B W Kennedy, CCLRC Rutherford Appleton Laboratory

VPTs for the CMS experiment

B W KennedyCCLRC Rutherford Appleton Laboratory

St Petersburg, 26 April 2005

Page 2: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Outline of talk

1. Description of CMS Electromagnetic Calorimeter

2. Experimental conditions and challenges3. Selection of Vacuum PhotoTriodes (VPTs)4. Experience with VPTs from RIE5. Conclusion

Page 3: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

The CMS Electromagnetic calorimeter

PbWO4 crystals

~61000 in barrel~15000 in endcaps

Pointing geometry

Length: 6m

Radius: 1.75m

Depth: 24-25 X0

Target energy resolution:<1% at E = 100 GeV

Endcap detail

(“supercrystals”)

Page 4: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Details of the SuperCrystal

Basic unit –Supercrystal55 array of crystalsCarbon-fibre alveolar

PbWO4 crystalsRadiation hardX0 = 8.9 mm

Fast scintillation 90% light in

100ns

Mechanically fragile Low light yield

~50 /MeV

Page 5: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Challenges for ECAL photodetector

High radiation environmentDose is strong function of angle to proton beamsBarrel: Up to 4 kGy in 10 years of LHC runningEndcap: 4-200 kGy in 10 years

Fast response requiredInterval between LHC beam crossings is only 25ns

Low light yield from PbWO4

Approximately 50 photons/MeV need device with gain

CMS choicesBarrel – Avalanche PhotoDiodesEndcap – Vacuum PhotoTriodes (VPTs)

Page 6: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Structure of Vacuum PhotoTriode

Single-stage photomultiplier

Gain ~ 10

Light

0V

1000V

800V

Photocathode

Grid anode

Dynode

Page 7: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Specification for CMS VPTs

Radiation toleranceLess than 10% loss in output after 20 kGy

Quantum efficiency pp 15% for light of wavelength 420nm

Gain g at zero magnetic fieldg 7

Magnetic field responseLoss in yield at 4T < 15% with respect to 0TStable (10%) at angles up to 26° to magnetic

field

Page 8: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Selection of VPTs

Prototypes supplied by several manufacturersElectron TubesHamamatsuMeltzPhotonisResearch Institute Electron

Testing and evaluation up to 1999Most manufacturers were able to meet

specifications

Tendering for contract in 1999Successful tender from RIE to supply 15500 VPTs

Page 9: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Typical response of prototype VPTs

Typical magnetic response

0.0

0.2

0.4

0.6

0.8

1.0

1.2

0 0.5 1 1.5 2

Magnetic Field (Tesla)

Re

l. A

no

de

Re

sp

on

se

0.0

0.2

0.4

0.6

0.8

1.0

1.2

-90 -60 -30 0 30 60 90VPT angle (deg.)

Rel.

An

od

e R

esp

on

se

VPT response vs angle in 1.8T magnetic field

VPT response vs field at 15°

0

2

4

6

8

10

12

0 200 400 600 800 1000

Dynode Voltage

Gai

n

V(A)=1000V

V(A)=800V

Variation of gain with applied voltage

Page 10: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Radiation tolerance of faceplate glass

dose varies strongly with angle.

All VPT faceplateglass tested to 20 kGy at Brunel University.

Glass batch accepted if <10% transmission loss (convoluted over PbWO4 spectrum) after 20kGy.

Nb-doped Lead tungstate

0.000

0.050

0.100

0.150

0.200

0.250

300 350 400 450 500 550 600 650

Wavelength (nm)

Rela

tive e

mis

sio

n i

nte

nsit

y

YT-49 Batch 30832a28.2 kGy dose

-0.02

0

0.02

0.04

0.06

0.08

0.1

0.12

300 400 500 600 700 800

Wavelength (nm)

Ind

uc

ed

ab

so

rba

nc

e

Page 11: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Delivery of VPTs

VPT progress

0

2000

4000

6000

8000

10000

12000

14000

16000

18000

01/01/2000 01/01/2001 01/01/2002 01/01/2003 01/01/2004 01/01/2005 01/01/2006 01/01/2007

Date

Nu

mb

er

of

VP

Ts

Total deliveredTotal scheduledTested at 1.8TTested at 4T

Page 12: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

VPT testing procedure

Two UK testing facilitiesUp to 1.8T at RAL

All VPTs, variable angle

4T system at Brunel University Sample testing (10%), fixed angle

Page 13: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Measurements on 10000 VPTs

Response measured at RAL in 1.8T magnetic field.

Consistent VPT response from batch to batch.

Correlation between 0T measurements at RIE and 1.8T measurements at RAL

Page 14: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

4T to 0T relative response

Measured at Brunel University. VPT at 15° to magnetic field.

Page 15: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Resolution of problems

Regular meetings between RIE and CMS groups at CERN, RAL, RIE.

Some problems inevitable in long production run.Most significant – discharges

seen in small fraction VPTs.Exchange of information

between CMS groups and RIE, and careful investigation by RIE experts

improvement in manufacturing procedure to reduce number of VPTs with discharges.

VPT 790

Initial spike

Example of discharge in VPT response

Fraction of VPTs showing discharge

Change to procedure

Page 16: B W Kennedy, CCLRC Rutherford Appleton Laboratory VPTs for the CMS experiment B W Kennedy CCLRC Rutherford Appleton Laboratory St Petersburg, 26 April

B W Kennedy, CCLRC Rutherford Appleton Laboratory

Conclusions

• VPTs supplied by RIE meet needs of the CMS experiment.

• Delivery of VPTs is on schedule.• Regular and open meetings between CMS

and RIE enable problems to be solved.• We look forward to the successful

conclusion of the contract in February 2006.