au 5200 alat - kovalent€¦ · test parameters item code gpt function select no. 1. operation yes...
TRANSCRIPT
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code GPT
Function select No.
1. Operation Yes
2. Sample volume 15
3. Sample volume at repeat run 4
4. Reagent volume R1 200
Reagent volume R2 50
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method Rate
7 Wavelength 1 340
Wavelength 2 380
8 Reaction slop -
9 Measuring point start 9
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % 15
12 Normal value H 37
Normal value L 0
13 Dynamic range H 800
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type MB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor **3865
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
**) Factor must be checked by calibration
serum
ALAT(GPT) FS (IFCC Mod.)
Order information
Cat. No. Kit size
10 270 021 R1 5 x 20 ml + R2 1 x 25 ml
10 270 022 R1 5 x 80 ml + R2 1 x 100 ml
10 270 023 R1 1 x 800 ml + R2 1 x 200 ml
For determination with pyridoxal-5-phosphate
activation additionally required:
10 501 030 6 x 3 ml
Notes
1. Please refer to the package insert for
ALAT(GPT) FS (IFCC Mod.) for the detailed
information about the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by DiaSys Diagnostic Systems GmbH
& Co.KG Alte Strasse 9, 65558 Holzheim,
Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code ALP
Function select No.
1. Operation Yes
2. Sample volume 4
3. Sample volume at repeat run 2
4. Reagent volume R1 200
Reagent volume R2 50
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method Rate
7 Wavelength 1 410
Wavelength 2 660+
8 Reaction slop +
9 Measuring point start 10
Measuring point end 15
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % 15
12 Normal value H 141
Normal value L 28
13 Dynamic range H 850
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor *
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
ALKALINE PHOSPHATASE FS IFCC
Order information
Cat. No. Kit size
10 041 021 R1 5x 20 ml + R2 1x 25 ml
10 041 022 R1 5x 80 ml + R2 1x 100 ml
10 041 023 R1 1x 800 ml + R2 1x 200 ml
Notes
1. Please refer to the package insert for
Alkaline phosphatase FS IFCC for the
detailed information about the test on the
following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board of the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG
Alte Strasse 9, 65558 Holzheim, Germany.
OLYMPUS AU 5200
Chemistry setting Temperature : 37°C
Test parameters
Item code AMY
Function select No.
1. Operation Yes 2. Sample volume 3
3. Sample volume at repeat run 2 4. Reagent volume R1 200 Reagent volume R2 50 5. W 3 parameter
Prime (bottle to valve) R1 Yes Prime (bottle to valve) R2 Yes Dispense level (valve to cuvette) R 1 100 Dispense level (valve to cuvette) R 2 100
6 Method Rate 7 Wavelength 1 410 Wavelength 2 660 8 Reaction slop +
9 Measuring point start 10 Measuring point end 15 10 O.D. value range max 2.000 O.D. value range min 0.000
11 Limit of Linearity % 15 12 Normal value H 100 Normal value L 0 13 Dynamic range H 2000
Dynamic range L 0 14 Panic value H # Panic value L # 15 Reagent OD range
First Point H 2.000 First Point L -2.000 Last Point H 2.000 Last Point L -2.000
16 Decision level for repeat run H 9999 Decision level for repeat run L 0.0 17 R. dispence by black rack 0
18 R. dispence by black interval 0
Calibration parameters
Calibration type AB ODO – Conc. 0.000 Cal. No. {x}> #
Conc. > # Factor range H Factor range L Factor *
Correlation correction
Correlation coefficient A Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
-AMYLASE CC FS
Order information Cat. No. Kit size
10 050 021 R1 5 x 20 ml + R2 1 x 25 ml
10 050 022 R1 5 x 80 ml + R2 1 x 100 ml
10 050 023 R1 1 x 800 ml + R2 1 x 200 ml
10 050 717 R1 5x 80 ml + R2 5 x 20 ml
Notes
1. Please refer to the package insert for -Amylase CC
FS for the detailed information about the test on the following:
Clinical Relevance Method and Principle Composition and Stability of the Reagents
Specimens Calibrators and Controls Performance Characteristics concerning; Measuring Range
Specificity/Interferences Sensitivity/Limit of Detection
Precision (Reproducibility, Repeatability) Method Comparison
Reference Ranges Literature 2. The stability of the reagent on board of the analyser is
at least one month provided that contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG Alte Strasse 9, 65558 Holzheim, Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code GOT
Function select No.
1. Operation Yes
2. Sample volume 15
3. Sample volume at repeat run 4
4. Reagent volume R1 200
Reagent volume R2 50
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method Rate
7 Wavelength 1 340
Wavelength 2 380
8 Reaction slop -
9 Measuring point start 9
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % 15
12 Normal value H 41
Normal value L 0
13 Dynamic range H 800
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type MB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor **3515
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
**) Factor must be checked by calibration
serum
ASAT(GOT) FS (IFCC Mod)
Order information
Cat. No. Kit size
10 260 021 R1 5 x 20 ml + R2 1 x 25 ml
10 260 022 R1 5 x 80 ml + R2 1 x 100 ml
10 260 023 R1 1 x 800 ml + R2 1 x 200 ml
For determination with pyridoxal-5-phosphate
activation additionally required:
10 501 030 6 x 3 ml
Notes
1. Please refer to the package insert for
ASAT(GOT) FS (IFCC Mod) for the detailed
information about the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by DiaSys Diagnostic Systems GmbH
& Co.KG Alte Strasse 9, 65558 Holzheim,
Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Bili D
Function select No. #
1. Operation Yes
2. Sample volume 20
3. Sample volume at repeat run 2
4. Reagent volume R1 280
Reagent volume R2 70
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method End
7 Wavelength 1 540
Wavelength 2 660
8 Reaction slop +
9 Measuring point start 2
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % -
12 Normal value H 0.3
Normal value L 0
13 Dynamic range H 10.0
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor *
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
BILIRUBIN AUTO DIRECT FS
Order information
Cat. No. Kit size
10 082 021 R1 5 x 20 ml + R2 1 x 25 ml
10 082 022 R1 5 x 80 ml + R2 1 x 100 ml
10 082 023 R1 1 x 800 ml + R2 1 x 200 ml
Notes
1. Please refer to the package insert for
Bilirubin Auto Direct FS for the detailed
information about the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board of the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG
Alte Strasse 9, 65558 Holzheim, Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Bili T
Function select No. #
1. Operation Yes
2. Sample volume 6
3. Sample volume at repeat run 2
4. Reagent volume R1 280
Reagent volume R2 70
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method End
7 Wavelength 1 540
Wavelength 2 660
8 Reaction slop +
9 Measuring point start 2
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % -
12 Normal value H 1.1
Normal value L 0
13 Dynamic range H 30.0
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor *
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
BILIRUBIN AUTO TOTAL FS
Order information
Cat. No. Kit size
10 081 021 R1 5 x 20 ml + R2 1 x 25 ml
10 081 022 R1 5 x 80 ml + R2 1 x 100 ml
10 081 023 R1 1 x 800 ml + R2 1 x 200 ml
Notes
1. Please refer to the package insert for
Bilirubin Auto Total FS for the detailed
information about the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board of the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG
Alte Strasse 9, 65558 Holzheim, Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Calc
Function select No. #
1. Operation Yes
2. Sample volume 3
3. Sample volume at repeat run 2
4. Reagent volume R1 250
Reagent volume R2
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
6 Method end
7 Wavelength 1 660
Wavelength 2 800
8 Reaction slop +
9 Measuring point start 0
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % -
12 Normal value H 10.3
Normal value L 8.6
13 Dynamic range H 25.0
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H -
Factor range L -
Factor *
Correlation correction
Correlation coefficient A -
Correlation coefficient B -
#) Data entry by the user
*) Calculated by the analyzer
CALCIUM AS FS Order information Cat. No. Kit size 1 1130 99 10 021 R 5 x 25 ml + 1 x 3 ml Std
1 1130 99 10 026 R 6 x 100 ml
1 1130 99 10 023 R 1 x 1000 ml
1 1130 99 10 030 6 x 3 ml Standard
1 1130 99 10 717 R 6 x 100 ml
Notes
1. Please refer to the package insert for Calcium AS FS for
detailed information about the test on the following:
Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding
- Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison
Reference Ranges Literature
2. The stability of the reagent on board the analyser is at
least one month provided that contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG Alte Strasse 9, 65558 Holzheim, Germany
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Chol
Function select No. #
1. Operation Yes
2. Sample volume 3
3. Sample volume at repeat run 2
4. Reagent volume R1 250
Reagent volume R2
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
6 Method end
7 Wavelength 1 520
Wavelength 2 660
8 Reaction slop +
9 Measuring point start 0
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % -
12 Normal value H 220
Normal value L 120
13 Dynamic range H 750
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H -
Factor range L -
Factor *
Correlation correction
Correlation coefficient A -
Correlation coefficient B -
#) Data entry by the user
*) Calculated by the analyzer
CHOLESTEROL FS
Order information
Cat. No. Kit size
10 130 021 R 5 x 25 ml + 1 x 3 ml Std
1 1300 99 10 026 R 6 x 100 ml
10 130 023 R 1 x 1000 ml
10 130 030 6 x 3 ml Standard
Notes
1. Please refer to the package insert for
Cholesterol FS for the detailed information
about the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board the
analyser is at least one month provided
that contamination and evaporation are
avoided.
3. Manufactured by DiaSys Diagnostic Systems
GmbH & Co.KG Alte Strasse 9, 65558 Holzheim,
Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code CKMB
Function select No.
1. Operation Yes
2. Sample volume 10
3. Sample volume at repeat run 5
4. Reagent volume R1 200
Reagent volume R2 50
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method Rate
7 Wavelength 1 340
Wavelength 2 380
8 Reaction slop +
9 Measuring point start 10
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % 15
12 Normal value H 24
Normal value L 0
13 Dynamic range H 1000
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type MCAL type
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor **8250
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
**) Factor must be checked by calibration
serum
CK-MB FS
Order information
Cat. No. Kit size
10 165 021 R1 5 x 20 ml + R2 1 x 25 ml
10 165 022 R1 5 x 80 ml + R2 1 x 100 ml
10 165 023 R1 1 x 800 ml + R2 1 x 200 ml
Notes
1. Please refer to the package insert for CK-MB
FS for the detailed information about the
test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by DiaSys Diagnostic Systems GmbH
& Co.KG Alte Strasse 9, 65558 Holzheim,
Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code CKNAC
Function select No. #
1. Operation Yes
2. Sample volume 7
3. Sample volume at repeat run 2
4. Reagent volume R1 200
Reagent volume R2 50
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method Rate
7 Wavelength 1 340
Wavelength 2 380
8 Reaction slop +
9 Measuring point start 11
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % 15
12 Normal value H 190
Normal value L 0
13 Dynamic range H 1000
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type MB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor **
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
**) Factor must be checked by calibration
serum
CK-NAC FS
Order information
Cat. No. Kit size
10 160 021 R1 5 x 20 ml + R2 1 x 25 ml
10 160 022 R1 5 x 80 ml + R2 1 x 100 ml
10 160 023 R1 1 x 800 ml + R2 1 x 200 ml
Notes
1. Please refer to the package insert for CK-
NAC FS for the detailed information about
the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board of the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG
Alte Strasse 9, 65558 Holzheim, Germany.
OLYMPUS AU 5200
Chemistry setting Temperature : 37°C
Test parameters
Item code Creat
Function select No. #
1. Operation Yes 2. Sample volume 12
3. Sample volume at repeat run 2 4. Reagent volume R1 240 Reagent volume R2 60 5. W 3 parameter
Prime (bottle to valve) R1 Yes Prime (bottle to valve) R2 Yes Dispense level (valve to cuvette) R 1 100 Dispense level (valve to cuvette) R 2 100
6 Method Rate 7 Wavelength 1 520 Wavelength 2 660 8 Reaction slop +
9 Measuring point start 9 Measuring point end 13 10 O.D. value range max 2.000 O.D. value range min 0.000
11 Limit of Linearity % 25 12 Normal value H 0.6 Normal value L 1.3 13 Dynamic range H 15.0
Dynamic range L 0 14 Panic value H # Panic value L # 15 Reagent OD range
First Point H 2.000 First Point L -2.000 Last Point H 2.000 Last Point L -2.000
16 Decision level for repeat run H 9999 Decision level for repeat run L 0.0 17 R. dispence by black rack 0
18 R. dispence by black interval 0
Calibration parameters
Calibration type AB ODO – Conc. 0.000 Cal. No. {x}> #
Conc. > # Factor range H Factor range L Factor *
Correlation correction
Correlation coefficient A Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
CREATININE FS
Order information Cat. No. Kit size
10 171 021 R1 4 x 20 ml + R2 1 x 20 ml + 1 x 3ml Std
1 1711 99 10 026 R1 5 x 80 ml + R2 1 x 100 ml
10 171 023 R1 1 x 800 ml + R2 1 x 200 ml
10 170 030 6 x 3 ml Standard
Notes
1. Please refer to the package insert for Creatinine FS for
the detailed information about the test on the following:
Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens
Calibrators and Controls Performance Characteristics concerning; Measuring Range Specificity/Interferences
Sensitivity/Limit of Detection Precision (Reproducibility, Repeatability)
Method Comparison Reference Ranges
Literature 2. The stability of the reagent on board of the analyser is
at least one month provided that contamination and
evaporation are avoided. 3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG
Alte Strasse 9, 65558 Holzheim, Germany.
CRP FS*
Order information Cat. No. Kit size 1 7002 99 10 015 R1 2 x 25 ml + R2 1 x 10 ml 10 700 021 R1 5 x 25 ml + R2 1 x 25 ml 10 700 023 R1 1 x 1000 ml + R2 1 x 200 ml 1 7000 99 10 039 1 x 2 ml Calibrator Level 1 1 x 2 ml Calibrator Level 2 1 x 2 ml Calibrator Level 3 1 x 2 ml Calibrator Level 4 1 x 2 ml Calibrator Level 5
Method Immunological turbidimetric test
Endpoint determination of the concentration of CRP through photometric measurement of antigen-antibody-reaction. Calculation: 1-point or multi-point calibration
Reagent preparation and stability The reagents and calibrators are ready-to-use and stable up to the end of the indicated month of expiry, if contamination is avoided and stored at 2 – 8 °C.
Specimen Serum, heparinized or EDTA plasma. Anticoagulants in usual concentration do not interfere. Avoid hemolysis! Stability: 8 days at 2 - 8 °C 3 months at - 20 °C (not in case of repeated deep freezing)
Prozone limit up to at least 200 mg/dl
Components and concentration in the test R1: Buffer TRIS pH 7.5 100 mmol/l PEG, detergents and stabilizers R2: Antiserum TRIS pH 8.0 100 mmol/l Anti-human CRP antibodies (goat) with stabilizers Calibrators:
stabilized human sera, concentrations are indicated on the label
Notes 1. The reagents contain Sodium Azide (0.095 %) as preservative.
Do not swallow! Avoid contact with skin and mucous membranes.
2. The calibrators were only produced with human sera from donors, where no HIV-antibodies and no HbsAg were detected by FDA-approved tests. As a risk of infection cannot be completely excluded, the calibrators should be handled as carefully as patient samples.
Normal range (see reference 3,4)
Adults up to 0.8 mg/dl Newborn up to 3 days up to 1.5 mg/dl Infants from the 4
th day and children up to 1.0 mg/dl
References 1. Skurk, A., Thomas, L., Lab. Med., 14, (1990), 111 – 115 2. Grützmeyer, S., von Schenck, H., Clin. Chem., 35,
(1989), 461 - 463 3. Claus, D. R., Osmand, A. P., Gewurz, H., J. Lab. Clin. Med., 87,
(1976), 120 – 128 4. Sabel, K.-G., Wadsworth, Ch., Acta Paediatr. Scand., 68,
(1979), 825 – 831
* fluid stable
OLYMPUS AU 5200
Chemistry setting Temperature : 37°C
Test parameters
Item code
Function select No.
1. Operation yes 2. Sample volume 15 3. Sample volume at repeat run 5 4. Reagent volume R1 250 Reagent volume R2 50 5. W 3 parameter Prime (bottle to valve) R1 yes Prime (bottle to valve) R2 yes Dispense level (valve to cuvette) R 1 100 Dispense level (valve to cuvette) R 2 100 6. Method end 7. Wavelength 1 340 Wavelength 2 660 8. Reaction slop + 9. Measuring point start 6 Measuring point end 16 10. O.D. value range max 2.000 O.D. value range min 0.000 11. Limit of Linearity % - 12. Normal value H 0.8 Normal value L 0.0 13. Dynamic range H 25.0 Dynamic range L 0.0 14. Panic value H # Panic value L # 15. Reagent OD range First Point H 2.000 First Point L -2.000 Last Point H 2.000 Last Point L -2.000 16. Decision level for repeat run H 9999 Decision level for repeat run L 0.0 17. R. dispence by black rack 0 18. R. dispence by black interval 0
Calibration parameters
Calibration type AB ODO – Conc. 0.000 Cal. No. {x}> # Conc. > * Factor range H - Factor range L - Factor **
Correlation correction
Correlation coefficient A Correlation coefficient B
#) Data entry by the user *) Calculated by the analyzer
11.02.11
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Gluc
Function select No. #
1. Operation Yes
2. Sample volume 3
3. Sample volume at repeat run 2
4. Reagent volume R1 300
Reagent volume R2
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
6 Method End
7 Wavelength 1 520
Wavelength 2 660
8 Reaction slop +
9 Measuring point start 0
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % -
12 Normal value H 115
Normal value L 75
13 Dynamic range H 400
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor *
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
GLUCOSE GOD FS
Order information
Cat. No. Kit size
10 250 021 R 5 x 25 ml + 1 x 3 ml Std
1 2500 99 10 026 R 6 x 100 ml
10 250 023 R 1 x 1000 ml
10 250 030 6 x 3 ml Standard
Notes
1. Please refer to the package insert for
Glucose GOD FS for the detailed information
about the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board of the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG
Alte Strasse 9, 65558 Holzheim, Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Gluc
Function select No. #
1. Operation Yes
2. Sample volume 3
3. Sample volume at repeat run 2
4. Reagent volume R1 280
Reagent volume R2 70
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method End
7 Wavelength 1 340
Wavelength 2 380
8 Reaction slop +
9 Measuring point start 5
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % 15
12 Normal value H 115
Normal value L 75
13 Dynamic range H 500
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor *
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
GLUCOSE HEXOKINASE FS
Order information
Cat. No. Kit size
10 251 021 R1 4 x 20 ml + R2 1 x 20 ml
+ 1 x 3ml Std
1 2511 99 10 026 R1 5 x 80 ml + R2 1 x 100 ml
10 251 023 R1 1 x 800 ml + R2 1 x 200 ml
10 250 030 6 x 3 ml Standard
Notes
1. Please refer to the package insert for
Glucose Hexokinase FS for the detailed
information about the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board of the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG
Alte Strasse 9, 65558 Holzheim, Germany.
OLYMPUS AU 5200
Chemistry setting Temperature : 37°C
Test parameters
Item code Iron
Function select No. #
1. Operation Yes 2. Sample volume 15
3. Sample volume at repeat run 3 4. Reagent volume R1 280 Reagent volume R2 70 5. W 3 parameter
Prime (bottle to valve) R1 Yes Prime (bottle to valve) R2 Yes Dispense level (valve to cuvette) R 1 100 Dispense level (valve to cuvette) R 2 100
6 Method End 7 Wavelength 1 600 Wavelength 2 800 8 Reaction slop +
9 Measuring point start 2 Measuring point end 16 10 O.D. value range max 2.000 O.D. value range min 0.000
11 Limit of Linearity % - 12 Normal value H 168 Normal value L 23 13 Dynamic range H 1000
Dynamic range L 0 14 Panic value H # Panic value L # 15 Reagent OD range
First Point H 2.000 First Point L -2.000 Last Point H 2.000 Last Point L -2.000
16 Decision level for repeat run H 9999 Decision level for repeat run L 0.0 17 R. dispence by black rack 0
18 R. dispence by black interval 0
Calibration parameters
Calibration type AB ODO – Conc. 0.000 Cal. No. {x}> #
Conc. > # Factor range H Factor range L Factor *
Correlation correction
Correlation coefficient A Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
IRON FS
Order information
Cat. No. Kit size 10 190 021 R1 4 x 20 ml + R2 1 x 20 ml + 1 x 3ml Std
1 1901 99 10 026 R1 5 x 80 ml + R2 1 x 100 ml
10 190 023 R1 1 x 800 ml + R2 1 x 200 ml
10 190 030 6 x 3 ml standard
10 190 717 R1 5x 80 ml + R2 5 x 20 ml
Notes
1. Please refer to the package insert for Iron FS for the
detailed information about the test on the following:
Clinical Relevance Method and Principle Composition and Stability of the Reagents
Specimens Calibrators and Controls Performance Characteristics concerning; Measuring Range
Specificity/Interferences Sensitivity/Limit of Detection
Precision (Reproducibility, Repeatability) Method Comparison
Reference Ranges Literature
2. The stability of the reagent on board of the analyser is at least one month provided that contamination and evaporation are avoided.
3. Manufactured by DiaSys Diagnostic Systems GmbH & Co.KG Alte Strasse 9, 65558 Holzheim, Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code LDH
Function select No.
1. Operation Yes
2. Sample volume 5
3. Sample volume at repeat run 2
4. Reagent volume R1 200
Reagent volume R2 50
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method Rate
7 Wavelength 1 340
Wavelength 2 380
8 Reaction slop -
9 Measuring point start 9
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % 15
12 Normal value H 480
Normal value L 0
13 Dynamic range H 2000
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type MB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor **
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
**) Factor must be checked by calibration
serum
LDH FS DGKC
Order information
Cat. No. Kit size
10 420 021 R1 5 x 20 ml + R2 1 x 25 ml
10 420 022 R1 5 x 80 ml + R2 1 x 100 ml
10 420 023 R1 1 x 800 ml + R2 1 x 200 ml
Notes
1. Please refer to the package insert for LDH
FS DGKC for the detailed information about
the test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by DiaSys Diagnostic Systems GmbH
& Co.KG Alte Strasse 9, 65558 Holzheim,
Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Phos
Function select No. #
1. Operation Yes
2. Sample volume 3
3. Sample volume at repeat run 1
4. Reagent volume R1 250
Reagent volume R2
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
6 Method end
7 Wavelength 1 340
Wavelength 2 380
8 Reaction slop +
9 Measuring point start 0
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % -
12 Normal value H 5.0
Normal value L 2.5
13 Dynamic range H 20.0
Dynamic range L 0.0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H -
Factor range L -
Factor *
Correlation correction
Correlation coefficient A -
Correlation coefficient B -
#) Data entry by the user
*) Calculated by the analyzer
PHOSPHORUS UV FS
Order information
Cat. No. Kit size
10 520 021 R 5 x 25 ml + 1 x 3 ml Standard
1 5200 99 10 026 R 6 x 100 ml
10 520 023 R 1 x 1000 ml
10 520 030 6 x 3 ml Standard
10 520 717 R 6 x 100 ml
Notes
1. Please refer to the package insert for
Phosphorus UV FS for the detailed
information about the test on the
following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board the
analyser is at least one month provided
that contamination and evaporation are
avoided.
3. Manufactured by DiaSys Diagnostic Systems
GmbH & Co.KG Alte Strasse 9, 65558 Holzheim,
Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Trig
Function select No. #
1. Operation Yes
2. Sample volume 3
3. Sample volume at repeat run 2
4. Reagent volume R1 250
Reagent volume R2
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
6 Method end
7 Wavelength 1 520
Wavelength 2 660
8 Reaction slop +
9 Measuring point start 0
Measuring point end 16
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % -
12 Normal value H 200
Normal value L 0
13 Dynamic range H 1000
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H -
Factor range L -
Factor *
Correlation correction
Correlation coefficient A -
Correlation coefficient B -
#) Data entry by the user
*) Calculated by the analyzer
TRIGLYCERIDES FS
Order information
Cat. No. Kit size
10 571 021 R 5 x 25 ml + 1 x 3 ml Std
1 5710 99 10 026 R 6 x 100 ml
10 571 023 R 1 x 1000 ml
10 571 030 6 x 3 ml Standard
Notes
1. Please refer to the package insert for
Triglycerides FS for the detailed
information about the test on the
following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board the
analyser is at least one month provided
that contamination and evaporation are
avoided.
3. Manufactured by DiaSys Diagnostic Systems
GmbH & Co.KG Alte Strasse 9, 65558 Holzheim,
Germany.
OLYMPUS AU 5200
Chemistry setting
Temperature : 37°C
Test parameters
Item code Urea
Function select No. #
1. Operation Yes
2. Sample volume 3
3. Sample volume at repeat run 2
4. Reagent volume R1 240
Reagent volume R2 60
5. W 3 parameter
Prime (bottle to valve) R1 Yes
Prime (bottle to valve) R2 Yes
Dispense level (valve to
cuvette) R 1
100
Dispense level (valve to
cuvette) R 2
100
6 Method Rate
7 Wavelength 1 340
Wavelength 2 380
8 Reaction slop -
9 Measuring point start 9
Measuring point end 12
10 O.D. value range max 2.000
O.D. value range min 0.000
11 Limit of Linearity % 25
12 Normal value H 43
Normal value L 17
13 Dynamic range H 300
Dynamic range L 0
14 Panic value H #
Panic value L #
15 Reagent OD range
First Point H 2.000
First Point L -2.000
Last Point H 2.000
Last Point L -2.000
16 Decision level for repeat
run H
9999
Decision level for repeat
run L
0.0
17 R. dispence by black rack 0
18 R. dispence by black
interval
0
Calibration parameters
Calibration type AB
ODO – Conc. 0.000
Cal. No. {x}> #
Conc. > #
Factor range H
Factor range L
Factor *
Correlation correction
Correlation coefficient A
Correlation coefficient B
#) Data entry by the user
*) Calculated by the analyzer
UREA FS
Order information
Cat. No. Kit size
10 310 021 R1 4 x 20 ml + R2 1 x 20 ml
+ 1 x 3ml Std
1 3101 99 10 026 R1 5 x 80 ml + R2 1 x 100 ml
10 310 023 R1 1 x 800 ml + R2 1 x 200 ml
10 310 030 6 x 3 ml standard
Notes
1. Please refer to the package insert for Urea
FS for the detailed information about the
test on the following:
Clinical Relevance
Method and Principle
Composition and Stability of the Reagents
Specimens
Calibrators and Controls
Performance Characteristics concerning;
Measuring Range
Specificity/Interferences
Sensitivity/Limit of Detection
Precision (Reproducibility,
Repeatability)
Method Comparison
Reference Ranges
Literature
2. The stability of the reagent on board of the
analyser is at least one month provided that
contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH & Co.KG
Alte Strasse 9, 65558 Holzheim, Germany.
nUric acid Fs*
TBHBA
Order information Cat. No. Kit size 10 302 021 R1 4 x 20 ml + R2 1 x 20 ml + 1 x 3 ml std. 1 3021 99 10 026 R1 5 x 80 ml + R2 1 x 100 ml 10 302 023 R1 1 x 800
ml + R2 1 x 200 ml
10 300 030 6 x 3 ml standard
Method Enzymatic colorimetric test, “TBHBA“
Uric acid is oxidized to Allantoine by Uricase. The generated Hydrogen Peroxide reacts with 4-Aminoantipyrine and 2,4,6-Tribromo-3-hydroxybenzoic acid (TBHBA) to Chinonimine.
Reagent preparation and stability The reagents and the standard are ready-to-use and stable up to the end of the indicated month of expiry, if contamination is avoided and stored at 2 – 8 °C. The reagents must be protected from light.
Specimen
Serum, heparinized or EDTA plasma, urine. For serum / plasma: Avoid hemolysis! Dilute urine 1 + 10 with distilled water.
Components and concentration in the test
R1: Phosphate buffer pH 7.0 100 mmol/l TBHBA (2,4,6-Tribromo-3-Hydroxy-
benzoic acid) 1 mmol/l
R2: Phosphate buffer pH 7.0 100 mmol/l 4-Aminoantipyrine 0,3 mmol/l Potassiumhexacyanoferrate (II) 10 µmol/l POD > 2 KU/l Uricase > 30 U/l Standard: 6 mg/dl (357 µmol/l)
Notes
Ascorbic acid interferes.
Normal range (see reference 1,2)
Serum: Men: 3.4 - 7.0 mg/dl (200 - 420 µmol/l) Women: 2.4 - 5.7 mg/dl (140 - 340 µmol/l) Urine: 250 - 750 mg/24 h (1.5 - 4.5 mmol/24 h)
References
1. Thefeld, W., Hoffmeister, H., Busch, E.-W., Koller, P. U., Vollmar, J., Dtsch. Med. Wschr., 98, (1973), 380 – 384
2. Henry, R. J., Cannon, D. C., Winkelmann, J.: Clinical Chemistry, Principles and Technics, 2
nd edition, Harper & Row Publishers
Inc., Hagerstown Maryland 1974, 534 3. Trinder, P., Ann. Clin. Biochem., 6, (1969), 24 - 27 4. Town, M.-H., Gehm, S., Hammer, B., Ziegenhorn, J.,
J. Clin. Chem. Clin. Biochem., 23, (1985), 591 * fluid stable
OLYMPUS AU 5200
Chemistry setting Temperature : 37°C
Test parameters
Item code
Function select No.
1. Operation yes 2. Sample volume 5 3. Sample volume at repeat run 3 4. Reagent volume R1 240 Reagent volume R2 60 5. W 3 parameter Prime (bottle to valve) R1 yes Prime (bottle to valve) R2 yes Dispense level (valve to cuvette) R 1 100 Dispense level (valve to cuvette) R 2 100 6. Method end 7. Wavelength 1 520 Wavelength 2 800 8. Reaction slope + 9. Measuring point start 6 Measuring point end 16 10. O.D. value range max 2.000 O.D. value range min 0.000 11. Limit of Linearity % - 12. Normal value H 7.0 Normal value L 2.4 13. Dynamic range H 20.0 Dynamic range L 0.0 14. Panic value H # Panic value L # 15. Reagent OD range First Point H 2.000 First Point L -2.000 Last Point H 2.000 Last Point L -2.000 16. Decision level for repeat run H 9999 Decision level for repeat run L 0.0 17. R. dispence by black rack 0 18. R. dispence by black interval 0
Calibration parameters
Calibration type AB ODO – Conc. 0.000 Cal. No. {x}> # Conc. > * Factor range H - Factor range L - Factor *
Correlation correction
Correlation coefficient A Correlation coefficient B
#) Data entry by the user *) Calculated by the analyzer
11.02.11
March 2010/3
OLYMPUS AU 5400 Specific Test Parameters General
Test Name CA Type Serum Operation Yes
Sample Volume 2.0 µl Dilution 0 µl
Reagents R1 Volume 160 µl Dilution 0 µl
R2 Volume 40 µl Dilution 0 µl
Wavelength Pri 660 Sec 800
Method END
Reaction Slope -
Measuring point 1 First 0 Last 27
Measuring point 2 First 0 Last 10
Linearity %
No-Lag-Time
Pre-dilution Rate
Min OD Max OD
Reagent OD Limit
First L -0.100 First H 2.50
Last L -0.100 Last H 2.50
Dynamic Range
L 0.0 H 25.0
Correlation Factor
A 1 B 0
On-board stability period 30 Days
Value/Flag # Level L # Level H #
Normal Ranges Age L Age H Sex Year Month Year Month
1 # # # # #
2 # # # # #
3 # # # # #
4 # # # # #
5 # # # # #
6 # # # # #
7 # # # # #
L H
Panic Value # # Unit mg/dl
Calibration Type AB Formula Y=AX+B Counts #
Factor Factor Cal.No OD CONC OD-L OD-H Point 1 * * Point 2 * * Point 3 * *
Point 4 * * Point 5 * * Point 6 * * Point 7 * *
1-Point Cal. Point
MB Type Factor Calibration Stability Period
#) Data entry by the user *) Enter calibration or standard value and position
CALCIUM P FS Order information
Cat. No. 1 1181 .. .. …
Notes
1. Please refer to the package insert for Calcium P FS for
detailed information about the test on the following:
Clinical Relevance Method and Principle Composition and Stability of the Reagents Specimens Calibrators and Controls Performance Characteristics regarding
- Measuring Range - Specificity/Interferences - Sensitivity/Limit of Detection - Precision (Reproducibility, Repeatability) - Method Comparison
Reference Ranges Literature
2. The stability of the reagent on board the analyser is 7 weeks
provided that contamination and evaporation are avoided.
3. Manufactured by
DiaSys Diagnostic Systems GmbH Alte Strasse 9, 65558 Holzheim, Germany
**This application proposal is for guidelines only. To avoid
misinterpretation measured results have to be validated and assessed with caution.
845 1181 10 02 25