Basler acA645-100gm Camera Specification Measurement protocol using the EMVA Standard 1288 Document Number: BD000636 Version: Preliminary Release Date: June 21, 2012 PRELIMINARY VERSION

Upload: others

Post on 15-Sep-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

Basler acA645-100gm

Camera SpecificationMeasurement protocol using the EMVA Standard 1288

Document Number: BD000636

Version: Preliminary

Release Date: June 21, 2012

PRELIMINARY VERSION

Page 2: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

For customers in the U.S.A.

This equipment has been tested and found to comply with the limits for a Class A digital device,pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protec-tion against harmful interference when the equipment is operated in a commercial environment.This equipment generates, uses, and can radiate radio frequency energy and, if not installedand used in accordance with the instruction manual, may cause harmful interference to radiocommunications. Operation of this equipment in a residential area is likely to cause harmful in-terference in which case the user will be required to correct the interference at his own expense.

You are cautioned that any changes or modifications not expressly approved in this manualcould void your authority to operate this equipment.

The shielded interface cable recommended in this manual must be used with this equipment inorder to comply with the limits for a computing device pursuant to Subpart J of Part 15 of FCCRules.

For customers in Canada

This apparatus complies with the Class A limits for radio noise emissions set out in RadioInterference Regulations.

Pour utilisateurs au Canada

Cet appareil est conforme aux normes Classe A pour bruits radioelectriques, specifiees dans leReglement sur le brouillage radioelectrique.

Life Support Applications

These products are not designed for use in life support appliances, devices, or systems wheremalfunction of these products can reasonably be expected to result in personal injury. Baslercustomers using or selling these products for use in such applications do so at their own riskand agree to fully indemnify Basler for any damages resulting from such improper use or sale.

Warranty Note

Do not open the housing of the camera. The warranty becomes void if the housing is opened.

All material in this publication is subject to change without notice and is copyright BaslerVision Technologies.

Page 3: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

Contacting Basler Support Worldwide

Europe:

Basler AG

An der Strusbek 60 - 62

22926 Ahrensburg

Germany

Tel.: +49-4102-463-500

Fax.: +49-4102-463-599

[email protected]

Americas:

Basler, Inc.

855 Springdale Drive, Suite 160

Exton, PA 19341

U.S.A.

Tel.: +1-877-934-8472

Fax.: +1-610-280-7608

[email protected]

Asia:

Basler Asia Pte. Ltd

8 Boon Lay Way

# 03 - 03 Tradehub 21

Singapore 609964

Tel.: +65-6425-0472

Fax.: +65-6425-0473

[email protected]

www.baslerweb.com

Page 4: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European
Page 5: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION CONTENTS

Contents

1 Overview 7

2 Introduction 8

3 Basic Information 93.1 Illumination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

3.1.1 Illumination Setup for the Basler Camera Test Tool . . . . . . . . . 103.1.2 Measurement of the Irradiance . . . . . . . . . . . . . . . . . . . . 10

4 Characterizing Temporal Noise and Sensitivity 114.1 Basic Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

4.1.1 Total Quantum Efficiency . . . . . . . . . . . . . . . . . . . . . . . 114.1.2 Temporal Dark Noise . . . . . . . . . . . . . . . . . . . . . . . . . . 134.1.3 Dark Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144.1.4 Doubling Temperature . . . . . . . . . . . . . . . . . . . . . . . . . 144.1.5 Inverse of Overall System Gain . . . . . . . . . . . . . . . . . . . . 154.1.6 Inverse Photon Transfer . . . . . . . . . . . . . . . . . . . . . . . . 164.1.7 Saturation Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . 174.1.8 Spectrogram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184.1.9 Non-Whiteness Coefficient . . . . . . . . . . . . . . . . . . . . . . 21

4.2 Derived Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224.2.1 Absolute Sensitivity Threshold . . . . . . . . . . . . . . . . . . . . 224.2.2 Signal-to-noise Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . 234.2.3 Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

4.3 Raw Measurement Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264.3.1 Mean Gray Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264.3.2 Variance of the Temporal Distribution of Gray Values . . . . . . . . 274.3.3 Mean of the Gray Values Dark Signal . . . . . . . . . . . . . . . . 284.3.4 Variance of the Gray Value Temporal Distribution in Darkness . . . 294.3.5 Light Induced Variance of the Temporal Distribution of Gray Values 304.3.6 Light Induced Mean Gray Value . . . . . . . . . . . . . . . . . . . . 314.3.7 Dark Current Versus Housing Temperature . . . . . . . . . . . . . 32

5 Characterizing Total and Spatial Noise 335.1 Basic Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

5.1.1 Spatial Offset Noise . . . . . . . . . . . . . . . . . . . . . . . . . . 335.1.2 Spatial Gain Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . 345.1.3 Spectrogram Spatial Noise . . . . . . . . . . . . . . . . . . . . . . 355.1.4 Spatial Non-whiteness Coefficient . . . . . . . . . . . . . . . . . . 38

5.2 Raw Measurement Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 395.2.1 Standard Deviation of the Spatial Dark Noise . . . . . . . . . . . . 395.2.2 Light Induced Standard Deviation of the Spatial Noise . . . . . . . 40

Basler acA645-100gm 5

Page 6: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

CONTENTS PRELIMINARY VERSION

Bibliography 41

6 Basler acA645-100gm

Page 7: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 1 Overview

1 Overview

Basler acA645-100gm

Item Symbol Typ.1 Unit Remarks

Temporal Noise Parameters

Total Quantum Efficiency (QE) η 37 % λ = 545 nm

Inverse of Overall System Gain 1K 6.2 e−

DN

Temporal Dark Noise σd0 21 e−

Saturation Capacity µe.sat 24200 e−

Derived Parameters

Absolute Sensitivity Threshold µp.min 57 p∼ λ = 545 nm

Dynamic Range DYNout.bit 10.2 bit

Maximum SNR SNRy.max.bit 7.3 bit

SNRy.max.dB 43.8 dB

Item Symbol Typ. Unit Remarks

Spatial Noise Parameters

Spatial Offset Noise, DSNU1288 σo 6.0 e−

Spatial Gain Noise, PRNU1288 Sg 1.1 %

Table 1: Most Important Specification Data

Operating Point

Item Symbol Remarks

Video output format 12 bits/pixel

Gain Register raw 200

Offset Register raw 66

Exposure time Texp 20.0µs to 12.2ms

Table 2: Operating Point for the Camera Used

1The unit e− is used in this document as a statistically measured quantity.

Basler acA645-100gm 7

Page 8: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

2 Introduction PRELIMINARY VERSION

2 Introduction

This measurement protocol describes the specification of Basler acA645-100gm cam-eras. The measurement methods conform to the 1288 EMVA Standard, the Standardfor Characterization and Presentation of Specification Data for Image Sensors andCameras (Release A1.03) of the European Machine Vision Association (EMVA) [1].

The most important specification data for Basler acA645-100gm cameras is sum-marized in table 1.

8 Basler acA645-100gm

Page 9: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 3 Basic Information

3 Basic Information

Basic Information

Vendor Basler

Model acA645-100gm

Type of data presented Typical

Number of samples 70

Sensor Sony ICX414AL

Sensor type CCD

Sensor diagonal Diagonal 8 mm , Optical Size 1/2”

Indication of lens category to be used C-Mount

Resolution 659 x 490 pixel

Pixel width 9.90 µm

Pixel height 9.90 µm

Readout type Progressive scan

Transfer type Interline transfer

Shutter type -

Overlap capabilities Overlapping

Maximum readout rate 100 frames/second

General conventions -

Interface type Gigabit Ethernet

Table 3: Basic Information

Basler acA645-100gm 9

Page 10: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

3.1 Illumination PRELIMINARY VERSION

3.1 Illumination

3.1.1 Illumination Setup for the Basler Camera Test Tool

The illumination during the testing on each camera was fixed. The drift in the illumina-tion over a long period of time and after the lamp is changed is measured by a referenceBasler A602fc camera. The reference camera provides an intensity factor that was usedto calculate the irradiance for each camera measurement.

Light Source

Item Symbol Typ. Unit Remarks

Wavelength λ 545 nm

Wavelength Variation ∆λ 50 nm

Distance sensor to light source d 280 mm

Diameter of the light source D 35 mm

f-Number f# 8 f# = dD

Table 4: Light Source

3.1.2 Measurement of the Irradiance

The irradiance was measured using an IL1700 Radiometer from International Light Inc.(Detector: SEL033 #6285; Input optic: W #9461; Filter: F #21487; regular calibration).The accuracy of the Radiometer is specified as ±3.5%.

The measured irradiance is plotted in figure 1.

50x10-3

40

30

20

10

0

Irra

dian

ce [W

/m^2

]

706050403020100

Measurement

'acA645-100gm' (70 cameras), Irradiance

Figure 1: Irradiance for Each Camera Measurement.

The error for each calculated value using the amount of light falling on the sensor isdependent on the accuracy of the irradiance measurement.

10 Basler acA645-100gm

Page 11: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4 Characterizing Temporal Noise and Sensitivity

4 Characterizing Temporal Noise and Sensitivity

4.1 Basic Parameters

4.1.1 Total Quantum Efficiency

Total Quantum Efficiency for One Fixed Wavelength Total quantum efficiency η(λ)in [%] for monochrome light at λ = 545 nm with a wavelength variation of ∆λ = 50 nm.

40

30

20

10

0

Qua

ntum

Effi

cien

cy [%

]

706050403020100

Camera

'acA645-100gm' (70 cameras), Quantum Efficiency

16

14

12

10

8

6

4

2

0

Num

ber

38.538.037.537.036.536.0

Quantum Efficiency [%]

'acA645-100gm' (70 cameras), Quantum Efficiency Histogram

Figure 2: Total Quantum Efficiency (QE)

Item Symbol Typ. Std. Dev. Unit Remarks

Total Quantum Efficiency (QE) η 37 TBD % λ = 545 nm

Table 5: Total Quantum Efficiency (QE)

The main error in the total quantum efficiency ∆η is related to the error in the mea-surement of the illumination as described in section 3.1.

Basler acA645-100gm 11

Page 12: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.1 Basic Parameters PRELIMINARY VERSION

Total Quantum Efficiency Versus Wavelength of the Light Total quantum effi-ciency η(λ) in [%] for monochrome light versus wavelength of the light in [nm] .

50

40

30

20

10

0

Qua

ntum

Effi

cien

cy [%

]

1000900800700600500400

Wavelength [nm]

'acA645-100gm', Quantum Efficiency

Figure 3: Total Quantum Efficiency Versus Wavelength of the Light

The curve of the total quantum efficiency versus the wavelength as shown in figure3 was calculated from the single measured total quantum efficiency as presented insection 4.1.1. For the shape of the curve, the data from the sensor data sheet wasused.

12 Basler acA645-100gm

Page 13: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.1 Basic Parameters

4.1.2 Temporal Dark Noise

Standard deviation of the temporal dark noise σd0 referenced to electrons for exposuretime zero in [ e−].

25

20

15

10

5

0

Std

. Dev

. Tem

pora

l Dar

k N

oise

[e-]

706050403020100

Camera

'acA645-100gm' (70 cameras), Std. Dev. Temporal Dark Noise

20

15

10

5

0

Num

ber

22.021.521.020.5

Std. Dev. Temporal Dark Noise [e-]

'acA645-100gm' (70 cameras), Std. Dev. Temporal Dark Noise Histogram

Figure 4: Temporal Dark Noise

Item Symbol Typ. Std. Dev. Unit Remarks

Temporal Dark Noise σd0 21 0.4 e−

Table 6: Temporal Dark Noise

Basler acA645-100gm 13

Page 14: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.1 Basic Parameters PRELIMINARY VERSION

4.1.3 Dark Current

Dark current Nd30 for a housing temperature of 30◦ C in [e−/s] .Not measured!

4.1.4 Doubling Temperature

Doubling temperature kd of the dark current in [◦ C].Not measured!

14 Basler acA645-100gm

Page 15: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.1 Basic Parameters

4.1.5 Inverse of Overall System Gain

Inverse of overall system gain 1K

in [ e−DN

].

7

6

5

4

3

2

1

0

Inve

rse

of O

vera

ll S

yste

m G

ain

[e-/

DN

]

706050403020100

Camera

'acA645-100gm' (70 cameras), Inverse of Overall System gain

16

12

8

4

0

Num

ber

6.356.306.256.206.15

Inverse of Overall System Gain [e-/DN]

'acA645-100gm' (70 cameras), Inverse of Overall System Gain Histogram

Figure 5: Inverse of Overall System Gain

Item Symbol Typ. Std. Dev. Unit Remarks

Inverse of Overall System Gain 1K 6.2 0.06 e−

DN

Table 7: Inverse of Overall System Gain

Basler acA645-100gm 15

Page 16: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.1 Basic Parameters PRELIMINARY VERSION

4.1.6 Inverse Photon Transfer

Inverse photon transfer 1ηK

in[

p∼DN

].

16

12

8

4

0Inve

rse

Pho

ton

Tra

nsfe

r [p

~/D

N]

706050403020100

Camera

'acA645-100gm' (70 cameras), Inverse Photon Transfer

16

14

12

10

8

6

4

2

0

Num

ber

17.217.016.816.616.416.2

Inverse Photon Transfer [e-/DN]

'acA645-100gm' (70 cameras), Inverse Photon Transfer Histogram

Figure 6: Inverse Photon Transfer

Item Symbol Typ. Std. Dev. Unit Remarks

Inverse Photon Transfer 1ηK 16.8 TBD p∼

DN λ = 545 nm

Table 8: Inverse Photon Transfer

The main error in the inverse photon transfer 1ηK

is related to the error in the mea-surement of the illumination as described in section 3.1.

16 Basler acA645-100gm

Page 17: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.1 Basic Parameters

4.1.7 Saturation Capacity

Saturation capacity µe.sat referenced to electrons in [ e−].

25000

20000

15000

10000

5000

0

Sat

urat

ion

Cap

acity

[e-]

706050403020100

Camera

'acA645-100gm' (70 cameras), Saturation Capacity

16

14

12

10

8

6

4

2

0

Num

ber

2460024400242002400023800

Saturation Capacity [e-]

'acA645-100gm' (70 cameras), Saturation Capacity Histogram

Figure 7: Saturation Capacity

Item Symbol Typ. Std. Dev. Unit Remarks

Saturation Capacity µe.sat 24200 240 e−

Table 9: Saturation Capacity

Basler acA645-100gm 17

Page 18: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.1 Basic Parameters PRELIMINARY VERSION

4.1.8 Spectrogram

Spectrogram referenced to photons in [p∼] is plotted versus spatial frequency in [1/pixel]for no light, 50% saturation, and 90% saturation.

140

120

100

80

60

40

20

0

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), FFT for No Light

All Mean

5

6

7

8

9

100

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), FFT for No Light

All Mean

Figure 8: Spectrogram Referenced to Photons for No Light

18 Basler acA645-100gm

Page 19: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.1 Basic Parameters

7000

6000

5000

4000

3000

2000

1000

0

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), FFT for 50% Saturation

All Mean

3

456

1000

2

3

456

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), FFT for 50% Saturation

All Mean

Figure 9: Spectrogram Referenced to Photons for 50% Saturation

Basler acA645-100gm 19

Page 20: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.1 Basic Parameters PRELIMINARY VERSION

14000

12000

10000

8000

6000

4000

2000

0

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), FFT for 90% Saturation

All Mean

4

6

810

3

2

4

6

810

4

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), FFT for 90% Saturation

All Mean

Figure 10: Spectrogram Referenced to Photons for 90% Saturation

20 Basler acA645-100gm

Page 21: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.1 Basic Parameters

4.1.9 Non-Whiteness Coefficient

The non-whiteness coefficient is plotted versus the number of photons µp in [p∼] col-lected in a pixel during exposure time.

6

5

4

3

2

1

0

Non

Whi

tene

ss

800006000040000200000

Mean Photon [Photons/pixel]

'acA645-100gm' (70 cameras), Non Whiteness

Figure 11: Non-whiteness Coefficient

Basler acA645-100gm 21

Page 22: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.2 Derived Data PRELIMINARY VERSION

4.2 Derived Data

4.2.1 Absolute Sensitivity Threshold

Absolute sensitivity threshold µp.min(λ) in [ p∼] for monochrome light versus wavelengthof the light in [nm] .

µp.min =σd0

η(1)

70

60

50

40

30

20

10

0Abs

olut

e S

ensi

tivity

Thr

esho

ld [p

~]

706050403020100

Camera

'acA645-100gm' (70 cameras), Absolute Sensitivity Threshold

16

14

12

10

8

6

4

2

0

Num

ber

60595857565554

Absolute Sensitivity Threshold [p~]

'acA645-100gm' (70 cameras), Absolute Sensitivity Threshold Histogram

Figure 12: Absolute Sensitivity Threshold

Item Symbol Typ. Std. Dev. Unit Remarks

Absolute Sensitivity Threshold µp.min 57 TBD p∼ λ = 545 nm

Table 10: Absolute Sensitivity Threshold

22 Basler acA645-100gm

Page 23: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.2 Derived Data

4.2.2 Signal-to-noise Ratio

Signal-to-noise ratio SNRy(µp) is plotted versus number of photons µp collected in apixel during exposure time in [p∼] for monochrome light with the wavelength λ given in[ nm]. The wavelength should be near the maximum of the quantum efficiency.

A : SNRy =µy − µy.dark

σy

(2)

B : SNRy =ηµp√

(ηµp + σ2d0

)(3)

Figure 13 shows the signal-to-noise ratio SNRy for monochrome light with the wave-length λ = 545 nm.

8

6

4

2

0

SN

R [b

it]

1614121086420

Mean Photon [bit]

'acA645-100gm' (70 cameras), SNR

A B

Figure 13: Signal-to-noise Ratio

The maximum achievable image quality is given as SNRy.max .

SNRy.max =√

µe.sat (4)

SNRy.max.bit = ld SNRy.max =log SNRy.max

log 2(5)

SNRy.max.dB = 20 log SNRy.max ≈ 6.02 SNRy.max.bit (6)

Basler acA645-100gm 23

Page 24: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.2 Derived Data PRELIMINARY VERSION

1

2

46

10

2

46

100

2

4

SN

R

100

101

102

103

104

105

Mean Photon [Photons/pixel]

'acA645-100gm' (70 cameras), SNR

A B

Figure 14: Signal-to-noise Ratio

Item Symbol Typ. Std. Dev. Unit Remarks

Maximum achievable SNR [bit] SNRy.max.bit 7.3 0.01 bit

Table 11: Maximum achievable SNR [bit]

Item Symbol Typ. Std. Dev. Unit Remarks

Maximum achievable SNR [dB] SNRy.max.dB 43.8 0.04 dB

Table 12: Maximum achievable SNR [dB]

24 Basler acA645-100gm

Page 25: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.2 Derived Data

4.2.3 Dynamic Range

Dynamic range DYNout.bit in [ bit].

DYNout =µe.sat

σd0

(7)

DYNout.bit = log2 (DYNout) (8)

12

10

8

6

4

2

0

Out

put D

ynam

ic R

ange

[bit]

706050403020100

Camera

'acA645-100gm' (70 cameras), Output Dynamic Range

14

12

10

8

6

4

2

0

Num

ber

10.1810.1610.1410.1210.10

Output Dynamic Range [bit]

'acA645-100gm' (70 cameras), Output Dynamic Range Histogram

Figure 15: Output Dynamic Range

Item Symbol Typ. Std. Dev. Unit Remarks

Output Dynamic Range DYNout.bit 10.2 0.03 bit

Table 13: Output Dynamic Range

Basler acA645-100gm 25

Page 26: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.3 Raw Measurement Data PRELIMINARY VERSION

4.3 Raw Measurement Data

4.3.1 Mean Gray Value

Mean gray value µy(µp) in [DN] is plotted versus number of photons µp in [p∼] collectedin a pixel during exposure time.

5000

4000

3000

2000

1000

0

Mea

n G

ray

Val

ue B

right

[DN

]

800006000040000200000

Mean Photon [Photons/pixel]

'acA645-100gm' (70 cameras), Mean Gray Value Bright

Figure 16: Mean Gray Values of the Cameras with Illuminated Pixels

26 Basler acA645-100gm

Page 27: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.3 Raw Measurement Data

4.3.2 Variance of the Temporal Distribution of Gray Values

The variance of the temporal distribution of gray values σ2y.temp(µp) in [DN2] is plotted

versus number of photons µp in [p∼] collected in a pixel during exposure time.

700

600

500

400

300

200

100

0Var

ianc

e G

ray

Val

ue B

right

[DN

^2]

800006000040000200000

Mean Photon [Photons/pixel]

'acA645-100gm' (70 cameras), Variance Gray Value Bright

Figure 17: Variance Values for the Temporal Distribution of Gray Values with IlluminatedPixels

Saturation Capacity The saturation point is defined as the maximum of the curve infigure 17. The abscissa of the maximum point is the number of photons µp.sat where thecamera saturates. The saturation capacity µe.sat in electrons is computed according tothe mathematical model as:

µe.sat = ηµp.sat (9)

Basler acA645-100gm 27

Page 28: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.3 Raw Measurement Data PRELIMINARY VERSION

4.3.3 Mean of the Gray Values Dark Signal

Mean of the gray values dark signal µy.dark(Texp) in [DN] is plotted versus exposuretime in [s] .

16

12

8

4

0

Mea

n G

ray

Val

ue D

ark

[DN

]

121086420

Exposure Time [ms]

'acA645-100gm' (70 cameras), Mean Gray Value Dark

Figure 18: Mean Gray Values for the Cameras in Darkness

28 Basler acA645-100gm

Page 29: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.3 Raw Measurement Data

4.3.4 Variance of the Gray Value Temporal Distribution in Darkness

The variance of the temporal distribution of gray values in darkness σ2y.temp.dark(Texp) in

[DN2] is plotted versus exposure time Texp in [s] .

14

12

10

8

6

4

2

0Var

ianc

e G

ray

Val

ue D

ark

[DN

^2]

121086420

Exposure Time [ms]

'acA645-100gm' (70 cameras), Variance Gray Value Dark

Figure 19: Variance Values for the Temporal Distribution of Gray Values in Darkness

Temporal Dark Noise The dark noise for exposure time zero is found as the offset ofthe linear correspondence in figure 19. Match a line (with offset) to the linear part of thedata in the diagram. The dark noise for exposure time zero σ2

d0is found as the offset of

the line divided by the square of the overall system gain K.

σd0 =

√σ2

y.temp.dark(Texp = 0)

K2(10)

Basler acA645-100gm 29

Page 30: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.3 Raw Measurement Data PRELIMINARY VERSION

4.3.5 Light Induced Variance of the Temporal Distribution of Gray Values

The light induced variance of the temporal distribution of gray values in [DN2] is plottedversus light induced mean gray value in [DN] .

600

500

400

300

200

100

0

Var

ianc

e G

ray

Val

ue (

Brig

ht -

Dar

k) [D

N^2

]

3500300025002000150010005000

Mean Gray Value (Bright - Dark) [DN]

'acA645-100gm' (70 cameras), Diff. Variance vs Diff. Mean Gray Value

Figure 20: Light Induced Variance of the Temporal Distribution of Gray Values VersusLight Induced Mean Gray Value

Overall System Gain The overall system gain K is computed according to the math-ematical model as:

K =σ2

y.temp − σ2y.temp.dark

µy − µy.dark

(11)

which describes the linear correspondence in figure 20. Match a line starting at theorigin to the linear part of the data in this diagram. The slope of this line is the overallsystem gain K.

30 Basler acA645-100gm

Page 31: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 4.3 Raw Measurement Data

4.3.6 Light Induced Mean Gray Value

The light induced mean gray value µy − µy.dark in [ DN] is plotted versus the number ofphotons collected in a pixel during exposure time Kµp in [ p ∼].

4000

3500

3000

2500

2000

1500

1000

500

0

Mea

n G

ray

Val

ue (

Brig

ht -

Dar

k) [D

N]

6000050000400003000020000100000

Mean Photon [Photons/pixel]

'acA645-100gm' (70 cameras), Difference Mean Gray Value

Figure 21: Light Induced Mean Gray Value Versus the Number of Photons

Total Quantum Efficiency The total quantum efficiency η is computed according tothe mathematical model as:

η =µy − µy.dark

Kµp

(12)

which describes the linear correspondence in figure 21. Match a line starting at theorigin to the linear part of the data in this diagram. The slope of this line divided by theoverall system gain K yields the total quantum efficiency η.

The number of photons µp is calculated using the model for monochrome light. Thenumber of photons Φp collected in the geometric pixel per unit exposure time [p∼/s] isgiven by:

Φp =EAλ

hc(13)

with the irradiance E on the sensor surface [W/m2] , the area A of the (geometrical)pixel [m2] , the wavelength λ of light [m] , the Planck’s constant h ≈ 6.63 · 10−34 Js, andthe speed of light c ≈ 3 · 108 m/s. The number of photons can be calculated by:

µp = ΦpTexp (14)

during the exposure time Texp. Using equation 12 and the number of photons µp, thetotal quantum efficiency η can be calculated as:

η =hc

ATexp

1

E

1

λ

µp − µy.dark

K. (15)

Basler acA645-100gm 31

Page 32: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

4.3 Raw Measurement Data PRELIMINARY VERSION

4.3.7 Dark Current Versus Housing Temperature

The logarithm to the base 2 of the dark current in [e−/s] versus deviation of the housingtemperature from 30◦C in [ ◦ C]

Not measured!

32 Basler acA645-100gm

Page 33: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 5 Characterizing Total and Spatial Noise

5 Characterizing Total and Spatial Noise

5.1 Basic Parameters

5.1.1 Spatial Offset Noise

Standard deviation of the spatial offset noise σo referenced to electrons in [ e−].

7

6

5

4

3

2

1

0

DS

NU

1288

[e-]

706050403020100

Camera

'acA645-100gm' (70 cameras), DSNU1288

16

14

12

10

8

6

4

2

0

Num

ber

6.26.05.85.6

DSNU1288 [e-]

'acA645-100gm' (70 cameras), DSNU1288 Histogram

Figure 22: Spatial Offset Noise ( DSNU1288 )

Item Symbol Typ. Std. Dev. Unit Remarks

Spatial Offset Noise ( DSNU1288 ) σo 6.0 0.2 e−

Table 14: Spatial Offset Noise ( DSNU1288 )

Basler acA645-100gm 33

Page 34: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

5.1 Basic Parameters PRELIMINARY VERSION

5.1.2 Spatial Gain Noise

Standard deviation of the spatial gain noise Sg in [ %].

2.5

2.0

1.5

1.0

0.5

0.0

PR

NU

1288

[%]

706050403020100

Camera

'acA645-100gm' (70 cameras), PRNU1288

35

30

25

20

15

10

5

0

Num

ber

2.01.81.61.41.21.0

PRNU1288 [%]

'acA645-100gm' (70 cameras), PRNU1288 Histogram

Figure 23: Spatial Gain Noise ( PRNU1288 )

Item Symbol Typ. Std. Dev. Unit Remarks

Spatial Gain Noise ( PRNU1288 ) Sg 1.1 0.2 %

Table 15: Spatial Gain Noise ( PRNU1288 )

34 Basler acA645-100gm

Page 35: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 5.1 Basic Parameters

5.1.3 Spectrogram Spatial Noise

Spectrogram referenced to photons in [p∼] is plotted versus spatial frequency in [1/pixel]for no light, 50% saturation, and 90% saturation.

60

50

40

30

20

10

0

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), Spatial FFT for No Light

All Mean

2x101

3

4

5

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), Spatial FFT for No Light

All Mean

Figure 24: Spectrogram Referenced to Photons for No Light

Basler acA645-100gm 35

Page 36: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

5.1 Basic Parameters PRELIMINARY VERSION

7000

6000

5000

4000

3000

2000

1000

0

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), Spatial FFT for 50% Saturation

All Mean

100

2

4

68

1000

2

4

6

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), Spatial FFT for 50% Saturation

All Mean

Figure 25: Spectrogram Referenced to Photons for 50% Saturation

36 Basler acA645-100gm

Page 37: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 5.1 Basic Parameters

14000

12000

10000

8000

6000

4000

2000

0

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), Spatial FFT for 90% Saturation

All Mean

2

4

68

103

2

4

68

104

FF

T A

mpl

itude

[p~

]

0.50.40.30.20.10.0

Frequency [Cycles/pixel]

'acA645-100gm' (70 cameras), Spatial FFT for 90% Saturation

All Mean

Figure 26: Spectrogram Referenced to Photons for 90% Saturation

Basler acA645-100gm 37

Page 38: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

5.1 Basic Parameters PRELIMINARY VERSION

5.1.4 Spatial Non-whiteness Coefficient

The non-whiteness coefficient is plotted versus the number of photons µp in [p∼] col-lected in a pixel during exposure time.

20

15

10

5

0

Spa

tial N

on-W

hite

ness

6000050000400003000020000100000

Mean Photon [Photons/pixel]

'acA645-100gm' (70 cameras), Spatial Non-Whiteness

Figure 27: Spatial Non-whiteness Coefficient

38 Basler acA645-100gm

Page 39: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION 5.2 Raw Measurement Data

5.2 Raw Measurement Data

5.2.1 Standard Deviation of the Spatial Dark Noise

Standard deviation of the spatial dark noise in [DN] versus exposure time in [s] .

1.2

1.0

0.8

0.6

0.4

0.2

0.0

Spa

tial S

td. D

ev. G

ray

Val

ue D

ark

[DN

]

86420

Exposure Time [ms]

'acA645-100gm' (70 cameras), Spatial Std. Dev. Gray Value Dark

Figure 28: Standard Deviation of the Spatial Dark Noise

From the mathematical model, it follows that the variance of the spatial offsetnoise σ2

o should be constant and not dependent on the exposure time. Check that thedata in the figure 28 forms a flat line. Compute the mean of the values in the diagram.The mean divided by the conversion gain K gives the standard deviation of the spatialoffset noise σo .

DSNU1288 = σo =σy.spat.dark

K(16)

The square of the result equals the variance of the spatial offset noise σ2o .

Basler acA645-100gm 39

Page 40: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

5.2 Raw Measurement Data PRELIMINARY VERSION

5.2.2 Light Induced Standard Deviation of the Spatial Noise

Light induced standard deviation of the spatial noise in [DN] versus light induced meanof gray values [DN] .

80

70

60

50

40

30

20

10

0

Std

. Dev

. Gra

y V

alue

(B

right

- D

ark)

[DN

]

3500300025002000150010005000

Mean Gray Value (Bright - Dark) [DN]

'acA645-100gm' (70 cameras), Spatial Gain Noise

Figure 29: Light Induced Standard Deviation of the Spatial Noise

The variance coefficient of the spatial gain noise S2g or its standard deviation

value Sg respectively, is computed according to the mathematical model as:

PRNU1288 = Sg =

√σ2

y.spat − σ2y.spat.dark

µy − µy.dark

, (17)

which describes the linear correspondence in figure 29. Match a line through theorigin to the linear part of the data. The line’s slope equals the standard deviation valueof the spatial gain noise Sg .

40 Basler acA645-100gm

Page 41: Basler acA645-100gmapisc.com/basler/PDF/area/aca645-100gm.pdffor Characterization and Presentation of Specification Data for Image Sensors and Cameras (Release A1.03) of the European

PRELIMINARY VERSION REFERENCES

References

[1] EUROPEAN MACHINE VISION ASSOCIATION (EMVA): EMVA Standard 1288 - Stan-dard for Characterization and Presentation of Specification Data for Image Sensorsand Cameras (Release A1.03). 2006

Basler acA645-100gm 41