Axiom™ 2.0 Assay 384HT-Array FormatSITE PREPARATION GUIDE
Automated Protocol for Beckman Biomek FXP
Publication Number 703269
Revision 5
For Research Use Only. Not for use in diagnostic procedures.
The information in this guide is subject to change without notice.
DISCLAIMER
TO THE EXTENT ALLOWED BY LAW, LIFE TECHNOLOGIES AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Important Licensing Information
This product may be covered by one or more Limited Use Label Licenses. By use of this product, you accept the terms and conditions of all applicable Limited Use Label Licenses.
Corporate entity
Life Technologies | Carlsbad, CA 92008 USA | Toll Free in USA 1 800 955 6288
TRADEMARKS
All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. Microsoft, Windows, Internet Explorer, and Excel are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries. Jitterbug is a trademark of Boekel Scientific. Biomek and Allegra are registered trademarks of Beckman Coulter, Inc. Vortex-Genie is a registered trademark of Scientific Industries Inc. Bio-Rad, Hard-Shell, and Microseal are registered trademarks of Bio-Rad Laboratories, Inc. UV-Star is a registered trademarks of GREINER BIO-ONE. Symbol is a registered trademark of Symbol Technologies, Inc.
©2014, 2016-2017 Thermo Fisher Scientific Inc. All rights reserved.
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Prepa
Contents
CHAPTER 1 Biomek® FXP Target Prep Express Configuration . . . . . . . 4
Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Instrument specifications and settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Related Beckman Coulter documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Axiom™ 2.0 Target Prep Method for 384 Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Importing the method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
CHAPTER 2 Equipment and supplies required . . . . . . . . . . . . . . . . . . 10
Applied Biosystems equipment, software, reagents, and arrays required . . . . . . . . . . . . . . 10
Labware required for the Biomek FXP Target Prep Express. . . . . . . . . . . . . . . . . . . . . . . . 13
Axiom™ 2.0 Target Prep 384HT Samples Templates Kit . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Other equipment, consumables, and reagents required . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Pre-amplification area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Oven requirements for target prep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Spectrophotometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Thermal cyclers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Shakers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Plate centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Other common lab equipment required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Other reagents and gels required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Pipettes and tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Other Consumables Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
APPENDIX A Supplier contact list. . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Documentation and support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Related documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Customer and technical support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Limited product warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
ration Guide 3
1 Biomek® FXP Target Prep ExpressConfiguration
Equipment
The Beckman Coulter Biomek® FXP Target Prep Express (Cat. No. A83103) is required to run the Axiom™ 2.0 Assay for 384HT Samples with the automated target preparation. This workstation includes the accessories, software and deck configuration listed in Table 1. For information on upgrading a Biomek liquid handler already at your site, contact Beckman Coulter. In addition to this workstation, the components listed in Chapter 2 are also required. This chapter details the configuration required when using Windows® 7 operating system.
Table 1 Biomek FXP Target Prep Express for Running the Axiom™ 2.0 Assay for 384HT Samples Workflow
Item Quantity BeckmanCat. No.
Biomek FXP Target Prep ExpressThe part number for this workstation includes the components listed below.
1 A83103
• Biomek FXP Dual Multichannel Span-8 1 A31844
• Biomek Windows 7 New Installation kit with Biomek Software 4.1 1 B40917
• Packaged Tip Loader 1 719356
• Span-8 Disposal ALP 1 719590
• Low Volume Disposable Mandrel Kit 8-Pack 1 A39380
• BFX 96-channel Disp, Tip Head, 200U 1 719368
• Pkg Static ALP Platform 4 719357
• 4 × 3 ALP Kit 1 719948
• Static Peltier ALP 1 A93938
• Shaking Peltier ALP 1 A93942
• Adapter, Deep Well Plate (required for Axiom™ assays) 1 A83050
• Adapter, Thin PCR (required for HT 3' IVT PLUS and HT WT PLUS assays) 1 A19709
• Frame for Modular Reservoir 4 372795
• 24-Position Tube Rack 1 373661
• Inserts, Tube, 11 mm (case of 25) 1 373696
• 24 Position Cold Tube Block(referred to as Reagent Cold Block in this document)
1 A83054
4 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 1 Biomek® FXP Target Prep Express ConfigurationInstrument specifications and settings 1
Instrument specifications and settings
In order for the Axiom 2.0 assay for 384HT Samples to be successfully performed on the Biomek FXP Target Prep Express Instrument the system must meet the following specifications and be properly configured.
• Biomek FXP Target Prep Express system
– Hybrid dual armed Biomek FXP
– 96MC head left pod and Span-8 head right pod
– 1 mL syringes and low volume tubing on the Span8 system
– 384 Deep-Well Adapter (Cat. No. 101019) mounted on the Shaking Peltier Device with notches facing the front
– ALPs secured to deck as shown in Figure 1
• System fluid is degassed water
• Clean compressed air is at 40–80 psi
• Computer workstation specifications
– Windows® 7 operating system
– Internet Explorer® v8.0
– Microsoft® Excel® 2010 software; 32-bit version ONLY software
– Biomek software v4.1.41 (version 4.1 build 41)
• Biomek software
– Make sure that the Affymetrix Deck is the default deck
– Make sure deck location SPelt96_1 has the device ShakingPeltier1 associated with it
– Make sure deck location Pelt_1 has the device StaticPeltier1 associated with it
– Make sure deck location TR1 is ALP type Span8TrashRight and not Span8TipTrashRight
• Check for a properly framed deck for both the Biomek FXP 96MC and Span8 pods
– Framing tolerance is 0.03 cm. Reframe if this is exceeded when checked
– If deck was reframed, save a current copy of the instrument file for backup
– Exit and restart Biomek software after installing Affymetrix Deck and after any changes in deck framing in order to save the new values in the default instrument (.bif) file
• Additional Pod Settings
– Navigate to: Instrument Hardware Setup Pod2 Additional Pod Settings
– Pod 2 (Span-8) settings: Set System Trailing Airgap to 20 µL (Figure 2)
– Pod 1 (MC96) settings (Figure 3):
Check Always Rove at Z-Max box
Set Additional Roving Height to 0.45 cm or lower (default is 0.75 cm)
Set Gripper Offset in D Axis to 0.7 cm
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 5
Chapter 1 Biomek® FXP Target Prep Express ConfigurationInstrument specifications and settings1
Figure 1 Biomek FXP TPE System (top) and Deck Configuration (bottom)
6 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 1 Biomek® FXP Target Prep Express ConfigurationInstrument specifications and settings 1
Figure 2 Hardware Setup Changes for Span-8 Pod2
Figure 3 Hardware Setup Changes for Multichannel 96 Pod1
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 7
Chapter 1 Biomek® FXP Target Prep Express ConfigurationRelated Beckman Coulter documentation1
Related Beckman Coulter documentation
• Biomek® Liquid Handler User’s Manual, Beckman Coulter Pub. No. 987834
• Biomek® FX Laboratory Automation Workstation Preinstallation Manual, Beckman Coulter Pub. No. 719450, rev AH
• Biomek® Software User’s Manual Version 4.1, Beckman Coulter Pub. No. B30026AA
• Install Guide Biometra TRobot on Biomek FX Revision 2, 9/8/2014
Axiom™ 2.0 Target Prep Method for 384 Samples
The purpose of this section is to identify critical steps for importing the Axiom™ 2.0 Target Prep 384 Samples method into the Biomek® software for use on the Biomek® FXP Target Prep Express.
Importing the method
1. Run Setup.exe.
2. Click Setup.exe in method package to copy files from the package to the following locations and import the Biomek method.
Table 2 Directory Location
Source Destination
\Root\*.* C:\Users\Public\Documents\Affymetrix\Axiom 2.0 384
\Images\ C:\Users\Public\Documents\Axiom 2.0 384\Images
\Misc\ C:\Users\Public\Documents\Axiom 2.0 384
8 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 1 Biomek® FXP Target Prep Express ConfigurationAxiom™ 2.0 Target Prep Method for 384 Samples 1
3. Set Method Preferences:
a. Locate and click the method branch named Axiom 2.0 Target Prep 384 Samples (Figure 4).
b. Barcode Tracking [Enabled] to input barcode during runtime deck setup.
• Log Name [Default/Other]: Name of barcode log text file
• Log Path [Default/Other]: Location of barcode log file
c. Set the Validate Reagent Module Barcodes run option to check for the correct reagent module part number.
d. Set the Verify GeneTitan Labware Setup run option to check if the GT trays were placed in the correct position on the deck.
e. Set the Prepare gel and OD plates for QC run option.
f. Set the Test Mode run option which disables the timers.
Figure 4
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 9
2 Equipment and supplies required
This chapter includes the supplier and ordering information for the equipment, software, reagents, arrays, labware and other consumables that have been validated for use with the Applied Biosystems™ Axiom™ 2.0 Assay for 384HT Samples.
Information on suppliers of the equipment and supples is listed in Appendix A, ʺSupplier contact listʺ on page 27.
Applied Biosystems equipment, software, reagents, and arrays required
Table 3 Applied Biosystems equipment, software, reagents and arrays required
Item Cat. No.
Equipment
GeneTitan™ Multi-Channel Instrument* Contact Thermo Fisher Scientific, Inc.
Software
GeneChip™ Command Console™ (AGCC) version 4.3 or later
Axiom Analysis Suite version 1.1 or later
Reagents
Axiom™ 2.0 384HT Reagent Kit 902245 (Sufficient for one 384HT array plate)
Arrays
Axiom™ 384HT Genotyping Solution Array Plates:
Please visit www.thermofisher.com for a complete list of array plates available for use with the Axiom™ 2.0 Assay for 384 Samples.
* For a complete list of all equipment and supplies required for GeneTitan Instrument installation and operation, please consult theGeneTitan™ Multi-Channel Instrument Site Preparation Guide (Pub. No. 08-0305).
10 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required 2
Consumables
Each Axiom™ 384HT High Volume Consumable Kit is sufficient to process 5 Axiom 384-format plates. • 10 x 384 Layout GeneTitan™ Stain Tray (Stain 1)• 5 x 384 Layout Axiom™ Stain 2 Tray• 5 x 384 Layout Axiom™ Stab. Tray• 5 x 384 Layout Axiom™ Ligation Tray• 5 x 384 Layout GeneTitan™ Hyb Tray• 5 x 384 Layout GeneTitan™ Scan Tray• 25 x 384 Layout GeneTitan™ Scan and Stain Tray Cover These trays are required for processing Axiom 384 array plates on the GeneTitan™ Multichannel Instrument. Please use the new Axiom 384HT High Volume Consumable Kit when running the Applied Biosystems™ Axiom™ 2.0 Assay for 384HT Samples.
902629
Axiom™ 384HT Consumable Kit for Biomek FXp (sufficient for 5 x 384 rxn runs)1 2
Each Kit Includes:• 10 x 96 PCR Plate (Part No. 203015)• 20 x 96 Deep-Well Square Plate (Part No. 203016)• 15 x 384 Deep-Well 240 μL Plate (Part No. 203017)• 15 x 384 Standard 120 μL Plate (Part No. 203018)• 5 x 384 PCR Plate (Part No. 902288)• 35 x Shallow 96 Full Reservoir (Part No. 203020)• 5 x Deep 96 Full Reservoir (Part No. 203021)
902288
Axiom™ 384HT Consumable Kit for QC (sufficient for 10 x 384 rxn runs)2, Each Kit Includes:• 10 x 384 UV Plate (Part No. 203022)
902289
Axiom™ 384HT Starter kit for Biomek FXP Each Kit Includes:• 1 x Thermoshake 384 Adapter (Part No. 101019)• 1 x Plastic Plate Lid (Part No. 101018)• Plate collar• 1 set Reservoir Stickers (Part No. 90-0972)• 1 x Cold Block Template #1 (Part No. 15-0448)• 1 x Cold Block Template #2 (Part No. 15-0451)• 1 x 2D Handheld Barcode Reader (Part No. 00-0179) • 1 x External Barcode Reader Holder (Part No. 00-0093)
902290
Training Modules Available
Axiom™ 384HT Training Kit, Each Kit Includes:• Axiom™ 2.0 384HT Reagent Kit.• Axiom™ Human AIMs Array Plate• Axiom™ 384HT Consumable Kit for Biomek FXP • Axiom™ 384HT Consumable Kit for QC • Axiom™ 384HT GeneTitan™ MC Consumable Kit• HapMap Samples3
902291
1 For quantity required per run, please refer to Table 4.2 The components of this kit may be purchased directly from the vendor, as listed in Table 4.
Table 3 Applied Biosystems equipment, software, reagents and arrays required (Continued)
Item Cat. No.
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 11
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required2
3 Note: Per Thermo Fisher Scientific agreement with Coriell, all residual HapMap DNA samples must be destroyed immediately followingcompletion of training.
12 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required 2
Labware required for the Biomek FXP Target Prep Express
The labware required to run the Axiom™ 2.0 Assay for 384HT Samples Workflow on the Biomek FXP Target Prep Express is listed in Table 4. Photographs of the labware are provided in Table 5 on page 14.
Table 4 Labware for the Biomek FXP Target Prep Express
Item Quantity required Supplier Cat. No.
Pipette tips, barrier, 96 tips/rack1
AP96 P50 Tips (pink) 1152 tips
Beckman Coulter
A21586
AP96 P250 Tips (aqua) 1440 tips 717253
Span-8 P250 Tips (green) 50 tips 379503
Span-8 P1000 Tips (yellow) 154 tips 987925
Plates
96 PCR Plate2 2 BIO-RAD HSP9631
384 Deep-Well 240 μL Plate2 3 3 E&K Scientific EK-2010-SBC
384 Standard 120 μL Plate2 3 Axygen/Corning P-384-120SQ-C
384 PCR Plate2,3 1 Thermo Fisher Scientific
902288
96 Deep-Well Square Plate2, 3 4 Thermo Scientific
AB-0932
384 UV Plate4 1 E & K Scientific EK-30801
Reservoirs for Reagents
Shallow 96 Full Reservoir2 7 E & K Scientific EK-2036
Deep 96 Full Reservoir2 1 E & K Scientific EK-2035
Half Reservoir (75 mL module) 2Beckman Coulter
372786
Quarter Reservoir (40 mL module) 13 372790
Quarter Reservoir Divided by Width (19 mL module) 1 372792
Other
Microseal 'P' Replacement Pads BIO-RAD MSP-1003
Plastic Plate Lid Thermo Fisher Scientific
101018
384 Deep-Well Adapter Thermo Fisher Scientific
101019
1 For pipette tips, the “Quantity required” column lists the total number of tips required and not the number of racks.2 Available in the Axiom™ 384HT Consumable kit for Biomek FXP (Cat. No. 902288). See Table 3 on page 10 for details.3 This plate is also part of the Axiom™ 384HT Consumable kit for Biomek® FXP. If purchased directly from the supplier, a unique barcode
must be added.4 Available in the Axiom™ 384HT Consumable Kit for QC (Cat. No. 902289). See Table 3 on page 10 for details.
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 13
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required2
Photographs of the labware requiredThe photographs in Table 5 are provided to help visualize the labware required on the Biomek FXP Target Prep Express to run the automated target preparation portion of the Axiom™ 2.0 Assay Automated Workflow.
Table 5 Labware Used on the Biomek Workstation deck
Labware Supplier and Cat. No. Labware image
Biomek AP96 – P250 Pipette Tips (aqua box; pre-sterile, barrier)
Beckman CoulterCat. No. 717253
Biomek Span P250 Pipette Tips (green box; pre-sterile, barrier)
Beckman CoulterCat. No. 379503
Biomek Span P1000 Pipette Tips (yellow box; pre-sterile, barrier, conductive)
Beckman CoulterCat. No. 987925
Biomek Span P50 Pipette Tips (pink box; pre-sterile, barrier)
Beckman CoulterCat. No. A21586
14 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required 2
Frame for reservoirs Beckman CoulterCat. No. 372795
Half reservoirHalf module, 75 mL capacity
Beckman CoulterCat. No. 372786
Quarter reservoirs• Quarter module, 40 mL capacity• Quarter module divided by width,
19 mL capacity each receptacle
Beckman Coulter• Cat. No. 372790 (40 mL)• Cat. No. 372792 (19 mL)
Table 5 Labware Used on the Biomek Workstation deck (Continued)
Labware Supplier and Cat. No. Labware image
Frame
Divided by width19 mL capacity
Undivided40 mL capacity
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 15
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required2
Reagent block, chilled to 4°C Beckman CoulterCat. No. A83054
24-position tube rackwith two 11 mm tube inserts in positions A1 and B1.
Beckman CoulterCat. No. 373661 (rack)Cat. No. 373696 (insert)
Plate collar Plate collar used with ABgene 384-well PCR plate. Contact your Genotyping Technical Sales Specialist for additional plate collars.
96 PCR plateHard-Shell® Thin-Wall96-Well Skirted PCR Plates(available in multiple colors)
Bio-RadCat. No. HSP9631
Also available in the Axiom™ 384HT Consumable kit for Biomek FXP (Cat. No. 902288)
Table 5 Labware Used on the Biomek Workstation deck (Continued)
Labware Supplier and Cat. No. Labware image
Metal posts on block circled in red.A1
16 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required 2
384 deep-well 240 μL plate1
Axygen® 384 Well Clear V-Bottom 240 μL Polypropylene Deep Well Not Treated Plate, Sterile
Axygen/CorningCat. No. P-384-240SQ-C-S
Also available in the Axiom™ 384HT Consumable kit for Biomek FXP (Cat. No. 902288)
384 standard 120 μL plateAxygen® 384 Well Clear V-Bottom 120 μL Polypropylene Deep Well Not Treated Plate
Axygen/CorningCat. No. P-384-120SQ-C
Also available in the Axiom™ 384HT Consumable kit for Biomek FXP (Cat. No. 902288)
384 PCR Plate2 ABgene 384 Well PCR Microplate, Clear, Nonsterile
Thermo Fisher ScientificCat. No. AB0937
Also available in the Axiom™ 384HT Consumable kit for Biomek FXP (Cat. No. 902288)
96 deep-well square plate1 ABgene® Mark II 96-Well Storage Plate, Square Well, 2.2 mL
Thermo ScientificCat. No. AB-0932
Also available in the Axiom™ 384HT Consumable kit for Biomek FXP (Cat. No. 902288)
Table 5 Labware Used on the Biomek Workstation deck (Continued)
Labware Supplier and Cat. No. Labware image
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 17
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required2
384 UV plate384 Well, UV-Star Plates 120 μL, Non-Sterile
E & K ScientificCat. No. EK-30801
Also available in the Axiom™ 384HT Consumable kit for QC (Cat. No. 902289)
Shallow 96 full reservoir Individual Well Format Reservoirs - 96 Well, Low Profile
E & K ScientificCat. No. EK-2036
Also available in the Axiom™ 384HT Consumable kit for Biomek FXP (Cat. No. 902288)
Deep 96 full reservoir Individual Well Format Reservoirs - 96 Well, Sterile
E & K ScientificCat. No. EK-2035
Also available in the Axiom™ 384HT Consumable kit for Biomek FXP (Cat. No. 902288)
Plastic plate lid Thermo Fisher Scientific Cat. No. 101018
Also available in Axiom™ 384HT Starter kit for Biomek FXP (Cat. No. 902290)
Table 5 Labware Used on the Biomek Workstation deck (Continued)
Labware Supplier and Cat. No. Labware image
Plastic plate lid (top side)
Plastic plate lid (underside)
P-pad
18 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required 2
P PadMicroseal® ‘P’ PadPCR plate-sealing pad for automation, resusable
Bio-RadCat. No. MSP-1003
384 deep-well adapterPlate adapter for Inheco Thermoshake, fits Axygen P-384-240SQ-C-S 384 deep-well plate
Thermo Fisher ScientificCat. No. 101019
Also available in Axiom™ 384HT Starter kit for Biomek FXP (Cat. No. 902290)
1 The plates come unbarcoded from the vendor. Barcode labels should be applied before use. Barcoded versions of the plates areavailable in the Axiom 384HT Consumable Kit for Biomek FXP.
2 At the current time this plate is unbarcoded. If desired, barcode label must be manually applied.
Table 5 Labware Used on the Biomek Workstation deck (Continued)
Labware Supplier and Cat. No. Labware image
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 19
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required2
Axiom™ 2.0 Target Prep 384HT Samples Templates Kit
The Tube Block and Reservoir Frames templates are provided in the Axiom™ 384HT Starter Kit for Biomek FXP (Cat. No. 902290) and are ordered by your Thermo Fisher Scientific FAS prior to beginning training. These templates are used with the automated target prep protocol.
Tube block templateThe Axiom 2.0 Target Prep 384 Samples Reagent Kit tube block templates were designed to fit precisely onto the top of the Beckman Coulter Tube Block (Cat. No. A83054). They are held in place by the metal posts on the block (Figure 5). Using these templates helps to ensure the proper placement of reagent tubes onto the block for each method.
Reservoir labelsThe reservoir labels are stick-on labels for the modular reagent reservoir frames (used at different stages in the automated workflow) and are provided in the Axiom™ 2.0 Target Prep 384HT Samples Templates Kit. These stick-on labels are color-coded to match the colors found on the caps of the reagent tubes in the Axiom 2.0 384HT Reagent Kit. Using these labels helps ensure the proper placement of reservoirs and reagents for each step.
Figure 5 Axiom Reagent Templates for Tube Block
Tube block templates
Tube block
20 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 2 Equipment and supplies requiredApplied Biosystems equipment, software, reagents, and arrays required 2
There are three reservoir holders used in the Axiom 2.0 Target Prep 384 Samples method. Two of these will have templates on two sides and the remaining reservoir holder will have a template on one side for a total of 7 templates.
Remove the protective surface from the back of the label and place on the reservoir frames as directed in the Figure 6 through Figure 8.
Reservoir frame 1
Reservoir frame 2
Figure 6 Reservoir frame 1
Figure 7 Reservoir frame 2
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 21
Chapter 2 Equipment and supplies requiredOther equipment, consumables, and reagents required2
Reservoir frame 3
Other equipment, consumables, and reagents required
Pre-amplification area
Precautions are required when manipulating genomic DNA to avoid contamination with foreign DNA amplified in other reactions and procedures. It is recommended that genomic DNA manipulations are performed in a dedicated pre-amplification room or area separate from the main laboratory.
This pre-amplification area should have a dedicated set of pipettes and plasticware. If no dedicated area is available, use of a dedicated bench or a dedicated biosafety hood and dedicated pipettes is suggested. If no dedicated bench or biosafety hood is available, a set of dedicated pipettes is recommended.
Oven requirements for target prep
We recommend using the ED 56 BINDER oven listed in Table 6. If another oven is used, it must be able to maintain a constant temperature of 37°C for at least 24 hr, and have a temperature accuracy of 1°C.
Figure 8 Reservoir frame 3
Table 6 Ovens Required for the Axiom 2.0 Assay for 384HT Samples
Item Supplier Cat. No.
Oven:
Required if processing more than 3 array plates per week.
• ED 56 Drying and Heating Chamber1 BINDER ED056UL-120VVoltage: 120 V 1~60 Hz
P/N 9010-0334
ED056-230VVoltage: 230 V 1~50/60
HzP/N 9010-0333
Optional — for low throughput of 3 or fewer array plates per week:• GeneChip™ Hybridization Oven 6452 Thermo Fisher
Scientific00-0331
1 Replaces BINDER Model ED 56.2 The GeneChip™ Hybridization Oven 640 is currently not supported with the Axiom 2.0 Assay; however, if you want to utilize it in the
workflow please contact your Field Service Engineer (FSE) or Thermo Fisher Scientific Technical Support regarding the compatibility ofthis oven with the Axiom 2.0 Assay.
22 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 2 Equipment and supplies requiredOther equipment, consumables, and reagents required 2
Spectrophotometer We recommend using one of the spectrophotometers listed in Table 7.
Thermal cyclers We have verified the performance of this assay using the following off-deck thermal cyclers in their 384-well block configurations:
• Dual 384-Well GeneAmp® PCR System 9700, Cat. No. N8050002
• For existing GeneAmp® PCR System 9700 the following 384 block module can be used with the assay
– Dual 384-Well Sample Block Module, Cat. No. N8050400 BI 9700 (with a gold, silver or aluminum block)
The performance of this assay has not been verified with other thermal cyclers. Use of other thermal cyclers may result in assay failure and may violate the Axiom Array and Reagent replacement policy.
Shakers We recommend using one of the following shakers listed in Table 8.
Table 7 Spectrophotometers
Item Supplier Cat. No.
One of the following spectrophotometers:• DTX 880 Multimode Detector, with:
Genomic Filter SlideBeckman Coulter 987921
A30184
SpectraMax High throughput Microplate Spectrophotometer
Molecular Devices
Plus384
Table 8 Shakers
Shaker Supplier Cat. No.
Digital Microplate Shaker Thermo Scientific 88880023 or 88880024
Jitterbug™ Boekel Scientific Model 130 000
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 23
Chapter 2 Equipment and supplies requiredOther equipment, consumables, and reagents required2
Plate centrifuge One plate centrifuge is required for the Axiom™ 2.0 Assay 384 Samples Workflow. We recommend the plate centrifuges listed in Table 9. When centrifuging and drying pellets, the centrifuge must be able to spin down plates at:
• rcf: 3200 xg with an appropriate rotor - bucket combination (4000 RPM for the Eppendorf 5810R configuration described below)
• temperature: 4°C
In addition, the bottom of the rotor buckets should be soft rubber to ensure that the deep well plates do not crack. Do not use buckets where the plates sit directly on a metal or hard plastic bottom. For the Eppendorf 5810R, do not use the A-4-62 rotor with a WO-15 plate carrier (hard bottom).
Relative centrifugal force (rcf) can be calculated as follows:
rcf = (1.118 × 10-5) R S2
where R is the radius of the rotor in centimeters, and S is the speed of the centrifuge in revolutions per minute.
Table 9 Plate centrifuges recommended for the Axiom™ 2.0 Assay for 384HT Samples Automated Workflow
Item Supplier Cat. No.
One of the following centrifuges:
• Allegra® 25R Refrigerated Benchtop Centrifuge Beckman Coulter 369434 (60 Hz, 280 V)369435 (50/60 Hz, 200 V)369436 (50 Hz, 230 V)
368954 S5700 Microplate rotor
• Centrifuge 5810R (refrigerated)
– A-4-81 MTP/Flex swinging bucket rotor with 4 microtest-plate buckets
Eppendorf 022625501 (60 Hz, 120 V)022625101 (50 Hz, 120 V, 20 AMP version)022625551 (50 Hz, 230 V)
022638807
• Sorvall® Legend® XTR (refrigerated), with: Thermo Scientific 75004521 (60 Hz, 120 V)75004520 (50 Hz, 230 V)75004523 (50/60 Hz, 230 V USA and Canada)
TX-750 high visibility swinging rotor bucket 75003607
Carrier for microplates (including plate trays and neoprene pads)
One of either:• Set of two carriers (75003795)• Set of four carriers (75003617)
24 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Chapter 2 Equipment and supplies requiredOther equipment, consumables, and reagents required 2
Other common lab equipment required
Other reagents and gels required
Table 10 Other equipment required
Item Supplier Cat. No.
Freezer, –20°C Any vendor —
Refrigerator, 2 to 8°C Any vendor —
Vortex-Genie® (for plates and microtubes)1
1 Equivalent items from other manufacturers are acceptable.
Scientific Industries
SI-0236 (120 V/60 Hz)SI-0246 (230 V/50 Hz)
Mini Microcentrifuge, for 2 mL tubes1 VWR 93000-196 (120V)93000-196 (230V)
Bel-Art Cryo-Safe Mini Cooler, –15°C1 VWR 47751-730
Ice bucket, 4 to 9 liters Any vendor —
Table 11 Other Reagents Required
Item Supplier Cat. No.
2-Propanol, anhydrous, 99.5% (Isopropanol) Sigma-Aldrich 278475
Reduced EDTA TE Buffer (10 mM Tris-HCl pH 8.0, 0.1 mM EDTA)
Thermo Fisher Scientific
75793
Table 12 Reagents and Gels Required to Run QC Steps
Item Supplier Cat. No.
Mother E-Base™ Device
Thermo Fisher Scientific
EB-M03
Daughter E-Base™ Device EB-D03
E-Gel® 48 4% agarose gels (for Axiom QC) G8008-04
TrackIt™ 25 bp DNA Ladder (for Axiom QC) 10488-022
TrackIt™ Cyan/Orange Loading Buffer (for Axiom QC) 10482-028
E-Gel® 48 1% agarose gels (for gDNA QC) G8008-01
RediLoad™ (for gDNA QC) 750026
E-Gel® 96 High Range DNA Marker (for gDNA QC) 12352-019
Water, nuclease-free, ultrapure MB grade (only required for performing rehybridization protocols and gel QC)
71786
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 25
Chapter 2 Equipment and supplies requiredOther equipment, consumables, and reagents required2
Pipettes and tips Pipettes and tips recommended for performing the gel QC steps in the Axiom™ 2.0 Assay 384 Samples Workflow are listed in Table 13.
Other Consumables Required
Table 13 Pipettes and tips
Item Supplier Cat. No.s
The following single or multichannel pipettes are required: Single Channel Multichannel
Pipette, 0.5–10 μL Rainin L-10 L12-10
Pipette, 2–20 μL Rainin L-20 L12-20
Pipette tips, LTS filter Rainin GP-L10F GP-L10F
Table 14 Consumables Required
Item Supplier Cat. No.
Adhesive film for 96-well plates – use one of the following:
• MicroAmp® Clear Adhesive Film Thermo Fisher Scientific
4306311
• Microseal 'B' Film Bio-Rad MSB1001
Kimwipes® Your choice —
Markers, permanent, fine point Your choice —
26 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
A Supplier contact list
Table 15 Supplier contact list
Supplier Website Address
Axygen/Corning www.corning.com/lifesciences
Beckman Coulter www.beckmancoulter.com
Bio-Rad www.bio-rad.com
E&K Scientific www.eandkscientific.com
Eppendorf www.eppendorf.com
Molecular Devices www.moleculardevices.com
Rainin www.rainin.com
Scientific Industries www.scientificindustries.com
Sigma-Aldrich www.sigmaaldrich.com
Thermo Fisher Scientific(Thermo Scientific, Applied Biosystems, Invitrogen, Life Technologies, Affymetrix)
www.thermofisher.com
VWR www.vwr.com
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 27
Documentation and support
Related documentation
Table 16 Documents related to the Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP
Document Publication number Description
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP User Guide
703268 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP User Guide (Windows XP and 7)
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
703269 Provides guidance on reagents, instruments, and supplies required to run the Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP.
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Quick Reference (Windows XP)
703270 An abbreviated reference for the target preparation step of the Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP running on the Windows XP operating system. This quick reference document is intended for experienced users.
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Quick Reference (Windows 7)
703371 An abbreviated reference for the target preparation step of the Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP running on the Windows 7 operating system. This quick reference document is intended for experienced users.
Axiom™ 2.0 384HT gDNA Sample Prep Quick Reference
703163 An abbreviated reference on the genomic DNA sample preparation protocol for 384HT-array format.
GeneTitan™ MC Protocol for Axiom™ 384HT Array Plate Processing Quick Reference
703164 An abbreviated reference for processing 384HT-array format plates on the GeneTitan Multi-Channel (MC) Instrument.
GeneTitan™ Multi-Channel Instrument User Guide
08-0308 The GeneTitan Multi-Channel (MC) Instrument automates array processing from target hybridization to data generation by combining a hybridization oven, fluidics processing, and state-of-the art imaging device into a single bench-top instrument. This document detailing the use, care, and maintenance for the GeneTitan MC Instrument.
28 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide
Documentation and supportRelated documentation
GeneTitan™ Multi-Channel Instrument Site Preparation Guide
08-0305 Provides guidance on creating and maintaining the proper environment required for the GeneTitan Multi-Channel Instrument.
Analysis and software
Axiom™ Genotyping Solution Data Analysis Guide
702961 This guide provides information and instructions for analyzing Axiom genotyping array data. It includes the use of Axiom™ Analysis Suite, Applied Biosystems Microarray Power Tools (formerly APT) and SNPolisher R package to perform quality control analysis (QC) for samples and plates, SNP filtering prior to downstream analysis, and advanced genotyping methods.
Applied Biosystems ™ GeneChip™ Command Console™ Software User Guide
702569 This user guide provides instructions on using Applied Biosystems GeneChip Command Console Software (AGCC) used to control GeneChip instrument systems. Command Console Software provides an intuitive set of tools for instrument control and data management used in the processing of GeneChip Arrays.
Axiom™ Analysis Suite User Guide 703307 This user guide provides instructions on using Axiom™ Analysis Suite—a single-source software package to enable complete genotyping analysis of all Axiom arrays.
Other
Biomek® Liquid Handler User’s Manual 987834 This document is installed at the same time as the Biomek FXP Target Prep Express software. To access, click Start All Programs Beckman Coulter Manuals.
Biomek® Software User’s Manual B30026AA This document is installed at the same time as the Biomek FXP Target Prep Express software. To access, click Start All Programs Beckman Coulter Manuals.
Symbol® Cyclone™ M2000 Series Product Reference Guide
72E-39696-06 Barcode scanners documentation from manufacturer.
Table 16 Documents related to the Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP
Document Publication number Description
Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide 29
Documentation and supportCustomer and technical support
Customer and technical support
Visit thermofisher.com/support for the latest in services and support, including:
• Worldwide contact telephone numbers
• Product support, including:
– Product FAQs
– Software, patches, and updates
• Order and web support
• Product documentation, including:
– User guides, manuals, and protocols
– Certificates of Analysis
– Safety Data Sheets (SDSs; also known as MSDSs)
Note: For SDSs for reagents and chemicals from other manufacturers, contact the manufacturer.
Limited product warranty
Life Technologies Corporation and/or its affiliate(s) warrant their products as set forth in the Life Technologies’ General Terms and Conditions of Sale found on Life Technologies’ website at www.thermofisher.com/us/en/home/global/terms-and-conditions.html. If you have any questions, please contact Life Technologies at thermofisher.com/support.
30 Axiom™ 2.0 Assay 384HT-Array Format Automated Protocol for Beckman Biomek FXP Site Preparation Guide