Page 1: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

AVEA Enhanced Pulmonary

Mechanics

Page 2: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced SettingsUser Interface Module

Advanced Settings Feature

Page 3: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Accessing Advanced Features

Press “Adv Settings” hard key on left side panel of the Avea monitor to access the Advanced Settings Window

Page 4: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings Window

Yellow triangle indicates this parameter has Advanced Settings

Page 5: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Activating an Advanced Setting

Press the desired parameter. The parameter will then be high-lighted with a blue background.

The Advanced Settings features will then appear in the window above for the parameter

Page 6: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings

Universal Features:

Sensitivity & Pressure Support

Page 7: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Sensitivity: Flow & Pressure

Flow Trigger Advanced SettingsBias Flow

Pressure Trigger Sensitivity

Page 8: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Sensitivity• “Bias Flow”

– Bias Flow enables Flow Triggering– To have adequate Flow Triggering, the Bias Flow

should be set 0.5 liters above the Flow Trigger setting• Example: Bias Flow set at 2.0 litres, Flow Trigger should be

set no higher than 1.5 litres

• “Pressure Trigger”– The Avea has both a pressure and a flow trigger

active at the same time. Whichever signal (flow or pressure) is triggered first, the ventilator will then cycle into inspiration

Page 9: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Pressure Support VentilationPSV Setting has 4 Advanced Settings

Volume Limit

Rise Time

Flow Cycle

%

Max Inspiratory Time

Page 10: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Pressure Support Ventilation• “Volume Limit”

– Every pressure type breath on the Avea has “Volume Limit” as an Advanced Setting feature

– The clinician can enter a maximum volume limit. Once the ventilator has reached this limit during the delivery of the breath, it will then cycle the breath to exhalation using volume as the cycle criteria

– When the volume limit is reached, a visual display appears in the alarm area. The display remains illuminated as long as the volume limit is active

• “PSV Rise Time”– Setting range is from 1 to 9, 1 being fastest and 9 slowest.– Allows the clinician to adjust the speed at which flow is delivered to

the patient during a Pressure Support breath

Page 11: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Pressure Support Ventilation• “PSV Flow Cycle %”

– The Flow Cycle works as a percentage of the peak flow of the PSV breath

– The Flow Cycle can be set in increments of 5% up to 45%• Example: If the flow cycle is set at 10% and the peak flow for the

PSV breath is 60 lpm, the breath will cycle when the decelerating inspiratory flow curve decays to 6 lpm

• “PSV Tmax”– The PSV Maximum Inspiratory Time setting allows for an

adjustable back up inspiratory time setting in circumstances where flow cycling is unreliable (i.e. endotracheal tube leaks)

– PSV Tmax should be set at 1.5-2 X the set Inspiratory Time for Control breaths

– The PSV Tmax is adjustable from 0.2 to 5.0 seconds

Page 12: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings

Pressure Control Ventilation - PCV

Page 13: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

PCV - Pressure Control

Inspiratory Pressure

Inspiratory Time

Flow Trigger

Page 14: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Inspiratory Pressure in PCVTouch Insp Pres and these settings become available

Machine Volume

Volume Limit

Inspiratory Rise Time

Page 15: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings Inspiratory Pressure in PCV

• “Mach Vol” (Machine Volume)– This feature allows the clinician to set a target tidal

volume during a Pressure Control breath– When Machine Volume is active, the “Mach Vol” icon

appears in the modes display window

Machine Volume icon

Page 16: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Machine Volume

• Is a Dual Control Mode; Ventilator delivers Volume Controlled Flow signal and Pressure Control signal to gas delivery system simultaneously.– Sets the minimum tidal volume delivered

from the ventilator in a pressure control breath.

– Always used with the time cycling criterion in pressure control ventilation.

– Circuit compliance compensated in adult and pediatric applications.

Page 17: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Machine Volume

• Allows for the maintenance of traditional Volume Control Ventilation while allowing the ventilator to vary peak flow to meet the patient’s inspiratory flow demands

• Incorporates a decelerating/ variable inspiratory flow waveform

• Can be thought of as a variable flow-volume ventilation

Page 18: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Machine Volume• Ventilator calculates the decelerating inspiratory flow

required to deliver the Machine Volume (Minimum Tidal Volume) in the set inspiratory time.

• When Peak Flow decelerates to this calculated peak inspiratory flow– If Machine Volume has not been delivered

• transition to a continuous flow until the Machine Volume has been delivered– Inspiratory Time will remain constant

» maximum Peak Inspiratory Pressure is determined by the High Peak Pressure alarm setting

– If the Machine Volume is met or exceeded during delivery of the pressure control breath, the ventilator will complete the breath as a normal Pressure Control breath.

Page 19: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Pressure Control- Volume Control Transition

The constant, square wave flow continues until the pre-set tidal volume is delivered

Pressure is increased within the breath

Page 20: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Pressure Control Adjustments with Machine Volume Guarantee

• Pressure Control – Decreased

Compliance• Lung• Chest Wall

– Neuromuscular Blockers

– Increased AW Resistance• Mucous

• Bonchospasm

• Pressure Control – Increased

Compliance• Lung

• Chest Wall

– Spontaneous Breathing

– Decreased AW Resistance• Post Suctioning

• Post Bronchodilators

Page 21: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Machine Volume

Note Inspiratory Pressure set at zero, breath is decelerating flow pressure controlled breath. (Set Mach Volume 350cc) Inspiratory time is constant

Page 22: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Machine Volume

Note transitions in flow between Pressure Control and Volume Control flow waveforms

Page 23: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Machine Volume Vs. PRVC

Machine Volume can respond to changes in compliance within 2-4 msec. PRVC will update on next breath.

Compliance decreases ventilator responds within the breath

Compliance increases ventilator responds within the breath

Page 24: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Inspiratory Pressure in PCV

• “Vol Limit” (Volume Limit)– This feature allows the clinician to set a maximum volume

limit that cannot be exceeded during a Pressure Control breath

– When the volume limit is reached, a visual display appears in the alarm area. The display remains illuminated as long as the volume limit is active

• “Insp Rise” (Inspiratory Rise Time)– Setting range from 1 to 9, 1 being fastest and 9 slowest.– Allows the clinician to adjust the speed at which flow is

delivered to the patient during a Pressure Control breath

Page 25: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Volume Limit• Active in:

– Pressure, – PRVC – Vsync – TCPL– PSV breaths only. – Neonatal applications

Volume LimitTM requires the use of a wye flow sensor.

– If a proximal flow sensor is used • Volume Limit is activated

by the inspiratory tidal volume measured by the wye flow sensor.

• If Volume Limit threshold has been reached:– Alarm status indicator will

change to yellow and display the words Volume Limit.

– Cannot be reset until the ventilator has delivered a breath, which does not meet the Volume Limit threshold.

– To reset the alarm status window use the alarm-reset button

– Do not use with leaky ET tube

Page 26: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Volume limit with Neonatal Flow Sensor

Inspiratory VT limit 16 ml = 8 cc/kg

Volume limit can be used in conjunction with Machine volume to allow volume bracketing in Pressure Control. Machine volume is the minimum volume guarantee and Volume Limit will terminate the breath based on a maximum inspiratory tidal volume threshold. When used with only PCV, as compliance improves, so will volume. Patient will be protected from inadvertent volutrauma. Volume Limit may limit gas flow during hiccups or during increased flow demands of the patient.

2 KG

Page 27: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Inspiratory Time in PCV

“Flow Cycle” is the only Advanced Setting feature for Inspiratory Time.

Page 28: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Inspiratory Time in PCV

• “Flow Cycle”– In addition to time cycling in PCV, the Avea allows the

clinician the ability to add flow cycling to the inspiratory cycle criteria of the PCV breath

– The Flow Cycle works as a percentage of the peak flow of the PCV breath (mimicking a Pressure Support Breath)

– When a Flow Cycle percentage is set, the PCV breath can either be cycled into expiration by the percentage of Flow Cycle or Inspiratory Time depending on which criteria is met first

Page 29: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Flow Cycle• The default setting is 0 (off)

– Can be set in increments of 5% to 45%

–Flow cycle can be more comfortable for spontaneously breathing patients–It can reduce mean airway pressure and VT

100

50

Breath termination at 45%

Breath termination at 5%

Page 30: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Flow Cycle

• Allows patient to determine their own I- time by terminating the breath once a certain percentage of the peak inspiratory flow is met

• May improve preload and eliminate V/Q mismatching

• Improves patient/ventilator dsy-synchrony• May tremendously improve oxygenation

and ventilation in spontaneously breathing patients

Page 31: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Expiratory Flow Cycling

V

T

Peak Inspiratory Flow

5%20%

40%

Page 32: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Fixed Inspiratory Time

No Flow Cycle

Flow Cycle OnFlow Cycled Breath

Page 33: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Flow Cycle

If Flow Cycle is set with Machine Volume the breath will not Flow Cycle unless the Machine Volume (Minimum Tidal Volume) has been delivered

Volume Cycled Breath

Flow Cycled Breath

Page 34: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Patient Comfort

• Dial in a Flow Cycle %– May apply Flow Cycle transiently to measure

pt’s true I time and set IT accordingly– Flow Cycling can only occur if target tidal

volume has been reached in PRVC and Machine Volume

– Should only be used with patients that have low FiO2 requirements because by limiting IT Mean Airway Pressure may be affected

Page 35: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Breath Termination Hierarchy

AlarmSupercedes all other criteria

Volume LimitIf set, supercedes all lesser

criteria

Flow Cycle*In volume targeted and controlled

modes only after set volume is achieved

Time / Volume CycledIf no other criteria intercede

Page 36: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings

Volume Control Ventilation - VCV

Page 37: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

VCV - Volume Control

Volume Peak Flow PSV Flow Trigger

Page 38: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Volume Parameter in VCV

Vsync On/Off

Vsync Rise Time

Sigh Select

Page 39: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Volume Parameter in VCVWhen Vsync is selected, Volume Limit becomes apparent and the Vsync icon appears on screen.

0=Off

1=On

Page 40: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced SettingsVsync

• Vsync operates similar to PRVC– Flow is only difference

• Set in Vsync• Automatic in PRVC

• The maximum step change between two consecutive breaths is 3 cm H2O.

Page 41: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Volume Limit

• This feature allows the clinician to set a maximum volume limit that cannot be exceeded during a Vsync breath.

• When the volume limit is reached, a visual display appears in the alarm area. The display remains illuminated as long as the volume limit is active

Page 42: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced SettingsVolume Parameter in VCV

• “Vsync Rise Time”– Setting range from 1 to 9, 1 being fastest and 9

slowest.– With Vsync active, this control allows the clinician to

adjust the speed at which flow is delivered to the patient.

• “Sigh”– A sigh volume breath is delivered every 100th breath

in place of the next normal volume breath.– The sigh volume breath is 1.5 times the set tidal

volume

Page 43: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced SettingsPeak Flow in VCV

Waveform Type

1 = Ramp0 = Square

Demand FlowOn/Off

Page 44: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Intra-Breath Demand Flow

• Peak Inspiratory Pressure (Ppeak) every 2 milliseconds throughout the breath cycle

• “Virtual” Pressure Support Target of the greater of: – PEEP + 2 cmH2O or

– Ppeak – 2 cmH2O • Maximum PEEP x 2

• Ppeak decrease by 2 cmH2O

– “switch over” to Pressure Support at virtual PSV Target. • Once Tidal volume has been delivered

– “looks” at the inspiratory flow • Inspiratory Flow > set peak flow

– flow cycled at 25% of peak inspiratory flow • Inspiratory Flow = set flow

– ends as Volume Control breath

Page 45: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Intra-Breath Demand Flow in VCVNote augmented flow to meet patients inspiratory demand. Demand ends before set volume has been delivered.

Demand continues through end of breath. Breath is FLOW CYCLED at 25% of PIFR.

Page 46: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Intra-Breath Demand Flow ON / OFF

Double cycling in periods of high demand

Page 47: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

PSV & Flow Trigger in VCV

These Advanced Settings are the same for all modes and types of ventilation

Page 48: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings

Pressure Regulated Volume Control - PRVC

Page 49: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

PRVC – Pressure Regulated Volume Control

Volume Inspiratory Time

Flow Trigger

Page 50: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Volume Parameter in PRVC

Similar settings to PCV since PRVC is a hybrid of PCV

Page 51: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

PRVC / Vsync

• Logic controlled Volume TARGETED breath– Decelerating flow, VC test breath with a 40 msec pause– End inspiratory pressure is Target Pressure for the first pressure

control breath– Subsequent breaths are delivered as pressure control breaths

based on the dynamic compliance of the previous breath. • Inspiratory pressure is adjusted automatically by the ventilator to

maintain the target volume. • Maximum step change between two consecutive breaths is 3 cmH2O. • Maximum tidal volume delivered in a single breath is determined by

the Volume Limit setting.

– Vsync NOTE:• The Peak Flow control sets the flow rate, which is used by the

ventilator for the test breath only. The ventilator uses the Peak Flow setting and Inspiratory Pause to determine the maximum inspiratory time during Vsync ventilation.

– Flow Cycling can only occur if target tidal volume has been reached

Page 52: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

PRVC / Vsync• This test breath sequence is initiated in the event any of the following

occur:– Entering the Mode (PRVC or Vsync)– Changing the set tidal volume while in PRVC or Vsync– Reaching the Volume Limit setting– Delivered tidal volume > 1.5 times the set volume– Flow termination of the test breath

• Intrabreath demand system is active for all test breaths

– Exiting Standby– Activation of any of the following alarms

• High Peak Pressure Alarm• Low Peak Alarm• Low PEEP Alarm• Patient Circuit Disconnect Alarm• I-Time Limit• I:E Limit

– Flow Cycling can only occur if target tidal volume has been reached

Page 53: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings

Airway Pressure Release Ventilation - APRV

Page 54: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

APRV• Patient breathes spontaneously at a

CPAP level

• Periodically “released” to a lower CPAP level to flush CO2 from the FRC

P Question:

What two features would make this type of ventilation better?

Page 55: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

APRV – Airway Pressure Release Ventilation

Answer: Synchronize the changes and add PSV

That’s what these Advanced Settings do.

Page 56: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced SettingsTime High in APRV

Page 57: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

APRV

• Allows spontaneous breathing over a high mean airway pressure.

• High mean airway pressure may improve lung recruitment.

• Pressure support available .

• Apnea back-up parameters available.

Page 58: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

APRVAdvanced settings

• Time high sync period-allows the operator to allow a period in which the ventilator will synchronize (during expiration) from high to low pressure phase.– Adjustable from 0-50% of Time high.

• Time high PSV-on or off-allows the user to have pressure support on or off during the time high phase.

• Time low sync- allows the operator to allow a period in which the ventilator will synchronize (during inspiration) from 0-50%– Adjustable from 0-50% of Time high

Page 59: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

APRVProposed Benefits

• 1. Lower Paw for a given tidal volume compared• with volume-targeted modes, e.g., AC, SIMV• 2. Lower minute ventilation, i.e., less dead space• ventilation• 3. Limited adverse effects on cardio-circulatory• function• 4. Spontaneous breathing possible throughout• entire ventilatory cycle• 5. Decreased sedation use• 6. Near elimination of neuromuscular blockade

Page 60: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Pressure

Time High

Time Low

Spontaneous breaths

APRV

Page 61: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings Time High in APRV

• “Time High Sync %”– Synchronizes transition from Pressure High to

Pressure Low – Occurs with the first end of inspiration detected after

the Time High Sync % window opens.– The Time High Sync %window is adjustable from 0 to

50%, in 5% increments of set Time High.– The default setting for Time High Sync % is 0 (off)

Page 62: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced Settings Time High in APRV

• “Time High PSV”– Time High PSV is either turned on (position 1) or off (position 0)– The PSV is delivered above the current phase baseline

pressure. – If Time High PSV is activated as Time High

Advanced Setting, the ventilator will deliver the same PSV level for both Pressure High and Pressure Low.

• Example: If Plow= 5cmH2O and Phigh= 25cmH2O, then a PSV breath set at 5 cmH2O will result in a Plow increase to 10cmH2O and Phigh increase to 30cmH2O

Page 63: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced SettingsTime Low in APRV

Page 64: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Advanced SettingsTime Low in APRV

• “Time Low Sync %”– The ventilator synchronizes the change from

Pressure Low to Pressure High with the detection of inspiratory flow or the first inspiratory effort detected within the Time Low Sync % window.

– The Time Low Sync % window is adjustable from 0 to 50%, in 5% increments of set Time Low.

– The default setting for Time Low is 0 (off)

Page 65: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Synchronization Window

• Actual Time High or Time Low can change if Sync is activated

• Depends on when the patient triggers to go from High to Low or Low to High

Gives a range of possible I-times if Sync is activated

Page 66: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

PSV & Flow Trigger in APRV

Both PSV & Flow Trigger Advanced Settings in VCV function as previously discussed on earlier slides

Page 67: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Automatic Airway Compensation (AAC)

• AAC when applied, will supplement spontaneous pressure support or or pressure control flow cycled breaths based on calculated endotracheal tube resistance.

• Has no role in other pressure control or volume breath types.

• To activate, select “AAC On” and dial in tube size and length. An indicator saying “AAC” will appear in the mode bar.

Page 68: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

How Does Tube Compensation Work?

• Positive pressure added based on the inspired flow and tube curvature, diameter, and length

• Patient doesn’t have to experience the resistive work due to the artificial airway

Page 69: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

AAC

• Calculation based on:– ETT Size

• 2.0 to 10.0• Length

– flow – pharyngeal curvature – FiO2 / gas density– transitions from turbulent to laminar flow  

• Only ventilator with AAC available for controlled breaths

Page 70: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

AAC 

(length) x (viscosity) (radius4) 

This equation is used during periods of laminar flow where resistance is primarily determined by gas viscosity.

 

  (density 0.75) x (viscosity .025) x (flow 1.75) x (length) (radius 4.75)

This equation is used during periods of turbulent flow where resistance is primarily determined by gas density.

Page 71: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Circuit Pressure

What The Ventilator Circuit Sees

• Flow triggering can be effective in minimizing imposed work

Paw

Page 72: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

What The Carina Sees

Circuit pressure

Lower pressure at the carina

Paw

Page 73: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

• AAC adds appropriate pressure to keep carina pressure at preset inspiratory pressure

What the Carina Sees

Higher circuit pressure

No decrease in pressure at carina

Paw

Page 74: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Considerations when using AAC

• Monitor pressure support level to achieve desired volumes. Supported breaths may be augmented above pressure support, so less pressure support may be needed.

• Compensatory support varies with many factors, and will vary with patient effort and other factors.

• Monitor volumes and pressures carefully, with special considerations with neonates.

Page 75: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

Set-up for Patient Use (cont)

Automatic Airway Compensation-suggested

reading: Wrigge H, Zinerling J, Hering R,Swafenberg N, Stuber F, et al,

Cardiopulmonary effects of automatic tubing compensation during airway pressure release ventilation in patients with acute lung injury Anesthesia 2001 Aug;95(2)382-9

Guttmann J, Haberthur C, Mols G.Related Articles

Automatic tube compensation.Respir Care Clin N Am. 2001

Sep;7(3):475-501

Page 76: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

AutoPEEP Airway

Page 77: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

AutoPEEP Airway

AutoPEEPaw is the airway pressure at the end of exhalation immediately prior to the beginning of the next mandatory inspiration.

Page 78: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

AutoPEEP Airway Timing

The ventilator will establish the AutoPEEP measurement when the system pressure reaches equilibration, at the next mandatory breath interval or 6 seconds whichever is shorter.

Page 79: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

AutoPEEP Airway Gas Trapping

Normal: 0 cm H2O above the applied PEEPNote:Requires a passive patient and a cuffed endotracheal tube.Occurs when either insufficient expiratory time is present or in dynamic flow limitation conditions such as asthma or severe COPD. This results in gas trapping in the lungs.

Page 80: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

MIP / P100 Maneuver

Page 81: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

MIP Maneuver

The MIP (Maximum Inspiratory Pressure) / P100 maneuver measures the negative deflection in the pressure tracing during the patient’s active effort to demand a breath.

Page 82: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

MIP Maneuver

Maximum Inspiratory Pressure (MIP), is the maximum negative airway pressure that is achieved by the patient, during an expiratory hold maneuver.

Page 83: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

MIP Maneuver

Normal: Adults < -70 to -100 cm H2O Pediatrics < -20 to -100 cm H2O Readiness for extubation < -20 cm H2O

Note: Patient effort variable.Indicator of inspiratory muscle strength.Useful in weaning and progression of neuromuscular disease.Will be decreased in conditions such as kyphoscoliosis, advanced age, COPD and neuromuscular disease. .

Page 84: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

P100 Maneuver

Respiratory Drive (P100), is the negative pressure that occurs 100 ms after an inspiratory effort has been detected.

Page 85: AVEA Enhanced Pulmonary Mechanics. Advanced Settings User Interface Module Advanced Settings Feature

P100 Maneuver

Normal: -1 to -4 cm H2O Adults -0.5 to -4 cm H2O Pediatrics

Note:The pressure that occurs 100 milliseconds (ms) after the onset of inspiration while the inspiratory valve remains closed. Because it normally takes at least 300 ms for the patient to become aware of the occluded airway this is a good test of the respiratory center output. There is no change in lung volume or airflow during this initial 300 ms, therefore abnormalities in lung mechanics have no effect on the measurement.

Values in excess of -5 cm H2O may indicate a high respiratory drive that may result in increased work of breathing and subsequent fatigue.