Basic chemistry, POC, November, 4-2018 xxxxx Copyright © Labquality Oy Labquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to Labquality EQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents, testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading. Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unless otherwise agreed. 17.12.2018 1/22 Round Sample x pt Result diff% z-score 18/4 Sample S002 142.1 143.0 0.6% 0.74 18/4 Sample S001 161.4 162.0 0.4% 0.52 18/3 Sample S002 142.1 144.0 1.4% 1.31 18/3 Sample S001 165.2 168.0 1.7% 1.13 18/2 Sample S002 164.9 167.0 1.3% 1.25 18/2 Sample S001 141.0 142.0 0.7% 1.33 18/1 Sample S002 140.2 140.0 -0.2% -0.14 18/1 Sample S001 165.8 167.0 0.7% 0.61 17/4 Sample S002 141.5 143.0 1.1% 0.99 17/4 Sample S001 134.6 135.0 0.3% 0.27 17/3 Sample S002 138.5 136.0 -1.8% -1.41 17/3 Sample S001 141.2 139.0 -1.5% -1.24 x pt sd SEM CV% n Vitros 250-5600, 5.1.FS 142.1 mmol/l 1.2 0.5 0.8 6 All methods 141.9 mmol/l 0.9 0.3 0.7 11 x pt sd SEM CV% n Vitros 250-5600, 5.1.FS 161.4 mmol/l 1.1 0.4 0.7 6 All methods 156.8 mmol/l 5.5 1.7 3.5 11 Sodium |VITROS 5,1 FS

Upload: others

Post on 23-Mar-2022

0 views

Category:

Documents


0 download

TRANSCRIPT

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 1/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 142.1 143.0 0.6% 0.7418/4 Sample S001 161.4 162.0 0.4% 0.5218/3 Sample S002 142.1 144.0 1.4% 1.3118/3 Sample S001 165.2 168.0 1.7% 1.1318/2 Sample S002 164.9 167.0 1.3% 1.2518/2 Sample S001 141.0 142.0 0.7% 1.3318/1 Sample S002 140.2 140.0 -0.2% -0.1418/1 Sample S001 165.8 167.0 0.7% 0.6117/4 Sample S002 141.5 143.0 1.1% 0.9917/4 Sample S001 134.6 135.0 0.3% 0.2717/3 Sample S002 138.5 136.0 -1.8% -1.4117/3 Sample S001 141.2 139.0 -1.5% -1.24

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

142.1mmol/l

1.2 0.5 0.8 6

All methods 141.9mmol/l

0.9 0.3 0.7 11

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

161.4mmol/l

1.1 0.4 0.7 6

All methods 156.8mmol/l

5.5 1.7 3.5 11

Sodium |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 2/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 4.77 4.81 0.77% 0.7918/4 Sample S001 6.26 6.31 0.83% 1.0018/3 Sample S002 6.76 6.85 1.38% 1.3718/3 Sample S001 6.38 6.45 1.15% 1.2518/2 Sample S002 6.30 6.37 1.18% 1.3018/2 Sample S001 6.84 6.84 -0.06% -0.0418/1 Sample S002 4.70 4.82 2.55% 1.3318/1 Sample S001 6.24 6.40 2.56% 1.5617/4 Sample S002 6.89 6.85 -0.63% -0.5817/4 Sample S001 3.92 3.86 -1.44% -1.0217/3 Sample S002 7.02 6.92 -1.37% -0.8517/3 Sample S001 4.80 4.76 -0.75% -0.45

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

4.77mmol/l

0.05 0.02 1.0 6

All methods 4.69mmol/l

0.11 0.03 2.4 12

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

6.26mmol/l

0.05 0.02 0.8 6

All methods 6.08mmol/l

0.20 0.06 3.3 12

Potassium |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 3/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 103.6 102.0 -1.5% -1.4018/4 Sample S001 129.5 129.6 0.1% 0.0518/3 Sample S002 101.5 100.4 -1.1% -1.0518/3 Sample S001 134.9 134.7 -0.2% -0.1718/2 Sample S002 134.1 134.3 0.2% -18/2 Sample S001 100.8 100.2 -0.6% -18/1 Sample S002 101.9 100.5 -1.4% -1.0418/1 Sample S001 129.0 127.2 -1.4% -0.9917/4 Sample S002 100.9 101.1 0.2% 0.0917/4 Sample S001 113.7 113.6 -0.1% -0.0617/3 Sample S002 102.3 103.0 0.7% 0.5817/3 Sample S001 101.1 102.3 1.2% 1.04

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

103.6mmol/l

1.1 0.5 1.1 5

All methods 105.0mmol/l

2.4 0.8 2.3 9

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

129.5mmol/l

1.5 0.7 1.2 5

All methods 128.1mmol/l

2.2 0.7 1.7 9

Chloride |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 4/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 5.54 5.60 1.06% 0.7518/4 Sample S001 14.75 14.67 -0.55% -0.3218/3 Sample S002 3.13 3.19 1.87% 1.1918/3 Sample S001 14.63 14.95 2.20% 1.3818/2 Sample S002 14.93 15.57 4.27% 1.6818/2 Sample S001 2.77 2.86 3.39% 0.8018/1 Sample S002 5.28 5.43 2.82% 1.6718/1 Sample S001 14.74 15.12 2.59% 1.2017/4 Sample S002 2.74 2.86 4.26% 1.7617/4 Sample S001 3.26 3.36 3.15% 1.4317/3 Sample S002 6.18 6.10 -1.28% -0.9417/3 Sample S001 4.90 4.82 -1.72% -1.32

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

5.54mmol/l

0.08 0.03 1.4 6

All methods 5.39mmol/l

0.26 0.07 4.9 17

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

14.75mmol/l

0.25 0.10 1.7 6

All methods 14.88mmol/l

1.01 0.25 6.8 17

Glucose |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 5/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 6.25 6.50 4.07% 1.5418/4 Sample S001 17.12 17.49 2.19% 1.2118/3 Sample S002 5.61 5.66 0.97% 0.3818/3 Sample S001 17.38 17.66 1.61% 0.7518/2 Sample S002 17.37 17.49 0.69% 0.3118/2 Sample S001 6.23 6.33 1.62% 0.5318/1 Sample S002 6.07 5.66 -6.80% -1.7518/1 Sample S001 17.05 16.49 -3.30% -1.4617/4 Sample S002 6.15 6.00 -2.48% -0.5417/4 Sample S001 5.25 5.16 -1.67% -0.2817/3 Sample S002 5.18 5.16 -0.38% -0.1117/3 Sample S001 5.62 5.66 0.72% 0.50

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

6.25mmol/l

0.17 0.07 2.7 6

All methods 6.22mmol/l

0.17 0.05 2.8 11

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

17.12mmol/l

0.31 0.13 1.8 6

All methods 18.83mmol/l

2.01 0.60 10.7 11

Urea |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 6/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 67.24 69.60 3.51% 1.2818/4 Sample S001 96.37 98.60 2.32% 1.6618/3 Sample S002 71.06 72.10 1.46% 0.9818/3 Sample S001 91.62 92.70 1.18% 0.4018/2 Sample S002 90.43 91.40 1.08% -18/2 Sample S001 69.43 71.60 3.13% -18/1 Sample S002 65.76 66.00 0.36% 0.2018/1 Sample S001 96.44 96.10 -0.35% -0.2317/4 Sample S002 70.69 70.80 0.16% 0.0617/4 Sample S001 55.78 55.30 -0.85% -0.2917/3 Sample S002 70.08 68.20 -2.68% -1.4717/3 Sample S001 68.58 67.10 -2.16% -1.24

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 67.24 g/l 1.84 0.82 2.7 5All methods 66.88 g/l 1.79 0.68 2.7 7

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 96.37 g/l 1.35 0.60 1.4 5All methods 95.60 g/l 1.87 0.71 2.0 7

Protein |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 7/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 2.40 2.45 1.95% 1.1218/4 Sample S001 3.19 3.27 2.57% 1.5918/3 Sample S002 2.47 2.54 2.71% 1.2718/3 Sample S001 3.24 3.32 2.63% 1.4718/2 Sample S002 3.24 3.31 2.08% -18/2 Sample S001 2.46 2.51 2.14% -18/1 Sample S002 2.36 2.40 1.82% 1.2718/1 Sample S001 3.18 3.23 1.52% 0.8917/4 Sample S002 2.43 2.44 0.26% 0.1217/4 Sample S001 2.39 2.37 -0.90% -0.6817/3 Sample S002 2.37 2.40 1.42% 0.4517/3 Sample S001 2.37 2.40 1.16% 0.31

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

2.40mmol/l

0.04 0.02 1.7 5

All methods 2.35mmol/l

0.07 0.02 3.0 9

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

3.19mmol/l

0.05 0.02 1.6 5

All methods 3.13mmol/l

0.08 0.03 2.5 9

Calcium |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 8/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 0.82 0.83 1.69% 0.5218/4 Sample S001 1.52 1.53 0.90% 0.4618/3 Sample S002 0.84 0.83 -1.45% -0.2118/3 Sample S001 1.51 1.50 -0.55% -0.1918/2 Sample S002 1.56 1.54 -1.31% -18/2 Sample S001 0.90 0.88 -2.09% -18/1 Sample S002 0.79 0.76 -3.75% -1.3918/1 Sample S001 1.51 1.43 -5.61% -1.4217/4 Sample S002 0.88 0.83 -5.82% -1.4217/4 Sample S001 0.79 0.74 -6.90% -1.7917/3 Sample S002 0.86 0.84 -2.44% -0.4217/3 Sample S001 0.80 0.76 -5.06% -1.41

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

0.82mmol/l

0.03 0.01 3.3 5

All methods 0.82mmol/l

0.03 <0.01 3.4 9

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

1.52mmol/l

0.03 0.01 1.9 5

All methods 1.53mmol/l

0.05 0.02 3.2 9

Magnesium |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 9/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 348.98 349.98 0.29% 0.0918/4 Sample S001 576.54 571.99 -0.79% -0.4218/3 Sample S002 365.67 358.31 -2.01% -1.1318/3 Sample S001 574.90 570.20 -0.82% -0.3518/2 Sample S002 587.72 589.84 0.36% -18/2 Sample S001 364.91 360.69 -1.16% -18/1 Sample S002 348.03 351.76 1.07% 0.3118/1 Sample S001 575.99 578.53 0.44% 0.1717/4 Sample S002 357.94 352.36 -1.56% -0.5017/4 Sample S001 293.48 291.05 -0.83% -0.2617/3 Sample S002 330.58 332.72 0.65% 0.3017/3 Sample S001 357.30 361.29 1.12% 0.35

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

348.98µmol/l

11.12 4.97 3.2 5

All methods 353.86µmol/l

14.74 4.44 4.2 11

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

576.54µmol/l

10.87 4.86 1.9 5

All methods 582.04µmol/l

22.67 6.84 3.9 11

Uric acid |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 10/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 4.22 4.19 -0.62% -0.3318/4 Sample S001 6.15 6.20 0.79% 0.4618/3 Sample S002 4.34 4.24 -2.23% -0.7518/3 Sample S001 5.62 5.61 -0.20% -0.0618/2 Sample S002 5.64 5.32 -5.64% -18/2 Sample S001 4.18 4.08 -2.40% -18/1 Sample S002 4.20 4.16 -1.05% -0.6018/1 Sample S001 5.95 5.81 -2.42% -0.5017/4 Sample S002 4.15 4.11 -0.86% -0.2617/4 Sample S001 3.53 3.51 -0.58% -0.1617/3 Sample S002 4.14 4.19 1.16% 0.4417/3 Sample S001 4.43 4.39 -0.80% -0.24

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

4.22mmol/l

0.08 0.04 1.9 5

All methods 4.01mmol/l

0.40 0.04 9.9 85

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

6.15mmol/l

0.11 0.05 1.7 5

All methods 5.62mmol/l

0.46 0.05 8.2 86

Cholesterol |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 11/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 1.69 1.68 -0.76% -0.4918/4 Sample S001 3.05 3.06 0.34% 0.4218/3 Sample S002 1.28 1.24 -3.09% -1.7118/3 Sample S001 3.54 3.36 -5.22% -1.7518/2 Sample S002 3.47 3.42 -1.42% -18/2 Sample S001 1.05 1.06 1.26% -18/1 Sample S002 1.63 1.57 -3.91% -1.8718/1 Sample S001 2.97 2.85 -4.09% -1.5117/4 Sample S002 1.05 1.08 2.38% 0.8417/4 Sample S001 0.97 0.99 1.68% 0.5017/3 Sample S002 1.61 1.57 -2.44% -0.5617/3 Sample S001 1.77 1.72 -3.02% -0.68

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

1.69mmol/l

0.03 0.01 1.6 5

All methods 1.32mmol/l

0.25 0.05 18.9 22

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

3.05mmol/l

0.02 0.01 0.8 5

All methods 3.04mmol/l

0.34 0.07 11.2 22

Triglycerides |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 12/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 66.40 66.00 -0.60% -0.1418/4 Sample S001 339.44 328.00 -3.37% -0.9818/3 Sample S002 83.60 83.00 -0.72% -0.4518/3 Sample S001 328.80 317.00 -3.59% -1.1418/2 Sample S002 327.25 317.00 -3.13% -18/2 Sample S001 88.50 90.00 1.69% -18/1 Sample S002 62.33 70.00 12.30% 1.5618/1 Sample S001 334.50 347.00 3.74% 1.1817/4 Sample S002 84.00 89.00 5.95% 1.0117/4 Sample S001 101.29 108.00 6.63% 1.7017/3 Sample S002 83.40 83.00 -0.48% -0.0717/3 Sample S001 66.80 67.00 0.30% 0.05

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 66.40 U/l 2.88 1.29 4.3 5All methods 59.85 U/l 7.18 2.07 12.0 12

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

339.44U/l

11.64 5.21 3.4 5

All methods 404.67U/l

77.40 22.34 19.1 12

Alkaline phosphatase |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 13/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 46.60 45.00 -3.43% -0.3818/4 Sample S001 302.00 275.00 -8.94% -1.1518/3 Sample S002 52.00 51.00 -1.92% -0.1718/3 Sample S001 313.00 312.00 -0.32% -0.0318/2 Sample S002 328.25 315.00 -4.04% -18/2 Sample S001 58.50 57.00 -2.56% -18/1 Sample S002 46.83 43.00 -8.19% -1.3118/1 Sample S001 317.67 310.00 -2.41% -0.6617/4 Sample S002 62.29 62.00 -0.46% -0.0617/4 Sample S001 77.04 77.00 -0.06% -0.0117/3 Sample S002 65.60 65.00 -0.91% -0.1217/3 Sample S001 56.40 58.00 2.84% 0.32

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 46.60 U/l 4.22 1.89 9.1 5All methods 43.58 U/l 6.71 2.12 15.4 10

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

302.00U/l

23.53 10.52 7.8 5

All methods 303.18U/l

37.91 11.99 12.5 10

Creatine kinase |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 14/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 60.00 57.00 -5.00% -18/4 Sample S001 250.00 259.00 3.60% -18/3 Sample S002 378.00 151.00 -60.05% -18/3 Sample S001 861.50 356.00 -58.68% -18/2 Sample S002 333.00 358.00 7.51% -18/2 Sample S001 113.50 118.00 3.96% -18/1 Sample S002 166.50 63.00 -62.16% -18/1 Sample S001 685.50 265.00 -61.34% -17/4 Sample S002 329.25 318.00 -3.42% -0.6117/4 Sample S001 327.50 321.00 -1.98% -0.4017/3 Sample S002 362.50 351.00 -3.17% -0.7117/3 Sample S001 180.00 174.00 -3.33% -0.71

xpt sd SEM CV% n

Vitros 250-5600 and 5,1FS, calculated to IFCC

60.00U/l

4.24 3.00 7.1 2

All methods 77.46U/l

29.61 11.19 38.2 7

xpt sd SEM CV% n

Vitros 250-5600 and 5,1FS, calculated to IFCC

250.00U/l

7.94 4.58 3.2 3

All methods 284.94U/l

142.72 50.46 50.1 8

Lactate dehydrogenase |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxx x

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 15/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 43.80 44.00 0.46% 0.4518/4 Sample S001 158.80 162.00 2.02% 0.5618/3 Sample S002 73.20 74.00 1.09% 0.3118/3 Sample S001 179.20 181.00 1.00% 0.3218/2 Sample S002 178.00 173.00 -2.81% -18/2 Sample S001 84.00 84.00 0.00% -18/1 Sample S002 43.17 42.00 -2.70% -1.0018/1 Sample S001 158.00 157.00 -0.63% -0.5617/4 Sample S002 83.14 84.00 1.03% 0.8017/4 Sample S001 48.57 50.00 2.94% 1.1217/3 Sample S002 64.25 64.00 -0.39% -0.2017/3 Sample S001 57.25 57.00 -0.44% -0.20

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 43.80 U/l 0.45 0.20 1.0 5All methods 42.84 U/l 2.23 0.60 5.2 14

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

158.80U/l

5.76 2.58 3.6 5

All methods 141.14U/l

23.59 6.31 16.7 14

Glutamyl transferase |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 16/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 1.01 1.03 1.50% 0.7418/4 Sample S001 1.43 1.42 -0.67% -0.1818/3 Sample S002 1.48 1.40 -5.63% -1.0518/3 Sample S001 1.41 1.37 -2.95% -0.9318/2 Sample S002 1.47 1.42 -3.37% -18/2 Sample S001 1.21 1.03 -15.05% -18/1 Sample S002 1.03 1.01 -2.16% -0.5018/1 Sample S001 1.48 1.40 -5.23% -1.6717/4 Sample S002 1.20 1.27 6.14% 0.9717/4 Sample S001 0.91 0.90 -0.97% -0.1717/3 Sample S002 1.40 1.34 -4.56% -0.8417/3 Sample S001 1.10 1.06 -3.21% -1.40

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

1.01mmol/l

0.02 <0.01 2.0 5

All methods 1.41mmol/l

0.56 0.12 40.1 23

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

1.43mmol/l

0.05 0.02 3.7 5

All methods 1.39mmol/l

0.36 0.08 26.1 23

Cholesterol, high density liporoteins |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 17/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 41.18 39.50 -4.08% -1.3418/4 Sample S001 54.52 52.30 -4.07% -1.6318/3 Sample S002 43.94 43.50 -1.00% -1.7518/3 Sample S001 51.42 50.20 -2.37% -1.4918/2 Sample S002 49.50 51.20 3.43% -18/2 Sample S001 41.97 43.30 3.18% -18/1 Sample S002 39.90 39.10 -2.01% -0.6818/1 Sample S001 53.60 53.10 -0.93% -0.4717/4 Sample S002 41.83 41.30 -1.26% -0.3617/4 Sample S001 29.04 29.40 1.25% 0.7717/3 Sample S002 41.80 41.60 -0.48% -0.3617/3 Sample S001 40.74 40.50 -0.59% -0.43

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 41.18 g/l 1.25 0.56 3.0 5All methods 40.73 g/l 1.13 0.38 2.8 9

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 54.52 g/l 1.36 0.61 2.5 5All methods 55.28 g/l 1.56 0.52 2.8 9

Albumin |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 18/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 1.18 1.15 -2.25% -1.1018/4 Sample S001 3.27 3.19 -2.35% -1.2218/3 Sample S002 1.54 1.52 -1.09% -1.0418/3 Sample S001 3.12 3.07 -1.50% -1.0618/2 Sample S002 3.08 3.00 -2.69% -18/2 Sample S001 1.45 1.41 -2.48% -18/1 Sample S002 1.13 1.12 -0.83% -0.4418/1 Sample S001 3.19 3.06 -4.14% -0.9817/4 Sample S002 1.45 1.44 -0.85% -0.4517/4 Sample S001 1.50 1.48 -1.11% -0.6417/3 Sample S002 1.27 1.25 -1.67% -0.4017/3 Sample S001 1.20 1.17 -2.14% -0.62

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

1.18mmol/l

0.02 0.01 2.0 5

All methods 1.15mmol/l

0.06 0.02 5.2 9

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

3.27mmol/l

0.06 0.03 1.9 5

All methods 3.09mmol/l

0.22 0.07 7.0 9

Phosphorus |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 19/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 59.60 51.00 -14.43% -0.9918/4 Sample S001 294.00 276.00 -6.12% -1.1518/3 Sample S002 67.40 66.00 -2.08% -0.3818/3 Sample S001 291.00 273.00 -6.19% -1.4718/2 Sample S002 281.00 277.00 -1.42% -18/2 Sample S001 60.75 58.00 -4.53% -18/1 Sample S002 56.00 52.00 -7.14% -0.6518/1 Sample S001 290.50 276.00 -4.99% -1.4817/4 Sample S002 63.14 69.00 9.28% 1.2817/4 Sample S001 43.63 44.00 0.86% 0.0917/3 Sample S002 65.40 74.00 13.15% 1.0517/3 Sample S001 67.20 74.00 10.12% 1.11

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 59.60 U/l 8.68 3.88 14.6 5All methods 59.01 U/l 5.87 1.86 10.0 10

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

294.00U/l

15.62 6.99 5.3 5

All methods 356.00U/l

68.22 21.57 19.2 10

Amylase |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 20/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 16.00 14.00 -12.50% -1.0718/4 Sample S001 187.00 181.00 -3.21% -1.2218/3 Sample S002 30.40 29.00 -4.61% -1.0418/3 Sample S001 206.80 208.00 0.58% 0.2018/2 Sample S002 209.75 209.00 -0.36% -18/2 Sample S001 28.75 28.00 -2.61% -18/1 Sample S002 16.17 16.00 -1.03% -0.2218/1 Sample S001 188.50 185.00 -1.86% -1.2817/4 Sample S002 29.00 30.00 3.45% 0.8417/4 Sample S001 46.13 50.00 8.40% 1.9817/3 Sample S002 48.80 51.00 4.51% 0.8217/3 Sample S001 20.00 21.00 5.00% 1.00

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 16.00 U/l 1.87 0.84 11.7 5All methods 15.81 U/l 7.07 2.13 44.7 11

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

187.00U/l

4.90 2.19 2.6 5

All methods 202.08U/l

24.81 7.48 12.3 11

Aspartate aminotransferase |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 21/22

Round Sample xpt Result diff% z-score

18/4 Sample S001 116.40 105.00 -9.79% -0.8918/3 Sample S002 29.20 32.00 9.59% 0.5318/3 Sample S001 144.40 145.00 0.42% 0.0718/2 Sample S002 149.00 153.00 2.68% -18/2 Sample S001 27.50 23.00 -16.36% -18/1 Sample S002 17.17 17.00 -0.97% -0.0418/1 Sample S001 135.83 134.00 -1.35% -0.3417/4 Sample S002 28.13 30.00 6.67% 0.5917/4 Sample S001 60.25 65.00 7.88% 0.7417/3 Sample S002 43.80 41.00 -6.39% -0.7917/3 Sample S001 21.00 16.00 -23.81% -1.1617/2 Sample S002 150.84 156.00 3.42% 3.65

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 13.33 U/l 8.62 4.98 64.7 3All methods 7.46 U/l 6.77 2.40 90.9 8

# - manually removed

Coordinator: We dont handle results < or >U/I.Results agreewell with other results.

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

116.40U/l

12.76 5.71 11.0 5

All methods 117.63U/l

16.07 4.64 13.7 12

Alanine aminotransferase |VITROS 5,1 FS

Basic chemistry, POC, November, 4-2018 xxxxx

Copyright © Labquality OyLabquality does not permit any reproduction for commercial purposes of any portion of the material subject to this copyright. Labquality prohibits any use of its name, or reference to LabqualityEQA program, or material in this report in any advertising, brochures or other commercial publications. Labquality EQA data do not necessarily indicate the superiority of instruments, reagents,testing equipments or materials used by participating laboratories. Use of Labquality EQA data to suggest superiority or inferiority of equipments or materials may be deceptive and misleading.Proficiency test results are handled confidentially. Labquality will not issue any statements to third parties of the performance of laboratories in external quality assessment schemes unlessotherwise agreed.

17.12.2018 22/22

Round Sample xpt Result diff% z-score

18/4 Sample S002 81.6 85.0 4.2% 1.0118/4 Sample S001 421.4 432.7 2.7% 1.0918/3 Sample S002 82.2 83.2 1.2% 0.4818/3 Sample S001 415.6 428.3 3.0% 1.1918/2 Sample S002 418.1 425.6 1.8% 0.5518/2 Sample S001 83.6 86.7 3.7% 0.8918/1 Sample S002 81.5 83.2 2.0% 0.4518/1 Sample S001 413.4 423.0 2.3% 1.0717/4 Sample S002 84.7 87.6 3.5% 0.6817/4 Sample S001 91.2 92.9 1.9% 0.5717/3 Sample S002 75.1 75.2 0.2% 0.0817/3 Sample S001 85.8 89.4 4.2% 0.59

xpt sd SEM CV% n

Vitros 250-5600, 5.1.FS 81.6 µmol/l 3.4 1.4 4.1 6All methods 78.3 µmol/l 6.9 1.8 8.8 14

xpt sd SEM CV% n

Vitros 250-5600,5.1.FS

421.4µmol/l

10.3 4.2 2.5 6

All methods 387.0µmol/l

38.8 10.4 10.0 14

Creatinine |VITROS 5,1 FS