is Normal. The RO Will Start When The Water

Level In The Reservoir Falls Below The Middle Float.

Interlock Circuit Interrupted Verify If Any Pre-Treatment Equipment (i.e. Carbon,

Softener) are in Backwash Or Regeneration.

Alarm Lamp Is Burned Out Replace Bulb.

ALARM-MUTE Switch On The Level

Control Box Is In MUTE Position

Place ALARM-MUTE Switch In The ALARM

Position

Alarm Lamp Is Burned Out Replace Bulb.

Loose Connection In Remote Alarm

Wiring

Inspect All Wiring

Tank-Off-Direct Switch

It should be in the TANK position

High Level Switch Malfunction Gently tap the solenoid. This may clear any obstruction.

Defective Switch – Replace

Water Quality Monitor Not Within Set-

Points Of Water Quality Monitor.

Check the Water Quality Monitor Set-Points.

If this does not correct the problem, contact technical

support.

Faulty Water Quality Monitor Or Water

Quality Monitor Control Board

Contact Technical Support.

Product Purge Valve May Be In The

Open Position If the Water Quality is good, the valve is stuck open;

replace the valve.

If this does not correct the problem, contact technical

support.

Man Central RO Operator's Rev Dec 2014 Page 83 of 99

Page 92: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

System Trouble

Shooting

Con’t

RO Shuts Down

Without Alarming

RO Audible Alarm

Is Sounding But No

Lights Illuminated

Low Level Alarm

Does Not Sound

When Water Level

Falls Below Bottom

Float Switch

Lights Do Not

Illuminate In Alarm

Condition

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Possible Causes Possible Solutions Pre-Treatment Equipment Is Not

Operating Properly Check All Pre-Treatment Equipment To Verify

Proper Operation.

RO Requires Cleaning /

Disinfection

Reference Section – Cleaning And Disinfection

Membranes Need To Be Replaced

Call Technical Assistance To Verify

Filter Fouled Replace Filter Every 30 Days Or When ∆ Pressure Is

Equal Or Greater Than 15 PSI.

Filter Inlet Hose Is Kinked Or

Obstructed.

Check Hose. Remove Kinks Or Obstructions.

Faulty Gauge Verify The Accuracy Of The Gauge. Replace As

Needed.

Man Central RO Operator's Rev Dec 2014 Page 84 of 99

Page 93: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Example

Checklist

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

Example Checklist

The Daily Checklists are examples of what can be used

when evaluating a system and troubleshooting. They are

generic.

While many customers will use this checklist as it is

designed, some customers will choose to design their own

checklist, specific to the individual clinic.

Central R.O. System Operation

Central R.O. System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 85 of 99

Page 94: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 86 of 99

Page 95: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Example of CENTRAL R.O. SYSTEM QUALITY ASSURANCE CHECKLIST

ITEM(S TO BE CHECKED) MON

TUE

WED

THU

FRI

SAT

SUN

DATE

R.O. SWITCH POSITIONS:

Check Normal Operation TANK

Check Normal Operation DIRECT

R.O. FLUSHES Correctly

R.O. Goes into DISINFECT

R.O. LIGHTS & ALARMS:

Operate Light (ON)

High Feed Temp. (OFF)

Low Pressure shuts off at what psi?

High Product Pressure shuts off at what psi?

Poor Water Quality Monitor functions properly?

Interlock Fault Light (OFF)

FLOWMETERS:

Product Flow in GPM

Reject Flow in GPM

Recirculation Flow in GPM

WATER QUALITY METER:

Feed Water TDS (ppm)

Product Water TDS (ppm)

Rejection Rate (>90%)

Poor Water Quality Set-Point

R.O. PRESSURE GAUGES:

Pre-Filter Pressure

Membrane Pressure

Reject Pressure

Product Pressure

FLUSH TIMER:

Time R.O. is set to Flush

How Long is the R.O. set to Flush?

Set to Correct time and day?

INCOMING POWER

Adjusted to correct voltage?

L1 - L2

L1 - L3

L2 - L3

L1 - GROUND

L2 - GROUND

L3 - GROUND

L1 AMP Draw

L2 AMP Draw

L3 AMP Draw

Thermal Overload Setting

Interlock Circuit Tested?

Man Central RO Operator's Rev Dec 2014 Page 87 of 99

Page 96: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 88 of 99

Page 97: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Limited

Warranty

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

LIMITED WARRANTY TERMS and CONDITIONS a. This limited warranty is given only to the original buyer and covers the equipment delivered with this limited warranty. b. The buyer shall be barred from any recovery on this limited warranty or otherwise for damages due in whole or in part to… … unreasonable use … improper operation … use beyond normal fashion … failure to follow instructions … failure to maintain the product in good condition and repair … or the like. c. If the buyer discovers or should have discovered a defect in which it is reasonable to conclude that damage, either personal, property, or economic, may result, the buyer's continued use of the product shall constitute any assumption of risk by the buyer and a bar to any recovery for breach of this limited warranty or otherwise. d. No oral or written representation, information, or advice given by Better Water LLC or any of its representatives shall create a warranty or in any way increase the scope of this express limited warranty and shall not form a part of the basis for bargain.

WHAT IS WARRANTED AND FOR HOW LONG?

a. All equipment, excluding ion exchange and filtration media and cartridges, are warranted to be free from factory defects in materials, and workmanship under normal use for a period of one (1) year from the date of shipment. b. It is a condition precedent to recovery on this limited warranty that the buyer strictly comply with all operating and maintenance guidelines established by Better Water LLC and that the serial number (if applicable) is intact and legible on the equipment. c. It is a condition precedent to recovery on this limited warranty for damage to the external finish of the equipment that the buyer notifies Better Water LLC at the time of the installation that the finish is damaged.

WHAT IS REMEDY FOR BREACH OF THIS LIMITED

WARRANTY or NEGLIGENCE BY BETTER WATER LLC

a. Buyer's sole and exclusive remedy for any breach of this limited warranty or negligence by Better Water LLC shall be repair or replacement of the defective part, at the option of Better Water LLC, provided such defective part is returned to Better Water LLC for inspection. b. Better Water LLC shall not be obligated to supply an exact replacement of the defective part and reserves the right to substitute new and improved parts.

Central RO System Operation

Central RO System Operator Manual

Man Central RO Operator's Rev Dec 2014 Page 89 of 99

Page 98: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Limited

Warranty Con’t

Better Water LLC 698 Swan Drive Smyrna, TN 37167 Phone 615-355-6063 Fax 615-355-6065

www.betterwater.com

c. Better Water LLC shall provide at no cost to buyer, labor to remove and/or replace defective parts covered by this limited warranty for a period of ninety (90) days from the date of installation by Better Water LLC of the equipment. d. After such ninety (90) day period, buyer shall be responsible for any labor or service charge for the removal and/or replacement of any defective parts. e. Buyer shall be responsible for all travel expenses and freight charges at all times. f. Better Water LLC shall have no obligation to repair or replace any defective part if buyer fails to follow the procedure set forth in “HOW TO OBTAIN A REPLACEMENT PART UNDER LIMITED WARRANTY”. IN NO EVENT SHALL THIS LIMITED WARRANTY BE CONSTRUED TO COVER, NOR SHALL BETTER WATER LLC BE LIABLE TO BUYER AS ANY OTHER PERSON FOR, ANY CONSEQUENTIAL, INCEDENTAL, ECONOMIC, DIRECT, INDIRECT, GENERAL OR SPECIAL DAMAGES, WHICH ARE HEREBY EXPRESSLY DISCLAIMED.

HOW TO OBTAIN A REPLACEMENT PART UNDER LIMITED

WARRANTY a. Buyer should contact the Customer Service or Technical Support Departments and request a Return Goods Authorization. b. Described part(s) will be sent with a purchase order. c. The returned part(s) will be returned to the factory for limited warranty consideration. If part(s) are not covered under the limited warranty, part(s) will be considered billable against the purchase order supplied. WHAT IS NOT COVERED BY THIS LIMITED WARRANTY:

By way of example and not limitation, this limited warranty does not cover:

• Damage to or replacement of any ion exchange resin of filter media

• Labor or service charges for the removal and/or replacement of any defective parts after the ninety (90) day period from the date of installation or sale by Better Water LLC

• Freight charges and travel expenses

• Damage from inadequate or defective wiring, improper voltage, improper connections or electrical service, inadequate or defective plumbing, water supply, or water pressure, or in violation of applicable building, plumbing or electrical codes, laws, ordinances or regulations.

• Damage from improper installation or operation, including but not limited to, abuse, accident, neglect, improper maintenance, freezing and fires, or abnormal use.

• Damage caused by contaminants in Buyer’s water supply, including hardness, chlorine, chloramines, sulfur, bacterial iron, tannin, algae, oil, organic matter or other unusual substances, if special equipment has not been installed by Better Water LLC to remove such contaminants

• Damage to or caused by filters/membranes or other replacement parts not purchased from Better Water LLC or damage caused by modification, alteration, repair or service of the equipment or any of its parts by anyone other than Better Water LLC or its expressly authorized representatives.

Man Central RO Operator's Rev Dec 2014 Page 90 of 99

Page 99: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

BeBeBeBetter Watertter Watertter Watertter Water llc llc llc llc

Appendix 1

Central RO Technical

Service Bulletins

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 91 of 99

Page 100: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Relocation of Wire on Replacement VFDs TSB 2012009

Created/Last Revised Date 10/18/2012

Last Reviewed Date 10/18/2012

Page 1 of 1

OVERVIEW:

The manufacturer discontinued the model of the VFD used in Better Water LLC’s 2436 ROs and Pretreatment devices in October of 2012. The replacement model is functionally the same with the exception of the use of one of the terminal posts, which will require the relocation of a wire if the older model VFD is replaced with a newer model. Better Water LLC’s part number of EQMOBO01353 which is the “3 Phase-In / 3 Phase-Out / 3 Horsepower

VFD“ will remain the same. The difference between these two models is mostly cosmetic. See the pictures below of the old and new models.

WIRE RELOCATION:

If the older model VFD has to be replaced, do the following, using the before and after pictures as reference: 1. Take note of the location of the white wire (from the pressure transducer) on the older model VFD’s terminal

post labeled VIA.

2. Replace the older model VFD with the newer model VFD. 3. On the newer model VFD, relocate the white wire (from the pressure transducer) to the terminal post labeled

VIC.

TECHNICAL SERVICE BULLETIN

Old Model VFD

White/Cream color with hinged cover Front View

New Model VFD

Black color, no hinged cover Front View

Old Model VFD

White wire is located on the VIA

terminal post

New Model VFD

White wire is located on the VIC terminal post

Man Central RO Operator's Rev Dec 2014 Page 92 of 99

Page 101: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Relocation of Wire on Replacement VFDs TSB 2012010

Created/Last Revised Date 11/09/2012

Last Reviewed Date 11/09/2012

Page 1 of 1

OVERVIEW:

The manufacturer discontinued the model of the VFD used in Better Water LLC’s 3046 ROs and Post-treatment devices in October of 2012. The replacement model is functionally the same with the exception of the use of one of the terminal posts and the location of grounding posts, both requiring the relocation of wires if the old model VFD is replaced with the new model. Better Water LLC’s part number of EQMOBO01350 which is the “3 Phase-In / 3 Phase-Out / 5 Horsepower VFD“ will remain the same. The difference between these two models is mostly cosmetic. See the pictures below of the old and new models.

TERMINAL POST & GROUND WIRE RELOCATION:

1. Turn off the power, and unplug the device from the electrical outlet before opening the control box. 2. On the old model VFD, take note of the location of the wire (from the pressure transducer) on the terminal

post labeled VIA and where the three ground wires are located. 3. Replace the old model VFD with the new model VFD, reconnecting the wires with the following exceptions.

4. On the new model VFD, relocate the wire (from the pressure transducer) to the terminal post labeled VIC.

5. Relocate the three ground wires to the three grounding posts on the bottom of the new model VFD.

TECHNICAL SERVICE BULLETIN

Old Model VFD

White/Cream color with hinged cover Front View

New Model VFD

Black color, no hinged cover Front View

Old Model VFD

Wire located on terminal post VIA New Model VFD

Wire located on terminal post VIC

Old Model VFD

Three ground wires New Model VFD

Three grounding posts

Man Central RO Operator's Rev Dec 2014 Page 93 of 99

Page 102: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 94 of 99

Page 103: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Replacing Chromalox Temperature Controller with Omron Temperature Controller

TSB 2014002

Created/Last Revised Date 12/05/14

Last Reviewed Date 12/05/14

Page 1 of 3

OVERVIEW: Effective December 5, 2014, Better Water will start using the Omron Temperature Controller instead of the Chromalox Temperature Controller because of easier use and programmability. Prior to this, there have been two Chromalox models used. When replacing a Chromalox with an Omron, the only difference is the position of the connecting wires. The part number ELCPD100001 remains the same. The differences between these models and the wiring connection changes are detailed below. Refer to the pictures below to determine which model you have: If replacing and older model Chromalox model 1600 Temperature Controller you will also need to order the new model’s bracket whose part number is FXBROO00894, which is slightly larger. PROCEDURE: 1. Turn OFF power to the RO. 2. Open the cover to the Control Box. 3. Unclip the old model Chromalox Temperature Controller, and gently pull it forward so the attached wires can be accessed.

TECHNICAL SERVICE BULLETIN

WARNING To avoid electrical shock, turn the power to the RO OFF, and unplug it from the electrical outlet.

A lock-out tag should be placed on the unit to prevent accidental use while this Replacement is underway.

Chromalox model 1600 Part#

ELCPD100001

Chromalox model 6040 Part#

ELCPD100001

New Model Bracket Part#

FXBROO00894

New Omron model Part#

ELCPD100001

Man Central RO Operator's Rev Dec 2014 Page 95 of 99

Page 104: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Replacing Chromalox Temperature Controller with Omron Temperature Controller

TSB 2014002

Created/Last Revised Date 12/05/14

Last Reviewed Date 12/05/14

Page 2 of 3

4. Before releasing the attached wires make note of the individual wire numbers and which post-screws they are attached to on the old model, using the following schematic label for reference. 4. Carefully release each of the seven wires by loosening the post-screws they are attached to. 5. Once all the wires have been released, carefully remove the old model Chromalox Temperature Controller from the unit. 6. If replacing the bracket, remove the three screws that secure the old bracket. 7. Install the new bracket. 8. Each of the seven wires that were attached are individually numbered. Use the following diagram to attach the correct wire to the correct post-screws on the new Omron Temperature Controller. The schematic label on the side of the Controller should also be referenced to assist with the installation.

Chromalox Model 1600

Wire #1 to Post #1 Wire #2 to Post #2 Wire #3 to Post #3 Wire #4 to Post #4 Wire #5 to Post #5 Wire #9 to Post #9

Wire #10 to Post #10

Chromalox Model 6040

Wire #1 to Post #6 Wire #2 to Post #5 Wire #3 to Post #4

Wire #4 to Post #13 Wire #5 to Post #14 Wire #9 to Post #9

Wire #10 to Post #10

Omron Model Wire #1 to Post #4 Wire #2 to Post #5 Wire #3 to Post #6 Wire #4 to Post #1 Wire #5 to Post #2

Wire #9 to Post #11 Wire #10 to Post #12

Man Central RO Operator's Rev Dec 2014 Page 96 of 99

Page 105: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Replacing Chromalox Temperature Controller with Omron Temperature Controller

TSB 2014002

Created/Last Revised Date 12/05/14

Last Reviewed Date 12/05/14

Page 3 of 3

9. Install the new Omron Temperature Controller into the Control Box, by carefully pushing it into place and using the clip to properly secure it. 10. Turn ON the power to the RO. 11. Verify the new Omron Temperature Controller is working properly. 12. Close and secure the cover to the Control Box.

Man Central RO Operator's Rev Dec 2014 Page 97 of 99

Page 106: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

This Page Left Blank Intentionally

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 98 of 99

Page 107: Better Water LLC Central R.O. System Operation ·  · 2015-12-1823 .03 Pyrogen final ... Dimensions Approximately 39" W x 42” D x 77” H ... Central R.O. System Operator Manual

Better WaterBetter WaterBetter WaterBetter Water llc llc llc llc

CENTRAL R.O. TEST

VERIFICATION

AND RESULTS

Better Water LLC 698 Swan Drive, Smyrna, TN 37167

Ph. (615) 355-6063, Fax (615) 355-6065

WWW.BETTERWATER.COM

Man Central RO Operator's Rev Dec 2014 Page 99 of 99