Basic physiology of fMRI: signal and noise
![Page 2: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/2.jpg)
Overview
• Brief overview of fMRI • Contrast mechanisms
– Source of the fMRI signal
• Sources of Noise (& corrections) – Motion
– Cardiac
– Respiration
![Page 3: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/3.jpg)
Brain activity during cued finger tapping
![Page 4: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/4.jpg)
Measuring Brain Function with MRI
Task
Neuronal Activity
Red Blood Cells
![Page 5: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/5.jpg)
How can we see brain function with MRI?
red blood cells
oxygenated
deoxygenated
L. Pauling, C. D. Coryell, Proc.Natl. Acad. Sci. USA 22, 210-216, 1936.
Oxygenated and deoxygenated red blood cells have different magnetic properties
diamagnetic
paramagnetic
![Page 6: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/6.jpg)
Detecting “activation”
Capillaries!
Veins!Arteries!
Capillaries!
Veins!Arteries!
χ1 χ2 χ2 χ2
ΔS =
0
0.2
0.4
0.6
0.8
1
1.2
0 50 100 150 200
M0a e-t/T2*a - M0r e-t/T2*r
time
Sig
nal
ΔS S0
= M0a
M0r e-tΔR2* - 1
![Page 7: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/7.jpg)
Measuring Brain Function with MRI
Task
t
t
task Hemodynamics Neuronal Activity
Red Blood Cells
Task Blood Flow deoxy-Hb MRI signal
BOLD = Blood Oxygenation Level Dependent
![Page 8: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/8.jpg)
The Blood Oxygenation Level Dependent Response
Task
neuronal activity
BF
deoxy-Hb
MR signal
![Page 9: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/9.jpg)
The Blood Oxygenation Level Dependent Response
Task
neuronal activity
BF
HbO2 deoxy-Hb
MR signal
CMRO2 BV - - +
0 20 40 60
Time (sec)
Sig
nal
0 5 10 15 Time (sec)
Sig
nal
![Page 10: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/10.jpg)
Temporal dynamics of fMRI signal change
Delay
Activation amplitude
Arbitrary Scale
+ 2 sec
- 2 sec
0 sec 0 10 20 30 40 50
-2 -1.5 -1 -0.5 0 0.5 1 1.5 2
Num
ber o
f Vox
els
Delay (sec)
t
task fMRI signal
0 10 s
![Page 11: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/11.jpg)
Functional Magnetic Resonance Imaging (fMRI)
Typical MRI t
Functional MRI
Look for changes in MRI signal over time
Task Blood Flow MRI signal deoxy-Hb
![Page 12: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/12.jpg)
Brain activity during cued finger tapping
t
t
task
Map the changes in MRI signal over time
![Page 13: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/13.jpg)
Functional MRI
t
t
Task-related Neuronal Activity
Physiologic noise (cardiac, respiration)
Subject motion
Scanner instabilities Image reconstruction errors
Spontaneous Neuronal Activity
Hemodynamics resting state
![Page 14: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/14.jpg)
Problem: Subject Motion
motion
t
t
t
Signal changes occur at edges
t
![Page 15: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/15.jpg)
Correcting for subject motion
• Image registration
t = 0 t = 1
• Prevent Motion
• Regress out motion-related changes
![Page 16: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/16.jpg)
Problem: motion during the volume
![Page 17: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/17.jpg)
Problem: motion during the volume
slices are actually acquired like this… expected acquisition:
![Page 18: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/18.jpg)
Problem: motion during the volume
Potential solution: Prospective motion correction
slice tracks with the motion
• C.C. Lee, et al., Magn. Res. Med., 39, 1998. • H.A. Ward, et al., Magn. Res. Med., 43, 2000. • O. Speck, et al., Magma 19, 2006.
![Page 19: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/19.jpg)
Physiological noise
0 40 80 120 160 200 240 280 320 time (s)
Cardiac
M.S. Dagli et al., NeuroImage 9, 1999
Respiration
time 00 TIME = 110 s
ΔΦ =90 o 400
200
B0
Field Map
Figure courtesy of J. Bodurka
![Page 20: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/20.jpg)
Can filtering (<0.1Hz) reduce cardiac noise?
0 5 10 15 20 time (s)
Sign
al
Cardiac fluctuation (1.1 Hz) Sampled every TR (2s)
0 0.5 1 1.5 2 Frequency (Hz)
0
0.1
0.2
0.3
Pow
er
• Not always: Aliasing
![Page 21: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/21.jpg)
Correction of physiological noise
RETROICOR (G. Glover et al., Magn. Reson. Med. 44, 2000.)
Pulse Oximeter
0 1 2 3 4 5 6 7 8 9 10 time (s)
Respiration
MR Images
φc
φr
sin( φc )
cos( φc )
sin( 2φc )
cos( 2φc )
sin( φr )
cos( φr )
sin( 2φr )
cos( 2φr )
Additional Regressors:
![Page 22: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/22.jpg)
Correction of physiological noise
time
Sig
nal
Reshuffle the data based on its cardiac or respiration phase
0 1 2 3 4 5 6
Sig
nal
phase
![Page 23: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/23.jpg)
Correction: Cardiac + Respiration
X. Hu et al., Magn. Res. Med. 1995
![Page 24: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/24.jpg)
At what stage should physio correction be done?
• Preprocessing – Motion correction – Physiological noise correction – Slice time correction – Nuisance Regression – Spatial Smoothing – Convert to percent signal change – Temporal filter
• Define ROI (seed) • Average EPI time course over ROI • Regression T.B. Jones et al., Neuroimage 42, 2008
![Page 25: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/25.jpg)
Physio. fluctations can occur at low frequencies
Variations in breathing during rest ΔBOLD signal
PETCO2
BOLD (GM)
BOLD (WM)
R.G. Wise et al., NeuroImage 21, 2004
ΔPETCO2 correlated w/ ΔBOLD
![Page 26: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/26.jpg)
Resting fluctuations in respiration
RVT = Respiration Volume per Time
50 100 150 200 250 300 350
0
6
|Z|
group (n=11)
RVT-related changes
RVT
![Page 27: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/27.jpg)
Heart Rate
Birn et al. (unpublished)
![Page 28: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/28.jpg)
RVHRcor (Respiration Volume + Heart Rate)
C. Chang et al., Neuroimage 2009
![Page 29: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/29.jpg)
Nuisance Regression
H.J. Jo et al., NeuroImage 52, 2010
![Page 30: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/30.jpg)
The effect of global signal regression
The global signal is highly correlated w/ RVT
-1
-0.5
0
0.5
1
CC
: RV
T ~
glob
al s
igna
l
without global regression
with global regression
Connectivity maps
0.5
0.3
-0.1
0
(separate study, n = 18)
![Page 31: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/31.jpg)
Global Signal Regression
K. Murphy, et al., Neuroimage, 2008
Can introduce anti-correlations
![Page 32: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/32.jpg)
Global Signal Regression
Z.S. Saad, et al., Brain Connectivity, 2012
Can alter group differences …under certain conditions
![Page 33: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/33.jpg)
Nuisance Regression
H.J. Jo et al., NeuroImage 52, 2010
![Page 34: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/34.jpg)
PSTCor (Phase-shifted Soft Tissue Corr.)
J.S. Anderson et al., HBM 32, 2011
![Page 35: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/35.jpg)
“FIX” (FMRIB’s ICA-based X-noisifier)
L. Griffanti, et al., Neuroimage 2014
![Page 36: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/36.jpg)
ICA-AROMA
![Page 37: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/37.jpg)
ICA-AROMA
R.H.R. Pruim et al., NeuroImage 112 (2015)
![Page 38: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/38.jpg)
Tools to correct for physiological noise
• Filtering - Works only if TR is short enough (< 400ms)
• IMPACT (Chuang et al., 2001) – if TR short enough
• Retrospective correction (k-space) (X. Hu et al., 1995)
• RETROICOR (Glover et al., 2000)
• CORSICA (V. Pelbarg, et al., 2007)
• PESTICA (E.B. Beall, et al., 2007)
• RVTcor (R.M. Birn, et al., 2006) • RVHRcor (C. Chang, et al., 2009) • ANATICOR (H-J. Jo, et al., ) • APPLECOR, PEARCOR (M Marx, et al., 2013) • PSTCor (J.S. Anderson, et al., 2011) • CompCor (Y. Behzadi, et al., 2007)
• FIX (L. Griffanti, et al., 2014)
• ICA-AROMA (###, et al., 2015)
![Page 39: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/39.jpg)
Should we perform physio corrections?
• Reduces fluctuations related to heart beat and respiration – Reduce false positives – Reduce false negatives
• Some physiological fluctuations are
associated with neuronal activity – E.g. Heart rate variability
This is still an open question in the field
![Page 40: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/40.jpg)
Signal / Physiologic Noise
Tem
pora
l Sig
nal t
o N
oise
Rat
io
TSNR vs SNR
Signal-to-Noise Ratio
![Page 41: Basic physiology of fMRI: signal and noise Materials...(cardiac, respiration) Subject motion Scanner instabilities Image reconstruction errors Spontaneous Neuronal ... R.G. Wise et](https://reader036.vdocuments.us/reader036/viewer/2022063016/5fd609c361d21146d321b3c8/html5/thumbnails/41.jpg)
J. Bodurka
TSNR vs SNR