avs navigator -...

36
AVS Navigator ® TLIF Surgical Technique

Upload: vuongque

Post on 20-Nov-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

AVS Navigator®

TLIF Surgical Technique

Position MattersTM

AVS Navigator PEEK Spacer

Multiple Pathologies

Multiple Approaches

Multiple Techniques

…one solution.

MIS Applications

Multiple Insertion Angles Mini - Open Applications

2

AVS Navigator TLIF Surgical Technique

3

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Step 1: Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Step 2: Preparation of Facet Joints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Step 3: TLIF Site Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Step 4: Distraction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Step 5: Discectomy and Endplate Preparation . . . . . . . . . . . . . . . . . . . . . . . . . 11

Step 6: Sizing the Disc Space . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Step 7: AVS Navigator Implant Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Step 8: Locking the Implant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Step 9: AVS Navigator Implant Insertion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Step 10: Placement of Bone Graft . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Step 11: Posterior Fusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Step 12: Closure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Implants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Instructions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Table of Contents

AVS Navigator TLIF Surgical Technique

4

Introduction

The following technique describes an open TLIF technique . This technique may be applied unilaterally to any lumbar interbody space, based on the pathology being addressed and surgeon preference . This technique may be performed through an open or minimally invasive approach . In an open or minimally invasive TLIF approach, the interbody space is generally accessed from a posterior approach, and an AVS Navigator Implant is inserted into the disc space .

Description

The AVS Navigator PEEK Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation . These hollow implants are offered in a variety of lengths, heights, and lordotic angles designed to adapt to different patient anatomies . They have serrations on the superior and inferior surfaces designed for multidirectional fixation, ergonomically shaped anterior edges, and flat posterior edges .* Radiopaque (Tantalum) markers have been embedded within the implants, which are designed to allow for visualization in radiographic images .

*These implants feature the clinically proven UniLIF wedge nose to help the leading edge slide along the ventral annulus . In addition, the back rail on the trailing end of the implant has been designed to help facilitate an 80° interdiscal turn .

Multiple Insertion Angles. . . One Implant.

AVS Navigator TLIF Surgical Technique

5

Step 1: Exposure

A TLIF procedure can be performed using a standard open or minimally invasive approach . Typically, cage insertion is performed from the symptomatic side in patients with radiculopathy or the concave side in patients with coronal deformity .

This technique will describe an L4-L5 TLIF . The incision is made over the spinous process of L3 and extends to L5 (Fig . 1) . Sharp paraspinal, subperiosteal dissection is performed exposing the L3-L4 and L4-L5 facets . Care must be taken to avoid damage to the L3-L4 facet capsules . Unless the surgeon intends to perform concomitant intertransverse fusion, the transverse processes need not be exposed .

As a note, when inserting all instruments into the disc space, care should be taken to avoid any unnecessary damage to soft tissues .

Note: The remaining steps in this surgical technique are similar in both open and minimally invasive surgical approaches. All images in this technique guide depict an L4-L5 open approach for image clarity. Please referto the Stryker Luxor, Phantom or AriaRetractor System surgical techniquesfor additional information and detailedimages on minimally invasiveapproaches.

Figure 1

AVS Navigator TLIF Surgical Technique

6

Step 2: Preparation of Facet Joints

Both L4-L5 facet capsules should be removed circumferentially . This is typically accomplished using a cautery device . The L4-L5 facet is prepared for fusion by removing the articular cartilage from the facet joint with a burr or other appropriate instrument (Fig . 2) .

Step 3: TLIF Site Preparation

An osteotome may be used to remove the left inferior articular process of L4 with two cuts (Fig . 3):

• One vertical cut just medial to the facet extending superiorly to the superior border of the facet .

• One horizontal cut directed laterally towards the foramen .

Once both cuts are made, the inferior articulating process of L4 may be removed with a kerrison . The left lateral edge of the ligamentum flavum can then be visualized . A curette may be used to release (but not resect) the ligamentum flavum from the superior lamina of L5 allowing for distraction . The ligamentum flavum should be preserved when possible, but resection of redundant flavum may be necessary in patients with neurologic compression .

Figure 2

Figure 3

AVS Navigator TLIF Surgical Technique

7

Step 4: Distraction

Effective distraction facilitates later steps of the procedure, including removal of the superior articular process of L5, decompression of the neuroforamen, preparation of the disc space, and insertion of the spacer . This can be accomplished through several techniques: pedicle screw distraction, distraction between boney elements, and/or distraction with a positioning device .

If pedicle screw distraction is chosen, the screws should be positioned at this time using a standard technique . Apply distraction between pedicle screws placed at L4 and L5 (Fig . 4) .

Boney Element DistractionIn an open procedure, boney distraction may be applied between the boney elements of L4 and L5, such as spinous processes or laminae .

Positioning DeviceThe patient’s lordosis can be adjusted with a positioning device, such as a Wilson frame (open and minimally invasive procedures) .

With sufficient distraction, the L4-L5 foramen will be opened, and the entire superior articular process of L5 will be visualized . A small probe may be used to palpate the superior margin of the left L5 pedicle, and the superior articular process of L5 should be resected to this point, which should expose the L4-L5 disc in the L4-L5 foramen . This may be accomplished with an osteotome, kerrison, rongeur, or other appropriate instrument . A consistent leash of veins traverses the foramen over the disc space . The surgeon should be prepared to control bleeding from these veins with bipolar cautery . Hemostasis may be enhanced with topical hemostatic agents and cottonoid pledgets . The exiting L4 nerve root is rostral at this point, under the pars interarticularis, and in many instances is not visible (Fig . 5) .

Figure 4

Figure 5

AVS Navigator TLIF Surgical Technique

8

The following technique is recommended when using theDynamic Distractor.

1) Thread the Extension Handle into the Dynamic Distractor .

Note: It is mandatory that the surgeon use the Extension Handle in order to prevent the application of too much torque on the distal end of the instrument.

2) Prior to insertion, ensure the Dynamic Distractor is completely closed and the dial reads 7mm .

Note: Prior to use, lubricate the Indicator Dial and rotate the T-Handle until Milk is absorbed into instrument. While the Instrument is fully open, lubricate the Paddles with Instrument Milk (Fig. 7).

Disc Space Distraction

In patients with poor bone quality, itmay be useful to size the disc space withpaddle distractors, reamer distractors,or trials before locking the distractiondown through the pedicle screws . Discspace distraction gives an excellent senseof annular tension restoration andavoids pedicle screw preloading, whichcould potentially cause post-operativeloosening (Fig . 6) .

Note: Choices for supplementalspinal fixation systems include, but arenot limited to, Stryker Spine pediclescrew and rod or plate systems (Xia,Radius, Trio, Techtonix, andMantis). Please refer to the respectivesurgical techniques of the abovementioned Stryker Spine supplementalfixation systems for additionalinformation on implantation.

Dynamic Distractor48394005Extension Handle48394006

Extension Handle

Lubricate entire dial and Distal Paddles

Figure 6

Figure 7

AVS Navigator TLIF Surgical Technique

9

3) Insert the Dynamic Distractor into the disc space (Fig .8) . A Slaphammer may be used to aid with insertion (Fig . 9) . It is necessary to ensure that the distractor paddles contact the cortical rim in at least one place, which can be confirmed using f luoroscopy .

Note: Special care should be taken toprevent the distractor from going too far anterior.

4) To distract, rotate clockwise, until the desired height is achieved . Special care should be taken to prevent over distracting the disc space and over torquing the distractor paddles (Fig . 10) .

Slaphammer48061001

7-16mm

Figure 8

Figure 9

Figure 10

AVS Navigator TLIF Surgical Technique

10

5) The numeric dial on the Dynamic Distractor identifies the height of the disc space (Fig . 11) .

6) To remove the Dynamic Distractor, rotate counterclockwise until the distractor is completely closed and the dial reads 7mm . Remove from the disc space (Fig . 12) .

Note: The Dynamic Distractor must be sterilized with the distractor paddles completely open with the dial reading 16mm. Flushing of large debris in paddles may be necessary prior to sterilization. Refer to Cleaning Instructions in Package Insert.

Note: It is highly recommended that shipment of the Distractor be within the AVS Navigator container. During transportation outside of the container, the distractor paddles must be fully closed prior to shipping within a well padded container.

Insertion DistractionFigure 11

Figure 12

AVS Navigator TLIF Surgical Technique

11

Step 5: Discectomy and Endplate Preparation

Access to the disc space is achieved through an annulotomy made lateral to the posterior longitudinal ligament . Using a scalpel, vertical cuts are made parallel to the dura and laterally in the foramen from the endplate of L4 to the endplate of L5 (Fig . 13) . Additional cuts extend horizontally along the endplates of L4 and L5, connecting the vertical cuts . The annulus and any accessible disc material are removed with a pituitary rongeur .

Note: It is recommended that the annulotomy be at least 12mm wide in order to facilitate insertion of the implant, which is 12mm wide.

Commonly, a large osteophyte will protrude inferiorly off the endplate of L4, and sometimes superiorly off the endplate of L5 . Sharp excision of these osteophytes with an osteotome or kerrison punch will provide a larger entry portal if desired .

Note: Throughout the remainder of the procedure, care must be taken to not damage the remaining lateral, anterior, or posterior annulus.

A curette or narrow Cobb elevator is used to elevate disc material from the endplates of L4 and L5 on the left lateral side . Next, the angled curettes are used to elevate the disc from the endplates of L4 and L5 centrally and on the right side . An easily missed portion of the disc lies posteriorly and centrally within the disc space, just ventral to the spinal canal . Special effort should be directed to disc removal in this zone to provide optimal surface area for interbody fusion . Straight and up-biting pituitary rongeurs should be used to remove the disc (Fig . 14) . Additionally, multiple passes with the straight and angled curettes may be necessary to ensure adequate discectomy . Fluoroscopy may help in ensuring an adequate discectomy while limiting the risk of perforating the ventral, lateral, or posteromedial annulus .

Figure 13

Figure 14

AVS Navigator TLIF Surgical Technique

12

Step 6: Sizing the Disc Space

The disc space height can be sized using a series of paddle distractors, reamer distractors or trials (Fig . 15) . The paddle distractor, reamer distractor, or trial size is serially increased until the appropriate fit within the disc space is achieved . The paddle distractor, reamer distractor, or trial should fit snugly within the disc space with distraction released, but care must be taken to not oversize the implant . Undersizing may contribute to loss of lordosis and inadequate stability for fusion, while oversizing may result in difficulty inserting the implant and disruption of the vertebral boney endplate with subsequent graft subsidence .

TrialsAttach the chosen AVS Navigator Trial to the T-Handle and insert into the disc space (Fig . 16) .

The AVS Navigator Trials range from 6mm - 15mm, are 26mm in length, and 10mm in width . A ridge exists on the post of the Trial at 31mm, and is designed to assist the surgeon in determining which implant length is appropriate from radiographic images . The 26mm and 31mm lengths correspond to implant lengths (Fig . 17) .

Note: The trial width is 2mm narrower than the implant.

Note: Be sure to trial the anterior third of the disc space, as this will be the implant’s final position.

Note: AVS Navigator trials were not designed to be used as paddle distraction devices (Fig. 18). For paddle distractors or reamer distractors please consult the Surgical Techniques for Reliance PLIF or Reliance LITE. AVS Navigator trials are made distinctly to measure the height and length of your interbody spacer to the disc space.

T-Handle48061000

AVS Navigator Trial483922(06-15)

26mm31mm

26mm31mm

Figure 15

Figure 16

Figure 17

Figure 18

AVS Navigator TLIF Surgical Technique

13

Step 7: AVS Navigator Implant Preparation

Select the appropriately sized AVS Navigator Implant and assemble it to the desired AVS Navigator Inserter – Straight or Bayoneted (Fig. 19) .

Prior to assembly, confirm the following:

1) The slide lock is in the “rotate” position . To change the position of the button, first press down, then slide forward to lock or backwards to rotate (Fig . 20) .

2) The locking knob is to be turned counterclockwise all the way (Fig . 21) .

To assemble the implant to the inserter, one of the following two technique options is recommended .

Option 1: Assembly Using Assembly Graft Block

Choose appropriate Assembly Graft Block 26 or 31 (Fig . 22) .

• Assembly Graft Block 26 is used with AVS Navigator Implants that are 26mm in length .

• Assembly Graft Block 31 is used with AVS Navigator Implants that are 31mm in length .

Place implant in the appropriate slot . Slots correspond to implant heights 6mm - 15mm .

Line up inserter to:

A) Grab around the back rail (Fig . 23) . B) Rotate to the left to fully grasp the

rail (Fig . 24) .

Proceed to Step 8: Locking the Implant

Assembly Graft Block 26mm 48393003Assembly Graft Block 31mm 48393013

Step A Step B

Straight Inserter 48393001

Bayoneted Inserter 48393002

Side Rail

Rail Slot Back Rail

1SlideLock 2

Locking Knob

Figure 19

Figure 20

Figure 21

Figure 22

Figure 23 Figure 24

AVS Navigator TLIF Surgical Technique

14

Option 2: Manual Assembly Slide implant back rail on inserter track .

Rotate until locking bar and rail slot are aligned (Fig . 25) .

Locking bar and rail slot are aligned (Fig . 26) .

Step 8: Locking the Implant

Provisionally lock the implant by pressing down and pushing the slide lock forward to the “lock” position . This pushes the locking bar into the rail on the implant (Fig . 27) .

Rotate the locking knob clockwise until the outer sleeve on the distal end of the implant is flush with the implant and the locking knob is in the “lock” range (Fig . 28) .

The hard stop on the system is the tactile feel the operator feels when rotating the knob . This may happen at the beginning of the lock band or the middle, or the end . Do not over tighten . Stop once it feels secure .

Final Lock

Locking Knob

LockRange

Rail Slot

Back Rail

Locking Bar

Inserter Track

Figure 25

Figure 26

Figure 27

Figure 28

AVS Navigator TLIF Surgical Technique

15

Place the AVS Navigator Implant in the Assembly Graft Block and pack it with autogenous bone graft using the Impactor (Fig . 29 and 30) . The compacted graft should be flush with the upper and lower surfaces of the implant, to later be in contact with the vertebral endplates .

Impactor48393004

Step 9: AVS Navigator Implant Insertion

The AVS Navigator Inserters are specifically designed to allow a surgeon to impact into the disc space with a fixed engagement between the inserter and implant, and then activate a rotational component between the inserter and implant for final placement .

1) Prior to insertion, confirm the inserter is fully locked to the implant and the rail is directed medial . At this position, the implant and inserter are fixed in position (Fig . 31) .

2) Carefully insert the AVS Navigator Implant gently and progressively into the disc space (Fig . 32) .

3) Optimal positioning may be facilitated by first directing the implant obliquely until it contacts the ventral annulus (Fig . 33) .

4) Maintain posterior distraction until the implant clears the posterior annulus .

Note: Fluoroscopy may be useful in determining the appropriate trajectory for insertion of the implant and appropriate final positioning. Care should be taken to prevent the instrument from pushing too far anteriorly.

AVS Navigator Graft Volume (in cc)26mm 31mm

6mm 0 .34 0 .417mm 0 .39 0 .488mm 0 .44 0 .549mm 0 .49 0 .6110mm 0 .55 0 .6711mm 0 .60 0 .7412mm 0 .65 0 .8013mm 0 .70 0 .8714mm 0 .76 0 .9415mm 0 .81 1 .00

Assembly Graft Block48393003 26mm48393013 31mm

Figure 29

Figure 30

Figure 31

Figure 32

Figure 33

AVS Navigator TLIF Surgical Technique

16

5) To activate the rotational component, release the provisional lock by

A . Pressing down on the slide lock

and moving to the “rotate” position (Fig . 34) .

B . Turning the locking knob a 1/4 turn counterclockwise from “lock” to “rotate (Fig . 35) .”

Note: Only a quarter turn is needed to initiate the rotation of the implant. If desired, the surgeon may retighten the locking knob at any rotational position to further impact the inserter.

At this position, the implant is designed to rotate on the rail as the leading edge contacts the ventral annulus (Fig . 36) .

Figure 34

Figure 35

Figure 36

AVS Navigator TLIF Surgical Technique

17

6) Continue to impact gently and progressively into the disc space until the implant reaches desired positioning (Fig . 37) .

Note: If difficulty is encountered, it most likely represents a mechanical block to the passage of the implant. Check for adequacy of the discectomy and be sure distraction is maintained. Then, reposition the implant as needed. The distraction may be released slightly to facilitate final positioning.

Note: When deemed medically necessary, for intraoperative rescue when the inserter is still attached to the implant, turn the locking knob clockwise to the “lock” position and use a Slaphammer to remove (Fig. 38) .

Figure 37

Figure 38

AVS Navigator TLIF Surgical Technique

18

Assembly Graft Block 26mm 48393003Assembly Graft Block 31mm 48393013

Step A Step B

Side Rail

Rail Slot Back Rail

Alternative Inserter Option: AVS Navigator Straight or Bayoneted Inserters with the Final Lock Knob

Note: If using the AVS Navigator Straight or Bayoneted Inserter with the final lock knob, please use the following steps:

1. Prior to assembly, confirm the following:

1) The slide lock is in the “rotate” position (Fig . 39) .

2) The locking knob is to be turned counterclockwise all the way . This can be visually confirmed by identifying that the outer sleeve is flush with the main sleeve (Fig . 40) .

2. To assemble the implant to the inserter, one of the following two technique options is recommended .

Option A: Assembly Using Assembly Graft Block Choose appropriate Assembly Graft Block 26 or 31 (Fig . 41) .

• Assembly Graft Block 26 is used with AVS Navigator Implants that are 26mm in length .

• Assembly Graft Block 31 is used with AVS Navigator Implants that are 31mm in length .

Place implant in the appropriate slot . Slots correspond to implant heights 6mm - 15mm .

Line up inserter to:

A) Grab around the back rail (Fig . 42) . B) Rotate to the left to fully grasp the rail

(Fig . 43) .

Proceed to Step 3: Locking the Implant

21 Final LockSlide

Lock

LockingKnob

Figure 39 Figure 40

Figure 41

Figure 42 Figure 43

AVS Navigator TLIF Surgical Technique

19

Option B: Manual Assembly Slide implant back rail on inserter track .

Rotate until locking bar and rail slot are aligned (Fig . 44) .

Locking bar and rail slot are aligned (Fig . 45) .

3. Locking the Implant

Provisionally lock the implant to the inserter by pushing the slide lock forward to the “lock” position . This pushes the locking bar into the rail slot on the implant . Tighten the final lock (Fig . 46) .

Rotate the locking knob clockwise until the outer sleeve of the inserter is flush with the implant and the locking knob is in the “lock” range (Fig . 47) .

LockingScrew

Final Lock

Locking Knob

LockRange

Rail Slot

Back Rail

Locking Bar

Inserter Track

Figure 44

Figure 45

Figure 46

Figure 47

AVS Navigator TLIF Surgical Technique

20

4. AVS Navigator Implant Insertion

The AVS Navigator Inserters are specifically designed to allow a surgeon to impact into the disc space with a fixed engagement between the inserter and implant, and then activate a rotational component between the inserter and implant for final placement .

1) Prior to insertion, confirm the inserter is fully locked to the implant and the rail is directed medial . At this position, the implant and inserter are fixed in position (Fig . 48) .

2) Carefully insert the AVS Navigator Implant gently and progressively into the disc space (Fig . 49) .

3) Optimal positioning may be facilitated by first directing the implant obliquely until it contacts the ventral annulus (Fig . 50) .

4) Maintain posterior distraction until the implant clears the posterior annulus .

Note: Fluoroscopy may be useful in determining the appropriate trajectory for insertion of the implant and appropriate final positioning. Care should be taken to prevent the instrument from pushing too far anteriorly.

Figure 48

Figure 49

Figure 50

AVS Navigator TLIF Surgical Technique

21

5) To activate the rotational component, release the provisional lock by

A . Unlocking the final lock by

turning counterclockwise . (Fig . 51)

B . Pulling the slide lock to the “rotate” position (Fig . 52) .

C . Turn the locking knob a 1/4 turn counterclockwise from “lock” to “rotate (Fig . 53) .”

Note: Only a quarter turn is needed to initiate the rotation of the implant. If desired, the surgeon may retighten the locking knob at any rotational position to further impact.

At this position, the implant is designed to rotate on the rail as the leading edge contacts the ventral annulus (Fig . 54) .

Figure 51

Figure 52

Figure 53

Figure 54

AVS Navigator TLIF Surgical Technique

22

5. Inserter Removal

To disengage and remove the inserter, turn the locking knob counterclockwise fully until it stops, or if visible, turn until the main shaft is flush with the outer shaft (Fig . 57) . A hard stop will be felt . Once it is, the system has been fully released . Gently remove the inserter from the disc space .

Note: It is important to fully unlock the locking knob in order to prevent partial removal or compromise of the implant’s final position.

6) Continue to impact gently and progressively into the disc space until the implant reaches desired positioning (Fig . 55) .

Note: If difficulty is encountered, it most likely represents a mechanical block to the passage of the implant. Check for adequacy of the discectomy and be sure distraction is maintained. Then, reposition the implant as needed. The distraction may be released slightly to facilitate final positioning.

Note: When deemed medically necessary, for intraoperative rescue when the inserter is still attached to the implant, turn the locking knob clockwise to the “lock” position and use a Slaphammer to remove (Fig. 56) .

Figure 55

Figure 56

Figure 57

AVS Navigator TLIF Surgical Technique

23

Footed Tamp 48393007

Straight Tamp 48393008

Based on surgeon preference, tamps, straight and footed, can be used to gently tamp the implant to its final position (Fig . 60) . The Straight Tamp is designed to engage the rail for posterior adjustments (Fig . 58) . The Footed Tamp is designed for midline adjustments (Fig . 59) .

There are three tantalum markers embedded within the PEEK material to help visually confirm its position under fluoroscopy (Fig . 61) .

• A 4mm tantalum marker is positioned as the leading marker towards the nose of the implant . This marker is 6 .8mm from the nose of the implant .

• A 2 .5mm tantalum marker is positioned anteriorly, and is 2 .2mm from the most anterior aspect of the implant .

• A 3mm tantalum marker is positioned on the trailing end, and is 6 .9mm from the back rail of the implant .

2.2mm

6.8mm 6.9mm

Figure 58

Figure 60

Figure 61

Figure 59

AVS Navigator TLIF Surgical Technique

24

Step 10: Placement of Bone Graft

Autogenous bone graft may be placed lateral, ventral, and/or dorsal to the implanted AVS Navigator Implant .

INTRAOPERATIVE RESCUE

If the surgeon chooses to realign or remove the implant prior to releasing the implant, and the implant is still engaged with the inserter, turn the locking dial clockwise to tighten back onto the implant . The implant may then be removed by using the Slaphammer to reverse impact the implant out of the disc space (Fig . 62) .

If the implant is disengaged from the inserter, the inserter should not be reattached to the implant for rescue . In this scenario, a blunt nerve hook may be used to remove the implant .

Step 11: Posterior Fusion

As the interbody portion of the procedure is now complete, distraction can be released . If pedicle screws had not been inserted earlier in the procedure, they may be inserted at this point . Vertical rods are connected to the screws and gentle compression is applied between the L4 and L5 screws bilaterally . This facilitates restoration of segmental lordosis and compression of the AVS Navigator Implant and bone graft . Apply appropriate final torque tightening to all screws . The remaining bone graft may be placed in the decorticated right L4-L5 facet to promote fusion . Intraoperative radiographs should be performed to confirm satisfactory position of the AVS Navigator Implant, pedicle screws, and bone graft (Fig . 63) .

Step 12: Closure

Check the left L4-L5 foramen and TLIF site for any bone fragments or extraneous soft tissue . Once satisfactory decompression of the exiting and traversing nerve roots is confirmed, the wound should be closed .

Figure 62

Figure 63

AVS Navigator TLIF Surgical Technique

25

AVS Navigator TLIF Surgical Technique

26

Top Insert

Middle Insert

Bottom Insert

AVS Navigator TLIF Surgical Technique

27

Top Insert

Dynamic Distractor Container

Bottom Insert

AVS Navigator TLIF Surgical Technique

28

Instruments

Reference # Instrument Description

48393001 Straight Inserter

48393002 Bayoneted Inserter

48392206 6mm Trial, 0° Lordosis

48392207 7mm Trial, 0° Lordosis

48392208 8mm Trial, 0° Lordosis

48392209 9mm Trial, 0° Lordosis

48392210 10mm Trial, 0° Lordosis

48392211 11mm Trial, 0° Lordosis

48392212 12mm Trial, 0° Lordosis

48392213 13mm Trial, 0° Lordosis

48392214 14mm Trial, 0° Lordosis

48392215 15mm Trial, 0° Lordosis

48061000 Reliance T-Handle

48393003 Assembly Graft Block 26mm

48393013 Assembly Graft Block 31mm

48393004 Impactor

48394005 Dynamic Distractor

48394006 Extension Handle

48061001 Reliance LITE Slaphammer

48393007 Footed Tamp

48393008 Straight Tamp

48393010 Container

48393011 Dynamic Distractor Container

AVS Navigator TLIF Surgical Technique

29

Implants

Reference # Implant Description

48392060 6mm x 26mm x 0° AVS Navigator Implant

48392070 7mm x 26mm x 0° AVS Navigator Implant

48392080 8mm x 26mm x 0° AVS Navigator Implant

48392090 9mm x 26mm x 0° AVS Navigator Implant

48392100 10mm x 26mm x 0° AVS Navigator Implant

48392110 11mm x 26mm x 0° AVS Navigator Implant

48392120 12mm x 26mm x 0° AVS Navigator Implant

48392130 13mm x 26mm x 0° AVS Navigator Implant

48392140 14mm x 26mm x 0° AVS Navigator Implant

48392150 15mm x 26mm x 0° AVS Navigator Implant

48392064 6mm x 26mm x 4° AVS Navigator Implant

48392074 7mm x 26mm x 4° AVS Navigator Implant

48392084 8mm x 26mm x 4° AVS Navigator Implant

48392094 9mm x 26mm x 4° AVS Navigator Implant

48392104 10mm x 26mm x 4° AVS Navigator Implant

48392114 11mm x 26mm x 4° AVS Navigator Implant

48392124 12mm x 26mm x 4° AVS Navigator Implant

48392134 13mm x 26mm x 4° AVS Navigator Implant

48392144 14mm x 26mm x 4° AVS Navigator Implant

48392154 15mm x 26mm x 4° AVS Navigator Implant

AVS Navigator TLIF Surgical Technique

30

Implants

Reference # Implant Description

48393060 6mm x 31mm x 0° AVS Navigator Implant

48393070 7mm x 31mm x 0° AVS Navigator Implant

48393080 8mm x 31mm x 0° AVS Navigator Implant

48393090 9mm x 31mm x 0° AVS Navigator Implant

48393100 10mm x 31mm x 0° AVS Navigator Implant

48393110 11mm x 31mm x 0° AVS Navigator Implant

48393120 12mm x 31mm x 0° AVS Navigator Implant

48393130 13mm x 31mm x 0° AVS Navigator Implant

48393140 14mm x 31mm x 0° AVS Navigator Implant

48393150 15mm x 31mm x 0° AVS Navigator Implant

48393064 6mm x 31mm x 4° AVS Navigator Implant

48393074 7mm x 31mm x 4° AVS Navigator Implant

48393084 8mm x 31mm x 4° AVS Navigator Implant

48393094 9mm x 31mm x 4° AVS Navigator Implant

48393104 10mm x 31mm x 4° AVS Navigator Implant

48393114 11mm x 31mm x 4° AVS Navigator Implant

48393124 12mm x 31mm x 4° AVS Navigator Implant

48393134 13mm x 31mm x 4° AVS Navigator Implant

48393144 14mm x 31mm x 4° AVS Navigator Implant

48393154 15mm x 31mm x 4° AVS Navigator Implant

AVS Navigator TLIF Surgical Technique

31

IMPORTANT PRODUCT INFORMATION FORSTRYKER SPINE AVS® NAVIGATOR PEEK SPACERS

DESCRIPTIONThe AVS® Navigator PEEK Spacers are intended for use as interbody fusion devices . They are offered in a variety of lengths, heights and lordotic angles . The hollow implant has serrations on the top and bottom for fixation . Radiopaque markers have been embedded within the implant to help allow for visualization in radiographic images .

MATERIALAll components of all AVS® PEEK Spacers systems are manufactured out of the following materials:• Implant: Polyetheretherketone (PEEK)

(ASTM F2026)• Radiopaque markers: Tantalum (ASTM

F560)

INDICATIONSThe Stryker Spine AVS® Navigator PEEK Spacers are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1 .

DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies . The DDD patients may also have up to Grade I spondylolisthesis at the involvedlevel(s) . These patients should be skeletally mature and have six months of nonoperative therapy .

Additionally, the AVS® Navigator PEEK Spacers can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis .

The AVS® Navigator PEEK Spacers are to be implanted via a posterior or posterolateral approach .

The AVS® Navigator PEEK Spacers are intended to be used with supplemental fixation systems that havebeen cleared for use in the lumbosacral spine .

GENERAL CONDITIONS OF USEThe implantation of intervertebral body fusion devices must be performed only by experienced spinal surgeons having undergone the necessary specific training in the use of such systems because this is a technically demanding procedure presenting a risk of serious injury to the patient .

The information contained in the Package Insert is necessary but not sufficient for the use of this device . This information is in no sense intended as a substitute for the professional judgment, skill and experience of the surgeon in careful patient selection, preoperative planning and device selection, knowledge of the anatomy and biomechanics of the spine, understanding of the materials and the mechanical characteristics of the implants used, training and skill in spinal surgery and the use of associated instruments for implantation, securing the patient’s cooperation in following an appropriately defined post-operative management program and conducting scheduled post-operative follow-up examinations .

CAUTION• Federal (USA) law restricts this device to

sale by or on the order of a physician (or properly licensed practitioner) with appropriate training or experience .

• Based on the fatigue testing results, the physician/surgeon must consider the levels of implantation, patient weight, patient activity level, other patient conditions, etc . which may impact on the performance of the intervertebral body fusion device .

• The implantation of the intervertebral body fusion device must be performed only by experienced spinal surgeons with specific training in the use of this device because this is a technically demanding procedure presenting a risk of serious injury to the patient .

• Potential risks identified with the use of this intervertebral body fusion device, which may require additional surgery, include: device component fracture, loss of fixation, pseudoarthrosis (i .e . non-union), fracture of the vertebrae, neurological injury, and vascular or visceral injury .

• Specialized instruments are provided by STRYKER Spine and must be used to assure accurate implantation of the intervertebral body fusion device . While rare, intraoperative fracture or breakage of instruments can occur, instruments, which have experienced extensive use or extensive force, are more susceptible to fracture depending on the operative precaution, number of procedures, and disposal attention . Instruments must be examined for wear or damage prior to surgery . Instruments for implantation of the AVS® PEEK Spacers are provided non-sterile and must be sterilized prior to use .

• The components of the system should not be used with components of any other system or manufacturer . Any such use will negate the responsibility of Stryker Spine for the performance of the resulting mixed component implant .

• The AVS® PEEK Spacers have not been evaluated for safety and compatibility in the MR environment . The AVS® PEEK Spacers have not been tested for heating or migration in the MR environment .

INFECTIONTransient bacteremia can occur in daily life . Dental manipulation, endoscopic examination and other minor surgical procedures have been associated with transient bacteremia . To help prevent infection at the implant site, it may be advisable to use antibiotic prophylaxis before and after such procedures .

INSTRUMENTSInstruments are provided by STRYKER Spine and must be used to assure accurate implantation of the device . While rare, intraoperative fracture or breakage of instruments can occur . Instruments which have experienced extensive use or extensive force are more susceptible to fracture depending on the operative precaution, number of procedures, disposal attention . Instruments must be examined for wear or damage prior to surgery .

REUSENever reuse or re-implant spinal surgical implants . These could become contaminated resulting in infection . In addition, even though the device appears undamaged, it may have small defects which could compromise structural integrity reducing its service life and/or leading to patient injury .

Surgeons must verify that the instruments are in good condition and operating order prior to use during surgery .

HANDLINGCorrect handling of the implant is extremely important . The operating surgeon must avoid notching or scratching the device .

ALLERGY AND HYPERSENSITIVITY TO FOREIGN BODIESWhen hypersensitivity is suspected or proven, it is highly recommended that the tolerance of the skin to the materials that make up the implants be checked before they are implanted .

CONTRA-INDICATIONSContraindications may be relative or absolute . The choice of a particular device must be carefully weighed against the patient’s overall evaluation . Circumstances listed below may reduce the chances of a successful outcome:• The AVS® PEEK Spacers should not be

implanted in patients with an active infection at the operative site .

AVS Navigator TLIF Surgical Technique

32

• The AVS® PEEK Spacers are not intended for use except as indicated .

• Marked local inflammation .• Any abnormality present which affects the

normal process of bone remodeling including, but not limited to, severe osteoporosis involving the spine, bone absorption, osteopenia, primary or metastatic tumors involving the spine, active infection at the site or certain metabolic disorders affecting osteogenesis .

• Any mental or neuromuscular disorder which would create an unacceptable risk of fixation failure or complications in postoperative care .

• Open wounds .• Pregnancy .• Inadequate tissue coverage over the

operative site .• Any neuromuscular deficit which places an

unsafe load level on the device during the healing period .

• Obesity . An overweight or obese patient can produce loads on the spinal system which can lead to failure of the fixation of the device or to failure of the device itself . Obesity is defined according to the W .H .O . standards .

• A condition of senility, mental illness, or substance abuse . These conditions, among others, may cause the patient to ignore certain necessary limitations and precautions in the use of the implant, leading to failure or other complications .

• Foreign body sensitivity . Where material sensitivity is suspected, appropriate tests must be made prior to material selection or implantation .

• Other medical or surgical condition which would preclude the potential benefit of spinal implant surgery, such as the presence of tumors, congenital abnormalities, elevation of sedimentation rate unexplained by other diseases, elevation of white blood cell count (WBC), or marked left shift in the WBC differential count .

• Prior fusion at the levels to be treated

These contra-indications can be relative or absolute and must be taken into account by the physician when making his decision . The above list is not exhaustive . Surgeons must discuss the relative contraindications with the patients .

INFORMATION FOR PATIENTSThe surgeon must discuss all physical and psychological limitations inherent to the use of the device with the patient . This includes the rehabilitation regimen, physical therapy, and wearing an appropriate orthosis as prescribed by the physician . Particular discussion should be directed to the issues of premature weight bearing, activity levels, and the necessity for periodic medical follow-up .

The surgeon must warn the patient of the surgical risks and make them aware of possible adverse effects .

The surgeon must warn the patient that the device cannot and does not replicate the flexibility, strength, reliability or durability of normal healthy bone, that the implant can break or become damaged as a result of strenuous activity or trauma, and that the device may need to be replaced in the future . If the patient is involved in an occupation or activity which applies inordinate stress upon the implant (e .g ., substantial walking, running, lifting, or muscle strain) the surgeon must advice the patient that resultant forces can cause failure of the device . Patients who smoke have been shown to have an increased incidence of nonunions . Such patients should be advised of this fact and warned of the potential consequences . For diseased patients with degenerative disease, the progression of degenerative disease may be so advanced at the time of implantation that it may substantially decrease the expected useful life of the appliance . In such cases, orthopaedic devices may be considered only as a delaying technique or to provide temporary relief . Patients with previous spinal surgery at the level(s) to be treated may have different clinical outcomes compared to those without a previous surgery .

PRE-OPERATIVE PRECAUTIONSThe surgical indication and the choice of implants must take into account certain important criteria such as:• Patients involved in an occupation or

activity that applies excessive loading upon the implant (e .g ., substantial walking, running, lifting, or muscle strain) may be at increased risk for failure of the fusion and/or the device .

• Surgeons must instruct patients in detail about the limitations of the implants, including, but not limited to, the impact of excessive loading through patient weight or activity, and be taught to govern their activities accordingly . The procedure will not restore function to the level expected with a normal, healthy spine, and the patient should not have unrealistic functional expectations .

• A condition of senility, mental illness, chemical dependence or alcoholism . These conditions among others may cause the patients to ignore certain necessary limitations and precautions in the use of the implant, leading to failure and other complications .

• Foreign body sensitivity . Where material sensitivity is suspected appropriate tests should be made prior to material implantation .

• Surgeons must advise patients who smoke have been shown to have an increased incidence of nonunions . Such patients must be advised of this fact and warned of the potential consequences .

• Care must be taken to protect the components from being marred, nicked, or notched as a result of contact with metal or abrasive objects .

THE CHOICE OF IMPLANTSThe choice of proper shape, size and design of the implant for each patient is crucial to the success of the surgery . The surgeon is responsible for this choice, which depends on each patient .

Patients who are overweight may add additional stresses and strains on the device which can lead to implant fatigue and/or lead to deformation or failure of the implants .

The size and shape of the bone structures determine the size, shape and type of the implants . Once implanted, the implants are subjected to stresses and strains . These repeated stresses on the implants must be taken into consideration by the surgeon at the time of the choice of the implant, during implantation as well as in the post-operative follow-up period . Indeed, the stresses and strains on the implants may cause fatigue, fracture or deformation of the implants, before the bone graft has become completely consolidated . This may result in further side effects or necessitate the early removal of the osteosynthesis device .

INTRA-OPERATIVE PRECAUTIONS• The insertion of the implants must be

carried out using instruments designed and provided for this purpose and in accordance with the specific implantation instructions for each implant . Those detailed instructions are provided in the surgical technique brochure supplied by STRYKER Spine .

• Discard all damaged or mishandled implants .

• Never reuse an implant, even though it may appear undamaged .

PATIENT CARE FOLLOWING TREATMENTPrior to adequate maturation of the fusion mass, implanted spinal instrumentation may need additional help to accommodate full load bearing . External support may be recommended by the physician from two to four months postoperatively or until x-rays or other procedures confirm adequate maturation of the fusion mass; external immobilization by bracing or casting be employed . Surgeons must instruct patients regarding appropriate and restricted

AVS Navigator TLIF Surgical Technique

33

activities during consolidation and maturation for the fusion mass in order to prevent placing excessive stress on the implants which may lead to fixation or implant failure and accompanying clinical problems . Surgeons must instruct patients to report any unusual changes of the operative site to his/her physician . The physician must closely monitor the patient if a change at the site has been detected .

ADVERSE EFFECTSInclude but are not limited to:• Late bone fusion or no visible fusion mass

and pseudarthrosis;• While the expected life of spinal implant

components is difficult to estimate, it is finite . These components are made of foreign materials which are placed within the body for the potential fusion of the spine and reduction of pain . However, due to the many biological, mechanical and physicochemical factors which affect these devices but cannot be evaluated in vivo, the components cannot be expected to indefinitely withstand the activity level and loads of normal healthy bone;

• Superficial or deep-set infection and inflammatory phenomena;

• Allergic reactions to the implanted materials, although uncommon, can occur;

• Decrease in bone density due to stress shielding;

• Dural leak requiring surgical repair;• Peripheral neuropathies, nerve damage,

heterotopic bone formation and neurovascular compromise, including paralysis, loss of bowel or bladder function, or foot-drop may occur .

• Cessation of growth of the fused portion of the spine;

• Loss of proper spinal curvature, correction, height and/or reduction;

• Delayed Union or Nonunion: Internal fixation appliances are load sharing devices which are used to obtain alignment until normal healing occurs . In the event that healing is delayed, does not occur, or failure to immobilize the delayed/nonunion results, the implant will be subject to excessive and repeated stresses which can eventually cause loosening, bending or fatigue fracture . The degree or success of union, loads produced by weight bearing, and activity levels will, among other conditions, dictate the longevity of the implant . If a nonunion develops or if the implants loosen, bend or break, the device(s) must be revised or removed immediately before serious injury occurs;

• Neurological and spinal dura mater lesions from surgical trauma;

• Early loosening may result from inadequate initial fixation, latent infection, premature loading of the device or trauma . Late loosening may result from trauma, infection, biological complications or

mechanical problems, with the subsequent possibility of bone erosion, or pain .

• Serious complications may occur with any spinal surgery . These complications include, but are not limited to, genitourinary disorders; gastrointestinal disorders; vascular disorders, including thrombus; bronchopulmonary disorders, including emboli; bursitis, hemorrhage, myocardial infarction, infection, paralysis or death .

• Inappropriate or improper surgical placement of this device may cause distraction or stress shielding of the graft or fusion mass . This may contribute to failure of an adequate fusion mass to form .

• Intraoperative fissure, fracture, or perforation of the spine can occur due to implantation of the components . Postoperative fracture of bone graft or the intervertebral body above or below the level of surgery can occur due to trauma, the presence of defects, or poor bone stock .

Adverse effects may necessitate reoperation or revision .

The surgeon must warn the patient of these adverse effects as deemed necessary .

REMOVALIf fusion / bone graft growth occurs, the device will be deeply integrated into the bony tissues . As a result, the AVS® PEEK Spacers are not intended to be removed unless the management of a complication or adverse event requires the removal . Any decision by a physician to remove the device should take into consideration such factors as:• The risk to the patient of the additional

surgical procedure as well as the difficulty of removal .

• Migration of the implant, with subsequent pain and/or neurological, articular or soft tissue lesions

• Pain or abnormal sensations due to the presence of the implants

• Infection or inflammatory reactions• Reduction in bone density due to the

different distribution of mechanical and physiological stresses and strains

PACKAGING• The implants are single use devices,

provided non-sterile, and delivered in individual packages . The typical packaging used is clear plastic tubes and polyethylene bags . The packages must be intact at the time of receipt .

• Implants must be removed entirely from their packaging prior to sterilization .

• The implants may also be supplied as a complete set: implants and instruments are arranged on trays and placed in specially designed storage boxes

FURTHER INFORMATIONA surgical technique brochure is available on request through your STRYKER representative or directly from STRYKER Spine . Users with brochures that are over two years old at the time of surgery are advised to ask for an updated version .

COMPLAINTSAny health professional having a complaint or grounds for dissatisfaction relating to the identity, quality, durability, reliability, safety, effectiveness or performance of a device should notify STRYKER Spine or its representative . Moreover, if a device has malfunctioned, or is suspected of having malfunctioned, STRYKER Spine or its representative must be advised immediately .

If a STRYKER Spine product has ever worked improperly and could have caused or contributed to the death of or serious injury to a patient, the distributor or STRYKER Spine must be informed as soon as possible by telephone, fax or in writing .

For all complaints, please include the name and reference along with the batch number of the component(s), your name and address and an exhaustive description of the event to help STRYKER Spine understand the causes of the complaint .

For further information regarding services, please contact:

STRYKER Spine2 Pearl Court, Allendale, NJ 07401-1677 USATel: +1-201-749-8000

AVS Navigator TLIF Surgical Technique

34

Description / Material CompositionSurgical instruments supplied by STRYKER Spine are manual medical tools designed solely for use in the fitting of STRYKER Spine implants . They are made of different materials including stainless steel, aluminum, titanium and plastics (silicone, acetal, etc) that comply with the standards applicable to their specific material composition . However these materials are not implantable . STRYKER Spine instruments do not contain natural rubber (such as: natural rubber latex, dry natural rubber, and synthetic latex or synthetic rubber that contains natural rubber in its formulation) .

Intended Use The AVS Navigator Inserter instruments (references #48393001 and #48393002) are intended to be used exclusively with the AVS Navigator PEEK Spacers .

STRYKER Spine instruments must be used in the manner described in the Surgical Techniques brochures provided by STRYKER Spine . Unless labelled for single use, this instrument may be re-used . Prior to using the instruments, the surgeon shall have given careful consideration to all aspects of the surgical intervention as well as to the limits of the instrumentation . Recommendations for use are provided in the Surgical Technique brochures available from STRYKER Spine representatives .

Under no circumstances may STRYKER Spine instruments be used to fit components from other manufacturers, if STRYKER Spine instruments are used in such cases the company shall not be liable for the operation of the resulting unit . Likewise, STRYKER Spine implants may only be implanted with the appropriate instruments supplied by STRYKER Spine . STRYKER Spine shall not be liable for implants implanted using other manufacturer’s instruments .

ReuseInstruments labelled as single use shall not be reused . While a single-use instrument may appear undamaged, the instrument may have acquired contaminants that compromise sterility and/or blemishes, nicks or latent compromise of its integrity .

Potential Adverse EffectsIncorrect maintenance, cleaning or handling may render the instruments unsuitable for

their intended use, cause corrosion, dismantling, distortion and/or breakage or cause injury to the patient or operating staff .

Below is a list, albeit not exhaustive, of potential complications:• Neurological lesion, paralysis, pain, lesion

of the soft tissues, the visceral organs or the joints, in the event of incorrect use or breakage of the instruments .

• Infection, if the instruments are not properly cleaned and sterilized .

• Dural leaks, compression of vessels, damage to nerves or nearby organs as a result of slippage or poor positioning of a faulty instrument .

• Damage caused by the involuntary releasing of the springs of certain instruments .

• Damage caused by the instruments used to bend or cut in-situ due to excessive forces occurring when they are used .

• Cutting the gloves or the skin of surgical staff .

• Tissue lesions on the patient or surgical staff and/or an increase in operating time as a result of having to dissemble the instruments during surgery .

• Crack, fracture or involuntary perforation of the bone .

As a result of the mechanical features required, most of the instruments are made of non implantable materials . In the event an instrument breaks, no fragment must remain in the patient as this could cause post-operative complications such as allergies, infections, or complications of a biological nature associated with the release of metal components, possibly requiring further intervention .

Packaging• STRYKER Spine instruments may be

supplied either sterile or non-sterile in instrument containers or individual packaging . Instruments sold sterile are always presented in individual packaging and will be clearly labelled as sterile on the package label . The sterilization method applied is indicated on the product label . The sterile instrument packaging must be intact at the time of receipt and the integrity of the packaging shall be checked prior to use .

The containers and the packaging of the instruments must be intact when received . The packaging materials must be removed prior to sterilization .

AVS NAVIGATOR INSERTER INSTRUMENTATION INSTRUCTIONS FOR USE

STRYKER SPINE AVS NAVIGATOR INSERTERSNON STERILE PRODUCTINSTRUCTIONS FOR USE PACKAGE INSERT

Examination Prior to UseFor instruments sold sterile:• Before utilization, it is necessary to verify

the sterility expiration date, which is indicated as the “use by” date . Stryker Spine cannot be held responsible for use of its products after their expiration date .

• Pre-sterilized instruments must not be allowed to come into contact with any disinfectant product or solution .

• It is recommended to verify the integrity of the original package before use . Sterility is ensured only if there is no trace of damage to the packaging .

• In case of damage to the packaging, or after opening of the packaging, re-sterilization of the instrument is strictly forbidden, regardless of the method that might be employed .

• Instruments should be visually examined for damage by doctors and staff in operating centers prior to and after surgery .

For instruments designed for single use (sold non-sterile):• It is recommended to verify the integrity of

the instrument and original package before use .

• Instruments should be visually examined for damage by doctors and staff in operating centers prior to surgery .

For instruments designed for re-use:• The life of the instrument depends on the

number of times they are used as well as the precautions taken in handling, cleaning and storage . Great care must be taken of the instruments to ensure that they remain in good working order .

• Instruments should be examined for wear or damage by doctors and staff in operating centers prior to surgery .

• The examination shall include a visual and functional inspection of the working surfaces, articulation points, and springs . It should also include verifying all welded connections, that all components are present, and the cleanliness of the orifices and cavities, as well as the absence of any cracks, distortion, impact, corrosion or other change . For instruments with articulations, lubrication may be necessary . Using a silicone lubricating cream is recommended .

• Special attention should be paid to the clamping keys, especially to the hexagonal shank bits . The latter must not be blunt as this could compromise the clamping of the fittings and lead to a risk of detaching .

AVS Navigator TLIF Surgical Technique

35

Similarly instruments designed to cut bone such as reamers, rakers etc, must be rigorously inspected for sharpness .

• Special attention should be paid to the screwdrivers . It is crucial that they are used for the purpose they were designed for, as specifically indicated in the Surgical Technique of each product line .

• Certain surgical intervention requires the use of instruments which include a measuring function . These are to be inspected for wear and the clear visibility of any surface markings .

• STRYKER Spine and its representatives are available to help carry out proper instrument inspections .

• STRYKER Spine shall not be responsible in the event of the use of instruments that are damaged, incomplete, show signs of excessive wear and tear, or that have been repaired or sharpened outside the control of STRYKER Spine . Any faulty instruments must be replaced prior to any intervention .

Recommendations For Instruments Provided Sterile• Products may be stored at room

temperature and withstand the normal transportation conditions .

• Products not labeled as sterile are not sterile .

• Do not use if package is opened or damaged or after the “Use by” date on the label has expired .

• Stryker shall not be responsible for the use of products presenting package deterioration or expiration of shelf life .

• Care must be taken to prevent contamination of instrument after opening of package .

PRE-OPERATIVE PRECAUTIONSAnyone using STRYKER Spine products can obtain a Surgical Technique by requesting one from a distributor or from STRYKER Spine directly . Those using brochures published more than two years before the surgical intervention are advised to request an updated version .

STRYKER Spine devices may only be used by doctors who are fully familiar with the surgical technique required . The doctor operating must take care not to use the instruments to exert inappropriate stress on the spine or the implants and must scrupulously comply with any operating procedure described in the surgical technique provided by STRYKER Spine . For example, the forces exerted when repositioning an instrument in-situ must not be excessive as this is likely to cause injury to the patient .

To reduce the risk of breakage, care must be taken not to distort the implants or nick, hit or score them with the instruments unless otherwise specified by the applicable STRYKER Spine Surgical Technique .Extreme care must be taken when the instruments are used near vital organs, nerves or vessels .

Unless otherwise specified on the label, the instruments can be reused after decontamination, cleaning and sterilization .

Any electrosurgical devices have the potential for providing an ignition source . Do not use in the presence of flammable substances .

CAUTIONFederal law (U .S .A) restricts this device to sale by or on the order of a licensed physician . Similar restrictions may apply in your jurisdiction .

StorageThe instruments are packaged in individual packages or in containers . After they are used they must be stored in a clean, dry and temperate place .

A surgeon must always rely on his or her own professional clinical judgment when deciding whether to use a particular product when treating a particular patient . Stryker does not dispense medical advice and recommends that surgeons be trained in the use of any particular product before using it in surgery .

The information presented is intended to demonstrate the breadth of Stryker product offerings . A surgeon must always refer to the package insert, product label and/or instructions for use before using any Stryker product . Products may not be available in all markets because product availability is subject to the regulatory and/or medical practices in individual markets . Please contact your Stryker representative if you have questions about the availability of Stryker products in your area .

Stryker Corporation or its divisions or other corporate affiliated entities own, use or have applied for the following trademarks or service marks: Aria, AVS Navigator, LITE, Luxor, Mantis, Phantom, Position Matters, Radius, Reliance, Stryker, Techtonix, Trio, UniLIF, Xia . All other trademarks are trademarks of their respective owners or holders .

Not intended for promotional or marketing use outside the United States . TLNAV-ST-1_Rev-4_11834SC/GS 09/16

Copyright © 2016 StrykerPrinted in USA

Stryker Spine2 Pearl Court Allendale, NJ 07401-1677 USAt: 201-749-8000www.stryker.com