avalight-hal-cal-mini and avalight-dh-cal calibrated light...

2
Calibrating your spectrometer has never been easier: the AvaLight-HAL- CAL-Mini and Avalight-DH-CAL are NIST traceable calibrated light sources to measure absolute spectral intensity. The AvaLight-HAL-CAL-Mini is a com- pact, affordable light source. It is cali- brated for the visible range (350-1095 nm). Optionally, an extended calibra- tion for the near-infrared spectral range (1100-2500 nm) can be ordered. It has a built-in diffuser, a cosine cor- rector with SMA adapter and comes with a calibration file in ASCII format. Calibration can be done using the AvaSoft software. The AvaLight-HAL-CAL-ISPxx-Mini is a special version of the Avalight-HAL- CAL-Mini, which enables coupling any of Avantes’ AvaSphere-xx-IRRAD integrating spheres to the lightsource (xx=30, 50 or 80) for calibration. This source is supplied with a special bot- tom plate to stabilize the AvaSphere. The Avalight-HAL-CAL-Mini and Avalight-HAL-CAL-ISPxx-Mini include a power supply. AvaLight-HAL-CAL-Mini and AvaLight-DH-CAL Calibrated Light Sources AvaLight-HAL-CAL-Mini AvaSphere-50-LS-HAL-CAL AvaLight-DH-CAL For radiance calibration in the visible spectral range (360-1100 nm), Avantes offers the AvaSphere-50-LS-HAL-CAL. This radiance calibration sphere is nor- mally used to calibrate a spectrometer and a bare fiber with an SMA connec- tor. This uniform source ensures that the fiber is homogeneously illuminated during calibration. The AvaSphere-LS-HAL-CAL is supplied with a power supply and a CD-ROM with the calibration file in ASCII- format. * Radiance calibration * Visible range * For homogeneous illuminated areas For calibrations in the ultraviolet and visible range (200-1099 nm), the AvaLight-DH-CAL is the best solu- tion. This source can be used with all AvaSpec spectrometers to calibrate for absolute spectral intensity. It is supplied with a built-in diffuser, a cosine corrector (CC-VIS/NIR) with SMA adapter and a CD-ROM containing the calibration files in ASCII format. Two calibration files are included: one for irradiance calibration over the full range (200-1099 nm) and one over the visible and near infrared range (350-1099 nm). For the first file, both the deuterium and the halogen bulb should be used during calibration. The second file is to be used with only the halogen light. The halogen only spec- trum provides a smoother black body calibration spectrum for the longer wavelengths. For a more balanced spectrum accross the 200-1100 nm range, Avantes offers the AvaLight-DH-BAL-CAL. For ultra- violet range calibration only (200-400 nm), the AvaLight-D-CAL is the right choice. The calibration files can be imported in the AvaSoft-IRRAD application soft- ware, for intensity calibration which turns your spectrometer into a spectro- radiometer. The AvaLight-DH-CAL-ISPxx is a spe- cial version of these calibrated light sources, meant to facilitate coupling of the AvaSphere-xx-IRRAD integrat- ing spheres to the light source. (xx is 30, 50 or 80). The AvaLight-DH-CAL, AvaLight-DH-BAL-CAL and AvaLight-D- BAL-CAL all include a power supply. * Flexible calibration * Ultraviolet and visible range * Built-in diffuser and cosine corrector * Field calibration * Visible and optional NIR range * Built-in diffuser * Versatile

Upload: others

Post on 26-Sep-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: AvaLight-HAL-CAL-Mini and AvaLight-DH-CAL Calibrated Light ...telab.vuse.vanderbilt.edu/docs/specs/Datasheet-cal-MINIv4.pdfturns your spectrometer into a spectro-radiometer. The AvaLight-DH-CAL-ISPxx

Calibrating your spectrometer has never been easier: the AvaLight-HAL-CAL-Mini and Avalight-DH-CAL are NIST traceable calibrated light sources to measure absolute spectral intensity.

The AvaLight-HAL-CAL-Mini is a com-pact, affordable light source. It is cali-brated for the visible range (350-1095 nm). Optionally, an extended calibra-tion for the near-infrared spectral range (1100-2500 nm) can be ordered. It has a built-in diffuser, a cosine cor-rector with SMA adapter and comes with a calibration file in ASCII format.

Calibration can be done using the AvaSoft software.

The AvaLight-HAL-CAL-ISPxx-Mini is a special version of the Avalight-HAL-CAL-Mini, which enables coupling any of Avantes’ AvaSphere-xx-IRRAD integrating spheres to the lightsource (xx=30, 50 or 80) for calibration. This source is supplied with a special bot-tom plate to stabilize the AvaSphere. The Avalight-HAL-CAL-Mini and Avalight-HAL-CAL-ISPxx-Mini include a power supply.

AvaLight-HAL-CAL-Mini and AvaLight-DH-CAL Calibrated Light SourcesAvaLight-HAL-CAL-Mini

AvaSphere-50-LS-HAL-CAL

AvaLight-DH-CAL

For radiance calibration in the visible spectral range (360-1100 nm), Avantes offers the AvaSphere-50-LS-HAL-CAL. This radiance calibration sphere is nor-mally used to calibrate a spectrometer and a bare fiber with an SMA connec-tor. This uniform source ensures that the fiber is homogeneously illuminated during calibration.

The AvaSphere-LS-HAL-CAL is supplied with a power supply and a CD-ROM with the calibration file in ASCII-format.

* Radiance calibration* Visible range* For homogeneous illuminated areas

For calibrations in the ultraviolet and visible range (200-1099 nm), the AvaLight-DH-CAL is the best solu-tion. This source can be used with all AvaSpec spectrometers to calibrate for absolute spectral intensity. It is supplied with a built-in diffuser, a cosine corrector (CC-VIS/NIR) with SMA adapter and a CD-ROM containing the calibration files in ASCII format.

Two calibration files are included: one for irradiance calibration over the full range (200-1099 nm) and one over the visible and near infrared range (350-1099 nm). For the first file, both the deuterium and the halogen bulb should be used during calibration. The second file is to be used with only the halogen light. The halogen only spec-trum provides a smoother black body

calibration spectrum for the longer wavelengths.

For a more balanced spectrum accross the 200-1100 nm range, Avantes offers the AvaLight-DH-BAL-CAL. For ultra-violet range calibration only (200-400 nm), the AvaLight-D-CAL is the right choice.

The calibration files can be imported in the AvaSoft-IRRAD application soft-ware, for intensity calibration which turns your spectrometer into a spectro-radiometer.

The AvaLight-DH-CAL-ISPxx is a spe-cial version of these calibrated light sources, meant to facilitate coupling of the AvaSphere-xx-IRRAD integrat-ing spheres to the light source. (xx is

30, 50 or 80). The AvaLight-DH-CAL, AvaLight-DH-BAL-CAL and AvaLight-D-BAL-CAL all include a power supply.

* Flexible calibration* Ultraviolet and visible range* Built-in diffuser and cosine corrector

* Field calibration* Visible and optional NIR range * Built-in diffuser* Versatile

Page 2: AvaLight-HAL-CAL-Mini and AvaLight-DH-CAL Calibrated Light ...telab.vuse.vanderbilt.edu/docs/specs/Datasheet-cal-MINIv4.pdfturns your spectrometer into a spectro-radiometer. The AvaLight-DH-CAL-ISPxx

Ligh

t So

urces

Technical DataAvaLight-HAL-CAL-Mini AvaSphere-50-LS-HAL-CAL AvaLight-DH-(BAL)-CAL

Calibration use Irradiance µW cm-2 nm-1 Radiance µW sr-1 cm-2 nm-1 Irradiance µW cm-2 nm-1

Calibrated surface CC-VIS/NIR or AvaSphere Bare fiber CC-UV/VIS or AvaSphere

Wavelength Range350-1095 nm / 1100-2500 nm*

350-1095 nm / 1100-2500 nm*

200-1099 nm

Calibration Repeatability ± 0.5 % ± 0.5 % ± 1.0%

Calibration Relative Uncertainty to NIST standard

±9.5% (350-1100 nm)±6.5% (1100-1950 nm)±9.5% (1950-2500 nm)

±9.5% (350-1100 nm)±10% (200-240 nm)±9% (240-350 nm)±10% (350-400 nm)±9.5% (400-1100 nm)

Calibration valid for 60 hrs 60 hrs 60 hrs

Warm-up Time Ca. 15 min. Ca. 15 min. Ca. 30 min.

Bulb Output 170µW cm-2 nm-1 (@800 nm )

100 µW sr-1 cm-2 nm-1

(@800 nm)80 µW cm-2 nm-1 (@215 nm)5 µW cm-2 nm-1 (@800 nm)

Power requirement 12 V / 2.08A 24 VDC / 0.3 A 100-240 VAC

Dimensions 150 x 78 x 37 mm 70 round x 75 mm height 315 x 165 x 140 mm

* optional extended range NIR calibration

Ordering InformationAvaLight-HAL-CAL-Mini • NIST traceable Halogen Lamp with CC- VIS/NIR diffuser, incl. PS-12V/2.08A

AvaLight-HAL-CAL-ISP30-Mini• NIST traceable Halogen Lamp for use with AvaSphere-30-IRRAD, incl. PS-12V/2.08A and

special sphere holder bottom plate

AvaLight-HAL-CAL-ISP50-Mini • As AvaLight-HAL-CAL-ISP30 for use with AvaSphere-50-IRRAD

AvaLight-HAL-CAL-ISP80-Mini • As AvaLight-HAL-CAL-ISP30 for use with AvaSphere-80-IRRAD

AvaSphere-50-LS-HAL-CAL• NIST traceable radiance calibration source, 10 mm and SMA adapter calibrated, incl

PS-24V/1.25A

HL-Recal • AvaLight-HAL-CAL recalibration service 350-1095 nm

HL-Recal-NIR • AvaLight-HAL-CAL extended or recalibration service 1100-2500 nm

AvaLight-D-CAL • NIST traceable UV Deuterium Lamp with CC-UV/VIS diffuser, -SR fibers recommended

AvaLight-DH-CAL• NIST traceable UV/VIS Deuterium/Halogen Lamp with CC-UV/VIS diffuser, -SR fibers

recommended

AvaLight-DH-BAL-CAL • As AvaLight-DH-CAL, but balanced UV/VIS Deuterium/Halogen

AvaLight-DH-CAL-ISP30• NIST traceable UV/VIS Deuterium/Halogen Lamp for use with AvaSphere-30-IRRAD, incl.

special sphere holder bottom plate, -SR fibers recommended

AvaLight-DH-CAL-ISP50 • As AvaLight-DH-CAL-ISP30 for use with AvaSphere-50-IRRAD

AvaLight-DH-CAL-ISP80 • As AvaLight-DH-CAL-ISP30 for use with AvaSphere-80-IRRAD

DH-Recal • AvaLight-DH-CAL recalibration service 200-1099 nm

AvaSoft-IRRAD • Irradiance add-on software, to be ordered with AvaSoft-Full

Did you know Avantes is specialized

in custom made fiber-optic cables?