1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May 27, 2015 ABSTRACT We describe using external SExtractor (v2.8.6) catalogs from crclean.fits images to align ACS/WFC images with DrizzlePac/TweakReg. Note that this example was originally created before a more recent update to ACS/WFC geometric distortion files. At the time of this writing, one must follow the advice on the ACS Geometric Distortion web page as the first step in the process. By late 2015, as part of OPUS 2015.3, this part will be included by default in the standard pipeline processing and this will no longer need to be manually done by the user. We describe the rest of the process of preparing images for SExtractor, running SExtractor, and using the ouput catalogs to feed to the TweakReg task for alignment, and show that reasonably good first-cut results can be obtained with mostly default parameters in SExtractor and TweakReg. Better results may be possible with more exacting methods. This describes a method for quick alignment, not the ultimate best alignment. Note also that the use of crclean.fits images may be more suited to provide better results for ACS/WFC and WFC3/UVIS than for WFC3/IR. Introduction This example describes the alignment of four ACS/WFC images that contain a large number of extended sources, but relatively few point sources. Dithers of roughly 60 pixels between images were performed between the images, in order to obtain good sky coverage in the area of the gap between the two CCDs. Where possible, it is very important to start with images taken in the same visit and

Upload: others

Post on 11-Apr-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

1

Instrument Science Report ACS 2015-04

Basic Use of SExtractor Catalogs With TweakReg - I

Ray A. Lucas and Bryan Hilbert May 27, 2015

ABSTRACT We describe using external SExtractor (v2.8.6) catalogs from crclean.fits images to align ACS/WFC images with DrizzlePac/TweakReg. Note that this example was originally created before a more recent update to ACS/WFC geometric distortion files. At the time of this writing, one must follow the advice on the ACS Geometric Distortion web page as the first step in the process. By late 2015, as part of OPUS 2015.3, this part will be included by default in the standard pipeline processing and this will no longer need to be manually done by the user. We describe the rest of the process of preparing images for SExtractor, running SExtractor, and using the ouput catalogs to feed to the TweakReg task for alignment, and show that reasonably good first-cut results can be obtained with mostly default parameters in SExtractor and TweakReg. Better results may be possible with more exacting methods. This describes a method for quick alignment, not the ultimate best alignment. Note also that the use of crclean.fits images may be more suited to provide better results for ACS/WFC and WFC3/UVIS than for WFC3/IR.

Introduction                        This  example  describes  the  alignment  of  four  ACS/WFC  images  that  contain  a    large  number  of  extended  sources,  but  relatively  few  point  sources.  Dithers  of  roughly  60  pixels  between  images  were  performed  between  the  images,  in  order  to  obtain  good  sky  coverage  in  the  area  of  the  gap  between  the  two  CCDs.  Where  possible,  it  is  very  important  to  start  with  images  taken  in  the  same  visit  and  

Page 2: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

2

in  the  same  filter.    Also,  until  OPUS  2015.3  is  active,  to  use  the  latest  geometric  distortion  data,  follow  the  instructions  at  the  ACS  Geometric  distortion  page  at  http://www.stsci.edu/hst/acs/analysis/distortion    and  then  continue  as  below.      Summary of Subsequent Steps After Distortion Correction  

1. Examine  the  data  &  observing  program  structure;  note  same  visit/filter    2. Do initial single drizzle to produce *_crclean.fits files (cosmic rays removed) 3. Download and install SExtractor; edit a few parameters as needed 4. Run SExtractor to produce catalogs for use in TweakReg alignment 5. Visually inspect and clean up SExtractor catalogs, iterating if needed 6. Prepare control file for inputting SExtractor catalogs into TweakReg 7. Run TweakReg, iteratively, if needed, inputting the SExtractor catalogs 8. Proceed with alignment of other same visit/same filter combinations if

available, then use TweakBack to align all images to a common reference frame across all visits and filters, concluding with final drizzle, per filter, of all co-aligned frames.

(1) Examine the data and observing program structure  The Data (from a single visit of HST Program 10092):

Table 1 (above): Data parameters and dither pattern in the field of view (same visit & filter). The images have many extended sources (faint galaxies) but few stellar point sources for alignment, so the default DAOFIND-like ImageFindPars task does not generally work as well as it does on stellar-sources. Thus, we try using SExtractor-generated catalogs generated from *_crclean.fits images as input to TweakReg for determining shifts, updating headers, and aligning the images. Description of Data                              These  data  are  ACS/WFC  exposures  of  a  field  of  galaxies  from  a  single  visit  (0X)  of  program  10092.  Each  of  the  four  exposures  were  collected  through  the  F814W  filter  and  had  an  exposure  time  of  507  seconds.  The  images  all  shared    the      same  nominal  pointing  (RA:  150.20233  Dec:  2.6773222)  but  were  dithered,  with      large  (many  pixel)  steps  between  each.  The  sizes    of  the  steps  can  be  seen  in  the      POSTARG  columns  of  the  table  in  Table  1,  above,  keeping  in  mind  that  the  plate  scale  of  the  WFC  channel  is  0.05”  per  pixel,  implying  dither  steps  of  about  60  pixels  (3.1”/0.05”/pixel)  between  consecutive  images.  These  data  were  collected  on  15    May,  2004.  The  RA  and  Dec  of  Aperture  columns  in  Table  1  above  show  the  pointing  location  listed  in  the  header,  after  taking  the  POSTARGs  into  account.    

Page 3: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

3

   

Figure 1 (above): Dither pattern in the field of view (same visit & filter). North is up, East is to the left.  (2) Initial single drizzle, making *_crclean.fits images Second, run AstroDrizzle through the single drizzle step using default WCS for alignment during preliminary cosmic ray rejection and production of *_crclean.fits. In PyRAF, if using line commands rather than the Teal GUI interface, this is the syntax:

import drizzlepac from drizzlepac import astrodrizzle unlearn astrodrizzle astrodrizzle.AstroDrizzle(‘j8xi0x*flc.fits’,driz_cr_corr=yes,driz_combine=‘no’)

- This produces a CR-cleaned version of each input image called *_crclean.fits. It is the *_crclean.fits files on which you run SExtractor in this method. - The quality of the initial CR-rejection depends on the quality of the original WCS alignment, and this is best for images taken within the same visit and same filter.

Page 4: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

4

Image full of cosmic rays before CR cleaning (above) Image cleaned of cosmic rays except for strip along bottom of chip (below)

Figures 2a, 2b (above): Original and CR-cleaned images (3) Download and Install SExtractor Third, download and install SExtractor. We recommend the latest version (currently v2.8.6) for this purpose since it handles multi-extension FITS files better than some earlier versions. SExtractor may be downloaded from the web site referenced at the end of this example (or your institution may have a site-specific version for you to download), and you may follow instructions in the various guides or documentation referenced there, or with any site-specific version. Included with the download, in the

Page 5: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

5

/config subdirectory, are 4 files which you will need for running SExtractor – default.sex, default.param, default.conv, and default.nnw. In general, it is best to leave the originals in your defined SExtractor directory and just copy the 4 default.* files into your working directory for a given project, for editing as desired for the specific purposes of your work since SExtractor looks first in the working directory to see if there are any there. For this example, we used the default.nnw – a neural network file - and default.conv – a convolution kernel file - without any modifications, and only a few were made to default.sex, as listed below – mainly to do with parameters for object detection and measurement, and default.param, which is where we define the names and contents and format of our output catalogs, etc.

(4) Run SExtractor on the *_crclean.fits images Fourth, experimenting as needed, run SExtractor at the unix prompt. In this case, we used PSF seeing_fwhm=0.08 for ACS/WFC, detect_minarea=3, and detect_thresh=7. These values are set in the SExtractor default.sex file. Note that SExtractor completely ignores the initial extension of a multi-extension FITS file, and therefore the group numbers [0] & [3] in the example below correspond to the 1st and 4th extensions of the *_crclean.fits files, which are the science data extensions [1] & [4] of ACS/WFC data. For example: sex –catalog_name j8xi0xs0q_crclean_ext1.cat –checkimage_name j8xi0xs0q_crclean_ext1_aperturecheck.fits j8xi0xs0q_crclean.fits’[0]’ sex –catalog_name j8xi0xs0q_crclean_ext4.cat –checkimage_name j8xi0xs0q_crclean_ext4_aperturecheck.fits j8xi0xs0q_crclean.fits’[3]’ Repeat this for all of your *_crclean.fits images in this visit/filter combination. (5) Visually inspect results, and clean up catalogs if needed At this point, you may want to visually inspect your SExtractor results. To do this, you will need to display the images in ds9 with PyRAF line commands, and then overplot the catalog for various images with the tvmark task, both of which are also documented within PyRAF. In this example, there were a number of spurious sources (cosmic rays) which were not removed in the earlier step since that area of the CCD chip (a strip along the bottom) did not overlap any other images to which comparison could be made for automatic identification and removal of the cosmic rays, so these were then edited from the catalog by hand before they were fed to the tweakreg task for alignment of the images. An alternate, more automated solution might have been to use a larger detect_minarea in SExtractor, however, this, too can have its drawbacks since that will result in fewer objects, and may also yield a poorer solution in some cases, as it did in this one. It can be very data-dependent, so experimentation is key. Yet another way to do this is to mask out the affected area before running SExtractor, but you may want to use replacement values which are near to some mean global sky value if the sky background is not

Page 6: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

6

strongly variable since this can otherwise also affect the global sky level which SExtractor uses in object detection and measurement. SExtractor can also be set up to use local sky values around each object, but other factors come into play then. You may also want to try deleting from your catalogs troublesome objects which have high residuals as shown in the TweakReg plots, but that, too, can be tricky, and the size of an object’s residuals may vary some from image to image and catalog to catalog. In general, it is perhaps better to use other methods, such as scripted awk commands to remove such objects on the periphery or elsewhere, or within specified magnitude ranges or stellarity values, or ellipticity, or galaxy morphology type, etc. (Small elliptical galaxies may be better than galaxies with more complex morphology when having to compare across different bandpasses, for example.) However, each dataset is different, and you may have to try a number of different things and tailor your approach to the particular situation. (See Paper II – ACS ISR 2015-05 - by R. A. Lucas for more details.) A final hint: to view the maps that SExtractor created as checkimages, you may need to insert the SIMPLE = T header keyword in the checkimages before you can display and view them. (6) Prepare control file for inputting SExtractor catalogs to tweakreg To feed the SExtractor catalogs to TweakReg, you must create a list file that relates the images to be updated to the catalog file for each chip of each image. In this example, we created a file called astdriz_catfile.list, shown below. (The name doesn’t matter as long as it matches the file name for it in the ‘catfile’ parameter in TweakReg. In the automated version of TweakReg, it is called coords_catfile.lis.) On each line (and the order on the line is important), the file must contain:   1) the name of the *_flc.fits file, 2) the name of the SExtractor catalog file corresponding to data extension [1], and 3) the name of the SExtractor catalog file corresponding to data extension [4]. (The SExtractor catalog names are the ones from your SExtractor runs, and should be ones that make sense to you and reflect the fact that they represent ACS/WFC extensions [1] and [4] in that order on each line.)    In this case, our file called astdriz_catfile.list contains: j8xi0xs0q_flc.fits j8xi0xs0q_crclean_ext1.cat j8xi0xs0q_crclean_ext4.cat j8xi0xs3q_flc.fits j8xi0xs3q_crclean_ext1.cat j8xi0xs3q_crclean_ext4.cat j8xi0xs6q_flc.fits j8xi0xs6q_crclean_ext1.cat j8xi0xs6q_crclean_ext4.cat J8xi0xsaq_flc.fits j8xi0xsaq_crclean_ext1.cat j8xi0xsaq_crclean_ext4.cat (7) Run TweakReg (iteratively) using input catalogs from SExtractor Next, run TweakReg, telling TweakReg which columns in the SExtractor catalogs contained the x and y positions. (Those columns were determined earlier by the order in which you list them when editing your SExtractor default.param file. In this case, we used columns 5 and 6.) Run it first with updatehdr=no until you are satisfied (rms ~0.15,

Page 7: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

7

rms plots centered around 0) with a clear view that upper and lower bounds are not being artificially cut off by a sigma-clipping that is too aggressive in order to achieve an artificially lower rms. Finally, run it once more with updatehdr=yes in order to propagate the WCS updates into the headers of the input files. In the run below, we used default values of minobj=15, searchrad=1.0 (arcsec), nclip=3, sigma=3.0, fitgeometry=rscale, and x,y positions in units of pixels. Again, using line commands in PyRAF:

from drizzlepac import tweakreg unlearn tweakreg tweakreg.TweakReg(‘j8xi0x*flc.fits’,catfile=‘astdriz_catfile.list’,xcol=5,ycol=6)

TweakReg will produce output that looks like the following plots in Figures 3a, 3b, and 3c.:

Figure 3a: Histogram showing offsets associated with peak number of source matches.

Page 8: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

8

Figure 3b: Vector plot. Random values are preferable to strong trends.

Figure 3c: RMS residuals. RMS values of ~0.14, 0.14 are good; plots well-centered at ~0.0.

Page 9: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

9

(8) Proceed with further alignments of visit/filter groups (if available), then across visits and filters, concluding with all images from all visits and filters co-aligned and drizzled to a common reference frame Once you are happy with the alignment solution for that single visit and filter combination and have re-run tweakreg with updatehdr=yes, then you can continue by using astrodrizzle to drizzle together all of the images for that given visit and filter combination which you have aligned. Again, remember that the alignment process works best when starting with all images in the same visit taken in the same filter. Finally, if data from other visits and filters are available, repeat steps 4-7 for data from each filter within your visit. The TweakBack task can then be used to align the data from all of the filters in the visit, and to align and combine images taken across multiple visits and filters, drizzling them all to a common reference frame, as shown in Sections 7.5.5 and 7.5.6 in the DrizzlePac Handbook.

Conclusions We have shown that using mostly default parameter values in DrizzlePac and SExtractor tasks can result in reasonably good astrometric solutions, although there can still be room for improvement in most cases. However, in this example, we are addressing the most basic usage for a quick, rough alignment which may be good enough for some purposes.

Recommendations We recommend use of SExtractor version 2.8.6 since it is in general more user-friendly for use with tweakreg than earlier SExtractor versions, though even version 2.8.6 has some quirks, and other earlier versions can also be made to work reasonably well. See the AAS poster by R.A. Lucas (2012) from the Anchorage, Alaska AAS Meeting for more on the general methods, as well the poster by R.A. Lucas, B. Hilbert, and W. Hack (2013) from the Long Beach, California AAS Meeting for more details of a comparison between use of SExtractor version 2.5.0 vs SExtractor version 2.8.6.

Acknowledgements We thank our colleagues in the STScI DrizzlePac Working Group for useful discussions. References   SExtractor: SExtractor web site: http://www.astromatic.net/software/sextractor Bertin, E., & Arnouts, S., 1996: SExtractor: Software for Source Extraction, Astronomy & Astrophysics Supplement 317, 393

Page 10: Basic Use of SExtractor Catalogs With TweakReg - I · 1 Instrument Science Report ACS 2015-04 Basic Use of SExtractor Catalogs With TweakReg - I Ray A. Lucas and Bryan Hilbert May

10

Bertin, E., SExtractor v2.13 User’s Manual, at https://www.astromatic.net/pubsvn/software/sextractor/trunk/doc/sextractor.pdf Akhlaghi, M., 2012, Notes on SExtractor 2.8.6, at http://www.astr.tohoku.ac.jp/~akhlaghi/sextractor_notes.html Holwerda, B., SExtractor for Dummies, at http://astroa.physics.metu.edu.tr/MANUALS/sextractor/Guide2source_extractor.pdf DrizzlePac tasks and usage: Gonzaga et al., 2012, The DrizzlePac Handbook (Available on-line.) http://documents.stsci.edu/hst/HST_overview/documents/DrizzlePac/DrizzlePac.cover.html STScI DrizzlePac Web Site http://www.stsci.edu/hst/HST_overview/drizzlepac AAS Posters at DrizzlePac web site: Lucas, R.A., 2012, AstroDrizzle: Utilizing New CALACS Products, AAS-Anchorage, Alaska http://www.stsci.edu/hst/HST_overview/drizzlepac/resources/ral_aas-anchorage_2012_template-v7_Lucas.pdf Lucas, R.A., Hilbert, B., and Hack, W. 2013, Using SExtractor With the DrizzlePac Tweakreg Software for Aligning and Combining Sources in HST/ACS Images, AAS-Long Beach, California http://www.stsci.edu/hst/HST_overview/drizzlepac/resources/Lucas_AAS.pdf ISRs Lucas, R.A., 2015, Basic Use of SExtractor Catalogs With TweakReg – II, Instrument Science Report ACS 2015-05