bertan’602c’high’voltage’power’ supply’staon’explained’bohngrp/site/groupfiles/high...

7
Bertan 602C High Voltage Power Supply Sta9on Explained Loca9on: 323 S9nson Remick Hall Coupled with IX71 Epifluorescence Microscope

Upload: others

Post on 25-Jan-2021

0 views

Category:

Documents


0 download

TRANSCRIPT

  • Bertan  602C  High  Voltage  Power  Supply  Sta9on  Explained  

    Loca9on:  323  S9nson  Remick  Hall  Coupled  with  IX-‐71  Epifluorescence  

    Microscope  

  • 1  

    HV  supply  

    relay  box  switch  box  

    NI  BNC-‐2110  Card  +  Computer  

    2  

    3  

    4  

    5  

    6  

    7  

    9  10  

    8  

    Orange  line:  115V  AC      Red  line:  0-‐20kV  DC    Dashed  line:  0-‐5V  DC  (signaling  purposes)  

    Hardware  Configura9on  

  • 3kV  Connec9ons  Explained  

    1. TB1  Connec9on:  1-‐Line,  2-‐Neutral,  3-‐Ground  (115V  AC  IN)  2. MHV  (RF  coaxial)  Outlet  (0-‐3kV  DC  OUT)  3. Red  and  Black  (HV  and  Ground)  wire  inlet    4. Outlet  to  4  Pla9num  (Pt)  9p  electrodes    (HV  APPLIED  TO  MICROFLUIDICS  DEVICES)  5. Analog  Out  (ao1)  –  coaxial  out  (SENDS  SIGNAL  TO  CONTROL  HV  OUTPUT)  6. P2  Connec9ons:  3-‐GND,  4-‐PRGM  INPUT    7. Digital  and  Timing  I/O:  D  GND,  PO.0-‐PO.7  (SENDS  SIGNAL  TO  CONTROL  TIME  THAT  HV  IS  APPLIED  TO  EACH  Pt  ELECTRODE)  

    8. Serial  port  input  9. Serial  port  output  (SENDS  SIGNAL  TO  OPEN/CLOSE  SWITCHES  TO  Pt  ELECTRODES)  10. Serial  port  input    

  • 20kV  Connec9ons  Explained  

    1. TB1  Connec9on:  1-‐Line,  2-‐Neutral,  3-‐Ground  (115V  AC  IN)  2. A400Bxx  Outlet  (0-‐3kV  DC  OUT)  3. Red  and  Black  (HV  and  Ground)  wire  inlet    4. Outlet  to  4  Pla9num  (Pt)  9p  electrodes    (HV  APPLIED  TO  MICROFLUIDICS  DEVICES)  5. Analog  Out  (ao2)  –  coaxial  out  (SENDS  SIGNAL  TO  CONTROL  HV  OUTPUT)  6. P2  Connec9ons:  3-‐GND,  4-‐PRGM  INPUT    7. Digital  and  Timing  I/O:  D  GND,  PO.0-‐PO.7  (SENDS  SIGNAL  TO  CONTROL  TIME  THAT  HV  IS  APPLIED  TO  EACH  Pt  ELECTRODE)  

    8. Serial  port  input  9. Serial  port  output  (SENDS  SIGNAL  TO  OPEN/CLOSE  SWITCHES  TO  Pt  ELECTRODES)  10. Serial  port  input    

  • 12  7  

    5  

    2  

    1  

    8  

    Relay  Box  (serial  ports  10  &  13  on  back)    

    Switch  Box   Pt/Pd  Electrode    wires  (clear)  

    NI  BNC  2110  Card  

    9  

    Bertan  602C-‐30P,N  

  • Switch  Box  Outline  

         

         

         

         

         

         

         

         

    GND  1   HV  1   GND  2   HV  2  

    I   II   III   IV  

    1   2   3   4   5   6   7   8  

    GND1/Float/HV1   GND2/Float/HV2  

    LabView  program  controls  switches  1  through  8.  When  switch  is  closed  (electricity  is  poten9ally  flowing  to  electrodes)  and  the  light  on  the  relay  box  is  on.  

  • Switch  Scenarios  for  LabView  Program  

    Electrode   CondiWon   Switch  OperaWon  

    GND  1   Close  1  

    A   HV  1   Close  2  

    Float   1  &  2  Open  

    GND  1   Close  3  

    B   HV  1   Close  4  

    Float   3  &  4  Open  

    GND  2   Close  5  

    C   HV  2   Close  6  

    Float   5  &  6  Open  

    GND  2   Close  7  

    D   HV  2   Close  8  

    Float   7  &  8  Open